Skip to main content
The Journal of Automatic Chemistry logoLink to The Journal of Automatic Chemistry
. 1997;19(5):169–173. doi: 10.1155/S1463924697000187

Development and analytical performance of an automated screening method for cannabinoids on the Dimension clinical chemistry system

D M Obzansky 1, E G Gorman 1, S P Kramer 1, I S Masulli 1, E A Nuzzaci 1, W F Skogen 1
PMCID: PMC2548125  PMID: 18924804

Abstract

A fully automated, random access method for the determination of cannabinoids (UTHC) was developed for the Dimension AR and XL clinical chemistry systems. The method utilizes Abuscreen ONLINE reagents and a multianalyte liquid calibrator containing 11-nor-Δ9-THC-9-carboxylic acid. Within-run and total reproducibility, determined using NCCLS protocol EP5- T2, was less than 0.6% and 1.6% CV, respectively, at all concentrations. Calibration stability was retained for at least 30 days. An extensive evaluation of non-structurally related drugs and various physiological substances indicated lack of interference in the method. No sample carry-over was observed following a specimen containing 1886 ng/ml 11-nor-Δ9-THC-9-carboxylic acid. A 99.1% agreement (N = 445 samples) was found between an EMIT based method on the aca discrete clinical analyser and the Dimension UTHC method.

Dimension clinical chemistry system and aca discrete clinical analyser are registered trademarks of Dade International.

Full Text

The Full Text of this article is available as a PDF (437.4 KB).


Articles from The Journal of Automatic Chemistry are provided here courtesy of Wiley

RESOURCES