Evaluation of Toxoplasma gondii Immunoglobulin G (IgG) and IgM Assays Incorporating the New Vidia Analyzer System

Adriana Calderaro 1,*, Giovanna Piccolo 1, Simona Peruzzi 1, Chiara Gorrini 1, Carlo Chezzi 1, Giuseppe Dettori 1
PMCID: PMC2446631  PMID: 18480234

Abstract

The new Vidia system is a fully automated system based on chemiluminescence and antigen bound to magnetic microparticles, which allows a fast measurement of Toxoplasma gondii-specific immunoglobulin G (IgG) and IgM levels. The analytical performances of the Vidia Toxo IgG and IgM assays were compared with those of the automated Vidas, AxSYM, and Liaison Toxo IgG and IgM assays. The comparative evaluation was performed utilizing 204 frozen sera belonging to 166 subjects and 201 fresh sera collected from 198 subjects. For the Vidia Toxo IgG system, the sensitivities were 100% in both the retrospective and prospective studies, and specificities were 98.39% in the retrospective study and 100% in the prospective study, respectively. The sensitivities of the other three Toxo IgG assays were 100%, and the specificities ranged from 96.77% to 100%. For the Vidia Toxo IgM assay, the sensitivity and specificity were 100% in both the retrospective and prospective studies. The overall sensitivities and specificities of the other three Toxo IgM assays ranged from 80% to 100% and from 99.44% to 100%, respectively. In our study, the Vidia system revealed excellent sensitivity (100% for both IgG and IgM assays) and good specificity (99.25% for IgG and 100% for IgM assays).


Toxoplasma gondii is an obligate intracellular protozoan responsible for common parasitic infections throughout the world in a wide range of hosts including humans (2, 6, 8).

In immunocompetent subjects, acute infection is usually asymptomatic or characterized by mild, nonspecific clinical symptoms (e.g., cold or light case of the flu), and spontaneous recovery is the rule (8). However, in pregnant women, acute infection by T. gondii may result in congenital disease, causing abortion or severe damages to the fetus at birth or later in life (1), and/or sequelae that may be prevented or reduced by early treatment (10, 13). Toxoplasmic chorioretinitis can be seen in congenitally or perinatally acquired disease as a result of acute infection or reactivation (8).

In immunocompromised patients such as patients infected by human immunodeficiency virus and bone marrow or organ transplant recipients, toxoplasmosis almost always happens as a result of a reactivation of a chronic infection, and it can be a life-threatening condition presenting with encephalitis, chorioretinitis, pneumonitis, or multiorgan involvement (8).

During infection, IgM antibodies may appear earlier and decline more rapidly than IgG antibodies and are frequently the first class of antibodies detected after primary infection (7). Moreover, they may persist for months or years after infection, and only negative results in the vast majority of cases exclude acute infection (1).

The diagnosis of recently acquired toxoplasmosis is usually based on the detection of specific IgM antibodies, seroconversion, or a fourfold or greater rise in the titer of T. gondii-specific IgG antibodies (12). As seroconversion and a rise in IgG titers are seldom demonstrable, the detection of T. gondii-specific IgM antibodies has been the most frequently used serological marker for diagnosing acute infection. However, the interpretation of serological tests is complicated by the prevalence of high IgG titers among healthy individuals and the persistence of specific IgM in some people (12).

Thus, the serological diagnosis of toxoplasmosis is very complex and has been discussed extensively in the literature (8, 11); diagnosis is especially difficult for immunocompromised individuals, who have a reduced or absent production of antibodies, and pregnant women.

The absence of IgG antibodies before or early in pregnancy allows the identification of women at risk of acquiring infection, and the presence of IgG allows the identification of immunocompromised patients at risk for the reactivation of a latent infection.

Many serological tests for the detection of T. gondii-specific immunoglobulin are commercially available, and the enzyme-linked immunosorbent assay is one of the easiest tests to perform (3, 5, 7, 9). Originally, immunoassays were performed in a microtiter plate format using an enzyme-based color detection method, but more recently, they have been made available for autoanalyzers that use a range of detection methods including chemiluminescence and immunofluorescence.

