Abstract
We present a practical method for radioactivity distribution analysis in small-animal tumors and organs using positron emission tomography imaging with a calibrated source of known activity and size in the field of view. We reconstruct the imaged mouse together with a source under the same conditions, using an iterative method, Maximum Likelihood Expectation-Maximization with System Modeling, capable of delivering high resolution images. Corrections for the ratios of geometrical efficiencies, radioisotope decay in time and photon attenuation are included in the algorithm. We demonstrate reconstruction results for the amount of radioactivity within the scanned mouse in a sample study of osteolytic and osteoblastic bone metastasis from prostate cancer xenografts. Data acquisition was performed on the small-animal PET system which was tested with different radioactive sources, phantoms and animals to achieve high sensitivity and spatial resolution. Our method uses high resolution images to determine the volume of organ or tumor and the amount of their radioactivity, has the possibility of saving time, effort and the necessity to sacrifice animals. This method has utility for prognosis and quantitative analysis in small-animal cancer studies, and will enhance the assessment of characteristics of tumor growth, identifying metastases, and potentially determining the effectiveness of cancer treatment. The possible application for this technique could be useful for the organ radioactivity dosimetry studies.
1. Introduction
Complex clinical decisions on treatment are often guided by Positron Emission Tomography (PET) imaging principally using 18 FDG. PET is often used with other imaging techniques (by combining with CT or MRI) to obtain complementary information. Imaging with 18 FDG or other agents often requires quantitative measurements associated with the imaging data.
The factors that affect PET quantitation are resolution, photon attenuation and scattering, random coincidence rate, detector normalization, dead time and noise. It is very difficult to account for these factors for quantitative analysis of three-dimensional reconstructed radioactivity in tumors or mouse organs.
In most PET studies the Standardized Uptake Value (SUV) method is used to quantify tumor radioactivity. As the most widely used semi-quantitative parameter for tumor diagnosis, SUV determination involves measuring activity at a target site, with correction for injected dose, plasma glucose level, uptake period, body weight and, more important, correction for reconstruction method (DiChiro et al 1988, Keyes 1996, Thie et al 2000, Huang 2000, Truong et al 2004, Kok et al 2005, Popperl et al 2006). To eliminate the need for body-weight correction, SUV has been calculated on the basis of body weight: tissue concentration [MBq/g]/injected dose [MBq]/body weight [g] (see other SUV determinations in de Boer et al 2003). However, the SUV method does not correct for any inaccuracy in the measured dose, which may occur with injected dose extravasations, or with an elevated uptake elsewhere in the body. The accuracy of the SUV and the accuracy of relative change during treatment are not well documented and it might be a problem for diagnostic purposes in multicenter studies (Boellaard R et al 2004). Recent studies even find that SUV readings vary on different PET systems (Takahashi Y et al 2007) and Regions of Interest (ROI) can influence quantitative FDG-PET study results (Evilevich V et al 2007).
Thus, the microPET-R4 rodent scanner and ASIPro reconstruction software (both from CTI Concorde Microsystems Inc., Knoxville, TN) were used for semi-quantitative radioactivity analysis in two recent studies: to delineate the stages involved in the development of arthritis (Wipke et al 2004) and breast cancer metastasis (Liang et al 2005). In the first study quantitative analysis of ROI was performed over the selected mouse tissues and averaging the radioactive concentration over the contained voxels. In the second study a pixel ROI was outlined in the regions of increased FDG uptake, and after correction for radioactivity decay, the maximum SUV was semi-quantitatively calculated according to the method of Truong et al 2004.
