Abstract
A liquid chromatography-atmospheric pressure chemical ionization-tandem mass spectrometry (LC-APCI-MS/MS) method for quantification of ten amphetamine-related analytes in 1 g meconium is presented. Specimen preparation included homogenization and solid phase extraction. Two multiple reaction monitoring transitions were monitored per analyte. Ten and 1 µL injection volumes permitted quantification up to 10,000 ng/g, with sufficient sensitivity to quantify minor metabolites. Lower limits of quantification ranged from 1.25 to 40 ng/g. Precision was less than 14.2%, with accuracy between 79 – 115%. Meconium from a methamphetamine-exposed neonate was analyzed. Metabolites p-hydroxymethamphetamine, norephedrine and 4-hydroxy-3-methoxymethamphetamine were identified in meconium for the first time.
Keywords: meconium, amphetamines, LC-MS-MS
Introduction
According to the 2005 National Survey on Drug Use and Health National Findings, 3.9% of pregnant women, ages 15 to 44 years, reported illicit drug use within the previous month.[1] Drug exposure to the developing fetus has been associated with reduced birth weight, small head circumference, premature birth, fetal distress, potential birth defects, and perinatal complications.[2] Limited studies have considered specific adverse outcomes of human in utero methamphetamine (MAMP) or amphetamine (AMP) exposure, either alone or in combination with other drugs such as cocaine.[3–7] Human neonatal outcomes after in utero exposure to methylenedioxy- derivatives of AMP, such as 3,4-methylenedioxymethamphetamine (MDMA) and 3,4-methylenedioxyamphetamine (MDA) are not well understood. Exposed infants may be identified by maternal self-report or chromatographic confirmation of MAMP, AMP, MDMA and/or metabolites in meconium. Due to potential social and medical implications of in utero drug exposure, exposed neonates need to be accurately identified, so that appropriate resources and follow-up are available for the child throughout development, and drug treatment, child care training and other support also are available for the mother.
Meconium is the first neonatal bowel movement, typically excreted within a few days of birth. It is a complex matrix consisting of water, epithelial cells, lanugo, bile acids and salts, enzymes, lipids, sugars, proteins, and swallowed amniotic fluid.[8,9] Meconium formation begins around the twelfth week of gestation, offering a wider window of detection than other neonatal matrices.[10] Meconium analysis can detect maternal drug use over the second and third trimesters, but specimen weight is frequently quite limited, while positive results in hair or urine only indicate use during the third trimester or the 1– 3 days immediately preceding delivery, respectively. Obtaining meconium from diapers also is easier than obtaining neonatal urine with a special collection device that frequently falls off or irritates newborn skin. Hair is often present in quantities insufficient for drug analysis; when ample hair is available, many mothers are reluctant to consent to hair collection for cosmetic or cultural reasons. Analysis of meconium requires extensive sample preparation due to its complexity and heterogeneity, with homogenization commonly employed prior to liquid-liquid and/or solid phase-extraction (SPE). Cocaine and metabolites, cannabinoids, amphetamines, opiates, phencyclidine, methadone, nicotine, non-steroidal anti-inflammatory drugs, and ethanol metabolites, fatty acid ethyl esters, have previously been analyzed in meconium.[11–31] Gareri et al.[8], Moore et al.[9], and Gray et al.[32] have reviewed a number of these methods, which have primarily involved immunoassays, gas chromatography-mass spectrometry (GC-MS) and liquid chromatography-mass spectrometry (LC-MS).
Primary metabolic pathways of MAMP include N-demethylation to AMP and aromatic hydroxylation to p-hydroxymethamphetamine (pOHMAMP, pholedrine); additional minor metabolites include p-hydroxyamphetamine (pOHAMP), norephedrine (NOREPH) and p-hydroxynorephedrine (pOHNOREPH).[33,34] The main metabolic pathways of MDMA involve N-dealkylation to MDA, followed by O-demethylation to form intermediates of 3,4-dihydroxyamphetamine (HHA) and 3,4-dihydroxymethamphetamine (HHMA). HHA and HHMA undergo further O-demethylation, forming 4-hydroxy-3-methoxymethamphetamine (HMMA) and 4-hydroxy-3-methoxyamphetamine (HMA). 3,4-Methylenedioxyethylamphetamine (MDEA) is metabolized via N-deethylation to MDA and 4-hydroxy-3-methoxyethylamphetamine (HMEA), via the formation of 3,4-dihydroxyethylamphetamine (HHEA) following cleavage of the methylenedioxy- group.[34–37] HMA, HMMA, HMEA, HHA, HHMA, HHEA may also undergo conjugation to form glucuronide and/or sulfate metabolites.[34,37,38] Figure 1 summarizes the metabolic pathways of MAMP, AMP, MDMA and MDEA.
Figure 1.
Metabolic pathways of methamphetamine (MAMP), amphetamine (AMP), 3,4-methylenedioxymethamphetamine (MDMA) and 3,4-methylenedioxyethylamphetamine (MDEA); Abbreviations: 3,4-dihydroxyamphetamine (HHA), 3,4-dihydroxyethylamphetamine (HHEA), 3,4-dihydroxymethamphetamine (HHMA), 4-hydroxy-3-methoxyamphetamine (HMA), 4-hydroxy-3-methoxyethylamphetamine (HMEA), 4-hydroxy-3-methoxymethamphetamine (HMMA), 3,4-methylenedioxyamphetamine (MDA), norephedrine (NOREPH), p-hydroxyamphetamine (pOHAMP), p-hydroxynorephedrine (pOHNOREPH), p-hydroxymethamphetamine (pOHMAMP)
Meconium specimens screening positive for amphetamines by immunoassay sometimes cannot be confirmed for AMP, MAMP, MDMA, MDA and MDEA by GC-MS; between 60–100% of meconium specimens positive by AMP immunoassays are negative for AMP or MAMP after GC-MS reanalysis.[15,39–41] This raises the question of whether MAMP and AMP are the most appropriate biomarkers in meconium for the detection of in utero AMP exposure. Possibly, unidentified MAMP or AMP metabolites, other sympathomimetic amines, or endogenous meconium compounds cross-react with the immunoassay, triggering positive responses. The presence of MAMP and AMP metabolites in meconium cannot always be detected with current confirmatory analytical methods that focus on parent compounds, AMP, MAMP, MDMA, MDA, MDEA and N-methyl-1-(3,4-methylenedioxyphenyl)-2-butanamine (MBDB) [15,22,23]; therefore MAMP-exposed neonates could be misidentified and denied interventional measures. Additionally, identification and quantification of other MAMP and AMP metabolites in meconium may provide valuable data regarding the maternal/fetal transfer of AMP-related drugs.
The aim of this study was to develop and validate a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the simultaneous quantification of ten AMP-related analytes including four compounds never before investigated in meconium. AMP and MAMP concentrations exceeding 1,000 ng/g have been previously identified in meconium [23]; therefore, analysis of AMP-related analytes requires simultaneous quantification of high parent compound concentrations (that is, greater than 1,000 ng/g) and low minor metabolite concentrations in quantity-limited meconium specimens. To accommodate the large analytical range, two calibration curves were constructed using different injection volumes. The method will be employed to characterize AMP and related compound disposition in infant meconium.
Experimental
Chemicals and Reagents
All drug standards were obtained as racemic mixtures. Methanolic solutions (1.0 mg/mL) of AMP, AMP-d11, MAMP, MAMP-d14, phenylpropanolamine hydrochloride (also known as NOREPH), NOREPH-d3 hydrochloride, MDA, MDA-d5, MDMA, MDMA-d5, MDEA, MDEA-d5, and HMMA were purchased from Cerilliant (Austin, TX, USA). A 1.0 mg/mL methanolic standard HMA hydrochloride was obtained from Lipomed (Cambridge, MA, USA). pOHAMP hydrobromide and pOHMAMP hydrochloride were kindly provided by the National Institute on Drug Abuse (NIDA) Drug Inventory (maintained by Research Triangle Institute, Research Triangle Park, NC, USA), and pOHNOREPH hydrochloride was purchased from Sigma Aldrich (St. Louis, MO, USA).
