Skip to main content
Clinical Microbiology Reviews logoLink to Clinical Microbiology Reviews
. 2008 Jul;21(3):403–425. doi: 10.1128/CMR.00014-08

Evidence-Based Biosafety: a Review of the Principles and Effectiveness of Microbiological Containment Measures

Tjeerd G Kimman 1,*, Eric Smit 2, Michèl R Klein 1
PMCID: PMC2493080  PMID: 18625678

Abstract

We examined the available evidence on the effectiveness of measures aimed at protecting humans and the environment against the risks of working with genetically modified microorganisms (GMOs) and with non-GMO pathogenic microorganisms. A few principles and methods underlie the current biosafety practice: risk assessment, biological containment, concentration and enclosure, exposure minimization, physical containment, and hazard minimization. Many of the current practices are based on experience and expert judgment. The effectiveness of biosafety measures may be evaluated at the level of single containment equipment items and procedures, at the level of the laboratory as a whole, or at the clinical-epidemiological level. Data on the containment effectiveness of equipment and laboratories are scarce and fragmented. Laboratory-acquired infections (LAIs) are therefore important for evaluating the effectiveness of biosafety. For the majority of LAIs there appears to be no direct cause, suggesting that failures of biosafety were not noticed or that containment may have been insufficient. The number of reported laboratory accidents associated with GMOs is substantially lower than that of those associated with non-GMOs. It is unknown to what extent specific measures contribute to the overall level of biosafety. We therefore recommend that the evidence base of biosafety practice be strengthened.

INTRODUCTION

Work with pathogenic microorganisms and genetically modified microorganisms (GMOs) requires precautions that guarantee the safety of humans and the environment, including laboratory personnel, patients treated with GMOs, and other persons who could be exposed to these microorganisms. During the past decades, responsible authorities and researchers have therefore developed regulations and guidelines that in some detail describe containment measures and working instructions (Fig. 1). For GMOs such regulations and guidelines appear to be largely derived from those developed for working with the natural, genetically unmodified pathogenic microorganisms from which these GMOs have been derived.

FIG. 1.

FIG. 1.

Context of biosafety measures. Based on a risk assessment, wild-type biological agents and GMOs are assigned to one of four risk categories. Work is subsequently performed under conditions that reflect increasing containment demands, i.e., BSL-1 to -4. Risks are contained by a set of measures employing biological and physical barriers and laboratory practices.

Despite containment measures and guidelines, laboratory infections, usually involving non-GMOs, occur more or less commonly, suggesting that biosafety rules are not always effective or complied with. The guidelines and instructions for working with GMOs appear to be largely effective, as there have been no major accidents with GMOs or with their unintended release. Nonetheless, despite such regulations and the lack of major accidents with GMOs, there appears to be continuing concern about the health and safety of individuals and the environment exposed to potentially hazardous GMOs (65). It has also been noted that the laws and regulations governing the biotechnology world are outdated, are not comprehensive, and span too many agencies (65). Indeed, while the natures of the risks and the measures to handle these risks are largely identical for GMOs and non-GMOs, in many countries there are different regulations for GMOs and non-GMOs. This may be because the latter were presumed to carry greater risks for causing ecological disturbances upon unintended release. For example, in The Netherlands, the Ministry of the Environment oversees work with GMO, while the Ministry of Social Affairs oversees work with human pathogens. The different regulations and overseeing authorities may be confusing to workers in the field. Moreover, it is unknown to what extent specific factors contribute to a safe biosafety practice. Thus, it is often unclear if and to what extent measures aimed at providing biosafety are based on documented evidence of their effectiveness.

A central question in this study is whether containment measures are effective and evidence based. One may argue that the evidence for the effectiveness of containment measures is at best indirect, i.e., based on the lack of many overt laboratory-acquired infections (LAIs). In addition, we could question whether the criteria to judge effectiveness are sufficiently developed. Indeed, the objectives of containment measures are often not explicitly defined, and without (quantifiable) objectives, evaluation of effectiveness is difficult. Furthermore, in finding evidence for the effectiveness of biosafety measures, it is important to judge the quality of the evidence. For comparison, in evidence-based medicine, systematic reviews, hypothesis-driven controlled laboratory experiments, and prospective studies provide a higher quality of evidence than case reports and expert opinion.

In this review we give a brief historical overview of the development of the current biosafety practice, we will try to identify which principles and methods appear to underlie it, and we will describe this current biosafety practice. We then present an approach for evaluating the effectiveness of biosafety measures to contain pathogenic microorganisms, and finally we summarize experimental and observational data on the effectiveness of containment measures. Our primary goal is to evaluate the evidence-based containment measures for GMOs. However, because such measures are largely based on containment measures for non-GMOs and because more data, although scarce, are available for non-GMOs, we also examine evidence-based measures to contain non-GMO pathogens. These data may be extrapolated to GMOs. We thereby hope to contribute to a conceptual framework that helps in further developing an evidence-based biosafety practice.

DEVELOPMENT OF CONTAINMENT MEASURES: BRIEF HISTORIC OVERVIEW

While Robert Koch had already developed some kind of biosafety cabinet (BSC), A. G. Wedum of the U.S. Biological Research Laboratories at Fort Detrick, MD, can be regarded as one of the pioneers in developing biosafety measures after the Second World War. He evaluated the risks of handling hazardous biological agents and developed practices, equipment, and facility safeguards for their control (113, 147, 148). Following his initial work, it is now regarded as conventional wisdom that enclosure and ventilation of contaminated work areas are important factors in eliminating LAIs. Besides safe microbiological techniques, primary barriers (safety equipment and personal protective equipment) and secondary barriers (facility safeguards) are now regarded as vital elements of containment measures.

It was recognized early that examining LAIs could be informative about the risks involved with laboratory work. Several comprehensive reviews of LAIs have therefore been compiled (116, 131). These early examinations recognized that the primary route of transmission of many of the causative agents was by aerosol, and they led to the development of laminar-flow BSCs. Legislation and guidelines that were introduced over the years have probably reduced but not eliminated the risk of occupational exposure to infectious agents.

With the growing ability to manipulate DNA in the mid 1970s, there was also growing concern about the potential hazards associated with recombinant DNA research and technology (9). GMOs could display the intended properties, but could also have unpredictable and undesirable features. There still appears to be debate about the health and safety of laboratory workers and animals, as well as the environment, exposed to potentially hazardous GMOs (65, 98). At the Asilomar Conference in 1975, general principles for dealing with potential biohazards related to GMOs were drafted. It was suggested that containment should be an essential consideration in the experimental design and that the effectiveness of the containment should match the estimated risk. Adjustment of the level of precaution to the level of risk would prevent infection without unduly impeding operations.

The first edition of the National Institutes of Health (NIH) guidelines for research involving DNA molecules appeared in 1976. Now, 3 decades later, there are a number of authoritative international guidelines, instructions, and recommendations for the safe handling and manipulation of hazardous biological agents, including GMOs. In 1984, the U.S. NIH and Centers for Disease Control and Prevention (CDC) produced the first edition of a guidebook, called Biosafety in Microbiological and Biomedical Laboratories, that is now considered a major reference text. The NIH/CDC and the WHO manuals are based on historical accounts of incidents with infectious microorganisms and extensive experience of experts working in this field, and they have been developed and improved over the last 30 years (152, 156). Legislation has been implemented, for example, in European Union and national regulations (145). In Europe, national authorities have based their regulations on directives from the European Union, such as the directive on the protection of workers from risks related to exposure to biological agents at work and the directive on the contained use of GMOs. A summary of important guidelines and manuals is given in Table 1.

TABLE 1.

Summary of guidelines and directives on recombinant DNA research

Institution Publication Yr issued (edition) Website or publication no.
CDC/NIH Biosafety in Microbiological and Biomedical Laboratories 2007 www.cdc.gov/od/ohs/biosfty/bmbl5/bmbl5toc.htm
EEC/EC Directive on the protection of workers from risks related to exposure to biological agents at work 1990 90/679/EEC, 89/391/EEC, 2000/54/EC
Directive on the contained use of genetically modified microorganisms 1990 90/219/EEC, 98/81/EC, 2000/608/EC
AS/NZS Safety in Laboratories (part 3, Microbiological Aspects and Containment Facilities) 2002 2243.3
PHAC Laboratory Biosafety Guidelines 2004 (3rd) www.phac-aspc.gc.ca/publicat/lbg-ldmbl-04/index.html
WHO Laboratory Biosafety Manual 2004 (3rd) www.who.int/csr/resources/publications/biosafety/WHO_CDS_CSR_LYO_2004_11/en
VROM Intergrale versie van de Regeling genetisch gemodificeerde organismen en Besluit genetisch gemodificeerde organismen (in Dutch) 2004 www.vrom.nl/docs/milieu/regelingGGO_inclusiefbijlagen_1okt2003.pdf

PRINCIPLES AND METHODS OF BIOSAFETY

In judging the effectiveness of measures intended to ensure biosafety, it may be helpful to know which goals, principles, and methods of biosafety measures have been employed and to evaluate the scientific basis of their effectiveness. Because the natures of risks are largely identical for GMOs and non-GMOs, containment measures to handle these risks are largely identical for both. In the past biosafety measures have evolved step by step, and usually based on expert knowledge and experience, but without a unifying set of guiding principles. Explicit guiding principles are therefore usually lacking in most legal regulations and scientific papers. We here attempt to draft a hierarchy of such guiding principles and methods that may be applicable to both GMOs and non-GMO, as follows.

  1. Risk assessment is the first and central step, and includes hazard recognition and identification, understanding of exposure potentials, frequency of occurrence, evaluation of work tasks and equipment, and assigning protective measures to the specific tasks involved.

  2. The second principle is biological containment; wherever possible, risks for the workers and the environment may be minimized by reducing exposure potentials and their consequences by using attenuated microorganisms (GMO or non-GMO) that have reduced replicative capacity, infectivity, transmissibility, and virulence.

  3. Concentration and enclosure, while simple and often overlooked, obviously constitute one of the most important ways to provide biosafety, by “locking up” microorganisms as much as possible, limiting the microbiological work load, and concentrating it in as few as possible work sites. For example, if work associated with a harmful microorganism can be restricted to one BSC, there will be less risk than if this work is done in two BSCs. Detection of microorganisms by nucleic acid multiplication carries less risk than detection by culture.

  4. The fourth principle is exposure minimization; when enclosure has to be disturbed, exposure can be minimized by a set of behaviors collectively known as safe microbiological techniques, including an orderly and disciplined work fashion, wearing gloves, wearing a mask and eye covering, prevention of aerosol and droplet generation, no mouth pipetting, prevention of skin disruptions, etc. These measures in particular provide operator protection (121). In practice they often make use of special equipment providing physical containment. Laboratory workers should be trained in these aspects.

  5. In physical containment, further protection of the operator and the environment is provided by physical barriers that prevent or minimize escape of microorganisms from the working place and laboratory. These include a range of design requirements, such as doors and locks, and physical safety equipment, such as BSCs, isolators, air filtration, wastewater management systems, etc. In addition to protecting the laboratory workers, these measures provide environmental protection. Obviously such physical containment is combined with and complemented by the previous methods. Physical barriers have been designated primary and secondary barriers. Primary containment measures minimize occupational exposure of laboratory workers and thereby limit transmission of microorganisms from these workers to others. The secondary barriers provide supplementary microbiological containment, serving mainly to prevent the escape of infectious agents when a failure in the primary barriers occurs.

  6. Hazard minimization includes a set of activities to reduce the consequences of exposure should it occur; these may include the availability of emergency procedures, a contingency plan, and health and medical surveillance, but also vaccination to reduce the consequences of inadvertent exposure (152).

In the following sections we further elaborate on some of these principles and methods and on the way in which they have been incorporated in legislation and regulations.

BIOSAFETY MEASURES

Risk Assessment

From a thorough risk assessment procedure, considering all potentially harmful effects for humans and the environment, follows the risk classification of microorganisms and subsequent containment measures that should be taken to manage these risks. Both the nature and scale of activities need to be considered to estimate the possibility of exposure of humans and the environment and the consequences of such exposure. Examples of risk assessment procedures can be found at http://www.hse.gov.uk/biosafety/gmo/acgm/ecrisk.htm and in European Union Directive 90/219/EEC.

As indicated below, there are four risk categories for hazardous biological agents and four containment levels. Work with non-GMO microorganisms is usually done at the corresponding containment level. It is evident that a risk assessment for working with noncharacterized pathogens in a clinical laboratory may involve uncertainties. Through manipulation GMOs may acquire pathogenic properties that are unexpected and/or not well characterized or understood, necessitating a higher containment level than for work with the natural microorganism from which the GMO has been derived, or additional measures (see, for example, reference 145). Therefore, the nature of recombinant DNA sequences, vectors, and recipient organisms needs to be carefully evaluated, as well as any potential biohazard associated with particular experimental settings. It is particularly important to address whether or not genetic modification affects cell tropism, host range, virulence, or susceptibility to antibiotics or other effective treatments. Some considerations regarding the risk assessment and categorization of GMO activities are given in Table 2.

TABLE 2.

Summary of considerations in the risk assessment for GMOs

Category Factor(s) to be considered
Recipient microorganisms Virulence, transmissibility, host range, susceptibility to antivirals or antibiotics, availability of prophylaxis, control and treatment
Vectors Replicative capacity, integration into host genome
Insert or donor sequences Toxicity, biological properties, replicative capacity, properties (known/unknown), gene-gene and gene-environment interactions
Activities Scale, animal experiments, transport
Host factors Immunodeficiencies
Population factors (Vaccine-derived) immunity

Biological Containment

A subsequent step in the selection of measures to ensure biosafety is to minimize biological hazards associated with the work by employing host microorganisms with a reduced host range, strains with natural or genetically modified characteristics that diminish their invading capacity or virulence, self-inactivating vectors, etc. Thus, employing biological containment is not restricted to work with GMOs. Natural pathogenic microorganisms may be replaced by less pathogenic microorganisms; for example, Mycobacterium tuberculosis may be replaced with nonvirulent mycobacteria, such as Mycobacterium smegmatis (27). Approaches for acquiring biological containment by genetic modification are given in Table 3.

TABLE 3.

Principles and methods of establishing biological containment

Principle Method Example(s)
Attenuation Natural or genetically modified deletion of virulence genes Modified vaccinia virus Ankara, herpesvirus vectors, E. coli K-12, Salmonella aro mutants, Vibrio ctx mutants, Lactococcus lactis thyA mutant
Host range restriction Natural host-restricted viruses Canarypox virus, fowlpox virus, baculovirus
Host range alteration Ecotropic packaging cell lines, pseudotyping Retroviral vectors
Use of replication-defective vectors Deletion and provision of essential gene products in trans Herpesvirus vectors, alphavirus vectors, retroviral vectors, adenovirus vectors, adeno-associated virus vectors, respiratory syncytial virus, lentivirus vectors, Salmonella aro mutants, E. coli K-12
Prevention of gene transfer Expression of suicide functions E. coli relF

Viruses.

Examples of biological containment include the use of highly modified vaccinia virus Ankara, which has a significantly reduced pathogenicity, and the use of avian poxviruses, including canarypox and fowlpox, that have a restricted host range and do not replicate in mammals (93, 108). These viruses have a better safety profile than the classical vaccinia virus.

Biological containment for retroviral vectors has been obtained by providing gene products that are required for the production of progeny viruses, i.e., gag, pol, and env, in trans by packaging cell lines that stably express these trans-acting functions. Formation of replication-competent human immunodeficiency virus (HIV) is excluded when the env gene is missing from the packaging plasmid. Furthermore, the range of host species that can be transduced with such vectors can be manipulated by using particular env genes or by pseudotyping with envelope proteins of other viruses (29, 39, 118). Various envelope proteins are associated not only with a varying host range but also with varying stability and intrinsic toxicity. A concern has been that recombination events may give rise to replication-competent lentiviruses, but further modifications have been made to reduce this opportunity. These include the use of different transcriptional units, further attenuation by deleting nonessential genes, and self-inactivating vectors (17, 29). Self-inactivating vectors contain an inactive long terminal repeat, resulting in a lack of promoter activity and preventing potential transcriptional activation of (onco)genes downstream of the integration site. These vectors cannot be rescued by wild-type HIV. No replication-competent lentiviruses have been detected when such systems are used (29, 37). Altogether, the risk associated with the newly developed lentivectors is therefore minimal, while they have distinct advantages such as stable integration in nondividing and dividing cells, long-term expression of the transgene, and absence of an immune response. However, although viral infection using state-of-the art lentivectors is very unlikely, their high transduction efficiency may increase the risk of accidental exposure of lab workers, leading to a positive anti-p24 HIV antibody response and accidental transduction of potentially hazardous genes. Contact transmission thus should still be avoided.

