Abstract
Significant improvements have made it possible to add the technology of time-of-flight (TOF) to improve PET imaging, particularly for oncology applications. The goals of this work were to investigate the benefits of TOF in experimental phantoms and to determine how these benefits translate into improved performance for patient imaging.
Methods
The Gemini TF scanner (Philips Medical Systems) used in this study is a fully 3D scanner and utilizes the scintillator LYSO with a system timing resolution of ~600 ps. The data are acquired in list-mode and reconstructed with an ML-EM algorithm; the system model includes the TOF kernel and corrections for attenuation, detector normalization, randoms, and scatter. The scatter correction is an extension of the modelbased single-scatter simulation to include the time domain. Phantom measurements to study the benefit of TOF include 27-cm and 35-cm diameter distributions with spheres ranging in size from 10 mm to 37 mm. To assess the benefit of TOF PET for clinical imaging, patient studies are quantitatively analyzed.
Results
The lesion phantom studies demonstrate the improved contrast of the smallest spheres with TOF compared to non-TOF and also confirm the faster convergence of contrast with TOF. These gains are evident from visual inspection of the images, as well as a quantitative evaluation of contrast recovery of the spheres and noise in the background. The gains with TOF are higher for larger objects. These results correlate with patient studies in which lesions are seen more clearly and with higher uptake at comparable noise for TOF than with non-TOF.
Conclusion
TOF leads to a better contrast vs. noise trade-off than non-TOF but one that is difficult to quantify in terms of a simple sensitivity gain improvement: a single gain factor for TOF improvement does not include the increased rate of convergence with TOF, nor does it consider that TOF may converge to a different contrast than non-TOF. The experimental phantoms results agree with those of prior simulations and help explain the improved image quality with TOF for patient oncology studies.
Keywords: Positron Emission Tomography, PET, Time-of-Flight PET, fully-3D Whole-Body Imaging
INTRODUCTION
There has been considerable advancement of the technology and instrumentation in positron emission tomography (PET) over the last 30 years since the first tomography ring systems were developed. Significant improvements have been made in detectors, hardware, and image processing that impact both image quality and accuracy of quantification. Some of the major achievements include i) the development and incorporation of new scintillators and detector configurations for higher spatial resolution and sensitivity, ii) the evolution from two-dimensional (2D) systems with septa to three-dimensional (3D) systems with larger axial fields-of-view for improved sensitivity, iii) the transition from analytic filtered back-projection reconstruction algorithms to fully 3D iterative techniques with data corrections included in the system model for improved image quality and quantification, and iv) the combination of a computed tomography (CT) scanner with the PET instrument for both attenuation correction and anatomical correlation. The data presented in this paper demonstrate that the technology of time-off-light (TOF) can be added upon these achievements to further improve PET.
The idea to use TOF information was originally proposed in the early stages of PET scanner development (1–3), and the first TOF PET systems were developed in the 1980s (4–8). A good summary of this early work is included in (9, 10). These early systems used cesium fluoride (CsF) or barium fluoride (BaF2) scintillators and were capable of meeting the high count-rate demands of research brain and heart studies with short-lived isotopes, but they could match neither the spatial resolution nor the sensitivity of conventional PET scanners with bismuth germanate (BGO) scintillators. By the early 1990s, these early TOF scanners were retired from use - just before whole-body oncology studies with [18F]-fluorodeoxyglucose (FDG) became common for the clinical diagnosis and staging of cancer. TOF PET is particularly advantageous for whole-body imaging, since the theoretical improvement with TOF is predicted to increase with the size of the patient. This is fortuitous, since conventional PET image quality degrades noticeably for large patients due to increased attenuation, which leads to both the loss of true counts and the increase of scattered counts. The promise of TOF for clinical PET is that it has the potential to improve the image quality in heavy patients, precisely where it is needed most.
There are a number of reasons why TOF is making a resurgence in PET today. First is the development of new scintillators that combine fast timing decay with high light output and high stopping power (11, 12). The advantage of scintillators such as lutetium oxyorthosilicate (LSO) and lutetium-yttrium oxyorthosilicate (LYSO) is that they have desirable properties for PET imaging without TOF, so their very good timing resolution should enhance already good performance. In contrast, CsF and BaF2 have very good timing resolution, but their low light output limits the detector design choices for light sharing and decoding crystals. The older TOF scanners with these scintillators generally had low sensitivity resulting from poor packing fraction and poor spatial resolution of about 10–15 mm, since the crystals were large in cross-section. The last of these first-generation TOF PET scanners, the SuperPETT 3000 scanner, used a block-encoding scheme with improved spatial resolution of about 8 mm (13). It is common today in conventional PET scanners to use smaller crystals with encoding schemes for crystal identification to achieve spatial resolutions of 4 to 6 mm. Although modern clinical scanners all have spatial resolutions in this range, the encoding ratio (number of crystals:number of photomultiplier tubes, PMTs) depends on the light output of the crystal, so BGO is typically configured with a lower encoding ratio than LSO or LYSO. For TOF PET a high encoding ratio is desirable because this reduces the number of PMTs in the system; this is of practical importance, since a TOF PET system cannot use the same PMTs as a conventional PET scanner, and fast PMTs tend to be more expensive due to the more complicated plano-concave photocathode design that achieves the shorter transit time spread and faster rise time needed for TOF measurements.