A more sophisticated method to determine the maximum radioactivity concentration within a tumor or an organ was described in Zhang et al 2006. In this method, mean pixel values within the multiple ROI volumes were converted to µCi/ mL/ min by using a calibration constant (Wu et al 2005). Recently, an analytic semi-automated approach to calculate body distribution of PET tracers using co-registration a digital mouse phantom with small-animal images was proposed in Kesner et al 2006. The main goal in Domenico et al 2003 was to quantify the activity measured in a ROI within a reconstructed image of a small-animal and compare results with the ones derived from standard biodistribution methods: sacrificing the animal and putting each organ of interest in a calibrated gamma counter. Note that all PET devices are calibrated periodically for detector sensitivity using a calibrated source (generally a syringe filled with a known amount of radioisotope). From this scan, the number of counts per radioactivity detected by each detector pair is recorded. These numbers are used for subsequent scans to normalize the number of counts for detector efficiencies and to determine the amount of activity within the scanned object, but it not possible with high resolution to determine activity in small-animal tumors or organs.
We present a new practical method (Slavine and Antich ©2007) to determine radioactivity distribution in ROI from reconstructed PET images with a source of known activity and size in the field of view (FOV) in an example using osteolytic and osteoblastic bone metastasis from prostate cancer xenografts. Our method is different, more precise and yet simpler than that described above. It is based on a 3D reconstruction method capable of delivering high resolution images, has the possibility of saving time, effort and the necessity to sacrifice animals. The 3D high resolution reconstruction and radioactivity analysis can help to analyze the size and aggressiveness of the tumor, determine its growth in time, and the effectiveness of the treatment.
2. Biological objectives and experimental methods
The prostate cancer (CaP) is the second leading cause of cancer-related death among men (Gao et al 1997, Ghosh et al 2004). Prostate-specific antigen testing has been widely adopted for screening prostate cancer, which have a propensity to metastasize to bone (Wu et al 1998, Hall et al 2005, Foss et al 2005 and Guise et al 2006). Prostate cancer may cause osteolysis or abnormal new bone formation (Roodman 2004). As shown by histologic examination (Roudier et al 2003) the same prostate cancer patient often has evidence of osteolytic and osteoblastic disease. Therefore it is desirable to develop a noninvasive imaging technique to monitor individual tumors in individual patients.
To study both osteolytic and osteoblastic bone metastasis we used a mouse xenograft model (Wu et al 1998, Weerden 2000) and applied a multimodality imaging technique (Lin et al 2005) by combining our PET with CT, PET or MRI with our home built bioluminescent tomographic system (see details in Richer et al 2004, Dikmen et al 2005) or X-ray radiography. MR images were collected using a Varian Unity INOVA 4.7 T MR system equipped with actively shielded gradients. A spin echo sequence was used to obtain T1- and T2- weighted images of the control and tumor-bearing femurs. Note that these studies, in addition, have shown the feasibility and importance of multimodality imaging applied to investigations of tumor development. Each method can reveal different aspects of tumor development and together they have the potential to monitor tumor metastasis quantitatively, providing insight into tumor proliferation and differentiating the phenotypes of prostate cancer in mice.
We used the main results of these studies to confirm the ability of our method to achieve radioactivity distribution analysis in small-animal PET cancer studies.
3. Small-animal PET studies
3.1 Calibrated radioactivity source
In this practical method for radioactivity distribution analysis we used a 22Na radioactive source (half life is 2.6003 years, average positron energy is 0.215 MeV, positron decay ratio of 89.8%, see National Nuclear Data Center, 2004) produced by Amersham Buchler GmbH & Co KG (Braunschweig, Germany) and certified by Lloyd’s Register Quality Assurance (LRQA, Houston, TX). The active material is absorbed in a 1 mm diameter ion exchange bead, and sealed by ultrasonic welding between 0.5 mm polystyrene windows which are mounted in a plastic frame 23.5×11.0×2.0 mm3 volume. The active bead is visible and at the geometric center of the radioactive source. The nominal activity of the source on reference day May, 1 1998 was 370.0 ± 11.1 kBq. The reported calibration accuracy (overall uncertainty) of 3% was based on standard uncertainty multiplied by a coverage factor of 2, providing a level of confidence of approximately 95%. The photon emission and purity of the source were checked using a Ge (Li) detector. On the day of PET tumor imaging the source activity was about 67.0 ± 2.0 kBq.
