Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2009 Oct 1.
Published in final edited form as: J Am Soc Mass Spectrom. 2008 Jun 28;19(10):1527–1534. doi: 10.1016/j.jasms.2008.06.013

Construction of a Versatile High Precision Ambient Ionization Source for Direct Analysis and Imaging

Jason S Sampson 1, Adam M Hawkridge 1, David C Muddiman 1,*
PMCID: PMC2605079  NIHMSID: NIHMS73851  PMID: 18657438

Abstract

The design and construction of a high precision ambient ionization source matrix-assisted laser desorption electrospray ionization (MALDESI) is described in full detail including a complete parts list. The computer controlled high precision motion control system and high repetition rate Explorer laser are demonstrated during MALDESI-FT-ICR analysis of peptides and proteins ranging from 1–17 kDa. The high stability ionization source platform described herein demonstrates both the advantages of the new MALDESI source and versatility for application to numerous desorption and ionization techniques.

Keywords: MALDESI, DESI, Ambient Ionization, FT-ICR

Introduction

The introduction and development of hybrid ambient ionization sources such as laser desorption atmospheric pressure ionization (LDAPI)[1, 2], fused droplet electrospray ionization/extraction electrospray (FD-ESI, EESI)[3, 4], direct analysis in real time (DART)[5], desorption electrospray ionization (DESI)[6], electrospray assisted laser desorption ionization (ELDI)[711], matrix-assisted laser desorption electrospray ionization (MALDESI)[12, 13] and infrared laser desorption electrospray ionization [14, 15] has advanced the capabilities of mass spectrometry. MALDESI for example is a pulsed ionization source that holds promise in areas ranging from top-down proteomics, tissue imaging, and ionization mechanism elucidation. The pulsed nature of MALDESI combined with its ability to generate multiply-charged ions are characteristics that are particularly well suited for Fourier transform ion cyclotron resonance (FT-ICR) and Orbitrap (LTQ-Orbi) mass spectrometry due to the inverse relationship of frequency to m/z and √(m/z), respectively. The consequence of this relationship is high resolving power, < 3 ppm mass measurement accuracy, and amenability to a multitude of tandem MS/MS techniques for bottom-up and top-down proteomics (e.g., CID, ETD, ECD, SORI, and IRMPD).[1618] Many of these MS/MS techniques in FT-ICR and Orbitrap instruments would benefit from a pulsed ionization source where intact proteins and polypeptides with complex post-translational modifications could be carefully interrogated rather than the typical “continuous” ionization encountered during a LC-MS/MS analysis where peak widths (i.e. analysis times) are on the order of 10–30 seconds.

The advancement and acceptance of MALDESI and related hybrid ionization techniques are critically dependent on the widespread dissemination of detailed source designs such that results can be reproduced and improved in a variety of laboratories. Furthermore, there are certainly unanticipated potential applications that could benefit from novel or improved source designs. Herein we provide a detailed description and characterization of the third version of the MALDESI source utilized in this laboratory. The further improvement to this design both within our lab as well as other laboratories should enable MALDESI and related hybrid ambient ionization sources to mature to a level of analytical robustness now widely enjoyed for ESI and MALDI.

Experimental

Materials

Bradykinin, angiotensin I, melittin, glucagon, bovine ubiquitin, lysozyme, myoglobin and 2,5-dihydroxybenzoic acid were purchased from Sigma-Aldrich (St. Louis, MO, USA) and used without further purification. HPLC-grade acetonitrile and high purity water were purchased from Burdick and Jackson (Muskegon, MI, USA). The electrospray solution was prepared by mixing acetonitrile and water 1:1 (v:v). The organic matrix solution was prepared by dissolving 150 mg DHB into 1 mL of the electrospray solution. All samples were prepared from 200 µM stock solutions and mixed 1:1 (v:v) with the matrix solution. Sample spots of 0.8 µL of the analyte matrix solution were deposited for each sample, equivalent to ~80 picomoles analyte per spot.

