INTRODUCTION
Breakthroughs in biology often arise from the development of new microscopy tools. Our understanding of biological systems has greatly benefited from the studies with electron microscopy, phase contrast microscopy, confocal fluorescence microscopy, scanning probe microscopy, and other imaging tools. In recent years, the coupling of nonlinear optics with scanning microscopy opened up a new window through which to watch the biological world. The tight focusing condition in microscopy allows an efficient generation of nonlinear optical (NLO) signals even with low pulse energies. Meanwhile, the nonlinear dependence on excitation intensity ensures that the NLO signal is only generated in the focal center, providing an inherent three-dimensional (3D) spatial resolution. In this sense, nonlinear optics was “born” for imaging use.
The earliest development of NLO microscopy can be traced back to the 1970s when second-harmonic imaging of crystals was reported as a proof-of-principle of second-harmonic generation (SHG) microscopy.1,2 In 1982, Duncan and co-workers reported coherent anti-Stokes Raman scattering (CARS) imaging of D2O in onion cells using a dye laser system.3 However, these early papers did not prompt a lot activity in the 1980s. The prosperity of the multiphoton and NLO microscopy field was triggered by the demonstration of two-photon excitation fluorescence (TPEF) imaging with a mode-locked femtosecond dye laser in 1990,4 followed by demonstrations of high-quality third harmonic generation (THG) imaging5 and CARS imaging6 in the late 1990s. Nowadays, TPEF and SHG microscopy have become powerful tools for biological research.7,8
The first systematic study of the CARS phenomenon was published in 1965 by two scientists from the Ford Motor company.9 CARS spectroscopy emerged in the early 1970s10 and has been one of the most important coherent Raman spectroscopy techniques.11–13 In a CARS process, three laser fields at the pump (ωp), Stokes (ωs), and probe ( ) frequencies interact with a medium to generate a new field at the anti-Stokes frequency . In most experiments, the pump field Ep and probe field come from the same laser beam. The CARS signal arises from the third-order-induced polarization, . CARS occurs in any medium with nonzero susceptibility χ(3). Importantly, the CARS signal can be significantly enhanced when the beating frequency, (ωp − ωs), is in resonance with a molecular vibration (Fig. 1A). By scanning (ωp − ωs), one can acquire a CARS spectrum that carries essentially the same information as the spontaneous Raman spectrum of the same sample. However, because of the interference between the Raman component and the nonresonant component, the CARS line profile is dispersed, with a shift of the peak to lower frequency and appearance of a dip at higher frequency (Fig. 1B).
In a CARS experiment, other coherent Raman processes occur simultaneously, including coherent Stokes Raman scattering (CSRS), stimulated Raman gain (SRG), stimulated Raman loss (SRL), and Raman-induced Kerr effect (RIKE).11,14 In CSRS, a new field at frequency ωs − (ωp − ωs) is generated. In SRG, the ωs field experiences a gain while in SRL the ωp field experiences a loss. In RIKE, new polarization components of the excitation fields are generated. In most CARS imaging experiments, two parallel-polarized beams are used, where RIKE is negligible (χ2111 = 0 for an isotropic medium). In contrast to SRG, SRL, and RIKE, the CARS signal is spectrally separated from the excitation beams, making it easy to filter out the linear scattering from a heterogeneous biological system. Compared with CSRS, the CARS signal appears at a wavelength shorter than excitation wavelengths, usually in the visible region where detectors of high quantum efficiency are available.
In CARS microscopy,15–18 the pump and Stokes beams are tightly focused into a sample and a CARS image is generated by scanning either the sample or the laser beams. Besides the 3D sectioning capability, CARS microscopy offers several unique advantages.
CARS microscopy permits nondestructive molecular imaging without any labeling. This advantage is important for imaging small molecules such as lipids or drugs for which the labeling may significantly affect the molecular properties.
The coherent addition of CARS fields results in a quadratic signal increase with respect to the number of molecular oscillators, versus the linear increase of spontaneous Raman signal. The coherent addition also results in a highly directional output, which greatly facilitates the signal collection. The large CARS signal allows high-speed vibrational imaging, which is not possible with Raman microscopy.
CARS signals appear at a wavelength that is shorter than excitation wavelengths, spectrally separated from the one-photon fluorescence background.
The major disadvantage of CARS is the existence of a nonresonant background which arises from the electronic contribution to χ (3). This nonresonant background is coherently mixed with the vibrationally resonant signal, which limits the sensitivity of CARS in detecting weak Raman bands. Additionally, the resonant CARS signal decreases quadratically with the molecular concentration, making the detection of low-concentration molecules difficult. In the past few years, tremendous efforts have been spent on suppressing the nonresonant background and pushing the detection sensitivity limit.
This paper aims to evaluate the potential and limitations of CARS microscopy through a review of recent advances. The rest of the paper will discuss the technical developments, summarize the biological and biomedical applications, and finally present an outlook.
TECHNICAL DEVELOPMENTS
The first attempt toward using CARS as a microscopy technique was made by Duncan and co-workers in the 1980s.3,19,20 Using visible dye lasers, non-collinear beam geometry, and a charge-coupled device (CCD) detector to record the signal at the phase matching direction, the authors acquired CARS images of D2O distribution in onion cells. Xie and co-workers revived CARS microscopy in 1999 using near-infrared laser pulses and a collinear beam geometry.6 Significant advances in the last few years have matured CARS microscopy, as summarized below.
Construction of a Robust Coherent Anti-Stokes Raman Scattering Microscope
The combined use of the following strategies has produced a CARS microscope for imaging highly dynamic systems with repeatable results. A schematic of such a microscope is shown in Fig. 2. It is notable that most of these strategies involved the breakdown of conventional wisdoms rooted from CARS spectroscopy.
Near-Infrared Laser Excitation
The use of a near-infrared laser pulse first by Zumbusch et al. avoided two-photon electronic enhancement of the nonresonant contribution.6 Because the generation of CARS does not need one-photon or two-photon electronic resonance, even longer excitation wavelengths (i.e., 1.2 μm) can be used to reduce tissue scattering and increase the optical penetration depth accordingly.21 The use of longer wavelengths could also reduce the photo-damage induced by multiphoton absorption.22
Collinear Beam Geometry
As a four-wave-mixing process, the generation of CARS needs to fulfill the phase matching condition, |kas − (2kp − ks)|L < π, where L is the light–sample interaction length. kp, ks, and kas are wave vectors of the pump, Stokes, and CARS fields, respectively. With a collinear beam geometry, dispersion of the refractive index (nj, j = p, s, and as) introduces a wave vector mismatch of (nasωas − 2npωp + nsωs)/c. In order to maximize the interaction length, non-collinear beam geometries were used in CARS spectroscopy to minimize the wave vector mismatch. This method was also used in the 1982 CARS microscope.3 Xie and co-workers realized that the phase mismatch could be relaxed with a tight focus using a high numerical aperture objective in microscopy.6 Because the interaction length is around several micrometers or less under the tight focusing condition, the phase matching condition can be fulfilled with a collinear beam geometry for forward detection. The collinear beam geometry greatly simplified the implementation and has been a key step in producing high-quality CARS images.
