Accuracy and Reproducibility of Conformal Radiotherapy using Data from a Randomised Controlled Trial of Conformal Radiotherapy in Prostate Cancer (MRC RT01, ISRCTN47772397)

S Stanley ∗,, S Griffiths , MR Sydes , AR Moore , I Syndikus §, DP Dearnaley ; RT01 Radiographer Trial Implementation Group on behalf of all the RT01 Collaborators
PMCID: PMC2568874  PMID: 18565744

Abstract

Aims

The MRC RT01 trial used conformal radiotherapy to the prostate, a method that reduces the volume of normal tissue treated by 40–50%. Because of the risk of geographical miss, the trial used portal imaging to examine whether treatment delivery was within the required accuracy.

Material and methods

In total, 843 patients were randomly assigned to receive 64 Gy in 32 fractions over 6.5 weeks or 74 Gy in 37 fractions over 7.5 weeks. Field displacements and corrections were recorded for all imaged fractions. Displacement trends and their association with time, disease and treatment set-up characteristics were examined using univariate and multivariate analyses. A Radiographer Trial Implementation Group (RTIG) was set up to inform the quality assurance process and to promote the development of best practice.

Results

Treatment isocentre positioning was within 5 mm in every direction on 6238 (83%) of the 7535 fractions imaged. In total, 532 (81%) of 695 included patients had at least one ≥ 3mm displacement and 415 (63%) had at least one ≥ 5mm displacement. Univariate, multivariate and stepwise models of ≥ 5mm displacements showed an increased likelihood of displacement in weeks 1 and 2 with low melting point alloy (LMPA) blocks compared with multileaf collimators, film verification compared with electronic portal imaging (EPI) and increased number of fractions imaged. Except for LMPA, this was also seen for ≥ 5mm displacements in weeks 3–6.

Conclusions

Accurate conformal treatment was delivered. The use of EPI was associated with increased reported accuracy. The RTIG was a crucial part of the quality assurance process.

Key words: Conformal radiotherapy, controlled trial, portal imaging, prostate cancer, reproducibility

Introduction

Conformal radiotherapy is now a standard treatment for localised carcinoma of the prostate [1]. The MRC RT01 trial investigated the efficacy and safety of delivering dose-escalated conformal field radiotherapy for prostate cancer [2–4]. Between January 1998 and December 2001, 843 patients at 19 radiotherapy centres (17 in the UK) were randomised to receive either the standard dose (64 Gy/32 fractions) or an escalated dose (74 Gy/37 fractions). This was given as 64 Gy to the prostate and seminal vesicles (phase I) according to risk group, with or without a 10 Gy boost to the prostate only (phase II). Compared with conventional radiotherapy, conformal radiotherapy reduces the volume of normal tissue treated by 40–50% [5]. However, with this comes the risk of geographical miss due to set-up uncertainties. Therefore, regular portal imaging and image analysis was carried out to ensure that treatment delivery was within the margin of tolerance. Displacements were identified and corrections confirmed according to the protocol.

The trial quality assurance group initiated a dosimetric and geometric quality assurance review process [6]. Completion of a quality assurance questionnaire was a prerequisite to participation. The validity of the questionnaire responses was confirmed by visits to all UK centres during the trial where under experimental conditions using a phantom a low and acceptable variation in dosimetric and geometric uncertainty was indicated [7]. The quality assurance processes showed that all centres could deliver and verify conformal radiotherapy to the standard required to participate in the trial.

A Radiographer Trial Implementation Group (RTIG) involving radiographers from all UK participating centres was established to co-ordinate clinical aspects of the trial quality assurance process. RTIG roles included the implementation of radiographer-led portal imaging analysis, ensuring that data on treatment accuracy were collected according to protocol, monitoring technique accuracy in treatment delivery and developing and sharing best practice methods for the radiotherapy process. Displacement and correction data were collected on a radiographers' log case report form for each patient [2,8]. This paper uses the displacement and correction data submitted by the participating centres to assess the variability in the patient set-up and to comment on the accuracy of conformal radiotherapy treatments delivered during only phase I of trial radiotherapy, which was common to all trial patients.

Materials and Methods

Of 831 patients who started radiotherapy, 824 had radiographer logs returned. However, one participating centre used online imaging to verify and correct before each fraction and so no displacement data were submitted. Therefore, patients from this centre were excluded from these analyses and displacement and correction data for 695 patients were analysed.

Treatment Technique

The trial protocol allowed for three- or four-field techniques using either multileaf collimators (MLCs) or low melting point alloy (LMPA) shaped blocks (Table 1). All fields were to be treated daily on a linear accelerator of ≥5 MV. All patients were treated supine with locally standard immobilisation.

Table 1.