Profilometery Measurements

A Tencor Alpha Step profilometer (5 nm resolution) was used to measure the diameter of the laser ablation crater in a gold coated QCM electrode. The distance was measured across the width of the laser ablation craters with 3 repeats at a scan speed of 50 µm/second. The QCM electrode (International Crystal Manufacturing, Oklahoma City, OK, USA) consisted of 10 MHz A/T cut quartz crystal with a 1000 angstrom gold layer coated on a chromium base layer.

MALDESI-LTQ-FT Mass Spectrometry

MALDESI mass spectra were obtained using a hybrid LTQ-FT Ultra mass spectrometer (Thermo Electron Inc., San Jose, CA, USA and Bremen, Germany) equipped with an actively-shielded 7 T superconducting magnet (Oxford Instruments, Concord, MA, USA). The MALDESI source (Figure 1) was placed in front of the LTQ-FT fitted with a modified extended ion transfer capillary (Part#54). Solvent was electrosprayed at 2.8 kV through a 75 µm i.d. fused silica capillary (Part#40) with a 30 µm fused silica tapered PicoTip (Part#44) using a Harvard PHD-2000 syringe pump (Part#28) at a flow rate of 400 nL/minute. The analyte was laser desorbed from each sample spot actively-dried[ 19] onto a stainless steel sample target (Part#36) located directly below and between the ESI emitter and the ion transfer capillary. The stainless steel sample target was biased at 300 volts. Each mass spectrum was single-acquisition with resolving power at 400 m/z set to 200,000fwhm and the AGC is set to 1 × 106.

Figure 1.

Figure 1

Photograph of the new MALDESI ionization source coupled to a LTQ-FT mass spectrometer with major parts indicated and part numbers that correspond to those listed in Table 1.

Source Design

MALDESI Source Design and Construction

A new version of the MALDESI source was constructed based in part on the previous source design.[12, 13] A photo of the new source is shown in Figure 1 with major parts indicated. A complete parts list, grouped by manufacturer, is included in Table 1, with specifications for custom fabricated parts described in Table 2. A schematic of the ionization source labeled by part number is included in the Supplemental Material (Figure S1). The manual linear XYZ sample positioning stages were replaced with computer controlled motorized sample positioning stages (Parts#2–5) and control unit (Part#6) for precise sample positioning during analysis. The high speed long travel translational stages (Parts#2, 3) allow for interrogation of samples across the full length and width of a standard MALDI target (Part#36). Custom translation programming was accomplished using the supplied software (Newport, ESP); enabling one touch program execution for whole sample spot analysis (a diagram of the ablation path is shown in the Supplemental Material (Figure S2)) as well as allowing the user to operate the source remotely. The position control can be actuated linearly without program execution for spot to spot analysis, as typically performed in MALDI. The high precision (0.035 µm resolution, 100 nm increments) of the positioning system enables accurate repeatable (± 600 nm, bi-directional) positioning amenable to applications requiring a high degree of positioning control (e.g., tissue imaging).

Table 1.

