. 1995;17(6):213–218. doi: 10.1155/S1463924695000332

Evaluation of a random access analyser: BM/Hitachi 911

Taweesook Kanluan 1, Surapon Tangvarasittichai 1, Orathai Tangvarasittichai 1
PMCID: PMC2548101  PMID: 18925048

Abstract

The performance of Boehringer Mannheim's BM/Hitachi 911 was evaluated for three months. The mean coeffcient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1.16% (range 0.47-2.38%) and 1.35% (range 0.62-2.93,%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an eight-hour assay nor was any sample-related carry-over detected. In all cases, the regression analyses (slopes) of the results obtainedfrom BM/Hitachi 911 and 717 were between the extreme values of 0.94 and 1.05. During the three months of operation, no major problem was encountered. The BM/Hitachi 911 was found to be easily operated, to require minimal attention and simple daily maintenance during operation.