Picosecond Pulse Excitation
It is generally valid that femtosecond pulses provide a high peak intensity that is needed for NLO processes. However, this wisdom becomes invalid in CARS microscopy.23 In the spectral domain, the spectral width of a femtosecond pulse is much broader than the width of most Raman lines. As a result, most of the energy only contributes to the generation of nonresonant background. Instead, the spectral width of a picosecond pulse matches the Raman line width, so that the excitation energy can be focused on a single Raman band. Thus, the use of picosecond pulse excitation not only provides sufficient spectral resolution,24 but also increases the ratio of resonant signal to non-resonant background.23 Currently picosecond pulse excitation can be realized by using two synchronized Ti:sapphire lasers23 or using an optical parametric oscillator pumped by a pulsed laser.21,25
Epi- (i.e., Backward-) Detection
For isotropic bulk samples studied in CARS spectroscopy, the signal goes forward in the phase matching direction. Unlike spectroscopy, microscopy studies heterogeneous samples. For objects that are much smaller than the excitation wavelengths, the phase matching condition is relaxed and the CARS radiation pattern becomes symmetrical forward and backward.23,26 Systematic study showed that epi-detected CARS (E-CARS) could arise from objects of small interaction length, discontinuity of χ(3) at an interface, and back-reflection of forward CARS.27 Because the CARS signal from the surrounding solvent goes forward, E-CARS provided a means to detect small objects embedded in a bulk medium.23,26 Another significant application of E-CARS is live animal imaging for which it is impossible for the forward CARS signal to pass through a thick tissue.28,29
Laser Scanning
In early papers, CARS images were generated by using low repetition rate (in KHz) pulses that had a high peak power in order to ensure a large signal level. However, the use of low repetition rate pulses necessitated minutes for recording an image on a sample scanning platform, which was not suitable for live cell or in vivo imaging. Later, using a confocal microscope (Olympus, FV300/IX71), it was shown that the same sensitivity could be achieved by scanning picosecond lasers of high repetition rates (in MHz), with an image acquisition speed of a frame per second.30 More recently, video-rate scanning CARS has been demonstrated.28 The implementation of CARS imaging on a laser-scanning microscope platform not only ensures high-speed imaging of highly dynamic systems, but also makes CARS microscopy compatible with a common confocal fluorescence microscope and other NLO imaging modalities. Similar to TPEF microscopy, the confocal pinhole is not used for laser-scanning CARS imaging. Additionally, non-descanned large area detectors can be used to increase the detection sensitivity for tissue imaging. A major difference between CARS and TPEF is that CARS utilizes two excitation laser beams at different wavelengths. Therefore, microscope objectives with minimal chromatic aberrations should be used to ensure effective signal generation as well as collection.
Suppression of Nonresonant Background
For fingerprint Raman bands such as the amide I band, the resonant CARS signal could be buried in the large nonresonant background from the object of interest as well as the surrounding medium. Several strategies were used to improve the signal to background ratio in CARS spectroscopy, including polarization-sensitive detection,31,32 time-resolved detection,33 electronic resonance excitation,34,35 and interferometry.36 Except for resonance CARS, which may induce serious damage under the tight focusing condition, these methods have been demonstrated on the microscope platform for background-free vibrational imaging, as summarized below.
Polarization-Sensitive Detection
Polarization-sensitive detection37,38 utilizes the different polarization properties of the resonant and nonresonant components of χ(3). Away from one-photon electronic resonance, the depolarization ratio of the nonresonant CARS field equals 1/3 following the Kleinman’s symmetry.39 The depolarization ratio of the resonant CARS field equals the Raman depolarization ratio and varies in the range of 0 to 0.75. The nonresonant background can be rejected by placing an analyzer before the detector while a projection of the resonant signal along the analyzer polarization can be detected.
Time-Resolved Detection
Time-resolved detection40 makes use of the different free-induction decay rates of the resonant and nonresonant contributions. With a flat profile in the spectral domain, the nonresonant background contributes to an instantaneous spike in the time domain due to electronic dephasing. The resonant part involves a vibrational transition and has a longer dephasing time (on the order of one picosecond) that is related to the spectral width of the corresponding Raman line. Thus, the nonresonant background can be eliminated by introducing a suitable time delay between the femtosecond pump/Stokes pulses and the probe pulse.
Spectral and Spatial Phase Control
In spectral phase control,41,42 the pulse shaping technique was applied to modulate the phase of different spectral components in a femtosecond pulse with a periodic function (period Ωm). As a result, a Raman-resonant signal with vibrational frequency of NΩm (N is an integer) gets constructive interference, while the nonresonant signal is canceled by destructive interference. This method allowed CARS imaging with a single femtosecond beam.41 Recently, spatial phase shaping of focus was implemented to cancel the coherent addition of the nonresonant signal from a bulk medium, which permitted the visualization of interfaces.43,44 It should be noted that the nonresonant signal from bulk solvent can also be canceled by using counter-propagating pump and Stokes laser beams to introduce a large spatial phase mismatch inside the focal volume.27
Interferometric Coherent Anti-Stokes Raman Scattering
Interferometric coherent anti-Stokes Raman scattering45–52 separates the real and imaginary parts of χ(3) by using a local oscillator to interfere with the CARS signal from a sample. Moreover, this approach amplifies the signal level because the detected signal is the interference term between the resonant signal and the local oscillator.
Despite nice proof-of-principle demonstrations using standard samples such as polymer beads or solvents, the complexity of these advanced techniques may hamper their use as an analytical tool to study real tissue samples in which the laser intensity, phase, and polarization can be significantly changed. Time-resolved detection is not affected by tissue scattering, but its ultimate sensitivity is limited by the use of femtosecond pulses, which increases the nonresonant background in the first place.
