Skip to main content
The Journal of Automatic Chemistry logoLink to The Journal of Automatic Chemistry
. 1988;10(2):95–100. doi: 10.1155/S1463924688000185

A computer-controlled potentiometric/spectrophotometric titrator

John D Stong 1
PMCID: PMC2547744  PMID: 18925194

Abstract

A laboratory computer controlled potentiometric titrator interfaced to a diode array spectrophotometer is described. The titrator consists of widely used, commercially available components; therefore, major attention is given to modes of interconnection and software implementation in data format and system control. Replicate potentiometric titrations of glycines gave a relative standard deviation in titre of 1.035% and a relative standard deviation in pH of 0.745%. Replicate spectrophotometric titrations of bromophenol blue were analysed at three wavelengths to yield pKa= 3.898 ± 0.075 (1.9% rsd).

Methods of data presentation and manipulation are presented.