Abstract
End tidal carbon dioxide (ETCO2) monitoring is the non‐invasive measurement of exhaled CO2. The Intensive Care Society guidelines include (ETCO2) monitoring as one of the objective standards required for monitoring patients in transport, and the American Heart Association recommends that all intubations must be confirmed by some form of ETCO2 measurement. The physiological principles and technology underlying ETCO2 measurement and the clinical indication for its use in the prehospital environment are reviewed. ETCO2 monitoring has been widely established in the prehospital environment and is of particular use for verification of endotracheal tube placement. It is non‐invasive and easy to apply to breathing circuits. The units now available are compact and rugged, with extended battery operating times, which are ideally suited for prehospital use and should be considered as an essential item for advanced airway management.
Keywords: End‐tidal, carbon dioxide, monitoring, prehospital, retrieval
End tidal carbon dioxide (ETCO2) monitoring is the non‐invasive measurement of exhaled CO2, first studied clinically by Smallhout and Kalenda in the 1970s.1 It has been used extensively in operating theatres and intensive care units for the past 25 years and increasingly in emergency departments and the prehospital setting.2,3,4,5 ETCO2 monitoring is one of the objective standards set in the Intensive Care Society guidelines 2002 for the transport of the critically ill adult,6 and is mandated by the American Heart Association in the new paediatric advanced life support guidelines, which state that all intubations must be confirmed by some form of ETCO2 measurement.7
This article reviews the physiological principles and technology underlying ETCO2 measurement and the clinical indication for its use in the prehospital environment.
PHYSIOLOGY
Aerobic cell metabolism produces CO2, which diffuses out of the cell and into the tissue capillaries, where it is transported to the right side of the heart by the venous circulation.2,8 The right ventricle pumps the venous blood to the lungs, where diffusion of carbon dioxide across the blood gas barrier of the alveolar wall, interstitial fluid, and pulmonary capillary endothelium is governed by Fick's law of diffusion (see definitions). The partial pressure of venous CO2 is 46 mmHg (6.1 kPa) and the alveolar partial pressure of CO2 is 40 mmHg (5.3 kPa) which gives a pressure gradient of only 6 mmHg (0.8 kPa).8,9 This is a much smaller gradient than that for O2 (65 mmHg/8.7 kPa), but the diffusion of both gases is almost equal, because CO2 is much more soluble and has a diffusion constant 20 times greater than O2.
When CO2 diffuses out of the lungs in to the exhaled air it can be measured by a capnometer, which measures the partial pressure or maximum concentration of CO2 at the end of exhalation. The graphic representation of expired CO2 over time is termed capnography, and the waveform produced is a capnogram.10
ETCO2 is a reflection of metabolism, circulation, and ventilation, and is usually equal to arterial CO2 if there is no ventilation perfusion mismatch. The difference is usually 2–5 mmHg lower in ETCO2 than arterial CO2 (PaCO2).11 This difference can be markedly increased if there is a derangement of either the perfusion of the lung (as in compromised blood flow to the lung secondary to hypotension12 or pulmonary embolus) or following the application of high levels of positive end expiratory pressure during mechanical ventilation, or in states that compromise ventilation of the alveoli, such as any of the lung diseases. Any conditions causing ventilation perfusion mismatch will cause the net ETCO2 to underestimate the PaCO2.13 It is important to remember this, because the majority of patients in the prehospital environment will have significant physiological changes to their haemodynamic or ventilatory status as a result of trauma or anaesthetic agents used to aid tracheal intubation,14 which may interfere with quantitative interpretation of ETCO2.
TECHNOLOGY
There are several different methods for measuring CO2 in respiratory gases but in the prehospital environment these are limited to two common methods: (a) monitors that incorporate infrared (IR) spectroscopy (either handheld or incorporated in to multimodal monitors) and (b) colorimetric detectors. IR spectroscopy works on the principle that infrared light is only absorbed by CO2 and water vapour in the exhaled gases. The water vapour is removed by dehumidification, and in the sensor unit, light from the infrared measuring source is filtered to include only the bandwidth corresponding to the absorption peak of CO2. The amount of gas absorbed at the CO2 bandwidth is compared with a known CO2 concentration from a reference sample, and the partial pressure of ETCO2 in the sample is determined.15 IR spectroscopy monitor technology allows measurement of CO2 qualitatively, semi‐qualitatively, or quantitatively. The ETCO2 is measured in pressure (torr) or volume concentration by dividing the ETCO2 in torr by atmospheric pressure.
