Abstract
Correlative light/electron microscopy (CLEM) allows the simultaneous observation of a given subcellular structure by fluorescence light microscopy (FLM) and electron microscopy. The use of this approach is becoming increasingly frequent in cell biology. In this study, we report on a new high data output CLEM method based on the use of cryosections. We successfully applied the method to analyze the structure of rough and smooth Russell bodies used as model systems. The major advantages of our method are (i) the possibility to correlate several hundreds of events at the same time, (ii) the possibility to perform three-dimensional (3D) correlation, (iii) the possibility to immunolabel both endogenous and recombinantly expressed proteins at the same time and (iv) the possibility to combine the high data analysis capability of FLM with the high precision–accuracy of transmission electron microscopy in a CLEM hybrid morphometry analysis. We have identified and optimized critical steps in sample preparation, defined routines for sample analysis and retracing of regions of interest, developed software for semi/fully automatic 3D reconstruction and defined preliminary conditions for an hybrid light/electron microscopy morphometry approach.
Keywords: confocal microscopy, correlative microscopy, cryosection, electron microscopy, immunofluorescence, immunogold, morphometry
In recent years, fluorescence light microscopy (FLM) has moved beyond the qualitative viewing of labeled objects. Improved optical and laser equipments, fluorescence-based labeling methods, computer technology and image analysis software have made possible the study of the location, orientation and dynamics of specific biomolecular structures from organelles to single particles (1). Despite the success and wide application (2) of wide-field microscopy (WFM), two-photon excitation microscopy and confocal laser scanning microscopy (CLSM), some limitations still remain. In particular, for typical wavelengths in the visible range, CLSM optimally resolves 180 nm in the focal plane (x,y) and 500–800 nm along the optical axis (z). In addition, fluorescence microscopy is unable to identify relationships between labeled and unlabeled cellular compartments.
Electron microscopy (EM) and immunoEM can complement FLM analysis because of their higher resolution and ability to visualize labeled and unlabeled cellular compartments. A number of correlative light/EM microscopy (CLEM) methods have been developed to fill the FLM–EM gap (3–11). CLEM can be ‘ideally’ defined as an imaging platform able to capture the same objects by means of multimodal microscopy techniques. This allows for deeper insights in the true size, shape and interconnections of subcellular structures as well as precise mapping of antigen localization. Coupling a dynamic (FLM) to a high-resolution (transmission electron microscopy, TEM) approach, CLEM methods can also add temporal dimensions to the analysis of in vivoevents (3–6,12–14). Notwithstanding the overall potentiality, CLEM methods still exhibit a number of problems. Among these are the low rate of data output, the low label sensitivity and specificity, the difficulty to perform simultaneous multiple labeling of endogenous and recombinantly expressed antigens or to correlate three-dimensional (3D) FLM with 3D TEM images (15).
We have designed and realized a novel CLEM approach (high data output CLEM, HDO-CLEM), able to potentially perform CLEM analysis on hundreds of events per study session. Our method allows the simultaneous labeling of both recombinantly tagged and endogenous antigens, at the FLM and TEM level, and to obtain the correlation of 3D FLM with 3D TEM images (3D CLEM). In this study, we provide the complete description and validation of the HDO-CLEM ‘package’ that includes process architecture, with defined sample preparation/labeling protocols and dedicated software for image analysis and 3D modeling.
Results
Model system
Rough endoplasmic reticulum (RER) and ER–Golgi intermediate compartment (ERGIC) (16) display a tubular network and assume diverse morphologies under specific conditions, such as increased protein storage (17). We have previously shown (17,18) that HeLa cells stably transfected with a mouse immunoglobulin μ chain, lacking the first constant domain (μ-ΔCh1), together with a wild-type immunoglobulin λ light chain (λ), display spherical dilated RER cisternae resembling the Russell bodies identified in plasmocitoma cells (rough Russell bodies, RRBs). RRBs are about 1 μm in diameter and contain protein aggregates of mutant immunoglobulin M (17,18). Conversely, when μ-ΔCh1 is transfected in the absence of λ, it aggregates in the smooth ERGIC tubular compartments (smooth Russell bodies, SRBs), as shown by the association with the ERGIC marker ERGIC-53 (17). SRBs are composed of narrow (∼100 nm wide, n-SRB) or dilated (∼200 nm wide, d-SRB), curled, tubular cisternae. When analyzed by 3D reconstruction of optical CLSM sections, SRBs, identified by immunofluorescence labeling of μ, showed tightly packed and fused tubular structures. TEM analysis recognized instead well-separated individual tubular structures (Figure 1A–C). The SRB morphological differences between CLSM and TEM are likely because of imaging artifacts linked to the low z-axis resolution of the CLSM.
