Summary
This article aims to:
provide a broad overview of the three stage process of introducing a new or modified method into the clinical diagnostic laboratory
outline general principles and strategies that can be utilised in the process of selecting a suitable method
give a practical example of how these principles and strategies can be applied.
Introduction
The introduction of a new or modified method (assay) is a core value adding activity of the clinical diagnostic laboratory. Historically, the main drivers of these activities have been clinical demand and the desire to improve the quality of an existing assay. More recently, with the ever-increasing demand on limited healthcare resources, the need to reduce assay-associated costs may now also be a factor driving the change process.
The selection of a suitable assay is an important step in the process of introducing a new or modified method into the clinical diagnostic laboratory.
The aim of this paper is to provide a broad overview of the process of introducing a new or modified method into the clinical diagnostic laboratory, primarily to define where in the overall process selection of a suitable assay fits. Following this, the paper will then examine in more detail the specific factors to be considered in the selection of a suitable assay.
Process Overview
The process of introducing a method into routine use in a clinical diagnostic laboratory is shown in Figure 1.1–3 This process may be initiated either by the introduction of a new test or because an improvement is required in the quality or cost effectiveness of an existing test.
Figure 1.
High-level overview of the process. Adapted from Westgard J. Method Validation. Selecting a method to validate.1 Westgard QC Inc. copyright 2008.
Introduction of New Tests
New putative markers are typically the result of research undertaken in public sector institutes; hospitals; established In Vitro Diagnostic (IVD) companies; pharmaceutical companies; and, more recently, from a new breed of diagnostic companies that have emerged as a result of studies on the human genome and proteome.
The caveats to be considered when introducing a new assay are i) does the result provide information of significant clinical utility; and ii) does the assay facilitate the re-capture of resources, resulting in cost savings or reinvestment of freed up resources.3
Existing Assay
A new method for an existing assay may provide the laboratory with opportunities to improve quality by expanding the linear range, reducing imprecision, removing interferences, improving turnaround time, or allowing a new scenario, e.g. POCT.
New methods for existing assays may also provide opportunities to reduce costs, for example, by transferring the test to the core laboratory’s automated platform.
It is important that the clinical requirement (in the case of a new test) or the improvement required (in the case of an existing assay) is well defined before any assays are selected.
Once selected, the process moves into the second stage, method validation. ISO 15189 states that “method validations shall be as extensive as necessary to meet the needs of the given application”.4 It is also a requirement of the standard that only validated procedures be used to confirm that the method is suitable for the intended use.
The validation phase may identify opportunities to improve method performance. Improvements could be as simple as introducing a multipoint calibration curve, or changing a secondary wavelength.
The next step in this stage is implementation of the assay and would include activities such as updating Standard Operating Procedures and incorporating them into the laboratory manual, training staff and educating clinicians.
The third and final stage is performing the test and, through the laboratory’s quality management system, the continuous monitoring of the performance in order to verify the quality of patient results.
The remainder of this paper will focus on the steps involved in the selection of a suitable assay.
Selecting a Suitable Assay
Selecting a suitable assay for the clinical laboratory is a three-step process.
Step 1: Defining the clinical need and/or improvement required
The clinical need or improvement required could be well established as a result of:
-
Clinical outcome trials
For example, the UK Prospective Diabetes study and the Diabetes Control and Complications trial that established HbA1c as the standard measure of long-term glycaemic control.
-
Regulatory requirements that set analytical performance criteria
For example, the Centers for Disease Control and Prevention (CDC) in the USA has established, through the Cholesterol Reference Method Laboratory Network, a process whereby manufacturers obtain certification for Total, HDL and LDL Cholesterol methods as having met National Cholesterol Education Programme (NCEP) bias and coefficient of variation (CV) specifications. Certified methods are publicly available on the CDC website.5
-
Recommendations of relevant professional bodies
Contemporary examples of professional bodies making recommendations with respect to the performance criteria of various assays include -
National Kidney Disease Education Programme - plasma creatinine.
US CDC and the American Heart Association - CRP
The American College of Cardiology and European Society of Cardiology - Troponin
Published Literature
Discussions with Clinicians
Step 2: Review available methods
Information on a range of methods is available from:
Published Literature
-
IVD companies
Reagent package inserts.
A number of IVD companies provide users with access to web based libraries.
User groups.
Conferences - e.g. AACB annual scientific meeting
Professional Organisations – e.g. IFCC, AACB & AACC
External Quality Assessment (EQA) Program - An EQA end of cycle report is a valuable independent guide to the performance of a range of methods in common use.
Step 3: Select methods with characteristics that best suit the laboratory’s service requirements
Method characteristics can be categorised as1:
-
Application Characteristics: These are factors that determine if a method can be implemented into a particular laboratory and would typically include things such as:
Where is the test to be run: Core Lab, Satellite Lab or at the Point of Care
Cost per test
Suitable sample types
Expected test volume
Equipment required
Turn around time required
-
Occupation, Health and Safety Issues
Operator, Patient, Environment, and Third party risk assessment
Product supply and support
Personnel requirements and training
-
Methodology Characteristics: These are factors that contribute to best performance and include:
-
Method: Is there a
IFCC recommended method; or
Reference method available.
-
Standardisation and Calibration
Metrological Traceability to certified primary reference material
-
Regulatory Requirements
Laboratory accreditation requirements
Australian Register of Therapeutic Goods (ARTG) listing
-
Other
Minimum number of steps
Ease of use
Robustness
Kit configuration
-
-
Performance Characteristics: Demonstrate how well a method performs and will include:
Precision (imprecision)
Interference (analytical specificity)
Trueness (bias)
Measuring range
Clinical validation
Reference interval validation
Combined, the method characteristics form assay boundary conditions (Figure 2).