The simplest and most robust method for background removal is to digitally subtract the off-resonance CARS image from the on-resonance one. This method was first used by Duncan et al. in CARS imaging of deuterated liposomes.20 By comparing the difference between the two images acquired at the peak and dip positions of a CARS band, quantitative detection of a specific deuterated lipid in a single bilayer was recently demonstrated.53 To facilitate the study of more dynamic systems, the on- and off-resonance signals can be acquired simultaneously or in a frequency modulation manner by using three pulsed lasers.21,25
For C–H stretch vibrations, the Raman-resonance enhancement of χ(3) is significant because of the high density of C–H bonds in biological molecules, particularly in lipids. A resonant signal to nonresonant background ratio of more than 10:1 was obtained through multiplex CARS spectral analysis of multilamellar lipid membranes.38 With the large signal from C–H vibration, CARS imaging of lipid bodies,28,54–56 single cell membranes,30,57 myelin sheath in central and peripheral nervous systems,29,58 and drug molecules59 has been demonstrated. Virtually no sophisticated background reduction schemes are needed for imaging C–H abundant objects.
Improving the Detection Sensitivity
In CARS, the coherent addition of anti-Stokes Raman scattering photons results in a large and directional signal. For a bulk liquid, the CARS signal was estimated to be larger than the spontaneous Raman signal by nine orders of magnitude.12 Thus, CARS microscopy is naturally sensitive to molecular assemblies such as lipid membranes. On the other hand, the resonant CARS signal drops quadratically with the reduction of the concentration of target molecules and can be buried in the nonresonant background. The sensitivity of CARS microscopy was significantly improved by using picosecond pulse excitation to focus all the energy on a single Raman band.23 The use of epi-detection to reject the solvent background allowed detection of 100 nm polystyrene bead that contains 106 aromatic C–H oscillators.30 The sensitivity of epi-detection was further enhanced by mixing the epi-scattered resonant CARS field with the back-reflected nonresonant CARS field.57 For a d62-DPPC bilayer supported by a cover glass, such coherent mixing caused a highly dispersed spectral profile superimposed on a flat nonresonant background. By measuring the E-CARS intensity acquired at the peak and dip frequencies, Li et al. was able to probe 10% d62-DPPC in a single bilayer,53 which corresponded to 2 × 104 DPPC molecules inside the focal area of 0.25 μm2.
For forward detection, frequency-modulation CARS allowed the detection of 5 × 105 methanol molecules in the excitation volume of 0.1 fL, where the CARS intensities on-and off-resonance with the CH3 stretch vibration were used.60 It is worth noting that the detection sensitivity depends on the target molecule and the vibrational mode as well. For CARS imaging of natural paclitaxel molecules, the forward CARS signal from the aromatic CH vibration was used for detecting paclitaxel dissolved in N,N,-dimethyl-formamide solution. The detection limit was 1.8 attomole paclitaxel in the excitation volume of 0.06 fL, corresponding to 29 mM.59 Such a sensitivity allowed imaging of paclitaxel dispersed in a polymer film59 but is still inadequate for imaging the drug distribution inside of a paclitaxel-treated mammalian cell.
Multimodal Nonlinear Optical Imaging on a Coherent Anti-Stokes Raman Scattering Microscope
Multimodality is important because different NLO imaging methods each have their distinctive advantages: TPEF can be used to visualize proteins, ions with fluorescent labeling or specific auto-fluorescent structures; SHG and sum-frequency generation (SFG) are selective to non-centrosymmetric molecular assemblies such as collagen fibrils; and CARS is naturally sensitive to lipid-enriched structures such as adipocytes. Fu et al. realized that the two picosecond laser beams for CARS imaging could also be used for electronic SFG imaging.61 The authors demonstrated multimodal NLO imaging of ex vivo spinal tissues by CARS imaging of myelin sheath, SFG imaging of astrocyte processes, and TPEF imaging of calcium indicators on the same platform.61 At the single cell level, Nan et al. combined CARS with TPEF to characterize the relationship between lipid droplets and mitochondria.55 Multimodal NLO imaging has recently been used to map the molecular composition of atheroma62 and to evaluate the impact of obesity on adipocyte size and collagen fibril density in mammary tumor stroma.56
Integrating Coherent Anti-Stokes Raman Scattering Microscopy with Spectroscopy
Spectral analysis provides important information that is not available in a single-frequency CARS image. The high-speed imaging on a laser-scanning CARS microscope was achieved by using picosecond pulses to focus all the excitation power on a single Raman band. In this configuration, a CARS spectrum has to be recorded in a time-consuming manner by tuning the Stokes laser frequency point by point. The past few years saw a great interest in the development of multiplex CARS microscopy and microspectroscopy with pico/femto excitation,38,63 chirped pulse,64,65 fiber-based super-continuum,66–70 and single broadband pulses.41,48,71 Recently hyper-Raman microspectroscopy has also been coupled with CARS imaging using picosecond pulses.72
Multiplex CARS microscopy is usually implemented by raster-scanning a sample with a pixel dwell time on the order of milliseconds, limiting its application to dynamic biological systems. On the other hand, if one only needs to record the spectrum at a point of interest, microspectrometry can be implemented with the point-scan mode of a laser-scanning microscope,62 where high-speed imaging and spectral recording could be obtained from the same sample. It should be noted that interpretation of CARS spectra is complicated by the nonresonant background. Recently, the maximum entropy method was successfully used to fit the CARS spectra and extract the imaginary part of the susceptibility without prior knowledge of the sample composition.73
In Vivo Coherent Anti-Stokes Raman Scattering Imaging
In comparison to imaging of cultured cells or extracted tissues, in vivo optical imaging provides a much better means to monitor cells over time without significant change or perturbation of their environment. CARS microscopy allows label-free molecular imaging. This advantage is particularly important for in vivo studies for which the labeling is complicated by inefficient diffusion and non-specific binding in the tissue environment. Evans et al. demonstrated E-CARS imaging of sebaceous glands, corneocytes, and adipocytes in the skin of a live mouse.28 By using minimal surgery to cut open the skin, Huff et al. demonstrated in vivo E-CARS imaging of sciatic nerve and surrounding adipocytes.29 The contrast mechanisms for in vivo CARS have been carefully investigated. The E-CARS signals from the sciatic nerve tissues were found to arise from interfaces as well as back reflection of forward CARS. For the densely packed skin tissue, the forward-propagating photons would experience a significantly greater number of scattering events. The skin could also distort the point spread function at the foci,74 which eliminates the E-CARS signal from interfaces. In such a case, multiple reflection of forward CARS becomes the major source of E-CARS.28
Analysis and Interpretation of a Coherent Anti-Stokes Raman Scattering Image
The quantitative information that can be extracted from a CARS image is the intensity at each pixel. Converting the CARS intensity into the information of molecular composition is, however, not straightforward because of the non-resonant background that depends on the sample property at each pixel. Only in well-defined samples such as supported lipid bilayers was quantitative CARS imaging demonstrated.53 Because the generation of CARS critically depends on the tight overlap of the pump and Stokes laser foci, it is suggested to routinely check the spatial overlap of the two laser beams using the transmission channel of the microscope. Calibration of the system using the nonresonant signal from a standard sample such as a cover glass is also recommended when recording a CARS spectrum or quantitatively comparing the CARS intensities between different samples. It is important to mention that for a complex biological sample, the tightness of the foci may be deteriorated by objects in the focusing path. For instance, when the foci were above a lipid droplet, the forward CARS signal was significantly reduced and a dark hole was observed in the image.30 For the same reason, the efficiency of CARS generation also highly depends on the optical penetration depth. Thus, careful attention must be paid when interpreting the depth intensity profile.
