Skip to main content
The Journal of Automatic Chemistry logoLink to The Journal of Automatic Chemistry
. 1993;15(5):177–181. doi: 10.1155/S1463924693000240

Fluorimetric methods for the measurement of intermediate metabolites (lactate, pyruvate, alanine, β-hydroxybutyrate, glycerol) using a COBAS FARA centrifugal analyser

L D Monti 1, P E Sandoli 1, S Costa 1, V C Phan 1, P M Piatti 1
PMCID: PMC2548020  PMID: 18924974

Abstract

Intermediate products of the metabolism of glucose, fat and amino-acid are important in the evaluation of such metabolic disorders as diabetes mellitus, liver disease and metabolic acidosis. In the present study, methods for the measurement of intermediate metabolites (lactate, pyruvate, alanine, β-hydroxybutyrate and glycerol) have been adapted to a fast centrifugal analyzer: the COBAS FARA. Correlation coeffcients rangedfrom 0.90 to 0.99, compared to established manual spectrophotometric methods. Within-run coeffcients of variation (CVs) ranged between 2.9 and 8.8% at low levels, between 1.5 and 5.7% at medium levels and between 1.2 and 5.6% at high levels. Between-run CVs were between 4.0 and 15.0% at low levels, between 1.7 and 7.0% at medium levels and between 1.3 and 2.7% at high levels. These fluorimetric assays for the determination of intermediate metabolites on COBAS FARA (Roche) have a good sensitivity and precision, are less costly than manual methods and can be used on a routine basis.

Full Text

The Full Text of this article is available as a PDF (355.3 KB).


Articles from The Journal of Automatic Chemistry are provided here courtesy of Wiley