Methods used for treatment and verification by centre

Centre* Number of fields Beam modification Image type Analysis method
2 Four-field MLC EPI Software
3 Both Both Both Software
4 Four-field MLC EPI Software
7 Three-field MLC EPI Software
8 Four-field MLC EPI Software
9 Three-field MLC Film Manual
10 Three-field MLC EPI Software
11 Three-field Both EPI Software
15 Three-field LMPA Film Manual
16 Four-field MLC Film Manual
17 Three-field MLC EPI Software
18 Four-field LMPA EPI Software
19 Three-field MLC EPI Manual
20 Four-field LMPA Film Manual
22 Three-field LMPA Film Manual
23 Three-field MLC EPI Software
25 Three-field Both Film Manual
30 Four-field MLC Both Software

MLC, multileaf collimator; LMPA, low melting point alloy; EPI, electronic portal imaging.

Sites have been coded. The same codes are used in Tables 1 and 2.

Computed Tomography Planning and Safety Margins

Gross tumour volume and clinical target volume were to be defined on computed tomography scans taken at 5 mm intervals from the bottom of the sacro-iliac joints to the penile urethra (1 cm below ischial tuberosities). The clinical target volume was defined as gross tumour volume + 0.5 cm and planning target volume with a three-dimensional safety margin around the clinical target volume of 0.5–1.0 cm. Each participating centre could specify their own planning target volume margin within this range to account for local set-up uncertainties. No oral, rectal or intravenous contrast agents were allowed.

Verification Protocol, Radiographers' Log Displacement and Correction Data

Although in 1998 electronic portal imaging (EPI) was a recent innovation, used in relatively few UK centres, it is now an established method for determining set-up accuracy [9–13]. For the RT01 trial, an image-based verification protocol was devised to measure set-up displacements and corrections at regular intervals throughout the course of treatment. The radiographers' log used in a previous single-centre pilot study [14] was adapted to record displacement and correction data.

The trial imaging protocol defined a field placement tolerance of 3 mm in any field axis. Positioning errors ≥5 mm were considered unacceptable and were required to have a correction applied before the subsequent fraction being delivered. Images taken, measurements and corrections made and accuracy on the fraction after correction were recorded on the radiographers' log for each fraction imaged. Displacements were recorded for lateral, longitudinal and vertical directions from anterior/posterior and lateral/oblique views. Rotational errors were not recorded as not all centres were capable of accurately quantifying this type of error.

The trial imaging protocol also defined image frequency and megavoltage images were acquired on at least two consecutive fractions during the first week of treatment and once weekly thereafter, with repeat images after any corrections to verify the change. The timing of the image acquisition (before, during or after treatment delivery) was not defined in the protocol. All megavoltage images were compared with either a simulator film or a digitally reconstructed radiograph to determine displacements. The method of image registration used was the choice of the centre. As both film and electronic images were acquired, manual (light box) or software analysis methods were used to measure displacements.

Statistical Considerations

The direction and size of displacements and corrections were summarised according to the three cardinal axes. Fractions were classified as having a ≥3 mm or ≥5 mm displacement or correction in any one dimension (direction) or over all three dimensions. The number of fractions with and without a displacement, and the number of patients with and without displacement in a given week of treatment were summarised graphically.

It was anticipated that displacements might occur throughout treatment, that set-up errors would most probably occur in the first 2 weeks of treatment, and that corrections would prevent systematically occurring errors thereafter. Univariate and multivariate ordered logistic regression models were run for: patients with at least one ≥3 mm displacement; patients with at least one ≥5 mm displacement; data in weeks 1 and 2; data in weeks 3 onwards. The effects of disease and treatment set-up factors were investigated. The risk group, which was calculated from T-stage, differentiation and prostate-specific antigen, was a stratification factor for the trial randomisation. Therefore, this was included in the multivariate analyses and, consequently, T-stage and differentiation were not included in the multivariate analyses (prostate-specific antigen was not included in either uni- or multivariate analyses). The models considered the impact baseline bladder symptoms, which were dichotomised to none or some according to the Royal Marsden Hospital scale [3]. The treatment centre was not included as a variable in the analyses because the number of centres was too large and there were too many variables within the data from each centre. The allocated dose group was included in the univariate analyses for completeness and assurance, but excluded from the multivariate analyses: these analyses include only data from phase I of treatment, which was identically planned regardless of the allocated trial treatment (treatment differed in terms of the inclusion or not of phase II treatment). Therefore, any effect from the dose group could only be due to chance. There was no evidence of an effect in the univariate analyses for allocated dose group and the factor was not included in the multivariate model. Patients were excluded from the (ordered) logistic regression analyses where imaging was carried out using both EPI and film (n = 41) or a six-field technique (n = 1) was used. Therefore, data were available for these analyses from 657 of 695 patients.

All analyses were carried out using the statistical package Stata 9 (Stata Corporation, College Station, TX, USA).

Results

Translational Displacements

Displacements ≥3 mm were recorded for 2797 (37%) of the 7535 fractions reported. The displacements were similarly distributed across the three directions (Fig. 1) with no good evidence of a difference from zero. The absolute lateral mean displacement was 0.2 mm to the right (standard deviation 2.2 mm); the longitudinal absolute mean displacement was 0.2 mm superiorly (standard deviation 2.4 mm); and the vertical absolute mean displacement was 0.1 mm anteriorly (standard deviation 1.2 mm).

Fig. 1.

Fig. 1

Size and direction of reported corrections and displacements in imaged fractions.

