Correction

PMCID: PMC2467472  PMID: 18670604

Information was printed inaccurately in the 'How to’ article on applanation tonometry in Issue 64, December 2007. The corrections are shown here in bold.

Preparation

  • Ensure that the prism has been disinfected with isopropyl alcohol 70 per cent (methylated spirit) or sodium hypochlorite 1 per cent. The prism must be rinsed in sterile water and wiped dry with a clean swab (as residue of the disinfectant may cause a caustic burn on the cornea).

Method

When the applanation tonometry procedure has been completed:

  • Wipe the prism with a clean, dry swab and replace it in the receptacle containing the disinfectant.