MALDESI Parts List

Label Description Distributor Part # Quantity
1 Explorer Q-Switched DPSS Laser Newport (Irvine, CA) EXPL-349-120-1KE 1
2 High performance low profile linear stage Newport (Irvine, CA) 436 3
3 Motorized actuator (X and Y axes) Newport (Irvine, CA) LTA-HS 2
4 Vernier Micrometer (Z axis) Newport (Irvine, CA) SM-50 1
5 Angle bracket (90 deg) Newport (Irvine, CA) 360-90 1
6 2 Axis motion controller/driver Newport (Irvine, CA) ESP300-11N1N1 1
7 19 in. rack mount brackets (ESP300) Newport (Irvine, CA) ESP300-R 1
8 Slotted base Newport (Irvine, CA) B-05A 3
9 2 in. post holder Newport (Irvine, CA) VPH-2 3
10 2 in. ss post Newport (Irvine, CA) SP-2 1
11 4 in. ss post Newport (Irvine, CA) SP-4 1
12 6 in. ss post Newport (Irvine, CA) SP-6 2
13 12 in. ss post Newport (Irvine, CA) SP-12 2
14 Right angle post connector Newport (Irvine, CA) CA-1 1
15 Adjustable angle post connector Newport (Irvine, CA) CA-2 1
16 1 in. lens mount Newport (Irvine, CA) LH-1 1
17 UV fused silica plano convex lens AR10 Newport (Irvine, CA) SPX017 + AR.10 1
18 UV enhanced aluminum mirror Newport (Irvine, CA) 10D20AL.2 2
19 Performance plus breadboard Thorlabs (Newton, NJ) PBI11111 1
20 Passive support frame (27.5"H × 36"L × 30"W) Thorlabs (Newton, NJ) PFP51505 1
21 Periscope assembly Thorlabs (Newton, NJ) RS99 1
22 4 in. stainless steel post Thorlabs (Newton, NJ) RS4 1
23 4 in. mounting post Thorlabs (Newton, NJ) P4 4
24 Swivel casters (3 in.) Grainger 1G196 4
25 Caster brake kit Grainger 4X698 4
26 Power strip (6 plug) Grainger 6X953 2
27 DC power supply Analytica of Branford 103510 1
28 PHD 2000 Syringe pump Harvard Apparatus 70–2000 1
29 Fiber-Lite Light source Dolan Jenner MI-150 1
30 Dell laptop computer (Pentium M, 1.4 GHz) Dell N/A 1
31 Dell 19 in. monitor Dell N/A 1
32 CCD Camera Hitachi KP-M1AN 1
33 Leica optical amplifier Vashaw Scientific 312996 1
34 Leica Monozoom 7 N/A N/A 1
35 Happauge Win-TV USB-2 Circuit City N/A 1
36 SS 192 well MALDI plate Applied Biosystems 4333375 1
37 RS232 to USB converter Tiger Direct N/A 1
38 High Voltage wire (5kV, 2 pcs. 12 in., 24 in.) N/A N/A 1
39 LTQ high voltage plug Connectronics Corp. 10334-02 1
40 24 in. fused silica capillary (75 um i.d.) Polymicro Technologies 2000019 1
41 Stainless steel union Valco Instrument Co. ZUIXC 1
42 Syringe adapter Valco Instrument Co. VISF-2 1
43 PEEK tubing sleeve orange 0.062×0.016 Upchurch Scientific F-230 2
44 Silica tip (30um i.d. tapered) New Objective Inc. FS360-75-30-N-20 1
45 Hamilton gas tight syringe 100 uL Fisher Scientific 14-813-138 1
46 Breadboard Fabricated in house N/A 1
47 Camera support bracket Fabricated in house N/A 1
48 Teflon ESI holder Fabricated in house N/A 1
49 Sample target high voltage clip Fabricated in house N/A 1
50 19 in. rack support bracket Fabricated in house N/A 2
51 Sample target right angle bracket Fabricated in house N/A 1
52 Sample plate Teflon insulator Fabricated in house N/A 1
53 Computer shelf Fabricated in house N/A 1
54 Extended capillary Fabricated in house N/A 1
55 Laser spacing block Fabricated in house N/A 1

Table 2.

Description of parts fabricated in-house

46 16 in. × 16 in. × 3/8 in. thickness aluminum (thickness may be reduced to 1/4 in.), drilled and tapped 1/4 × 20 (1" on center)
47 8 in. × 1/2 in. thickness aluminum, two piece clamp for Monozoom lens support
48 2 in. × 1/2 in. × 1/4 in. thickness Teflon; cut-out to receive a stainless steel union, threaded tube inserted across width to attach aluminum retainer and electrical contact for ESI voltage
49 1¼ in. × 1/8 in. × 1/32 in. thickness stainless steel with custom bend for electrical contact with sample target
50 16 in. × 2½ in. (1½ in. along component attachment region) × 1/4 in. thickness aluminum, slots at each end for attachment to support frame (Part#20)
51 3 in. length × 2 ½ in. height, 1/4 in. thickness aluminum, (2) vertical slots cut in short side for attachment to XYZ stage, (2) clear holes (#8) in long side for Teflon target insulator attachment
52 2½ in. × 2½ in. × 1/4 in. thickness Teflon with (2) small stainless steel posts to accommodate the sample target and (2) holes drilled and tapped 8/32 in. from bottom to attach to sample target bracket (Part#51)
53 1/4 in. thickness aluminum, cut to fit across upper cross members of support frame (Part#20)
54 2 in. extended stainless steel capillary with lug to fit LTQ atmospheric pressure interface
55 7½ in. × 3¾ in. × 1½ in. aluminum block including ½ in. slots on each end for ¼ × 20 screw attachment to the breadboard (Part#19)

