Abstract
Two-photon-enhanced dendritic nanoprobes are being developed for two-photon (2P) laser scanning microscopy of oxygen [1]. In these molecular constructs, phosphorescence of metalloporphyrins is coupled to two-photon absorption (2PA) of electronically separate antenna dyes via intramolecular Förster-type resonance energy transfer (FRET). In the originally developed probes, competing electron transfer (ET) between the antennae and the long-lived triplet states of metalloporphyrins partially quenched the phosphorescence, reducing the probe's sensitivity and dynamic range. The rate of such ET can be reduced by tuning the redox potentials of the chromophores. In order to identify the optimal metalloporphyrin-2P antenna pairs, we performed screening of several phosphorescent Pt porphyrins (FRET acceptors) and 2P dyes (FRET donors) using dynamic quenching of phosphorescence. Phosphorescence lifetimes of Pt porphyrins were measured as a function of the dye concentration in organic solutions. The obtained Stern-Volmer quenching constants were correlated with the corresponding ET driving forces (ΔGET), calculated using the Rehm-Weller equation. FRET-pairs with minimal quenching rates were identified. The developed approach allows convenient screening of candidate-compounds for covalent assembly of 2P-enhanced triplet nanodevices. Systematic electrochemical measurements in a series of Pt porphyrins with varying peripheral substitution and conjugation pathways are presented.
Keywords: Phosphorescence, Oxygen, Porphyrin, Two-photon, Electron transfer
1. Introduction
Molecular devices capable of generation of triplet states upon two-photon (2P) excitation have been the focus of several recent studies, including construction of 2P agents for photodynamic therapy [2, 3, 4, 5] and 2P phosphorescent oxygen nanosensors [1, 6]. Phosphorescent probes used in biological oxygen measurements [7] are based on metalloporphyrins, whose 2P absorption (2PA) cross-sections (σ2) are very small, and their ability to generate emissive triplet states upon 2P excitation is unacceptably low for consideration in imaging applications [1, 8]. 2P-enhanced phosphorescent probes make use of the Förster resonance energy transfer (FRET) as a coupling mechanism between the absorption by an array of 2P-chromophores (2P antenna) and the excitation of a metalloporphyrin (core). The latter is followed by the intersystem crossing (isc) within the porphyrin and the emission of phosphorescence [1].
One problem inherent to the design of FRET-based 2PA phosphorescent probes is quenching of the triplet states by photoinduced electron transfer (ET), involving the metalloporphyrin and the 2P antenna [6]. Metalloporphyrin triplet states are very long-lived (tens to hundreds of microseconds), and even slow intramolecular quenching processes can be deleterious to their emission. For example, phosphorescence quantum yields of some of the earlier reported probes were reduced by as much as 50–100 times due to intramolecular ET's [1, 6].
Tuning intramolecular distances between the core and the 2P antennae provides an effective mechanism for manipulating the rates of ET and FRET in order to increase the sensor performance [6]. This strategy is general and can be applied even in those cases when the ET is strongly favored thermodynamically. However, by far the most effective way to eliminate the unwanted quenching processes is to cancel their driving forces by adjusting the redox potentials of the participating chromophores.
One approach to evaluate potential 2PA-antenna/phosphorecent core pairs in photoinduced ET reactions is based on dynamic quenching of phosphorescence. Dynamic quenching [9, 10] of long-lived triplet states [11, 12] by freely diffusing quenching species in solution is a well-known phenomenon. Practical aspects of dynamic quenching as analytical technique, e.g. for studies of protein structure and dynamics, were extensively reviewed [9, 10, 13, 14]. In particular, quenching of excited states of aromatic molecules via electron and energy transfer was studied in detail in 1960’s–1970’s, when the fundamentals of quenching kinetics were established [15, 16, 17]. Quenching by photoinduced ET [17] has been extensively researched in relation to solar energy conversion and biological photosynthetic systems [18].
In a dynamic quenching experiment, the rate of ET can be estimated by measuring the dependence of the excited state lifetime on the concentration of quencher, unless the ET rate is greater than the rate of diffusion. Dynamic quenching is especially well suited for measuring triplet quenching by the ET because of inherently high sensitivity of long phosphorescence lifetimes (~10−3–10−5 s) to the presence of quenching species in the environment [19]. In the present work we used dynamic quenching of phosphorescence to study ET reactions between sets of phosphorescent Pt porphyrins and 2P dyes in order to identify FRET pairs with minimal ET driving forces for construction of covalent 2P-enhanced nanosensors.
2. Experimental
Materials
Pt porphyrin 3 (PtTPP) was obtained from Frontier Scientific, Inc. Pt porphyrins 1, 2, 4–11 were synthesized by inserting platinum into the corresponding free bases in refluxing benzonitrile using the method proposed by Buchler et al. [20]. Non-extended free-base porphyrins for syntheses of PtP's 1, 2, 4–8 were prepared using the Lindsey method [21]. Free-base porphyrins for syntheses of PtTBP's 9–11 were prepared using the method developed earlier [22].
Dye A (AF-50) was a courtesy of Dr. L.-S. Tan (AFRL/MLBP). Dyes C and D were synthesized using the method described by Blanchard-Desce et al. [23] and Ulman et al. [24] with minor modifications. Dyes B, [24] E [1] and F [25] were prepared as described previously.
Detailed description of the synthetic procedures and characterization data are available in Supporting Information.
Spectroscopic and electrochemical measurements
Optical spectra were recorded on a Perkin-Elmer Lambda 35 UV-Vis spectrophotometer. Steady state fluorescence and phosphorescence measurements were performed on a SPEX Fluorolog-2 spectrofluorometer (Jobin-Yvon Horiba). Emission spectra were obtained using solutions with no more than 0.1 OD in the absorbance maxima. Typically, at the excitation wavelength, the absorbance of a sample was 0.05–0.06 OD. Quartz fluorometric cells (Starna, Inc, 1 cm optical path length) were used in linear optical experiments. Emission quantum yields were measured relative to the fluorescence of Rhodamine B in MeOH (ϕfl=0.5) [26, 27]. Time resolved phosphorescence measurements were performed using an in house constructed phosphorometer [28] modified for time-domain operation. The excitation sources in this instrument are 1W LED's (Hewlett Packard, λmax=530 nm and λmax=635 nm for excitation of PtP's and PtTBP's respectively), whose output is controlled by a 333 kHz D/A board (MIO-16, National Instruments, Inc.).