Treatment was delivered within 5 mm of the planned position in every dimension for 6238 (83%) of the 7535 fractions reported, based on the position of bony landmarks. Table 2 shows data by centre. In total, 532 (81%) of 695 patients included had at least one ≥3 mm displacement and 415 (63%) had at least one ≥5 mm displacement.

Table 2.

Corrections and displacements overall

Radiotherapy treatment centre* Patients with data Total fractions reported Fractions reported per patient
Total films/images Displaced fractions reported [n (%)]
Corrected fractions reported [n (%)]
Median Quartiles 3 mm+ 5 mm+ 3 mm+ 5 mm+
2 49 653 13 12–14 1396 246 (38) 38 (6) 11 (2) 3 (0)
3 149 1614 11 9–12 3259 983 (61) 558 (35) 173 (11) 48 (3)
4 35 376 10 10–12 1193 228 (61) 137 (36) 21 (6) 8 (2)
7 22 299 14 12–15 702 102 (34) 41 (14) 34 (11) 7 (2)
8 40 481 11 10–14 968 169 (35) 68 (14) 10 (2) 2 (0)
9 14 137 10 9–11 213 63 (46) 22 (16) 11 (8) 1 (1)
10 40 433 11 10–12 1054 123 (28) 40 (9) 10 (2) 3 (1)
11 139 1518 11 9–13 3204 495 (33) 171 (11) 127 (8) 33 (2)
15 5 29 8 2–8 60 3 (10) 1 (3) 0 (0) 0 (0)
16 31 280 9 8–10 528 85 (30) 49 (18) 35 (13) 19 (7)
17 19 152 8 7–9 364 64 (42) 33 (22) 2 (1) 1 (1)
18 3 23 8 6–9 73 6 (26) 0 (0) 1 (4) 0 (0)
19 92 1028 11 10–12 3213 38 (4) 33 (3) 34 (3) 31 (3)
20 7 61 8 8–10 139 5 (8) 1 (2) 3 (5) 1 (2)
22 22 177 8 7–9 372 9 (5) 2 (1) 1 (1) 1 (1)
23 3 29 10 9–10 87 19 (66) 6 (21) 0 (0) 0 (0)
25 2 23 12 11–12 44 16 (70) 8 (35) 6 (26) 6 (26)
30 23 222 10 8–12 524 143 (64) 89 (40) 11 (5) 11 (5)
Total 695 7535 11 9–12 17 393 2797 (37) 1297 (17) 490 (7) 175 (2)

Sites have been coded. The same codes are used in Tables 1 and 2.

3 mm+ displacements include 5 mm+ displacements.

The number of fractions imaged reduced over the treatment weeks. This is also reflected in a decrease in the number of patients with fractions imaged. This cannot be explained by patients stopping treatment early. The proportion of reported displacements in a given treatment week is summarised in Fig. 2. The first two weeks had a higher reported proportion of patients with displaced fractions. A similar pattern was seen for the proportion of displacements of ≥3 mm and ≥5 mm. There were fewer corrections than displacements, but the same trend towards decreased corrections carried out by week of treatment was observed (data not shown).

Fig. 2.

Fig. 2

Four-way panel graphs of displacements by week.

Association of Disease and Treatment Set-up Characteristics with Displacements

Radiotherapy delivery methods were used in only a small number of combinations across the trial centres and associations were present between the beam modification and treatment verification methods. For example, 407 (96%) patients whose treatment was verified with EPI had beam modification with MLC, whereas this was the case for only 71 (31%) patients verified with film. As expected, the number of images taken was much higher for patients imaged with EPI (mean 27.8, standard deviation 9.0) rather than film (mean 19.6, standard deviation 5.2; P < 0.001) and also for MLC (mean 26.8, standard deviation 9.6) over LMPA (mean 21.0, standard deviation 4.4) as the beam modification method (P < 0.001).

In the univariate models of ≥3 mm reported displacements in weeks 1 and 2 (Table 3), there was evidence of an increased likelihood of a displacement being reported with higher T-stage, lower age, moderate risk group, beam modification with LMPA, treatment verification with film, four phase I fields and increasing number of imaged fractions reported. In the multivariate model of ≥3 mm reported displacements in weeks 1 and 2, lower age, moderate risk group, treatment verification with film and an increased number of fractions imaged were significant. In weeks 3–6 (Table 4), only an increasing number of fractions reported and treatment verification with film were associated with increased occurrence of at least one reported displacement.

Table 3.