Similarly, replication-deficient adenoviruses have been developed using a viral DNA vector and a helper cell line that has been stably transfected with the E1 region (E1A and E1B) of the adenoviral genome. Vectors prepared from the cell line lack the E1 region and remain replication defective (72).

Safe replication-deficient herpesvirus vectors have been developed by deleting the viral glycoprotein gD. Envelope glycoprotein gD is essential for virus entry but is not required for subsequent steps in the viral replication cycle. Phenotypically complemented gD null mutants can infect cells and can spread, both in vitro and in vivo, by direct cell-to-cell transmission. However, progeny virions released by the infected cells are noninfectious because they lack gD (109). Thus, an accidental infection remains restricted to a single round of replication. The same principle has been used for other viruses, for example, respiratory syncytial virus. Other ways to generate safe herpesvirus vectors are deletion of genes that are essential for virulence in vivo but that are nonessential in vitro or deletion of immediate-early genes that activate early and late gene expression and subsequent propagation of the crippled virus in complementing cell lines (68, 75). Replication-incompetent herpesvirus vectors have also been generated by the use of cosmid DNAs to provide the necessary viral gene products for propagation of defective viruses or amplicons. These approaches reduce the opportunities for generation of replication-competent viruses through recombination.

Safe vectors have been developed from the nonpathogenic adeno-associated virus. The vector lacks all viral genes and requires coinfection with a helper adenovirus or a helper-free packaging system (24, 120).

A high level of biological containment is also acquired by using the Autographa californica nuclear polyhedrosis virus, a member of the baculovirus family. These viruses normally replicate in insect cells but not in mammalian cells. Furthermore, by deleting the nonessential polyhedron gene, the virus becomes noninfectious for its natural host (106).

Bacteria and protozoa.

The classical example of a biologically contained bacterium is Escherichia coli K-12. E. coli K-12 is a debilitated strain that does not normally colonize the human intestine. The strain survives poorly in the environment and has a history of safe commercial use. E. coli K-12 is considered an enfeebled organism as a result of being maintained in the laboratory environment for over 70 years (154). E. coli K-12 is defective in at least three cell wall characteristics. First, the outer membrane has a defective lipopolysaccharide core which affects the attachment of the O-antigen polysaccharide side chains (26). Second, it does not have the type of glycocalyx required for attachment to the mucosal surface of the human colon (S. Edberg, unpublished report, U.S. Environmental Protection Agency, 1991) as a result of the altered O-antigen properties noted above. Finally, K-12 strains do not appear to express capsular (K) antigens, which are heat-labile polysaccharides important for colonization and virulence. K-12 thus is not able to recognize and adhere to the mucosal surface of colonic cells (26). The normal flora in residence in the colon thus can easily exclude K-12. Furthermore K-12 lacks other virulence factors (26, 46; Edberg, unpublished report).

Several approaches have been used to develop safe bacterial vectors for use as vaccines or for gene or protein delivery, in particular, strains of Salmonella, Shigella, Listeria, Mycobacterium, Vibrio, and lactic acid bacteria (71, 137). When developing vector vaccines, the challenge is to develop strains that are well tolerated by the recipient host and no longer persist in the environment, yet still induce protective immune responses. This is not always achieved, however (32, 141). Elucidation of biosynthetic pathways has led to the development of Salmonella vectors that were attenuated by disruption of genes encoding metabolic functions or genes located in a pathogenicity island (Salmonella pathogenicity island 2, which encodes a type III secretion system). Salmonella pathogenicity island 2 is required for survival and growth within macrophages (67). For example, attenuating mutations in Salmonella strains included msbB, galE, via, rpoS+, aroCD, htrA, cya, crp, cdt, asd, phoPQ, purB, sifA, and ssaV (33, 67, 84, 90). The best-characterized live attenuated salmonellae have mutations in the prechorismate pathway. These are the so-called aro mutants, which are defective in the production of chorismate, which is essential in the synthesis of aromatic compounds (90). It is evident that strains carrying different mutations differ in properties such as invasiveness and survival. To reduce the possibility of reversion to virulence, strains carrying at least two attenuating distantly located mutations have been produced (67, 141). Salmonella enterica serovar Typhimurium VNP20009 was developed to deliver potential therapeutic proteins to tumor sites. It was created by chromosomal deletion of two genes, purI (purine biosynthesis) and msbB (lipopolysaccharide biosynthesis) and was attenuated at least 10,000-fold in mice compared with the parental wild-type strain (84).

A promising Listeria vector vaccine is an L. monocytogenes auxotrophic mutant with deletions in alanine racemase (daI) and d-amino acid aminotransferase (dat) genes, two genes required for the biosynthesis of bacterial cell walls. The strain was highly attenuated in mice (142). The strain requires d-alanine to grow and survive. Another L. monocytogenes candidate vaccine strain (LH1169) contained deletions in actA and plcB, genes that are necessary for cell-to-cell spread and escape from secondary vacuoles, respectively (2). Deleting lecithinase activity in L. monocytogenes also results in inhibited cell-to-cell spreading (31).

The virulence of Vibrio cholerae is due mainly to the expression of cholera toxin (CT). Hence, strategies for attenuating V. cholerae for use as an expression vector for heterologous antigens have been to engineer mutants in which the CT gene (ctx) or the CT genetic element has been partially or completely deleted (71).

Stable mutant strains of Bacillus licheniformis, an industrially exploited species, were obtained by introducing defined deletions in recA and/or an essential sporulation gene (spoIV). These strains are totally asporogenous and severely affected in DNA repair, and they therefore are UV hypersensitive. In liquid media these strains grow equally well as the wild type. Hence, such genes appear to be suitable disruption targets for achieving biological containment (96).

Construction of a genetically modified Lactococcus lactis strain for intestinal delivery of human interleukin-10 (IL-10) employed a biological containment system by replacing the thymidylate synthase gene thyA with a synthetic human IL-10 gene. When deprived of thymidine or thymine, the viability of the strain dropped considerably, preventing its survival in the environment. Transgene escape through acquisition of an intact thyA gene is very unlikely and would recombine the transgene out of the genome. The system was validated in vivo in pigs (136) and was used in a gene therapy study (13).

Further improvement in enhancing the safety profile of bacterial vectors for gene transfer can be achieved by removing undesirable properties of plasmids, such as a prokaryote origin of replication and antibiotic resistance markers. These elements could lead to dissemination of prokaryotic replicative recombinant DNA. Darquet et al. (28) therefore developed so-called minicircles, which are supercoiled DNA molecules that lack such elements and contain only an expression cassette carrying the gene of interest. Furthermore, efficient suicide functions have been developed to ensure biological containment of bacteria (69, 128). Such systems achieve their goals when the GMOs self-destruct by expression of killing genes after fulfilling their jobs. Suicide systems are based on “lethal genes” that are triggered by preprogrammed conditions. Such systems, however, appear to differ in efficiency of the suicide function, and the less efficient ones may lead to selection of mutants that have lost their suicide function. However, no system can provide complete efficiency. One efficient system was based on the lethal E. coli relF gene, which prevents the transfer of plasmids to wild-type bacteria (69).

To enhance the safety of genetically modified yeast (Saccharomyces cerevisiae), genes encoding bacterial toxins have been used for containment control. Expression of the E. coli relE toxin gene was highly toxic to yeast cells, and this could be counteracted by expression of the relB gene (73).

In conclusion, we believe that the mechanisms underlying biological containment are usually well examined and understood. Evidence for their effectiveness is usually available, although only seldom in quantitative terms of infectivity and transmission. It is also important to note that not all GMOs are by definition biologically contained. They have reduced transmissibility or virulence only upon disruption of virulence factors. For example, very virulent recombinant influenza viruses and herpesviruses have been regenerated (30a, 47).

Physical Containment

Categorization of microorganisms (non-GMOs).

Central to biosafety programs is the concept of universal precautions (16). For that purpose, microorganisms are categorized into four risk categories on the basis of a risk analysis. Subsequent risk containment is focused not on specific infectious agents but on standard practices for handling infectious material that will prevent the transmission of all pathogens of that risk category. It is important to note that the principles, guidelines, and recommendations are basically the same for natural pathogens and GMOs. However, as would be logical, this conclusion did not lead to a single set of guidelines and recommendations for GMOs and non-GMOs, and thus there is some redundancy in guidelines and regulations.

The four categories for biological agents are based on their relative risk to laboratory workers and the community. In general, the following factors are considered in classifying biological agents: (i) the virulence of the biological agent or the severity of disease (in humans) (18), (ii) the mode of transmission (spread in the community and host range), (iii) the availability of effective preventive measures (e.g., vaccines), and (iv) the availability of effective treatment (e.g., antibiotics or antiviral drugs). The hazard of the infectious (non-GMO) agent increases from risk group 1, consisting of microorganisms not associated with disease, to risk group 4. Risk group 4 microorganisms can cause serious disease and can be readily transmitted, and effective treatments are usually not available (123, 152, 156). However, there are differences in the exact definitions as used by certain countries and/or organizations (such as NIH/CDC, WHO, and European Union) (Table 4), which result in differences in the exact listings of biological agents in each risk category (43; http://www.absa.org/XriskgroupsX/index.html). The main difference between the NIH/CDC classification and the WHO classification is that the latter includes hazards to animals and the environment. Another difference is that for risk group 3, the NIH/CDC states that “… therapeutic interventions may be available,” whereas the WHO and European Union state that “… effective treatment and preventive measures are available.” There also are differences in the description of transmission properties between the different classifications. In assigning an agent to a risk group, one must take into account that there are in all groups of microorganisms naturally occurring strains that vary in virulence and that may thus need a higher or lower level of containment (43). In general, regulators deal with this concept by taking the highest level of virulence into account.

TABLE 4.

Classifications of infectious (non-GMO) agents into risk groups by NIH/CDC, WHO, and the European Uniona

Risk group Classification according to:
NIH/CDC WHO European Union
1 Agents that are not associated with disease in healthy adult humans A microorganism that is unlikely to cause human or animal disease A microorganism that is unlikely to cause human disease
2 Agents that are associated with humans disease which is rarely serious and for which preventive or therapeutic interventions are often available A pathogen that can cause human or animal disease but is unlikely to be a serious hazard to laboratory workers, the community, livestock, or the environment; laboratory exposures may cause serious infection, but effective treatment and preventive measures are available and the risk of spread of infection is limited A pathogen that can cause human disease but is unlikely to be a serious hazard to laboratory workers; the risk of spread of infection is limited, and effective treatment and preventive measures are available
3 Agents that are associated with serious or lethal human disease for which preventive or therapeutic interventions may be available A pathogen that causes serious human or animal disease but does not ordinarily spread from one infected individual to another; effective treatment and preventive measures are available A pathogen that causes serious human disease and poses a serious hazard to laboratory workers; it is likely to spread from one infected individual to another; effective treatment and preventive measures are available
4 Agents that are likely to cause serious or lethal human disease for which preventive or therapeutic interventions are not usually available A pathogen that usually causes serious human or animal disease and that can be readily transmitted from one individual to another, directly or indirectly; effective treatment and preventive measures are not usually available A pathogen that usually causes serious human disease and that can be readily transmitted in the population; effective treatment and preventive measures are not usually available
a

Based on data from references 19, 145, 152, and 156.

There has been debate about the classification of particular microorganisms, in particular the Flaviviridae, variola virus, avian influenza A/H5N1 virus, and extremely drug-resistant M. tuberculosis strains, especially as to whether they should be categorized as category 3 or 4 biological agents (1, 20, 25, 40, 70, 97, 143). Because vaccination was stopped in the 1970s, variola viruses are now classified as category 4 biological agents. Although avian influenza A/H5N1 virus strains initially were classified as risk category 4 biological agents, susceptibility to antiviral drugs and the availability of effective vaccines may downgrade them to category 3 for further studies. Extremely drug-resistant M. tuberculosis strains (25) should be regarded as risk category 4 biological agent. Of note, to date no microorganisms other than viruses have been classified in category 4.

Categorization of GMOs and definition of harmful gene products and microorganisms.

As for non-GMOs, GMOs are assigned to specific risk categories based on the risk assessement. For GMOs biosafety containment levels are assigned depending on the risk category of the donor organism, unless the modification may result in a higher or unknown risk. Therefore, the nature and function of insert sequences and the properties of acceptor microorganisms are considered. It is remarkable that definitions of properties of GMOs are sometimes quite strict (for example, the toxicities of vertebrate toxins, which are expressed as 50% lethal dose/body weight), while other properties of harmful gene products are not exactly defined. In the regulations, both the nature and level of virulence and transmissibility of GMOs are not well defined, at least not in quantitative terms (such as the basic reproduction ratio [R0] [14]). Sometimes host range is taken into account (for example, for baculovirus, ecotropic murine retroviruses, and papillomaviruses when they are used in nonpermissive host/vector systems). Properties may be unknown, such as the capacity of microbial DNA to integrate into the host genome (as, for example, for HIV) and the availability of vector organisms in the environment and therefore the possibility of persistence in the environment. A point for consideration is therefore the possibility of monitoring replication and survival of the GMO. In case of scientific uncertainty, the precaution principle is leading, resulting in higher categorization of the GMO or additional measures on a case-by-case base. Although the possibility of microbial transmission is taken into account, the possibility that transmission is reduced by herd immunity, either vaccine derived or not, is not mentioned explicitly by legislators. The assessment should include whether properties of inserted sequences can be expressed in the background of the host organism. Thus, gene-gene and gene-environment interactions should always be considered. As an example, IL-4 is not normally considered a harmful gene product, but when expressed by a murine ectromeliavirus, it drastically enhanced the virulence of this virus by inducing changes in cytotoxic T-cell function (58). Importantly, the regulations, being not completely detailed, largely function as a framework that must be elaborated by researchers and laboratory directors in their risk assessment, in the risk categorization of GMOs, and in the implementation of procedures and regulations in their laboratories (79). While the considerations in the risk assessment and subsequent risk categorization may show differences for GMOs and non-GMOs, many of the subsequent containment procedures are similar.

Laboratory design and primary and secondary containment.

Phillips and Runkle (113) describe some principles underlying laboratory design. They introduce two concepts in designing laboratories. The first is the concept of primary and secondary barriers as described above, and the second provides the designers a logical division of major functional zones within a laboratory building. (Note that although elements of primary and secondary barriers are clearly recognizable, they are not mentioned as such in the European and Dutch regulations.) Phillips and Runkle (113) identify five functional zones in the facility (clean and transition, research area, animal holding, laboratory support, and engineering support). Primary containment measures minimize occupational exposure of laboratory workers. In addition to strict adherence to good microbiological practice, the primary containment barrier include physical separation of the biohazardous agent from the laboratory worker using closed vessels, personal protective equipment (e.g., gloves or full-body suits) and additional equipment (e.g., BSCs, enclosed centrifuge containers, or pipetting aids). The secondary barriers provide supplementary containment, serving mainly to protect other facility employees and to prevent the escape of infectious agents from the laboratory if and when a failure occurs in the primary barriers. These provide a separation between potentially contaminated areas in the building and the outside community. These measures may comprise special procedures (e.g., validated decontamination methods, training of personnel, strictly controlled access zones, interlocked doors, etc.) and special engineering and facility design features (e.g., decontamination equipment, showers, autoclaves, dedicated air handling system with filters, etc).

Secondary containment is very strict in high-containment laboratories. The high-containment laboratories (biosafety level 3 [BSL-3] and BSL-4) are airtight and have airlocks and a unidirectional airflow so that potentially contaminated air is kept inside. Thus, BSL-3 and -4 laboratories need to be negatively pressured, resulting in an airflow from adjacent areas into the laboratory and a filtered exhaust airflow outside the building without recirculation. Ten to 12 air changes per hour have been recommended, which removes approximately 99% of airborne particles in 23 min (92). Release of air into the environment is possible only through HEPA (high-efficiency particulate air) filters (rated 99.99% efficient with particles 0.3 μm in diameter and larger) or ULPA (ultra-low-penetration air) filters (rated 99.999% efficient with particles 0.12 μm in diameter). Filters are used for BSCs, autoclaves, incinerators, chemical decontamination showers, etc. In addition, HEPA filters are used to provide clean air to laboratory workers in full-body suits.