In addition to the availability of new scintillators, the improvements in performance and reliability of PMTs and electronics make TOF more practical today than in the past. Although fast PMTs were available 25 years ago, there are now more choices of cost-effective PMTs in the range of 1- to 2-inch diameter from the major PMT manufacturers. In addition, stability of electronics was reportedly very difficult to achieve with early TOF scanners. Current electronics have been observed to be very stable with little or no drift in the system timing resolution over periods of months (14).
Finally, there has been progress in image reconstruction algorithms. Early image reconstruction methods were developed that made use of the TOF information, including the most likely position (MLP) (15) and confidence-weighted (CW) back-projection (15–17) algorithms. Politte and Snyder (18, 19) then adapted the maximum likelihood expectation maximization (ML-EM) algorithm to include the TOF probability response function to reconstruct list-mode data and found that it yielded superior image quality (fewer artifacts and better contrast for the same reconstructed resolution) than the CW or MLP methods.
Although the early studies of TOF reconstruction considered only 2D data, it is natural today to combine list-mode acquisition and iterative reconstruction for 3D data, since list-mode acquisition preserves the intrinsic spatial and temporal information of the data. In addition, list-mode storage is often more efficient than binned format for 3D data because of the large number of measured lines of response with respect to the number of detected events. In the last decade there have been notable developments in list-mode reconstruction methods (20–22) and, more recently, inclusion of both TOF and other physical effects in the system model (23–25). Although list-mode reconstruction is a computationally intensive method, fast computers and parallelization make a practical implementation feasible today (26).
Given these recent advances in technology, there have been a number of experimental studies of TOF PET imaging and interest from both academic research groups as well as the commercial sector in the last couple of years. Results from a Siemens scanner based on LSO have been reported (27). However, since this system was not initially designed for TOF, limitations in the hardware result in a timing resolution of 1.2 ns, and the impact on clinical studies has been seen to be modest (28). Our group has been developing a research TOF PET scanner (29) based on a new scintillator, lanthanum bromide (LaBr3), that promises to have superior timing resolution, while Philips has developed a new scanner designed with TOF capability (Gemini TF) based on LYSO with an intrinsic timing resolution of ~600 ps. First results with the Gemini TF showed improved performance resulting from TOF measurement capability of the tomograph (14).
Our group has previously investigated the benefit of TOF for both lesion estimation and detectability tasks using simulated data (30). It was found that as the timing resolution improves, fewer iterations are required to achieve the same or higher maximum contrast recovery coefficient (CRC); for small lesions in larger objects in particular, higher values of Crc were obtained with TOF at the same noise level than without TOF. Lesion detectability, as judged by the non-prewhitening matched filter signal-to-noise ratio, was seen to increase nonlinearly as a function of both statistics and timing resolution. The goals of this work were to investigate the benefits of TOF in experimental phantoms and to determine how these benefits translate into improved patient image quality.
MATERIALS AND METHODS
Theory of TOF PET
To generate 3D images in PET, coincident lines-of-response (LORs) are normally detected and recorded at many angles, and tomographic images are generated through traditional filtered back-projection or iterative reconstruction. In a TOF PET system, for each annihilation event the difference in arrival times between the two coincident photons is also measured. Hypothetically, with perfect TOF information reconstruction would be unnecessary, as the location of each annihilation event could be identified based only on line pair and time difference information. However, even imperfect timing information helps to improve the reconstruction because TOF information serves to localize the coincidence, which reduces the propagation of noise along the LOR.
The distance that a given event is projected along the LOR is defined by an uncertainty Δx that can be much smaller (with good timing resolution) than the distance D (diameter of the patient) over which counts would be distributed equally in the forward- and back-projection steps of non-TOF reconstruction. A reduction of noise can be equated to an increase in sensitivity, and this effective sensitivity gain was estimated (15, 31) at the center of a uniform distribution to be proportional to D/Δx where Δx = c • Δt/2, c is the speed of light, and Δt is the timing resolution (full-width-at-half-maximum, fwhm). Table 1 lists D/Δx for several object sizes and timing resolutions. The diameters chosen represent thin (20 cm), average (27 cm), and heavy (35 cm) patients. In (14) we showed that phantom and patient data have the same relationship of NEC count-rate vs. size and that these phantom sizes can be used to represent patients ranging from 40 kg to 100 kg. A more involved derivation of TOF gain that accounts for filtering in the reconstruction process was given in (16), also for a uniform distribution, whereby the variance reduction (equal to sensitivity gain) is given by D/(1.6 • Δx). These simple estimates may not be accurate for the realistic, non-uniform activity distributions found in patients, but they are useful to provide an idea of the relative importance of timing resolution. In addition, these metrics argue that TOF gain increases not only as timing resolution improves but also as the object diameter increases, so we can predict the relative benefit of TOF as the patient size increases. However, it may be too simplistic to characterize the TOF gain as a single value, since we expect it to depend on both the task and methods of data correction and image reconstruction. The simple estimates of TOF gain, which were derived for analytical reconstruction and may not extend directly to iterative reconstruction, also do not consider the issue of faster convergence for an iterative reconstruction algorithm with TOF, an additional benefit of TOF that was shown in (30) to depend on both the timing resolution and statistics.