Specimen preparation protocol and immunolabeling
We based our CLEM method on the use of ultrathin cryosections. Ultrathin cryosections, pioneered by Tokuyasu (19), widely used in combination with immunogold labeling (20), had already proven to be the method of choice for multiple TEM immunolabeling of endogenous and recombinantly expressed antigens and an excellent substrate for a highly detailed immunofluorescence microscopy (8,10,21).
Cryosections were cut randomly through a pellet of RRB- or SRB-expressing HeLa cells. Sixty or 200 nm thick sections, used for detailed two-dimensional (2D) or 3D analysis, respectively, were placed on gold finder grids and used for both FLM and TEM observations (Figure 1D–K). We labeled μ-ΔCh1 by immunofluorescence and immunogold on the same section either alone or in combination with the ER marker calreticulin (22). CY3- and CY2-labeled second-step antibodies and protein A-gold (PAG) were used as tracer for FLM and TEM, respectively (Figure 1). To reduce reflection background and spherical aberration artifacts during FLM observation, finder grids were mounted between a microscope slide and a coverslip. Application of a layer of methylcellulose on the grids guaranteed their prolonged and repeated use for FLM imaging, without significant fluorescence loss, avoiding potential damage during recovery for TEM analysis. Unexpectedly, the methylcellulose film greatly reduced the fluorescence photobleaching, making superfluous the use of anti-fading agents. The methylcellulose coat was washed out before TEM observation to allow for EM staining. Additional examples of labeling for recycling compartments, late endosomes/lysosomes and Golgi apparatus are shown in Figure S1.
FLM imaging: routine procedure for region of interest identification
To identify the highest number of region of interest (ROI) on a cryosection, we acquired FLM maps of the fields of interest. As proof of principle, we systematically screened the grids by the eyepiece (‘conventional’ approach) to generate a low-magnification (×20 objective) fluorescence map of selected groups of ROIs (Figure 2A). A subsequent high-magnification/high-resolution imaging (×100 objective) of each individual ROI was then recorded for the FLM/TEM correlation analysis (Figure 2B–I).
Subsequently, we designed an optimized ‘work-on-map’ procedure to increase the efficiency of FLM ROI identification and retracing at the TEM. Spatially consecutive high-magnification images (×60 objective) of single, or serial, sections were automatically acquired and merged by the ‘MosaicJ’ software (23) into a single high-resolution fluorescence map (Figure 2J). The automated work-on-map high-resolution procedure greatly shortened the time needed for image acquisition. Figure 2J shows a CLSM montage of an area including three serial cryosections and containing several potential ROIs (i.e. about 100 cells/grid mesh). Finder grid reference points, that is mesh bars and mesh letters, were included in the map to identify ROI positions (Figure 2J). Mesh position and nuclei, or RBs, patterns were essential to retrace selected ROIs at the TEM. Further FLM acquisition steps were decisive in the case of signal saturation, instability of focus position, poor pixel sampling or chromatic aberration problems in the initial imaging (Figure 2K–M).
FLM imaging: 3D volume rendering
Fluorescence light optical sectioning microscopy techniques (24) and in particular CLSM (24) are common approach to 3D analysis of biological samples. The 3D modeling obtained from optical sectioning helps to define the spatial organization of subcellular organelles inside the sample. However, the relative poor axial resolution of CLSM, because of out-of-focus fluorescence background, limits the definition of subcellular compartments. CLEM approach, combined with the use of serial 200 nm cryosections, can solve this problem because the thickness below the resolution of the CLSM reduces the out-of-focus background, improving the fidelity of the 3D FLM modeling (Figure S3).
We used the automated work-on-map approach to track down corresponding ROIs on ribbons of three to eight consecutive serial cryosections, laid on a single finder grid. The pattern of cell nuclei, or RBs, allowed an a posteriorimanual screening of the montage to identify the same ROI in serial sections. The 3D surface rendering from physical sections required a registration (e.g. aligning) process. For this reason, we developed a user-friendly software for registering series of bidimensional multichannel images. Registration (25) and rendering software were based on ImageJ(26) plug-ins, either modified or originally developed, and collected together in a software package named MicroSCoBiOJ.
Registration was performed between consecutive pair of images (a reference image and a source image). We set the ROI identified on the first section as reference image (Figure 3A,F) and the corresponding ROI on the following section as the source image (Figure 3B,G). For each consecutive pair of serial images, the software automatically iterated the process. Nuclei, or RBs, pattern was used as reference to fix landmarks and to register the respective channel of consecutive serial section images (Figure 3A–C,F–H). The software automatically refined landmarks to minimize the mean square differences between the target image and the warped image (25). These landmarks were successively used to register other objects displayed on separate channels, within the same ROI. When occasional dramatic differences in reference patterns occurred between consecutive cryosections, a manual mode of registration was preferred, disengaging the automatic refinement procedure. Surface rendering and navigation functions embedded in MicroSCoBiOJ originated the final 3D modeling of the ROI and the rotation along different axis (Figure 3D,E,I,J and Videos S2 and S3). As expected, 3D reconstruction from physical sectioning leads to a substantial increase in axial resolution of FLM (8) (Figure S3). As a result of the 3D reconstruction, we identified connections occurring between portions of the tubular network of SRBs, otherwise not identifiable by optical sectioning or analysis of single cryosections (Figure S4).