The CARS intensity is also affected by the interference of two adjacent Raman bands. For this reason, when parallel-polarized pump and Stokes beams were used, the CARS peak for the symmetric CH2 stretch vibration was much more pronounced than that for the asymmetric CH2 vibration in lipids.38 Special attention should be paid to the CARS signal from OH vibration, which was shown to be 40 times smaller than that from CH vibration.75 In liposomes, the CARS contrast at 3200 cm−1, originally assigned to the interlamellar water,76 was later shown to be overwhelmed by the destructive interference term between the CH vibration and the nonresonant background.77
The dependence of CARS intensity on excitation beam polarization provides a means to determine the orientation of the oscillators. For the symmetric CH2 stretch vibration, the CARS intensity is maximized when the excitation beams are polarized along the dipole of the CH2 group.57,76 This method has been used to determine the ordering degree of lipids in myelin sheath,58 where standard samples were used to calibrate the microscope response to different excitation polarization. For a homogeneous sample, the CARS signal was shown to be linearly polarized along the excitation laser polarization.37 In contrast, the CARS polarization was significantly scrambled when generated from a tissue sample.58
Cell and Tissue Damage During Coherent Anti-Stokes Raman Scattering Imaging
In general, because CARS does not need one-photon or two-photon electronic resonance, multi-photon absorption induced damage is diminished in CARS microscopy with near-infrared picosecond pulse excitation. Nondestructive CARS imaging of live cells over a period of 300 s has been demonstrated.55 The mechanisms underlying the photo-damage in CARS microscopy have been studied using high excitation power and scanning over a small area.22,55 The damage to cultured cells was found to be nonlinear, featured by cell membrane blebbing22 or cell retraction.55 The damage to compact myelin sheath involved both linear and nonlinear mechanisms, depending on the peak power and pulse repetition rate applied.22 It was also found that the damage mainly arose from the pump laser beam, which has a shorter laser wavelength.22 Therefore, the use of a higher ratio of Stokes to pump power or longer excitation wavelengths helps the reduction of damage. Besides linear and nonlinear absorption of the incident photons, an extra source of photo-damage during CARS imaging arises from vibrational excitation. Theoretical analysis by Wang et al. showed that such vibrational excitation is due to the concomitant Raman gain and Raman loss processes, which contribute to an efficient light–matter energy transfer under the tight focusing condition.78
BIOLOGICAL AND BIOMEDICAL APPLICATIONS
The ultimate goal of developing CARS microscopy is to use it as an imaging tool to address scientific questions. Encouragingly, the number of application publications is rising and several significant applications to biology and medicine have been demonstrated, as summarized below.
Lipid Domains in Model Membranes
Lateral heterogeneity of lipid distribution in biological membranes plays a critical role in signal transduction, membrane traffic, and diseases.79 Using the CARS signal from the CH2 stretch vibration, detection of single bilayers was demonstrated.57,80 Potma et al. showed CARS imaging of lipid domains in giant unilamellar vesicles.81 Li et al. demonstrated quantitative E-CARS imaging of specific lipids in a supported bilayer.53 Molecular specificity was achieved by using lipids with deuterated acyl chains (i.e., d62-DPPC). The effect of deuteration on phase segregation was explored by comparing d62-DPPC-containing and DPPC-containing bilayers, with the same domain patterns being observed.53 An example of using E-CARS to study lipid domains is described below.
The E-CARS spectrum for the CD2 stretch vibration band of a single supported d62-DPPC bilayer is shown in Fig. 3A. Coherent mixing of the resonant signal from the bilayer and the nonresonant signal from the glass–water interface resulted in a peak at 2080 cm−1 and a dip at 2125 cm−1. The nonresonant background was removed by subtracting the dip CARS intensity from the peak CARS intensity. The (peak − dip) signal contained the cross term between the resonant and nonresonant CARS fields, which enhanced the detection sensitivity. As a result, a clear vibrational contrast was obtained from 10% d62-DPPC in a single bilayer. Moreover, Li et al. developed a theoretical model to convert the (peak − dip) intensity into molecular concentration.53 These efforts allowed quantitative measurement of the partition of d62-DPPC between coexisting domains, as shown in Figs. 3B and 3C. In summary, the ability to probe molecular compositions of lipid domains makes CARS microscopy a useful tool for the study of membrane organization.
Cellular Analysis
Laser-scanning CARS microscopy has been applied to vibrational imaging of cellular processes. By line-scanning the laser beams and using the resonant CARS signal from water, Potma et al. measured the intracellular water diffusion coefficient.82 By using a laser-scanning CARS microscope with simultaneous forward and backward detection, Cheng et al. characterized the morphology of mitosis and apoptosis in unstained NIH3T3 cells.30 Of all the cellular organelles, the lipid droplets produce the strongest CARS signal at the frequency of the C–H stretch vibration. Using the CARS signal, Nan et al. monitored adipogenesis in 3T3-L1 cells and also explored intracellular trafficking of lipid droplets in Y-1 cells.54,55 These studies showed the potential of CARS microscopy in characterization of 3D cell morphology as well as the study of intracellular lipid bodies.
Myelin Sheath and Demyelination
The myelin sheath, an extended plasma membrane wrapping around an axon, is crucial for high-speed impulse conduction. Numerous neurological diseases specifically attack the myelin sheath with disabling and often fatal results.83 The lamellar structure of myelin sheath has been well characterized by polarized light microscopy, electron microscopy, and X-ray diffraction. However, visualizing intact myelin in live tissues and in vivo cannot be accessed with these techniques.
