Assessing the ABL 500 blood gas analyser

B Gouget 1, R Andriamahatratra 1, Y Gourmelin 1, A Truchaud 1
PMCID: PMC2547913  PMID: 18924891

Abstract

The ABL 500 blood gas analyser from Radiometer has cordless electrodes and does not use a humidifier for calibrating gases. During the evaluation of the analytical performance of this instrument, the problem of p02 accuracy was approached by comparing the values obtained with two kinds of tonometry (film and bubble). An acceptable level of imprecision was demonstrated for all measured parameters. For within-run precision, with tonometry, coefficients of variation (CV) were ≆0.37% for pO2 and ≆0.52% for pCO2. A CV of 1.76% was found for day-to-day precision for both p02 and pCO2. In the linearity study, with both tonometry methods, and in the inter-instrument comparisons (the ABL was compared with the Ciba Corning 178), pO2 values obtained on the ABL 500 exhibited a slight overestimation above 150 mmHg (2.2-3.4% at 600 mmHg). This minor discrepancy is discussed with reference to the new design of the pO2 electrode, the algorithm for pO2 correction and the tonometry procedure. The results reported in this paper stress the importance of pO2 accuracy assessment for the evaluation of blood gas analysers.