Table 1.
Calculated sensitivity gain for a uniform distribution (15, 31) as defined by the relation D/Δx where D is the object diameter and Δx = c • Δt/2 with c being the speed of light and Δt being the (fwhm) timing resolution.
Δt = 300 ps | Δt = 600 ps | Δt = 1000 ps | |
---|---|---|---|
D = 20 cm | 4.4 | 2.2 | 1.3 |
D = 27 cm | 6.0 | 3.0 | 1.8 |
D = 35 cm | 7.8 | 3.9 | 2.3 |
In this paper, we examine the benefits of TOF in the clinical oncology imaging task of estimation of the uptake value in a region-of-interest (ROI). TOF has the potential to improve the accuracy of a lesion uptake measurement through better signal localization and reduced noise propagation. While the accuracy of this task is of primary importance, an added requirement is to perform it with the shortest possible scan time that, in turn, impacts the list-mode reconstruction time.
TOF scanner
The Gemini TF scanner (Philips Medical Systems) was used in this study. This fully 3D scanner has good intrinsic performance, as described in (14). The energy resolution of 11.5% (fwhm) allows the lower energy threshold to be raised to 440 keV. The system timing resolution measured with a point source in air is 585 ps, and daily measurements of the timing resolution have demonstrated that this value is very stable over a period of many months. However, there is a count-rate effect on the timing resolution that causes the timing resolution to degrade to about 650–700 ps at singles count-rates of 15–25 Mcps, which correspond to typical clinical count-rates. A study on the impact of using an inaccurate TOF kernel in iterative TOF PET reconstruction found that using a TOF kernel that was narrower than the actual timing resolution led to degraded contrast (32). Therefore, an estimated value of the timing resolution based on the singles rate during a study, rather than the intrinsic timing resolution measured at low activities, is used during image reconstruction.
Data corrections and reconstruction
To correct the measured timing differences between crystals, timing calibration methods have been investigated, using either a rotating line source (33) or stationary, shielded point source at the center of the scanner (34). For the Gemini a 68Ge rod source is used to measure the timing offsets, following the method described in (34). The range (minimum to maximum) of offset factors for the entire scanner is +/− 1 ns.
For scatter correction, it has been suggested that model-based single scatter simulation (SSS) should be modified to estimate the scatter distribution as a function of radial and time domains (35). We have extended our implementation of SSS (36, 37) in a similar fashion, but with a TOF-dependent scatter distribution (36, 37). Our approach uses fine timing bins to allow accurate modeling of scatter estimates while maintaining the intrinsic timing resolution of the system without creating impractical demands on storage requirements or reconstruction time.
The reconstruction approach implemented for TOF is a list-mode version of the ML-EM algorithm (24) with a TOF kernel applied in both the forward- and back-projection operations. The Poisson statistics of the data are preserved by using the data in their original form (i.e., without binning) and incorporating the physical effects of PET imaging in the system model. The TOF response function is modeled as a one-dimensional Gaussian function along the LOR. The ML-EM algorithm is accelerated by dividing the data into chronologically ordered subsets (24).
A fully 3D list-mode iterative reconstruction algorithm is computationally intensive and has proven to be challenging to implement for clinical purposes (26). Using a 10-node (dual CPU) computer cluster, the image processing with this algorithm proceeds in parallel with data acquisition and is typically completed for clinical whole-body (multi-bed) studies 10–30 minutes after the end of the acquisition, depending on the number of counts collected.
Phantom measurements
For comparison with prior simulation studies (30), 27- and 35-cm diameter phantoms were constructed for experimental measurements. As described earlier, these diameters represent average and heavy patients, respectively 63kg and 100 kg according to the correlation of count-rate vs. weight presented in (14). We modified the caps of 5-gallon and 13-gallon ‘Carboy’ bottles (Cole Parmer) to accept the ring of six spheres (diameters of 37, 28, 22, 17, 13, and 10 mm) from the NEMA image quality phantom (38). The spheres were filled with FDG with a 6:1 uptake ratio compared to the background activity concentrations of 8.2 kBq/mL (0.22 µCi/mL) and 4.1 kBq/mL (0.11 µCi/mL) in the 27- and 35-cm diameter phantoms, respectively. Three replicate acquisitions were performed. The prompt collect rates were 470 kcps (27-cm) and 250 kcps (35-cm); the singles rates were 21 Mcps (27-cm) and 16 Mcps (35-cm). For these studies the corresponding scatter fractions were 36% (27-cm) and 38% (35-cm), while the randoms fractions (delays/prompts) were 54% (27-cm) and 51% (35-cm). A CT scan was used for attenuation correction, and the data were reconstructed for up to 20 iterations (20 subsets/iteration) with the list-mode ML-EM algorithm described previously.