The myelin membrane contains a high lipid to protein ratio with 70% lipid and 30% protein by weight. Wang et al. demonstrated CARS imaging of intact myelin sheath using the resonant CARS signal from the high-density CH2 groups in the myelin membrane.58 In their work, the spinal cord white matter strips were isolated from guinea pigs as illustrated in Fig. 4A. The CARS spectrum of a single axonal myelin is shown in Fig. 4B. The peak for the symmetric CH2 stretch vibration appeared at 2840 cm−1, with a resonant signal to nonresonant background ratio of 10:1. Additionally, a weak band at 2930 cm−1 was observed and assigned to the CH3 stretch mode of the myelin proteins. CARS images of myelin sheath surrounding the parallel axons and myelin around a node of Ranvier are shown in Figs. 4C and 4D, respectively. Importantly, the CARS intensity strongly depends on the excitation polarization because of the lamellar structure of the myelin membranes. Such polarization dependence as well as the CARS intensity provides means to distinguish healthy myelin from degraded myelin. Upon the treatment of lysophosphatidylcholine, a widely used demyelination agent, the myelin became swelled (Fig. 4E), characterized by a three-fold decrease of CARS intensity and loss of polarization dependence.84 These capabilities open up exciting opportunities to the pathological study of demyelinating diseases such as multiple sclerosis.
Native Drug Molecules in a Polymeric Delivery System
Visualization of three-dimensional distribution of drug molecules and subsequent changes during the release process is critical for understanding drug delivery mechanisms as well as designing tailor-made release profiles. Kang et al. demonstrated that CARS microscopy could be effectively utilized to examine paclitaxel distribution in various polymer films with lateral resolution of 0.3 μm and depth resolution of 0.9 μm.59 In a more recent study, Kang et al. studied the distribution of paclitaxel in the blend of poly(ethylene glycol)/poly(lactide-co-glycolic acid) (PEG/PLGA).85 As shown in Fig. 5A, CARS signals from the aromatic CH, CH2, and CH3 stretch vibration bands allowed selective imaging of paclitaxel, PEG, and PLGA.85 CARS imaging revealed that PTX preferentially partitioned into the PEG domains in the phase-segregated PLGA/PEG film (Figs. 5B–5D). This result provided the first visual evidence of PTX dissolution in the PEG matrix. The re-distribution of PTX inside the films during the release process was also visualized by time-lapse CARS microscopy.85 These studies demonstrate the potential of CARS microscopy in drug delivery formulation research.
Obesity and Cardiovascular Disease
With a strong signal from lipid bodies that are enriched in C–H bonds, CARS microscopy allowed selective imaging of adipocytes in vivo.28,29 By integrating TPEF and SHG imaging modalities into a CARS microscope, Le et al. demonstrated label-free molecular imaging of atheroma induced by a high-fat diet.62 As shown in Fig. 6, an atherosclerotic plaque in an iliac artery was identified by the strong CARS signal from lipids. The lipid core also showed a two-photon autofluorescence signal, but a lack of collagen fibrils based on SFG imaging results. The observed degradation of collagen fibrils is consistent with the increased activity of matrix metalloproteinases in an atheroma.86 Multimodal NLO imaging was also utilized to evaluate the impact of obesity on molecular composition of mammary tumors in a rat model,56 where obesity was found to increase the diameter of adipocytes by twofold and collagen fibril content by five-fold. These studies showcase the potential of NLO microscopy in understanding the impact of obesity on cancer and cardiovascular disease.
Skin and Cosmetics
The skin is the largest organ of the body and is an important target of the cosmetics industry. CARS imaging of skin in a hairless mouse ear has been demonstrated on a video-rate scanning microscope.28 The resonant CARS signal for the CH2 stretch vibration allowed the visualization of polygonal corneocytes in stratum corneum,87 sebaceous gland, and adipocytes within the depth of ~100 μm underneath the skin surface.28 In combination with TPEF imaging of elastin and SFG imaging of collagen,88 one should be able to visualize both cells and extracellular matrix in the skin tissue. These capabilities promise to provide new knowledge about skin and to help evaluate the efficacy of cosmetic agents.
OUTLOOK FOR COHERENT ANTI-STOKES RAMAN SCATTERING MICROSCOPY
With chemical selectivity, inherent 3D resolution, and capability for high-speed data acquisition, CARS microscopy is poised to become a potent tool for biological imaging. Additionally, the ability to couple CARS with other modalities such as TPEF, SFG, and electrophysiological recording provides a powerful multimodal platform to spread the range of applications. Although biomedical applications of CARS microscopy are still in their infancy, initial works have shown great potential for this technique in medicine, especially in understanding the roles of lipids in health and diseases. One area of high impact would be obesity and associated health risks. Obesity is reaching epidemic proportions worldwide and is associated with an increased risk of premature death89 and cardiovascular diseases.90 CARS is extremely sensitive to lipid bodies that are enriched in triglyceride.54 This capability can be used to study adipogenesis in single cells91 and dietary fat absorption in the intestine. Multimodal CARS microscopy has been used to evaluate the impact of obesity on mammary tumor stroma composition56 and to map the composition of atheroma induced by a high-fat diet.62 Another high-impact area lies in ex vivo and in vivo imaging of the nervous system, where the abundant lipids exist in the form of myelin sheath. The large and polarization-dependent CARS signal from the lamellar myelin membrane58 opens up a new avenue to the study of multiple sclerosis,84 a human demyelinating disease that affects 2.5 million people worldwide. Other important systems include polymer materials, drug delivery systems, skin, and cosmetic materials. Additionally, CARS microscopy has shown unique advantages for biophysical study of lipid membrane organization.53,81,92 Along with the above areas, in which CARS microscopy has shown great promise, the applications of CARS imaging are expected to continue to grow during the years to come. For instance, development of CARS probes would be attractive because they do not undergo photobleaching with proper laser power.93
The availability of an affordable, easy-to-operate multimodal CARS microscope will accelerate the development of new applications. A prohibitive factor in CARS imaging remains the cost and complexity associated with synchronizing two pulsed lasers and coupling them into a scanning microscope. The development of compact, cost-effective laser sources and alternatives to free-space coupling such as by fiber delivery94 will make the CARS imaging system more attractive to researchers in biological and biomedical areas. Meanwhile, there is still a vast amount of space for pushing the fundamental limits of the technique, e.g., increasing the sensitivity by surface-enhanced CARS,95 the spatial resolution by near-field CARS,96 and the imaging speed by wide-field CARS microscopy.97
One area in which CARS will prove greatly useful would be label-free in vivo imaging of live animals. Cellular structure and functions are best preserved in the in vivo environment. In vivo imaging also avoids trauma injuries to the tissue during the extraction process. As magnetic resonance imaging lacks single-cell resolution, in vivo CARS imaging will fill an important niche for study of disease processes. Although in vivo CARS imaging of skin tissues28 and peripheral nerves29 has been demonstrated, several technical hurdles such as minimizing surgery, compensating for animal motion, and increasing imaging depth have to be overcome. The development of new technologies such as miniaturized microscope objectives,98 video-rate scanning,28 and longer wavelength excitation21 will help to facilitate CARS imaging of live animals.