Morphine-6-β-D-glucuronide·H2O solution (1.0 mg base/mL dimethyl sulfoxide) was sourced from Lipomed (Cambridge, MA, USA). m-Hydroxycocaine, p-hydroxycocaine, m-hydroxybenzoylecgonine, p-hydroxybenzoylecgonine were obtained from Research Biochemicals International (Natick, MA, USA). (±)-Norcotinine and trans-3’- hydroxycotinine were purchased as powders from Toronto Research Chemicals (North York, Ontario, Canada). Acetylsalicylic acid, acetaminophen, brompheniramine maleate, caffeine, chlorpheniramine maleate, and ibuprofen were obtained as powders from Sigma Aldrich (St. Louis, MO, USA). All other potential interference standards were obtained from Cerilliant (Austin, TX, USA) as 100 µg/mL or 1.0 mg/mL racemic solutions (where applicable) in methanol or acetonitrile (Table 1).
Table 1.
Compounds evaluated as potential interferences with amphetamine-related compounds in meconium. Low quality control samples were fortified with 2,500 ng/g potential interferent to determine if accurate quantification (±20% of target) could be obtained.
| Compounds | |
|---|---|
| Opiates | Buprenorphine, norbuprenorphine, morphine, morphine-3-β-D-glucuronide, morphine-6-β-D-glucuronide, normorphine, 6-acetylmorphine, codeine, norcodeine, 6-acetylcodeine, hydromorphone, hydrocodone, oxymorphone, noroxymorphone, oxycodone, noroxycodone, methadone, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, 2-ethyl-5-methyl-3,3-diphenylpyrroline, methadol |
| Cocaine | Cocaine, benzoylecgonine, norcocaine, norbenzoylecgonine, ecgonine ethyl ester, ecgonine methyl ester, anhydroecgonine methyl ester, ecgonine, cocaethylene, norcocaethylene, m-hydroxycocaine, p-hydroxycocaine, m-hydroxybenzoylecgonine, p-hydroxybenzoylecgonine |
| Cannabinoids | Δ9-Tetrahydrocannabinol, 11-nor-9-carboxyl-Δ9-tetrahydrocannabinol, 11-hydroxy-Δ9-tetrahydrocannabinol |
| Amphetamine | N-methyl-1-(3,4-methylenedioxyphenyl)-2-butanamine, 3,4-methylenedioxyphenyl-2-butanamine, cathinone, N-ethylamphetamine, 4-bromo-2,5-dimethoxyphenethylamine, ephedrine, fenfluramine, pseudoephedrine, norpseudoephedrine, p-methoxyamphetamine, p-methoxymethamphetamine, phentermine |
| Benzodiazepines | Alprazolam, 7-aminoclonazepam, 7-aminoflunitrazepam, 7-aminonitrazepam, bromazepam, clonazepam, diazepam, flunitrazepam, flurazepam, lorazepam, nitrazepam, nordiazepam, oxazepam, temazepam |
| Nicotine | Nicotine, cotinine, norcotinine, trans-3’-hydroxycotinine |
| Antidepressants | Clomipramine, fluoxetine, imipramine, norfluoxetine, paroxetine |
| Antihistamines | Brompheniramine, chlorpheniramine, diphenhydramine |
| Miscellaneous | Acetylsalicylic acid, acetaminophen, ibuprofen, pentazocine, propoxyphene, caffeine, clonidine, phencyclidine |
All solvents were HPLC grade and obtained from the following suppliers: methanol (Fisher Scientific, Pittsburgh, PA, USA) and acetonitrile and water (JT Baker, Phillipsburg, NJ, USA). All reagents were analytical reagent grade or higher. Ammonium acetate was obtained from Sigma Chemicals (St. Louis, MO, USA); ammonium hydroxide, hydrochloric acid and phosphoric acid were purchased from JT Baker (Phillipsburg, NJ, USA). Formic acid was sourced from Merck KGaA (Darmstadt, Germany).
Specimens
Pools of meconium screening negative for amphetamines, cannabinoids, cocaine and opiates were obtained from ElSohly Laboratories (Oxford, MS, USA); 1 g aliquots of each blank pool were analyzed to verify the absence of potential interferences prior to use as a matrix base for calibrators and quality control (QC) samples.
Authentic MAMP and/or AMP positive meconium specimens were obtained from United States Drug Testing Laboratory (USDTL, Des Plaines, IL, USA) to demonstrate the applicability of the validated method. Specimens were stored at −20 °C at USDTL and transferred frozen to NIDA with additional storage at −80°C until analysis.
Preparation of Calibrators and QC Samples
Stock solutions (1.0 mg base/mL) of pOHAMP, pOHMAMP and pOHNOREPH were prepared in methanol. Methanolic standards (1.0 mg/mL) of the ten analytes were combined and diluted in methanol to prepare a 50,000 ng/mL mixed calibrator stock solution. Working solutions were prepared at concentrations of 25, 50, 250, 500, 2,500, 5,000, 10,000, 25,000 and 50,000 ng/mL. Fifty, 100 or 200 µL aliquots of working solutions were added to 1.0 ± 0.1 g blank meconium to yield final concentrations of 1.25 – 10,000 ng/g. Two calibration curves, A and B, were generated to extend the dynamic analytical range and allow quantification of minor metabolites, as well as high concentrations of AMP and MAMP from a single meconium extract. Calibration curve A was generated using 10 µL injections of 1.25 – 2,500 ng/g calibrators to quantify analytes from their lower limit of quantification to 2,500 ng/g; calibration curve B was constructed with 1 µL injections of 125 – 10,000 ng/g calibrators for quantification of concentrations greater than 2,500 ng/g. Calibrator concentrations shared by both curves were injected from the same extracted sample. Presently, there are no estimates of minor metabolite concentrations in meconium. In the event that minor metabolite concentrations exceed 2,500 ng/g in authentic meconium specimens, the method was validated at higher concentrations with the 1 µL injection curve for minor metabolites, as well as parent metabolites.
QC working solutions were prepared from different ampules of 1.0 mg/mL methanolic standards than were used to prepare the calibrator solutions; when available, different manufacturer lot numbers were employed. Blank meconium samples were fortified with 160, 800, 8,000 and 50,000 ng/mL working QC solutions yielding low, medium and high QC samples for each analyte. For the 10 µL injection curve, low, medium and high QC concentrations were 8 ng/g for HMMA and MDEA or 40 ng/g for pOHAMP, pOHMAMP, NOREPH, HMA, AMP, MAMP, MDA and MDMA, 400 ng/g, and 2,500 ng/g, respectively. The 1 µL injection curve low, medium and high QC concentrations were 400, 2,500, and 10,000 ng/g, respectively. The same 400 or 2,500 ng/g extract was injected at 1 and 10 µL, serving as QC samples for both calibration curves.
Individual 1.0 mg/mL methanolic standards of six internal standards were combined and diluted in methanol to prepare a 100,000 ng/mL mixed internal standard stock solution. A 4,000 ng/mL working internal standard solution was prepared in methanol. Twenty-five µL of the working solution was added to each calibrator, quality control and authentic specimen, yielding internal standard concentrations of 100 ng/g. When available, the corresponding deuterated analogue was utilized as an internal standard; the internal standard for pOHAMP, pOHMAMP, HMA and HMMA was NOREPH-d3, the closest eluting deuterated internal standard.
All stock solutions were stored at −80°C prior to use; working solutions were prepared daily.
Instrumentation
Experiments were performed on a Shimadzu HPLC system, consisting of two LC- 10ADvp pumps, a CTO-10ACvp column heater and a SIL-HTC autosampler (Columbia, MD, USA). The HPLC was interfaced with an Applied Biosystems MDS Sciex API 3000 triple quadrupole mass spectrometer (Foster City, CA, USA), equipped with an atmospheric pressure chemical ionization (APCI) source. All data were acquired and quantified using Analyst software, Version 1.4.
A Tekmar ultrasonic disrupter (Cincinnati, OH, USA), Branson 3510 ultrasonic bath (Branson Ultrasonic Corporation, Danbury, CT, USA) and Eppendorf Refrigerated 5804R centrifuge (Brinkmann, Westbury, NY, USA) were utilized during specimen preparation. Samples were concentrated with a Zymark TurboVap LV evaporator (Hopkinton, MA, USA).
