Controlled comparison of the BACTEC high-blood-volume fungal medium, BACTEC Plus 26 aerobic blood culture bottle, and 10-milliliter isolator blood culture system for detection of fungemia and bacteremia.

M L Wilson 1, T E Davis 1, S Mirrett 1, J Reynolds 1, D Fuller 1, S D Allen 1, K K Flint 1, F Koontz 1, L B Reller 1
PMCID: PMC263579  PMID: 8463399

Abstract

The BACTEC high-blood-volume fungal medium (HBV-FM) (Becton Dickinson Diagnostic Instrument Systems, Sparks, Md.) was compared with the Isolator (IS) tube and the BACTEC Plus 26 (BP26) blood culture bottle for the ability to recover fungi from the blood of adult patients suspected of having fungemia. A total of 6,836 blood culture sets that fulfilled criteria for inclusion in the study were received. Three separate comparisons were performed: 4,907 HBV-FM versus IS, 4,886 BP26 versus HBV-FM, and 4,949 BP26 versus IS. For the HBV-FM versus IS comparison, 218 isolates were recovered: 125 (57.3%) were bacteria and 93 (42.7%) were fungi. HBV-FM was comparable to IS for recovery of yeasts, but IS was superior for recovery of Histoplasma capsulatum (25 versus 0 isolates recovered [P < 0.001]). Growth of Torulopsis glabrata was detected earlier (P < 0.05) in HBV-FM bottles. For the BP26 versus HBV-FM comparison, 229 isolates were recovered: 161 (70.3%) were bacteria, and 68 (29.7%) were fungi. HBV-FM was superior for recovery of T. glabrata (P < 0.025) and all fungi combined (P < 0.025). There were no statistically significant differences in the speed of detection of microbial growth. For the BP26 versus IS comparison, 251 isolates were recovered: 165 (65.7%) were bacteria, and 86 (34.2%) were fungi. IS was superior for recovery of H. capsulatum (P < 0.001), T. glabrata (P < 0.05), and fungi other than H. capsulatum (P < 0.025). BP26 was superior for recovery of all bacteria combined (P < 0.001) and viridans group streptococci (P < 0.01). Growth of T. glabrata (P < 0.05) was detected earlier in IS tubes. Growth of Staphylococcus aureus (P < 0.01), viridans group streptococci (P < 0.01), Pseudomonas aeruginosa (P < 0.05), and all microorganisms combined (P < 0.05) was detected earlier in BP26 bottles. For yeast, 57 of 59 (96.6%), 79 of 80 (98.7%), and 64 of 67(95.5%) were recovered from BP26 bottles, HBV-FM bottles, and IS tubes, respectively, by day 14; for H. capsulatum, 14 of 36 (38%) isolates were recovered from IS tubes by day 14. Mean times of recovery were similar for BACTEC bottles and IS. We conclude that (i) for recovery of fungi from blood cultures, HBV-FM is equivalent to IS (with the exception of H. capsulatum); (ii) for recovery of bacteria, BP26 is superior to IS; (iii) BP26 bottles are inferior to both HBV-FM bottles and IS tubes for recovery of T. glabrata; and (iv) HBV-FM bottles must be paired with another blood culture bottle or system to optimize detection of bacteremia.