Correlation analysis: retracing and superimposition of ROI at TEM level
To efficiently retrace at the TEM level the ROIs identified at the FLM, we generated composite low-magnification TEM map images employing the EM compustage and the automated functions embedded in the EM camera software.
Grid mesh landmarks and nuclei patterns allowed an easy recognition of the ROI at the TEM level. TEM and FLM spatial ROI map overlay enabled the correlation of ROIs. This operation required resampling and registration of FLM images with TEM images as target (Figure 4A–I). FLM/TEM image superimposition and correlation showed that CLSM was not able to completely resolve the tubular shape of SRB structures. This was mostly because of the limited resolution of the CLSM compared with TEM (Figure 4G–I and Table S1).
However, because TEM images are 2D superimposed projections of structures located at different z-levels within the cryosection thickness, they are still unable to fully resolve the complete 3D tubular architecture (Figure S3). For an improved analysis, we performed electron tomography (ETM) of selected ROI areas. The same RBs were first analyzed by FLM 3D reconstruction. Subsequently, any given section of the series was further analyzed by 3D TEM modeling at nanometer resolution (Figure 4J–L and Videos S4–S9). As predicted (Figure S3), the correlation of the 3D FLM with 3D ETM models (3D CLEM) revealed finer details of tubular connections (Figure 4L). Of note, the HDO-CLEM analysis was also successfully applied to cells expressing green fluorescent protein (GFP)-tagged proteins (Figure S5). An approximate timetable of all possible steps of HDO-CLEM is shown in Table S2.
Correlation analysis: HDO-CLEM morphometry
Morphometry analysis of biological samples can lead to quantitative geometrical information (e.g. surface and volume) and/or antigen density. While FLM provides high-throughput morphometry, with poor accuracy, TEM provides accurate morphometry with low efficiency. FLM and TEM morphometry analysis on the same structures can lead to a potential systematic error between the two approaches. This systematic error could potentially be used to optimize and/or calibrate FLM morphometry analysis, setting the bases for a high-throughput, high-resolution, hybrid FLM/TEM morphometry method.
To validate such an approach, we selected the three different morphologies of RBs, that is RRBs, d-SRBs and n-SRBs, and performed CLSM imaging (Figure 5A–C). Each RB structure was imaged by EM (Figure 5D–F), and a manual membrane contour was obtained for each image (insets in Figure 5D–F). An automatic segmentation plug-in, developed in our laboratory (weight adaptive threshold MicroSCoBiOJ,WAT-MicroSCoBiOJ), embedded in the MicroSCoBiOJ package (see Segmentation in File S1), was applied to CLSM images (insets in Figure 5A–C) of the same RBs. Figure 6 shows the fidelity of the CLSM surface measurements compared with the ‘real’ TEM measurements. The low mean relative error statistics (RRB 32 ± 6 SEM, d-SRB 37 ± 5 SEM and n-SRB 79 ± 4 SEM; RRB n= 8, d-SRB n= 5 and n-SRB n= 7) indicated the systematic nature of the relative error distribution. These data demonstrated that the major source of the systematic error was because of blurring (related to the band-limited nature of the CLSM) and, as expected, that it decreased for comparatively bigger structures. We next convoluted the three different models of RB, with a theoretical CLSM point spread function (see Image process formation in File S1), to simulate the image process formation (Figure 5G–I). Comparison between the real and the simulated images provided validation to our approach (Figure 6 and Figure S6). CLSM and CLSM simulation relative errors showed a similar behavior for the subresolved d-SRB and n-SRB structures, indicating that they are mostly because of the lower spatial resolution and the blurring introduced by the CLSM imaging. On the contrary, for the non subresolved RRBs, the discrepancy observed between CLSM and CLSM simulation errors indicated that the segmentation algorithm, used for the automatic segmentation of CLSM images to measure the surface of the RB structures, is the major source of the systematic error.
Knowledge of the systematic error allowed defining the fidelity of the FLM morphometry. Therefore, morphometry analysis can be performed on a large scale of samples by CLSM and analyzed with the accuracy of TEM analysis by correcting for the calculated systematic error.
Discussion
CLEM has proven to be a powerful approach to fill the gap between FLM and TEM imaging, providing greater insight into biological structures and their dynamic interconnections and a precise spatial information of protein localization. Several promising CLEM methods have been developed in the recent years; however, they still exhibit a number of problems and are not used on a routine basis by standard equipped imaging laboratories. Current CLEM, applied on tissue culture cells imaging, use plastic-embedded samples and either are performed on single cells (3), penalizing the possibility to obtain statistically relevant data, or are not suitable for multiple labeling of nonrecombinantly tagged antigens (5,7,12) or 3D analysis.