Strata XC 33 µm cation mixed-mode polymer (100mg/6mL) SPE cartridges, Synergi Polar RP 4 µm, 150 × 2.0 mm analytical columns and Synergi Polar RP 4.0 × 2.0 mm guard columns were obtained from Phenomenex (Torrance, CA, USA).
Specimen Preparation
Twenty-five µL of the 4,000 ng/mL working internal standard solution and 50, 100 or 200 µL of working calibrator or QC solution (when appropriate) were added to 1.0 ± 0.1 g of meconium. Specimens were centrifuged at 290 × g at room temperature for 5 min to drive meconium to the bottom of the 15 mL polypropylene centrifuge tube prior to adding 3 mL 17 mM methanolic hydrochloric acid. After vortex mixing, ultrasonic homogenization using 60 amp, output 7–10 W was performed for 1 min in an ice-water bath, followed by sonication in an ultrasonic bath for 30 min, and centrifugation at 8230 × g at room temperature for 10 min. The supernatant was decanted into a clean glass centrifuge tube and evaporated under nitrogen at 37°C to approximately 0.5 mL. Prior to SPE, 3 mL 2% phosphoric acid were added to each sample, and vortexed.
Strata XC SPE cartridges were conditioned with 3 mL methanol, followed by 3 mL water. After acidified samples were applied, cartridges were washed with 3 mL 0.1% phosphoric acid and allowed to dry under approximately 10 mm Hg vacuum for 3 min. After a 3 mL methanol wash, analytes were eluted with freshly prepared 3 mL methanol:ammonium hydroxide (95:5, v/v). Twenty-five µL 1% HCl were added to each extract prior to evaporation to dryness under nitrogen at 37°C and reconstitution in 200 µL of 10 mM ammonium acetate with 0.01% (v/v) formic acid.
Liquid Chromatography-Mass Spectrometry
Synergi Polar RP 4 µm, 150 × 2.0 mm analytical and 4.0 × 2.0 mm guard columns adequately separated the ten analytes. Gradient elution with a binary mobile phase system of (A) 10 mM ammonium acetate containing 0.01% (v/v) formic acid and (B) acetonitrile was performed with a 30°C column temperature and 200 µL/min flow rate. The gradient profile was 10% B for 1 min, linear increase to 50% B over 13 min, hold at 50% B for 2 min, ramp to 90% B over 2 min, hold at 90% B for 2 min, and decrease to 10% B over 2 min. The column was re-equilibrated for 18 min at 10% B after each injection. Injection volume was either 10 µL (Curve A) or 1 µL (Curve B). The autosampler was maintained at 15°C.
MS/MS parameters were optimized with acidified aqueous standard solutions of each analyte infused at 10 µL/min into the MS/MS via a Hamilton syringe infusion pump. Nebulizing (50 psi) and curtain gases (30 psi) were nitrogen. The corona discharge current was 3.0 µA and the source temperature was 375°C.
Multiple reaction monitoring (MRM) scans in positive ion mode of the molecular ion and the two most predominant fragments for each analyte were utilized. Selected MRM transitions for each analyte and internal standard, with corresponding collision energies, are given in Table 2. The 20 min analysis time was divided into two periods. Dwell times for all analyte and internal standard transitions was 200 ms. Quadrupoles operated at unit resolution for all experiments.
Table 2.
Precursor and product ions with corresponding collision energies for each analyte and internal standard.
| Analyte/ Internal Standard |
Precursor Ion (m/z) |
Quantifier transition production (m/z) |
Collision Energy (volts) |
Quantifier transition production (m/z) |
Collision Energy (volts) |
|---|---|---|---|---|---|
| Period 1 | |||||
| pOHAMP | 152 | 107 | 24 | 77 | 45 |
| pOHMAMP | 166 | 135 | 17 | 107 | 28 |
| NOREPH | 152 | 117 | 24 | 134 | 14 |
| NOREPH-d3 | 155 | 119 | 26 | 137 | 15 |
| HMA | 182 | 105 | 31 | 165 | 14 |
| HMMA | 196 | 165 | 18 | 105 | 32 |
| Period 2 | |||||
| AMP | 136 | 91 | 24 | 119 | 12 |
| AMP-d11 | 147 | 98 | 23 | 130 | 14 |
| MAMP | 150 | 119 | 17 | 91 | 22 |
| MAMP-d14 | 164 | 98 | 30 | 130 | 17 |
| MDA | 180 | 105 | 32 | 163 | 11 |
| MDA-d5 | 185 | 110 | 32 | 168 | 15 |
| MDMA | 194 | 163 | 17 | 105 | 35 |
| MDMA-d5 | 199 | 165 | 18 | 107 | 37 |
| MDEA | 208 | 163 | 19 | 105 | 36 |
| MDEA-d5 | 213 | 163 | 20 | 105 | 40 |
Abbreviations: amphetamine (AMP), 4-hydroxy-3-methoxyamphetamine (HMA), 4-hydroxy-3-methoxymethamphetamine (HMMA), methamphetamine (MAMP), 3,4-methylenedioxyamphetamine (MDA), 3,4-methylenedioxyethylamphetamine (MDEA), 3,4-methylenedioxymethamphetamine (MDMA), norephedrine (NOREPH), p-hydroxyamphetamine (pOHAMP), p-hydroxymethamphetamine (pOHMAMP)
Validation
Following bioanalytical validation guidelines [42–44], the method was fully validated in four days for the following parameters: selectivity, linearity, including limit of detection (LOD) and lower limit of quantification (LLOQ), accuracy and precision, extraction efficiency, matrix effect, and stability, including autosampler stability and freeze/thaw stability at −80°C.
Selectivity was established by analysis of eight different lots of blank meconium and through evaluation of potential interferences (Table 1). Duplicate low QC samples were fortified with 2,500 ng/g potential interferences. A lack of interference was documented if analytes of interest quantified within 20% of target, and qualifier/quantifier transition ratios were within permitted tolerances.
LOD and LLOQ were evaluated using triplicates of samples at or below the lowest calibrator for each analyte. LOD was defined as the concentration producing a Gaussian-shaped peak eluting within ± 2% of mean calibrator retention time, a minimum signal to noise (S/N) ratio of 3:1, and qualifier/quantification transition ratio within the maximum tolerance of mean calibrator qualifier/qualification transition ratio. LLOQ was defined as the concentration that met LOD criteria and had a minimum S/N ratio of 10:1 and QC precision and accuracy within ± 20%.
To assess linearity, calibration curves were obtained by analyzing drug-free meconium fortified with working calibrator solutions yielding 1.25 – 2,500 ng/g final concentrations for Curve A (10 µL injection) or 125 – 10,000 ng/g for Curve B (1 µL injection). Calibration curves were constructed using linear regression with 1/x weighting based on a minimum of six calibrator peak area ratios, excluding the blank. Calibrators and duplicate QC samples at low, medium and high concentrations were analyzed daily in each set of specimens.
Accuracy and precision were evaluated with five replicates at low, medium and high concentrations. Intra-assay (n = 5) precision and accuracy were assessed within the same validation batch on four separate days, while inter-assay precision and accuracy were evaluated over four validation batches (n = 20). Accuracy is the percentage of calculated to target concentration, while precision, expressed as % coefficient of variation (% CV), is the mean concentration divided by the standard deviation of the concentrations of 5 or 20 replicates.
Extraction efficiency and matrix effect for each analyte and internal standard were assessed through the preparation of three sets of samples, as outlined by Matuszewski et al.[42] Set 1 included unextracted standards reconstituted in mobile phase A; set 2 was extracted negative meconium samples with QC and internal standard solutions added after extraction during reconstitution; and set 3 was fortified meconium samples with QC and internal standard solutions added before extraction prior to homogenization. Each set had five replicates at four QC concentration levels: low, medium, and highs for the 10 and 1 µL injection methods. Extraction efficiency was calculated as the ratio of the average peak area in set 3 to set 2, expressed as a percentage. Matrix effect was defined as the percentage ratio of the average peak areas for set 2 and set 1.
Stability of extracted QC samples under 15°C autosampler storage was evaluated at low and high QC concentrations for each injection method. Each QC sample was injected every 3 hours, up to 78 hours. The % CV of the peak area of each analyte and internal standard was calculated for each QC sample. Analyte stability in unprocessed meconium samples was also evaluated after three freeze-thaw cycles with 5 replicates at low and high QC concentrations using the 10 µL injection method. Each cycle consisted of freezing at −80°C for a minimum of 24 hours, then thawing unassisted at room temperature. The freeze-thaw stability was calculated as the ratio of the average concentration of freeze-thawed QC samples and freshly prepared QC samples, expressed as a percentage.