CO2 sensors are located in line with the patient's breathing circuit (mainstream) or remote from the circuit (sidestream). Sidestream capnometers draw a continuous sample of gas from the respiratory circuit via tiny aspiration pumps for analysis. The major advantage of this method is that the patient does not require a closed breathing circuit to operate—that is, the patient does not need to be intubated and can monitored after administration of sedation or analgesia,16 for example, during a prolonged entrapment. The disadvantages of this system are that the sampling tubing can become occluded with water and secretions, and gas analysis may take fractionally longer. Mainstream capnometers use specially designed airway adapters with windows, which allow the measurement head to be placed directly in line with the breathing circuit, usually attached to, or close to, the endotracheal tube. The major advantage of this technology is almost instantaneous gas analysis (“first breath analysis”), but the disadvantage is that it requires the patient to be intubated.
Modern IR capnometers are small handheld devices that have long battery operating times and tough rubberised shells, making them particularly suited to the prehospital environment. They have been shown to function effectively within the medical air transport setting.17
Colorimetric detectors are disposable devices that attach directly to the endotracheal tube. They incorporate a pH sensitive chemical that changes from purple (ETCO2 0.03%–<0.05%) to yellow (ETCO2 2–5%), and colour cycles can be observed during inspiration and expiration if the endotracheal tube is correctly placed in the trachea. The detector has a dead space of 38 ml, which precludes its continuous use in infants, but it can be safely used in patients weighing more than 2 kg to verify endotracheal tube position.18 A dedicated paediatric device with a dead space of 3 ml does exist, which allows brief usage in infants up to 1 kg.19 The advantages of the colorimetric devices are they are light and require no calibration or batteries. However, they can be difficult to read in poor lighting and give no indication of the quality of ventilation, and once soiled with blood, gastric contents, or tracheally administered drugs, they cease to function.
CLINICAL INDICATIONS FOR PREHOSPITAL ETCO2 MONITORING
Determination of endotracheal tube placement
Intubation in the field is often performed in a less than ideal environment, with limitations on available equipment.20 Unrecognised oesophageal intubation can prove rapidly fatal, and rates as high as 17–25% have been reported in emergency airway management of non‐arrest patients.21,22 Review of the literature is inconclusive for the actual incidence of missed and unrecognised oesophageal intubation in the prehospital setting in severely injured patients.23,24,25,26,27,28,29 There are multiple methods for determining correct tube placement but no single method has been shown to be foolproof30 (see Table 1). It is thus recommended that multiple methods are employed to confirm tube position, as no single method has perfect accuracy.31 Capnography has proved a powerful adjunct, and in combination with auscultation has emerged as the gold standard in prehospital determination of tube position, as auscultation alone has not been sufficiently reliable.23
Table 1 Methods for confirming ETT placement in the prehospital setting.