We have developed a robust CLEM ‘package’ (HDO-CLEM) matching the following requirements: (i) high data output rate, (ii) multiple labeling of both endogenous and recombinantly expressed antigens, (iii) 3D analysis of the structures of interest and (iv) affordability by many TEM laboratories. HDO-CLEM package includes the specific protocols for sample preparation/labeling and a complete software toolbox (MicroSCoBiOJ) for image analysis and 3D reconstruction.
Although several very promising hybrid FLM/TEM labels, able to penetrate an intact cell, have been developed in the recent years [e.g. fluorescence photo-oxidation of diaminobenzidine (5,12) and quantum dots (4,6)], we have pointed to the physical sectioning of the sample, as a solution to antigen accessibility. The use of ultrathin cryosections, pioneered by Tokuyasu (19) and optimized mostly thanks to the founding work of Slot and Geuze (20), is an established method of choice for TEM immunolabeling, with respect to membrane visibility, preservation and displaying of ultrastructural details, maintenance of immunoreactivity and antibody penetration. In particular, excellent immunogold procedures to label cryosections have been developed (20). Ultrathin cryosections serve also as an excellent substrate for high-resolution immunofluorescence microscopy, providing several advantages, that is sharper images, higher z-axis resolution and lower fluorescence background (8,10,21). Finally, cryosections do not suffer from the sample shrinkage because of the dehydration procedures used for plastic embedding and are suitable for 3D analysis both at the FLM and at the TEM level. Consequently, we could develop MicroSCoBiOJ software and set conditions for a rapid HDO-CLEM 3D analysis.
The are at least three relevant aspects about the use of HDO-CLEM that make this method novel and possibly able to overcome some limits of other CLEM approaches.
The first is the combination of (i) the well-known suitability of ultrathin cryosections for immunogold and immunofluorescence labeling to obtain images of multiple endogenous and recombinant antigens, identified by both labeling methods, on the same section and (ii) the original use of automated composition of large field maps of cells, both at the light and at the electron microscope level to allow for an extensive subsequent correlation of the images. The method has been applied successfully using both standard fluorochromes and/or GFP-tagged proteins for immunofluorescence, showing a certain versatility of use, and solving some of the problems encountered by the use of other hybrid-labeling agents for CLEM.
In particular, quantum dots fluorescence is quenched by heavy metal staining so that CLEM has to be performed on samples with suboptimal ultrastructural preservation because of the omission of glutaraldehyde fixation and osmium postfixation (4). Fluorescence photo-oxidation of diaminobenzidine, instead, allows the labeling of recombinantly expressed proteins but not of endogenous antigens (5,12). Finally, fluoronanogolds, an alternative FLM/TEM labeling method for cryosections (19,27), rely on silver enhancement for TEM observation. The efficiency of silver enhancement is dramatically reduced by a prolonged illumination (28), making the use of fluoronanogolds difficult for CLEM analysis that requires extensive FLM observation time.
The second is to provide new tools to perform 3D analysis and correlation. Automated ETM, on both plastic sections and cryosections (29–32), allows us to obtain 3D images of cells with a resolution of about 3–8 nm, providing valuable information about the structural basis of many cellular processes. Tomograms are generated from the tilt series using a variety of software, for example IMOD(33) or iTEM (Olympus SIS). However, the examination of large complex structures requires surface rendering or volume rendering modeling and rotational tools to facilitate interpretation and measurements (30). This modeling requires a manual segmentation approach to define the limits of the selected structures, especially for cryosections (30). By this approach, the user traces the limits of the feature of interest with a mouse or a pen directly on the screen, repeating this operation for the several tens to hundreds of slices present in a tomography volume. Although informative, this procedure is time consuming and may lead to distortions because of manual drawing. In addition, for the most common electron microscope voltage set (i.e. 120–300 kV), tomography is limited to a section thickness of 200–400 nm.
However, in several instances, the analysis of a complex structure may require a better resolution compared with CLSM but not as extreme as in EM analysis. For instance, the 3D SRB convoluted structure could not be discriminated by a higher data output method such as CLSM (hundreds to thousands of events analyzed per single study section of a few hours) and required a very low data output and time-consuming ETM analysis (about a week of full-time work to generate a single model). By using 200 and 60 nm thick sections, there is a reduction in the out-of-focus background with respect to our discrimination abilities in the x-yplane so that the convoluted structure of the SRBs can be resolved already at the light microscopy level (compare Figure 1B with S4E). The use of thin sections allows a more precise quantitative evaluation of the amount of labeled macromolecules by means of fluorescence CLSM to be correlated with EM data. In addition ‘Turboreg MicroSCoBiOJ’, in combination with WAT MicroSCoBiOJ, helps speeding up the process of serial section alignment and modeling; a model of a complex structure such as a SRB can be obtained in a few hours. We have modeled RBs by the alignment of up to eight serial sections; therefore, a total thickness of about 1600 nm has been analyzed. In addition, ETM analysis from any of the serial sections composing a specific object can still be performed if higher resolution analysis is necessary. In this latter instance, the use of the fluorescence map allowed us to easily retrace the desired object in the serial sections.