Results and Discussion
While electrospray and APCI ionization were evaluated for each analyte, APCI was selected on the basis of less matrix effect. Monitoring two MRM transitions per analyte adheres to the Commission of European Communities (EC) identification criteria requiring a minimum of three identification points (IP). Under these guidelines, the LC-MSn precursor ion has a value of 1.0 IP and each LC-MSn transition product earns 1.5 IP, resulting in a total of 4.0 IP.[45] Mass spectrometric conditions for each MRM transition, such as collision energy, were individually optimized. Selection of the quantifier and qualifier transitions was based on transitions from the molecular ion to the most and second-most predominant fragment ions, respectively. For NOREPH, HMA and MDA, selection of the quantifier and qualifier transitions was reversed; the second-most predominant fragment ion was chosen for quantification since the most predominant fragment corresponded to the loss of water. The MAMP quantification transition also was based on the second-most abundant fragment ion, as this transition had a lower level of background noise in blank meconium samples.
Currently, there is no consensus on acceptable relative abundance variability of LC-MS/MS MRM transition ratios, as several regulatory bodies have proposed differing guidelines.[45–47] In the current method, the EC guidelines for maximal permitted tolerances were followed, where permitted tolerances of relative ion intensities for LC-MSn techniques are dependent on relative transition intensities.[45] Mean relative abundance ratios of each analytes’ two MRM transitions ranged from 0.17 to 0.95 in calibrators, with % CVs of less than 12% observed over four days for both the 10 µL and 1 µL injection methods. Mean transition abundance ratios and permitted tolerances under EC guidelines are given in Table 3. Using the 10 and 1 µL injection methods, all analytes had acceptable MRM transition ratios within these permitted tolerances for all QC samples.
Table 3.
Mean relative abundance ratios of two MRM transitions monitored for amphetamine-related analytes in meconium
| 10 µL injection | 1 µL injection | ||||
|---|---|---|---|---|---|
| Analyte | Relative abundnce* |
Inter-day % CV (n=4) |
Relative abundance* |
Inter-day % CV (n=4) |
Permitted Tolerance (%)** |
| pOHAMP | 0.166 | 3.8 | 0.170 | 1.4 | 30 |
| pOHMAMP | 0.898 | 5.5 | 0.946 | 1.2 | 20 |
| NOREPH | 0.338 | 1.5 | 0.331 | 2.2 | 25 |
| HMA | 0.330 | 2.9 | 0.291 | 12.2 | 25 |
| HMMA | 0.323 | 3.3 | 0.309 | 3.9 | 25 |
| AMP | 0.787 | 3.0 | 0.786 | 2.1 | 20 |
| MAMP | 0.656 | 0.8 | 0.660 | 1.3 | 20 |
| MDA | 0.420 | 3.9 | 0.416 | 3.0 | 25 |
| MDMA | 0.436 | 8.2 | 0.382 | 1.3 | 25 |
| MDEA | 0.409 | 2.7 | 0.410 | 2.4 | 25 |
Abbreviations: amphetamine (AMP), coefficient of variation (CV), 4-hydroxy-3-methoxyamphetamine (HMA), 4-hydroxy-3-methoxymethamphetamine (HMMA), methamphetamine (MAMP), 3,4-methylenedioxyamphetamine (MDA), 3,4-methylenedioxyethylamphetamine (MDEA), 3,4-methylenedioxymethamphetamine (MDMA), norephedrine (NOREPH), p-hydroxyamphetamine (pOHAMP), p-hydroxymethamphetamine (pOHMAMP)
Ratio of qualifier to quantification transition peak areas, except for NOREPH, HMA, MAMP, and MDA which are the ratio of quantification to qualifier transition.
As specified by Commission of European Communities Council Directive 96/23/EC [45]
Representative extracted ion chromatograms for blank meconium and a low QC sample, and authentic meconium obtained from a neonate following in utero MAMP exposure are given in Figure 2 and Figure 3, respectively. All analytes eluted within 15 min; each analytical run was less than 33 min. The column was allowed to re-equilibrate for 18 minutes after the final step of the gradient program and the beginning of the next injection.
Figure 2.
Extracted ion chromatograms of a) a blank meconium sample and b) a low quality control sample. Quantification and qualifier transitions are denoted by solid and dashed lines, respectively; analyte retention time indicated by arrow i) p-hydroxyamphetamine (pOHAMP), ii) p-hydroxymethamphetamine (pOHMAMP), iii) norephedrine (NOREPH), iv) 4-hydroxy-3-methoxyamphetamine (HMA), v) 4-hydroxy-3-methoxymethamphetamine (HMMA), vi) amphetamine (AMP), vii) methamphetamine (MAMP), viii) 3,4-methylenedioxyamphetamine (MDA), ix) 3,4-methylenedioxymethamphetamine (MDMA), x) 3,4-methylenedioxyethylamphetamine (MDEA). QC concentrations: 8 ng/g for HMMA and MDEA, 40 ng/g for all other analytes.
Figure 3.
Extracted ion chromatograms of an authentic meconium specimen collected from a neonate following in utero drug exposure. Quantification and qualifier transitions are denoted by solid and dashed lines, respectively. Injection volume of 10 µL, except for methamphetamine (1 µL). Specimen contains: a) 129.6 ng/g p-hydroxymethamphetamine, b) 19.3 ng/g norephedrine, c) 4.2 ng/g 4-hydroxy-3-methoxymethamphetamine, d) 639.2 ng/g amphetamine, e) 4,638.8 ng/g methamphetamine.
Blank meconium specimens from eight different sources had minimal interferences with analytes of interest. Seventy-three common licit and illicit compounds were investigated as potential interferences with analytes of interest (Table 1). Duplicate low QC samples were fortified with 2,500 ng/g potential interferences. A concentration of 2,500 ng/g was chosen for the interference study because for most analytes evaluated, 2,500 ng/g exceeds the highest observed concentrations in human meconium reported in the literature. In only a small percentage of cocaine-positive specimens, concentrations of cocaine and some of its metabolites may be greater than 2,500 ng/g. Anhydroecgonine methyl ester (AEME), a pyrolytic biomarker of cocaine base, coelutes with HMA’s quantification transition, m/z 182→105; however, the qualification transition is unaffected. AEME’s presence results in a false negative HMA result as the qualifier/quantification transition ratio would be decreased. If HMA interference was suspected, AEME can be confirmed by monitoring AEME-specific transitions, m/z 182→91 and m/z 182→122. Norpseudoephedrine, a stereoisomer of NOREPH, gave a response for both NOREPH transitions at a retention time immediately after NOREPH, leading to a split peak. However, the two QC samples containing 40 ng/g of NOREPH and 2,500 ng/g norpseudoephedrine were 77% accurate. MAMP qualifier/quantification transition peak area ratio was considerably affected by coelution of phentermine, a structural isomer, with the MAMP qualifier m/z 150→91 transition. The average qualifier/quantification transition peak area ratio of two low QC samples increased from 1.5 to 32.3 in the presence of phentermine; therefore, MAMP false negatives due to the presence of phentermine could be identified by a dramatic increase in the qualifier/quantification transition ratio.