Patients with one or more displacement of at least 3 mm in weeks 1–2: univariate and multivariate logistic regression models

Category* Data
Univariate models
Multivariate models
Miss n 3 mm+ displacement OR (95% CI) z P n OR (95% CI) z P
T-stage 2 655 178 1.65 (1.26, 2.17) 3.61 <0.001 645
Differentiation 0 657 179 0.81 (0.57, 1.15) −1.18 0.240
Age (quartiles) 0 657 179 0.89 (0.76, 1.04) −1.52 0.129 0.84 (0.71, 0.99) −2.11 0.035
Risk group 0 657 179 1.41 (0.98, 2.02) 1.86 0.063 1.49 (1.02, 2.19) 2.06 0.039
Allocated treatment 0 657 179 1.04 (0.74, 1.46) 0.21 0.836 1.15 (0.80, 1.64) 0.74 0.461
Dose group 0 657 179 1.00 (0.26, 3.82) 0.00 0.998
Beam modification 6 651 179 1.41 (0.94, 2.11) 1.65 0.098 1.35 (0.74, 2.44) 0.98 0.326
Treatment verification 2 655 179 1.38 (0.96, 2.00) 1.73 0.084 2.59 (1.41, 4.75) 3.08 0.002
Phase I fields 2 655 179 1.59 (1.10, 2.30) 2.49 0.013 1.40 (0.95, 2.07) 1.72 0.086
Fractions imaged 0 657 179 1.99 (1.50, 2.64) 4.80 <0.001 1.47 (1.29, 1.69) 5.63 <0.001
Baseline bladder symptoms 6 651 179 1.05 (0.74, 1.50) 0.28 0.776 1.06 (0.73, 1.54) 0.29 0.770

T-stage and differentiation were not included in the multivariate models because they were jointly represented in the risk group.

LMPA, low melting point alloy; MLC, multileaf collimator; EPI, electronic portal imaging.

Category (reference category is depicted in bold): T-stage = T1 vs T2 vs T3 Differentiation = Good vs moderate vs poor (based on differentiation or Gleason sum score, where available); Age = 46–62, 63–66, 67–70, 71 and over; Risk group = Low vs moderate (trial stratification factor); Allocated treatment = 74Gy/37 fractions vs 64 Gy/32 fractions; Beam modification = MLC vs LMPA; Treatment verification = EPI vs film; Phase I fields = three vs four (patients with six phase I fields were excluded); Total dose = <64Gy vs 64 Gy+; Number of fractions reported = 0–3 vs 4–6 vs 7+.

Table 4.

Patients with one or more displacement of at least 3mm in weeks 3–6: univariate and multivariate logistic regression models

Category* Data
Univariate models
Multivariate models
Miss n 3 mm+ displacement OR (95% CI) z P n OR (95% CI) z P
T-stage 2 639 226 1.34 (1.04, 1.72) 2.28 0.023 630
Differentiation 0 641 226 0.76 (0.55, 1.07) −1.58 0.113
Age (quartile) 0 641 226 1.01 (0.87, 1.17) 0.13 0.900 0.96 (0.82, 1.13) −0.47 0.638
Risk group 0 641 226 1.56 (1.11, 2.19) 2.54 0.011 1.32 (0.90, 1.95) 1.43 0.153
Allocated treatment 0 641 226 1.14 (0.83, 1.58) 0.81 0.420 1.13 (0.78, 1.62) 0.65 0.518
Dose group 0 641 226 0.45 (0.10, 2.16) −0.99 0.321
Beam modification 5 636 224 2.24 (1.50, 3.34) 3.97 <0.001 1.00 (0.53, 1.86) −0.01 0.991
Treatment verification 1 640 226 2.05 (1.43, 2.93) 3.91 <0.001 2.01 (1.14, 3.56) 2.4 0.016
Phase I fields 1 640 226 1.92 (1.36, 2.71) 3.70 <0.001 1.27 (0.86, 1.88) 1.22 0.222
Fractions imaged 0 641 226 3.98 (2.97, 5.33) 9.28 <0.001 1.74 (1.53, 1.98) 8.29 <0.001
Baseline bladder symptoms 6 635 226 0.97 (0.69, 1.37) −0.15 0.882 0.92 (0.63, 1.35) −0.42 0.677

T-stage and differentiation were not included in the multivariate models because they were jointly represented in the risk group.

LMPA, low melting point alloy; MLC, multileaf collimator; EPI, electronic portal imaging.

Category (reference category is depicted in bold): T-stage = T1 vs T2 vs T3; Differentiation = Good vs moderate vs poor (based on differentiation or Gleason sum score, where available); Age = 46–62, 63–66, 67–70, 71 and over; Risk group = Low vs moderate (trial stratification factor); Allocated treatment = 74Gy/37 fractions vs 64 Gy/32 fractions; Beam modification = MLC vs LMPA; Treatment verification = EPI vs film; Phase I fields = three vs four; Total dose = <64Gy vs 64 Gy+; Number of fractions reported = 0–3 vs 4–5 vs 6–8 vs 9+.

In the univariate, multivariate and stepwise models of ≥5 mm reported displacements there was evidence of an increased likelihood of a displacement across weeks 1–2 and 3–6 with film and an increased number of fractions imaged. LMPA beam modification was associated with increased displacements in all the models except multivariate weeks 3–6 (Tables 5 and 6).

Table 5.