2PA cross-sections of compounds were determined by the relative emission method, using Rhodamine B in MeOH as a standard (λex=840 nm, σ2≈210 GM) [29, 30]. 2PA measurements were conducted using a system based on a femtosecond Ti:Sapphire oscillator (Coherent Mira 900, λmax=840 nm, 110 fs, 76 MHz rep. rate). The output of the laser was passed through a half-wave plate and a polarizing cube, so that the excitation power could be adjusted continuously, and directed into the objective (Nikon FLUOR X40, 170 µm focal distance, NA 1.3) of an inverted microscope (Nikon, Diaphot 300). The power of the excitation beam was measured before it entered the microscope using an optical power meter (Molectron). The sample solution was placed in a cylindrical cuvette, one side of which was made out of a thin microscope cover slip. The emission was collected by the objective, passed through a dichroic beam splitter (Nikon 835dcsp) and a short-pass cut-off filter (sp785nm, RazorEdge, Semrock) and directed into a monochromator (Acton SP150). The output of the monochromator was projected on a N2-cooled CCD camera (Roper Scientific) and recorded using WinSpec32 software (Roper Scientific). The 2P origin of the absorption was confirmed in each case by plotting quadratic dependence of the emission signal on the excitation power and selecting the power (about 10 mW).
The oxidation and reduction potentials were measured on a EG&G Potentiostat (Model 173) at 298 K. A three-electrode system was used, which consisted of a glassy carbon or platinum disk working electrode, a platinum wire counter electrode and a saturated calomel reference electrode (SCE). The SCE electrode was separated from the bulk of the solution by a fritted-glass bridge of low porosity that contained the solvent supporting electrolyte mixture. The solvent for the electrochemical experiments, absolute N,N-dimethylformamide (DMF, Fluka Chemical Co.), and the supporting electrolyte, tetra-n-butylammonium perchlorate (TBAP, Fluka Chemical Co.) were used as received without further purification. Half-wave potentials were calculated as E1/2= (Epa + Epc)/2 at a scan rate of 0.1 V/s and are referenced to the Fc/Fc+ (Ferrocene >98%, Fluka Chemical Co.)
Quenching experiments
The titration cell for the dynamic quenching measurements consisted of a glass vial (vol. ~ 2 ml) with a polypropylene stopper, in which two stainless steel needles were inserted to provide inlet and outlet ports for Ar. The inlet needle was immersed in the solution to allow effective bubbling. Solution of a Pt porphyrin (1 ml, OD512 nm=0.1±0.01) in DMF (Fisher Scientific, Inc., Biotech grade, 99.9+%) was placed inside the vial, which was positioned inside a cuvette holder shaded from the ambient light. The excitation and emission light was delivered to and from the vial by a pair of optical fibers (3 mm Ø), positioned at the right angle with respect to one another. Phosphorescence was excited by 5 µs-long pulses (λex=530 nm for non-extended PtP’s (1–8); λex=635 nm for PtTBP’s (9–11)), followed by 1 ms measurement period. 100–200 decays were averaged to give SNR's (signal-to-noise ratio) >100. The decays were analyzed by least-squares fitting to single-exponentials.
In a typical dynamic quenching experiment, oxygen was displaced from the solution by Ar until the phosphorescence lifetime, recorded at 10 s intervals, became constant (reached a plateau). An aliquot of a solution of 2PA dye in DMF was added, and Ar bubbling was resumed until the phosphorescence lifetime reached the next plateau. For each dye concentration (5–6 in total), the mean phosphorescence lifetime was calculated as the average across the plateau; the first plateau (no quencher added) corresponded to the native phosphorescence lifetime (τ0). Stern-Volmer quenching plots (1/τ-1/τ0 νs [Q]) were used to calculate the quenching constants (kq). Each kq value was derived as the average from three independent experiments. kq values from different experiments varied within no more than 10% of the average.
3. Results and Discussion
3.1. Approach
Under conditions of pure dynamic quenching, kinetic data are described by linear Stern-Volmer plots, which relate changes in the luminescence lifetimes τ to the concentration of quencher [Q]:
| (1) |
where τ0 is the intensity and the lifetime in the absence of quencher. The observed quenching rate constants kq are determined by the rate-limiting step (such as diffusion or electron transfer); and the presence of a correlation between kq's and the ET driving forces (ΔGET) in a series of similar emitter/quencher pairs suggests that the ET is the predominant quenching mechanism [31].
The triplet state of a chromophore can be either the electron donor or the acceptor, depending on the quencher used, and the ET driving force can be estimated from the ground state redox potentials and the excitation energy according to the Rehm-Weller equation [32]:
| (2) |
Here, ΔGET is the driving force; Eox(D) and Ered(A) are the reduction and the oxidation potentials of the donor and the acceptor in their ground states, respectively; ΔE00 is the excitation energy (donor or acceptor); and w=wP−wR is the work term, consisting of the Coulombic energies of reactants (wR) and products (wP) [32]. According to (2), ΔGET will become less negative (weaker driving force) upon: (i) an increase in the oxidation potential of the donor, (ii) a decrease in the reduction potential of the acceptor; and (iii) a decrease in the excitation energy ΔE00.
The rate of ET is related to ΔGET through the Arrhenius and Marcus equations [33], and is expected to grow exponentially with an increase in |ΔGET|, assuming ΔGET<0, up until it reaches the diffusion limit [34]. Formal kinetics can be used to express the observed Stern-Volmer quenching constant in terms of the ET rate.
In this study, dynamic quenching of the phosphorescence of Pt porphyrins by a series of 2P dyes was used to evaluate relative efficiencies of photo-induced ET's in the chromophore pairs. The ET driving forces were correlated with the phosphorescence quenching rate constants in order to establish that ET is the main quenching pathway [31, 35].
3.2. Triplet chromophores
A series of phosphorescent Pt(II) meso-tetraarylporphyrins with redox potentials varying in a wide range was selected (Fig. 1). Pt porphyrins exhibit strong room temperature phosphorescence with lifetimes τ0 in the range of 40–60 µs (see Table 1).
Fig. 1.