LODs and LLOQs ranged from 0.8 – 25 and 1.25 – 40 ng/g, respectively for the 10 µL injection curve; LLOQs for the 1 µL injection curve were 125 ng/g for all analytes, except HMA at 250 ng/g (Table 4). Linear dynamic ranges for the 10 and 1 µL injection methods were LLOQ - 2,500 ng/g and 125 (or 250) - 10,000 ng/g, respectively. A previous LC-MS method employing electrospray ionization and selective ion monitoring achieved LLOQs of 4 ng/g for MDA, MDEA and MDMA and 5 ng/g AMP, MAMP and HMMA; upper limits of quantification (ULOQ) were 1,000 ng/g.[23] The LLOQs observed in this method using the 10 µL injection were lower than those for MDEA and HMMA, but higher for AMP, MAMP, MDMA and MDA. The higher LLOQs for these analytes were associated with low level interferences present within blank meconium. Additional analytes included in this method that were not considered by the previous method included pOHMAMP, NOREPH, pOHAMP and HMA with LLOQs of 8, 12.5, 12.5 and 40 ng/g, respectively. The elevated LLOQ for HMA was due to background interference from blank meconium with the qualifier transition. The ULOQ of the 10 µL injection method was higher than the previously reported method and inclusion of the 1 µL injection method further increased the dynamic range to 10,000 ng/g for all analytes. Extending the ULOQ of the 10 µL injection method beyond 2,500 ng/g was not possible, as detector saturation was observed; therefore, the two injection volume system was employed to allow quantification up to 10,000 ng/g from a single extract. The 1 µL injection method need only be employed for specimens containing analyte concentrations greater than the ULOQ with the 10 µL injection method (2,500 ng/g); this is often the case with MAMP. Limited specimen amount is a significant concern in meconium analysis; extending the linear dynamic range reduces the need for dilution and reanalysis, and preserves the specimen for other testing, if necessary.
Table 4.
Mean retention times, limits of detection (LODs), lower limits of quantification (LLOQs) and calibration slope and intercept (n = 4) for amphetamine-related analytes in meconium using 10 µL and 1 µL injection methods
| 10 µL injection | 1 µL injection | |||||||
|---|---|---|---|---|---|---|---|---|
| Analyte | Retention time (± SD) |
LOD (ng/g) |
LLOQ (ng/g) |
Slope (± SD) |
Intercept (± SD) |
LLOQ (ng/g) |
Slope (± SD) |
Intercept (± SD) |
| pOHAMP | 5.94 (0.05) | 4 | 12.5 | 4.220 (0.786) | −0.109 (0.154) | 125 | 5.313 (0.977) | −2.204 (1.222) |
| pOHMAMP | 6.81 (0.05) | 2.5 | 8 | 4.485 (0.659) | −0.042 (0.088) | 125 | 5.745 (1.088) | −1.978 (1.419) |
| NOREPH | 7.21 (0.05) | 2.5 | 12.5 | 1.650 (0.226) | −0.044 (0.036) | 125 | 1.925 (0.321) | −0.694 (0.287) |
| HMA | 6.79 (0.05) | 25 | 40 | 1.403 (0.117) | −0.130 (0.060) | 250 | 1.568 (0.054) | −1.422 (0.382) |
| HMMA | 7.55 (0.04) | 1.25 | 2.5 | 4.665 (0.475) | −0.025 (0.040) | 125 | 5.365 (0.604) | −1.208 (1.451) |
| AMP | 9.69 (0.06) | 4 | 12.5 | 0.623 (0.055) | 0.018 (0.010) | 125 | 0.725 (0.089) | −0.0514 (0.180) |
| MAMP | 10.98 (0.09) | 2.5 | 12.5 | 0.702 (0.033) | 0.001 (0.016) | 125 | 0.878 (0.022) | −0.192 (0.183) |
| MDA | 10.64 (0.07) | 4 | 12.5 | 2.198 (0.477) | −0.074 (0.064) | 125 | 2.470 (0.524) | −1.063 (0.448) |
| MDMA | 11.87 (0.08) | 8 | 12.5 | 1.465 (0.245) | −0.016 (0.029) | 125 | 1.755 (0.265) | −0.306 (0.459) |
| MDEA | 13.54 (0.10) | 0.8 | 1.25 | 0.794 (0.049) | −0.001 (0.005) | 125 | 0.918 (0.045) | −0.206 (0.200) |
Abbreviations: amphetamine (AMP), 4-hydroxy-3-methoxyamphetamine (HMA), 4-hydroxy-3-methoxymethamphetamine (HMMA), methamphetamine (MAMP), 3,4-methylenedioxyamphetamine (MDA), 3,4-methylenedioxyethylamphetamine (MDEA), 3,4-methylenedioxymethamphetamine (MDMA), norephedrine (NOREPH), p-hydroxyamphetamine (pOHAMP), p-hydroxymethamphetamine (pOHMAMP)
Mean (± SD) calibration curve slope and intercept and mean (± SD) retention time for each analyte for the 10 µL and 1 µL injection curves over four batches are given in Table 4. Coefficients of determination (R2) were greater than 0.993 and 0.979 for all analytes for the 10 µL and 1 µL injection curves, respectively, using least squares regression and 1/x weighting. All calibrators used in each calibration were calculated to be within ± 20% of their target concentrations.
Accuracy and precision were evaluated with five replicates of low, medium and high concentrations for each analyte. Intra- (n = 5) and inter-assay (n = 20) accuracy and precision data for the 10 µL and 1 µL injection curves are shown in Table 5. Intra-assay (n = 5) precision was less than 6.4% and accuracy 79 – 115% for 10 µL injection QC samples. Inter-assay (n = 20) precision was less than 11.0% and accuracy 84 – 104%. One µL injection QC samples demonstrated intra-assay (n = 5) precision and accuracy of less than 9.7% and 83 – 116%, respectively; with inter-assay (n = 20) precision and accuracy of 2.7 – 14.2% and 90 – 101%, respectively.
Table 5.
Intra- and inter-assay precision and accuracy of amphetamine-related analytes in meconium for the 10 µL and 1 µL injection curves
| 10 µL injection | 1 µL injection | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Analyte | Concentration (ng/g) | Intra-assay (n=5) | Inter-assay (n=20) | Concentration (ng/g) |
Intra-assay (n=5) | Inter-assay (n=20) | ||||
| Accuracy (%) |
Precision (%CV) |
Accuracy (%) |
Precision (%CV) |
Accuracy (%) |
Precision (%CV) |
Accuracy (%) |
Precision (%CV) |
|||
| pOHAMP | 40 | 83 – 92 | 2.0 – 4.5 | 87 | 4.7 | 400 | 83 – 103 | 3.1 – 5.5 | 92 | 9.5 |
| 400 | 94 – 107 | 2.0 – 5.9 | 101 | 7.2 | 2,500 | 93 – 109 | 3.7 – 6.8 | 100 | 8.7 | |
| 2,500 | 90 – 101 | 2.1 – 5.0 | 96 | 5.3 | 10,000 | 88 – 97 | 4.0 – 6.3 | 93 | 6.1 | |
| pOHMAMP | 40 | 81 – 99 | 1.7 – 5.7 | 91 | 7.9 | 400 | 84 – 108 | 3.6 – 5.7 | 95 | 10.6 |
| 400 | 96 – 112 | 2.5 – 4.0 | 102 | 6.8 | 2,500 | 91 – 108 | 4.6 – 6.6 | 99 | 9.1 | |
| 2,500 | 92 – 99 | 1.8 – 3.5 | 96 | 3.8 | 10,000 | 89 – 97 | 4.0 – 5.4 | 94 | 5.6 | |
| NOREPH | 40 | 83 – 92 | 1.6 – 4.7 | 89 | 4.9 | 400 | 87 – 102 | 3.4 – 4.5 | 92 | 7.6 |
| 400 | 88 – 104 | 2.9 – 4.3 | 96 | 7.7 | 2,500 | 87 – 102 | 4.8 – 6.2 | 94 | 8.3 | |
| 2,500 | 92 – 98 | 1.8 – 4.3 | 96 | 3.8 | 10,000 | 90 – 96 | 4.3 – 6.3 | 93 | 5.5 | |
| HMA | 40 | 89 – 108 | 1.5 – 2.8 | 99 | 8.8 | 400 | 94 – 111 | 2.5 – 4.0 | 99 | 8.1 |
| 400 | 85 – 100 | 2.5 – 4.6 | 90 | 7.7 | 2,500 | 85 – 97 | 4.9 – 6.6 | 90 | 7.1 | |