Figure 2.

Assay boundary conditions determined by Method Characteristics.
Result
Based on the above analysis, Candidate Method 1 meets or exceeds the required boundary conditions and therefore would be the most suitable method to proceed to Stage 2 of the process.
Conclusion
The validation and implementation of a new or modified method is a resource intensive process, requiring adherence to well defined procedures in order to meet increasingly stringent laboratory accreditation requirements. Therefore, it is important that the most suitable assay is selected at the outset of the change process.
However, before a suitable method can be selected, it is essential that the clinical need or desired improvement be well defined. Only then is the clinical laboratory in a position to determine the context in which a new or modified assay is to be evaluated.
One approach to determining the context in which a new or modified method is to be assessed, which is outlined in this paper, is to identify applicable boundary conditions in terms of application, methodology and performance characteristics, required to be met in order to achieve the desired outcome. A suitable assay will be one that meets or exceeds all boundary conditions, and in so doing, becomes the ideal candidate to proceed to the second stage of the process of introducing a new method, validation and implementation.
Table.
The Table shows a worked example of how the process could be utilised to select a new Amylase method for the Core Laboratory.
Desired Outcome: Replace Saccharogenic amylase method with a Chromogenic amylase method (candidate methods 1 and 2).
| Application Characteristics | Boundary Condition | Candidate Method 1 | Candidate Method 2 | Comments |
|---|---|---|---|---|
| Availability | Core Laboratory | Suitable | Suitable | |
| Cost per Test | <8 cents | 5.0 cents | 5.5 cents | |
| Sample Type | Serum, Plasma & Urine | Yes | Yes | All acceptable. EDTA plasma not recommended (10% decrease in values) |
| Equipment Required | Validated Instrument application available | Yes | Yes | In house |
| verification required as per laboratory procedures | ||||
| Turn Around Time | Total Analysis Time less than 10 minutes (theoretical) | 8 minutes | 3 minutes | |
| OH&S | Acceptable level of risk to Operator, Patient, Environment and Third Party | Yes | Yes | Based on clinical laboratory’s risk assessment procedure |
| Product Supply and Support | Available directly from manufacturer or fully owned subsidiary | Yes | Yes | |
| Personnel requirements | No additional headcount requirements | Yes | Yes | |
|
| ||||
| Methodology Characteristics | Boundary Condition | Candidate Method 1 | Candidate Method 2 | Comments |
|
| ||||
| Method | Chromogenic, utilising a soluble p-nitrophenol (PNP) Oligosaccharide Substrate | 5 Ethylidene–G7- pNP (EPS) | 2-Chloro-4- nitrophenyl–α-D-Maltotrioside (CNP-G3) | |
| Metrological Traceability | Traceable to certified primary reference material CRM456 and IFCC reference measurement | Yes | No | Refer to traceability tree |
| Regulatory requirements | Conformité Européene marked | Yes | Yes | |
| Ease of use | Supplied ready to use | Liquid stable | Liquid stable | |
| Robustness | >18 months shelf life | 20 months from Date of Manufacture | 36 months from Date of Manufacture | |
| Kit Configuration | 400 –500 tests per kit | 425 | 450 | Based on supplied recommended procedure |
|
| ||||
| Performance Characteristics | Boundary Condition | Candidate Method 1 | Candidate Method 2 | Comments |
|
| ||||
| Repeatability | CV< 3% or SD <3 U/L | CV<1.5%
SD<2 U/L |
CV<2.5%
SD <2.5 U/L |
Package insert data |
| Intermediate Precision | CV<5% or SD<5 U/L | CV<2.5%
SD<2.5 U/L |
CV<3%
SD<3 U/L |
Package insert data |
| Analytical Specificity | Hb–No interference up to 10 g/L | 10 g/L | 10 g/L | Package insert data |
| Lipid - No interference up to 22.6 mmol/L | 22.6 mmol/L | 11.3 mmol/L | Package insert data | |
| Glucose–No interference up to 120 mmol/L | 120 mmol/L | 120 mmol/L | Package insert data | |
| Bilirubin- No interference up to 900 μmol/L | >900 μmol/L | 900 μmol/L | Package insert data | |
| Ascorbic Acid–No interference up to 2 g/L | >2 g/L | >2 g/L | Package insert data | |
| Measuring Range | Up to 2000 U/L | 4–2000 U/L | 4–1800 U/L | Package insert data |
| Lot to Lot Variability | <3 % | <3% | <3% | Information supplied by manufacturer |
| On Board Reagent Stability | >21 days | 30 days | 45 days | Package insert data |
Footnotes
Competing Interests: None declared.
References
- 1.Westgard J. Method validation. Selecting a method to validate. [(Accessed 12 March 2007)]. http://www.westgard.com/lesson20.htm.
- 2.Lumsden JH. Laboratory test method validation. Rev Med Vet. 2000;151:623–30. [Google Scholar]
- 3.O’Kane DJ, Ebert TA, Hallaway BJ, Roberts SG, Bhuiyan AK, Tenner KS. A laboratorian’s perspective on evaluation and implementation of new laboratory tests. Clin Chem. 1997;43:1771–80. [PubMed] [Google Scholar]
- 4.ISO 15189:2006(E) Medical Laboratories –Particular requirements for quality and competence. ISO, Geneva, Switzerland.
- 5.USA Centers for Disease Control and Prevention Website. [(Accessed 5 February 2008)]. http://www.cdc.gov/labstandards/crmln_clinical.htm#Introduction.