Categorization of biosafety containment levels.

Biosafety containment levels have been categorized in a range from 1 to 4. As indicated above, the containment levels are assigned depending on the risk group of the microorganism (GMO or non-GMO) and the scale and nature of activities. Importantly, the BSL designations are based on a composite of the design features, construction, containment facilities, equipment, and operational procedures required for working with agents from the various risk groups (152). It is important to stress that although handling microorganisms of a certain risk group usually requires working at the accompanying BSL, a risk assessment should be made to take other specific factors into consideration. For example, particular experiments may generate high-concentration aerosols, requiring a higher degree of safety. Thus, professional evaluation, based on personal responsibility, should always guide the BSL for the specific work (152). Sewell (131) formulated some broad recommendations. BSL-1 is recommended for teaching activities with agents that are not associated with disease. BSL-2 practices are used in diagnostic laboratories that manipulate agents that are not transmitted via aerosols (e.g., hepatitis B virus [HBV], HIV, enteric pathogens, and staphylococci). BSL-3 is recommended when working with agents that are highly infectious and are transmitted via aerosols (e.g., M. tuberculosis, Brucella spp., and Coccidioides immitis) and for large-scale work with BSL-2 agents. BSL-4 practices are required when working with unusual agents that cause life-threatening infections for which no treatment is available.

Depending on country and/or regulation authority, there are differences between the exact requirements for each of the four containment levels, not to mention the sometimes confusing differences in nomenclature for the (high-)containment laboratories. Nulens and Voss (104) reviewed the basic practice, equipment, and facilities necessary for each of the BSL, based on European Union Directive 98/81 and the WHO biosafety manual. This information is summarized in a general way in Table 5. A comprehensive listing of all requirements and equipment necessary for the four BSLs is presented in Table 6. This list is adapted from the WHO biosafety manual (152).

TABLE 5.

BSLs, practices, and safety equipment necessary for the four levels of microbial containmenta

BSL Practices Safety equipment Facilities
1 Standard microbiological practices None required Open bench top and sinkb
2 BSL-1 practices plus biosafety warning, biosafety manual, waste decontamination, medical surveillance policy Class I or II BSC, laboratory coats, gloves, face protection; face mask optional BSL-1 plus autoclave
3 BSL-2 practices plus controlled access, decontamination of waste and laboratory clothing, baseline serum sample Class I or IIc BSC, protective clothing, gloves; respiratory protection optional BSL-2 plus physical separation from corridors, self-closing double doors, no air recirculation, negative airflow in laboratory, HEPA-filtered air outletc
4 BSL-3 practices plus clothing change before entering, shower on exit, all material decontaminated on exit Class III BSC or class I or II BSC in combination with full-body, air-supplied positive-pressure suits BSL-3 plus separated building or isolated zone, dedicated supply, vacuum, and decontamination systems plus air filtration
a

Adapted from reference 152 with permission of the publisher and based on EU Directive 98/81 and reference 104.

b

According to EU Directive 98/81, an autoclave is required for ML-I.

c

According to EU Directive 98/81.

TABLE 6.

Requirements for equipment and design for laboratories of different BSLsa

Design factor or equipment Requirement at BSL:
1 2 3 4
Isolation of laboratory No No Yes Yes
Rooms sealable for decontamination No No Yes Yes
Ventilation
    Inward airflow No Desirable Yes Yes
    Controlled ventilation system No Desirable Yes No
    HEPA filtered exhaust No No Yesb/noc Yes
Double-door entry No No Yes Yes
Airlock No No No Yes
Airlock with shower No No No Yes
Anteroom No No Yes
Anteroom with shower No No Yes/nod No
Effluent treatment No No Yes/nod Yes
Autoclave
    On site Yesb/no Yesb/desirable Yes Yes
    In laboratory No No Desirable Yes
    Double ended No No Desirable Yes
BSCs No Yesb/desirable Yes Yes
Personnel safety monitoring capability No No Desirable Yes
a

Adapted from reference 152 with permission of the publisher.

b

According to Dutch regulations (145).

c

Depending on location of exhaust.

d

Depending on agents used.

At BSL-1, safety is achieved mainly by applying good microbiological techniques. To achieve a higher degree of containment and thus a higher degree of protection against LAIs, the number of requirements increases up to the maximum containment at BSL-4. Besides equipment, there are several codes of practice for laboratory access, personal protection, and working procedures. Scientific evidence for the efficiency of these measures is scarce, and most of these measures have been developed based on a long history of microbiological practice and common sense. Table 7 describes the codes of practice regarding access and personal protection for BSL-1, -2, and -3 laboratories.

TABLE 7.

Codes of practice regarding access and personal protection for BSL-1, -2, and -3 laboratoriesa

BSL(s) Access Personal protection
1 and 2 The international biohazard sign must be displayed on doors of laboratories in which microorganisms of risk group 2 or higher are handled; only authorized personnel are allowed to enter the laboratories; laboratory doors should be kept closed; children under age 16 are not allowed to enter; access to animal houses should be specially authorized; animals not involved in the work are not allowed; signs prohibiting smoking, drinking, and eating should be displayed in and outside the laboratory (not required in some countries, including The Netherlands) Laboratory coveralls, gowns, or uniforms must be worn in the laboratory; appropriate gloves must be worn during all work involving contact with infected material, and after use gloves should removed aseptically and hands must be washed; personnel must wash their hands after handling infectious material or animals and before they leave the laboratory; safety glasses or face shields must worn when it is necessary to protect the eyes and face from splashes, impacting objects, or artificial UV radiation; it is prohibited to wear protective clothing outside the laboratory; open-toe footwear should not be worn; eating, drinking, applying cosmetics, and handling contact lenses in the laboratory are prohibited; storing human foods or drinks in the laboratory is prohibited; protective clothing should not be stored in the same lockers as street clothing
3 The international biohazard warning symbol and sign displayed on the laboratory access door must identify the biosafety level and the name of the laboratory supervisor who controls access and must indicate any special conditions for entry into the area (e.g., immunization) Laboratory protective clothing must be of the type with solid-front or wrap-around gowns, scrub suits, coveralls, head covering, and, where appropriate, shoe covers or dedicated shoes; front-buttoned standard laboratory coats are unsuitable, as are sleeves that do not fully cover the forearms; laboratory protective clothing must not be worn outside the laboratory, and it must be decontaminated before it is laundered; removal of street clothing and change into dedicated laboratory clothing may be warranted when working with certain agents (agricultural or zoonotic agents);open manipulations of all potentially infectious material must be conducted within a BSC or other primary containment device; respiratory protective equipment may be necessary for some laboratory procedures or when working with animals infected with certain pathogens
a

Based on data from reference 152. Note that for BSL-3 laboratories the codes of practice for BSL-1 and -2 apply as well, except when modified as described in the table.

In The Netherlands, BSL-1 to -4 are called MLI to -IV for work with GMOs. These are based on the European Union Directives 90/219 and 98/81 and are implemented in national regulations (145). Many technical requirements and access and personal protection rules are similar, but a major difference is that according to Dutch regulations, at the MLII level (BSL-2) a class II BSC is not optional but is required. Another difference is that the Dutch regulations provide more detail with respect to procedures. For instance, it is stated, “prepare your work carefully limiting the necessarily movement from one place to another during the microbiological work. Collect all material and equipment before you start and arrange them in an orderly fashion.”

The first BSL-4 high-containment laboratories were built in the 1970s. Until then, researchers had handled extremely hazardous biological agents in so-called glove boxes: hermetically sealed, transparent cabinets fitted with rubber gloves, compatible with a class III BSC. To day, most BSL-4 high-containment laboratories operate as suit laboratories, where researchers wear full-body positive-pressure (“space”) suits (156; European Union Directives 90/218/EC, 98/81/EC, and 2000/54/EC). Key features of BSL-4 high-containment laboratories are the safeguards to prevent failure and faults of containment systems and measures. There is thus redundancy of critical systems and biosafety procedures.

Although the objectives that the legislators want to reach by using containment levels and procedures are not always explicitly mentioned, one could deduce some guiding objectives, as we have summarized in Table 8.

TABLE 8.

Objectives of microbiological containment rulesa

Objective(s) Achievement at BSL:
1 2 3 4
Reduction of direct and oro-fecal transmission of nonpathogenic microorganisms to lab personnel, reduction of transmission of nonpathogenic microorganisms outside the laboratory, general hygiene + + +
Reduction oro-fecal transmission of enteric pathogens to lab personnel + +
Reduction of airborne transmission of pathogenic microorganisms to lab personnel + ++
Reduction of transmission of pathogenic microorganisms outside the laboratory by direct contact, environmental spread, and airborne spread + ++
Strict prevention of transmission of very virulent microorganisms to lab personnel +++
Strict prevention of transmission of very virulent microorganisms outside the laboratory, calamity procedures +++
a

This table specifies objectives that may be achieved by containment measures at BSL-1 to -4, as specified by EU 90/679 and VROM (145). At increasing safety levels, there are additional demands regarding equipment and procedures, while those of the lower levels are maintained. At BSL-1 equipment and procedures are directed at providing general hygiene, which probably mainly diminish infection by the oro-fecal route. At BSL-2 a BSC class II is optional or required (as in The Netherlands), providing mainly respiratory protection to the laboratory workers; environmental spread is diminished by keeping windows closed. At BSL-3 further environmental protection is provided by a sluice, disinfection facilities, and a unidirectional HEPA-filtered airflow. At BSL-4 all procedures and equipment are directed toward preventing any microbial transmission. +, adequate; ++, good; +++, excellent.

APPROACHES FOR BIOSAFETY EVALUATION

The scientific literature on evaluation of the effectiveness of biosafety measures is very scarce and does not provide a consensus approach. The effectiveness of biosafety measures may be evaluated by different approaches and at different levels. In Table 9 we have given a simple classification of approaches that collectively may provide guidance in evaluation activities. This table presents questions and purposes of single evaluation activities. A first level of evaluation may be directed at measuring the effectiveness of single containment equipment and procedures, such as the filtering capacity of face masks and BSCs under experimental circumstances. Such an evaluation could be directed at physical or, preferably, microbiological criteria. Subsequently these single apparatuses and procedures should be evaluated during practical work. It is evident that by taking this step from experimental challenge to practical work, unforeseen circumstances that may occur during practical work may be detected. For example, turbulences caused by movements of personnel during practical work may lower the protection afforded by BSCs. Air leakage may occur along respiratory masks during work. Masks may not fit perfectly.

TABLE 9.

Approaches for biosafety evaluationa

Category Exptl approach Practical conditions
Equipment and procedures Do single devices and procedures function effectively upon exptl challenge with particles or model microorganisms? Do single devices and procedures function effectively during practical work?
Laboratory Does the laboratory as a whole afford effective containment upon exptl challenge with particles or model microorganisms? Does the laboratory as a whole afford effective containment during practical work?
Laboratory workers and environment NA Are laboratory workers and the environment protected against infection?
a

Classification of approaches to evaluate containment measures, giving examples of the major questions and purposes of the evaluation steps. NA, not applicable.

A subsequent level of evaluation would be the laboratory as a whole, including its design and construction, the equipment, and working instructions. Again, such an evaluation can be done experimentally during a validation of the laboratory process or actually in a working laboratory setting. An experimental approach may use the deliberate release of indicator particles or model microorganisms. Evaluation of laboratory safety under field circumstances may include analysis of environmental samples taken inside and near the laboratory. In this case the effectiveness of biosafety measures and working instructions are actually evaluated during practical work and includes compliance of workers with working instructions, their experience and training, unintentional incidents, and efficacy of containment measures.

Finally, one may evaluate the effectiveness of measures at the clinical-epidemiological level, examining the overall effectiveness of measures in their capacity to prevent infection of laboratory workers and others. It is evident that laboratory workers play a central role in such an evaluation, as they are the persons who both are at high risk and may pass infections to others. Such epidemiological studies may follow a passive or an active searching approach. Clearly, for ethical reasons this level of evaluation is usually not suitable for an experimental approach.

Compliance with Procedures and Training

We consider that optimizing the training of personnel and monitoring their compliance with procedures are important, as the best biosafety measures are only as good as the participation and discipline of laboratory workers themselves. In many reports, extensive training of laboratory workers and the proper execution of guidelines are mentioned as one of the most important measures to prevent incidents (85). However, poor compliance has also been reported (45, 144). In Argentina, incidents with pathogenic microorganisms were reduced after setting up a training program and providing protocols (85). The importance of training was also emphasized by experiences during the outbreak of severe acute respiratory syndrome (SARS) in 2003 (81, 100). Laboratory escapes of the virus from BSL-3 laboratories occurred in Singapore, Taiwan, and Beijing because of breaches in good laboratory practice rather than failure of the facilities. Extensive contamination occurred because gloves were inappropriately worn and contaminated surfaces were not disinfected (81).

EXPERIMENTAL AND OBSERVATIONAL DATA ON THE EFFECTIVENESS OF CONTAINMENT MEASURES

Do Single Devices and Procedures Function Effectively?

BSCs.

Because most infections in the laboratory occur via aerosols and infected material and surfaces (123), equipment directed at minimizing airborne infections has received most attention. There are three classes of BSCs with different levels of protection, i.e., classes I, II, and III (152). In addition, within BSC II there are three subtypes. Class II BSCs consist of a chamber with a small open front in which an airflow is generated to prevent microorganisms from escaping the chamber. Laboratory workers sitting behind the cabinet insert their hands and arms into the chamber. All objects and the arms of the worker can disturb the airflow and cause the microorganisms to escape. Class III BSCs are basically similar, but the front is completely closed and can be accessed via attached gloves.

BSCs should meet legal standards, as, for example, defined by the European Union (EN12469). BSCs have been improved significantly during recent years due to this EN12469 standard, among others. Unfortunately, literature on the containment efficiency of class II BSCs is scarce and is addressed mainly in older publications, and such literature is virtually absent for class III BSCs.

In general BSCs provide a good level of protection when operated and maintained correctly (107). However, in several older studies, before the introduction of EN12469, it was shown that personnel working with open-front BSCs can still be exposed to infectious doses of microorganisms (7, 74, 112). Barbeito and Taylor (6) investigated the efficiency of containment of a BSC under three different closure conditions and different air velocities. In the cabinet, between 105 and 106 microorganisms per cubic foot were released in 5 min. When the glove panel was removed, a human infectious dose was released, and the number of microorganisms that escaped containment increased with decreased air velocity. Moreover, an increase in human activity in the cabinet resulted in increased numbers of microorganisms escaping the cabinet. When the glove panel was attached, no microorganisms could be detected outside the cabinet. A remarkable finding was that when the glove panel was installed without the gloves attached, no microorganisms escaped from the cabinet. Their main conclusions were that laboratory workers are protected from infectious microorganisms only when they use closed BSCs with high airflow velocities and that the effectiveness of BSCs is compromised by the activity of the workers. Macher and First (86) performed measurements on exposure of workers to bacterial spores using a class II BSC with an adjustable work opening. Aerosols of bacterial spores were created inside the cabinet, and the escaping spores were measured. The work opening height appeared to be a significant predictor of spore concentrations outside the BSC (86). Similarly, airflow velocity was negatively correlated with the concentration of escaping spores. Human activity in the cabinet, such as hands moving through the opening, also resulted in the escape of spores. Spore concentrations in the operator's breathing zone were about 24 times higher than acceptable levels. Surprisingly, working in the rear of the cabinet was less safe than working in the front, since the close proximity of the body to the cabinet influenced the airflow. Thus, for safe working conditions it is essential to limit the movement of arms and hands by arranging the equipment in the most practical way. Heidt (54) also tested the efficiency of a class II BSC. This author concluded that the cabinet provided sufficient protection, since microorganisms escaped only at the highest densities of the test aerosol created inside the cabinet. Since the number of bacteria detected was very low, this was considered to be acceptable. Osborne and coworkers (107) investigated a number of BSCs and calculated the operator protection factors (OPFs), as assessed by still and latterly limited “in-use” KI-Discus tests. The OPF is defined as the ratio of the exposure to airborne contamination generated on the open bench to the exposure resulting from the same dispersal of airborne contamination generated within the cabinet (66). Most BSCs had OPFs higher than 100,000, except when room pressure changed or when drafts occurred in the laboratory. The performance of class II BSCs was shown to be affected by the movements of the worker, and some movements reduced OPF results as found before by Macher and First (86). However, the levels of failure were marginal. The OPF tests revealed that a selected class II unit provided the same OPF as a class I unit when properly used.