The demonstrated capability of CARS microscopy in ex vivo tissue imaging is expected to stir interests in the development of CARS endoscopy for in vivo diagnosis. The advantages of label-free chemical imaging of CARS endoscopy are abundantly obvious. First, label-free imaging overcomes the challenges of inefficient diffusion and non-specific targeting of exogenous fluorophores in the complex in vivo environment. Second, because CARS does not require electronic resonance, longer excitation wavelengths can be used to minimize multiphoton absorption induced photo-damage. Third, the intrinsic 3D sectioning capability of CARS imaging enables the analysis of spatial organization of bio-structures with molecular details. The preliminary work on fiber-based CARS imaging94,99 implied the possibilities of “driving” CARS into hospitals. CARS endoscopy holds great promise for rapid identification of malignancy in vivo. For diagnosis of cardiovascular diseases, it is anticipated that CARS endoscopy can be coupled with existing imaging modalities such as optical coherent tomography and intravascular ultrasound on the same catheter-based platform. This combination would create a multimodal imaging tool that is capable of two-dimensional (2D) mapping of morphology with optical coherent tomography and intravascular ultrasound and 3D sub-micrometric imaging of molecular composition with CARS endoscopy.
Acknowledgments
The author acknowledges Haifeng Wang, Li Li, Yan Fu, Terry B. Huff, Thuc Le, Eunah Kang, Han-Wei Wang, and Joshua Robinson for their contributions to the data presented in the paper. The author is grateful for the collaborations with professors Riyi Shi, Kinam Park, Michael Sturek, and Ignacio G. Camarillo in biomedical applications of CARS microscopy. This work was supported by NSF grant 0416785-MCB and NIH grants R21 EB004966, R01 EB007243, and R01 HL078715.
References
- 1.Gannaway JN, Sheppard CJR. Opt Quant Electron. 1978;10:435. [Google Scholar]
- 2.Hellwarth R, Christensen P. Opt Commun. 1974;12:318. [Google Scholar]
- 3.Duncan MD, Reintjes J, Manuccia TJ. Opt Lett. 1982;7:350. doi: 10.1364/ol.7.000350. [DOI] [PubMed] [Google Scholar]
- 4.Denk W, Strickler JH, Webb WW. Science. 1990;248:73. doi: 10.1126/science.2321027. [DOI] [PubMed] [Google Scholar]
- 5.Barad Y, Eisenberg H, Horowitz M, Silberberg Y. Appl Phys Lett. 1997;70:922. [Google Scholar]
- 6.Zumbusch A, Holtom GR, Xie XS. Phys Rev Lett. 1999;82:4142. [Google Scholar]
- 7.Zipfel WR, Williams RM, Webb WW. Nat Biotechnol. 2003;21:1369. doi: 10.1038/nbt899. [DOI] [PubMed] [Google Scholar]
- 8.Campagnola PJ, Loew LM. Nat Biotechnol. 2003;21:1356. doi: 10.1038/nbt894. [DOI] [PubMed] [Google Scholar]
- 9.Maker PD, Terhune RW. Phys Rev. 1965;137:A801. [Google Scholar]
- 10.Begley RF, Harvey AB, Byer RL. Appl Phys Lett. 1974;25:387. [Google Scholar]
- 11.Shen YR. The Principles of Nonlinear Optics. John Wiley and Sons; New York: 1984. [Google Scholar]
- 12.Levenson MD, Kano SS. Introduction to Nonlinear Laser Spectroscopy. Academic Press; San Diego: 1988. [Google Scholar]
- 13.Mukamel S. Principles of Nonlinear Optical Spectroscopy. Oxford University Press; New York: 1995. [Google Scholar]
- 14.Levenson MD, Song JJ. In: Coherent Nonlinear Optics: Recent Advances. Feld MS, Letokhov VS, editors. Springer-Verlag; Berlin: 1980. p. 293. [Google Scholar]
- 15.Holtom GR, Thrall BD, Chin BY, Wiley HS, Colson SD. Traffic. 2001;2:781. doi: 10.1034/j.1600-0854.2001.21106.x. [DOI] [PubMed] [Google Scholar]
- 16.Cheng JX, Xie XS. J Phys Chem B. 2004;108:827. [Google Scholar]
- 17.Volkmer A. J Phys D: Appl Phys. 2005;38:R59. [Google Scholar]
- 18.Rodriguez LG, Lockett SJ, Holtom GR. Cytometry. 2006;69A:779. doi: 10.1002/cyto.a.20299. [DOI] [PubMed] [Google Scholar]
- 19.Duncan MD. Opt Comm. 1984;50:307. [Google Scholar]
- 20.Duncan MD, Reintjes J, Manuccia TJ. Opt Eng. 1985;24:352. [Google Scholar]
- 21.Ganikhanov F, Carrasco S, Xie XS, Katz M, Seitz W, Kopf D. Opt Lett. 2006;31:1292. doi: 10.1364/ol.31.001292. [DOI] [PubMed] [Google Scholar]
- 22.Fu Y, Wang H, Shi R, Cheng JX. Opt Exp. 2006;14:3942. doi: 10.1364/oe.14.003942. [DOI] [PubMed] [Google Scholar]
- 23.Cheng JX, Volkmer A, Book LD, Xie XS. J Phys Chem B. 2001;105:1277. [Google Scholar]
- 24.Hashimoto M, Araki T, Kawata S. Opt Lett. 2000;25:1768. doi: 10.1364/ol.25.001768. [DOI] [PubMed] [Google Scholar]
- 25.Burkacky O, Zumbusch A, Brackmann C, Enejder A. Opt Lett. 2006;31:3656. doi: 10.1364/ol.31.003656. [DOI] [PubMed] [Google Scholar]