The images were analyzed by drawing ROIs with diameters equal to the sphere diameters on the hot lesions in the central slice. The count density in these ROIs was used as an estimate of the uptake in each lesion. The background count density was obtained by drawing 12 ROIs at a radial distance of 11 cm within the central slice and in slices ±1 and ±2 cm away, for 60 background ROIs of each sphere size. The contrast recovery coefficient (CRC) was calculated as
(1) |
where H and B are the average counts in the sphere and background ROIs, respectively, and a is the true activity ratio (=6), determined using the method described in (38). Noise was calculated as the average pixel-to-pixel percent standard deviation (SD) in the 60 background ROIs, rather than the background variability measure prescribed by NEMA, because the latter measure is sensitive to non-uniformities in the image but relatively insensitive to statistical noise. The CRC and noise values were averaged over the three replicate acquisitions.
Patient oncologic imaging
The standard imaging protocol for clinical oncology studies with the Gemini TF at the University of Pennsylvania is to inject a 555-MBq (15-mCi) dose of FDG and scan for 3 min/bed position following a 60-min uptake period. Since the axial field-of-view of the scanner is 18 cm with an incremental bed movement between frames of 9 cm to allow for a 50% overlap between bed positions, a whole-body scan from the base of the skull to mid-thigh typically requires between 8 and 11 bed positions. The average prompt rate is 220–550 kcps, which leads to approximately 40–100 Mcts (prompts) for a single bed position in a typical study. The singles rate for clinical studies ranges between 15–25 Mcps. The range in the singles and prompt rates represent values for varying patient sizes, with the low values representative of larger patients. The prompt and singles rates for the patients agree well with the rates at which the 27- and 35-cm diameter phantoms were acquired. A CT scan was used for attenuation correction. The data were reconstructed for up to 10 iterations with 33 subsets, using an average TOF resolution of 660 ps, depending on the singles count-rate of the study. Data corrections were performed as described previously.
Five patient studies spanning the range of sizes (46–140 kg, BMI=19–46) were selected for quantitative analysis of lesion uptake. Six to nine lesions (1–2 cm diameter) were selected for analysis in each patient for a total of 37 lesions. Small circular ROIs (8-mm diameter) were drawn on the lesions and the average value in each ROI recorded. A 40-mm diameter ROI drawn in a region of the liver that was visually determined to be uniform was used as a background region to normalize out any bias. The ratio of average lesion counts to average background counts (L/B ratio) was calculated for each lesion. Noise was determined as the pixel-to-pixel percent SD in the liver ROI.
RESULTS
Phantom measurements
Figures 1A and 1B show images from the 35-cm phantom study for a 5-minute scan reconstructed without (A) and with (B) TOF information for different iterations of ML-EM. These images demonstrate visually the improved contrast of the smallest spheres with TOF compared to non-TOF; even after 20 iterations it is difficult to visualize the 10-mm sphere without TOF while the images continue to get noisier with more iterations. With TOF the 10-mm sphere is easily seen after only 1 iteration. The images also confirm the faster convergence of contrast with TOF found in the simulation studies.
Figure 1.
Images from the 35-cm diameter phantom measurement with two cold spheres (28, 37 mm) and four hot spheres (10, 13, 17, 22 mm) with 6:1 contrast. (A) non-TOF and (B) TOF images for 1, 2, 5, 10, and 20 iterations (left to right) for 5-min scan time. (C) non-TOF images after 10 iterations and (D) TOF images after 5 iterations for 5, 3, 2, and 1-min scan times (left to right).
Figures 1C and 1D show the same phantom reconstructed without (C) and with (D) TOF but now as a function of scan time, ranging from 5 min down to 1 min. The images with TOF are after 5 iterations; those for non-TOF are after 10 iterations, since these images appear closer to convergence (i.e., there is not a significant visual improvement in contrast but only an increase in noise with more iterations). For the TOF reconstruction there is an increase in background noise as the scan time is decreased, but all spheres are clearly visible for scan times > 2 min, and only the 10-mm sphere is not visible in the 1-min image. In contrast, for the non-TOF reconstruction the 10-mm sphere is not visible even in the 5-min image, and the 13-mm sphere is barely seen in the 2-min image and not seen in the 1-min image.