Although the literature is not unequivocal, it appears that the use of BSCs decreases LAIs significantly (54, 86, 107, 119). However, in a recent publication, Rusnak et al. (119) examined illness surveillance data archived from the U.S. offensive biological warfare program (from 1943 to 1969) and concluded that BSCs and other measures failed to sufficiently prevent illness caused by agents with lower infective doses in a high-risk research setting.

Though required in some countries (including The Netherlands), cabinet performance is not generally assessed. However, on-site containment tests indicated that 37 class II BSCs (all with adequate type test certification and including 18 new installations) failed to meet the OPF requirements as defined in BS 7526. Thus, testing for containment using an OPF test appears to be essential both at commissioning and during routine maintenance (22, 107).

While formaldehyde gas has been used for over 100 years for decontamination, the efficacy of this process remains controversial (95). Moreover, because of its toxicity, the use of formaldehyde itself requires containment procedures (66). Formaldehyde decontamination of BSCs is usually validated using spore strips and culture. Therefore, poliovirus, Mycobacterium bovis strain BCG, or Bacillus spores have been used. Bacterial spores on stainless steel appear to be resistant to decontamination, and using bacterial spores to validate decontamination is too slow. Therefore, commercial biological indicator tests have been developed, which may be an aid in detecting incomplete decontamination. Difficulties in obtaining effective decontamination by using formaldehyde gas have been demonstrated. Factors contributing to the effectiveness of decontamination by formaldehyde include the formaldehyde level, the relative humidity, the temperature levels, and the medium to be contaminated. Locations beyond the exhaust filters of BSCs were the most difficult to decontaminate.

Cell sorters.

Modern cell sorting equipment has become an important tool in microbiological laboratories (77, 110). Because cell sorters lead to aerosol formation and are not easily accommodated by regular BSCs, this kind of apparatus could cause the operators to become contaminated (77, 110, 125). In addition, their high costs often prohibit their incorporation within BSL facilities. To solve this problem, Lennartz and coworkers (77) integrated a fluorescence-activated cell sorter (FACS) into a specially developed class II BSC. Biosafety was subsequently tested by using T4 bacteriophage aerosols and shown to be excellent. Bacteriophages were readily detected inside and outside when the airflow of the BSC was off, but when the BSC was turned on no bacteriophages could be detected outside (77). Many FACS protocols include inactivation steps, including the use of fixatives based on alcohols or formaldehyde. Some of these protocols have been evaluated for antimicrobial activity directed against specific pathogens, in particular HIV. Formaldehyde at concentrations of 0.5 to 2% is effective in inactivating HIV, but the ability of fixatives to inactivate other microorganisms in FACS equipment, including HBV, has not been demonstrated. In addition some protocols employ nonfixed cells. While analytic cytometers are engineered not to produce aerosols, jet-in-air cell sorters generate droplets and microdroplets that may be aerosolized. Recently high-speed cell sorting using high operating pressures with an increased potential for aerosol generation and an enhanced risk of sample splashes at the sample introduction port has become more prevalent. At the same time, instrument manufacturers have become more safety conscious and have developed novel devices for containment of aerosols and splashes, modified sample uptake ports on cell sorters, and installed mechanisms to stop sample flow in case of a nozzle clog to reduce operator risks (111, 125). A Vantage FACS was thus modified for safe use with potentially HIV-infected cells. Safety tests with bacteriophages were performed to evaluate the potential spread of biologically active material during cell sorting. The bacteriophage sorting showed that the biologically active material was confined to the sorting chamber. A failure mode simulating a nozzle blockage resulted in detectable droplets inside the sorting chamber, but no droplets could be detected when an additional air suction from the sorting chamber had been put on (133). While these observations may be reassuring, some recommendations regarding the use of FACS equipment are important (124-126). The recently published International Society for Analytical Cytology biosafety standard for sorting of unfixed cells states that droplet-based sorting of infectious or hazardous biological material requires a higher level of containment than the one recommended for the risk group classification of the pathogen (126). Training has to include performing aerosol containment testing of instruments to be used for biohazardous sorting. In addition, waste fluid has to be collected in 10% sodium hypochlorite, and fluid lines should be disinfected using a 1:10 dilution of 5.25% sodium hypochlorite. Notwithstanding their potential hazards, no documented disease transmission through the use of a cytometer has occurred (124, 125).

Respiratory protection devices.

A few papers examined the efficacy of face respirators and surgical masks. For example, Balazy et al. (5) examined the performance of two types of N95 half-mask, filtering face piece respirators and two types of surgical masks. The collection efficiency of these respiratory protection devices was investigated using MS2 virus (a nonharmful simulant of several pathogens) in a particle size range of 10 to 80 nm. Penetration of virions through N95 respirators, which are certified by the National Institute for Occupational Safety and Health (NIOSH), can exceed an expected level of 5%. The tested surgical masks showed a much higher particle penetration of the MS2 virions: 20.5% and 84.5%.

Does the Laboratory as a Whole Afford Effective Containment?

The proper functioning of equipment should be evaluated not only in isolation but also in the context of the entire laboratory. In considering biosafety of laboratories, the proper functioning of autoclaves may be overlooked. Barbeito and Brookey (7) and Marshall et al. (87) emphasized the potential of autoclaves to release viable microorganisms into the atmosphere and the importance of proper sterilizer location, ventilation, containment of heavily contaminated loads, and adequate sterilizer maintenance.

One of the few studies to assess contamination of the laboratory environment with pathogens found in blood examined 800 environmental samples taken from 10 clinical and research laboratories working at the BSL-2 level at the NIH. Thirty-one samples from 11 work stations in three laboratories contained HBV surface antigen. Factors associated with environmental contamination included flawed laboratory techniques (mouth pipetting, splashing, placing pens in the mouth, improper use of equipment, and improper instrument design requiring external wash steps) (odds ratio [OR], 9.78; 95% confidence interval [CI], 1.46 to 65.49), high work loads (OR, 5.06; 95% CI, 0.8 to 31.96), and inappropriate behaviors (including not wearing gloves) (OR, 2.75; 95% CI, 0.44 to 17.4). Flow cytometry was identified as the technique with the most frequent occurrence of overt spills (38). Indeed, HBV infection was among the most commonly reported LAIs. Laboratory workers in urban medical centers may have been at almost three times the risk of acquiring HBV infection than other hospital employees due to exposure to patients' blood and at 7 to 10 times the risk than that of the general public (38).

Some evidence of the effectiveness of a BSL-3 laboratory environment may be derived from experiences with a specially designed BSL-3 laboratory for autopsies of patients with SARS (79). SARS coronavirus is highly infectious, and during the outbreak of SARS more than 30% of the approximately 8,000 infected persons were health care workers. The autopsy laboratory was established in Beijing Ditan Hospital (which was designated the SARS hospital during the outbreak of SARS in China) in May 2003. Remarkably, the efficiency of decontamination in this laboratory was evaluated by a sarin simulant test. A sarin simulant aerosol of 0.3-μm particles at 4 mg/liter was generated and spread by a special device in the contaminated area. Sarin could not be detected in either the semicontaminated area or the clean area, and particles of >0.3 μm in size were not detected in the exhaust air. Twenty-three pathologists and technicians participated in 16 complete autopsies that were performed on patients with clinically confirmed or suspected SARS, of which seven cases were later confirmed to be SARS infections. None of these personnel demonstrated any evidence of SARS infection.

The set of biocontainment measures that define BSL-4 biosafety is comfortably the most comprehensive and stringent, but each setting and laboratory design is unique and comparative data on their containment effectiveness are nonexistent. A problem in assessing the effectiveness of BSL-4 containment measures is that isolated containment measures are considered insufficient. Thus, individual components such as autoclaves, incinerators, chemical decontamination showers, gaseous decontamination systems, air ventilation systems, and HEPA filters can be tested for physical parameters during normal operation and under extreme conditions, but it remains unclear how closely the simulated test conditions resemble the real-life situation. The effectiveness of HEPA filters is typically validated using bacterial spore strips or particles.

Are Laboratory Workers and the Environment Protected against Infection?

The analysis of laboratory accidents may illustrate what can go wrong and point the way to improvements. Such accidents are one of the most relevant parameters to evaluate the overall effectiveness of integrated biosafety measures. However, the epidemiology of the incidence and severity of LAIs is largely unknown, as there are neither national surveillance and monitoring systems with complete coverage nor many systematic studies on their occurrence (60, 131). Denominator data that are necessary to calculate the actual incidence of LAIs are usually lacking. In addition, LAIs may be subclinical and may have an atypical incubation period and route of infection, and laboratory workers and directors may be reluctant to report them because of fear of reprisal and stigma (52, 131). Therefore, much information is obtained from anecdotal case reports and some retrospective questionnaires. Such case reports do not always report on possible failure of biosafety procedures or unintended accidents. While accidental parenteral inoculation of infectious material appears to be one of the leading causes of LAIs, most LAIs appear to occur even with the best safety precautions in place (116, 131). A summary of some recent LAIs is given in Table 10.

TABLE 10.

Causes and factors involved in LAIs

Microorganism No. of cases Setting Possible primary cause Other associated factor(s) Reference(s)
GMO
    Recombinant vaccinia virus 1 Research lab Needle stick injury while handling mice 59
1 Research lab Unknown; no apparent failures in handling and no disruptions of epidermal barrier Possible enhanced infectivity due to immunomodulating insert (LFA/ICAM-1); long interval after vaccination 95
1 Research lab Needle stick injuries 105
1 Research lab Unknown Eczema as predisposing factor; noncompliance with procedures (no wearing of gloves, no recent vaccination) 89
1 Research lab Unknown Infrequent eye protection, laboratory coat sleeves were not elasticized and did not always cover the wrist, waste pipettes were not disinfected before removal from the biosafety cabinet, work (including vortexing) outside the BSC, no vaccination 78
Non-GMOs
    Vaccinia virus 1 Research lab Accidental cut wound on a coverslip No vaccination 158
1 Research lab Accidental needle stick injury No recent vaccination 94
    SARS coronavirus 1 Research lab Cross-contaminated West Nile virus sample, but no clear recognized laboratory accident Insufficient training and noncompliance with BSL-3 procedures 80
1 (90 persons quarantined) Research lab Spilling accident (in BSL-4 laboratory) 101
2, plus 8 secondary cases outside the lab, of which 1 was lethal Research lab Inadequate inactivation of SARS virus batch followed by transport to low-safety lab Noncompliance with procedures (no check on inactivation, no monitoring of workers' health status); in addition, there were problems with improper air circulation, poorly located autoclaves and freezers, and training and record-keeping (100) 36
2 Research lab Unknown; perhaps inadequate inactivation and infection outside the BSL-3 area 102
    Marburg virus 25, plus 6 secondary cases among medical staff (7 deaths total) Research lab Handling Vervet monkeys originating from Uganda before the virus was designated a risk group 4 organism 41, 132, 140
2 (Marburg, Germany, and Belgrade) Research lab Accidental infection before the virus was designated a risk group 4 organism 10, 99
    Ebola virus 1 Accidental needle inoculation while processing material from patients in Africa 35
5 people exposed but not ill, likely because the strains used were not very virulent or because of infinitesimal doses (Reston, VA, and Fort Detrick, Frederick, MD) Two accidents, occurring while handling nonhuman primates and mice 3
1 (lethal) (Novosibirsk, Siberia) Needle stick injury while handling guinea pigs 4
    Lassa virus 1 Handling tissue cultures and infected mice 76
    Sabiá virus 2 (Brazil and Yale University) Diagnostic and research laboratories Handling of unknown virus (in diagnostic laboratory) and leakage from centrifuging tubes (in BSL-3 laboratory) 8, 44, 82
    Burkholderia mallei 1 Research lab Noncompliance with biosafety practices (no routine use of latex gloves) Researcher had type 1 diabetes mellitus 135
    E. coli O157:H7 5 Different diagnostic labs Not identified or not reported Infections occurred before E. coli O157:H7 was reclassified as a risk group 3 organism 23
4 Different diagnostic labs Noncompliance with biosafety practices (handling without latex gloves, hands were not washed each time that gloves were removed, open laboratory coat) Sudden increase in volume of specimens, low infectious dose, prolonged survival on stainless-steel surfaces (155) 134
    Brucella melitensis 5 Diagnostic lab Not identified or not reported 49
7 Diagnostic lab No apparent failure of recommended safety practices Large number of isolates of Brucella spp. handled per year in area of endemicity 159
    Brucella melitensis/ B. abortus 75 (retrospective survey of 30 years; attack rate, 11.9 %) Different diagnostic labs Break in biosafety measures in 80% of the cases (including lack of recognition of an isolate of Brucella sp. and failure to work in a biological safety cabinet) Large number of isolates of Brucella spp. handled per year 12
    Mycobacterium tuberculosis 7 (retrospective survey) Different public health labs Needle stick injury in 1 case; in the other cases, the source of infection could not be determined Inadequate isolation procedures, high volume of specimens, faulty ventilation 63
    Neisseria meningitidis 16 (retrospective survey) Different research labs No respiratory protection in 15/16 cases 130
2 Two clinical labs High volume of specimens and formation of microaerosols 50
5 (retrospective survey; relative risk for laboratory workers, 184 [95% CI, 60 to 431]) Different labs Working outside a BSC 11

Reviews.

Most LAIs are caused by microorganisms that are very pathogenic or that need a very low infectious dose, including arboviruses, Venezuelan equine encephalitis virus, hantavirus, HBV, HCV, Brucella sp., Coxiella burnetii, Francisella tularensis, Mycobacterium tuberculosis, Salmonella spp., Shigella spp., Chlamydia psittaci, Blastomyces dermatitidis, Coccidioides immitis, Cryptosporidium spp., and organisms causing typhus, streptococcal infections, histoplasmosis, leptospirosis, tularemia, coccidiomycosis, and dermatomycosis (52, 116, 149). A direct link to accidents or exposure events, such as aspiration, injection, cut, spill, or bite, appears to be apparent in only a minority of the LAIs, while the majority are likely caused by undefined exposure to aerosols (52, 116, 149-151, 159). Indeed aerosols have been responsible for major outbreaks of LAIs caused by Brucella spp., Coxiella burnetti (Q fever), Chlamydia psittaci (psittacosis), and M. tuberculosis (89, 103, 129). The main hazards for inoculation (114-116, 129, 131) include (i) parenteral inoculation, (ii) inhalation of infectious aerosols, and less commonly, (iii) accidental oral ingestion and (iv) direct contact with mucous membranes or (broken) skin. Special hazards occur when working with infected animals. The presence of highly pathogenic microorganisms in unknown clinical samples likely explains the high incidence of tuberculosis among laboratory workers. In different studies the incidence of tuberculosis in laboratory personnel is estimated to be 3 to 100 times the frequency observed in the general population. The high infectivity of M. tuberculosis is related to its low infective dose (i.e., a 50% infective dose of <10 bacilli) (1, 117). Schellekens (123) calculated that 1 out of 100 to 1,000 laboratory workers per year are infected, but recent studies suggest that the rate of LAIs per person per year is decreasing (107, 123, 156).

Pike (116) tabulated the most common sources of LAIs from published literature and survey data. In the period from 1924 to 1977, there were 4,079 reported cases of LAIs with 168 casualties. In the subsequent period from 1980 to 1991, there were 375 reported cases with 5 casualties. At the time of Pike's survey, most LAIs (59%) occurred in research laboratories, compared with 17% in diagnostic laboratories. The highest mortality rate (7.8%) was associated with psittacosis. At that time, approximately 70% of LAIs resulted from work with the infectious agents (21%) or animals (17%), exposure to aerosols (13%), and accidents (18%). Less frequent sources of infection included clinical specimens (7%), autopsies (2%), and contaminated glassware (1%). Most causes of LAI were unknown (82%), and in only 18% of the reported cases could the cause be attributed to accidents, associated with the use of sharps such as needles (25%), injuries by glass (16%), splashes or spills (27%), mouth pipetting (13%), and bites by laboratory animals (14%). Many of the LAIs of unknown origin were likely caused by exposure to an infectious aerosol.