Patients with one or more displacement of 5mm in weeks 1–2: univariate and multivariate logistic regression models

Category* Data
Univariate models
Multivariate models
Miss n 3 mm+ displacement OR (95% CI) z P n OR (95% CI) z P
T-stage 2 655 318 1.07 (0.85, 1.36) 0.59 0.558 645
Differentiation 0 657 320 0.81 (0.59, 1.12) −1.28 0.202
Age (quartile) 0 657 320 1.02 (0.88, 1.17) 0.23 0.819 0.98 (0.84, 1.14) −0.28 0.778
Risk group 0 657 320 0.98 (0.71, 1.36) −0.12 0.901 1.07 (0.75, 1.52) 0.37 0.708
Allocated treatment 0 657 320 0.83 (0.61, 1.12) −1.21 0.225 0.91 (0.66, 1.26) −0.57 0.566
Dose group 0 657 320 0.60 (0.17, 2.06) −0.82 0.414
Beam modification 6 651 319 1.72 (1.21, 2.45) 3.03 0.002 1.85 (1.09, 3.14) 2.29 0.022
Treatment verification 2 655 319 1.49 (1.08, 2.05) 2.43 0.015 2.69 (1.55, 4.68) 3.51 <0.001
Phase I fields 2 655 319 1.22 (0.89, 1.67) 1.24 0.214 1.16 (0.82, 1.63) 0.84 0.403
Fractions imaged 0 657 320 1.87 (1.45, 2.40) 4.88 <0.001 1.56 (1.38, 1.77) 6.92 <0.001
Baseline bladder symptoms 6 651 319 0.80 (0.58, 1.10) −1.40 0.161 0.78 (0.56, 1.10) −1.4 0.161

T-stage and differentiation were not included in the multivariate models because they were jointly represented in the risk group.

LMPA, low melting point alloy; MLC, multileaf collimator; EPI, electronic portal imaging.

Category (reference category is depicted in bold): T-stage = T1 vs T2 vs T3; Differentiation = Good vs moderate vs poor (based on differentiation or Gleason sum score, where available); Age = 46–62, 63–66, 67–70, 71 and over; Risk group = Low vs moderate (trial stratification factor); Allocated treatment = 74Gy/37 fractions vs 64 Gy/32 fractions; Beam modification = MLC vs LMPA; Treatment verification = EPI vs film; Phase I fields = three vs four; Total dose = <64Gy vs 64 Gy+; Number of fractions reported = 0–3 vs 4–6 vs 7+.

Table 6.

Patients with one or more displacement of 5 mm in weeks 3–6: univariate and multivariate logistic regression models

Category* Data
Univariate models
Multivariate models
Miss n 3 mm+ displacement OR (95% CI) z P n OR (95% CI) z P
T-stage 2 639 365 1.13 (0.89, 1.44) 1.01 0.313 630
Differentiation 0 641 366 0.80 (0.58, 1.11) −1.35 0.178
Age (quartile) 0 641 366 1.00 (0.87, 1.16) 0.06 0.953 0.94 (0.80, 1.11) −0.68 0.493
Risk group 0 641 366 1.40 (1.00, 1.97) 1.94 0.052 1.21 (0.81, 1.80) 0.95 0.344
Allocated treatment 0 641 366 1.06 (0.78, 1.45) 0.39 0.698 1.06 (0.74, 1.52) 0.31 0.755
Dose group 0 641 366 0.75 (0.21, 2.61) −0.46 0.649
Beam modification 5 636 363 2.99 (2.09, 4.29) 5.96 <0.001 1.10 (0.60, 2.00) 0.3 0.765
Treatment verification 1 640 366 2.94 (2.10, 4.10) 6.30 <0.001 3.12 (1.77, 5.48) 3.95 <0.001
Phase I fields 1 640 366 1.31 (0.95, 1.80) 1.65 0.099 0.77 (0.52, 1.13) −1.33 0.185
Fractions imaged 0 641 366 3.83 (2.94, 5.00) 9.92 <0.001 1.70 (1.52, 1.91) 9.11 <0.001
Baseline toxicity 6 635 366 0.88 (0.64, 1.22) −0.76 0.446 0.82 (0.56, 1.19) −1.05 0.294

T-stage and differentiation were not included in the multivariate models because they were jointly represented in the risk group.

LMPA, low melting point alloy; MLC, multileaf collimator; EPI, electronic portal imaging.

Category (reference category is depicted in bold): T-stage = T1 vs T2 vs T3; Differentiation = Good vs moderate vs poor (based on differentiation or Gleason sum score, where available); Age = 46–62, 63–66, 67–70, 71 and over; Risk group = Low vs moderate (trial stratification factor); Allocated treatment = 74Gy/37 fractions vs 64 Gy/32 fractions; Beam modification = MLC vs LMPA; Treatment verification = EPI vs film; Phase I fields = three vs four; Total dose = <64Gy vs 64 Gy+; Number of fractions reported = 0–3 vs 4–5 vs 6–8 vs 9+.

There was no evidence of an effect from the allocated trial treatment, actual dose given or the presence of baseline bladder symptoms.

Discussion

MRC RT01 was a pragmatic trial that accompanied the initiation of conformal radiotherapy for prostate cancer in many UK centres. The results of external measurements of displacements and associated corrections show that it is possible to give accurate and reproducible treatment. There may be some factors that affect accuracy.