Clinical |
Lateral/epigastric auscultation: equal and symmetrical breath sounds |
Symmetrical bilateral hemi‐thorax elevation on ventilation |
Reservoir bag compliance and refilling |
Direct inspection of tube passing between the cords |
Mechanical |
Lighted stylet |
Oesophageal detector device |
Colorimetric capnography |
Electronic capnography (sidestream or mainstream) |
There have been no studies addressing whether colorimetric CO2 detectors are superior to capnography for confirming endotracheal tube position exclusively in the prehospital environment; however, the current literature suggests that both are equally accurate at detecting tracheal intubation, but colorimetric detection is potentially less accurate at detecting oesophageal intubation.32 It has been demonstrated that in emergency situations where air that contains CO2 may be present in the oesophagus, colorimetric devices may falsely indicate correct tracheal intubations, due to their sensitivity to very low CO2 values.33 Animal studies have also indicated that a diet cola drink placed in the stomachs of piglets produces sufficient CO2 to give false positive ETCO2 readings when the oesophagus was intubated.34
Control of ventilation
Traumatic brain injury is a leading cause of death in young adults.35 Interventions in the prehospital phase are critical for prevention of secondary brain injury. Positron emission tomography brain imaging studies have demonstrated that cerebral blood vessels are sensitive to changes in PaCO2, and hypocapnia induced by hyperventilation (PaCO2 <34 mmHg) can lead to vasoconstriction and a reduction in global cerebral blood flow with an increase in volume of hypoperfused brain tissue, thus worsening the secondary brain injury.36 Hyperventilation strategies to reduce hypertension induced by traumatic brain injury are no longer used, and ventilation parameters should be aimed at achieving “normocapnia”. This can be adequately achieved by utilisation of a capnograph after intubation to guide ventilation parameters and thus prevent inadvertent hyperventilation.37,38
Safety during transport
As mentioned previously, ETCO2 monitoring is one of the objective standards set in the Intensive Care Society guidelines 2002 for transport of the critically ill.6 A typical primary transfer may involve moving a ventilated patient from the scene to aircraft to ambulance to emergency department, with inherent risks of circuit disconnection at every stage. Endotracheal tubes have been shown to migrate as far as 50 mm during transportation.39 Because of the high levels of ambient noise in the prehospital environment, the capnograph is a more reliable indicator of integrity of the breathing circuit than reliance on audible alarms.14
ETCO2 as a prognostic indicator
The primary interventions of cardiopulmonary resuscitation (CPR) include establishing a patent airway, ventilation, and chest compressions. If these interventions are performed optimally, a cardiac output of 17–27% can be achieved.40,41 Several animal studies have found that the higher the coronary artery perfusion pressure achieved during CPR, the more likely the resuscitation is to be successful.42,43 ETCO2 has been shown to correlate linearly with coronary artery perfusion pressure, and hence this simple and non‐invasive method can be used to measure blood flow during CPR and can indicate a return of spontaneous circulation.44 It has been demonstrated that ETCO2 values of <10 mmHg at the start of resuscitative efforts is associated with a high risk of death.45 ETCO2 measurements may have some value in decision making about when to terminate resuscitative efforts, given their prognostic value.
ETCO2 as a reflection of cardiac output has also been applied to patients with blunt trauma to determine if there is an association with outcome from prehospital trauma. Patients requiring intubation following blunt trauma with low ETCO2 values at 20 minutes had lower rates of survival, thus indicating that ETCO2 is of value in predicting outcome.46
CONCLUSION
ETCO2 monitoring has been widely established for a number of roles in the prehospital environment and is of particular use for verification of endotracheal tube placement in difficult and noisy environments. It is non‐invasive and easy to apply to breathing circuits. Manufacturers have developed compact and rugged units with extended battery operating times, which are ideally suited for prehospital use, and should be considered as an essential item of kit for any prehospital care practitioners performing advanced airway management.
DEFINITIONS
Fick's law of diffusion: relates the flow of gas across a membrane to the area and thickness of the membrane, the partial pressure difference of the gas across the membrane, and the diffusion constant of the individual gas
End tidal CO2: measurement of CO2 at the very end of expiration. This is the maximum concentration of expired CO2.
Abbreviations
CPR - cardiopulmonary resuscitation
ETCO2 - end tidal carbon dioxide
IR - infrared
PaCO2 - arterial carbon dioxide
Footnotes
Competing interests: there are no competing interests
References
- 1.Gravenstein J H S, Paulus D A, Hayes T J. Clinical indications. In: Gravenstein JS, Paulus DA. Hayes TJ, ed. Capnography in clinical practice. Stoneham, MA: Butterworth, 198943–49.