The purpose of this study was to evaluate the analytical performance of the new automated Vidia Toxo IgG and IgM (bioMérieux, Marcy l'Etoile, France) immunoassays and compare the results with those of the Vidas Toxo IgG and IgM (bioMérieux), AxSYM Toxo IgG and IgM (Abbott Laboratories, Abbott Park, IL), and Liaison Toxo IgG and IgM (DiaSorin, Saluggia, Italy) assays.

MATERIALS AND METHODS

Samples. (i) Retrospective study for Toxoplasma IgG and IgM.

Frozen aliquots from 204 sera from 166 subjects (median age of 39 years, ranging from 1 month to 85 years) sent to the Section of Microbiology of the University Hospital of Parma (Italy) for the serological diagnosis of toxoplasmosis during 2005 and 2006 were analyzed in the retrospective study. Before testing, samples were thawed and clarified by centrifugation at 2,000 × g for 10 min. Thawed samples were kept at 4°C until use (no more than 5 days) before refreezing. In particular, the 204 samples were collected from 104 pregnant women, 55 immunocompromised patients, 27 children/infants, 11 blood donors, 6 pregnant human immunodeficiency virus-infected women, and 1 healthy adult.

(ii) Prospective study for Toxoplasma IgG and IgM.

From July to September 2006, a total of 201 samples from 198 subjects (median age of 43 years, ranging from 3 months to 86 years) prospectively obtained from the laboratory of diagnostic Toxoplasma IgG and IgM testing were stored at 4°C until use (no more than 5 days) and then frozen at −20°C. In particular, the 201 samples were collected from 139 pregnant women, 38 healthy adults, 16 immunocompromised patients, 4 children, and 4 blood donors.

Tests.

Each sample from both retrospective and prospective studies was tested in parallel or within 24 h at the same time by the four fully automated enzyme immunoassays. Testing was carried out according to the manufacturers’ instructions.

The Vidas, AxSYM, and Liaison systems were described elsewhere previously (3, 7, 9).

The Vidas system, which is routinely utilized in our laboratory, uses a solid-phase receptacle coated with membrane and cytoplasmic Toxoplasma antigen (RH Sabin strain) and an anti-human IgG or IgM conjugate and 4-methylumbelliferryl phosphate as a substrate (7). For the Vidas Toxo IgG and IgM tests, positive results are defined as value of ≥8 international units (IU)/ml and index values of ≥0.65, equivocal results range from 4 to 8 IU/ml and from index values of 0.55 to 0.65, and negative results are defined as <4 IU/ml and index values of <0.55, respectively.

The AxSYM system uses T. gondii-coated microparticles as the solid phase for IgG detection and microparticles coated with IgG to human IgM for IgM detection; bound T. gondii IgG or IgM is detected with anti-human IgG or IgM conjugate and 4-methillumbellilferyl phosphate as a substrate (3). For AxSYM Toxo IgG and IgM tests, positive results are defined as values of ≥3 IU/ml and index values of ≥0.800, equivocal results range from 2 to 3 IU/ml and from index values of 0.600 to 0.799, and negative results are defined as <2 IU/ml and index values of <0.600, respectively.

The Liaison immunoassay uses magnetic microparticles coated with inactivated T. gondii (RH strain) and monoclonal antibodies labeled with an isoluminol derivative (9). For Liaison Toxo IgG and IgM tests, positive results are defined as a value of ≥8 IU/ml and an index value of ≥0.800 index, equivocal results range from 6 to 8 IU/ml and from index values of 0.600 to 0.799, and negative results are defined as <6 IU/ml and index values of <0.600, respectively.

The Vidia system combines a two-step enzyme immunoassay method with paramagnetic microparticles and a final chemiluminescence detection step. The Vidia system has not previously been described fully.