| 2,500 | 91 – 98 | 1.8 – 4.1 | 94 | 3.7 | 10,000 | 90 – 96 | 3.9 – 8.3 | 94 | 6.3 | |
| HMMA | 8 | 79 – 91 | 3.2 – 5.9 | 84 | 7.7 | 400 | 88 – 99 | 2.3 – 5.8 | 92 | 6.2 |
| 400 | 96 – 111 | 2.7 – 5.4 | 104 | 8.0 | 2,500 | 92 – 108 | 4.5 – 6.4 | 100 | 7.9 | |
| 2,500 | 84 – 92 | 1.5 – 4.4 | 89 | 4.6 | 10,000 | 89 – 94 | 3.7 – 7.5 | 92 | 6.0 | |
| AMP | 40 | 84 – 98 | 1.7 – 4.9 | 93 | 7.2 | 400 | 88 – 102 | 1.9 – 4.7 | 93 | 7.1 |
| 400 | 99 – 114 | 2.6 – 5.1 | 103 | 7.2 | 2,500 | 91 – 111 | 1.0 – 6.2 | 100 | 8.2 | |
| 2,500 | 87 – 96 | 1.8 – 3.5 | 92 | 4.7 | 10,000 | 89 – 92 | 0.9 – 4.1 | 90 | 2.7 | |
| MAMP | 40 | 82 – 104 | 1.4 – 3.7 | 93 | 9.1 | 400 | 90 – 103 | 3.1 – 4.3 | 94 | 6.6 |
| 400 | 93 – 115 | 3.6 – 5.8 | 101 | 9.5 | 2,500 | 90 – 105 | 2.6 – 4.7 | 97 | 6.8 | |
| 2,500 | 85 – 97 | 1.6 – 2.7 | 92 | 5.6 | 10,000 | 91 – 95 | 2.3 – 4.9 | 92 | 3.9 | |
| MDA | 40 | 86 – 98 | 2.5 – 6.1 | 93 | 6.4 | 400 | 88 – 100 | 2.8 – 6.2 | 94 | 7.7 |
| 400 | 87 – 105 | 3.6 – 6.4 | 97 | 8.1 | 2,500 | 84 – 114 | 4.1 – 9.7 | 96 | 14.2 | |
| 2,500 | 89 – 109 | 2.1 – 6.2 | 101 | 9.1 | 10,000 | 84 – 107 | 2.8 – 5.4 | 95 | 10.3 | |
| MDMA | 40 | 84 – 97 | 2.4 – 7.3 | 93 | 7.3 | 400 | 90 – 106 | 3.0 – 7.1 | 96 | 7.7 |
| 400 | 96 – 113 | 3.0 – 4.8 | 104 | 8.5 | 2,500 | 89 – 116 | 2.7 – 5.3 | 101 | 11.5 | |
| 2,500 | 92 – 96 | 1.4 – 4.7 | 94 | 3.6 | 10,000 | 86 – 97 | 1.3 – 4.2 | 92 | 5.8 | |
| MDEA | 8 | 81 – 101 | 3.4 – 6.0 | 87 | 11.0 | 400 | 87 – 96 | 2.6 – 5.1 | 90 | 5.1 |
| 400 | 90 – 107 | 2.6 – 5.4 | 97 | 8.0 | 2,500 | 87 – 105 | 2.2 – 5.5 | 97 | 7.6 | |
| 2,500 | 89 – 98 | 2.9 – 4.7 | 93 | 4.9 | 10,000 | 91 – 93 | 1.9 – 5.0 | 92 | 3.2 | |
Abbreviations: amphetamine (AMP), 4-hydroxy-3-methoxyamphetamine (HMA), 4-hydroxy-3-methoxymethamphetamine (HMMA), methamphetamine (MAMP), 3,4-methylenedioxyamphetamine (MDA), 3,4-methylenedioxyethylamphetamine (MDEA), 3,4-methylenedioxymethamphetamine (MDMA), norephedrine (NOREPH), p-hydroxyamphetamine (pOHAMP), p-hydroxymethamphetamine (pOHMAMP)
Extraction efficiency and matrix effect for each analyte and internal standard were assessed at four concentrations, low, medium, and highs for the 10 µL and 1 µL injection methods (Table 6). Extraction efficiencies ranged from 74.2 – 80.6% for all analytes and 63.3 – 69.6% for internal standards. Matrix effect was observed in most analytes and internal standards at all concentrations, with values ranging from 85.2 – 149.4%.
Table 6.
Extraction efficiency and matrix effect of amphetamine-related analytes and internal standards in meconium
| Analyte/ Internal Standard |
Concentration (ng/g) | Extraction efficiency (%) (n=5) |
Matrix Effect (%) (n=5) |
|
|---|---|---|---|---|
| pOHAMP | Low | 40 | 71.9 | 117.7 |
| Medium | 400 | 73.9 | 109.7 | |
| High (10 µL) | 2,500 | 85.4 | 111.0 | |
| High (1 µL) | 10,000 | 80.4 | 113.2 | |
| pOHMAMP | Low | 40 | 76.3 | 117.6 |
| Medium | 400 | 77.2 | 108.4 | |
| High (10 µL) | 2,500 | 86.8 | 104.2 | |
| High (1 µL) | 10,000 | 81.9 | 109.8 | |
| NOREPH | Low | 40 | 72.6 | 108.9 |
| Medium | 400 | 72.8 | 104.3 | |
| High (10 µL) | 2,500 | 82.4 | 98.5 | |
| High (1 µL) | 10,000 | 78.2 | 108.0 | |
| HMA | Low | 40 | 74.2 | 144.0 |
| Medium | 400 | 75.4 | 123.9 | |
| High (10 µL) | 2,500 | 83.4 | 107.5 | |
| High (1 µL) | 10,000 | 74.8 | 115.5 | |
| HMMA | Low | 40 | 75.0 | 149.4 |
| Medium | 400 | 75.8 | 132.3 | |
| High (10 µL) | 2,500 | 83.3 | 114.9 | |
| High (1 µL) | 10,000 | 79.5 | 114.4 | |
| AMP | Low | 40 | 72.4 | 92.5 |
| Medium | 400 | 73.3 | 85.2 | |
| High (10 µL) | 2,500 | 81.0 | 86.7 | |
| High (1 µL) | 10,000 | 80.4 | 97.0 | |
| MAMP | Low | 40 | 69.1 | 101.8 |
| Medium | 400 | 71.3 | 91.3 | |
| High (10 µL) | 2,500 | 79.6 | 90.0 | |
| High (1 µL) | 10,000 | 77.9 | 97.9 | |
| MDA | Low | 40 | 70.5 | 100.5 |
| Medium | 400 | 70.4 | 96.0 | |
| High (10 µL) | 2,500 | 80.2 | 94.5 | |
| High (1 µL) | 10,000 | 75.6 | 106.0 | |
| MDMA | Low | 40 | 68.8 | 107.0 |
| Medium | 400 | 71.2 | 99.3 | |
| High (10 µL) | 2,500 | 79.8 | 98.6 | |
| High (1 µL) | 10,000 | 77.5 | 105.0 | |
| MDEA | Low | 40 | 70.1 | 106.0 |
| Medium | 400 | 73.9 | 99.1 | |
| High (10 µL) | 2,500 | 82.5 | 96.8 | |
| High (1 µL) | 10,000 | 79.2 | 103.8 | |
| NOREPH-d3 | Low | 100 | 68.6 | 119.3 |
| Medium | 100 | 67.0 | 114.5 | |
| High (10 µL) | 100 | 70.2 | 107.1 | |
| High (1 µL) | 100 | 71.8 | 109.9 | |
| AMP-d11 | Low | 100 | 66.0 | 100.1 |
| Medium | 100 | 65.7 | 95.9 | |
| High (10 µL) | 100 | 69.2 | 89.3 | |
| High (1 µL) | 100 | 72.9 | 102.5 | |
| MAMP-d14 | Low | 100 | 62.8 | 122.8 |
| Medium | 100 | 63.9 | 114.1 | |
| High (10 µL) | 100 | 66.5 | 103.7 | |
| High (1 µL) | 100 | 71.9 | 102.1 | |
| MDA-d5 | Low | 100 | 64.0 | 119.5 |
| Medium | 100 | 64.1 | 114.0 | |
| High (10 µL) | 100 | 66.1 | 107.1 | |
| High (1 µL) | 100 | 67.3 | 112.2 | |
| MDMA-d5 | Low | 100 | 60.4 | 131.9 |
| Medium | 100 | 62.9 | 120.4 | |
| High (10 µL) | 100 | 63.2 | 117.4 | |
| High (1 µL) | 100 | 66.6 | 112.7 | |
| MDEA-d5 | Low | 100 | 71.2 | 110.7 |
| Medium | 100 | 66.0 | 118.0 | |
| High (10 µL) | 100 | 68.9 | 110.8 | |
| High (1 µL) | 100 | 72.1 | 109.8 | |
Abbreviations: amphetamine (AMP), 4-hydroxy-3-methoxyamphetamine (HMA), 4-hydroxy-3-methoxymethamphetamine (HMMA), methamphetamine (MAMP), 3,4-methylenedioxyamphetamine (MDA), 3,4-methylenedioxyethylamphetamine (MDEA), 3,4-methylenedioxymethamphetamine (MDMA), norephedrine (NOREPH), p-hydroxyamphetamine (pOHAMP), p-hydroxymethamphetamine (pOHMAMP)
Stability studies suggest no relevant degradation of extracts under autosampler storage conditions. There was less than 10% change in absolute peak area over 78 hours at low and high QC concentrations for both injection methods. The meconium specimen demonstrating method applicability was initially analyzed only for MAMP and AMP at USDTL, where specimens were stored at −20°C prior to and following analysis. Specimens were shipped on dry ice and stored at −80°C at NIDA. USDTL validated their analytical method including stability at −20°C; analyte stability greater than 97% and 83% for six months and one year after one freeze/thaw cycle, respectively was observed. Stability after three freeze-thaw cycles at −80°C was demonstrated at two QC concentrations for the current method. Mean concentration of the five freeze-thawed replicates was within 14% of the mean concentration of freshly prepared QC samples.