A recent survey of symptomatic and asymptomatic LAIs has been conducted by Harding and Byers (52), who reviewed 270 publications from 1979 to 2004, a period during which much has been done to improve laboratory safety while the work load in laboratories increased. A decrease in the number of LAIs would therefore be expected; however, knowledge on the total population at risk and the total number of infections would be needed. Harding and Byers (52) found a total of 1,448 cases and 36 deaths, 6 of which were aborted fetuses. The infections occurred in clinical, research, teaching, public health, and production facility laboratories, with clinical and research laboratories accounting for approximately 76%. In recent years more LAIs from clinical laboratories were reported, probably due to a more active employee health program, the absence of biosafety containment equipment in a number of clinical laboratories, or the fact that during the early stages of culture identification, personnel are working with unknowns and may not be using adequate containment procedures. Like earlier findings, the authors report that only a small proportion of the LAIs resulted from actual accidents. Most were acquired by simply working in the laboratory or by exposure to infected animals.

Sewell (131) concluded that adherence to the guidelines promulgated by the various regulatory agencies decreases the risk of occupational exposure to infectious agents. However, he also recommended additional studies to evaluate the effectiveness of other safety measures implemented or mandated in the laboratory. Interestingly, Sewell (131) describes personal risk factors of laboratory workers that are associated with accidental infections. Characteristics of persons who have few accidents include adherence to safety regulations, a respect for infectious agents, “defensive” work habits, and the ability to recognize a potentially hazardous situation. In contrast, persons involved in laboratory accidents tend to have low opinions of safety programs, to take excessive risks, to work too fast, and to be less aware of the infectious risks of the agents they are handling. Also, men and younger employees (17 to 24 years old) are involved in more accidents than women and older employees (45 to 64 years old).

While many reports emphasize the importance of personal protection, there are indications that extensive personal protection by use of double gloves, face masks, and protective clothing is not the sole solution, since such measures can reduce the dexterity of the laboratory worker, leading to increased accidents (122). This indicates that the use of sharps should be minimized when workers wear extensive personal protection.

Laboratory-acquired parasitic infections, from both protozoa and helminths, have been extensively reviewed by Herwaldt (56). Importantly, because protozoa, in contrast to most helminths, multiply in humans, even a small inoculum can cause illness. The author summarizes 199 case reports on laboratory and health care workers. The most frequently reported parasitic infections were caused by Trypanosoma cruzi, Toxoplasma gondii, Plasmodium spp., Leishmania spp., and Cryptosporidium parvum. Two cases (one of Chagas' disease and one of toxoplasmosis) were fatal. However, as with other infections, accurate counts of accidental exposures and infections and information on the risk per person-year are unavailable. Some of the laboratory-acquired parasitic infections were directly linked to accidents (e.g., a bite by an escaped infected mosquito) and poor laboratory practices, such as recapping a needle, removing a syringe from a needle, working barehanded, mouth pipetting, and working too fast. For 105 cases an accident or a likely route of exposure could be presumed; 47 (44.8%) of these included a percutaneous exposure via a sharp object. In other cases no apparent accidents were recognized or reported, suggesting that subtle exposures (e.g., contamination of unrecognized microabrasions and exposure through aerosolization or droplet spread) resulted in infection.

Surveys.

Walker and Campbell (146) did one of the few systematic but retrospective studies. They carried out a retrospective questionnaire survey of 397 responding United Kingdom laboratories covering 1994 and 1995. Approximately 75% of these were diagnostic laboratories, 14% were research laboratories, and 9% were teaching laboratories. Over 55,000 person-years of occupational exposure were covered, and only nine cases of LAI were identified, giving an overall infection incidence rate of 16.2/100,000 person-years, compared with 82.7 infections/100,000 person-years found in a similar survey covering 1988 and 1989 which was conducted by Grist and Emslie (48). This decline in incidence continues the trend previously reported for the period from 1970 to 1989. Infections were most common in females (in contrast to the findings reported by Sewell [131]), in relatively young staff, in microbiology laboratory workers, and in scientific/technical employees. Gastrointestinal infections predominated, particularly shigellosis, but few specific etiological factors relating to working practices were identified. These included a broken glass leading to a hand cut, a rat bite, and aerosol contamination. In most cases no clear accident was reported. Lack of experience was cited as a definite factor in two of the cases. Single cases of HCV, E. coli O157, and M. tuberculosis infection were identified, in addition to single cases of nonspecified septicemia and gastroenteritis. The absence of any cases of HBV infection, as in 1988 to 1989, reflects a sharp decline since 1970 and was attributed to increased awareness, better technique, and the availability of immunization. Furthermore, the absence of eye infections and the paucity of skin infections may indicate good technique and use of protective equipment. Despite the shortcomings of this study (retrospective study design, no reliable denominator, potential underreporting or underrecording, and no detection of asymptomatic infections), the authors concluded that the small number of cases identified indicates high standards of infection control, although they still recognized room for improvement. Finally, the study emphasized that the notification system in place in the United Kingdom to report LAIs is inadequate for the task of monitoring their true incidence in a comprehensive way, a conclusion that probably holds true for many other countries. For comprehensive monitoring of the incidence of LAIs, it is necessary to establish a routine, active surveillance program or prospective survey which has the support and commitment of the laboratories themselves.

Recently, a report was drafted on the biosafety status of clinical laboratories in Japan (160). Data were obtained from 431 hospitals and 301 institutions. The authors found 28 cases of possible laboratory-associated tuberculosis infection, of which 25 could be associated with the lack of BSCs, which are required for work with M. tuberculosis. Other risk factors were insufficiently skilled equipment operation and rupture accidents during centrifugation of blood. Within the last 5 years 1,534 events of self-inflicted needle punctures were recorded (160).

A retrospective survey of incidents occurring during biotechnological and clinical work in Flanders, Belgium, indicated that on average 13.6 incidents occurred per year among 7,302 laboratory workers. As a result, 69 persons (<1%) were exposed to biological agents, resulting in 2 LAIs, caused, respectively, by L. monocytogenes and Brucella melitensis. Most incidents occurred in clinical laboratories, likely caused by the higher number of working hours actually spent in clinical laboratories and the sometimes unknown nature of microorganisms. Handling of experimental animals and waste was considered risky. Most incidents were caused by human failure, including prick accidents, spilling, breaking, and maintenance work carried out in the laboratory (30).

Sejvar et al. (130) undertook a systematic, retrospective evaluation of the risk of meningococcal disease among clinical microbiologists and an assessment of the laboratory procedures that might predispose technicians to infection. Cases of suspected or proven laboratory-acquired meningococcal disease were identified by placing an information request on e-mail discussion groups of infectious disease, microbiology, and infection control professional organizations. Sixteen cases of probable laboratory-acquired meningococcal disease occurring worldwide between 1985 and 2001 were identified, including six U.S. cases between 1996 and 2000. Nine cases (56%) were serogroup B; seven (44%) were serogroup C. Eight cases (50%) were fatal. In 15 cases (94%), isolate manipulation was performed without respiratory protection. An average of three microbiologists are estimated to be exposed to the 3,000 meningococcal isolates seen in U.S. laboratories, yearly resulting in an attack rate of 13/100,000 microbiologists between 1996 and 2001, compared to 0.2/100,000 among U.S. adults in general. The case/fatality rate of 50% seen among survey cases is substantially higher than that observed among community-acquired cases, which may be explained by ascertainment bias due to underreporting of mild cases of disease. However, an alternative possibility is that clinical microbiologists routinely work with highly virulent strains and high concentrations of organisms. All cases identified in this inquiry occurred among microbiologists and not among workers in other areas of the clinical laboratory. This suggests that exposure to isolates of Neisseria meningitidis, and not patient samples, represents the increased risk for infection. In addition, all isolates were derived from sterile sites. None of the microbiologists identified were working with isolates obtained from pharyngeal or respiratory secretions, suggesting that such pharyngeal isolates represent a lower risk, presumably due to their lower pathogenicity. The authors concluded that prevention should focus on the implementation of class II BSCs or additional respiratory protection during manipulation of suspected meningococcal isolates. Following two cases that prompted this survey, CDC has instituted a prospective surveillance for laboratory-acquired meningococcal disease (21).

In The Netherlands, two surveillance systems monitor the occurrence of labor-acquired infections, i.e., not exclusively LAIs. The number of reported LAIs is low, but both systems suffer from serious underreporting and do not provide details of the transmission route or accidental cause of infection (55).

GMO-associated laboratory accidents.

Fortunately, the number of accidental releases or LAIs with GMOs appears to be very low. Jones et al. (59) reported the first case of accidental vaccination with a recombinant vaccinia virus (Western Reserve [WR] strain) expressing the nucleoprotein gene of vesicular stomatitis virus. The infection had a relatively mild course, perhaps because the laboratory worker had received smallpox vaccination 30 years before the accident or because of attenuation of the virus by the insertional inactivation of the thymidine kinase gene. Openshaw et al. (105) reported an accidental infection of a laboratory worker with recombinant vaccinia virus (WR strain) expressing proteins of respiratory syncytial virus. The infection occurred through two separate needle accidents during the same work session, although the worker was experienced in the procedure. The procedure was subsequently modified to prevent further accidents. The laboratory worker had been vaccinated with standard smallpox vaccine, a practice that may have restricted the severity of symptoms to local redness and swelling. Mempel et al. (89) reported the case of a recombinant vaccinia virus infection in a previously vaccinated researcher working with various genetically modified strains. The isolated virus carried a functionally inactivated cytohesin-1 gene of human origin, which impairs leukocyte adhesion by interacting with the LFA/ICAM-1 axis. The immunomodulating nature of the inserted construct might have added to the infectivity of the virus. Although the paper does not detail safety procedures in the lab, the infection occurred while the handling was considered proper. Contact infections were not reported. Lewis et al. (78) reported a case of ocular vaccinia infection in an unvaccinated laboratory worker. The infecting virus was a unique form of recombinant WR vaccinia virus constructed in the laboratory. Although laboratory staff generally followed established biosafety precautions, several opportunities for virus exposure were identified. Experiments were performed partly outside a BSC. Staff infrequently wore eye protection. Laboratory coat sleeves were not elasticized and did not always cover the wrist. Waste pipettes were not disinfected before removal from the BSC. Instances occurred in which samples with low titers of live virus were removed from the BSC, transported to other parts of the facility, and manipulated. In addition, laboratory staff routinely vortexed tubes containing live virus outside the BSC. Finally, a laboratory-acquired recombinant vaccinia virus infection running a severe course was reported in a research laboratory technician who had a long history of eczema. She had been working with a thymidine kinase-deficient strain of vaccinia virus for use as a vector for gene therapy. She did not report an accidental inoculation, but she usually wore no protective gloves. She had been vaccinated with smallpox vaccine as a child but not again before beginning her laboratory work with vaccinia virus (83).

Another case of laboratory-acquired vaccinia virus (nonrecombinant) infection was reported by Wlodaver et al. (158). This infection occurred in a laboratory technician who had not been previously vaccinated and who developed generalized vaccinia. She had accidentally cut a finger on a coverslip while working with vaccinia virus. Evaluation of this accident in her laboratory prompted a review of procedures for handling contaminated glassware. Moussatché et al. (94) reported another accidental needle stick inoculation of a laboratory worker with vaccinia virus. Although the patient had previously been vaccinated against smallpox, severe lesions appeared on the fingers.

In total, there have been at least 19 reported cases of laboratory-acquired vaccinia virus infections, of which 5 were accidental infections with recombinant vaccinia (83; this paper). Several researchers emphasize the necessity of vaccinia vaccination of laboratory workers. In the United States and Canada, specific recommendations exist for laboratory personnel who conduct research with (recombinant) orthopoxviruses, including vaccinia virus (20, 153). The Advisory Committee on Immunization Practices recommends revaccination at least every 10 years for persons working with non-highly attenuated vaccinia viruses, recombinant viruses developed from non-highly attenuated vaccinia viruses, or other nonvariola orthopoxviruses. To ensure an increased level of protection against more virulent nonvariola orthopoxviruses (e.g., monkeypox virus), empirical revaccination every 3 years can be considered. In contrast, mandatory guidelines with respect to vaccinia vaccination do not exist in Europe (57).

In conclusion, LAIs with GMOs appear very seldom and appear to be restricted to infections with recombinant vaccinia virus. Although laboratory accidents with other GMOs may have been unnoticed due to a subclinical course of infection, this situation seems to reflect that vaccinia virus is very widely used. Perhaps more important, the recombinant virus is still pathogenic, and this might be enhanced by certain gene inserts. Vaccinia virus infection can be established via several routes, including breaks in the skin, and the infectious dose is probably low. Guidelines for working safely with vaccinia virus, which include vaccination, are available (57). We consider it nonetheless advisable to work with the highly attenuated strains of vaccinia virus (modified vaccinia virus Ankara and NYVAC) or with avian poxviruses that have a restricted host range and do not replicate in mammals (ALVAC and TROVAC) whenever possible.

Accidents with risk category 4 organisms.

LAIs with category 4 biological agents (filoviruses, arenaviruses, flaviviruses, and bunyaviruses) are extremely rare and usually occurred earlier in settings with lower levels of biocontainment and/or involved animal work (41, 42, 51, 91, 127, 138, 139, 114). Rare laboratory incidents with New World arenaviruses have been reported in earlier surveys more than 4 to 5 decades ago (e.g., with Junin virus and Machupo virus) (51). Such experiences illustrated the need for more effective measures to reduce hazards.

While this low number of BSL-4 laboratory accidents may be reassuring, the number of BSL-4 labs and workers is increasing. This appears to be in defiance of the “concentration and enclosure” principle, because the risks associated with this work may increase with the number of facilities and workers (61, 62). Concern is further fuelled by several incidents that included unreported infections (among others involving Brucella and Coxiella burnetti) and other biosafety breaches. In 2006, a Department of Health and Human Services Inspector General audit of security procedures found that 11 of 15 institutions had “serious weaknesses,” such as unlocked doors and freezers and lax inventory records (61, 62). Another incident, in 2007, was the escape of foot-and-mouth disease virus from the Pirbright facilities in the United Kingdom, which has been linked to an outdated effluent system and caused several outbreaks of this very contagious disease among cattle and sheep (53).

DISCUSSION

In this paper we have reviewed the principles underlying biosafety measures for work with pathogenic microorganisms and GMOs, and we have examined to what extent evidence for their effectiveness is available. Clearly the risks of working with GMOs are considered to be largely identical to those of working with pathogenic non-GMO microorganisms, and hence much of the knowledge of and containment measures for GMOs are derived from the latter. Regulations appear to be more strict for GMOs, however.

Among many reports on biosafety, we found only scarce information on the evaluation of effectiveness and on criteria to judge effectiveness. We must therefore keep in mind that safety cannot be expressed in absolute terms. It is a relative concept defined in terms of tolerability and acceptability limits (64). This notion implies that workers and regulators try to find a balance between the costs of safety measures and the potential benefits of the work for society. For example, in microbiological work, safety measures and associated costs increase from BSL-1 to BSL-4. Indeed, safety measures at BSL-1 and -2 are probably insufficient to prevent all infections with microorganisms of the corresponding risk categories, but their consequences at these levels are considered to be acceptable or negligible.

The current biosafety practice gradually evolved during the previous century. Therefore, it is not immediately obvious whether and what principles have been employed to ensure safe work and on which scientific basis they were built. In this paper we have tried to identify some principles that appear to underlie the current practice. Such principles clearly partly overlap and mutually enhance each other. A central activity, either implicit but preferably explicit, is a thorough risk assessment procedure that considers all potentially harmful effects and their possibility of occurrence. Other important underlying principles are the use of (wherever possible and appropriate) biological containment, concentration and enclosure, exposure minimization, physical containment, and hazard minimization.

Clearly, throughout the world regulators have adapted the model of universal precautions based on a classification of microorganisms in four hazard classes and accompanying standard safety practices (16, 156). The advantage of this model is that work with certain microorganisms can be grouped together to comply with the accompanying containment rules according to their classification. A disadvantage may be that this universal model may overlook the necessity to tailor safety measures for specific microorganisms or specific strains with particular routes of transmission or virulence properties. Researchers should also reckon with variability in human immunocompetence. Therefore, risk assessment remains at the core of any individual experiment. Such a risk assessment should in particular be based on (preferably quantitative) parameters of transmission, infectivity, and virulence. These should guide the subsequent measures aimed at reducing the amount of microorganisms to which individuals are exposed to below a minimal threshold level of infectivity. Nonetheless, this universal model of four biohazard classes appears to work well, but we recommend further harmonization of criteria for both non-GMO and GMOs and between different regulatory authorities, such as the European Union, WHO, and CDC.