The requirement to correct all errors ≥5 mm at or before the next fraction was applied inconsistently, as centres had differing local correction practices. This will have influenced the number of displacements and corrections reported. Only one centre corrected most of the displacements reported, although the number of displacements reported by this centre was comparatively small. In other centres, corrections were only made after two fractions showing errors of >3 or >5 mm had been measured. This would explain why some centres have more displacements recorded than corrections and is particularly relevant for weeks 2–6, as any systematic errors should have been identified and corrected during week 1 (or at least by week 2). Displacements are more useful here than corrections because displacements were more objective across the trial, whereas corrections required human interpretation and intervention: there were no consistent trial-wide guidelines on how these should be applied. The method of correcting only after errors were seen on two or more fractions is supported by several studies, suggesting that, in order to optimise the correction in terms of identifying the systematic component, repeated imaging is required [15–17]. Assessment of random and systematic errors should be included for future trials.

Local policies for the magnitude of the correction ranged between correcting 50 and 100% of the measured displacements. For this pragmatic trial it was appropriate to allow centres to be comfortable with their own verification policies as many were just starting to use EPI and associated software. To have been more specific may have resulted in poor compliance. However, for any new trial, a more detailed correction policy should be specified. Deviation from the weekly imaging protocol was identified in that the number of patients imaged in a given week was often less than the number of patients still undergoing radiotherapy. The quality assurance visits identified that at least one centre had interpreted the protocol incorrectly and had not imaged weekly during phase 1 for patients treated early in the trial.

The method used to image also had implications. Film verification increased the likelihood of at least one reported displacement compared with EPI, despite the fact that film was associated with significantly fewer displaced fractions reported than EPI and fewer images taken during treatment than EPI. The methods used to measure displacement may also have had an effect. The difficulty in the evaluation of lateral images was highlighted by RTIG. Perera et al. [18] found that human observers have difficulty in identifying displacements of <5 mm when using manual methods. It should be noted that where EPI was used in this trial, the software analysis was not based on fully automatic registration.

The displacement data are comparable with those reported in several smaller studies of prostate conformal radiotherapy reproducibility [19–27], where immobilisation methods have been compared. A broad range of displacements was seen between the participating centres and this can be linked to their practice. Further discussion on the methods used by participating centres is discussed in other RT01 publications with recommendations [6,8].

Only one centre was using clinician-led portal image analysis for the trial, although several undertook clinician review after radiographer-led analysis. For most radiographers, this trial provided an opportunity to develop skills in portal imaging and further showed that this is a role that can be successfully undertaken by radiographers [28]. The existence of RTIG provided an unprecedented opportunity for radiographers from a number of centres to discuss treatment delivery, verification and reproducibility issues and to learn from the experiences and practices of others for conformal prostate set-up, enabling developments within centres. It also showed that there is a considerable learning and preparation time and a need for ongoing work to implement new and complex practice, and the need for appropriate technology and associated user skills to be available. It is recommended that for future trials involving technical developments in radiotherapy, a radiographer group is included and this is being undertaken in the UK as part of the Academic Clinical Oncology and Radiobiology Research Network (ACORRN) project [29]. The RTIG forum allowed differences in practice to be explored and best practice was developed [6].

A time trend has been found to be significant in other much smaller studies [23,30]. This was not tested in this trial because of the structure of the available data, but the proportion of imaged fractions associated with a displacement seemed to have decreased over time. However, the clinical significance of any time trend may be questioned. Weekly imaging protocols must balance increased radiation dose and workload with the benefit of accurate treatment associated with the identification and subsequent correction of displacements. It also depends on the tolerance level applied. For a 5 mm tolerance level, the value of weekly imaging is reduced, but for 3 mm tolerance, weekly imaging is recommended.

From the (ordered) logistic regression modelling there was evidence that increased T-stage was associated with increased ≥3 mm displacements reported. We speculate that patients with more symptomatic advanced cancers may have had more difficulty in maintaining a full bladder and therefore increased difficulty in maintaining a stable position during treatment. A more manageable drinking protocol of less than 500 ml or treating with an empty bladder may improve reproducibility, but this would require further study to assess the effects on small bowel and bladder toxicity. This category of patients may be considered unsuitable for further margin reduction techniques unless image guidance is available.

Beam modification with LMPA rather than MLC was associated with increased identification of displacement. The use of LMPA and film may be associated with a longer overall fraction delivery time, so the likelihood of patient movement is increased, combining film and LMPA will compound this effect.

Another consistently important factor was the actual number of fractions reported during each period. There was clear evidence in univariate and multivariate models that an increased number of fractions imaged was statistically significantly associated with the finding of at least one displacement. The number of fractions imaged is not an entirely independent variable; finding displacements or making corrections required further imaging, affecting the total number of fractions imaged and possibly the number of further displacements reported. For these reasons the models focused on the reporting of one reported displacement or more rather than the actual number of displacements. Future trials recording displacement and correction data could pre-specify the fractions for imaging, these then being the primary focus of any analyses of displacement.