In brief, in the Vidia Toxo IgG assay, 10 μl of sample is used: specific IgG in the sample is bound to the beads, unbound material is removed by washing, and an alkaline phosphatase-labeled monoclonal anti-human IgG antibody is added. After incubation, a wash step removes the unbound conjugate. During the final detection step, the substrate is transformed into a luminescent product by the conjugate alkaline phosphatase. The intensity of the luminescence increases according to the quantity of Toxoplasma IgG in the sample. The instrument measures the relative luminescence value. Results are automatically analyzed and calculated by the instrument using the factory master calibration curve data, which are stored in the instrument and are expressed in IU/ml. Depending on the test value, positive results are defined as a value of ≥5 IU/ml, equivocal results range from 3 to 5 IU/ml, and negative results are defined as a value of <3 IU/ml.

In the Vidia Toxo IgM assay, 10 μl of sample is used: Toxoplasma IgM in the sample is bound to the beads, unbound material is removed by washing, and an inactivated Toxo antigen is added, which is itself revealed by an alkaline phosphatase-labeled anti-Toxo monoclonal antibody. After incubation, a wash step removes the unbound conjugate. During the final detection step, the substrate is transformed into a luminescent product by the conjugate alkaline phosphatase. The intensity of the luminescence increases according to the quantity of Toxoplasma IgM in the sample and is measured as the relative luminescence value. Results are automatically calculated by the instrument using the factory master data, which are stored in the instrument, and are expressed as an index. No international standard is available for testing of Toxoplasma IgM. The positive threshold is established and maintained using collection sera with known statuses. Positive results are defined as index values of ≥1.00, equivocal results range from index values of 0.68 to 1.00, and negative results are defined as index values of <0.68.

Analysis of results.

The samples for which at least one equivocal result has been found with any of the methods used were excluded from the analysis.

The “status” of samples (positive or negative) was defined on the basis of concordant results obtained by at least three of the four assays compared in this study.

Discrepancies were resolved by considering individual clinical data, when available, and/or by repetition of the test.

The analytical sensitivity and specificity of each assay were determined by expressing the results obtained as a ratio of samples with appropriately assigned positive or negative status, respectively. The accuracy of each assay was calculated based on the percentage of correct results obtained by the test under evaluation.

RESULTS

The sensitivity and specificity of the Vidia Toxo IgG and IgM tests were determined using all sera, excluding results obtained from 29 (7.1%) sera for IgG and 16 (3.9%) sera for IgM, for which at least one equivocal result or not-resolved status was observed (Table 1).

TABLE 1.

Details of equivocal results or nonresolved samples

Study No. of equivocal and/or nonresolved samples No. of samples with equivocal results for:
Vidia Vidas AxSYM Liaison
Retrospective study
    Toxoplasma IgG 20 8 13 5 5
    Toxoplasma IgM 12 9 6 4 0
Prospective study
    Toxoplasma IgG 9 6 3 2 0
    Toxoplasma IgM 4 2 2 2 0

Toxoplasma IgG.

Sensitivities were 100% for the Vidia Toxo IgG assay and for the other assays in both the retrospective and prospective studies. Specificities were 96.77% for the AxSYM assay (which gave four false-positive results), 98.39% for the Vidia Toxo IgG assay (which gave two false-positive results), and 100% for the Vidas and Liaison tests in the retrospective study. No false-positive results were resolved by repeating the discordant assays. In the prospective study, specificities were 100% for all assays (Table 2).

TABLE 2.