3.2 Small-animal PET device
Positron emission tomography of tumor-bearing mice was performed on our home built small-animal PET device based on two detectors with adjustable separations for optimizing system sensitivity and field of view (see more details in Tsyganov et al 2006). Our PET imager uses the two-fold coincident detection of a single event in two orthogonal fibers to detect the location and the energy transferred at a point within the detector. A single Position-Sensitive Photomultiplier Tube Hamamatsu R-2486 is used to perform a single-ended readout of all fiber arrays oriented in each direction. The scanner, with an effective field of view of 100 × 100 mm2 , has vertical and horizontal bed motion. The point source sensitivity is about 12.6 cps/kBq at 90 mm inner diameter. The current data transfer rate is about 6 MB/s (~40 K events per second). The two detectors are mounted on a gantry with rotational and translational degrees of freedom which can be chosen to suit a particular imaging protocol. Measurement of spatial resolution - Full-Width at Half-Maximum (FWHM) - for the calibrated positron emitting source was done in this PET device and was found to be about 1.0 mm.
3.3 Animal preparation and measurements
All animal studies were conducted in compliance with guidelines set by the Institutional Animal Care and Use Committee of the University of Texas Southwestern Medical Center at Dallas. Male athymic nude mice at the age of 6–8 weeks weighing 20g – 25g (Harlan Inc. Indianapolis, IN), were housed in a barrier filter room in a sterilized environment and fed Purina rodent chow ad lib. About one week latter, the mice were used for tumor implantation. Two human CaP cell lines, PC-3 (osteolytic) and C4-2 (osteoblastic), were stably transfected with luciferase gene. An intraosseous CaP mouse xenograft model was established by injection of a single-cell suspension (106 cells, 20 µ l) of C4-2 or PC-3 cells into the bone marrow of the right femur using a 23-gauge syringe. The left femur served as contralateral control. During the studies, prostate-specific antigen levels were measured every week in C4-2 tumor-bearing mice.
Food and water were removed from the mouse cage 30 min before the PET study. The mice were injected with no-carrier-added 18 FDG (was obtained from Petnet Solution, Knoxville, TN), then first anesthetized with 2% isoflurane and placed near the center of the FOV of the PET device where the highest sensitivity and image resolution are obtained. The total initial activity for PC-3 injected mouse was 30.01 ± 0.60 MBq (mean ± SD) and 34.78 ± 0.70 MBq for C4-2 injected mouse. One hour after injection, when data acquisition was started, the radioactivity was 20.57 ± 0.41 MBq and 23.83 ± 0.48 MBq respectively. The Capintec CRC-15R dose calibrator (Capintec Inc., NJ) was used to measure activity levels with accuracy better than 2%. The mouse was kept under anesthesia with 1.5% isoflurane via a nose cone during the scan. The mouse in the prone position and the 22Na calibrated source were placed between two detectors at a separation of 206 mm.
The detectors were rotated to obtain a complete 3D experimental raw data set. About 2.5 million valid list-mode PET events were collected in a 30 min acquisition. During all static imaging scans the 22Na source was situated below the mouse at a distance of about 40 mm.
4. Iterative 3D image reconstruction method
Prior to performing the 3D image reconstruction, we calculated corrections for the ratios of geometrical efficiencies for different detector positions and took into account the radioisotope’s decay at the time of imaging.
Our reconstruction algorithm consists of two main steps: 3D reconstruction of the imaged mouse together with a calibrated source under the same conditions used to determine the characteristics of calibrated source (size and activity), geometry and volume of each ROI -followed by calculation of the amount of radioactivity in each ROI using the radioactivity calculation method.