Conclusions

It is feasible to give accurate and reproducible treatment according to external measurements of displacement. The RT01 trial provided a unique opportunity to study the accuracy of treatment techniques being used for conformal radiotherapy on a large scale and showed that patient set-up uncertainties contributed the largest component of the reported errors. Moderate risk group patients were more likely to have a displacement >3 mm. The use of EPI and MLC are associated with fewer displacements reported and weekly imaging is recommended for departments using a 3 mm tolerance level for displacement correction. In future trials where set-up reproducibility is an outcome measure, it would be advantageous to use specified fractions to record displacements and to be more prescriptive in the protocol for the correction of errors to enable random and systematic errors to be assessed.

The RTIG was vital to ensuring that the trial protocol was followed and best practice developed and it is recommended that such a group is included in future similar trials. Radiographer-led assessment of treatment verifications has enabled a UK national multi-centre trial of conformal prostate radiotherapy to be accomplished with acceptably high standards of accuracy and reproducibility.

References

  • 1.National Institute for Health and Clinical Excellence . National Institute for Health and Clinical Excellence; London: 2002. NHS NICE Guidance on Cancer Services: Improving Outcomes in Urological Cancers – the Manual, 2002. [Google Scholar]
  • 2.Sydes M.R., Stephens R.J., Moore A.R. Implementing the UK Medical Research Council (MRC) RT01 trial (ISRCTN 47772397): methods and practicalities of a randomised controlled trial of conformal radiotherapy in men with localised prostate cancer. Radiother Oncol. 2004;72:199–211. doi: 10.1016/j.radonc.2004.04.007. [DOI] [PubMed] [Google Scholar]
  • 3.Dearnaley D.P., Sydes M.R., Langley R.E. The early toxicity of escalated versus standard dose conformal radiotherapy with neo-adjuvant androgen suppression for patients with localised prostate cancer: results from the MRC RT01 trial (ISRCTN47772397) Radiother Oncol. 2007;83(3):31–41. doi: 10.1016/j.radonc.2007.02.014. [DOI] [PubMed] [Google Scholar]
  • 4.Dearnaley D.P., Sydes M.R., Graham J.D. Escalated-dose versus standard-dose conformal radiotherapy in prostate cancer: first results from the MRC RT01 randomised controlled trial. Lancet Oncol. 2007;8(6):475–487. doi: 10.1016/S1470-2045(07)70143-2. [DOI] [PubMed] [Google Scholar]
  • 5.Dearnaley D.P. Radiotherapy of prostate cancer: established results and new developments. Semin Surg Oncol. 1995;11:50–59. doi: 10.1002/ssu.2980110108. [DOI] [PubMed] [Google Scholar]
  • 6.Mayles W.P.M., Moore A.R., Aird E.G.A. Questionnaire based quality assurance for the RT01 trial of dose escalation in conformal radiotherapy for prostate cancer (ISRCTN 47772397) Radiother Oncol. 2004;73:199–207. doi: 10.1016/j.radonc.2004.08.017. [DOI] [PubMed] [Google Scholar]
  • 7.Moore A.R., Stanley S., Bidmead A.M. Quality assurance of dosimetry in UK MRC RT01 trial of dose escalation using conformal radiotherapy in prostate cancer (ISRCTN 47772397) Clin Oncol. 2003;15(2):S8–S9. doi: 10.1016/j.radonc.2004.08.017. [DOI] [PubMed] [Google Scholar]
  • 8.Griffiths S., Stanley S., Sydes M., RT01 Radiographers Group on behalf of all the RT01 Collaborators Recommendations on best practice for radiographer set up of conformal radiotherapy treatment for patients with prostate cancer: experience developed during the MRC RT01 trial (ISRTCN 47772397) J Radiother Pract. 2005;4:107–117. [Google Scholar]
  • 9.Hurkmans C.W., Remeijer P., Lebesque J.V. Set-up verification using portal imaging: review of current clinical practice. Radiother Oncol. 2001;58:105–120. doi: 10.1016/s0167-8140(00)00260-7. [DOI] [PubMed] [Google Scholar]
  • 10.Mubata C.D., Bidmead A.M., Ellingham L.M. Portal imaging protocol for radical dose-escalated radiotherapy treatment of prostate cancer. Int J Radiat Oncol Biol Phys. 1998;40:221–231. doi: 10.1016/s0360-3016(97)00551-8. [DOI] [PubMed] [Google Scholar]
  • 11.De Neve W., Van den Heuvel F., Coghe M. Interactive use of portal imaging in pelvic irradiation. Int J Radiat Oncol Biol Phys. 1993;25:517–524. doi: 10.1016/0360-3016(93)90075-7. [DOI] [PubMed] [Google Scholar]
  • 12.Greer P.B., Mortesen T.M., Rad D.T. Comparison of two methods for anterior posterior isocenter localization in pelvic radiotherapy using electronic portal imaging. Int J Radiat Oncol Biol Phys. 1998;41:1193–1199. doi: 10.1016/s0360-3016(98)00160-6. [DOI] [PubMed] [Google Scholar]
