Skip to main content
The Journal of Automatic Chemistry logoLink to The Journal of Automatic Chemistry
. 1990;12(5):213–220. doi: 10.1155/S1463924690000281

Optimized methods to measure acetoacetate, 3-hydroxybutyrate, glycerol, alanine, pyruvate, lactate and glucose in human blood using a centrifugal analyser with a fluorimetric attachment

R Stappenbeck 1, A W Hodson 1, A W Skillen 1, L Agius 2, K G M M Alberti 2
PMCID: PMC2547834  PMID: 18925277

Abstract

Optimized methods are described for the analysis of glucose, lactate, pyruvate, alanine, glycerol, D-3-hydroxybutyrate and acetoacetate in perchloric acid extracts of human blood using the Cobas Bio centrifugal analyser. Glucose and lactate are measured using the photometric mode and other metabolites using the fluorimetric mode. The intra-assay coefficients of variation ranged from 0.7 to 4.1%, except with very low levels of pyruvate and acetoacetate where the coefficients of variation were 7.1 and 12% respectively. All seven metabolites can be measured in a perchloric acid extract of 20 μl of blood. The methods have been optimized with regard to variation in the perchloric acid content of the samples. These variations arise from the method of sample preparation used to minimize changes occurring in metabolite concentration after venepuncture.

Full Text

The Full Text of this article is available as a PDF (1.1 MB).


Articles from The Journal of Automatic Chemistry are provided here courtesy of Wiley