The third is to allow a CLEM hybrid morphometry analysis, based on integrated multimodal data set that combines the high data analysis capability of FLM, with the high precision–accuracy of TEM. Morphometry at the TEM can be particularly time consuming, especially on cryosections because of the low image contrast that hampers the use of reliable segmentation software to automatically identify the compartment borders. A solution to this bias is offered by the finding that systematic errors between FLM and TEM can be identified for different types of sample (i.e. RRBs, n-SRBs and d-SRBs in our model) when comparing surface estimates obtained by CLEM analysis on the same structures by CLSM and TEM. Therefore, once a systematic error is identified by CLEM on a limited number of structures, it can be subsequently applied to the large-scale measurements obtained by CLSM to correct these values to the accuracy of the EM. Potentially, this method could also be applied to compare the labeling density of cell compartments by relating the fluorescence density obtained by CLSM analysis with the surface values obtained by HDO-CLEM morphometry.
The choice of SRB and RRB, as a complex and a simple model, respectively, was instrumental to the development of the HDO-CLEM method. Several aspects of the structure of these organelles were already previously defined by an EM analysis (17). HDO-CLEM did not add on to this knowledge for RRBs, and the added value was limited to the possibility to perform morphometry on a large number of organelles by the use of CSLM corrected for the systematic error. However, the understanding of the SRB morphology was improved by HDO-CLEM that helped define the 3D structure of a large number of organelles, at a resolution intermediate between the current CLSM analysis and the EM tomography, and set conditions for further studies. In particular, we are currently using this approach to study the potential connections between the RER and the SRB, as a way to unravel potential RER/ERGIC interactions (Valetti C, Tacchetti C. Centro di Ricerca MicroSCoBiO, Dipartimento di Medicina Sperimentale, Università di Genova, 16132, Genoa, Italy, personal communication). This type of studies must rely on a statistically significant large body of data provided by CLSM, by EM resolution to clearly identify the connections and by a fast 3D analysis to visualize connections that would be lost in a 2D analysis; HDO-CLEM fulfills these requirements.
Materials and Methods
Cell culture
Tet-Off HeLa cells were transfected with pTRE-CH1 (50 μg) and pTK-Hyg (5 μg; Clontech Laboratories), as described (17). Clones were grown in the presence of 3 μg/mL tetracycline, 400 μg/mL hygromycin and 100 μg/mL geneticin. Hygromycin-resistant clones were screened by immunofluorescence 72 h after tetracycline removal.
CLEM specimen preparation
Cells fixed in 2% paraformaldehyde (Sigma-Aldrich) and 0.2% glutaraldehyde (PolyScience, Inc.) in PBS, pH 7.4, for 2 h were gently scraped-off the dish, embedded in 12% gelatin in PBS, cut into squared pieces and infiltrated with 2.3 m sucrose (Sigma-Aldrich) overnight at 4°C, mounted on pins, frozen in liquid nitrogen and sectioned at −100°C (200 nm cryosections) or −120°C (60 nm cryosections) with a Leica EM FCS Ultracut Microtome (Leica Microsystems). Flat ribbons of 60 or 200 nm thick cryosections, collected with 1.15 m sucrose and 1% 25 centipoise (cP) methylcellulose (Sigma-Aldrich), were transferred on formvar-coated 100-μm mesh, hexagonal gold finder grids (Agar Assing). For immunolabeling, samples were incubated on 2% gelatin in PBS for 20 min at 37°C, incubated with either rabbit anti-μ chain (Zymed) or rabbit anti-calreticulin (Stressgen) antibodies in 1% BSA in PBS for 30 min, rinsed in 1% BSA in PBS and incubated with donkey anti-rabbit Cy3-conjugated antibodies (Zymed) for 45 min in the presence of 4′,6-diamidino-2-phenylindole (DAPI) (Sigma) and PAG (10 nm) for 20 min (Department of Cell Biology, University Medical Center, Utrecht). For double labeling, after calreticulin labeling, samples were fixed for 5 min with 1% glutaraldehyde in PBS, washed in PBS and incubated with rabbit antibodies anti-μ chain (Zymed) for 30 min, rinsed in 1% BSA in PBS, incubated 45 min with donkey anti-rabbit Cy2-conjugated antibody (Zymed) in the presence of DAPI (Sigma), followed by PAG (15 nm) for 20 min, fixed for 2 min with 1% glutaraldehyde in PBS and washed in water. Unless otherwise indicated, all steps were at room temperature. For FLM observation, grids layered with a 200 nm coat of 2% methylcellulose were mounted between microscope slide and coverslip with 50% glycerol in water. For TEM observation, grids were unmounted, washed in water, incubated for 5 (200 nm cryosections) or 10 min (60 nm cryosections) in 2% uranyl acetate and 0.15 m oxalic acid and 5 min in 0.4% uranyl acetate and 1.8% 25ctp methylcellulose to increase contrast.