Pt porphyrins used as triplet chromophores in dynamic quenching experiments.
Table 1.
Selected photophysical and electrochemical properties of Pt porphyrins used in this study.
| # | λabs(Q), nm a | λphos, nm a | ϕ phos a,b | τ0, µs | E00, eVc | E1ox, Va,d | E1red, Va,d |
|---|---|---|---|---|---|---|---|
| 1 | 512, 542 | 662, 727 | 0.1 | 56 | 1.87 | e | −1.35 |
| 2 | 512, 540 | 663, 730 | 0.66 | 32 | 1.87 | e | −1.53 |
| 3 | 509 539 | 669, 732 | 0.75 | 52 | 1.85 | e | −1.65 |
| 4, 4a | 509 539 | 663, 727 | 0.86 | 55, 48 | 1.87 | 0.86f | −1.66 |
| 5 | 510, 539 | 671, 731 | 0.52 | 34 | 1.85 | e | −1.68 |
| 6, 6a | 511, 540 | 683, 742 | 0.66 | 38, 39 | 1.81 | 0.79 | −1.68 |
| 7, 7a | 510, 540 | 662, 730 | 0.11 | 76, 79 | 1.87 | 0.70 | −1.91 |
| 8 | 510, 540 | 668,732 | 0.11 | 74 | 1.86 | 0.67 | −1.94 |
| 9 | 617 | 783 | 0.10 | 29 | 1.58 | 0.37 | −1.61 |
| 10 | 620 | 787 | 0.15 | 45 | 1.57 | 0.50 | −1.55 |
| 11 | 624 | 767 | 0.07 | 14 | 1.62 | 0.60 | −1.23 |
In DMF.
E00 was calculated from the blue edge of the lowest vibrational band (T10→S00) in the phosphorescence spectrum.
νs Fc/Fc+(0.50 V νs SCE in DMF).
The oxidation potentials are beyond the positive limit of DMF.
Current-voltage curve was ill-defined due to a coupled reaction of the electrooxidized species with the solvent.
Porphyrins shown in Fig. 1 form two structurally different groups: “regular” Pt tetraarylporphyrins (PtP’s, 1–8) and Pt tetraaryltetrabenzoporphyrins (PtTBP’s, 9–11). Spectral characteristics within each group vary only slightly, whereas between the groups there are large differences (Fig. 2). Due to the extended π-conjugation, the lowest energy absorption bands (S0−S1, Q-bands) of PtTBP’s (9–11) are about 100 nm red-shifted relative to those of PtP's (1–8). Accordingly, T1−S0 phosphorescent bands of PtTBP’s also have lower energies; and this might result in lower driving forces for ET's involving PtTBP triplet states. Notably, PtTBP's possess higher 2PA cross-sections than PtP's [36, 6], which potentially makes them more suitable for 2P applications.
Fig. 2.
Absorption and phosphorescence (inset) spectra of representatives of two groups of Pt porphyrins: PtP 4 and PtTBP 10. Emission was excited at 510 nm (4) and 615 nm (10).
First oxidation and reduction potentials were measured using cyclic voltammetry (see Supplementary Material for examples). In some cases (1–3, 5), the first oxidation potentials were beyond the positive potential limit of DMF, (0.8 V νs Fc/Fc+ or 1.30 V νs SCE), while for other porphyrins (4, 4a), the first oxidation was within the potential limit of the solvent, but the current-voltage curve was ill-defined due to a coupled chemical reaction of the electrooxidized species with the solvent. In the latter case, the values for the peak potentials (Epa) are listed for the scan rate of 0.1 V/s. Notably, similar potentials were measured in PhCN and were reversible by both cyclic voltammetry and thin-layer spectroelectrochemistry. For this reason, the Epa values in DMF are assured to be close to E1/2 values in PhCN.
The first reduction of each porphyrin is reversible and well-defined in DMF. Importantly, the ring reduction potentials shift negatively with a decrease in the electron-withdrawing character of the substituents in the meso-aryl rings. The same shift in the potential was observed upon oxidation (or reduction), and the electrochemically measured separation between the potentials for the first ring-centered oxidation and the first ring-centered reduction was constant for all PtP's.
3.3. 2P dyes
The following characteristics were considered when selecting 2P dyes: (i) strong fluorescence, overlapping with absorption of Pt porphyrins for maximally efficient FRET; (ii) strong 2P absorption, shifted maximally to the red relative to the phosphorescence of Pt porphyrins in order to avoid possible interference of linear absorption via direct S0→T1 mechanism [37, 6]; (iii) synthetic availability and ease of functionalization.
Absorption Q-bands of PtP’s (1–8) are situated between 500 and 540 nm (Table 1). Unfortunately, the choice of suitable 2P absorbers emitting in this wavelength range was limited, mostly because many of them exhibit strong solvatochromism [38]. These dyes fluoresce near 500 nm in non-polar media, e.g. in toluene, but in polar solvents, their emission may shift by as much as several hundred nanometers. Such strong shifts would inevitably lead to significant losses in the FRET efficiency, and that would decrease performance of the sensors, designed to operate in biological aqueous environments.
Research in the area of 2P chromophores is mostly focused on their optical properties, whereas their electrochemical potentials are rarely known. A few exceptions include well-known fluorophores, such as Rhodamine B [39] and several recently developed dyes, whose redox potentials were examined in order to explain their 2PA behavior [38b].
In the absence of detailed data on solvatochromism and redox behavior, the 2PA antennae chromophores were chosen to represent basic classes of 2PA dyes. The selected molecules are shown in Fig. 3 and their absorption and emission spectra are shown in Fig. 4.
Fig. 3.
2PA dyes used as quenchers in dynamic quenching experiments.
Fig. 4.
Linear absorption (A) and fluorescence (B) spectra of 2PA dyes. The emission spectra are scaled to represent the relative emission quantum yields. The excitation wavelengths for each dye are shown in parentheses next to the labels.