Altogether, the regulations specifying the biosafety containment measures appear to be based on experience, expert judgment, and common sense. They are not motivated or supported (at least not explicitly) by scientific literature, however, and often are not based on precisely defined or specified properties of microorganisms and vector and insert sequences. In addition, the regulations do not exactly specify the level of protection that they aim to afford, for example, in terms of diminishing exposure of the laboratory workers below a threshold level of infectivity. Furthermore, it is clear that the physical containment classes 1 to 4 afford increasing levels of containment, but it is not sufficiently clear and scientifically supported to what extent they provide effective protection with regard to prevention of infection of laboratory personnel, prevention of airborne escape, etc. This, together with sometimes not very detailed regulations, puts much responsibility on researchers, lab directors, advisors, and regulatory authorities in further detailing working practices. The regulations also do not comprise evaluation procedures to monitor the compliance or effectiveness of the containment provisions. Table 11 summarizes our recommendations.

TABLE 11.

Summary of recommendations

Recommendation
Strengthen the evidence base wherever possible and feasible in modernizing biosafety measures; this may enhance the effectiveness of biosafety measures as well as compliance with these measures
Develop one set of guidelines and regulations for GMOs and non-GMOs
Develop explicit goals of biosafety measures that can be evaluated
Knowledge and measures of biosafety should be directed to (preferably quantitative) parameters of infectivity and transmission
Further harmonize biosafety guidelines between regulatory authorities (European Union, CDC, WHO)
Take gene-gene and gene-environment interactions into account in risk assessment
Optimize the use of biological containment, in particular with respect to the use of recombinant vaccinia virus
Monitor and evaluate biosafety aspects in laboratories and their compliance regularly
Optimize systematic surveillance of laboratory accidents and implement recommendations following such accidents; “blame-free” reporting may enhance the reporting rate, and serological monitoring may support the detection of laboratory infections and should match the risks involved
Promote education of laboratory personnel and compliance with the rules
Collect data to support the evidence base of the biosafety practice
Develop mathematical models to support the further development of knowledge of biosafety, to detect gaps in our knowledge, and to support the development and evaluation of biosafety measures

The hazard classification of work with GMOs follows the classification of work with non-GMOs. This extrapolation should be based on a risk estimation as precise as possible considering the genetic modifications involved. In case of doubt or uncertainty about the properties of the GMO involved, regulators and biosafety experts will choose a higher risk classification than the risk level of the microorganism from which the GMO has been derived, or they will demand additional safeguards. Basic research on transmission properties of GMOs, in comparison with those of the nonmodified organisms, may be helpful in such a risk assessment to further define the risks involved in the manipulation of GMOs. However, often properties, such as infective dose, may be difficult to obtain. In risk assessment we consider it important to take gene-gene and gene-environment interactions into account, because experience (for example, with the IL-4/ectromelia construct) has shown that specific gene products may have unwanted effects in a particular environment (58).

Regulators do not always require routine evaluation and monitoring of biosafety aspects in laboratories, which we would like to recommend. Routine monitoring of biosafety aspects, including monitoring of compliance and educational and behavioral aspects, may not easily be implemented, in particular in the many clinical laboratories with their high workload involving a wide variety of sometimes unknown microorganisms, but it may enhance overall safety awareness. For example, validation experiments using T4 bacteriophage, bacterial spores, or other indicator microorganisms could be useful for this purpose.

From the literature it appears obvious that there is little experience and no consensus on how the effectiveness of biosafety practices should be evaluated. Clearly, the effectiveness of biosafety measures can be assessed at different levels and under different circumstances that logically complement each other; i.e., one could question whether a single piece of equipment is effective under experimental conditions or, conversely, whether the population has not been accidentally exposed to LAIs. Data on the biological containment efficiency of equipment and laboratories are scarce and fragmented and are mainly limited to technical specifications. Monitoring of LAIs therefore appears to play a pivotal role in evaluating the effectiveness of containment and the potential exposure of laboratory workers and the population, but it suffers from serious underreporting throughout the world. Many reports of laboratory accidents are only anecdotal. We therefore recommend optimization of the systematic monitoring of laboratory accidents, including the serological monitoring of personnel. Infection with microorganisms, either GMO or not, that have a high infective dose or low virulence, usually belonging to risk category 1 and 2, may be difficult to detect. The extent of serological monitoring should therefore depend on the risks involved. A passive sampling strategy, i.e., collecting serum samples at the time of employment and following incidents, may be sufficient for work with low-virulence microorganisms, but an active sampling strategy at regular intervals may be considered for class 3 and 4 microorganisms. One clue to optimizing monitoring of accidents may be the introduction of “blame-free” reporting, which aims for workers to share experiences without being punished. In addition to systematic monitoring, retrospective surveys may be very useful, as they may identify certain risk factors, as shown for the occurrence of meningococcal disease (130).

Despite the methodological imperfections, it is clear that the number of GMO-associated laboratory accidents is very scarce in comparison with the number of non-GMO-associated infections and is practically restricted to accidental infection with recombinant vaccinia virus. We interpret this finding to mean that the biological containment obtained by attenuating GMOs is possibly a major factor in preventing their transmission. However, other factors contributing to the low number of GMO-associated accidents may be that GMOs are well characterized, implying that the worker has knowledge of the properties of the GMO and that the work load involving GMOs is likely much lower than that in clinical laboratories. Moreover, clinical samples may contain unknown pathogens. Other factors that may contribute to this low number of accidents involving GMOs are the stricter regulatory framework and a stricter compliance with containment rules. In many countries, including The Netherlands, both the researcher and the regulator make a risk assessment for each individual project that involves GMOs, a practice that is less developed for work with non-GMOs. In case of doubt or uncertainties regarding the properties of GMOs, biosafety experts and regulators will demand a higher risk category or additional measures. Both a local biosafety officer and a national inspectorate supervise this practice. Whenever possible, we consider it important to further optimize the possibilities of employing genetic modification to enhance the safety of GMOs. In particular, we recommend further definition of the genetic properties underlying the transmissibility and infectivity of microorganisms, measurement of the influence of specific mutations on infective dose and transmission properties of GMOs, and that containment rules be based on such findings. This is not an easy task but would provide a further scientific basis for the phenotypic properties of GMOs and the accompanying level of biological containment afforded by specific genetic alterations. Vaccinia virus or recombinant viruses developed from non-highly attenuated vaccinia viruses appear to be less well suited as vector organisms due to their retained virulence and low infectious dose and should be replaced by safer poxvirus vectors wherever possible.

In many reports of LAIs there has been a lack of compliance with biosafety practices. This observation may be reassuring regarding the effectiveness of such biosafety practices, at least if they are followed. It illustrates that education of laboratory personnel and compliance with the rules remain the top priority. Increased attention to these aspects may have caused a decrease in the rate of LAIs per person per year (107, 156). On the other hand, in the majority of cases of LAIs a direct cause could not be assigned (52, 116, 131, 149, 159), suggesting that a failure was not noticed in many cases or that containment may have been insufficient. This observation may warrant further research on the routes of exposure in such cases and on the effectiveness of measures. Finally, although monitoring of LAIs is an important element in evaluating the effectiveness of containment measures, it may overlook the risks associated with nonreplicating agents, such as transduction by nonreplicating viruses.

Many countries, including The Netherlands, regulate work with pathogenic microorganisms and GMOs differently. Because the regulations are derived from the same underlying principles and use the same instruments for biosafety and because the number of accidents involving GMOs is very low, we recommend harmonization, modernization, and simplification of the regulatory framework through developing a single set of regulations for both non-GMOs and GMOs.

Despite their presumed overall effectiveness in providing biosafety, it is often unclear to what extent the current set of specific biological or physical containment measures, alone or together, contribute to the prevention of transmission of pathogenic microorganisms or GMOs. In further developing and modernizing the biosafety practice, we therefore recommend developing evidence-based practices and criteria to evaluate effectiveness wherever this is possible and feasible. This may optimize and perhaps simplify future biosafety measures and stimulate compliance with the rules. Although scientific research may strengthen the evidence base for biosafety measures, such work is complicated and does not necessarily guarantee new findings on which further improvements can be based. To unravel complexities and to obtain further insight into the contribution of specific elements to biosafety, mathematical modeling, which is directed at quantitative parameters of infectivity and transmission, may be supportive, but modeling obviously needs confirmation by observational and experimental findings. Such an approach may, however, point to the data that are needed to further guide the development of evidence-based risk analysis and containment policy for both non-GMO pathogens and GMOs.

Acknowledgments

This work was financially supported by the Dutch Committee for Genetic Modification (COGEM).

We thank Marja Agterberg, Marjolein van Esschoten, Ben Peeters, Erik Schagen, Gijsbert van Willigen, and Dick van Zaane for their constructive support.