Microscopy
CLSM fluorescence images were collected with an Olympus FV1000 spectral CLSM, equipped with a multiline Ar laser (100 mW, 457–476–488–514 nm), an HeNe(G) laser (1 mW, 543 nm) and an ultraviolet laser diode (405 nm). Images were acquired with a PlanApo ×60/1.42 oil objective (Olympus Europa GMBH). DAPI was excited at the absorption tails using the 405 nm laser diode, and fluorescence was collected in the 425–475 nm spectral region. Cy2 was excited at 488 nm, and fluorescence was collected in the 500–540 nm spectral region. Cy3 was excited at 543 nm, and fluorescence was collected in the 555–655 nm spectral region. Transmission imaging was obtained using a 543 nm laser line source. WFM fluorescence images were collected with an Olympus IX70 equipped with a 100 Watt mercury arc lamp using a low-magnification UPlanFL ×20/0.50 objective (Olympus Europa GMBH) or a high-magnification PlanApo ×100/1.40 oil objective (Olympus Europa GMBH). DAPI Ex 360–385 Em BA420, FITC Ex 470–490 Em BA515 Ex and TRITC Ex 510–550 Em BA590 filter sets were used to collect DAPI, CY2 and CY3 signal, respectively. Images were recorded with a digital C4742-95 Hamamatsu charge-coupled device (CCD) camera. Occasionally (e.g. GFP imaging), the Nikon C1si CLSM was used. TEM and ETM were performed on a Tecnai 12-G2 EM (FEI Company), equipped with Lab6 filament and compustage. Planar images and Panorama Image Function maps were recorded with a SIS MegaView III CCD camera (Olympus-SIS) using the iTEM (Olympus-SIS) software. Photomicrographs were obtained on Kodak Film 4489.
Image processing and visualization
MicroSCoBiOJ was designed as a software toolbox collecting open source plug-ins for ImageJ(26). Some of the utilized plug-ins were optimized starting from exiting ones, and others were developed as original plug-ins. MosaicJ(23) assembles a mosaic of overlapping individual images and was used unmodified for high-resolution montages. Turboreg MicroSCoBiOJ, registers fluorescent physical section stacks, was modified from Turboreg to perform registration of serial multichannel images. WAT MicroSCoBiOJ, used for automatic surface area estimation on CLSM RB images, was fully developed in our laboratory using a segmentation algorithm based on a combined statically and dynamic threshold (see Segmentation in File S1). Basic functions of ImageJ were used for manual surface area estimation on TEM RB images. To obtain overlays of CLEM images, the pixel size of a FLM image was rescaled to that of the EM image and registered using the ‘rigid body rotation’ function of Turboreg (25) to avoid rotational differences. Merge of aligned/registered multichannel FLM image and grayscale TEM image was performed using Image5D plug-in. 3D fluorescence volume rendering was obtained combining two plug-ins developed in our laboratory: ‘Mesh Maker MicroSCoBiOJ’ implemented the Marching Cube (34) algorithm to extract isosurfaces from the 3D fluorescence volume data (see Surface rendering in File S1) and ‘Mesh Viewer MicroSCoBiOJ’ used the Java3D library (35) to render the set of the extracted isosurface. Tomograms of the tilt series were generated using the IMOD program package (33) or iTEM (Olympus SIS) software. PAG was used as fiducial marker.
Acknowledgments
We are deeply grateful to M. Wibbelt, P. Buescher and J. Kriegger (Olympus-SIS) for useful discussions and the use of the MegaView III camera and iTEM software, R. Parton (University of Queensland, Brisbane, Australia) and A. Sartori (Pasteur Institute, Paris, France) for critically reviewing the manuscript, P. Bianchini (MicroScoBiO, Genova, Italy) and S. Bruno (DIMES, University of Genova, Italy) for helpful discussions and G. Scita (IFOM, Milan Italy) for providing Rab5–GFP-transfected HeLa. This study was supported by Italian Cancer Research Foundation to C. T., Italian Minister of Research and University (FIRB project Nanomed to P. P. D. F. and PRIN to C. T. and A. D.) and SanPaolo Foundation to C. T.
Supporting Information
Additional Supporting Information may be found in the online version of this article:
Please note: Wiley-Blackwell are not responsible for the content or functionality of any supporting materials supplied by the authors. Any queries (other than missing material) should be directed to the corresponding author for the article.