Dyes A–E fluoresce in the region of 480–550 nm (Fig. 4B) and exhibit good spectral overlap with the Q-bands of PtP's (1–8) in polar solvents (e.g. DMF, see Table 2). Dye A, known as AF-50, is a well-characterized 2PA chromophore, a member of a class of 2PA dyes developed in the Air Force research laboratories [40]. Dyes B, C, and D represent a class of push-pull stilbenoids. The derivative of coumarin-343 (C343), dye E, although a relatively weak 2P absorber (σ2=28 GM at 840 nm) (Table 2), was chosen for its high quantum yield, efficient FRET onto PtP’s [1], availability and ease of functionalization. Importantly, E is a relatively small molecule, which makes it feasible to incorporate several C343 units into a single probe molecule without perturbing its overall structure and solubility.
Table 2.
Properties of two-photon absorbing dyes.
| # | λabs, nma (log ε) | λfluo, nma | ϕfluoa,b | λ2PA, nm | σ2, GMc[ref] | E1ox, Va,d | E1red, Va,d |
|---|---|---|---|---|---|---|---|
| A | 387 (4.63) | 552 | 0.9 | 840 | 55e, [41] | 0.43 | −2.28 |
| B | 388 (4.38) | 520 | 0.1 | 800 | ~10 [40b] | 0.23 | f |
| C | 386 (4.47) | 537 | 0.5 | 770 | 90 [23] | 0.47 | −2.30 |
| 750 | 50 [42] | ||||||
| 840 | 45e | ||||||
| D | 410 (4.69) | 552 | 0.2 | 815 | 195 [23] | 0.47 | −2.34 |
| 840 | 110e | ||||||
| E | 434 (4.60) | 480 | 1.0 | 840 | 28e | 0.51 | −2.09 |
| F | 564 (4.98) | 587 | 0.2 | 840 | 210 | 0.64 | −1.21f |
In DMF.
1 GM=10−50 cm4·s·photon−1.
Measured νs Fc/Fc+(0.50 V νs SCE in DMF). Irreversible waves were observed; the corresponding peak potentials Epa and Epc are shown instead of midpoint potentials.
Measured in this work in femtosecond regime using relative fluorescence method with Rhodamine B in MeOH as a standard. See Supporting Information for details.
The reduction potential was beyond the negative limit of DMF.
Reversible reduction was observed.
Both the oxidation and reduction of most 2P dyes (except for the reduction of F) were irreversible on the electrochemical time-scale. Therefore, the peak potentials Epa and Epc (listed in Table 2) were used as approximations for the positions of the true oxidation and reduction potentials for the purpose of the Rehm-Weller formula. The reduction potentials of dyes A–E are significantly lower than those of Pt porphyrins, which is consistent with the expectation that dyes A–E quench the phosphorescence of Pt porphyrins via the reductive mechanism.
The derivative of rhodamine B (RhB), dye F, is an effective FRET-donor for PtTBP’s [6]. The first oxidation potential of F (Eox=0.64 V) is substantially higher than the potentials of all the other dyes (Table 2), while its reduction potential (Ered=−1.21 V) suggests that F is likely to play the role of the oxidant in photoinduced ET reactions with PtTBP's. The 2PA maximum of RhB is located at 840 nm (σ2=200 GM) [29, 30]. It is known that excitation of RhB-PtTBP assemblies at 840 nm can lead to linear excitation of PtTBP's due to the presence of relatively strong direct S0−T1 absorption bands in their spectra [6]; but it is feasible to use RhB-donors with porphyrins, whose triplet bands do not interfere with excitation at 840 nm.
3.4. Dynamic quenching experiments
Phosphorescence lifetimes were used to monitor quenching of porphyrin triplet states by 2P dyes. Using lifetimes, as opposed to intensities, makes it easier to discriminate between static and dynamic quenching and exclude static quenching artifacts [10, 14]. The general procedure in our experiments involved titration of dilute solutions of Pt porphyrins with 2PA dyes in N,N-dimethylformamide (DMF) in the absence of oxygen. Phosphorescence lifetimes were measured as a function of the dye concentration, and quenching constants kq were calculated as slopes of the Stern-Volmer quenching plots.
Selected Stern-Volmer plots are shown in Fig. 5, and the values of the corresponding quenching constants are given in Table 3. The first graph (Fig. 5A) illustrates quenching of the phosphorescence of a representative PtP, Pt(II) meso-tetraphenylporphyrin (PtTPP, 3), by 2P dyes A–E. Among these dyes, stilbene B is the strongest quencher and C343-derivative E is the weakest quencher. This trend holds true for most of the PtP’s (1–8) in the series, while dyes A–E practically do not quench phosphorescence of PtTBP's (Table 3).
Fig. 5.
Selected Stern-Volmer quenching plots. In all the experiments, concentration of Pt porphyrins was about 10−6−10−5M. A) Quenching of phosphorescence of PtP (3) by dyes A–F; B) Quenching of phosphorescence of PtTBP's (9–11) by dye F; C) Quenching of phosphorescence of PtP’s (1–8) by dye E; D) Quenching of phosphorescence of PtP’s 1–6 and PtTBP (10) by dye F.
Table 3.
Rate constantsa kq (M−1 s−1) for dynamic quenching of phosphorescence of Pt porphyrins by 2PA dyes A–F in DMF.
| Pt porphyrin | 2PA dye |
|||||
|---|---|---|---|---|---|---|
| A | B | C | D | E | F | |
| 1 | 4.4×108 | 1.4×109 | 5.5×107 | 1.9×108 | 1.9×108 | 1.3×109 |
| 2 | 2.3×107 | 1.6×109 | 1.3×107 | 2.7×107 | 2.1×107 | 2.7×109 |
| 3 | 6.5×106 | 3.6×108 | 7.4×106 | 1.6×107 | 1.0×106 | 3.5×109 |
| 4 | 2.8×106 | 4.2×108 | - | 1.8×107 | 6.6×105 | 1.9×109 |
| 5 | - | - | 3.9×106 | 7.5×106 | 3.7×105 | - |
| 6 | 1.7×106 | 8.8×107 | 2.9×106 | 5.7×106 | 4.7×105 | 3.1×109 |
| 7 (7a) | 8.8×105 | 9.3×106 | 1.8×106 | 5.3×106 | 6.5×104 | - |
| (3.1×104) | ||||||
| 8b | - | 7.5×106 | - | - | 6.0×104 | - |
| 9 | 0 | 0 | 0 | 0 | 0 | 2.3×108 |
| 10 | 0 | 0 | 0 | 0 | 0 | 5.3×107 |
| 11c | - | - | - | - | 0 | 5.5×106 |
Each kq value was derived as the average from three independent experiments. kq values from different experiments varied within about 10% of the average.