Toxoplasma IgG performance

Study No. of samples No. of equivocal and/or nonresolved samples No. of negative results No. of positive results No. of false-positive results No. of false-negative results % Sensitivity (95% confidence interval) % Specificity (95% confidence interval) % Accuracy (95% confidence interval)
Retrospective Toxoplasma IgG 204 20
    Statusa 124 60
    Vidia 122 60 2 0 100.0 (93.75-100.0) 98.39 (94.18-99.57) 98.91 (96.03-99.70)
    Vidas 124 60 0 0 100.0 (93.75-100.0) 100.0 (96.88-100.0) 100.0 (97.87-100.0)
    AxSYM 120 60 4 0 100.0 (93.75-100.0) 96.77 (91.86-98.76) 97.82 (94.45-99.16)
    Liaison 124 60 0 0 100.0 (93.75-100.0) 100.0 (96.88-100.0) 100.0 (97.87-100.0)
Prospective Toxoplasma IgG 201 9
    Statusa 141 51
    Vidia 141 51 0 0 100.0 (92.73-100.0) 100.0 (97.24-100.0) 100.0 (97.95-100.0)
    Vidas 141 51 0 0 100.0 (92.73-100.0) 100.0 (97.24-100.0) 100.0 (97.95-100.0)
    AxSYM 141 51 0 0 100.0 (92.73-100.0) 100.0 (97.24-100.0) 100.0 (97.95-100.0)
    Liaison 141 51 0 0 100.0 (92.73-100.0) 100.0 (97.24-100.0) 100.0 (97.95-100.0)
Total Toxoplasma IgG 405 29
    Statusa 265 111
    Vidia 263 111 2 0 100.0 (96.52-100.0) 99.25 (97.23-99.8) 99.46 (98.03-99.85)
    Vidas 265 111 0 0 100.0 (96.52-100.0) 100.0 (98.51-100.0) 100.0 (98.94-100.0)
    AxSYM 261 111 4 0 100.0 (96.52-100.0) 98.49 (96.12-99.42) 98.93 (97.24-99.59)
    Liaison 265 111 0 0 100.0 (96.52-100.0) 100.0 (98.51-100.0) 100.0 (98.94-100.0)
a

The “status” of samples (positive or negative) was defined on the basis of concordant results obtained by at least three of the four assays compared in this study.

Toxoplasma IgM.

In the retrospective study, the sensitivities were 100% for the Vidia Toxo IgM assay as well as for the other methods except for the AxSYM Toxoplasma IgM test, which had a sensitivity of 83.33% (2 false-negative results out of 10 positive samples). In the prospective study, the sensitivities were 100% for the Vidia and Vidas tests and 80% for the AxSYM and Liaison tests (one false-negative result out of four positive samples).

The specificities in the retrospective study were 100% for the Vidia Toxo IgM, Vidas Toxo IgM, and Liaison Toxo IgM tests. The AxSYM assay gave one false-positive result, with a specificity of 99.44%. In the prospective study, specificities were 100% for all the assays (Table 3).

TABLE 3.

Toxoplasma IgM performance

Study No. of samples No. of equivocal and/or nonresolved samples No. of negative results No. of positive results No. of false-positive results No. of false-negative results % Sensitivity (95% confidence interval) % Specificity (95% confidence interval) % Accuracy (95% confidence interval)
Retrospective Toxoplasma IgM study 204 12
    Statusa 180 12
    Vidia 180 12 0 0 100.0 (75.0-100.0) 100.0 (97.83-100.0) 100.0 (98.97-100.0)
    Vidas 180 12 0 0 100.0 (75.0-100.0) 100.0 (97.83-100.0) 100.0 (98.97-100.0)
    AxSYM 179 10 1 2 83.33 (54.59-95.41) 99.44 (96.83-99.9) 98.43 (95.42-99.47)
    Liaison 180 12 0 0 100.0 (75.0-100.0) 100.0 (97.83-100.0) 100.0 (98.97-100.0)
Prospective Toxoplasma IgM 201 4
    Statusa 192 5
    Vidia 192 5 0 0 100.0 (55.56-100.0) 100.0 (97.96-100.0) 100.0 (98.00-100.0)
    Vidas 192 5 0 0 100.0 (55.56-100.0) 100.0 (97.96-100.0) 100.0 (98.00-100.0)
    AxSYM 192 4 0 1 80.0 (36.85-96.48) 100.0 (97.96-100.0) 99.49 (97.10-99.91)
    Liaison 192 4 0 1 80.0 (36.85-96.48) 100.0 (97.96-100.0) 99.49 (97.10-99.91)
Total Toxoplasma IgM 405 16 0 0 0 0
    Statusa 0 0 372 17 0 0
    Vidia 0 0 372 17 0 0 100.0 (80.95-100.0) 100.0 (98.94-100.0) 100.0 (98.98-100.0)
    Vidas 0 0 372 17 0 0 100.0 (80.95-100.0) 100.0 (98.94-100.0) 100.0 (98.98-100.0)
    AxSYM 0 0 371 14 1 3 82.35 (58.45-93.93) 99.73 (98.45-99.95) 98.97 (97.34-99.60)
    Liaison 0 0 372 16 0 1 94.12 (72.45-98.98) 100.0 (98.94-100.0) 99.74 (98.51-99.95)
a

The “status” of samples (positive or negative) was defined on the basis of concordant results obtained by at least three of the four assays compared in this study.

