A portable battery-powered flow injection monitor for the in situ analysis of nitrate in natural waters

N J Blundell 1, A Hopkins 1, P J Worsfold 1, H Casey 2
PMCID: PMC2548018  PMID: 18924971

Abstract

The design and performance of a portable, automated flow injection (FI)-based photometric monitor are described. The system is controlled by an in-house microcomputer system that enables the monitor (including a solid state detector) to operate from a 12 V battery supply. The monitor uses the cadmium reduction/diazotization method to analyse for nitrate with a linear range of 0 to 12 mg l-1 and a limit of detection of 0.05 mg l-1 (NO3-N). The hardware and software design, monitor performance and results obtained during unattended operation are presented.

Full Text

The Full Text of this article is available as a PDF (670.1 KB).