The quenching rates for porphyrin 8 and dyes A, C and D were not measured because these quenching experiments required very large amounts of the 2P dyes. On the other hand, dye B was found to be the strongest quencher for all the other PtP's, whereas dye E was the weakest for the PtP's with lower reduction potentials (3–7). By extrapolation, kq's for the pairs 8/B and 8/E provide the upper and the lower limits for the quenching rate constants of 8 by all the other dyes.
Phosphorescence of PtTBP 11 is quenched by dyes A–D very weakly.
Assuming that photoinduced ET is the main quenching process, such difference in the quenching behavior of PtP’s νs PtTBP’s can be rationalized by considering the effects of π-conjugation on the electronic structure of porphyrins. PtTBP’s (9–10) exhibit lower oxidation potentials E1ox and lower excitation energies E00: E1ox(PtP)>E1ox(PtTBP), E00(PtP)>E00(PtTBP) (Table 1) [43]. The latter is believed to be caused by destabilization of one of the porphyrin HOMO's and a decrease in the degree of the states mixing for TBP's [43b,c]. The decrease in the oxidation potential of TBP's νs regular porphyrins is also associated with the higher energy of the HOMO. The first reduction E1red potentials of PtTBP's and PtP's, however, are quite close: E1red(PtP)≈E1red(PtTBP).
When Pt porphyrins participate as donors in the ET with the same dye (Fig. 6A - oxidative quenching), the differences between the driving forces for similarly substituted PtP and PtTBP are expected to be small, provided the differences in E00's and E1ox's are close (E00(PtP)−E00(PtTBP)≈E1ox(PtP)−E1ox(PtTBP)) and cancel out in the Rehm-Weller formula (Eq. 2). On the other hand, if porphyrins play the role of acceptors (Fig. 6B - reductive quenching), the ET driving forces should be lower for PtTBP's than for PtP's because E1red(PtTBP)≈E1red(PtP) and E00(PtTBP)<E00(PtP).
Fig. 6.
Energy diagrams for oxidative (A) and reductive (B) quenching of phosphorescence by ET.
Interestingly, while it is expected that the quenching rates for PtP’s and PtTBP's by the oxidative mechanism (dye F) should be close, the measured constants kq for the PtP's appear to be much higher than those for PtTBP's and do not correlate with the PtP's redox potentials. This behavior suggests the presence of another efficient quenching process, which is likely to be the triplet-triplet energy transfer in RhB-PtP pairs. The T1−S0 energy gaps in PtP’s (1–8) and in RhB are quite close, i.e. 1.82–1.87 eV (662–680 nm, see Table 1) and 1.86 eV (667 nm) [44], respectively, making the triplet energy transfer feasible.
Within the PtP and PtTBP subsets, E00 values remain practically constant, and the ET driving forces change in parallel with the porphyrin’s oxidation or reduction potentials, depending on the ET direction. Graphs in Fig. 5B,C illustrate the effects of porphyrin redox potentials on the quenching rates in the two series: B) PtTBP’s - oxidative quenching by dye F; and C) PtP’s - reductive quenching by dye E. Plots in graph C reflect the fact that kq values correlate with reduction potentials (E1red) of PtP’s, and the most electron-rich porphyrins 6–8 are the least susceptible to quenching. For PtTBP’s (9–11) (graph B), dye F performs as the oxidant; and, consequently, constants kq correlate with the oxidation potentials (E1ox). It should be mentioned that T1−S0 energy gaps in PtTBP's are about 1.6 eV, making RhB←PtTBP triplet-triplet exchange unfavorable, thus leaving the ET the predominant quenching pathway.
The obtained data show that in the PtP-series quenching by the ET is minimal when the electron-rich porphyrins (e.g. 6–8) are paired with C343-derivative E. The observed rate constants are lower by as much as 105 times than those obtained for combinations of e.g. porphyrin 1, bearing eight electron-withdrawing carboxyl groups, with strongly reducing dyes such as B. Considering availability of E and its relatively small size, this coumarin-based chromophore will likely to be the primary choice for the construction of 2P-enhanced phosphorescent probes.
3.5. Kinetic model and correlation of quenching rate with ET driving force
In order to obtain evidence that the observed phosphorescence quenching was due to the ET, we correlated the values of experimental quenching constants kq with the driving forces, calculated using the Rehm-Weller formula.
Dynamic quenching of phosphorescence of Pt porphyrins by electro-active quenchers can be described as a sequence of the following reactions (Scheme 1):
Scheme 1.
In the Scheme, P is Pt porphyrin in its ground state; 3P* is Pt porphyrin in its excited triplet state; (3P*…Q) and (P−…Q+) refer to the encounter complex before and after the ET step, respectively; P− and Q+ refer to the ionic products of the ET reaction after their escape from the solvent cage. kform refers to the combined rate of formation of the triplet state 3P*, which includes the rates of absorption, fluorescence, internal conversion of the singlet state and intersystem crossing. Rate constant kphos is the actual rate of disappearance of the triplet state measured by the phosphorescence emission. This constant accounts for both the native radiative and non-radiate decays to the ground state. Constants kd (k−d), kET (k−ET) and kesc (k−esc) refer to the diffusion, electron transfer and cage escape processes and their reverses, respectively; and constant kreact refers to irreversible reaction(s) involving the ionic species P− and Q+. It should be mentioned that in Pt and Pd porphyrins, rates of intersystem crossing are usually very high [45, 37], and triplet states are formed with practically unity quantum yields [46].