The overall sensitivities considering the results of both retrospective and prospective studies were 100% for all the IgG assays, while they were 100% for the Vidia and Vidas IgM assays and 82.35% and 94.12% for the AxSYM and Liaison IgM assays, respectively.

DISCUSSION

The serological diagnosis of T. gondii infections usually does not present any interpretative problems for immunocompetent subjects, but there are four groups of individuals for which the diagnosis of toxoplasmosis is most critical: pregnant women who acquire the infection during gestation, fetuses and newborns who are congenitally infected, immunocompromised patients, and those with chorioretinitis (11).

In this study, the performances of four commercially available automated immunoassays for the detection of Toxoplasma IgG and IgM were evaluated using 405 collected serum samples, 204 frozen samples and 201 fresh samples, obtained by laboratory diagnostic Toxoplasma IgG and IgM testing.

For Toxoplasma IgG, the best overall accuracy was obtained with the Vidas and Liaison assays (100%). The Vidia and AxSYM assay accuracies were 99.46% and 98.93%, respectively. For Toxoplasma IgM, the best overall accuracy was obtained with the Vidia and Vidas assays (100%), followed by the Liaison (99.74%) and AxSYM (98.97%) assays.

Except for the Vidas assays, which gave no false-negative or false-positive results, Vidia assays reported no false-negative and the fewest false-positive results (two false-positive results for IgG) with respect to the AxSYM (five false positive for IgG and three false negative for IgM) and Liaison (one false negative for IgM) tests.

No previous data are available for Vidia assays for either IgG or IgM, whereas in a previous comparative study, the Liaison, AxSYM, and Vidas Toxo IgG and IgM assays were evaluated (9): the Liaison IgG assay showed agreements of 91% and 100% with the AxSYM IgG and Vidas IgG assays; the Liaison IgM assay showed agreements of 95% and 97% with the AxSYM IgM and Vidas IgM assays. The sensitivity and specificity of the Liaison system (combined IgG and IgM results compared to the Sabin-Feldman dye test, which measures total immunoglobulin) were 99.3% and 96.8%, respectively (9). The sensitivities of the Vidas assay were 97.3% for IgG and 70% for IgM, and specificities were 99.8% for IgG and 99.3% for IgM, as reported previously (4). In our study, for Toxoplasma IgG, all methods gave the best sensitivity (100%), and all the methods, except the Vidia (99.25%) and AxSYM (98.49%) tests, gave the best specificity (100%). For Toxoplasma IgM, all methods gave the best sensitivity (100%), except for the AxSYM (82.35%) and Liaison (94.12%) systems, and all methods gave the best specificity (100%), except for the AxSYM (99.73%) system.

In conclusion, in our study, the Vidia system revealed excellent sensitivity (100% for both IgG and IgM assays) and good specificity (99.25% for IgG and 100% for IgM assays).

Acknowledgments

This study was funded by bioMérieux, Marcy l'Etoile, France.

The findings and conclusions in this report are ours and do not necessarily represent the views of the funding manufacturer. We declare no conflicts of interest.

We thank the staff members of the Parasitology Unit of the University Hospital of Parma, Lucia Mazzani, and Maria Tortelli for excellent technical assistance.

Footnotes

Published ahead of print on 14 May 2008.