  • 13.Bel A., Vos P.H., Rodrigus P.T. High-precision prostate cancer irradiation by clinical application of an off-line patient setup verification procedure, using portal imaging. Int J Radiat Oncol Biol Phys. 1996;35:321–332. doi: 10.1016/0360-3016(95)02395-x. [DOI] [PubMed] [Google Scholar]
  • 14.Dearnaley D.P., Hall E., Lawrence D. Phase III pilot study of dose escalation using conformal radiotherapy in prostate cancer: PSA control and side effects. Br J Cancer. 2005;92:488–498. doi: 10.1038/sj.bjc.6602301. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Crook J.M., Raymond Y., Salhani D. Prostate motion during standard radiotherapy as assessed by fiducial markers. Radiother Oncol. 1995;37:35–42. doi: 10.1016/0167-8140(95)01613-l. [DOI] [PubMed] [Google Scholar]
  • 16.De Boer H.C.J., Heijmen B.J.M. A protocol for the reduction of systematic patient setup errors with minimal portal imaging workload. Int J Radiat Oncol Biol Phys. 2001;50:1350–1365. doi: 10.1016/s0360-3016(01)01624-8. [DOI] [PubMed] [Google Scholar]
  • 17.Denham J.W., Dally M.J., Hunter K. Effective decision making following a portal film: the results of a pilot study. Int J Radiat Oncol Biol Phys. 1993;26:869–876. doi: 10.1016/0360-3016(93)90503-n. [DOI] [PubMed] [Google Scholar]
  • 18.Perera T., Moseley J., Munro P. Subjectivity in interpretation of portal films. Int J Radiat Oncol Biol Phys. 1999;45(2):529–534. doi: 10.1016/s0360-3016(99)00204-7. [DOI] [PubMed] [Google Scholar]
  • 19.Baumert B.G., Zagralioglu O., Davis J.B. The use of a leg holder immobilisation device in 3D-conformal radiation therapy of prostate cancer. Radiother Oncol. 2002;65:47–52. doi: 10.1016/s0167-8140(02)00169-x. [DOI] [PubMed] [Google Scholar]
  • 20.Bentel G.C., Marks L.B., Sherouse G.W. The effectiveness of immobilisation during prostate irradiation. Int J Radiat Oncol Biol Phys. 1995;31(1):143–148. doi: 10.1016/0360-3016(94)00351-K. [DOI] [PubMed] [Google Scholar]
  • 21.Bieri S., Mirabell R., Nouet P. Reproducibility of conformal radiation therapy in localised carcinoma of the prostate without rigid immobilisation. Radiother Oncol. 1996;38:223–230. doi: 10.1016/0167-8140(95)01699-6. [DOI] [PubMed] [Google Scholar]
  • 22.Dawson L., Mah K., Franssen E. Target position variability throughout prostate radiotherapy. Int J Radiat Oncol Biol Phys. 1998;42(5):1155–1161. doi: 10.1016/s0360-3016(98)00265-x. [DOI] [PubMed] [Google Scholar]
  • 23.El-Gayed A.A., Bel A., Vijlbrief R. Time trend of patient setup deviations during pelvic irradiation using electronic portal imaging. Radiother Oncol. 1993;26:162–171. doi: 10.1016/0167-8140(93)90098-s. [DOI] [PubMed] [Google Scholar]
  • 24.Hanley J., Lumley M.A., Mageras G.S. Measurement of patient positioning errors in three-dimensional conformal radiotherapy of the prostate. Int J Radiat Oncol Biol Phys. 1997;37(2):435–444. doi: 10.1016/s0360-3016(96)00526-3. [DOI] [PubMed] [Google Scholar]
  • 25.Rosenthal S.A., Roach M., Goldsmith B.J. Immobilisation improves the reproducibility of patient positioning during 6-field conformal radiation therapy for prostate carcinoma. Int J Radiat Oncol Biol Phys. 1993;27:921–926. doi: 10.1016/0360-3016(93)90469-c. [DOI] [PubMed] [Google Scholar]
  • 26.Soffen E.M., Hanks G.E., Hwang C.C., Chu J.C.H. Conformal static field therapy for low volume low grade prostate cancer with rigid immobilisation. Int J Radiat Oncol Biol Phys. 1991;20:141–146. doi: 10.1016/0360-3016(91)90150-3. [DOI] [PubMed] [Google Scholar]
  • 27.Song P.Y., Washington M., Vaida F. A comparison of four patient immobilisation devices in the treatment of prostate cancer patients with three dimensional conformal radiotherapy. Int J Radiat Oncol Biol Phys. 1996;34(1):213–219. doi: 10.1016/0360-3016(95)02094-2. [DOI] [PubMed] [Google Scholar]
  • 28.Suter B., Shoulders B., MacClean M. Machine verification radiographs: an opportunity for role extension. Radiography. 2000;6:245–251. [Google Scholar]
  • 29.Academic Clinical Oncology and Radiobiology Research Network (ACORRN) www.acorrn.org [DOI] [PubMed]
  • 30.Rudat V., Scraube P., Oetzel Combined error of patient positioning variability and prostate motion uncertainty in 3D conformal radiotherapy of localised prostate cancer. Int J Radiat Oncol Biol Phys. 1996;35(5):1027–1034. doi: 10.1016/0360-3016(96)00204-0. [DOI] [PubMed] [Google Scholar]