The manually actuated pulsed nitrogen laser (337 nm) was replaced by the Explorer laser system (Part#1), a Q-switched diode pumped solid-state ultraviolet laser (349 nm) with first pulse suppression, precise internally measured laser power (0–120 µJ) and repetition rate control (0–5000 Hz) using the supplied software (Spectra-Physics, L-Win). The laser power, repetition rate and translation stage velocity control (0–5 mm/s) are invaluable for applications to various samples, surfaces and desorption conditions; preventing damage to solid substrate and depletion of analyte during analysis. The high resolution CCD camera (Parts#32–34) was connected using the WinTV PCI tuner (Part#35) to the laptop computer and monitor (Parts#30, 31) enabling real time on screen visualization and recording of the sample spot and tracking of the laser spot on target during laser ablation, which is particularly useful for analysis of small sample spots as well as tissue sections for imaging.

The XYZ stage (Parts#2–5, 38, 49, 51, 52), second stage laser beam positioning mirror (Part#18, mounted in the upper portion of the periscope assembly Parts#21, 11, 8), focusing lens assembly (Parts#8, 9, 11, 13, 14, 16, 17), ESI emitter assembly (Parts#8, 9, 12, 12, 15, 40–44, 48) and CCD camera assembly (Parts#8, 9, 10, 13, 32–34, 47) were all mounted onto the custom fabricated breadboard (Part#46). A schematic of the desorption and ionization region with a description of each sub-assembly are shown in Figure 2 and Figure 3. A clear hole drilled in the breadboard (Part#46) allows the passage of the laser beam from the first stage directional mirror (Parts#18, 21, 22) mounted on the main breadboard work surface (Part#19) to the second stage directional mirror (Parts#8, 11, 18, 21) mounted on the custom breadboard (Part#46). The breadboard (Part#46) with all mounted equipment was then mounted onto the vibrationally isolated main breadboard (Part#19) using 4 inch offset posts (Part#23). The laser (Part#1) mounted to a custom fabricated spacing block (Part#55), first stage laser beam positioning mirror (Part#18 mounted in lower portion of the periscope assembly, Part#21, attached to a 4 inch post, Part#22), syringe pump (Part#28), light source (Part#29) and monitor (Part#31) were mounted on the main breadboard (Part#19). The field of the main breadboard (Part#19) and custom breadboard (Part#46) were fabricated with tapped holes (1/4 × 20) on one inch center; a supply of appropriate fasteners is required to attach various components to the breadboards.

Figure 2.

Figure 2

Side view schematic of the ionization region of the MALDESI source with the numbers corresponding to the parts listed in Table 1.

Figure 3.

Figure 3

Top view schematic of the ionization region with a listing of the parts for each sub-assembly.

The high voltage power supply (Part#27) and motion controller (Part#6) were mounted to the support frame (Part#20) using custom fabricated rack mounting brackets (part#50). The high voltage power supply (Part#27) was attached to the sample target high voltage clip (Part#49) using 5 kV insulated wire (Part#38) for contact with the stainless steel sample target (Part#36) mounted on the Teflon insulator (Part#52). The keyboard, mouse and laptop computer (Part#30) were placed onto a custom fabricated shelf (part#53) situated on the upper cross beams of the support frame (Part#20). Power management was provided using two 6 plug power strips (Part#26) installed on the inside of the vertical supports on the rear of the support frame (Part#20). Prior to assembly, casters (Parts#24, 25) were installed to replace the existing adjustable feet on the support frame for mobility. The versatility of this ionization source platform allows for interchangeability of the source between instruments as well as installation of new components onto the existing platform.