Application of Method
A meconium specimen previously identified as MAMP- and AMP-positive by GC-MS was generously donated by USDTL to demonstrate method applicability. The specimen contained 4,638.8 ng/g MAMP, 639.2 ng/g AMP, 129.6 ng/g pOHMAMP, 19.3 ng/g NOREPH, and 4.2 ng/g HMMA; extracted ion chromatographs from this specimen are presented in Figure 3. The high concentration (> 2,500 ng/g) of MAMP demonstrates the importance of the 1 µL injection method for accurate quantification. pOHMAMP, NOREPH, and HMMA were detected for the first time in an authentic meconium specimen.
A low HMMA meconium concentration was found in the absence of MDMA. This is in contrast to one reported MDMA-positive meconium specimen that did not contain other methylenedioxy- derivatives (MDA or MDEA) or the HMMA metabolite [48]. We are unaware of any clinical data demonstrating HMMA after controlled MAMP administration, although it has been suggested as a possible MAMP metabolite.[37] Also, HMMA was reported as a metabolite of p-methoxymethamphetamine (PMMA) after administration to rats, although there are no supportive human data.[49] The presence of pOHMAMP in meconium may indicate maternal MAMP, pholedrine or PMMA exposure. Pholedrine is a therapeutic medication identified in blood, urine and liver of a fatal intoxication case.[50] Additionally, pOHMAMP was identified as a PMMA metabolite in rat urine [49], again with no available human data. Since pOHMAMP and HMMA can be products of several AMP-related precursors, examining the entire metabolic profile in meconium may be helpful to determine exposure to a specific drug. Also, co-ingestion of precursors cannot be excluded, since multiple AMP-related compounds are frequently found in illicit drugs. It is possible that in our case the fetus was exposed to MAMP and MDMA, pholedrine and/or PMMA. Unfortunately, no maternal self-report data are available for this specimen to provide additional information regarding exposure history.
While Pichini et al. applied their developed method to approximately 830 meconium specimens collected as part of the ‘Meconium Project’ by Hospital del Mar, Spain, none were AMP or MAMP positive, and only 1 MDMA positive was identified.[23,48] Additional meconium from neonates of self-reported heavy MAMP abusing mothers were analyzed by Pichini et al. using a reduced specimen aliquot, 0.3 – 0.5 g rather than 1 g.[23] High concentrations of AMP (38 – 1,019 ng/g) and MAMP (226 – 1,157 ng/g) were identified in these specimens, but no methylenedioxy- derivatives were detected, with the exception of one specimen with 12 ng/g MDMA. The advantage of the current method is that analyte concentrations up to 10,000 ng/g can be quantified simultaneously with lower concentrations of minor metabolites. This is accomplished by injecting 10 and 1 µL (if necessary) from a single 1 g aliquot extract and identifying and quantifying analyte concentrations from the appropriate concentration curve.
Conclusion
A validated LC-APCI-MS/MS method for the simultaneous quantification of ten AMP-related analytes in meconium was reported. Two MRM transitions were selected for each analyte for improved selectivity. Injection volumes of 1 µL and 10 µL allowed quantification of concentrations of parent analytes up to 10,000 ng/g, while sufficient sensitivity was available to quantify low minor metabolite concentrations. The validated method was applied to a meconium specimen obtained from a neonate following in utero exposure to AMP-related compounds. Three new biomarkers, pOHMAMP, NOREPH, and HMMA, were quantified in meconium for the first time. The validated method will be used to assess in utero drug exposure to amphetamine-related compounds and characterize disposition of these analytes in meconium.
Acknowledgements
This research was supported by the Intramural Research Program, National Institute on Drug Abuse, National Institutes of Health. The authors would like to thank United States Drug Testing Laboratory for donating authentic meconium specimens.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Substance Abuse & Mental Health Services Administration. Rockville M.D: Office of Applied Studies; Results from the 2005 National Survey on Drug Use and Health: National Findings, NSDUH Series H-30, DHHS Publication No. SMA 06-4194. 2006 http://www.oas.samhsa.gov/nsduh/2k5nsduh/2k5Results.pdf.
- 2.Huestis MA, Choo RE. Forensic Sci. Int. 2002;128:20. doi: 10.1016/s0379-0738(02)00160-3. [DOI] [PubMed] [Google Scholar]
- 3.Smith LM, LaGasse LL, Derauf C, Grant P, Shah R, Arria A, Huestis M, Haning W, Strauss A, Della Grotta S, Liu J, Lester BM. Pediatrics. 2006;118:1149. doi: 10.1542/peds.2005-2564. [DOI] [PubMed] [Google Scholar]
- 4.Smith L, Yonekura ML, Wallace T, Berman N, Kuo J, Berkowitz C. J. Dev. Behav. Pediatr. 2003;24:17. doi: 10.1097/00004703-200302000-00006. [DOI] [PubMed] [Google Scholar]
- 5.Little BB, Snell LM, Gilstrap LC., III Obstet. Gynecol. 1988;72:541. [PubMed] [Google Scholar]
- 6.Oro AS, Dixon SD. J. Pediatr. 1987;111:571. doi: 10.1016/s0022-3476(87)80125-7. [DOI] [PubMed] [Google Scholar]
- 7.Eriksson M, Larsson G, Zetterstrom R. Acta Obstet. Gynecol. Scand. 1981;60:253. doi: 10.3109/00016348109158127. [DOI] [PubMed] [Google Scholar]
- 8.Gareri J, Klein J, Koren G. Clin. Chim. Acta. 2006;366:101. doi: 10.1016/j.cca.2005.10.028. [DOI] [PubMed] [Google Scholar]
- 9.Moore C, Negrusz A, Lewis D. J. Chromatogr. B. 1998;713:137. doi: 10.1016/s0378-4347(97)00479-9. [DOI] [PubMed] [Google Scholar]
- 10.Ostrea EM, Romero A, Knapp DK, Ostrea AR, Lucena JE, Utarnchitt RB. J. Pediatr. 1994;124:477. doi: 10.1016/s0022-3476(94)70379-5. [DOI] [PubMed] [Google Scholar]
- 11.Pichini S, Pacifici R, Pellegrini M, Marchei E, Pérez-Alarcón E, Puig C, Vall O, García-Algar O. J. Chromatogr. B. 2003;794:281. doi: 10.1016/s1570-0232(03)00478-1. [DOI] [PubMed] [Google Scholar]
- 12.Pichini S, Marchei E, Pacifici R, Pellegrini M, Lozano J, García-Algar O. J. Chromatogr. B. 2005;820:151. doi: 10.1016/j.jchromb.2005.03.027. [DOI] [PubMed] [Google Scholar]