Figure 2 presents these visual impressions in a more quantitative manner. CRC vs. noise curves are shown for the 13- and 17-mm diameter spheres in the 27- (top) and 35-cm (bottom) cylinders for different scan times for TOF and non-TOF reconstructions. The data points correspond to 1, 2, 5, 10, 15, and 20 iterations (20 subsets/iteration). These curves demonstrate the improved convergence with TOF compared to non-TOF as well as with increased statistics. It can also be seen that the non-TOF reconstruction does not match TOF in the level of quantitative performance, as defined by the CRC/noise curve. Even a 5-min scan with non-TOF leads to a CRC/noise curve that is below that from the 2-min scan with TOF, consistent with the visual impressions. Results for the 22-mm sphere (not shown) were similar although the differences between TOF and non-TOF are smaller for this larger sphere. These CRC/noise curves show that convergence is reached faster with i) TOF, ii) a smaller diameter phantom, iii) a larger lesion, and iv) increased statistics or longer scan time. We believe that these trends are important, but the exact number of iterations needed to reach convergence will depend on details of the experimental set-up, the scanner characteristics, and the reconstruction algorithm, all of which are difficult to generalize. We also believe that there is evidence in these data that given a large enough phantom (or patient) and small enough lesion, the non-TOF reconstruction will not converge to the same value as the TOF reconstruction. This finding is supported by our visual impression of improved detectability of the 10-mm sphere with TOF as seen in Figure 1.
Figure 2.
CRC vs. noise curves for 13-mm (left) and 17-mm (right) hot spheres with 6:1 contrast in 27- (top) and 35-cm (bottom) cylinders. Scan times on the Gemini TF scanner were 1 ( ), 2 ( ), and 3 ( ) min (27-cm phantom) and 2 ( ), 3 ( ), 4 ( ), and 5 ( ) min (35-cm phantom) with closed symbols for non-TOF and open symbols for TOF reconstruction as a function of number of iterations (1, 2, 5, 10, 15, and 20).
Patient oncologic imaging
Figure 3 shows transverse images for two heavy patients: (top) with colon cancer (weight=119 kg, body mass index, BMI = 46.5) and (bottom) with abdominal cancer (115 kg, BMI = 38). In both cases the images reconstructed with TOF (right) have improved structural detail. The first example indicates uptake in a lesion, correlated with CT, that is difficult to see in the non-TOF image. Figure 4 shows a third patient (140 kg, BMI = 46) with non-Hodgkin’s lymphoma and multiple lesions. The different lesions are seen more clearly and with higher uptake in the TOF reconstruction (bottom) than in the non-TOF reconstruction (top).
Figure 3.
Representative transverse sections of two different patients: low dose CT (left), non-TOF ML-EM (middle), and TOF ML-EM (right). Patient 1 (top) with colon cancer (119 kg, BMI = 46.5) shows a lesion in the abdomen seen in CT much more clearly in the TOF image than in the non-TOF image. Patient 2 (bottom) with abdominal cancer (115 kg, BMI = 38) shows structure in the aorta seen in CT much more clearly in the TOF image than in the non-TOF image.
Figure 4.
Patient with non-Hodgkin’s lymphoma (140 kg, BMI = 46). The top row shows representative transverse, sagittal, and coronal images (not triangulated) for non-TOF reconstruction, while bottom row shows the same cross-sectional images for TOF reconstruction. In each of these images the different lesions are seen more clearly in the TOF reconstruction than the non-TOF reconstruction.
Figure 5 shows the improved contrast recovery achieved with TOF information in the third patient. Figure 5 (left) shows an anterior projection image where the letters (a–g) denote the lesions used in the L/B ratio analysis. Figure 5 (middle, right) show plots of the L/B ratio as a function of noise in the liver ROI for each of the lesions without (middle) and with (right) TOF. The L/B ratio curves with and without TOF are plotted on the same scale for direct comparison.
Figure 5.
Patient with non-Hodgkin’s lymphoma (140 kg, BMI = 46). Anterior projection image after 10 iterations of TOF ML-EM reconstruction is shown (left). The letters (a–g) denote the lesions that were used in the L/B ratio analysis. L/B ratio is plotted vs. noise in the liver ROI for 1–10 iterations for each of the lesions for non-TOF (middle) and TOF (right) reconstructions.
The TOF gain in CRC was calculated as the ratio of L/B ratio with TOF (3 iterations, 33 subsets/iteration) to that without TOF, where the number of iterations for the non-TOF reconstruction was chosen to match the pixel-to-pixel noise in the large liver ROI. Figure 6 shows a plot of TOF gain, averaged over the 6–9 lesions for a given patient, as a function of patient mass. The error bars show the range of TOF gains for that subject. There is a clear trend towards higher TOF gain with increasing patient size, which has been supported by a larger patient study of 30 patients (39).
Figure 6.
TOF gain as a function of patient mass. The TOF gain for matched noise levels, averaged over the 6–9 lesions (1–2 cm diameter) for each patient, is plotted as a function of patient mass. The error bars reflect the range of TOF gains seen for that patient.