REFERENCES

  • 1.Andersen, P., and T. M. Doherty. 2005. The success and failure of BCG—implications for a novel tuberculosis vaccine. Nat. Rev. Microbiol. 3:656-662. [DOI] [PubMed] [Google Scholar]
  • 2.Angelakopoulos, H., K. Loock, D. M. Sisul, E. R. Jensen, J. F. Miller, and E. L. Hohmann. 2002. Safety and shedding of an attenuated strain of Listeria monocytogenes with a deletion of actA/plcB in adult volunteers: a dose escalation study of oral inoculation. Infect. Immun. 70:3592-3601. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Anonymous. 2004. Ebola virus, laboratory accident—USA (Maryland). ProMED-mail, 2 February 2004. 20040220.0550. http://www.promedmail.org. Accessed 22 September 2007.
  • 4.Anonymous. 2004. Ebola, lab accident death—Russia (Siberia). ProMED-mail, 22 May 2004. 20040522.1377. http://www.promedmail.org. Accessed 22 September 2007.
  • 5.Bałazy, A. M., Toivola, A. Adhikari, S. K. Sivasubramani, T. Reponen, and S. A. Grinshpun. 2006. Do N95 respirators provide 95% protection level against airborne viruses, and how adequate are surgical masks? Am. J. Infect. Control 34:51-57. [DOI] [PubMed] [Google Scholar]
  • 6.Barbeito, M. S., and L. A. Taylor. 1968. Containment of microbial aerosols in a microbiological safety cabinet. Appl. Microbiol. 16:1225-1229. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Barbeito, M. S., and E. A. Brookey, Jr. 1976. Microbiological hazard from the exhaust of a high-vacuum sterilizer. Appl. Environ. Microbiol. 32:671-678. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Barry, M., M. Russi, L. Armstrong, D. Geller, R. Tesh, L. Dembry, J. P. Gonzalez, A. S. Khan, and C. J. Peters. 1995. Treatment of a laboratory-acquired Sabia virus infection. N. Engl. J. Med. 333:294-296. [DOI] [PubMed] [Google Scholar]
  • 9.Berg, P., D. Baltimore, S. Brenner, R. O. Roblin III, and M. F. Singer. 1975. Asilomar conference on recombinant DNA molecules. Science 188:991-994. [DOI] [PubMed] [Google Scholar]
  • 10.Borchert, M. 2001. Marburg hemorrhagic fever, 1967 to the present (02). ProMED-mail, 8 Feb 2001. 20010208.0248.] http://www.promedmail.org. Accessed 22 September 2007.
  • 11.Boutet, R., J. M. Stuart, E. B. Kaczmarski, S. J. Gray, D. M. Jones, and N. Andrews. 2001. Risk of laboratory-acquired meningococcal disease. J. Hosp. Infect. 49:282-284. [DOI] [PubMed] [Google Scholar]
  • 12.Bouza, E., C. Sanchez-Carrillo, S. Hernangomez, M. J. Gonzalez, et al. 2005. Laboratory-acquired brucellosis: a Spanish national survey. J. Hosp. Infect. 61:80-83. [DOI] [PubMed] [Google Scholar]
  • 13.Braat, H., P. Rottiers, D. W. Hommes, N. Huyghebaert, E. Remaut, J. P. Remon, S. J. van Deventer, S. Neirynck, M. P. Peppelenbosch, and L. Steidler. 2006. A phase I trial with transgenic bacteria expressing interleukin-10 in Crohn's disease. Clin. Gastroenterol. Hepatol. 4:754-759. [DOI] [PubMed] [Google Scholar]
  • 14.Breban, R., R. Vardavas, and S. Blower. 2007. Theory versus data: how to calculate R0? PLoS One 2:e282. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Reference deleted.
  • 16.Buesching, W. J., J. C. Neff, and H. M. Sharma. 1989. Infectious hazards in the clinical laboratory: a program to protect laboratory personnel. Clin. Lab. Med. 9:351-361. [PubMed] [Google Scholar]
  • 17.Bukovsky, A. A., J. P. Song, and L. Naldini. 1999. Interaction of human immunodeficiency virus-derived vectors with wild-type virus in transduced cells. J. Virol. 73:7087-7092. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Casadevall, A., and L. A. Pirofski. 1999. Host-pathogen interactions: redefining the basic concepts of virulence and pathogenicity. Infect. Immun. 67:3703-3713. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.CDC. 1974. Classification of etiological agents on the basis of hazard. U.S. Public Health Service, Washington, DC.
  • 20.CDC. 2001. Vaccinia (smallpox) vaccine: recommendations of the Advisory Committee on Immunization Practices (ACIP), 2001. Morb. Mortal. Wkly. Rep. 50:1-25. [PubMed] [Google Scholar]
  • 21.CDC. 2002. Laboratory-acquired meningococcal disease—United States, 2000. Morb. Mortal. Wkly. Rep. 51:141-144. [PubMed] [Google Scholar]
  • 22.Clark, R. P. 1997. Standards for safety cabinets. Nature 390:550. [DOI] [PubMed] [Google Scholar]
  • 23.Coia, J. E. 1998. Nosocomial and laboratory-acquired infection with Escherichia coli O157. J. Hosp. Infect. 40:107-413. [DOI] [PubMed] [Google Scholar]
  • 24.Collaco, R. E., X. Cao, and J. P. Trempe. 1999. A helper virus-free packaging system for recombinant adeno-associated virus vectors. Gene 238:397-405. [DOI] [PubMed] [Google Scholar]
  • 25.Corbett, E. L., C. J. Watt, N. Walker, D. Maher, B. G. Williams, M. C. Raviglione, and C. Dye. 2003. The growing burden of tuberculosis: global trends and interactions with the HIV epidemic. Arch. Intern. Med. 163:1009-1021. [DOI] [PubMed] [Google Scholar]
  • 26.Curtiss, R. 1978. Biological containment and cloning vector transmissibility. J. Infect. Dis. 137:668-675. [DOI] [PubMed] [Google Scholar]
  • 27.Danelishvili, L., L. S. Young, and L. E. Bermudez. 2006. In vivo efficacy of phage therapy for Mycobacterium avium infection as delivered by a nonvirulent mycobacterium. Microb. Drug Resist. 12:1-6. [DOI] [PubMed] [Google Scholar]
  • 28.Darquet, A. M., B. Cameron, P. Wils, D. Scherman, and J. Crouzet. 1997. A new DNA vehicle for nonviral gene delivery: supercoiled minicircle. Gene Ther. 12:1341-1349. [DOI] [PubMed] [Google Scholar]
  • 29.Debyser, Z. 2003. Biosafety of lentiviral vectors. Curr. Gene Ther. 3:517-525. [DOI] [PubMed] [Google Scholar]
  • 30.De Cock, B., and G. van Eetvelde. 2007. Biologische Veiligheid. Verkennende bio-incidentenanalyse in Vlaanderen. Report, Vlaamse milieumaatschappij, MIRA/2007/01. www.biorisico.UGent.be.
  • 30a.de Wind, N., B. P. Peeters, A. Zuderveld, A. L. Gielkens, A. J. Berns, and T. G. Kimman. 1994. Mutagenesis and characterization of a 41-kilobase-pair region of the pseudorabies virus genome: transcription map, search for virulence genes, and comparison with homologs of herpes simplex virus type 1. Virology 200:784-790. [DOI] [PubMed] [Google Scholar]
  • 31.Dietrich, G., A. Bubert, I. Gentschev, Z. Sokolovic, A. Simm, A. Catic, S. H. Kaufmann, J. Hess, A. A. Szalay, and W. Goebel. 1998. Delivery of antigen-encoding plasmid DNA into the cytosol of macrophages by attenuated suicide Listeria monocytogenes. Nat. Biotechnol. 16:181-185. [DOI] [PubMed] [Google Scholar]
  • 32.Dilts, D. A., I. Riesenfeld-Orn, J. P. Fulginiti, E. Ekwall, C. Granert, J. Nonenmacher, R. N. Brey, S. J. Cryz, K. Karlsson, K. Bergman, T. Thompson, B. Hu, A. H. Brückner, and A. A. Lindberg. 2000. Phase I clinical trials of aroA aroD and aroA aroD htrA attenuated S. typhi vaccines; effect of formulation on safety and immunogenicity. Vaccine 18:1473-1484. [DOI] [PubMed] [Google Scholar]
  • 33.DiPetrillo, M. D., T. Tibbetts, H. Kleanthous, K. P. Killeen, and E. L. Hohmann. 1999. Safety and immunogenicity of phoP/phoQ-deleted Salmonella typhi expressing Helicobacter pylori urease in adult volunteers. Vaccine 18:449-459. [DOI] [PubMed] [Google Scholar]
  • 34.Reference deleted.
  • 35.Emond, R. T., B. Evans, E. T. Bowen, and G. Lloyd. 1977. A case of Ebola virus infection. Br. Med. J. 2:541-544. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Enserink, M., and L. Du. 2004. SARS. China dumps CDC head, probes lab. Science 305:163. [DOI] [PubMed] [Google Scholar]
  • 37.Escarpe, P., N. Zayek, P. Chin, F. Borellini, R. Zufferey, G. Veres, and V. Kiermer. 2003. Development of a sensitive assay for detection of replication-competent recombinant lentivirus in large-scale HIV-based vector preparations. Mol. Ther. 8:332-341. [DOI] [PubMed] [Google Scholar]
  • 38.Evans, M. R., D. K. Henderson, and J. E. Bennett. 1990. Potential for laboratory exposures to biohazardous agents found in blood. Am. J. Public Health 80:423-427. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Farson, D., R. McGuinness, T. Dull, K. Limoli, R. Lazar, S. Jalali, S. Reddy, R. Pennathur-Das, D. Broad, and M. Finer. 1999. Large-scale manufacturing of safe and efficient retrovirus packaging lines for use in immunotherapy protocols. J. Gene Med. 1:195-209. [DOI] [PubMed] [Google Scholar]
  • 40.Fenner, F., D. A. Henderson, I. Arita, Z. Ježek, and I. D. Ladnyi. 1988. Small pox and its eradication. WHO, Geneva, Switzerland. http://whqlibdoc.who.int/smallpox/9241561106.pdf.
  • 41.Feldmann, H., W. Slenczka, and H. D. Klenk. 1996. Emerging and re-emerging of Filoviruses. Arch. Virol. Suppl. 11:77-100. [DOI] [PubMed] [Google Scholar]
  • 42.Feldmann, H., S. Jones, H. D. Klenk, and H. J. Schnittler. 2003. Ebola virus: from discovery to vaccine. Nat. Rev. Immunol. 3:677-685. [DOI] [PubMed] [Google Scholar]
  • 43.Flemming, D. O. 2000. Risk assessment of biological hazards, p. 57-64. In D. O. Flemming and D. L. Hunt (ed.), Biological safety: principles and practices, 3rd ed. ASM Press, Washington, DC.
  • 44.Gandsman, E. J., H. G. Aaslestad, T. C. Ouimet, and W. D. Rupp. 1997. Sabia virus incident at Yale University. Am. Ind. Hyg. Assoc. J. 58:51-53. [DOI] [PubMed] [Google Scholar]
  • 45.Gershon, R. R., D. Vlahov, S. A. Felknor, D. Vesley, P. C. Johnson, G. L. Declos, and L. R. Murphy. 1995. Compliance with universal precautions among health care workers at three regional hospitals. Am. J. Infect. Control 23:225-236. [DOI] [PubMed] [Google Scholar]
  • 46.Gorbach, S. L. 1978. Recombinant DNA: an infectious disease perspective. J. Infect. Dis. 137:615-623. [DOI] [PubMed] [Google Scholar]
  • 47.Grimm, D., P. Staeheli, M. Hufbauer, I. Koerner, L. Martínez-Sobrido, A. Solórzano, A. García-Sastre, O. Haller, and G. Kochs. 2007. Replication fitness determines high virulence of influenza A virus in mice carrying functional Mx1 resistance gene. Proc. Natl. Acad. Sci. USA 104:6806-6811. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Grist, N. R., and J. A. N. Emslie. 1991. Infections in British clinical laboratories, 1988-1989. J. Clin. Pathol. 44:667-669. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Gruner, E., E. Bernasconi, R. L. Galeazzi, D. Buhl, R. Heinzle, and D. Nadal. 1994. Brucellosis: an occupational hazard for medical laboratory personnel. Report of five cases. Infection 22:33-36. [DOI] [PubMed] [Google Scholar]
  • 50.Guibourdenche, M., J. P. Darchis, A. Boisivon, E. Collatz, and J. Y. Riou. 1994. Enzyme electrophoresis, sero- and subtyping, and outer membrane protein characterization of two Neisseria meningitidis strains involved in laboratory-acquired infections. J. Clin. Microbiol. 32:701-704. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Hanson, R. P., S. E. Sulkin, E. L. Beuscher, W. M. Hammon, R. W. McKinney, and T. H. Work. 1967. Arbovirus infections of laboratory workers. Extent of problem emphasizes the need for more effective measures to reduce hazards. Science 158:1283-1286. [DOI] [PubMed] [Google Scholar]
  • 52.Harding, A. L., and K. B. Byers. 2006. Epidemiology of laboratory-associated infections, p. 53-77. In D. O. Flemming and D. L. Hunt (ed.), Biological safety: principles and practices, 4th ed. ASM Press, Washington, DC.
  • 53.Health and Safety Executive. 2007. Final report on potential breaches of biosecurity at the Pirbright site 2007. http://www.hse.gov.uk/news/archive/07aug/finalreport.pdf.
  • 54.Heidt, P. J. 1982. Bacteriological testing of a modified laminar flow microbiological safety cabinet. Antonie van Leeuwenhoek 48:373-381. [DOI] [PubMed] [Google Scholar]
  • 55.Heimeriks, K. 2007. Surveillance van arbeidsgerelateerde infectieziekten. Report LCI/CIb. RIVM, Bilthoven, The Netherlands.
  • 56.Herwaldt, B. L. 2001. Laboratory-acquired parasitic infections from accidental exposures. Clin. Microbiol. Rev. 14:659-688. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Isaacs, S. N. 2004. Working safely with vaccinia virus: laboratory technique and the role of vaccinia vaccination. Methods Mol. Biol. 269:1-14. [DOI] [PubMed] [Google Scholar]
  • 58.Jackson, R. J., A. J. Ramsay, C. D. Christensen, S. Beaton, D. F. Hall, and I. A. Ramshaw. 2001. Expression of mouse interleukin-4 by a recombinant ectromelia virus suppresses cytolytic lymphocyte responses and overcomes genetic resistance to mousepox. J. Virol. 75:1205-1210. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Jones, L., S. Ristow, T. Yilma, and B. Moss. 1986. Accidental human vaccination with vaccinia virus expressing nucleoprotein gene. Nature 319:543. [DOI] [PubMed] [Google Scholar]
  • 60.Kahn, L. H. 2004. Biodefense research: can secrecy and safety coexist? Biosecur. Bioterror. 2:81-85. [DOI] [PubMed] [Google Scholar]
  • 61.Kaiser, J. 2007. Accidents spur a closer look at risks at biodefense labs. Science 317:1852-1854. [DOI] [PubMed] [Google Scholar]
  • 62.Kaiser, J. 2007. Lawmakers worry that lab expansion poses risks. Science 318:182. [DOI] [PubMed] [Google Scholar]
  • 63.Kao, A. S., D. A. Ashford, M. M. McNeil, N. G. Warren, and R. C. Good. 1997. Descriptive profile of tuberculin skin testing programs and laboratory-acquired tuberculosis infections in public health laboratories. J. Clin. Microbiol. 35:1847-1851. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Käppeli, O., and L. Auberson. 1997. The science and intricacy of environmental safety evaluations. Trends Biotechnol. 9:342-349. [DOI] [PubMed] [Google Scholar]
  • 65.Keatley, K. L. 2000. Controversy over genetically modified organisms: the governing laws and regulations. Qual. Assur. 8:33-36. [DOI] [PubMed] [Google Scholar]
  • 66.Kennedy, D. A., and C. H. Collins. 2000. Microbiological safety cabinets: selection, installation, testing, and use. Br. J. Biomed. Sci. 57:330-337. [PubMed] [Google Scholar]
  • 67.Khan, S. A., R. Stratford, T. Wu, N. Mckelvie, T. Bellaby, Z. Hindle, K. A. Sinha, S. Eltze, P. Mastroeni, D. Pickard, G. Dougan, S. N. Chatfield, and F. R. Brennan. 2003. Salmonella typhi and S. typhimurium derivatives harbouring deletions in aromatic biosynthesis and Salmonella pathogenicity island-2 (SPI-2) genes as vaccines and vectors. Vaccine 21:538-548. [DOI] [PubMed] [Google Scholar]
  • 68.Kimman, T. G., A. L. Gielkens, K. Glazenburg, L. Jacobs, M. C. de Jong, W. A. Mulder, and B. P. Peeters. 1995. Characterization of live pseudorabies virus vaccines. Dev. Biol. Stand. 84:89-96. [PubMed] [Google Scholar]
  • 69.Knudsen, S., P. Saadbye, L. H. Hansen, A. Collier, B. L. Jacobsen, J. Schlundt, and O. H. Karlstrom. 1995. Development and testing of improved suicide functions for biological containment of bacteria. Appl. Environ. Microbiol. 61:985-991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Kobasa, D., A. Takada, K. Shinya, M. Hatta, P. Halfmann, S. Theriault, H. Suzuki, H. Nishimura, K. Mitamura, N. Sugaya, T. Usui, T. Murata, Y. Maeda, S. Watanabe, M. Suresh, T. Suzuki, Y. Suzuki, H. Feldmann, and Y. Kawaoka. 2004. Enhanced virulence of influenza A viruses with the haemagglutinin of the 1918 pandemic virus. Nature 431:703-707. [DOI] [PubMed] [Google Scholar]
  • 71.Kochi, S. K., K. P. Killeen, and U. S. Ryan. 2003. Advances in the development of bacterial vector technology. Expert Rev. Vaccines 2:31-43. [DOI] [PubMed] [Google Scholar]
  • 72.Kost, T. A., J. P. Condreay, and C. A. Mickelson. 2000. Biosafety and viral gene transfer factors, p. 579-597. In D. O. Flemming and D. L. Hunt (ed.), Biological safety: principles and practices, 3rd ed. ASM Press, Washington, DC.
  • 73.Kristoffersen, P., G. B. Jensen, K. Gerdes, and J. Piskur. 2000. Bacterial toxin-antitoxin gene system as containment control in yeast cells. Appl. Environ. Microbiol. 66:5524-5526. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 74.Kruse, R. H. 1962. Potential aerogenic laboratory hazards of Coccidioides immitis. Am. J. Clin. Pathol. 37:150-158. [DOI] [PubMed] [Google Scholar]
  • 75.Laquerre, S., W. F. Goins, S. Moriuchi, T. J. Oligino, D. M. Krisky, P. Marconi, M. K. Soares, J. B. Cohen, J. C. Glorioso, and D. J. Fink. 1999. Gene transfer tool: herpes simplex virus vectors, p. 173-208. In T. Friedmann (ed.), The development of gene therapy. Cold Spring Harbor Laboratory, Cold Spring Harbor, NY.
  • 76.Leifer, E., D. J. Gocke, and H. Bourne. 1970. Lassa fever, a new virus disease of man from West Africa. II. Report of a laboratory-acquired infection treated with plasma from a person recently recovered from the disease. Am. J. Trop. Med. Hyg. 19:677-679. [DOI] [PubMed] [Google Scholar]
  • 77.Lennartz, K., L. Mengli, M. Flasshove, T. Moritz, and U. Kirstein. 2005. Improving the biosafety of cell sorting by adaptation of a cell sorting system to a biosafety cabinet. Cytometry 66A:119-127. [DOI] [PubMed] [Google Scholar]
  • 78.Lewis, F. M., E. Chernak, E. Goldman, Y. Li, K. Karem, I. K. Damon, R. Henkel, C. Newbern, P. Ross, and C. C. Johnson. 2006. Ocular vaccinia infection in laboratory worker, Philadelphia, 2004. Emerg. Infect. Dis. 12:134-137. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 79.Li, L., J. Gu, X. Shi, E. Gong, X. Li, H. Shao, X. Shi, H. Jiang, X. Gao, D. Cheng, L. Guo, H. Wang, X. Shi, P. Wang, Q. Zhang, and B. Shen. 2005. Biosafety level 3 laboratory for autopsies of patients with severe acute respiratory syndrome: principles, practices, and prospects. Clin. Infect. Dis. 41:815-821. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80.Lim, P. L., A. Kurup, G. Gopalakrishna, K. P. Chan, C. W. Wong, L. C. Ng, S. Y. Se-Thoe, L. Oon, X. Bai, L. W. Stanton, Y. Ruan, L. D. Miller, V. B. Vega, L. James, P. L. Ooi, C. S. Kai, S. J. Olsen, B. Ang, and Y. S. Leo. 2004. Laboratory-acquired severe acute respiratory syndrome. N. Engl. J. Med. 350:1740-1745. [DOI] [PubMed] [Google Scholar]