Formal treatment yields the following expression for quenching constant kq (for detailed derivation see Supporting Information):
| (3) |
We further assume that: (i) no significant interaction between the reagents occurs in the encounter complex (3P*…Q), i.e. k−d≈kd, which is suggested by the fact that the emission spectra of the mixtures with the quencher are identical to those of pure Pt porphyrins; (ii) the diffusion is much faster than the emission of phosphorescence, i.e. kd>>kphos; and (iii) the rate of the cage escape is much lower than the rate of the formation of the final products, i.e. k−esc/kreact>>1. Combining assumptions (i) and (ii) into k−d>>1/τ0+kd[Q] at low quencher concentrations (Q<<1), equation (3) can be rewritten as [47]:
| (4) |
In the last formula, kET was expressed as [ke°×exp(−ΔG*ET/RT)] from the Arrhenius equation; and k−ET/kET, which is the reciprocal of the electron transfer equilibrium constant (KET), was substituted for [exp(ΔGET/RT)]. Here ke° is the frequency factor, ΔG*ET is the free energy of activation for the ET step, and ΔGET is the driving force of the ET reaction. ΔG*ET can be further expressed using the Marcus-Hush theory [33] in terms of ΔGET and reorganization energy λ:
| (5) |
Experimentally determined kq's and their decimal logarithms (log kq) were plotted against the corresponding driving forces ΔGET (Fig. 7). In calculation of ΔGET we omitted the work terms w. In polar solvents Coulombic interactions in the ion pairs are typically expected to be small [48]. In addition, it is likely that the work terms were similar in the dye pairs studied (except for some cases, see below) and, therefore, they could be omitted for the purpose of our comparison. As expected, all the dependencies of log kq on ΔGET were linear through the most of the range; and for the quenching of phosphorescence of PtP’s (1–8) by the most reducing dye B (Fig. 7A) the plot began deviating from linearity. These data support the assumption that the phosphorescence quenching in our experiments was indeed caused by the ET [17], and the comparisons drawn in Section 3.4 were appropriate.
Fig. 7.
Decimal logarithms of experimental constants kq for quenching of phosphorescence of PtP’s (1–8) by dyes B (A) and E (B), plotted against the driving force of photoinduced ET (ΔGET).
Interestingly, in the cases involving large hydrophobic dyes (A, C, and D) in combination with porphyrins 7 and 8, bearing substituents in 2,6-positions of their meso-aryl rings, the observed quenching rates were higher than expected from the Marcus plots based on the other dye pairs. It is possible that in these cases the higher rates were caused by lower work terms w, since the effective separation between the ions in the corresponding ion pairs was larger due to the steric reasons. In addition, the geometry of the encounter complexes involving bulky dyes and 2,6-substituted PtP's could be different, affecting the reorganization energy. In principle, fitting experimental points to the function given by equation (3) should produce kinetic parameters kd, k−d/ke°, ΔG*ET(0), k−d/kesc, as well as the reorganization energy λ [47]. However, such fitting in practice would be highly unstable because of the large number of parameters, their implicit dependence and noise in the data.
4. Conclusion
The results of this study demonstrate that dynamic quenching of phosphorescence is a simple and effective screening method for selecting pairs of FRET-donors and FRET-acceptors for construction of 2P-enhanced triplet devices. Several conclusions from this work are summarized below.
First, most 2PA dyes studied in this work (A–E) act as strong reducing agents with respect to Pt porphyrin triplet states. Dyes A–E were selected as representatives of larger classes of structurally related chromophores, i.e., neutral push-pull stilbenoids with various degrees of conjugation. If such dyes are to be used as 2P-antennae, they would have to be paired with porphyrins (or other triplet chromophores) with lowest possible reduction potentials to avoid quenching.
Secondly, based on the data for dyes A–D, which belong to the class of push-pull (D-π-A) dyes [38b], predictions can be made for structurally related chromophores by taking into account how they differ in donor/acceptor character and degree of conjugation. For example, it can be inferred that: 1) all D-π-D dyes containing diphenylamino groups or stronger donors will reduce triplet states of PtP's; 2) D-π-A dyes similar to the studied ones in donor character (i.e. containing same or stronger donor groups) would quench PtP, but not PtTBP, triplet states.
Thirdly, in spite of the fact that all PtP’s (1–8) were to some extent quenched by all the dyes studied, the ET rates varied by as much as 105 times from the most strongly quenched to the most weakly quenched pair. Lowering the excited state energy (E00) by varying porphyrin structure, e.g. as in the case of π-extended PtTBP’s (9–11), points toward the possibility to decrease reductive quenching of triplet states. For example, partially extended porphyrins (mono-, di-, tribenzoporphyrins) should be more difficult to reduce compared to regular PtP’s.
Finally, for electron-rich PtP's, such as 6–8, rates of quenching by E were found to be very small (~104–105 M−1s−1), suggesting that in covalent assemblies, where ET can be additionally attenuated by spatial separation, these pairs of chromophores should exhibit superior performance.
Supplementary Material
Synthetic details and characterization data for new compounds described in this paper are available in the Supporting Information.