The advantages and disadvantages of using iterative reconstruction methods to produce accurate high resolution images were recently described in the literature (see, for example Leahy et al 2000, Vandenberghe et al 2001, Reader et al 2002 and Qi et al 2006). Our image reconstruction technique is based on the MLEM method (Shepp and Vardi 1982) which was improved for the One-Pass List-Mode high resolution algorithm with System Matrix modeling (MLSM) (Reader 2002 et al, Zinchenko et al 2003, Antich et al 2005) to take into account most of the finite resolution effects such as positron range, photon acollinearity, intrinsic detector resolution, scatter and random events. The method is generally applicable to any design geometry, to single (SPECT), or a double (PET) photon emission tomography.
The standard MLEM algorithm based on the measured list-mode data iterates the unknown activity value of voxel j (image is assumed to be discretized into J voxels) for each iteration step k:
| (1) |
where and are the voxel values for the new and old image estimates, is the expected count in Line of Response (LOR) i for the intensity estimate , I is the number of all possible system LORs and aij represents the probability that an emission from voxel j will be detected along LOR i. The measured data now consist of a list of M LOR definitions with implicitly equal to 1 for each acquired LOR. The normalization factor includes all possible measurable LORs. This algorithm is extendible to a data subset implementation to achieve significant acceleration and will generate a sequence of image estimates for each of N subsets with equal numbers (M/N) of list-mode events. The final image update is taken as the result of 3D reconstruction.
In order to take into account finite resolution effects we need to decompose the matrix of probabilities A = (aij)I×J into a product of three new matrices A=WXH. W = (wii ) IxI is the diagonal matrix of weighted factors for geometric sensitivity correction accounting for detector rotation, and is necessary to compensate for a sensitivity loss with increasing perpendicular distance di from the axis of rotation (Clark et al 1984):
| (2) |
where L is the transverse length of the detector and S is the separation of the PET detectors.
X = (xij)IxJ is a matrix whose elements correspond to the intersection length of LOR i with the voxel j and H = (hi j )JxJ is a square matrix which accounts for finite resolution effects. Attenuation estimate was carried out in backprojection space using attenuation correction factors for each voxel j (Reader 1999):
| (3) |
where lij is the path length of LOR i through voxel j, Li is the path length of LOR i through the attenuating object and the µ -linear attenuation coefficient.
Now, using the new expression for the matrix A, we can rewrite equation (1) in the form:
| (4) |
The expression (4) can be rewritten in terms of J-dimensional vectors with an element-by-element multiplication between vectors:
| (5) |
here s is the sensitivity correction factors and ck contains the multiplicative image correction value and can be given by:
| (6) |
where the matrix H is multiplied by a J-dimensional vector to give a new one, FPi is an operator which forward projects vector Hnk along LOR i to give a scalar value and, the operator BPi back-projects the scalar value along LOR i into a three-dimensional image for kth -iteration.
If we neglect the scattering effects, the blurring component H of the system matrix can be represented as a set of shift-invariant kernels ρ . In this case resolution blurring will be invariant across the image. For resolution modeling we use a convolution technique, so that the MLSM algorithm becomes:
| (7) |
where ρ is the resolution kernel and can be chosen as a Gaussian function (or more complex if needed) with standard deviation σ as parameter, ⊗ - denotes convolution procedure.
The algorithm described above, even using subsets, and especially for high amounts of list-mode data, requires many more iterations to image converge. To significantly increase reconstruction time savings, the image deblurring procedure with double Compton scattering model was developed (see more details in Antich et al 2005).
This iterative reconstruction algorithm with system matrix modeling was tested on several phantoms and successfully used for 3D image reconstruction for mice and rats, either bearing tumors or affected by genetically induced bone diseases (Antich et al 2002, Zinchenko et al 2003, Jennewein et al 2004–2008, Antich et al 2005, Lin et al 2005, Lewis et al 2005, Tsyganov et al 2006).
