Abstract
After His-bundle ablation, a 54-year-old pacemaker-dependent patient suffered from severe presyncopal attacks one year after pacemaker replacement. The attacks were caused by ventricular inhibition due to atrial far-field sensing during paroxysmal atrial flutter.
Keywords: atrial flutter, far field P-wave sensing, inhibition
Full text
PDF



Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Ben-Zur U. M., Gross J. N., Goldberger M. H., Tilkemeier P. L., Weyman A. K., Furman S. Oversensing of pacemakers in the bipolar pacing configuration: paradoxic resolution with programming to unipolar sensing. Am Heart J. 1994 Sep;128(3):617–619. doi: 10.1016/0002-8703(94)90641-6. [DOI] [PubMed] [Google Scholar]
- Cohen S. I., Koh D., Lau S. H., Rosen K. M., Damato A. N. P loops during common and uncommon atrial flutter in man. Br Heart J. 1977 Feb;39(2):173–180. doi: 10.1136/hrt.39.2.173. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lipchenca Igor, Garrigue Stéphane, Glikson Michael, Barold S. Serge, Clémenty Jacques. Inhibition of biventricular pacemakers by oversensing of far-field atrial depolarization. Pacing Clin Electrophysiol. 2002 Mar;25(3):365–367. doi: 10.1046/j.1460-9592.2002.00365.x. [DOI] [PubMed] [Google Scholar]
- Mond H. G., Sloman J. G. The malfunctioning pacemaker system. Part II. Pacing Clin Electrophysiol. 1981 Mar;4(2):168–181. doi: 10.1111/j.1540-8159.1981.tb06540.x. [DOI] [PubMed] [Google Scholar]
- Weinstock M., DeGuia R., Daniell M., Haft J. I. Inhibition of transvenous pacing through the coronary sinus by the atrial P wave: diagnosis with the aid of isoproterenol. Chest. 1971 May;59(5):563–566. doi: 10.1378/chest.59.5.563. [DOI] [PubMed] [Google Scholar]
- Wood M. A., Moskovljevic P., Stambler B. S., Ellenbogen K. A. Comparison of bipolar atrial electrogram amplitude in sinus rhythm, atrial fibrillation, and atrial flutter. Pacing Clin Electrophysiol. 1996 Feb;19(2):150–156. doi: 10.1111/j.1540-8159.1996.tb03306.x. [DOI] [PubMed] [Google Scholar]
- van Gelder L. M., el Gamal M. I., Tielen C. H. P-wave sensing in VVI pacemakers: useful or a problem? Pacing Clin Electrophysiol. 1988 Oct;11(10):1413–1418. doi: 10.1111/j.1540-8159.1988.tb04989.x. [DOI] [PubMed] [Google Scholar]