References
- 1.Hell SW. Far-field optical nanoscopy. Science. 2007;316:1153–1158. doi: 10.1126/science.1137395. [DOI] [PubMed] [Google Scholar]
- 2.Diaspro A. Confocal and Two-Photon Microscopy: Foundations, Applications and Advances. New York: Wiley-Liss; 2001. [Google Scholar]
- 3.Mironov AA, Polishchuk RS, Luini A. Visualizing membrane traffic in vivo by combined video fluorescence and 3D electron microscopy. Trends Cell Biol. 2000;10:349–353. doi: 10.1016/s0962-8924(00)01787-6. [DOI] [PubMed] [Google Scholar]
- 4.Giepmans BNG, Deerinck TJ, Smarr BL, Jones YZ, Ellisman MH. Correlated light and electron microscopic imaging of multiple endogenous proteins using Quantum dots. Nat Methods. 2005;2:743–749. doi: 10.1038/nmeth791. [DOI] [PubMed] [Google Scholar]
- 5.Grabenbauer M, Geerts WJC, Fernadez-Rodriguez J, Hoenger A, Koster AJ, Nilsson T. Correlative microscopy and electron tomography of GFP through photooxidation. Nat Methods. 2005;2:857–862. doi: 10.1038/nmeth806. [DOI] [PubMed] [Google Scholar]
- 6.Nisman R, Dellaire G, Ren Y, Li R, Bazett-Jones DP. Application of quantum dots as probes for correlative fluorescence, conventional, and energy-filtered transmission electron microscopy. J Histochem Cytochem. 2004;52:13–18. doi: 10.1177/002215540405200102. [DOI] [PubMed] [Google Scholar]
- 7.Ren Y, Kruhlak MJ, Bazett-Jones DP. Same serial section correlative light and energy-filtered transmission electron microscopy. J Histochem Cytochem. 2003;51:605–612. doi: 10.1177/002215540305100506. [DOI] [PubMed] [Google Scholar]
- 8.Robinson JM, Takizawa T, Pombo A, Cook PR. Correlative fluorescence and electron microscopy on ultrathin cryosections: bridging the resolution gap. J Histochem Cytochem. 2001;49:803–808. doi: 10.1177/002215540104900701. [DOI] [PubMed] [Google Scholar]
- 9.Takizawa T, Robinson JM. Correlative microscopy of ultrathin cryosections is a powerful tool for placental research. Placenta. 2003;24:557–565. doi: 10.1053/plac.2002.0942. [DOI] [PubMed] [Google Scholar]
- 10.Mori M, Ishikawa G, Takeshita T, Goto T, Robinson JM, Takizawa T. Ultrahigh-resolution immunofluorescence microscopy using ultrathin cryosections: subcellular distribution of caveolin-1alpha and CD31 in human placental endothelial cells. J Electron Microsc (Tokyo) 2006;55:107–112. doi: 10.1093/jmicro/dfl011. [DOI] [PubMed] [Google Scholar]
- 11.Rieder CL, Cassels G. Correlative light and electron microscopy of mitotic cells in monolayer cultures. Methods Cell Biol. 1999;61:297–315. doi: 10.1016/s0091-679x(08)61987-1. [DOI] [PubMed] [Google Scholar]
- 12.Gaietta G, Deerinck TJ, Adams SR, Bouwer J, Tour O, Laird DW, Sosinsky GE, Tsien RY, Ellisman MH. Multicolor and electron microscopic imaging of connexin trafficking. Science. 2002;296:503–507. doi: 10.1126/science.1068793. [DOI] [PubMed] [Google Scholar]
- 13.Roth J, Bendayan M, Orci L. FITC-protein A-gold complex for light and electron microscopic immunocytochemistry. J Histochem Cytochem. 1980;28:55–57. doi: 10.1177/28.1.6153194. [DOI] [PubMed] [Google Scholar]
- 14.Oorschot V, de Wit H, Annaert WG, Klumperman J. A novel flat-embedding method to prepare ultrathin cryosections from cultured cells in their in situ orientation. J Histochem Cytochem. 2002;50:1067–1080. doi: 10.1177/002215540205000809. [DOI] [PubMed] [Google Scholar]
- 15.Sims P, Albrecht R, Pawley JB, Centonze V, Deerinck T, Hardin H. When light microscopy resolution is not enough: correlational light microscopy and electron microscopy. In: Pawley JB, editor. Handbook of Biological Confocal Microscopy. 3rd edn. New York: Springer; 2006. pp. 846–860. [Google Scholar]
- 16.Schweizer A, Clausen H, van Meer G, Hauri HP. Localization of O-glycan initiation, sphingomyelin synthesis, and glucosylceramide synthesis in Vero cells with respect to the endoplasmic reticulum-Golgi intermediate compartment. J Biol Chem. 1994;269:4035–4041. [PubMed] [Google Scholar]