5. Radioactivity distribution value calculation
In our PET imaging system two high sensitivity detectors simultaneously record data in list-mode format. The data for each of the eight ( 45° angle step) positions was combined for all rotation angles and used for reconstruction. The total field of view of 110×130×120 mm3 in our 3D image reconstruction was subdivided into voxels of 0.5 mm3 in size. About 2.5 million events were reconstructed into a 220×260×240 array. Resolution was modeled with a Gaussian function. The resolution parameter σ was 1.2 mm during the 30 iterations. Results from our 3D reconstruction of the mice injected with 18 FDG in the presence of a 22 Na source in the field of view for different bone metastasis from prostate cancer xenografts is shown on Figure 1– Figure 6. The unit of the histogram’s colorbar is the voxel intensity after final iteration.
Figure 1.
Sagittal profile of all axial slices after 3D reconstruction for 18 FDG injected mouse with PC-3 tumor combined with calibrated 22 Na source (a) and transversal profile of the central axial slice for 22 Na source (b). Source was situated below the mouse at a distance of about 40 mm at Z=81.51 mm. The voxel size is 0.5 mm3.
Figure 6.
Number of voxels as a function of its reconstructed intensity for 18 FDG injected tumor-bearing mice of right femur with PC-3 tumor (a), C4-2 tumor (b) and for comparison of the two types of tumors (c). The intensities curves are normalized to the maximum intensity of the right femur.
Figure 1–Figure 2 and Table 1 show our reconstruction results separately for source areas of the mouse. Calibrated source was situated below the mouse at a distance of about 40 mm at Z=81.51mm for mouse with PC-3 tumor (see Fig. 1a and Table 1) and at Z=81.58 mm for the mouse with C4-2 tumor (see Fig. 2a and Table 1). Figure 1b shows transversal (for Y and Z profiles see Table 1) profile of the central axial slices for 22 Na source after 3D MLSM reconstruction for the mouse with PC-3 tumor. Figure 2b and Table 1 show the same for the mouse with C4-2 tumor. Each profile was fitted with a Gaussian function, and the FWHM of the distribution is taken as the reconstructed source size. It can be seen that reconstructed spatial resolution (source size) for each calibrated source is in excellent agreement with the actual size. Correspondence of reconstructed position coordinates for both sources show agreement when close to each other (see Table 1). Peaks of the reconstructed source intensities also demonstrate close agreement with each other with a difference of less than 1.0% (see legend box in Fig. 1b and Fig. 2b). Accuracy of the source reconstruction characteristics is very important for the activity calculation procedure.
Figure 2.
Sagittal profile of all axial slices after 3D reconstruction for 18 FDG injected mouse with C4-2 tumor combined with calibrated 22 Na source (a) and transversal profile of the central axial slice for 22 Na source (b). Source was situated below the mouse at a distance of about 40 mm at Z=81.58 mm. The voxel size is 0.5 mm3.
Table 1.
Reconstructed source positions and FWHM for X, Y and Z profiles in 18 FDG scan with PC-3 and C4-2 tumor-bearing right mice femurs. Each profile was fitted with a Gaussian function, and the FWHM of the distribution is taken as the reconstructed source size.
| PC-3 | C4-2 | |||
|---|---|---|---|---|
| Source Coordinate | position [mm] | FWHM [mm] | position [mm] | FWHM [mm] |
| X | 72.98 | 1.009 | 73.07 | 0.998 |
| Y | 114.31 | 0.999 | 114.73 | 1.006 |
| Z | 81.51 | 1.010 | 81.58 | 1.012 |
Once the animal with the calibrated 22Na source in the field of view is reconstructed with high precision (see results on Fig. 3 and Fig. 4) we can calculate the tumor’s or organ’s volume. On the Figure 3 we indicate the PC-3 tumor location and source position and on the Figure 4 same for the mouse with C4-2 tumor. Figure 3 and Figure 4 also demonstrate the good quality of 3D MLSM organs reconstruction. For image manipulations (rotation, chosen volume determination size, etc.) we applied the interactive utility for visualizing data on a PC graphic display Physics Analysis Workstation (PAW 1995) based on a statistical analysis and histogram package (HBOOK 1998). Note, the algorithm based on PAW was successfully used for volume size quantification in our bioluminescence imaging investigations (Richer et al 2004, Slavine et al 2006).