- 13.Xia Y, Wang PP, Bartlett MG, Solomon HM, Busch KL. Anal. Chem. 2000;72:764. doi: 10.1021/ac990201p. [DOI] [PubMed] [Google Scholar]
- 14.ElSohly MA, Kopycki W, Feng S, Murphy TP. J. Anal. Toxicol. 1999;23:446. doi: 10.1093/jat/23.6.446. [DOI] [PubMed] [Google Scholar]
- 15.ElSohly MA, Stanford DF, Murphy TP, Lester BM, Wright LL, Smeriglio VL, Verter J, Bauer CR, Shankaran S, Bada HS, Walls HC. J. Anal. Toxicol. 1999;23:436. doi: 10.1093/jat/23.6.436. [DOI] [PubMed] [Google Scholar]
- 16.Oyler J, Darwin WD, Preston KL, Suess P, Cone EJ. J. Anal. Toxicol. 1996;20:453. doi: 10.1093/jat/20.6.453. [DOI] [PubMed] [Google Scholar]
- 17.Lewis DE, Moore CM, Leikin JB, Koller A. J. Anal. Toxicol. 1995;19:148. doi: 10.1093/jat/19.3.148. [DOI] [PubMed] [Google Scholar]
- 18.Browne S, Moore C, Negrusz A, Tebbett I, Covert R, Dusick A. J. Forensic Sci. 1994;39:1515. [PubMed] [Google Scholar]
- 19.Murphey LJ, Olsen GD, Konkol RJ. J. Chromatogr. 1993;613:330. doi: 10.1016/0378-4347(93)80150-3. [DOI] [PubMed] [Google Scholar]
- 20.Clark GD, Rosenzweig IB, Raisys VA, Callahan CM, Grant TM, Streissguth AP. J. Anal. Toxicol. 1992;16:261. doi: 10.1093/jat/16.4.261. [DOI] [PubMed] [Google Scholar]
- 21.Browne SP, Tebbett IR, Moore CM, Dusick A, Covert R, Yee GT. J. Chromatogr. 1992;575:158. doi: 10.1016/0378-4347(92)80518-u. [DOI] [PubMed] [Google Scholar]
- 22.Franssen RME, Stolk LML, van den Brand W, Smit BJ. J. Anal. Toxicol. 1994;18:294. doi: 10.1093/jat/18.5.294. [DOI] [PubMed] [Google Scholar]
- 23.Pichini S, Pacifici R, Pellegrini M, Marchei E, Lozano J, Murillo J, Vall O, García-Algar Ó. Anal. Chem. 2004;76:2124. doi: 10.1021/ac035419x. [DOI] [PubMed] [Google Scholar]
- 24.Salem MY, Ross SA, Murphy TP, ElSohly MA. J. Anal. Toxicol. 2001;25:93. doi: 10.1093/jat/25.2.93. [DOI] [PubMed] [Google Scholar]
- 25.Choo RE, Murphy CM, Jones HE, Huestis MA. J. Chromatogr. B. 2005;814:369. doi: 10.1016/j.jchromb.2004.10.068. [DOI] [PubMed] [Google Scholar]
- 26.Stolk LM, Coenradie SM, Smit BJ, van As HL. J. Anal. Toxicol. 1997;21:154. doi: 10.1093/jat/21.2.154. [DOI] [PubMed] [Google Scholar]
- 27.Wingert WE, Feldman MS, Kim MH, Noble L, Hand I, Yoon JJ. J. Forensic Sci. 1994;39:150. [PubMed] [Google Scholar]
- 28.Ostrea EM, Knapp DK, Romero A, Montes M, Ostrea AR. J. Pediatr. 1994;124:471. doi: 10.1016/s0022-3476(94)70378-7. [DOI] [PubMed] [Google Scholar]
- 29.Alano MA, Ngougmna E, Ostrea EM, Konduri GG. Pediatrics. 2001;107:519. doi: 10.1542/peds.107.3.519. [DOI] [PubMed] [Google Scholar]
- 30.Bearer CF, Jacobson JL, Jacobson SW, Barr D, Croxford J, Molteno CD, Viljoen DL, Marais AS, Chiodo LM, Cwik AS. J. Pediatr. 2003;143:463. doi: 10.1067/S0022-3476(03)00442-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Moore C, Jones J, Lewis D, Buchi K. Clin. Chem. 2003;49:133. doi: 10.1373/49.1.133. [DOI] [PubMed] [Google Scholar]
- 32.Gray T, Huestis M. Anal. Bioanal. Chem. 2007;388:1455. doi: 10.1007/s00216-007-1228-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Caldwell J, Dring LG, Williams RT. Biochem. J. 1972;129:11. doi: 10.1042/bj1290011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Moffat AC, Osselton MD, Widdop B, Galichet LY. Clarke's analysis of drugs and poisons in pharmaceuticals, body fluids and postmortem material. London: Pharmaceutical Press; 2004. [Google Scholar]
- 35.Ensslin HK, Kovar K-A, Maurer HH. J. Chromatogr. B. 1996;683:189. doi: 10.1016/0378-4347(96)00129-6. [DOI] [PubMed] [Google Scholar]
- 36.Brunnenberg M, Lindenblatt H, Gouzoulis-Mayfrank E, Kovar KA. J. Chromatogr. B. 1998;719:79. doi: 10.1016/s0378-4347(98)00396-x. [DOI] [PubMed] [Google Scholar]
- 37.Kraemer T, Maurer HH. Ther. Drug Monit. 2002;24:277. doi: 10.1097/00007691-200204000-00009. [DOI] [PubMed] [Google Scholar]
- 38.de la Torre R, Farré M, Roset PN, Pizarro N, Abanades S, Segura M, Segura J, Camí J. Ther. Drug Monit. 2004;26:137. doi: 10.1097/00007691-200404000-00009. [DOI] [PubMed] [Google Scholar]
- 39.Moore C, Lewis D, Leikin J. Clin. Chem. 1995;41:1614. [PubMed] [Google Scholar]
- 40.Zhao Z, Liu J, LaGasse LL, Derauf C, Grant P, Shah R, Arria A, Haning W, Smith LM, Lester B, Huestis MA. presented at the Joint Meeting of SOFT and TIAFT. Washington DC: 2004. [Google Scholar]
- 41.Lester BM, ElSohly M, Wright LL, Smeriglio VL, Verter J, Bauer CR, Shankaran S, Bada HS, Walls HC, Huestis MA, Finnegan LP, Maza PL. Pediatrics. 2001;107:309. doi: 10.1542/peds.107.2.309. [DOI] [PubMed] [Google Scholar]
- 42.Matuszewski BK, Constanzer ML, Chavez-Eng CM. Anal. Chem. 2003;75:3019. doi: 10.1021/ac020361s. [DOI] [PubMed] [Google Scholar]
- 43.United States Department of Health and Human Services. Food and Drug Administration. Center for Drug Evaluation and Research. Center for Veterinary Medicine, Guidance for Industry: Bioanalytical Method Validation. 2001
- 44.International Conference on Harmonisation (ICH) Steering Committee. ICH Harmonised Tripartite Guideline. Validation of analytical procedures: Methodology; 1996. [Google Scholar]
- 45.Commission of the European Communities. Official Journal of the European Communities. 2002;221:8. [Google Scholar]
- 46.United States Department of Health and Human Services. Food and Drug Administration. Center for Veterinary Medicine, Guidance for Industry: Mass spectrometry for confirmation of the identity of animal drug residues. 2003 http://www.fda.gov/cvm/Guidance/guide118.pdf.
- 47.World Anti-Doping Agency. Identification criteria for qualitative assays incorporating chromatography and mass spectrometry. 2003 WADA Technical Document TD2003IDCR http://www.wada-ama.org/rtecontent/document/criteria_1_2.pdf.
- 48.Pichini S, Puig C, Zuccaro P, Marchei E, Pellegrini M, Murillo J, Vall O, Pacifici R, García-Algar Ó. Forensic Sci. Int. 2005;153:59. doi: 10.1016/j.forsciint.2005.04.013. [DOI] [PubMed] [Google Scholar]
- 49.Staack RF, Fehn J, Maurer HH. J. Chromatogr. B. 2003;789:27. doi: 10.1016/s1570-0232(02)01018-8. [DOI] [PubMed] [Google Scholar]
- 50.Römhild W, Krause D, Bartels H, Ghanem A, Schöning R, Wittig H. Forensic Sci. Int. 2003;133:101. doi: 10.1016/s0379-0738(03)00055-0. [DOI] [PubMed] [Google Scholar]