Acknowledgements
Support of the grants EB007279 and HL081273 from the NIH USA is gratefully acknowledged. We thank Dr. L.-S. Tan (AFRL/MLBP) for providing a sample of AF-50.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Brinas RP, Troxler T, Hochstrasser RM, Vinogradov SA. J. Am. Chem. Soc. 2005;127:11851–11862. doi: 10.1021/ja052947c. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.(a) Karotki A, Kruk M, Drobizhev M, Rebane A, Nickel E, Spangler CW. IEEE J. Sel. Top. Quant. Electron. 2001;7:971–975. [Google Scholar]; (b) Drobizhev M, Karotki A, Kruk M, Mamardashvili NZ, Rebane A. Chem. Phys. Lett. 2002;361:504–512. [Google Scholar]
- 3.(a) Dichtel WR, Serin JM, Edder C, Fréchet JMJ, Matuszewski M, Tan LS, Ohulchanskyy TY, Prasad PN. J. Am. Chem. Soc. 2004;126:5380–5381. doi: 10.1021/ja031647x. [DOI] [PubMed] [Google Scholar]; (b) Oar MA, Serin JM, Dichtel WR, Fréchet JMJ. Chem. Mater. 2005;17:2267–2275. [Google Scholar]
- 4.Morone M, Beverina L, Abbotto A, Silvestri F, Collini E, Ferrante C, Bozio R, Pagani GA. Org. Lett. 2006;8:2719–2722. doi: 10.1021/ol060742i. [DOI] [PubMed] [Google Scholar]
- 5.Belfield KD, Corredor CC, Morales AR, Dessources MA, Hernandez FE. J. Fluoresc. 2006;16:105–110. doi: 10.1007/s10895-005-0008-1. [DOI] [PubMed] [Google Scholar]
- 6.Finikova OS, Troxler T, Senes A, DeGrado WF, Hochstrasser RM, Vinogradov SA. J. Phys. Chem. A. 2007;111:6977–6990. doi: 10.1021/jp071586f. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.(a) Vanderkooi JM, Maniara G, Green TJ, Wilson DF. J. Biol. Chem. 1987;262:5476–5482. [PubMed] [Google Scholar]; (b) Rozhkov V, Wilson D, Vinogradov S. Macromolecules. 2002;35:1991–1993. [Google Scholar]; (c) Rietveld IB, Kim E, Vinogradov SA. Tetrahedron. 2003;59:3821–3831. [Google Scholar]
- 8.(a) Goyan RL, Cramb DT. Photochem. Photobiol. 2000;72:821–827. doi: 10.1562/0031-8655(2000)072<0821:nitpeo>2.0.co;2. [DOI] [PubMed] [Google Scholar]; (b) Kruk M, Karotki A, Drobizhev M, Kuzmitsky V, Gael V, Rebane A. J. Lumines. 2003;105:45–55. [Google Scholar]; (c) Karotki A, Khurana M, Lepock JR, Wilson BC. Photochem. Photobiol. 2006;82:443–452. doi: 10.1562/2005-08-24-RA-657. [DOI] [PubMed] [Google Scholar]
- 9.Eftink MR, Ghiron CA. Anal. Biochem. 1981;114:199–227. doi: 10.1016/0003-2697(81)90474-7. [DOI] [PubMed] [Google Scholar]
- 10.Webber SE. Photochem. Photobiol. 1997;65:33–38. [Google Scholar]
- 11.(a) Kalyanasundaram K. London: Academic Press; 1992. Photochemistry of Polypyridine and Porphyrin Complexes; Ch. 3–4. [Google Scholar]; (b) Roundhill DM. Photochemistry and Photophysics of Metal Complexes. New York: Plenum Press; 1994. and references therein. [Google Scholar]
- 12.Ma BW, Djurovich PI, Thompson ME. Coord. Chem. Rev. 2005;249:1501–1510. [Google Scholar]
- 13.For examples see: Mersol JV, Steel DG, Gafni A. Biochemistry. 1991;30:668–675. doi: 10.1021/bi00217a012. Schlyer BD, Lau E, Maki AH. Biochemistry. 1992;31:4375–4383. doi: 10.1021/bi00133a001. Beckham S, Cook MP, Karki L, Luchsinger MM, Whitlock VR, Wu Y, Zhang QF, Schuh MD. Arch. Biochem. Biophys. 1994;310:440–447. doi: 10.1006/abbi.1994.1190. Shen CY, Kostic NM. Inorg. Chem. 1996;35:2780–2784. Cioni P, Strambini GB. J. Am. Chem. Soc. 1998;120:11749–11757. Sokerina EV, Ullmann GM, van Pouderoyen G, Canters GW, Kostic NM. J. Biol. Inorg. Chem. 1999;4:111–121. doi: 10.1007/s007750050294. Crnogorac MM, Kostic NM. Inorg. Chem. 2000;39:5028–5035. doi: 10.1021/ic9910514.
- 14.(a) Eftink MR, Shastry MCR. Fluorescence Spectroscopy. 1997;278:258–286. doi: 10.1016/s0076-6879(97)78014-5. [DOI] [PubMed] [Google Scholar]; (b) Roy S. Energetics of Biological Macromolecules, Part D. 2004;379:175–187. doi: 10.1016/S0076-6879(04)79010-2. [DOI] [PubMed] [Google Scholar]
- 15.Birks JB. London: Wiley Interscience; 1970. Photophysics of Aromatic Molecules, Ch. 9–11. and references therein. [Google Scholar]
- 16.(a) Weller A. Discussions Faraday Soc. 1959;27:28–33. [Google Scholar]; (b) Förster T. Annalen Der Physik. 1948;2:55–75. [Google Scholar]
- 17.Kavarnos GJ, Turro NJ. Chem. Rev. 1986;86:401–449. [Google Scholar]
- 18.(a) Alstrum-Acevedo JH, Brennaman MK, Meyer TJ. Inorg. Chem. 2005;44:6802–6827. doi: 10.1021/ic050904r. [DOI] [PubMed] [Google Scholar]; (b) Guldi DM. J. Phys. Chem. B. 2005;109:11432–11441. doi: 10.1021/jp050984k. [DOI] [PubMed] [Google Scholar]; (c) Fukuzumi S, Guldi DM. Electron Transfer in Chemistry, v. 2, Ch. 8. In: Balzani V, editor. Weinheim: Wiley-VCH; 2001. [Google Scholar]; (d) Balzani V, Juris A. Coord. Chem. Rev. 2001;211:97–115. [Google Scholar]; (e) Gust D, Moore TA. The Porphyrin Handbook, v.8, Ch. 57. In: Kadish KM, Smith KM, Guilard R, editors. New York: Academic Press; 2000. [Google Scholar]; (f) Kalyanasundaram K, Gratzel M. Coord. Chem. Rev. 1998;177:347–414. [Google Scholar]
- 19.Assuming all the parameters equal, the sensitivity of the excited state towards quenching by diffusing molecules is proportional to the excited state lifetime, and, therefore, it is 4–5 orders of magnitude higher for phosphorescent than for fluorescent chromophores.
- 20.(a) Buchler JW, Dreher C, Herget G. Liebigs Annalen Der Chemie. 1988:43–54. [Google Scholar]; (b) Buchler JW, Lay KL, Stoppa H. Z. Naturforsch.(B) 1980;35:433–438. [Google Scholar]
- 21.Lindsey JS, Schreiman IC, Hsu HC, Kearney PC, Marguerettaz AM. J. Org. Chem. 1987;52:827–836. [Google Scholar]
- 22.Finikova OS, Cheprakov AV, Beletskaya IP, Carroll PJ, Vinogradov SA. J. Org. Chem. 2004;69:522–535. doi: 10.1021/jo0350054. [DOI] [PubMed] [Google Scholar]
- 23.Katan C, Terenziani F, Mongin O, Werts MHV, Porres L, Pons T, Mertz J, Tretiak S, Blanchard-Desce M. J. Phys. Chem. A. 2005;109:3024–3037. doi: 10.1021/jp044193e. [DOI] [PubMed] [Google Scholar]
- 24.Ulman A, Willand CS, Kohler W, Robello DR, Williams DJ, L. Handley L. J. Am. Chem. Soc. 1990;112:7083–7090. [Google Scholar]
- 25.Nguyen T, Francis MB. Org. Lett. 2003;5:3245–3248. doi: 10.1021/ol035135z. [DOI] [PubMed] [Google Scholar]
- 26.Karstens T, Kobs K. J. Phys. Chem. 1980;84:1871–1872. [Google Scholar]
- 27.Several values have been reported in the literature for the fluorescence quantum yield of Rhodamine B, ranging from 0.41 to 0.97: Demas JN, Crosby GA. J. Phys. Chem. 1971;75:991. We used the value of 0.5, reported by Karstens and Kobs [26] and confirmed by our own measurements.