Figure 3.
Coronal (a) and sagittal (b) slices through the 3D reconstruction for 18 FDG injected mouse with PC-3 tumor combined with calibrated 22 Na source (X=72.98 mm, Y=114.31 mm, Z=81.51 mm) in the field of view (about of 30 MBq of 18 FDG, 20 g nude mouse, 30 min acquisition time for 2.5×106 list-mode PET events). The voxel size is 0.5 mm3.
Figure 4.
Coronal (a) and sagittal (b) slices through the 3D reconstruction for 18 FDG injected mouse with C4-2 tumor combined with calibrated 22 Na source (X=73.07 mm, Y=114.73 mm, Z=81.58 mm) in the field of view (about of 35 MBq of 18 FDG, 25 g nude mouse, 30 min acquisition time for 2.5× 106 list-mode PET events). The voxel size is 0.5 mm3.
Now, after we reconstruct the intensity and size of the source and determine the geometry of interested ROI, we can calculate a radioactivity distribution value (RDV) for tumor and mouse organs. First we determine the number of LORs i (reconstructed intensity) through each voxel j (volume) of the tumor (organ) and the calibrated source volume. Radioactivity value of the tumor (organ) will be determined as a result of comparing both these values for tumor (organ) and source volumes.
6. Results
Figure 5a and Figure 5b demonstrate the transverse slices of the right femur tumor through the 3D reconstruction and evidently confirm the particular characteristics of osteolytic and osteoblastic cancer disease-damaged femur is visible for the PC-3 tumor (Fig. 5a) and abnormal new bone formation for the C4-2 tumor (Fig. 5b). The reconstructed PET results (Fig. 5–Fig. 6 and Table 2) strongly indicated a significantly elevated 18 FDG uptake in the right tumor-bearing femur compared to the left control in these mice. Our quantitative distribution activity analysis (see Table 2 and Fig. 6) also confirms that the PC-3 tumor was of a more aggressive nature than the C4-2 tumor and had a higher ratio of 18 FDG uptake between the right femur and the left control femur in the PC-3 tumor (2.61) than in the femur with C4-2 one (1.85).
Figure 5.
Transverse slices of the right femur tumor through the 3D reconstruction of the 18 FDG injected nude mouse and 22Na calibrated source in the field of view. Transaxial images demonstrated the damaged femur in the mouse with PC-3 tumor (a) and new bone branching for the femur with C4-2 tumor (b). Left femurs were left as controls. The voxel size is 0.5 mm3. Resolution was modeled with a 1.2 mm resolution kernel.
Table 2.
Reconstructed radioactivity in mice femurs with and without tumors and in other mouse organs calculated by using iterative 3D reconstruction algorithm and method for RDV calculation. On the day of PET tumor imaging the source activity was about 67.0 ± 2.0 kBq.