- 2.Sanders A B. Capnometry in emergency medicine. Ann Emerg Med 1989181287–1290. [DOI] [PubMed] [Google Scholar]
- 3.Santos L J, Varon J, Pic‐Aluas L.et al Practical uses of end‐tidal carbon dioxide monitoring in the emergency department. J Emerg Med 199412633–644. [DOI] [PubMed] [Google Scholar]
- 4.Helm M, Fischer S. The role of capnography in prehospital ventilation for trauma patients. Int Journ Int Care 200512124–130. [Google Scholar]
- 5.Bhende M S, LaCovey D. End‐tidal carbon dioxide monitoring in the prehospital setting. Prehosp Emerg Care 20015208–213. [DOI] [PubMed] [Google Scholar]
- 6.The Intensive Care Society Guidelines for the transport of the critically ill adult. 2002. www.ics.ac.uk/downloads/icstransport2002mem.pdf
- 7.Cummins R O, Hazinski M F. New guidelines on tracheal tube confirmation and prevention of dislodgement. Circulation 2000102(suppl 2)I380–I384. [DOI] [PubMed] [Google Scholar]
- 8.LaValle T L, Perry A G. Capnography: assessing end‐tidal CO2 levels. Dim Crit CareNurs19951470–77. [PubMed] [Google Scholar]
- 9.Trillo G, von Planta M, Kette F. ETCO2 monitoring during low flow states: clinical aims and limits. Resuscitation 1994271–8. [DOI] [PubMed] [Google Scholar]
- 10.Bhende M S. End‐tidal carbon dioxide monitoring in paediatrics: concepts and technology. J Postgrad Med 200147153–156. [PubMed] [Google Scholar]
- 11.Bhavani‐Shanker K, Moseley H, Kumar A Y.et al Capnometry and anaesthesia. Can J Anaesth 199239617–632. [DOI] [PubMed] [Google Scholar]
- 12.Askrog V F, Pender J W, Eckenhoff J E. Changes in physiological dead space during deliberate hypotension. Anaesthesiology 196425744–751. [DOI] [PubMed] [Google Scholar]
- 13.Gonzalez del Ray J A. End‐tidal CO2 monitoring. In: Henretig FM, King C, eds. Textbook of paediatric emergency procedures. Baltimore: Williams & Wilkins, 1997829–837.
- 14.Morley A P. Prehospital monitoring of trauma patients: experience of a helicopter emergency medical service. Br J Anaesth 199676726–730. [DOI] [PubMed] [Google Scholar]
- 15.Nobel J J. Carbon dioxide monitors: exhaled gas, capnographs, capnometers, end‐tidal CO2 monitors. Paediat Emerg Care 19939244–246. [PubMed] [Google Scholar]
- 16.Miner J R, Heegard W, Plummer D. End‐tidal carbon dioxide monitoring during procedural sedation. Acad Emerg Med 20029275–280. [DOI] [PubMed] [Google Scholar]
- 17.Peterson C, Budd R, Balazs K. Comparative evaluation of three end‐tidal CO2 monitors used during air medical transport. J Air Med Transp 1992117–10. [DOI] [PubMed] [Google Scholar]
- 18.Bhende M S. End‐tidal carbon dioxide monitors: are they useful in children? J Postgrad Med 19944078–82. [PubMed] [Google Scholar]
- 19.Aziz H F, Martin J B, Moore J J. The paediatric end‐tidal carbon dioxide detector role in endotracheal intubation in newborns. J Perinatol 199919110–113. [DOI] [PubMed] [Google Scholar]
- 20.Sloane C H, Vilke G M, Chan T C.et al Rapid sequence intubation in the field versus hospital in trauma patients. J Emerg Med 200019259–264. [DOI] [PubMed] [Google Scholar]
- 21.Ornato J P, Shipley J B, Racht E M.et al Multicenter study of a portable hand size colorimetric end‐tidal carbon dioxide device. Ann Emerg Med 199221518–523. [DOI] [PubMed] [Google Scholar]
- 22.Falk J L, Sayne M P. Confirmation of airway placement. Prehosp Emerg Care 19993273–278. [DOI] [PubMed] [Google Scholar]
- 23.Grmec S, Malley S. Prehospital determination of tracheal tube placement in severe head injury. Emerg Med J 200421518–520. [PMC free article] [PubMed] [Google Scholar]
- 24.Hussain L M, Redmond A D. Are prehospital deaths from accidental injury preventable? BMJ 19943081077–1080. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Omert L, Woodrow Y, Mizikowski S.et al Role of emergency medicine physician in airway management of the trauma patient. J Trauma 2001511065–1068. [DOI] [PubMed] [Google Scholar]
