Skip to main content
Critical Care logoLink to Critical Care
. 2008 Jul 7;12(4):421. doi: 10.1186/cc6928

Applying logic to pulmonary artery catheter use

Horst E Kunig 1,, Michael R Pinsky 2, Lothar Engelmann 3
PMCID: PMC2575551  PMID: 18620551

Mansour and colleagues recommend not routinely using the pulmonary artery catheter to guide hemodynamic management in the intensive care unit, because the perceived benefits are largely intangible [1]. Pulmonary artery catheter monitoring of the right ventricular ejection fraction (RVef) and of the right ventricular end-diastolic volume (EDV), however, reflects powerful yet underutilized relationships that assess right ventricular performance.

Since the cardiac output equals the product of the RVef, the EDV and the heart rate, one can assess the RVef to EDV relations as direct measures of right ventricular performance. A series of RVef, EDV and heart rate combinations can give the same cardiac output (Figure 1); monitoring or targeting cardiac output alone ignores this reality. For example, in hypovolemia the EDV is low and the RVef is increased, whereas in right ventricular failure the opposite is true. Furthermore, resuscitation from hypovolemia will increase the EDV and decrease the RVef [2], while reversal of cor pulmonale will decrease the EDV and increase the RVef.

Figure 1.

Figure 1

Right ventricular ejection fraction, right ventricular end-diastolic volume and heart rate relationship with cardiac output. CO, cardiac output; EDV, right ventricular end-diastolic volume; HR, heart rate; RVef, right ventricular ejection fraction.

No monitoring device other than the pulmonary artery catheter can continually assess these variables. These concepts are often used in the cardiac surgery suite, but are rarely codified for diagnosis and management. Monitoring devices can improve outcome only if they are coupled to treatments that improve outcome. Applying the above logic to previously proposed treatment protocols, such as using the mixed venous oxygen saturation [3], may therefore be the first step toward defining pulmonary artery catheter utility [4].

Abbreviations

EDV = right ventricular end-diastolic volume; RVef = right ventricular ejection fraction.

Competing interests

The authors declare that they have no competing interests.

See related journal club critique by Mansour et al., http://ccforum.com/content/12/1/301

References

  1. Mansour W, Milbrandt EB, Emlet LL. PAC in FACCT: time to PAC it in? Crit Care. 2008;12:301. doi: 10.1186/cc6767. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Dhainaut JF, Pinsky MR, Nouria S, Slomka F. Right ventricular function in human sepsis: a thermodilution study. Chest. 1997;112:1043–1049. doi: 10.1378/chest.112.4.1043. [DOI] [PubMed] [Google Scholar]
  3. Pinsky MR, Vincent J-L. Let us use the pulmonary artery catheter correctly and only when we need it. Crit Care Med. 2005;33:1119–1122. doi: 10.1097/01.CCM.0000163238.64905.56. [DOI] [PubMed] [Google Scholar]
  4. Vincent JL, Pinsky MR, Sprung C, Levy M, Marini JJ, Payen D, Rhodes A, Takala J. The pulmonary artery catheter: in medio virtus. Crit Care Med. [DOI] [PubMed]