DISCUSSION
The simulation results of (30) illustrated that TOF leads to a better CRC vs. noise trade-off but one that is difficult to quantify in terms of a simple sensitivity gain improvement: a single gain factor for TOF improvement does not include the increased rate of convergence with TOF, nor does it consider that TOF may converge to a different contrast than non-TOF, although the same statistical algorithm and optimal reconstruction parameters were used in both cases. These results are confirmed with the experimental phantom data. While any difference in CRC values to which TOF and non-TOF reconstructions converge is important, the increased rate of convergence for TOF is also of practical significance. One needs to fix the number of iterations for an iterative clinical reconstruction, and the faster and less spatially variant convergence of the TOF reconstruction leads to more stable results at earlier iterations.
The patient images demonstrate that the TOF advantages of higher contrast and faster convergence seen in simulated and phantom studies also translate to clinical studies. In a small study of 5 patients, it was seen that the L/B ratio after 10 iterations is higher with TOF than without TOF with only one exception of the 37 lesions studied. In routine practice, only three iterations are used to keep reconstruction times clinically practical (i.e., to complete reconstruction within 10–30 minutes after acquisition ends). After three iterations, however, the noise in the TOF image is higher than that in the non-TOF image after three iterations. If one chooses a non-TOF iteration where noise is matched with that in the corresponding TOF image (4–10 iterations, depending on the patient), the gain in L/B ratio with TOF ranges between 3–49%. In the heaviest patient (140 kg) the L/B ratio after 3 iterations with TOF is higher than that without TOF at matched noise by 36–49%. For the moderately heavy patients (91 and 105 kg) this improvement is in the range of 17–44% and 3–40%, respectively. For the average patient (77 kg) the improvement is in the range of 16–43%. It is worth pointing out that it is almost impossible to compare the noise at a fixed L/B ratio, since most lesions in non-TOF images never reach the TOF L/B ratio.
It can also be seen from the patient studies that the L/B ratio curves with TOF have reached or are very close to convergence by 10 iterations. In fact, the L/B ratio with TOF is within 5% of the 10-iteration value after three iterations in 30 of the 37 lesions; after four iterations, 95% of the lesions are within their 10-iteration value. The curves without TOF, on the other hand, show a greater spread in convergence rates. Many lesions, especially those close to other areas of high uptake, are quite slow to converge and still have not reached an asymptotic value by 10 iterations. For 23 of the 37 lesions studied, the L/B ratio after three iterations differs from the 10-iteration value by more than 5%. These results suggest that convergence with TOF is more spatially invariant, as noted in a recent study using simulated data (40).
As was observed from the phantom data, TOF allows one to image for shorter times and still achieve good image quality (as measured with the CRC/noise analysis and visual impressions). With or without TOF, image noise increases for shorter scans because fewer events are detected; however, TOF reconstruction leads to improved structural detail, so images with fewer counts may still have satisfactory image quality. This implies that TOF can be beneficial in situations where few counts are collected: e.g., dynamic imaging; respiratory gating; and imaging with non-pure positron-emitters, such as 124I, where the positron branching ratio is markedly lower than that of 18F and the complex decay scheme of gammas leads to additional random and cascade coincidences.
CONCLUSION
In this work, we have demonstrated that the benefit of TOF in PET imaging seen in simulation carries through to measured phantom and patient studies. Our findings in this paper lead us to conclude that TOF leads to a better CRC vs. noise trade-off in both phantoms and patients, but the TOF gain is not adequately expressed by a single gain factor, since the advantages of TOF are more than a simple increase in effective sensitivity. TOF reconstruction leads to a higher contrast recovery at matched noise with faster and more uniform convergence, and the benefit is even greater for larger patients. This paper attempts to quantify the TOF benefit using clinical data and shows results that are consistent with the phantom studies. Studies to assess how these results translate to clinical diagnosis and patient management are currently underway. In addition, we are investigating faster scintillators for better timing resolution and more computationally efficient reconstruction approaches.
ACKNOWLEDGMENTS
This work was supported by National Institutes of Health grants Nos. R33-EB001684 and R01-CA113941 and sponsored research agreements with Philips Medical Systems and Saint-Gobain Crystals. We would like to thank members of the Physics & Instrumentation Group at the University of Pennsylvania and the PET engineering team from the Nuclear Medicine Division of Philips Medical Systems for their technical and scientific help with this work. We would especially like to thank Matthew Werner for implementing the TOF reconstruction and SSS algorithms and for processing much of the data, Joshua Scheuermann for assistance with the phantom data, and Amy Perkins from Philips Research for analysis and helpful discussions about the patient data.