- 26.Papadopoulos I N, Bukis D, Karalas A.et al Preventable prehospital trauma deaths in a Hellenic urban health region. J Trauma 199641864–869. [DOI] [PubMed] [Google Scholar]
- 27.Winchel R J, Hoyt D B. Endotracheal intubation in the field improves survival in patients with severe head injury. Arch Surg 1997132592–597. [DOI] [PubMed] [Google Scholar]
- 28.Lockey D, Davies G. Survival of trauma patients who have prehospital intubation without anaesthesia or muscle relaxants: observational study. BMJ 2001323141–145. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Murray J A, Demetriades D, Berne T V.et al Prehospital intubation in patients with severe head injury. J Trauma 2000491065–1070. [DOI] [PubMed] [Google Scholar]
- 30.Cardoso M M, Banner M J, Melker R J. Portable devices used to detect endotracheal intubation during emergency situations: a review. Crit Care Med 199826957–964. [DOI] [PubMed] [Google Scholar]
- 31.Li J. Capnography alone is imperfect for endotracheal intubation during emergency intubation. J Emerg Med 200120223–229. [DOI] [PubMed] [Google Scholar]
- 32.Hogg K, Teece S. Colourimetric CO2 detector versus capnography for confirming endotracheal tube placement. Emerg Med J 200320265–266. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Puntervoll S A, Soriede E, Jacewicz W.et al Rapid detection of oesophageal intubation: take care when using colourimetric capnometry. Acta Anaesthesiol Scand 200246455–457. [DOI] [PubMed] [Google Scholar]
- 34.Qureshi S, Park K, Sturmann K.et al The effect of carbonated beverages on colourimetric end‐tidal CO2 determination. Acad Emerg Med 200071169. [PubMed] [Google Scholar]
- 35.Albanese J, Leane M, Martin C. Severe head injury with multiple trauma. In: Vincent JL, ed. Yearbook of intensive care and emergency medicine. Berlin: Springer Verlag, 2001353–375.
- 36.Winter C D, Adamides A A, Lewis P M.et al A review of the current management of severe traumatic brain injury. Surgeon 20053329–337. [DOI] [PubMed] [Google Scholar]
- 37.Helm M, Schuster R, Hauke J. Tight control of prehospital ventilation by capnography in major trauma victims. Br J Anaesth 200390327–332. [DOI] [PubMed] [Google Scholar]
- 38.Davis D P, Dunford J V, Ochs M.et al The use of quantative end‐tidal capnometry to avoid inadvertent severe hyperventilation in patients with head injury after paramedic rapid sequence intubation. J Trauma 200456808–814. [DOI] [PubMed] [Google Scholar]
- 39.Conrarady P A, Goodman C R, Lainge F. Alteration of endotracheal position. Flexion and extension of the neck. Crit Care Med 197647–12. [DOI] [PubMed] [Google Scholar]
- 40.Falk J L, Rackow E C, Weil M H. End‐tidal carbon dioxide concentration during cardiopulmonary resuscitation. N Engl J Med 1988318607–611. [DOI] [PubMed] [Google Scholar]
- 41.Gudipati C V, Weil M H, Bisera J.et al Expired carbon dioxide: a non‐invasive monitor of cardiopulmonary resuscitation. Circulation 198877234–239. [DOI] [PubMed] [Google Scholar]
- 42.Trevino R P, Bisera J, Weil M H. End‐tidal carbon dioxide as a guide to successful cardiopulmonary resuscitation. Crit Care Med 198513910–913. [DOI] [PubMed] [Google Scholar]
- 43.Bragg S. Expired PCO2 as a prognostic indicator of successful resuscitation from cardiac arrest. Ann Emerg Med 198514948–952. [DOI] [PubMed] [Google Scholar]
- 44.Steedman D J, Robertson C E. Measurement of end‐tidal CO2 concentration during cardiopulmonary resuscitation. Arch Emerg Med 19907129–134. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Levine R L, Wayne M A, Miller C L. End‐tidal carbon dioxide and outcome of out of hospital cardiac arrest. NEJM 1997337301–306. [DOI] [PubMed] [Google Scholar]
- 46.Deakin C D, Sado D M, Coats T J.et al Prehospital end‐tidal carbon dioxide concentration and outcome in major trauma. J Trauma 20045765–68. [DOI] [PubMed] [Google Scholar]