Desorption and Ionization Source Versatility

The high stability ionization source platform may be implemented using various desorption and ionization techniques for application specific analysis. Existing ionization sources may be installed onto the platform due to the inherent versatility of the breadboard for roll-up accessibility. Analyte desorption may be induced by one of a number of regimes including laser desorption as in MALDESI, laser induced acoustic desorption (LIAD),[20] heated nitrogen gas as in ASAP[21] and impact of high velocity charged droplets as in DESI[6] followed by a number of post-desorption ionization methods including atmospheric pressure photo ionization (APPI)[22], atmospheric pressure chemical ionization (APCI)[23] and electrospray ionization (ESI)[24]. In addition to the hybrid ionization sources listed above, this platform is amenable to basic electrospray ionization[24] and atmospheric pressure matrix-assisted laser desorption ionization.[25]

Results and Discussion

Explorer Laser Beam Characterization

The laser beam spot size was measured at a laser power of 112 µJ (measured internally) using a gold coated quartz crystal microbalance electrode as the laser target. A photo of the QCM electrode following laser ablation is shown in Figure 4B, with an expanded view of a single ablation crater shown in Figure 4C. In these experiments, the thin gold top layer (1000 angstroms) was removed exposing the chromium base layer with some re-deposition along the perimeter of the ablated crater. The elliptical laser beam spot size (60 µm × 80 µm) was determined by measuring the actual diameter of the ablation craters using a Tencor Alpha Step profilometer. A representative scan plot is shown in Figure 4A.

Figure 4.

Figure 4

A) Profilometer scan plot across the width (60 µm) of the laser ablation crater in the gold QCM electrode. B) Photograph of the QCM electrode after laser desorption. C) Expanded view of an elliptical ablation crater, the diameters (60 µm × 80 µm) were measured for multiple ablation craters across the QCM electrode.

MALDESI-LTQ-FT-ICR of Peptides and Proteins

MALDESI-FT-ICR mass spectra of peptides and proteins (1 – 8.6 kDa) including bradykinin, angiotensin I, melittin, glucagon and ubiquitin each mixed with organic matrix have been shown previously.[12, 13] Comparable data was obtained using the new source; representative MALDESI-FT-ICR mass spectra of bradykinin and melittin both mixed with DHB included in Supplemental Material (Figure S3). Lysozyme and myoglobin (not shown previously) were each mixed 1:1 (v:v) with DHB, 0.8 µL was deposited onto the sample target and actively dried.[19] The ESI solution flow-rate was set to 400 nL/minute and stable electrospray was obtained. The motion controller was pre-programmed to raster the sample surface under the laser beam as illustrated in the Supplemental Material (Figure S2). The laser was actuated using computer control at a laser power of 50 µJ (measured internally) and repetition rate of 10 Hz and the motion program initiated. Multiply-charged ions were generated and detected for each peptide and protein; demonstrating a molecular weight range from 1 – 17 kDa using this source. Representative MALDESI-FT-ICR mass spectra of myoglobin and lysozyme from this experiment are shown in Figure 5, with an observed resolving power of ~ 40,000.

Figure 5.

Figure 5

MALDESI-FT-ICR mass spectra of 0.8 µL actively dried spot of 200 µM A) myoglobin (16.9 kDa) and B) lysozyme C (14.3 kDa) mixed 1:1 (v:v) with 150 mg/mL DHB.

The amount of material ablated during analysis was calculated using the amount of material spotted (~80 picomoles), collection time (200 ms), laser repetition rate (10 Hz, 2 shots per collection), area of the laser spot (3846 µm2) and the area of the dried spot (A = 3.14 mm2); 196 femtomoles were removed per spectrum, assuming uniform distribution of analyte and complete removal at each laser shot. The development and implementation of molecular transport devices such as the air amplifier,[26, 27] air ejector[28] and RASTIR[29] which enable the efficient transport of ions into the mass spectrometer have been investigated and may prove important to increasing sensitivity.