REFERENCES
- 1.Anger HO. Survey of Radioisotope Cameras. ISA Trans. 1966;5(4):311. [Google Scholar]
- 2.Brownell GL, Burnham CA, Wilensky S, Aronow S, Kazemi H, Streider . New developments in positron scintigraphy and the application of cyclotron produced positron emitters. Vienna: 1969. [Google Scholar]
- 3.Budinger TF. Instrumentation trends in nuclear medicine. Semin Nucl Med. 1977;7(4):285–297. doi: 10.1016/s0001-2998(77)80037-8. [DOI] [PubMed] [Google Scholar]
- 4.Ter-Pogossian MM, Ficke DC, Hood JT, Sr, Yamamoto M, Mullani NA. PETT VI: A positron emission tomograph utilizing cesium fluoride scintillation detectors. J Comput Assist Tomogr. 1982;6:125–133. [PubMed] [Google Scholar]
- 5.Allemand R, Gresset C, Vacher J. Potential advantages of a cesium fluoride scintillator for a time-of-flight positron camera. J Nucl Med. 1980;21(2):153–155. [PubMed] [Google Scholar]
- 6.Wong WH, Mullani NA, Philippe EA, et al. Performance-Characteristics of the University-of-Texas TOFPET-I PET Camera. J Nucl Med. 1984;25(5):P46–P47. [Google Scholar]
- 7.Lewellen TK, Bice AN, Harrison RL, Pencke MD, Link JM. Performance-Measurements of the SP3000/UW Time-of-Flight Positron Emission Tomograph. IEEE Trans Nucl Sci. 1988 Feb;35(1):665–669. [Google Scholar]
- 8.Mazoyer B, Trebossen R, Schoukroun C, et al. Physical Characteristics of TTV03, a New High Spatial-Resolution Time-of-Flight Positron Tomograph. IEEE Trans Nucl Sci. 1990 Apr;37(2):778–782. [Google Scholar]
- 9.Lewellen TK. Time-of-Flight PET. Semin Nucl Med. 1998;28(3):268–275. doi: 10.1016/s0001-2998(98)80031-7. [DOI] [PubMed] [Google Scholar]
- 10.Moses WW. Time of flight in PET revisited. IEEE Trans Nucl Sci. 2003 Oct;50(5):1325–1330. [Google Scholar]
- 11.Moses WW. Current trends in scintillator detectors and materials. Nucl Instrum Meth A. 2002 Jul;487(1–2):123–128. [Google Scholar]
- 12.van Eijk CWE. Inorganic scintillators in medical imaging. Phys Med Biol. 2002;47:R85–R106. doi: 10.1088/0031-9155/47/8/201. [DOI] [PubMed] [Google Scholar]
- 13.Lewellen TK, Miyaoka RS, Kohlmyer SG. Improving the Performance of the SP-3000 PET Detector Modules. IEEE Trans Nucl Sci. 1992 Aug;39(4):1074–1078. [Google Scholar]
- 14.Surti S, Kuhn A, Werner ME, Perkins AE, Kolthammer J, Karp JS. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med. 2007 Mar;48(3):471–480. [PubMed] [Google Scholar]
- 15.Snyder DL, Thomas LJ, Terpogossian MM. A Mathematical-Model for Positron-Emission Tomography Systems Having Time-of-Flight Measurements. IEEE Trans Nucl Sci. 1981;28(3):3575–3583. [Google Scholar]
- 16.Tomitani T. Image-Reconstruction and Noise Evaluation in Photon Time-of-Flight Assisted Positron Emission Tomography. IEEE Trans Nucl Sci. 1981;28(6):4582–4589. [Google Scholar]
- 17.Snyder DL. Some Noise Comparisons of Data-Collection Arrays for Emission Tomography-Systems Having Time-of-Flight Measurements. IEEE Trans Nucl Sci. 1982;29(1):1029–1033. [Google Scholar]
- 18.Politte DG, Snyder DL. Results of a Comparative-Study of a Reconstruction Procedure for Producing Improved Estimates of Radioactivity Distributions in Time-of-Flight Emission Tomography. IEEE Trans Nucl Sci. 1984;31(1):614–619. [Google Scholar]
- 19.Snyder DL, Politte DG. Image-Reconstruction from List-Mode Data in an Emission Tomography System Having Time-of-Flight Measurements. IEEE Trans Nucl Sci. 1983;30(3):1843–1849. [Google Scholar]
- 20.Reader AJ, Erlandsson K, Flower MA, Ott RJ. Fast accurate iterative reconstruction for low-statistics positron volume imaging. Phys Med Biol. 1998 Apr;43(4):835–846. doi: 10.1088/0031-9155/43/4/012. [DOI] [PubMed] [Google Scholar]
- 21.Parra L, Barrett HH. List-mode likelihood: EM algorithm and image quality estimation demonstrated on 2-D PET. IEEE Trans Med Imag. 1998;17(2):228–235. doi: 10.1109/42.700734. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Huesman RH, Klein GJ, Moses WW, Qi JY, Reutter BW, Virador PRG. List-mode maximum-likelihood reconstruction applied to positron emission mammography (PEM) with irregular sampling. IEEE Trans Med Imag. 2000 May;19(5):532–537. doi: 10.1109/42.870263. [DOI] [PubMed] [Google Scholar]