  • 81.Lim, W., and D. N. C. Tsang. 2006. Laboratory containment of SARS virus. Ann. Acad. Med. Singapore 35:354-360. [PubMed] [Google Scholar]
  • 82.Lisieux, T., M. Coimbra, E. S. Nassar, M. N. Burattini, L. T. de Souza, I. Ferreira, I. M. Rocco, A. P. da Rosa, P. F. Vasconcelos, F. P. Pinheiro, et al. 1994. New arenavirus isolated in Brazil. Lancet 343:391-392. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 83.Loeb, M., I. Zando, M. C. Orvidas, A. Bialachowski, D. Groves, and J. Mahoney. 2003. Laboratory-acquired vaccinia infection. Can. Commun. Dis. Rep. 29:134-136. [PubMed] [Google Scholar]
  • 84.Low, K. B., M. Ittensohn, T. Le, J. Platt, S. Sodi, M. Amoss, O. Ash, E. Carmichael, A. Chakraborty, J. Fischer, S. L. Lin, X. Luo, S. I. MillerI, L. Zheng, I. King, J. M. Pawelek, and D. Bermudes. 1999. Lipid A mutant Salmonella with suppressed virulence and TNFalpha induction retain tumor-targeting in vivo. Nat. Biotechnol. 17:37-41. [DOI] [PubMed] [Google Scholar]
  • 85.Lucero, N. E., and F. Sineriz. 2005. The Argentine experience in enhancing biosafety through good laboratory practices. Asian Biotechnol. Dev. Rev. 8:99-120. [Google Scholar]
  • 86.Macher, J. M., and M. W. First. 1984. Effects of airflow rates and operator activity on containment of bacterial aerosols in a class II safety cabinet. Appl. Environ. Microbiol. 48:481-485. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 87.Marshall, B. M., H. Shin-Kim, D. Perlov, and S. B. Levy. 1999. Release of bacteria during the purge cycles of steam-jacketed sterilizers. Br. J. Biomed. Sci. 56:247-252. [PubMed] [Google Scholar]
  • 88.Memish, Z. A., and M. W. Mah. 2001. Brucellosis in laboratory workers at a Saudi Arabian hospital. Am. J. Infect. Control 29:48-52. [DOI] [PubMed] [Google Scholar]
  • 89.Mempel, M., G. Isa, N. Klugbauer, H. Meyer, G. Wildi, J. Ring, F. Hofmann, and H. Hofmann. 2003. Laboratory acquired infection with recombinant vaccinia virus containing an immunomodulating construct. J. Investig. Dermatol. 120:356-358. [DOI] [PubMed] [Google Scholar]
  • 90.Michael, A., R. Stratford, S. Khan, A. Dalgleish, and H. Pandha. 2004. Novel strains of Salmonella typhimurium as potential vectors for gene delivery. FEMS Microbiol. Lett. 238:345-351. [DOI] [PubMed] [Google Scholar]
  • 91.Miller, C. D., J. R. Songer, and J. F. Sullivan. 1987. A twenty-five year review of laboratory-acquired human infections at the National Animal Disease Center. Am. Ind. Hyg. Assoc. J. 48:271-275. [DOI] [PubMed] [Google Scholar]
  • 92.Mortland, K. K., and D. B. Mortland. 2003. Biosafety considerations in laboratory design. Clinical leadership and management. 17:175-177. [PubMed] [Google Scholar]
  • 93.Moss, B. 1996. Genetically engineered poxviruses for recombinant gene expression, vaccination, and safety. Poc. Natl. Acad. Sci. USA 93:11341-11348. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 94.Moussatché, N., M. Tuyama, S. E. Kato, A. P. Castro, B. Njaine, R. H. Peralta, J. M. Peralta, C. R. Damaso, and P. F. Barroso. 2003. Accidental infection of laboratory worker with vaccinia virus. Emerg. Infect. Dis. 9:724-726. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 95.Munro, K., J. Lanser, and R. Flower. 1999. A comparative study of methods to validate formaldehyde decontamination of biological safety cabinets. Appl. Environ. Microbiol. 65:873-876. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 96.Nahrstedt, H., J. Waldeck, M. Grone, R. Eichstadt, J. Feesche, and F. Meinhardt. 2005. Strain development in Bacillus licheniformis: construction of biologically contained mutants deficient in sporulation and DNA repair. J. Biotechnol. 119:245-254. [DOI] [PubMed] [Google Scholar]
  • 97.Nalca, A., A. W. Rimoin, S. Bavari, and C. A. Whitehouse. 2005. Reemergence of monkeypox: prevalence, diagnostics, and countermeasures. Clin. Infect. Dis. 41:1765-1771. [DOI] [PubMed] [Google Scholar]
  • 98.Nelson, C. H. 2001. Risk perception, behavior, and consumer response to genetically modified organisms. Toward understanding American and European public reaction. Am. Behav. Sci. 44:1371-1388. [Google Scholar]
  • 99.Nikiforov, V. V., I. U. I. Turovskiĭ, P. P. Kalinin, L. A. Akinfeeva, L. R. Katkova, V. S. Barmin, E. I. Riabchikova, N. I. Popkova, A. M. Shestopalov, V. P. Nazarov, et al. 1994. A case of a laboratory infection with Marburg fever. Zh. Mikrobiol. Epidemiol. Immunobiol. 3:104-106. (In Russian.) [PubMed] [Google Scholar]
  • 100.Normile, D. 2003. SARS experts want labs to improve safety practices. Science 302:31. [DOI] [PubMed] [Google Scholar]
  • 101.Normile, D. 2004. Infectious diseases. Second lab accident fuels fears about SARS. Science 303:26. [DOI] [PubMed] [Google Scholar]
  • 102.Normile, D. 2004. Severe acute respiratory syndrome: lab accidents prompt calls for new containment program. Science 304:1223-1225. [DOI] [PubMed] [Google Scholar]
  • 103.Noviello, S., R. Gallo, M. Kelly, R. J. Limberger, K. DeAngelis, L. Cain, B. Wallace, and N. Dumas. 2004. Laboratory-acquired brucellosis. Emerg. Infect. Dis. 10:1848-1850. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 104.Nulens, E., and A. Voss. 2002. Laboratory diagnosis and biosafety issues of biological warfare agents. Clin. Microbiol. Infect. 8:455-466. [DOI] [PubMed] [Google Scholar]
  • 105.Openshaw, P. J., W. H. Alwan, A. H. Cherrie, and F. M. Record. 1991. Accidental infection of laboratory worker with recombinant vaccinia virus. Lancet 338:459. [DOI] [PubMed] [Google Scholar]
  • 106.O'Reilly, D. R., L. K. Mileer, and V. A. Luckow. 1994. Baculovirus expression vectors: a laboratory manual. Oxford University Press, New York, NY.
  • 107.Osborne, R., T. Durkin, H. Shannon, E. Dornan, and C. Hughes. 1999. Performance of open fronted microbiological safety cabinets: the value of operator protection tests during routine servicing. J. Appl. Microbiol. 86:962-970. [DOI] [PubMed] [Google Scholar]
  • 108.Paoletti, E. 1996. Applications of poxvirus vectors to vaccination: an update. Proc. Natl. Acad. Sci. USA 93:349-353. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 109.Peeters, B., A. Bouma, T. de Bruin, R. Moormann, A. Gielkens, and T. Kimman. 1994. Non-transmissible pseudorabies virus gp50 mutants: a new generation of safe live vaccines. Vaccine 12:375-380. [DOI] [PubMed] [Google Scholar]
  • 110.Perfetto, S. P., D. R. Ambrozak, R. A. Koup, and M. Roederer. 2003. Measuring containment of viable infectious cell sorting in high-velocity cell sorters. Cytometry 52A:122-130. [DOI] [PubMed] [Google Scholar]
  • 111.Perfetto, S. P., D. R. Ambrozak, M. Roederer, and R. A. Koup. 2004. Viable infectious cell sorting in a BSL-3 facility. Methods Mol. Biol. 263:419-424. [DOI] [PubMed] [Google Scholar]
  • 112.Phillips, G. B. 1965. Microbiological hazards in the laboratory. I. Control. J. Chem. Educ. 42A:43-48. [DOI] [PubMed] [Google Scholar]
  • 113.Phillips, G. B., and R. S. Runkle. 1967. Laboratory design for microbiological safety. Appl. Microbiol. 15:378-389. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 114.Pike, R. M. 1976. Laboratory-associated infections: summary and analysis of 3921 cases. Health Lab. Sci. 13:105-114. [PubMed] [Google Scholar]
  • 115.Pike, R. M. 1978. Past and present hazards of working with infectious agents. Arch. Pathol. Lab. Med. 102:333-336. [PubMed] [Google Scholar]
  • 116.Pike, R. M. 1979. Laboratory-associated infections: incidence, fatalities, causes, and prevention. Annu. Rev. Microbiol. 33:41-66. [DOI] [PubMed] [Google Scholar]
  • 117.Richmond, J. Y., R. C. Knudsen, and R. C. Good. 1996. Biosafety in the clinical mycobacteriology laboratory. Clin. Lab. Med. 3:527-550. [PubMed] [Google Scholar]
  • 118.Romano, G., C. Pacilio, and A. Giordano. 1999. Gene transfer technology in therapy: current applications and future goals. Stem Cells 17:191-202. [DOI] [PubMed] [Google Scholar]
  • 119.Rusnak, J. M., M. G. Kortepeper, R. J. Hawley, A. O. Anderson, E. Boudreau, and E. Eitzen. 2004. Risk of occupationally acquired illnesses from biological threat agents in unvaccinated laboratory workers. Biosecur. Bioterror. 2:281-293. [DOI] [PubMed] [Google Scholar]
  • 120.Samulski, R. J., M. Sally, and N. Muzyczka. 1999. Adeno-associated viral vectors, p. 131-172. In T. Friedmann (ed.). The development of gene therapy. Cold Spring Harbor Laboratory, Cold Spring Harbor, NY.
  • 121.Sargent, E. V., and F. Gallo. 2003. Use of personal protective equipment for respiratory protection. ILAR J. 44:52-56. [DOI] [PubMed] [Google Scholar]
  • 122.Sawyer, J., A. Bennet, V. Haines, E. Elton, K. Crago, and S. Speight. 2007. The effect of microbiological containment systems on dexterity. J. Occup. Environ. Hyg. 4:166-173. [DOI] [PubMed] [Google Scholar]
  • 123.Schellekens, H. 2001. Veilig werken met microorganismen, parasieten en cellen in laboratoria en andere werkruimtes. Nederlandse Vereniging voor Microbiologie, Bilthoven, The Netherlands.
  • 124.Schmid, I., A. Kunkl, and J. K. Nicholson. 1999. Biosafety considerations for flow cytometric analysis of human immunodeficiency virus-infected samples. Cytometry 38:195-200. [DOI] [PubMed] [Google Scholar]
  • 125.Schmid, I., S. Merlin, and S. P. Perfetto. 2003. Biosafety concerns for shared flow cytometry core facilities. Cytometry A 56:113-119. [DOI] [PubMed] [Google Scholar]
  • 126.Schmid, I., C. Lambert, D. Ambrozak, G. E. Marti, D. M. Moss, S. P. Perfetto, and the International Society of Analytical Cytology. 2007. International Society for Analytical Cytology biosafety standard for sorting of unfixed cells. Cytometry A 71:414-437. [DOI] [PubMed] [Google Scholar]
  • 127.Schou, S., and A. K. Hansen. 2000. Marburg and Ebola virus infections in laboratory non-human primates: a literature review. Comp. Med. 50:108-123. [PubMed] [Google Scholar]
  • 128.Schweder, T., I. Schmidt, H. Herrmann, P. Neubauer, M. Hecker, and K. Hofmann. 1992. An expression vector system providing plasmid stability and conditional suicide of plasmid-containing cells. Appl. Microbiol. Biotechnol. 38:91-93. [DOI] [PubMed] [Google Scholar]
  • 129.Seidler, A., A. Nienhaus, and R. Driel. 2005. Review of epidemiological studies on the occupational risk of tuberculosis in low-incidence areas. Respiration 72:431-446. [DOI] [PubMed] [Google Scholar]
  • 130.Sejvar, J. J., D. Johnson, T. Popovic, J. M. Miller, F. Downes, P. Somsel, R. Weyant, D. S. Stephens, B. A. Perkins, and N. E. Rosenstein. 2005. Assessing the risk of laboratory-acquired meningococcal disease. J. Clin. Microbiol. 43:4811-4914. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 131.Sewell, D. L. 1995. Laboratory-associated infections and biosafety. Clin. Microbiol. Rev. 8:389-405. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 132.Smith, C. E., D. I. Simpson, E. T. Bowen, and I. Zlotnik. 1967. Fatal human disease from Vervet monkeys. Lancet ii:119-1121. [DOI] [PubMed] [Google Scholar]
  • 133.Sørensen, T. U., G. J. Gram, S. D. Nielsen, and J. E. Hansen. 1999. Safe sorting of GFP-transduced live cells for subsequent culture using a modified FACS vantage. Cytometry 37:284-290. [PubMed] [Google Scholar]
  • 134.Spina, N., S. Zansky, N. Dumas, and S. Kondracki. 2005. Four laboratory-associated cases of infection with Escherichia coli O157:H7. J. Clin. Microbiol. 43:2938-2939. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 135.Srinivasan, A., C. N. Kraus, D. DeShazer, P. M. Becker, J. D. Dick, L. Spacek, J. G. Bartlett, W. R. Byrne, and D. L. Thomas. 2001. Glanders in a military research microbiologist. N. Engl. J. Med. 345:256-258. [DOI] [PubMed] [Google Scholar]
  • 136.Steidler, L., S. Neirynck, N. Huyghebaert, V. Snoeck, A. Vermeire, B. Goddeeris, E. Cox, J. P. Remon, and E. Remaut. 2003. Biological containment of genetically modified Lactococcus lactis for intestinal delivery of human interleukin 10. Nat. Biotechnol. 21:785-789. [DOI] [PubMed] [Google Scholar]
  • 137.Steidler, L. 2004. Live genetically modified bacteria as drug delivery tools: at the doorstep of a new pharmacology? Expert Opin. Biol. Ther. 4:439-441. [DOI] [PubMed] [Google Scholar]
  • 138.Sulkin, S. E., and R. M. Pike. 1951. Laboratory infections. Science 114:3. [DOI] [PubMed] [Google Scholar]
  • 139.Sulkin, S. E., and R. M. Pike. 1951. Survey of laboratory-acquired infections. Am. J. Public Health Nations Health 41:769-781. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 140.Swanepoel, R. 2001. Marburg hemorrhagic fever, 1967 to the present. ProMED-mail, 7 February 2001. 20010207.0242. http://www.promedmail.org. Accessed 22 September 2007.
  • 141.Tacket, C. O., S. M. Kelly, F. Schodel, G. Losonsky, J. P. Nataro, R. Edelman, M. M. Levine, and R. Curtiss. 1997. Safety and immunogenicity in humans of an attenuated Salmonella typhi vaccine vector strain expressing plasmid-encoded hepatitis B antigens stabilized by the Asd-balanced lethal vector system. Infect. Immun. 65:3381-3385. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 142.Thompson, R. J., H. G. Bouwer, D. A. Portnoy, and F. R. Frankel. 1998. Pathogenicity and immunogenicity of a Listeria monocytogenes strain that requires d-alanine for growth. Infect. Immun. 66:3552-3561. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 143.Tumpey, T. M., C. F. Basler, P. V. Aguilar, H. Zeng, A. Solórzano, D. E. Swayne, N. J. Cox, J. M. Katz, J. K. Taubenberger, P. Palese, and A. García-Sastre. 2005. Characterization of the reconstructed 1918 Spanish influenza pandemic virus. Science 310:77-80. [DOI] [PubMed] [Google Scholar]
  • 144.Vaquero, M., P. Gomez, M. Romero, and M. J. Casal. 2003. Investigation of biological risk in mycobacteriology laboratories: a multicentre study. Int. J. Tuberc. Lung Dis. 7:879-885. [PubMed] [Google Scholar]
  • 145.VROM. 2004. Integrale versie van de Regeling genetisch gemodificeerde organismen en het Besluit genetische gemodificeerde organismen. Ministerie voor Volksgezondheid, Ruimtelijke Ordening en Milieu, Den Haag, The Netherlands.
  • 146.Walker, D., and D. Campbell. 1999. A survey of infections in United Kingdom laboratories, 1994-1995. J. Clin. Pathol. 52:415-418. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 147.Wedum, A. G. 1953. Bacteriological safety. Am. J. Public Health Nations Health 43:1428-1437. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 148.Wedum, A. G., E. Hanel, G. Briggs Phillips, and O. T Miller. 1956. Laboratory design for study of infectious disease. Am. J. Public Health Nations Health 46:1102-1113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 149.Wedum, A. G. 1961. Control of laboratory airborne infection. Bacteriol. Rev. 25:210-216. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 150.Wedum, A. G. 1964. Laboratory safety in research with infectious aerosols. Public Health Rep. 79:619-633. [PMC free article] [PubMed] [Google Scholar]
  • 151.Wedum, A. G. 1964. Airborne infection in the laboratory. Am. J. Public Health Nations Health 54:1669-1673. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 152.WHO. 2004. Laboratory biosafety manual, 3rd ed. World Health Organization, Geneva, Switzerland.
  • 153.Williams, N. R., and B. M. Cooper. 1993. Counselling of workers handling vaccinia virus. Occup. Med. (London) 43:125-127. [DOI] [PubMed] [Google Scholar]
  • 154.Williams-Smith, H. 1978. Is it safe to use Escherichia coli in recombinant DNA experiments? J. Infect. Dis. 137:655-660. [DOI] [PubMed] [Google Scholar]
  • 155.Willshaw, G. A., J. Thirlwell, A. P. Jones, S. Parry, R. L. Salmon, and M. Hickey. 1994. Vero cytotoxin-producing Escherichia coli O157 in beefburgers linked to an outbreak of diarrhoea, haemorrhagic colitis and haemolytic uraemic syndrome in Britain. Lett. Appl. Microbiol. 19:304-307. [DOI] [PubMed] [Google Scholar]
  • 156.Wilson, D. E., and L. C. Chosewood, ed. 2007. Biosafety in microbiological and biomedical laboratories, 5th ed. CDC and NIH, Washington, DC. www.cdc.gov/od/ohs/biosfty/bmbl5/bmbl5toc.htm.
  • 157.Reference deleted.
  • 158.Wlodaver, C. G., G. J. Palumbo, and J. L. Waner. 2004. Laboratory-acquired vaccinia infection. J. Clin. Virol. 29:167-170. [DOI] [PubMed] [Google Scholar]
  • 159.Yagupsky, P., N. Peled, K. Riesenberg, and M. Banai. 2000. Exposure of hospital personnel to Brucella melitensis and occurrence of laboratory-acquired disease in an endemic area. Scand. J. Infect. Dis. 32:31-35. [DOI] [PubMed] [Google Scholar]
  • 160.Yamashita, G. M., T. Misawa, T. Komori, K. Okuzumi, and T. Takahashi. 2007. Current biosafety in clinical laboratories in Japan: report of questionnaire data obtained from clinical laboratory personnel in Japan. Kanesnhogaku Zasshi 81:39-44. [DOI] [PubMed] [Google Scholar]

Articles from Clinical Microbiology Reviews are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES