Collaborative evaluation of the Radiometer Sensititre AP80 for identification of gram-negative bacilli.

J L Staneck 1, L S Weckbach 1, R C Tilton 1, R J Zabransky 1, L Bayola-Mueller 1, C M O'Hara 1, J M Miller 1
PMCID: PMC262899  PMID: 8501217

Abstract

A multicenter trial of the Sensititre AP80 panel read on the Sensititre AutoReader (Radiometer America, Westlake, Ohio) for the automated identification of gram-negative bacilli was conducted with 1,023 clinical isolates (879 members of the family Enterobacteriaceae plus 144 nonenteric organisms). Assignment of taxa was based on the computer-assisted interpretation of the results of a series of reactions with fluorogenic enzyme substrates after 5 h of incubation, with an incubation interval of approximately 18 h used when indicated. Accuracy was determined initially by comparison with the results obtained with the API 20E or Rapid NFT system (Analytab Products, Plainview, N.Y.). Isolates showing discrepancies were identified by using conventional biochemical profiles. Identifications were available after 5 h of incubation for 918 isolates (90%). Agreements with reference results for members of the family Enterobacteriaceae were 95.3 and 92.5% at the genus and species levels, respectively, and for the nonmembers of the family Enterobacteriaceae, the agreements with reference results were 95.1 and 84.7%, respectively. The Sensititre AP80 panel was found to be simple and convenient to use, allowed for the testing of three isolates per panel, required minimal supplementary testing for completion of identification, performed in a reproducible fashion, and demonstrated an accuracy of same-day identification comparable to that reported for other automated systems. The AP80 panel appears well suited for routine use in the clinical microbiology laboratory as an automated means of identifying both members of the family Enterobacteriaceae and nonenteric gram-negative bacilli.

Full text

PDF
1179

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. DiPersio J. R., Dyke J. W., Vannest R. D. Evaluation of the updated MS-2 Bacterial Identification system in comparison with the API 20E system. J Clin Microbiol. 1983 Jul;18(1):128–135. doi: 10.1128/jcm.18.1.128-135.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Doern G. V., Staneck J. L., Needham C., Tubert T. Sensititre autoreader for same-day breakpoint broth microdilution susceptibility testing of members of the family Enterobacteriaceae. J Clin Microbiol. 1987 Aug;25(8):1481–1485. doi: 10.1128/jcm.25.8.1481-1485.1987. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Nolte F. S., Krisher K. K., Beltran L. A., Christianson N. P., Sheridan G. E. Rapid and overnight microdilution antibiotic susceptibility testing with the Sensititre Breakpoint Autoreader system. J Clin Microbiol. 1988 Jun;26(6):1079–1084. doi: 10.1128/jcm.26.6.1079-1084.1988. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Pfaller M. A., Sahm D., O'Hara C., Ciaglia C., Yu M., Yamane N., Scharnweber G., Rhoden D. Comparison of the autoSCAN-W/A rapid bacterial identification system and the Vitek AutoMicrobic system for identification of gram-negative bacilli. J Clin Microbiol. 1991 Jul;29(7):1422–1428. doi: 10.1128/jcm.29.7.1422-1428.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Staneck J. L., Allen S. D., Harris E. E., Tilton R. C. Automated reading of MIC microdilution trays containing fluorogenic enzyme substrates with the Sensititre Autoreader. J Clin Microbiol. 1985 Aug;22(2):187–191. doi: 10.1128/jcm.22.2.187-191.1985. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Staneck J. L., Allen S. D., Harris E. E., Tilton R. C. Rapid MIC testing with the sensititre autoreader. J Clin Microbiol. 1988 Jan;26(1):1–7. doi: 10.1128/jcm.26.1.1-7.1988. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. York M. K., Brooks G. F., Fiss E. H. Evaluation of the autoSCAN-W/A rapid system for identification and susceptibility testing of gram-negative fermentative bacilli. J Clin Microbiol. 1992 Nov;30(11):2903–2910. doi: 10.1128/jcm.30.11.2903-2910.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES