Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 1994 Jun;32(6):1547–1549. doi: 10.1128/jcm.32.6.1547-1549.1994

Controlled clinical evaluation of Isolator and ESP aerobic blood culture systems for detection of bloodstream infections.

B A Kirkley 1, K A Easley 1, J A Washington 1
PMCID: PMC264035  PMID: 8077401

Abstract

A controlled clinical evaluation comparing the Isolator system (Wampole Laboratories, Cranbury, N.J.) and the ESP 80A blood culture bottle in the automated ESP system (Difco Laboratories, Detroit, Mich.) was performed with 10,535 blood culture sets from patients with suspected septicemia. Of 1,150 positive cultures, 844 positive cultures from 285 patients with 394 septic episodes fulfilled the study criteria for minimum blood sample requirements in each system and clinical significance of isolates. The Isolator system detected statistically significantly more positive cultures of Staphylococcus aureus (P < 0.001), Enterococcus spp. (P = 0.007), Escherichia coli (P = 0.001), Alcaligenes xylosoxidans (P = 0.02), Xanthomonas maltophilia (P = 0.01), Candida albicans (P < 0.001), and Candida glabrata (P = 0.05). The Isolator system detected significantly more septic episodes due to S. aureus (P < 0.001), X. maltophilia (P = 0.02), and C. albicans (P = 0.004) than did the ESP 80A bottle; however, the two systems did not otherwise significantly differ in their abilities to detect septic episodes due to other organisms.

Full text

PDF
1547

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Garner J. S., Jarvis W. R., Emori T. G., Horan T. C., Hughes J. M. CDC definitions for nosocomial infections, 1988. Am J Infect Control. 1988 Jun;16(3):128–140. doi: 10.1016/0196-6553(88)90053-3. [DOI] [PubMed] [Google Scholar]
  2. Kirkley B. A., Easley K. A., Basille B. A., Washington J. A. Controlled clinical comparison of two lysis-based blood culture systems, isolator and Septi-Chek Release, for detection of bloodstream infections. J Clin Microbiol. 1993 Aug;31(8):2114–2117. doi: 10.1128/jcm.31.8.2114-2117.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Morello J. A., Leitch C., Nitz S., Dyke J. W., Andruszewski M., Maier G., Landau W., Beard M. A. Detection of bacteremia by Difco ESP blood culture system. J Clin Microbiol. 1994 Mar;32(3):811–818. doi: 10.1128/jcm.32.3.811-818.1994. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Washington J. A., 2nd, Ilstrup D. M. Blood cultures: issues and controversies. Rev Infect Dis. 1986 Sep-Oct;8(5):792–802. doi: 10.1093/clinids/8.5.792. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)