Liquid Drop Sample Analysis

Liquid drop analysis of 0.8 µL droplets of 200 µM ubiquitin and myoglobin mixed 1:1 (v:v) with DHB (150 mg/mL) and deposited onto the sample target, biased at 300 V, for immediate analysis is demonstrated using MALDESI with ESI post-ionization. Liquid drop laser desorption analysis with ESI post-ionization has been demonstrated previously.[11, 30] The ESI emitter was biased at 2.8 kV to electrospray 50 % acetonitrile in water at a flow rate of 400 nL/minute. The liquid sample was continuously irradiated using the Explorer laser without moving the sample target, yielding relatively constant ion abundance over the lifetime of the droplet (approximately 30 seconds). liq-MALDESI-FT-ICR mass spectra of myoglobin and ubiquitin with ESI post-ionization are shown in Supplemental Material (Figure S4), A and B, respectively.

Conclusions

The MALDESI ionization source described herein couples high stability with precision motion and laser control. This design enables high mass resolving power analysis of biological molecules including intact and top-down characterization in addition to facilitating potential imaging applications.

The mobile design provides a stable analytical platform which may be used to implement a number of desorption regimes including laser desorption (e.g., MALDESI), high velocity solvent droplets (e.g., DESI), a stream of heated nitrogen gas (e.g., ASAP) and post desorption ionization using electrospray ionization (ESI), chemical (e.g., APCI) and photon ionization (e.g., APPI) for analysis of various types and classes of molecules on multiple MS platforms. The modular configuration of the MALDESI ionization source can facilitate the substitution of parts as necessary for specific applications or budgeting constraints. Furthermore, an IR laser could be mounted to the laser table breadboard for infrared laser MALDESI applications (IR-LDESI).[14, 15]