Full text

PDF
865

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Bille J., Edson R. S., Roberts G. D. Clinical evaluation of the lysis-centrifugation blood culture system for the detection of fungemia and comparison with a conventional biphasic broth blood culture system. J Clin Microbiol. 1984 Feb;19(2):126–128. doi: 10.1128/jcm.19.2.126-128.1984. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Bille J., Stockman L., Roberts G. D., Horstmeier C. D., Ilstrup D. M. Evaluation of a lysis-centrifugation system for recovery of yeasts and filamentous fungi from blood. J Clin Microbiol. 1983 Sep;18(3):469–471. doi: 10.1128/jcm.18.3.469-471.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Body B. A., Pfaller M. A., Durrer J., Koontz F., Gröschel D. H. Comparison of the lysis centrifugation and radiometric blood culture systems for recovery of yeast. Eur J Clin Microbiol Infect Dis. 1988 Jun;7(3):417–420. doi: 10.1007/BF01962353. [DOI] [PubMed] [Google Scholar]
  4. Brannon P., Kiehn T. E. Clinical comparison of lysis-centrifugation and radiometric resin systems for blood culture. J Clin Microbiol. 1986 Nov;24(5):886–887. doi: 10.1128/jcm.24.5.886-887.1986. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Brannon P., Kiehn T. E. Large-scale clinical comparison of the lysis-centrifugation and radiometric systems for blood culture. J Clin Microbiol. 1985 Dec;22(6):951–954. doi: 10.1128/jcm.22.6.951-954.1985. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Gill V. J., Zierdt C. H., Wu T. C., Stock F., Pizzo P. A., MacLowry J. D. Comparison of lysis-centrifugation with lysis-filtration and a conventional unvented bottle for blood cultures. J Clin Microbiol. 1984 Nov;20(5):927–932. doi: 10.1128/jcm.20.5.927-932.1984. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Gold J. W. Opportunistic fungal infections in patients with neoplastic disease. Am J Med. 1984 Mar;76(3):458–463. doi: 10.1016/0002-9343(84)90665-x. [DOI] [PubMed] [Google Scholar]
  8. Guerra-Romero L., Edson R. S., Cockerill F. R., 3rd, Horstmeier C. D., Roberts G. D. Comparison of Du Pont Isolator and Roche Septi-Chek for detection of fungemia. J Clin Microbiol. 1987 Sep;25(9):1623–1625. doi: 10.1128/jcm.25.9.1623-1625.1987. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Hamilton D. J., Amos D., Schwartz R. W., Dent C. M., Counts G. W. Effect of delay in processing on lysis-centrifugation blood culture results from marrow transplant patients. J Clin Microbiol. 1989 Jul;27(7):1588–1593. doi: 10.1128/jcm.27.7.1588-1593.1989. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Henry N. K., McLimans C. A., Wright A. J., Thompson R. L., Wilson W. R., Washington J. A., 2nd Microbiological and clinical evaluation of the isolator lysis-centrifugation blood culture tube. J Clin Microbiol. 1983 May;17(5):864–869. doi: 10.1128/jcm.17.5.864-869.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Horn R., Wong B., Kiehn T. E., Armstrong D. Fungemia in a cancer hospital: changing frequency, earlier onset, and results of therapy. Rev Infect Dis. 1985 Sep-Oct;7(5):646–655. doi: 10.1093/clinids/7.5.646. [DOI] [PubMed] [Google Scholar]
  12. Kellogg J. A., Manzella J. P., McConville J. H. Clinical laboratory comparison of the 10-ml isolator blood culture system with BACTEC radiometric blood culture media. J Clin Microbiol. 1984 Oct;20(4):618–623. doi: 10.1128/jcm.20.4.618-623.1984. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Kelly M. T., Roberts F. J., Henry D., Geere I., Smith J. A. Clinical comparison of isolator and BACTEC 660 resin media for blood culture. J Clin Microbiol. 1990 Sep;28(9):1925–1927. doi: 10.1128/jcm.28.9.1925-1927.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]
