Skip to main content
The Journal of Automatic Chemistry logoLink to The Journal of Automatic Chemistry
. 1997;19(1):9–13. doi: 10.1155/S1463924697000023

Evaluation of the Dimension XL clinical chemistry system

Neel V Neelkantan 1,, John P Mizzer 1, Jack Weinacht 1, David R Thompson 1, Bulent Sener 1, Diane K Stille 1, Wei G Zhao 1, David Miller 1
PMCID: PMC2548115  PMID: 18924790

Abstract

The analytical performance of the Dimension XL clinical chemical system was evaluated. The XL is the latest addition to the Dimension family of instruments; it is a random access analyser with a throughput up to 740 tests/hour. Regression analysis of method comparison studies with Dimension AR yielded slopes of 0.93 to 1.03 and correlation coefficients ≥0.96 for 28 assays. Excellent precision performance was also observed. New instrument features of the XL are discussed.

Full Text

The Full Text of this article is available as a PDF (416.8 KB).


Articles from The Journal of Automatic Chemistry are provided here courtesy of Wiley

RESOURCES