Supplementary Material

01

Acknowledgments

The authors thank Dr. John A. Mclean of Vanderbilt University for assistance in acquiring the Explorer laser system, Dr. Fred Stevie in the Analytical Instrumentation Facility at North Carolina State University for assistance with the profilometer measurements and the machine shop personnel in the College of Physical and Mathematical Sciences at North Carolina State University for their assistance with the MALDESI source construction. The authors gratefully acknowledge financial support received from the National Cancer Institute, National Institutes of Health (R33 CA105295), the W.M. Keck Foundation, the William R. Kenan, Jr. Fund for Engineering, Technology & Science, and North Carolina State University.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.Kolaitis L, Lubman DM. Detection of Nonvolatile Species by Laser Desorption Atmospheric-Pressure Mass-Spectrometry. Analytical Chemistry. 1986;58(11):2137–2142. [Google Scholar]
  • 2.Coon JJ, McHale KJ, Harrison WW. Atmospheric pressure laser desorption/chemical ionization mass spectrometry: a new ionization method based on existing themes. Rapid Communications in Mass Spectrometry. 2002;16(7):681–685. doi: 10.1002/rcm.626. [DOI] [PubMed] [Google Scholar]
  • 3.Chang DY, Lee CC, Shiea J. Detecting large biomolecules from high-salt solutions by fused-droplet electrospray ionization mass spectrometry. Analytical Chemistry. 2002;74(11):2465–2469. doi: 10.1021/ac010788j. [DOI] [PubMed] [Google Scholar]
  • 4.Chen HW, Venter A, Cooks RG. Extractive electrospray ionization for direct analysis of undiluted urine, milk and other complex mixtures without sample preparation. Chemical Communications. 2006;(19):2042–2044. doi: 10.1039/b602614a. [DOI] [PubMed] [Google Scholar]
  • 5.Cody RB, Laramee JA, Durst HD. Versatile new ion source for the analysis of materials in open air under ambient conditions. Analytical Chemistry. 2005;77(8):2297–2302. doi: 10.1021/ac050162j. [DOI] [PubMed] [Google Scholar]
  • 6.Takats Z, Wiseman JM, Gologan B, Cooks RG. Mass spectrometry sampling under ambient conditions with desorption electrospray ionization. Science. 2004;306(5695):471–473. doi: 10.1126/science.1104404. [DOI] [PubMed] [Google Scholar]
  • 7.Shiea J, Huang MZ, Hsu HJ, Lee CY, Yuan CH, Beech I, Sunner J. Electrospray-assisted laser desorption/ionization mass spectrometry for direct ambient analysis of solids. Rapid Communications in Mass Spectrometry. 2005;19(24):3701–3704. doi: 10.1002/rcm.2243. [DOI] [PubMed] [Google Scholar]
  • 8.Huang MZ, Hsu HJ, Lee JY, Jeng J, Shiea J. Direct Protein Detection from Biological Media through Electrospray-Assisted Laser Desorption Ionization/Mass Spectrometry. Journal of Proteome Research. 2006;5(5):1107–1116. doi: 10.1021/pr050442f. [DOI] [PubMed] [Google Scholar]
  • 9.Huang MZ, Hsu HJ, Wu CI, Lin SY, Ma YL, Cheng TL, Shiea J. Characterization of the chemical components on the surface of different solids with electrospray-assisted laser desorption ionization mass spectrometry. Rapid Communications in Mass Spectrometry. 2007;21(11):1767–1775. doi: 10.1002/rcm.3011. [DOI] [PubMed] [Google Scholar]
  • 10.Lin SY, Huang MZ, Chang HC, Shiea J. Using electrospray-assisted laser desorption/ionization mass spectrometry to characterize organic compounds separated on thin-layer chromatography plates. Analytical Chemistry. 2007;79(22):8789–8795. doi: 10.1021/ac070590k. [DOI] [PubMed] [Google Scholar]
  • 11.Peng IX, Shiea J, Loo RRO, Loo JA. Electrospray-assisted laser desorption/ionization and tandem mass spectrometry of peptides and proteins. Rapid Communications in Mass Spectrometry. 2007;21(16):2541–2546. doi: 10.1002/rcm.3154. [DOI] [PubMed] [Google Scholar]
  • 12.Sampson JS, Hawkridge AM, Muddiman DC. Generation and detection of multiply-charged peptides and proteins by matrix-assisted laser desorption electrospray ionization (MALDESI) Fourier transform ion cyclotron resonance mass spectrometry. Journal of the American Society for Mass Spectrometry. 2006;17(12):1712–1716. doi: 10.1016/j.jasms.2006.08.003. [DOI] [PubMed] [Google Scholar]
  • 13.Sampson JS, Hawkridge AM, Muddiman DC. Direct characterization of intact polypeptides by matrix-assisted laser desorption electrospray ionization quadrupole Fourier transform ion cyclotron resonance mass spectrometry. Rapid Communications in Mass Spectrometry. 2007;21(7):1150–1154. doi: 10.1002/rcm.2947. [DOI] [PubMed] [Google Scholar]