- 17.Mattioli L, Anelli T, Fagioli C, Tacchetti C, Sitia R, Valetti C. ER storage diseases: a role for ERGIC-53 in controlling the formation and shape of russell bodies. J Cell Sci. 2006;119:2532–2541. doi: 10.1242/jcs.02977. [DOI] [PubMed] [Google Scholar]
- 18.Valetti C, Grossi CE, Milstein C, Sitia R. Russell bodies: a general response of secretory cells to synthesis of a mutant immunoglobulin which can neither exit from, nor be degraded in, the endoplasmic reticulum. J Cell Biol. 1991;115:983–994. doi: 10.1083/jcb.115.4.983. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Tokuyasu KT. Immunochemistry on ultrathin frozen sections. Histochem J. 1980;12:381–403. doi: 10.1007/BF01011956. [DOI] [PubMed] [Google Scholar]
- 20.Slot JW, Geuze HJ. Cryosectioning and immunolabeling. Nat Protoc. 2007;2:2480–2491. doi: 10.1038/nprot.2007.365. [DOI] [PubMed] [Google Scholar]
- 21.Za L, Albertinazzi C, Paris S, Gagliani M, Tacchetti C, de Curtis I. {beta}PIX controls cell motility and neurite extension by regulating the distribution of GIT1. J Cell Sci. 2006;119:2654–2666. doi: 10.1242/jcs.02996. [DOI] [PubMed] [Google Scholar]
- 22.Ireland BS, Niggemann M, Williams DB. In vitro assays of the functions of calnexin and calreticulin, lectin chaperones of the endoplasmic reticulum. Methods Mol Biol. 2006;347:331–342. doi: 10.1385/1-59745-167-3:331. [DOI] [PubMed] [Google Scholar]
- 23.Thevenaz P, Unser M. User-friendly semiautomated assembly of accurate image mosaics in microscopy. Microsc Res Tech. 2007;70:135–146. doi: 10.1002/jemt.20393. [DOI] [PubMed] [Google Scholar]
- 24.Conchello JA, Lichtman JW. Optical sectioning microscopy. Nat Methods. 2005;2:920–931. doi: 10.1038/nmeth815. [DOI] [PubMed] [Google Scholar]
- 25.Thevenaz P, Ruttimann UE, Unser M. A pyramid approach to subpixel registration based on intensity. IEEE Transactions on Image Processing. 1998;7:27–41. doi: 10.1109/83.650848. [DOI] [PubMed] [Google Scholar]
- 26.Abramoff MD, Magelhaes PJ, Ram SJ. Image Processing with Image. J. Biophotonics International. 2004;11:36–42.
- 27.Takizawa T, Robinson JM. FluoroNanogold is a bifunctional immunoprobe for correlative fluorescence and electron microscopy. J Histochem Cytochem. 2000;48:481–485. doi: 10.1177/002215540004800405. [DOI] [PubMed] [Google Scholar]
- 28.Baschong W, Stierhof YD. Preparation, use, and enlargement of ultrasmall gold particles in immunoelectron microscopy. Microsc Res Tech. 1998;42:66–79. doi: 10.1002/(SICI)1097-0029(19980701)42:1<66::AID-JEMT8>3.0.CO;2-P. [DOI] [PubMed] [Google Scholar]
- 29.Subramaniam S. Bridging the imaging gap: visualizing subcellular architecture with electron tomography. Curr Opin Microbiol. 2005;8:316–322. doi: 10.1016/j.mib.2005.04.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Frey TG, Perkins GA, Ellisman MH. Electron tomography of membrane-bound cellular organelles. Annu Rev Biophys Biomol Struct. 2006;35:199–224. doi: 10.1146/annurev.biophys.35.040405.102039. [DOI] [PubMed] [Google Scholar]
- 31.Ladinsky MS, Howell KE. Electron tomography of immunolabeled cryosections. Methods Cell Biol. 2007;79:543–558. doi: 10.1016/S0091-679X(06)79021-5. [DOI] [PubMed] [Google Scholar]
- 32.McIntosh R, Nicastro D, Mastronarde D. New views of cells in 3D: an introduction to electron tomography. Trends Cell Biol. 2005;15:43–51. doi: 10.1016/j.tcb.2004.11.009. [DOI] [PubMed] [Google Scholar]
- 33.Kremer JR, Mastronarde DN, McIntosh JR. Computer visualization of three-dimensional image data using IMOD. J Struct Biol. 1996;116:71–76. doi: 10.1006/jsbi.1996.0013. [DOI] [PubMed] [Google Scholar]
- 34.Lorensen WE, Cline HE. Marching cubes: A high resolution 3D surface construction algorithm. Proceedings of the 14th annual conference on computer graphics and interactive techniques 1987. Vol. 21. ACM Press; New York, USA; 1987. pp. 163–169. [Google Scholar]
- 35.Vicidomini G, Mondal PP, Diaspro A. Fuzzy logic and maximum a posteriori-based image restoration for confocal microscopy. Opt Lett. 2006;31:3582–3584. doi: 10.1364/ol.31.003582. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.