- 23.Kimdon JA, Qi J, Moses WW. Paper presented at: 2003 IEEE Nuclear Science Symposium and Medical Imaging Conference. Portland, OR: 2003. Effect of random and scatter fractions in variance reduction using time-of-flight information. [Google Scholar]
- 24.Popescu LM. Paper presented at: 2004 IEEE Nuclear Science Symposium and Medical Imaging Conference. Rome, Italy: 2004. Iterative image reconstruction using geometrically ordered subsets with list-mode data. [Google Scholar]
- 25.Popescu LM, Lewitt RM. Paper presented at: 2004 IEEE Nuclear Science Symposium and Medical Imaging Conference. Rome, Italy: 2004. Tracing through a grid of blobs. [Google Scholar]
- 26.Wang W, Hu Z, Gualtieri EE, et al. Paper presented at: 2006 IEEE Nuclear Science Symposium and Medical Imaging Conference. San Diego, CA: 2006. Systematic and Distributed Time-of-Flight List Mode PET Reconstruction. [Google Scholar]
- 27.Conti M, Bendriem B, Casey M, et al. First experimental results of time-of-flight reconstruction on an LSO PET scanner. Phys Med Biol. 2005 Oct 7;50(19):4507–4526. doi: 10.1088/0031-9155/50/19/006. [DOI] [PubMed] [Google Scholar]
- 28.Casey M, Panin V, Bendriem B. Early clinical experience with time-of-flight PET. J Nucl Med. 2006;47(5):184P. [Google Scholar]
- 29.Karp JS, Kuhn A, Perkins AE, et al. Paper presented at: 2005 IEEE Nuclear Science Symposium and Medical Imaging Conference. San Juan, Puerto Rico: 2005. Characterization of TOF PET scanner based on Lanthanum Bromide. [Google Scholar]
- 30.Surti S, Karp JS, Popescu LA, Daube-Witherspoon ME, Werner M. Investigation of time-of-flight benefit for fully 3-D PET. IEEE Trans Med Imag. 2006 May;25(5):529–538. doi: 10.1109/TMI.2006.871419. [DOI] [PubMed] [Google Scholar]
- 31.Budinger TF. Time-of-Flight Positron Emission Tomography - Status Relative to Conventional PET. J Nucl Med. 1983;24(1):73–76. [PubMed] [Google Scholar]
- 32.Daube-Witherspoon ME, Surti S, Matej S, Werner M, Jayanthi S, Karp JS. Paper presented at: 2006 IEEE Nuclear Science Symposium and Medical Imaging Conference. San Diego, CA: 2006. Influence of Time-of-Flight Kernel Accuracy in TOF-PET Reconstruction. [Google Scholar]
- 33.Perkins AE, Werner M, Kuhn A, Surti S, Muehllehner G, Karp JS. Paper presented at: 2005 IEEE Nuclear Science Symposium and Medical Imaging Conference. San Juan, PR: 2005. Time of flight coincidence timing calibration techniques using radioactive sources. [Google Scholar]
- 34.Griesmer J, Laurence T, Cooke S, Karp J, Perkins A, Kolthammer J. Time-of-flight quality control for a new Philips Gemini PET/CT scanner. J Nucl Med. 2006;47(5):391P. [Google Scholar]
- 35.Watson CC. Paper presented at: 2005 IEEE Nuclear Science Symposium and Medical Imaging Conference. San Juan, PR: 2005. Extension of single scatter simulation to scatter correction of time-of-flight PET. [Google Scholar]
- 36.Accorsi R, Adam LE, Werner ME, Karp JS. Optimization of a fully 3D single scatter simulation algorithm for 3D PET. Phys Med Biol. 2004 Jun;49(12):2577–2598. doi: 10.1088/0031-9155/49/12/008. [DOI] [PubMed] [Google Scholar]
- 37.Werner ME, Surti S, Karp JS. Paper presented at: 2006 IEEE Nuclear Science Symposium and Medical Imaging Conference. San Diego, CA; 2006. Implementation and Evaluation of a 3D PET Single Scatter Simulation with TOF Modeling. [Google Scholar]
- 38.Daube-Witherspoon ME, Karp JS, Casey ME, et al. PET performance measurements using the NEMA NU 2-2001 standard. J Nucl Med. 2002;43:1398–1409. [PubMed] [Google Scholar]
- 39.Perkins AE, Saffer JR, S SJ, Werner ME, Karp JS, Divgi CR. Clinical optimization of the acquisition time of FDG time-of-flight PET. J Nucl Med. 2007;48(5):91P. [Google Scholar]
- 40.Vandenberghe S, Karp J, Lemahieu I. Influence of TOF resolution on object dependent convergence in iterative listmode MLEM. J Nucl Med. 2006;47(5):58P. [Google Scholar]