REFERENCES

  • 1.Alvarado-Esquivel, C., S. Sethi, K. Janitschke, H. Hahn, and O. Liesenfeld. 2002. Comparison of two commercially available avidity tests for Toxoplasma-specific IgG antibodies. Arch. Med. Res. 33:520-523. [DOI] [PubMed] [Google Scholar]
  • 2.Calderaro, A., G. Piccolo, C. Gorrini, S. Peruzzi, L. Zerbini, S. Bommezzadri, G. Dettori, and C. Chezzi. 2006. Comparison between two real-time PCR assays and a nested-PCR for the detection of Toxoplasma gondii. Acta Biomed. 77:75-80. [PubMed] [Google Scholar]
  • 3.Diepersloot, R. J., H. Dunnewold-Hoekstra, J. K. D. Hollander, and F. Vla. 2001. Antenatal screening for hepatitis B and antibodies to Toxoplasma gondii and rubella virus: evaluation of two commercial immunoassay systems. Clin. Diagn. Lab. Immunol. 8:785-787. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Hofgartner, W. T., S. R. Swanzy, R. M. Bacina, J., Condon, M. Gupta, P. E. Matlock, D. L. Bergeron, J. J. Plorde, and T. R. Fritsche. 1997. Detection of immunoglobulin G (IgG) and IgM antibodies to Toxoplasma gondii: evaluation of four commercial immunoassay systems. J. Clin. Microbiol. 35:3313-3315. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Jenum, P. A., and B. Stray-Pedersen. 1998. Development of specific immunoglobulins G, M, and A following primary Toxoplasma gondii infection in pregnant women. J. Clin. Microbiol. 36:2907-2913. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Kaul, R., P. Chen, and S. R. Binder. 2004. Detection of immunoglobulin M antibodies specific for Toxoplasma gondii with increased selectivity for recently acquired infections. J. Clin. Microbiol. 42:5705-5709. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Montoya, J. G. 2002. Laboratory diagnosis of Toxoplasma gondii infection and toxoplasmosis. J. Infect. Dis. 185:S73-S82. [DOI] [PubMed] [Google Scholar]
  • 8.Montoya, J. G., and O. Liesenfeld. 2004. Toxoplasmosis. Lancet 363:1965-1976. [DOI] [PubMed] [Google Scholar]
  • 9.Petersen, E., M. V. Borobio, E. Guy, O. Liesenfeld, V. Meroni, A. Naessens, E. Spranzi, and P. Thulliez. 2005. European multicenter study of the LIAISON automated diagnostic system for determination of Toxoplasma gondii-specific immunoglobulin G (IgG) and IgM and the IgG avidity index. J. Clin. Microbiol. 43:1570-1574. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Pinon, J. M., H. Dumon, C. Chemla, J. Franck, E. Petersen, M. Lebech, J. Zufferey, M.-H. Bessieres, P. Marty, R. Holliman, J. Johnson, V. Luyasu, B. Lecolier, E. Guy, D. H. M. Joynson, A. Decoster, G. Enders, H. Pelloux, and E. Candolfi. 2001. Strategy for diagnosis of congenital toxoplasmosis: evaluation of methods comparing mothers and newborns and standard methods for postnatal detection of immunoglobulin G, M, and A antibodies. J. Clin. Microbiol. 39:2267-2271. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Remington, J. S., P. Thuillez, and G. Montoya. 2004. Recent developments for diagnosis of toxoplasmosis. J. Clin. Microbiol. 42:941-945. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Suzuki, L. A., R. J. Rocha, and C. L. Rossi. 2001. Evaluation of serological markers for the immunodiagnosis of acute acquired toxoplasmosis. J. Med. Microbiol. 50:62-70. [DOI] [PubMed] [Google Scholar]
  • 13.Villena, I., D. Aubert, B. Leroux, D. Dupouy, M. Talmud, C. Chemla, T. Trenque, G. Schmit, C. Quereux, M Guenounou, M. Pluot, A. Bonhomme, and J. M. Pinon. 1998. Pyrimethamine-sulfadoxine treatment of congenital toxoplasmosis: follow-up of 78 cases between 1980 and 1997. Scand. J. Infect. Dis. 30:295-300. [DOI] [PubMed] [Google Scholar]

Articles from Clinical and Vaccine Immunology : CVI are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES