Skip to main content
The Journal of Automatic Chemistry logoLink to The Journal of Automatic Chemistry
. 1988;10(4):188–191. doi: 10.1155/S1463924688000409

Adaptation of a cold vapour mercury analyser to flow injection analysis

Celio Pasquini 1, Wilson F Jardim 1, Lourival C de Faria 1
PMCID: PMC2547765  PMID: 18925214

Abstract

Minor modifications to a Coleman MAS-50A Mercury Analyser System allowed the determination of mercury by flow injection analysis. Using sample volumes of 600 μl it was possible to analyse up to 120 samples per hour, with a detection limit of 0.2 μg. l-1 (120 pg) of mercury. The authors also report on a simple digestion procedure which replaces the time- and reagent-consuming EPA procedure, when the sample content permits.

Full Text

The Full Text of this article is available as a PDF (680.2 KB).