- 28.Vinogradov SA, Fernandez-Seara MA, Dugan BW, Wilson DF. Rev. Sci. Instrum. 2001;72:3396–3406. [Google Scholar]
- 29.Xu C, Webb WW. J. Opt. Soc. Am. B-Opt. Phys. 1996;13:481–491. [Google Scholar]
- 30.A value about 7 times lower was recently reported for σ2 of Rhodamine B by Kauert and co-workers: Kauert M, Stoller PC, Frenz M, Ricka J. Optics Express. 2006;14:8434–8447. doi: 10.1364/oe.14.008434.
- 31.See [ref.17], p. 419.
- 32.Rehm D, Weller A. Isr. J. Chem. 1970;8:259. [Google Scholar]
- 33.R. A. Marcus RA. J. Chem. Phys. 1956;24:966–978. [Google Scholar]
- 34.Marcus inverted region is not observed in dynamic quenching (see [ref.17]).
- 35.A more rigorous proof, requiring complete spectroscopic signatures of ionic species formed as a result of the ET reaction, was left for the next stage, when the covalent assemblies are constructed from the selected chromophores.
- 36.Karotki A, Drobizhev M, Kruk M, Spangler C, Nickel E, Mamardashvili N, Rebane A. J. Opt. Soc. Am. B-Opt. Phys. 2003;20:321–332. [Google Scholar]
- 37.Knyukshto VN, Shul'ga AM, Sagun EI, Zen'kevich EI. Optics and Spectroscopy. 2006;100:590–601. [Google Scholar]
- 38.For example see: Narang U, Zhao CF, Bhawalkar JD, Bright FV, Prasad PN. J. Phys. Chem. 1996;100:4521–4525. ; (b) Pond SJK, Rumi M, Levin MD, Parker TC, Beljonne D, Day MW, Bredas JL, Marder SR, Perry JW. J. Phys. Chem. A. 2002;106:11470–11480. [Google Scholar]; (c) Terenziani F, Painelli A, Katan C, M. Charlot M, Blanchard-Desce M. J. Am. Chem. Soc. 2006;128:15742–15755. doi: 10.1021/ja064521j. [DOI] [PubMed] [Google Scholar]; (d) Belfield KD, Bondar MV, Kachkovsky OD, Przhonska OV, Yao S. J. Lumines. 2007;126:14–20. [Google Scholar]
- 39.Fischer AB, Bronstein-Bonte I. J. Photochem. 1985;30:475–485. [Google Scholar]
- 40.(a) Reinhardt BA, Brott LL, Clarson SJ, Dillard AG, Bhatt JC, Kannan R, Yuan LX, He GS, Prasad PN. Chem. Mater. 1998;10:1863–1874. [Google Scholar]; (b) He GS, Yuan LX, Cheng N, Bhawalkar JD, Prasad PN, Brott LL, Clarson SJ, Reinhardt BA. J. Opt. Soc. Am. B-Opt. Phys. 1997;14:1079–1087. [Google Scholar]
- 41.Literature values for 2PA cross-section of AF-50 vary greatly. In an early work, the value as high as 11,500 GM was reported ([ref.40a]). More recently measured values are in the range 30–50 GM: Kim DY, Alm TK, Kwon JH, Kim D, Ikeue T, Aratani N, Osuka A, Shigeiwa M, Maeda S. J. Phys. Chem. A. 2005;109:2996–2999. doi: 10.1021/jp050747h. Saita S, Yamaguchi T, Kawai T, Irie M. ChemPhysChem. 2005;6:2300–2306. doi: 10.1002/cphc.200500254.
- 42.Lin TC, He GS, Prasad PN, Tan LS. J. Mater. Chem. 2004;14:982–991. [Google Scholar]
- 43.(a) Gouterman M. J. Mol. Spectros. 1961;6:138–163. [Google Scholar]; (b) Nguyen KA, Pachter R. J. Chem. Phys. 2001;114:10757–10767. [Google Scholar]; (c) Rosa A, Ricciardi G, Baerends EJ, van Gisbergen SJA. J. Phys. Chem. A. 2001;105:3311–3327. [Google Scholar]; (d) Kadish KM, Finikova OS, Espinosa E, Gros CP, De Stefano G, Cheprakov AV, Beletskaya IP, Guilard R. J. Porph. Phthal. 2004;8:1062–1066. [Google Scholar]
- 44.Vuorimaa E, Lemmetyinen H, VanderAuweraer M, DeSchryver FC. Thin Solid Films. 1995;268:114–120. [Google Scholar]
- 45.(a) Eastwood D, Gouterman M. J. Mol. Spectros. 1970;35:359–375. [Google Scholar]; (b) Kim DH, Holten D, Gouterman M, Buchler JW. J. Am. Chem. Soc. 1984;106:4015–4017. [Google Scholar]
- 46.Rogers JE, Nguyen KA, Hufnagle DC, McLean DG, Su WJ, Gossett KM, Burke AR, Vinogradov SA, Pachter R, Fleitz PA. J. Phys. Chem. A. 2003;107:11331–11339. [Google Scholar]
- 47.Mackay I, Cai LZ, Kirk AD, McAuley A. Inorg. Chem. 1999;38:3628–3633. doi: 10.1021/ic990093p. [DOI] [PubMed] [Google Scholar]
- 48.See ref. 17, p. 418.
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Synthetic details and characterization data for new compounds described in this paper are available in the Supporting Information.