- 26.Volkmer A, Cheng JX, Xie XS. Phys Rev Lett. 2001;87:023901. [Google Scholar]
- 27.Cheng JX, Volkmer A, Xie XS. J Opt Soc Am B. 2002;19:1363. [Google Scholar]
- 28.Evans CL, Potma EO, Puoris’haag M, Côté D, Lin CP, Xie S. Proc Natl Acad Sci USA. 2005;102:16807. doi: 10.1073/pnas.0508282102. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Huff TB, Cheng JX. J Microsc. 2007;225:190. doi: 10.1111/j.1365-2818.2007.01729.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Cheng JX, Jia YK, Zheng G, Xie XS. Biophys J. 2002;83:502. doi: 10.1016/S0006-3495(02)75186-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Akhmanov SA, Bunkin AF, Ivanov SG, Koroteev NI. JETP Lett (Engl Transl) 1977;25:416. [Google Scholar]
- 32.Oudar JL, Smith RW, Shen YR. Appl Phys Lett. 1979;34:758. [Google Scholar]
- 33.Laubereau A, Kaiser W. Rev Mod Phys. 1978;50:607. [Google Scholar]
- 34.Hudson B, Hetherington W, Cramer S, Chabay I, Klauminzer GK. Proc Natl Acad Sci USA. 1976;73:3798. doi: 10.1073/pnas.73.11.3798. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Andrews JR, Hochstrasser RM, Trommsdorff HP. Chem Phys. 1981;62:87. [Google Scholar]
- 36.Yacoby Y, Fitzgibbon R, Lax B. J Appl Phys. 1980;51:3072. [Google Scholar]
- 37.Cheng JX, Book LD, Xie XS. Opt Lett. 2001;26:1341. doi: 10.1364/ol.26.001341. [DOI] [PubMed] [Google Scholar]
- 38.Cheng JX, Volkmer A, Book LD, Xie XS. J Phys Chem. 2002;106:8493. [Google Scholar]
- 39.Brakel R, Schneider FW. In: Advances in Non-linear Spectroscopy. Clark RJH, Hester RE, editors. John Wiley and Sons; New York: 1988. p. 149. [Google Scholar]
- 40.Volkmer A, Book LD, Xie XS. Appl Phys Lett. 2002;80:1505. [Google Scholar]
- 41.Dudovich N, Oron D, Silberberg Y. Nature (London) 2002;418:512. doi: 10.1038/nature00933. [DOI] [PubMed] [Google Scholar]
- 42.Oron D, Dudovich N, Silberberg Y. Phys Rev Lett. 2003;90:213902. doi: 10.1103/PhysRevLett.90.213902. [DOI] [PubMed] [Google Scholar]
- 43.Krishnamachari VV, Potma EO. J Opt Soc Am A. 2007;24:1138. doi: 10.1364/josaa.24.001138. [DOI] [PubMed] [Google Scholar]
- 44.Krishnamachari VV, Potma EO. Chem Phys, paper. 2007 in press. [Google Scholar]
- 45.Evans CL, Potma EO, Xie XS. Opt Lett. 2004;29:2923. doi: 10.1364/ol.29.002923. [DOI] [PubMed] [Google Scholar]
- 46.Marks DL, Boppart SA. Phys Rev Lett. 2004;92:123905. doi: 10.1103/PhysRevLett.92.123905. [DOI] [PubMed] [Google Scholar]
- 47.Marks DL, Vinegoni C, Bredfeldt JS, Boppart SA. Appl Phys Lett. 2004;85:5787. [Google Scholar]
- 48.Lim SH, Caster AG, Leone SR. Phys Rev A. 2005;72:041803. [Google Scholar]
- 49.Potma EO, Evans CL, Xie XS. Opt Lett. 2006;31:241. doi: 10.1364/ol.31.000241. [DOI] [PubMed] [Google Scholar]
- 50.Kee TW, Zhao HX, Cicerone MT. Opt Exp. 2006;14:3631. doi: 10.1364/oe.14.003631. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.von Vacano B, Buckup T, Motzkus M. Opt Lett. 2006;31:2495. doi: 10.1364/ol.31.002495. [DOI] [PubMed] [Google Scholar]
- 52.Lim SH, Caster AG, Nicolet O, Leone SR. J Phys Chem B. 2006;110:5196. doi: 10.1021/jp057493k. [DOI] [PubMed] [Google Scholar]
- 53.Li L, Wang H, Cheng JX. Biophys J. 2005;89:3480. doi: 10.1529/biophysj.105.065607. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Nan X, Cheng JX, Xie XS. J Lipid Res. 2003;40:2202. doi: 10.1194/jlr.D300022-JLR200. [DOI] [PubMed] [Google Scholar]
- 55.Nan X, Potma EO, Xie XS. Biophys J. 2006;91:728. doi: 10.1529/biophysj.105.074534. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Le TT, Rehrer CR, Huff TB, Nichols MB, Camarillo IG, Cheng JX. Mol Imaging. 2007 paper in press. [PMC free article] [PubMed] [Google Scholar]
- 57.Potma EO, Xie XS. J Raman Spectrosc. 2003;34:642. [Google Scholar]
- 58.Wang H, Fu Y, Zickmund P, Shi R, Cheng JX. Biophys J. 2005;89:581. doi: 10.1529/biophysj.105.061911. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Kang E, Wang H, Kwon IK, Robinson J, Park K, Cheng JX. Anal Chem. 2006;78:8036. doi: 10.1021/ac061218s. [DOI] [PubMed] [Google Scholar]
- 60.Ganikhanov F, Evans CL, Saar BG, Xie XS. Opt Lett. 2006;31:1872. doi: 10.1364/ol.31.001872. [DOI] [PubMed] [Google Scholar]
- 61.Fu Y, Wang H, Shi R, Cheng JX. Biophys J. 2007;92:3251. doi: 10.1529/biophysj.106.089011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Le TT, Langohr IM, Locker MJ, Sturek M, Cheng JX. J Biomed Opt. 2007 doi: 10.1117/1.2795437. paper in press. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Wurpel GWH, Schins JM, Müller M. Opt Lett. 2002;27:1093. doi: 10.1364/ol.27.001093. [DOI] [PubMed] [Google Scholar]
- 64.Hellerer T, Enejder AMK, Zumbusch A. Appl Phys Lett. 2004;85:25. [Google Scholar]
