Skip to main content
The Journal of Automatic Chemistry logoLink to The Journal of Automatic Chemistry
. 1988;10(3):140–143. doi: 10.1155/S1463924688000264

Automated continuous monitoring of inorganic and total mercury in wastewater and other waters by flow-injection analysis and cold-vapour atomic absorption spectrometry

S E Birnie 1
PMCID: PMC2547753  PMID: 18925201

Abstract

An automated continuous monitoring system for the determination of inorganic and total mercury by flow-injection analysis followed by cold-vapour atomic absorption spectrometry is described. The method uses a typical flow-injection manifold where digestion and reduction of the injected sample takes place. Mercury is removed by aeration from the flowing stream in a specially designed air-liquid separator and swept into a silica cell for absorption measurement at a wavelength of 253.7 nm. A calibration curve up to 10 μg Hg ml-1 using three different path length cells is obtained with a detection limit of 0.02 μg Hg ml-1. The sampling rate of an injection every 3 min produces 20 results per hour from a flowing stream.

Full Text

The Full Text of this article is available as a PDF (659.6 KB).


Articles from The Journal of Automatic Chemistry are provided here courtesy of Wiley