  • 14.Nemes P, Vertes A. Laser ablation electrospray ionization for atmospheric pressure, in vivo, and imaging mass spectrometry. Analytical Chemistry. 2007;79(21):8098–8106. doi: 10.1021/ac071181r. [DOI] [PubMed] [Google Scholar]
  • 15.Rezenom YH, Dong J, Murray KK. Infrared laser-assisted desorption electrospray ionization mass spectrometry. Analyst. 2008;133(2):226–232. doi: 10.1039/b715146b. [DOI] [PubMed] [Google Scholar]
  • 16.Marshall AG, Guan SH. Advantages of high magnetic field for Fourier transform ion cyclotron resonance mass spectrometry. Rapid Communications in Mass Spectrometry. 1996;10(14):1819–1823. doi: 10.1002/(SICI)1097-0231(199611)10:14<1829::AID-RCM697>3.0.CO;2-U. [DOI] [PubMed] [Google Scholar]
  • 17.Kelleher NL, Lin HY, Valaskovic GA, Aaserud DJ, Fridriksson EK, McLafferty FW. Top down versus bottom up protein characterization by tandem high-resolution mass spectrometry. Journal of the American Chemical Society. 1999;121(4):806–812. [Google Scholar]
  • 18.Kelleher NL. Top-down proteomics. Analytical Chemistry. 2004;76(11):196A–203A. [PubMed] [Google Scholar]
  • 19.Williams TI, Saggese DA, Wilcox RJ, Martin JD, Muddiman DC. Effect of matrix crystal structure on ion abundance of carbohydrates by matrix-assisted laser desorption/ionization Fourier transform ion cyclotron resonance mass spectrometry. Rapid Communications in Mass Spectrometry. 2007;21(5):807–811. doi: 10.1002/rcm.2904. [DOI] [PubMed] [Google Scholar]
  • 20.Golovlev VV, Allman SL, Garrett WR, Taranenko NI, Chen CH. Laser-induced acoustic desorption. International Journal of Mass Spectrometry. 1997;169:69–78. [Google Scholar]
  • 21.McEwen CN, McKay RG, Larsen BS. Analysis of solids, liquids, and biological tissues using solids probe introduction at atmospheric pressure on commercial LC/MS instruments. Anal Chem. 2005;77(23):7826–7831. doi: 10.1021/ac051470k. [DOI] [PubMed] [Google Scholar]
  • 22.Robb DB, Covey TR, Bruins AP. Atmospheric pressure photoionization: an ionization method for liquid chromatography-mass spectrometry. Anal Chem. 2000;72(15):3653–3659. doi: 10.1021/ac0001636. [DOI] [PubMed] [Google Scholar]
  • 23.Shahin MM. Mass-Spectrometric Studies of Corona Discharges in Air at Atmospheric Pressures. Journal of Chemical Physics. 1966;45(7) 2600-&. [Google Scholar]
  • 24.Fenn JB, Mann M, Meng CK, Wong SF, Whitehouse CM. Electrospray Ionization for Mass-Spectrometry of Large Biomolecules. Science. 1989;246(4926):64–71. doi: 10.1126/science.2675315. [DOI] [PubMed] [Google Scholar]
  • 25.Laiko VV, Baldwin MA, Burlingame AL. Atmospheric pressure matrix assisted laser desorption/ionization mass spectrometry. Analytical Chemistry. 2000;72(4):652–657. doi: 10.1021/ac990998k. [DOI] [PubMed] [Google Scholar]
  • 26.Zhou L, Yue BF, Dearden DV, Lee ED, Rockwood AL, Lee ML. Incorporation of a venturi device in electrospray ionization. Analytical Chemistry. 2003;75(21):5978–5983. doi: 10.1021/ac020786e. [DOI] [PubMed] [Google Scholar]
  • 27.Dixon RB, Muddiman DC, Hawkridge AM, Fedorov AG. Probing the mechanisms of an air amplifier using a LTQ-FT-ICR-MS and fluorescence spectroscopy. J Am Soc Mass Spectrom. 2007;18(11):1909–1913. doi: 10.1016/j.jasms.2007.08.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Dixon RB, Bereman MS, Muddiman DC, Hawkridge AM. Remote mass spectrometric sampling of electrospray- and desorption electrospray-generated ions using an air ejector. J Am Soc Mass Spectrom. 2007;18(10):1844–1847. doi: 10.1016/j.jasms.2007.07.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Dixon RB, Sampson JS, Hawkridge AM, Muddiman DC. Novel Ambient Aerodynamic Ionization Source for Remote Analyte Sampling and Mass Spectrometric Analysis. Analytical Chemistry. 2008 doi: 10.1021/ac800289f. Accepted. [DOI] [PubMed] [Google Scholar]
  • 30.Sampson JSH, A M, Muddiman DC. Development and Characterization of a New Ionization Technique for Analysis of Biological Macromolecules: Liquid Matrix-Assisted Laser Desorption Electrospray Ionization. Analytical Chemistry. 2008 doi: 10.1021/ac8001935. Accepted. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

01

RESOURCES