Distinction and quantification of carry-over and sample interaction in gas segmented continuous flow analysis

Jia-Zhong Zhang 1
PMCID: PMC2548138  PMID: 18924810

Abstract

The formulae for calculation of carry-over and sample interaction are derived for the first time in this study. A scheme proposed by Thiers et al. (two samples of low concentration followed by a high concentration sample and low concentration sample) is verified and recommended for the determination of the carry-over coeffcient. The derivation demonstrates that both widely used schemes of a high concentration sample followed by two low concentration samples, and a low concentration sample followed by two high concentration samples actually measure the sum of the carry-over coeffcient and sample interaction coefficient. A scheme of three low concentration samples followed by a high concentration sample is proposed and verified for determination of the sample interaction coeffcient. Experimental results indicate that carry-over is a strong function of cycle time and a weak function of ratio of sample time to wash time. Sample dispersion is found to be a function of sample time. Fitted equations can be used to predict the carry-over, absorbance and dispersion given sample times, and wash times for an analytical system. Results clearly show the important role of intersample air segmentation in reducing carry-over, sample interaction and dispersion.

Full Text

The Full Text of this article is available as a PDF (715.2 KB).


Articles from The Journal of Automatic Chemistry are provided here courtesy of Wiley

RESOURCES