  14. Kiehn T. E., Wong B., Edwards F. F., Armstrong D. Comparative recovery of bacteria and yeasts from lysis-centrifugation and a conventional blood culture system. J Clin Microbiol. 1983 Aug;18(2):300–304. doi: 10.1128/jcm.18.2.300-304.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Koontz F. P., Flint K. K., Reynolds J. K., Allen S. D. Multicenter comparison of the high volume (10 ml) NR BACTEC PLUS system and the standard (5 ml) NR BACTEC system. Diagn Microbiol Infect Dis. 1991 Mar-Apr;14(2):111–118. doi: 10.1016/0732-8893(91)90044-g. [DOI] [PubMed] [Google Scholar]
  16. Murray P. R. Comparison of the lysis-centrifugation and agitated biphasic blood culture systems for detection of fungemia. J Clin Microbiol. 1991 Jan;29(1):96–98. doi: 10.1128/jcm.29.1.96-98.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  17. Muñoz P., Bernaldo de Quirós J. C., Berenguer J., Rodríguez Créixems M., Picazo J. J., Bouza E. Impact of the BACTEC NR system in detecting Candida fungemia. J Clin Microbiol. 1990 Mar;28(3):639–641. doi: 10.1128/jcm.28.3.639-641.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Nielsen H., Stenderup J., Bruun B. Fungemia in a university hospital 1984-1988. Clinical and mycological characteristics. Scand J Infect Dis. 1991;23(3):275–282. doi: 10.3109/00365549109024310. [DOI] [PubMed] [Google Scholar]
  19. Paya C. V., Roberts G. D., Cockerill F. R., 3rd Laboratory methods for the diagnosis of disseminated histoplasmosis: clinical importance of the lysis-centrifugation blood culture technique. Mayo Clin Proc. 1987 Jun;62(6):480–485. doi: 10.1016/s0025-6196(12)65473-1. [DOI] [PubMed] [Google Scholar]
  20. Prevost-Smith E., Hutton N. Value of extended agitation and subculture of BACTEC NR 660 aerobic resin blood culture bottles for clinical yeast isolates. J Clin Microbiol. 1992 Dec;30(12):3239–3242. doi: 10.1128/jcm.30.12.3239-3242.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. Rantala A., Niinikoski J., Lehtonen O. P. Yeasts in blood cultures: impact of early therapy. Scand J Infect Dis. 1989;21(5):557–561. doi: 10.3109/00365548909037885. [DOI] [PubMed] [Google Scholar]
  22. Smith R. L., Meixler S. M., Simberkoff M. S. Excess mortality in critically ill patients with nosocomial bloodstream infections. Chest. 1991 Jul;100(1):164–167. doi: 10.1378/chest.100.1.164. [DOI] [PubMed] [Google Scholar]
  23. Stockman L., Roberts G. D., Ilstrup D. M. Effect of storage of the du Pont lysis-centrifugation system on recovery of bacteria and fungi in a prospective clinical trial. J Clin Microbiol. 1984 Feb;19(2):283–285. doi: 10.1128/jcm.19.2.283-285.1984. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Tarrand J. J., Guillot C., Wenglar M., Jackson J., Lajeunesse J. D., Rolston K. V. Clinical comparison of the resin-containing BACTEC 26 Plus and the Isolator 10 blood culturing systems. J Clin Microbiol. 1991 Oct;29(10):2245–2249. doi: 10.1128/jcm.29.10.2245-2249.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  25. Telenti A., Roberts G. D. Fungal blood cultures. Eur J Clin Microbiol Infect Dis. 1989 Sep;8(9):825–831. doi: 10.1007/BF02185855. [DOI] [PubMed] [Google Scholar]
  26. Weinstein M. P., Mirrett S., Wilson M. L., Harrell L. J., Stratton C. W., Reller L. B. Controlled evaluation of BACTEC Plus 26 and Roche Septi-Chek aerobic blood culture bottles. J Clin Microbiol. 1991 May;29(5):879–882. doi: 10.1128/jcm.29.5.879-882.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  27. Weinstein M. P., Reller L. B., Murphy J. R., Lichtenstein K. A. The clinical significance of positive blood cultures: a comprehensive analysis of 500 episodes of bacteremia and fungemia in adults. I. Laboratory and epidemiologic observations. Rev Infect Dis. 1983 Jan-Feb;5(1):35–53. doi: 10.1093/clinids/5.1.35. [DOI] [PubMed] [Google Scholar]
  28. Young R. C., Bennett J. E., Geelhoed G. W., Levine A. S. Fungemia with compromised host resistance. A study of 70 cases. Ann Intern Med. 1974 May;80(5):605–612. doi: 10.7326/0003-4819-80-5-605. [DOI] [PubMed] [Google Scholar]