- 65.Knutsen KP, Johnson JC, Miller AE, Peterson PB, Saykally RJ. Chem Phys Lett. 2004;387:436. [Google Scholar]
- 66.Paulsen HN, Hilligsoe KM, Thogersen J, Keiding SR, Larsen JJ. Opt Lett. 2003;28:1123. doi: 10.1364/ol.28.001123. [DOI] [PubMed] [Google Scholar]
- 67.Kano H, Hamaguchi H. Appl Phys Lett. 2004;85:4298. [Google Scholar]
- 68.Kee TW, Cicerone MT. Opt Lett. 2004;29:2701. doi: 10.1364/ol.29.002701. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Petrov GI, Yakovlev VV. Opt Exp. 2005;13:1299. doi: 10.1364/opex.13.001299. [DOI] [PubMed] [Google Scholar]
- 70.Andresen ER, Paulsen HN, Birkedal V, Thøgersen J, Keiding SR. J Opt Soc Am B. 2005;22:1934. [Google Scholar]
- 71.Ogilvie JP, Beaurepaire E, Alexandrou A, Joffre M. Opt Lett. 2006;31:480. doi: 10.1364/ol.31.000480. [DOI] [PubMed] [Google Scholar]
- 72.Shimada R, Kano H, Hamaguchi H. Opt Lett. 2006;31:320. doi: 10.1364/ol.31.000320. [DOI] [PubMed] [Google Scholar]
- 73.Vartiainen EM, Rinia HA, Muller M, Bonn M. Opt Exp. 2006;14:3622. doi: 10.1364/oe.14.003622. [DOI] [PubMed] [Google Scholar]
- 74.Tsai TH, Tai SP, Lee WJ, Huang HY, Liao YH, Sun CK. Opt Exp. 2006;14:749. doi: 10.1364/opex.14.000749. [DOI] [PubMed] [Google Scholar]
- 75.Kennedy AP, Sutcliffe J, Cheng JX. Langmuir. 2005;21:6478. doi: 10.1021/la046820x. [DOI] [PubMed] [Google Scholar]
- 76.Cheng JX, Pautot S, Weitz DA, Xie XS. Proc Natl Acad Sci USA. 2003;100:9826. doi: 10.1073/pnas.1732202100. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Wurpel GWH, Müller M. Chem Phys Lett. 2006;425:336. [Google Scholar]
- 78.Wang H, Fu Y, Cheng JX. J Opt Soc Am B. 2007;24:544. [Google Scholar]
- 79.Edidin M. Annu Rev Biophys Biomol Struct. 2003;32:257. doi: 10.1146/annurev.biophys.32.110601.142439. [DOI] [PubMed] [Google Scholar]
- 80.Wurpel GWH, Schins JM, Müller M. J Phys Chem. 2004;2004:3400. [Google Scholar]
- 81.Potma EO, Xie XS. Chem Phys Chem. 2005;6:77. doi: 10.1002/cphc.200400390. [DOI] [PubMed] [Google Scholar]
- 82.Potma EO, Boeij WPD, MvHaastert PJ, Wiersma DA. Proc Natl Acad Sci USA. 2001;98:1577. doi: 10.1073/pnas.031575698. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Lazzarini RA, editor. Myelin Biology and its Disorders. Vol. 1 Elsevier Academic Press; San Diego, CA: 2004. [Google Scholar]
- 84.Fu Y, Wang H, Huff TB, Shi R, Cheng JX. J Neurosci Res. 2007 doi: 10.1002/jnr.21403. paper in press. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Kang E, Robinson J, Park K, Cheng JX. J Control Release paper. 2007 doi: 10.1016/j.jconrel.2007.05.007. in press. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86.Galis ZS. Circ Res. 2000;86:1. doi: 10.1161/01.res.86.1.1. [DOI] [PubMed] [Google Scholar]
- 87.Djaker N, Lenne PF, Marguet D, Colonna A, Hadjur C, Rigneault H. Nucl Instrum Methods Phys Res, Sect A. 2007;571:177. [Google Scholar]
- 88.Zipfel WR, Williams RM, Christie R, Nikitin AY, Hyman BT, Webb WW. Proc Natl Acad Sci USA. 2003;100:7075. doi: 10.1073/pnas.0832308100. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Adams KF, Schatzkin A, Harris TB, Kipnis V, Mouw T, Ballard-Barbash R, Hollenbeck A, Leitzmann MF. N Engl J Med. 2006;355:763. doi: 10.1056/NEJMoa055643. [DOI] [PubMed] [Google Scholar]
- 90.Van Gaal LF, Mertens IL, De Block CE. Nature (London) 2006;444:875. doi: 10.1038/nature05487. [DOI] [PubMed] [Google Scholar]
- 91.Yamaguchi T, Omatsu N, Morimoto E, Nakashima H, Ueno K, Tanaka T, Satouchi K, Hirose F, Osumi T. J Lipid Res. 2007;48:1078. doi: 10.1194/jlr.M600493-JLR200. [DOI] [PubMed] [Google Scholar]
- 92.Li L, Cheng JX. Biochemistry. 2006;45:11819. doi: 10.1021/bi060808h. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Tong L, Lu Y, Lee RJ, Cheng JX. J Phys Chem B. 2007 doi: 10.1021/jp073478z. paper in press. [DOI] [PubMed] [Google Scholar]
- 94.Wang H, Huff TB, Cheng JX. Opt Lett. 2006;31:1417. doi: 10.1364/ol.31.001417. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95.Koo TW, Chan S, Berlin AA. Opt Lett. 2005;30:1024. doi: 10.1364/ol.30.001024. [DOI] [PubMed] [Google Scholar]
- 96.Ichimura T, Hayazawa N, Hashimoto M, Inouye Y, Kawata S. Phys Rev Lett. 2004;92:220801. doi: 10.1103/PhysRevLett.92.220801. [DOI] [PubMed] [Google Scholar]
- 97.Heinrich C, Bernet S, Ritsch-Marte M. New J Phys. 2006;8:1. [Google Scholar]
- 98.Wang H, Huff TB, Fu Y, Jia KY, Cheng JX. Opt Lett. 2007 doi: 10.1364/ol.32.002212. paper in press. [DOI] [PubMed] [Google Scholar]
- 99.Légaré F, Evans CL, Ganikhanov F, Xie XS. Opt Exp. 2006;14:4427. doi: 10.1364/oe.14.004427. [DOI] [PubMed] [Google Scholar]