| Femur/Organ | Radioactivity value [MBq] | |
|---|---|---|
| PC-3 tumor | C4-2 tumor | |
| Total (t=0 h) | 30.01 ± 0.60 | 34.78 ± 0.70 |
| Total (t=1 h) | 20.57 ± 0.41 | 23.83 ± 0.48 |
| Total (recon) | 20.36 ± 0.65 | 23.47 ± 0.75 |
| Femur tumor | 0.60 ± 0.02 | 0.37 ± 0.01 |
| Femur Left | 0.23 ± 0.01 | 0.20 ± 0.01 |
| Heart | 1.16 ± 0.04 | 1.26 ± 0.04 |
| Tail | 2.55 ± 0.08 | 2.93 ± 0.18 |
| Head | 4.29 ± 0.14 | 4.49 ± 0.09 |
| Bladder* | 0.72 ± 0.02 | 0.76 ± 0.02 |
| Kidney Right | 0.22 ± 0.01 | 0.28 ± 0.01 |
| Kidney Left | 0.21 ± 0.01 | 0.27 ± 0.01 |
| Back | 2.08 ± 0.07 | 1.82 ± 0.06 |
| Other | 8.34 ± 0.27 | 11.09 ± 0.35 |
Relatively low activity in the bladder because we forced a decreased urine level in the bladder for these measurements
Note, in addition, our multimodality investigation results (Lin et al 2005) clearly showed that the PC-3 tumor grew aggressively over time. Both T1- and T2-weighted MRI images were used to differentiate bone and tumor from other soft tissues. The T2-weighted images exhibited the osteolytic nature of the PC-3 tumor and revealed that both PC-3 and C4-2 tumors progressed to the whole femur and disseminated to the surrounding tissues after five weeks p.c.i. X-ray radiography further demonstrated the damaged femur in mice bearing PC-3 tumors and new bone branching from the femur containing C4-2 tumor in close agreement with our 3D reconstruction results (Fig. 5).
7. Conclusion
In Table 2 and Figure 6 we compare our quantitative results for mice femurs and other organs calculated by using the iterative MLSM algorithm and RDV calculation method. The results of this example study of osteolytic and osteoblastic bone metastasis from prostate cancer xenografts have provided quantitative data, show the tomographic capabilities of our small-animal PET imaging system and clearly show perfect practice ability of this method for radioactivity distribution calculations in tumors and small-animal organs.
The radioactivity distribution value calculation method is applicable to complex geometries, can be used with any 3D high resolution reconstruction algorithm (Liu et al 2001 McColl et al 1990), and is amenable to fast imaging calculations (depending on number of voxels, iterations and amount of list-mode data from PET system). The average processing time per image update for MLSM reconstruction procedure was a few min (time dependent upon 3D convolution algorithm). The imaging procedure and reconstruction software were verified using a single source and several phantoms (Antich et al 2005). The number of iterations (image updates) in the 3D reconstruction procedure was chosen to obtain close agreement between calibrated source actual dimensions with reconstructed values - thus saving time, and effort and providing the opportunity for the reconstruction procedure to be semi-automated. By using this method we don’t need to sacrifice animals to calculate the radioactivity in mice organs.
The method to determine the radioactivity distribution value in tumors or organs proposed here has utility for prognosis and quantitation in small-animal cancer studies to better understand the effects of human disease in animal models (Lin et al 2005, Mason 2007, Jennewein et al 2008). As shown in Table 2 and Figure 6 this method could be used as an additional useful tool to the semi-quantitative SUV method or to any other method (Kesner et al 2006, Zhong et al 2008) to estimate radioactivity value in tumor or any mouse organs. The possible application for this technique could be useful for the organ radioactivity dosimetry studies (Smith T 1992, Virta et al 2008) that required before mass balance study (Argenti et al 2000, Beumer et al 2005) may be performed in humans.
In case of the future improvements we will extend the method to our SPECT small-animal studies (Soesbe et al 2007). The quantitative accuracy in PET/SPECT reconstruction is particularly poor for small tumors because of partial volume effect (Quarantelli 2004, Wells 2007). This effect arises because of the blurring induced by the imaging system’s point spread function (Barrett and Myers 2004), producing intra-voxel mixing of the activity arising from different functional tissue classes. In this case we have to include partial volume correction techniques in our image quantification method.
Acknowledgments
We are very grateful to Drs. Xiankai Sun, R. P. Mason, G. Arbique, J.Anderson, Jer-Tsong Hsieh, V. D. Kodibagkar, M.A. Lewis, R. McColl, O.Oz, R. Parkey, Zhengwang Zhang , A. Zinchenko and our student Mai Lin for their advice and the valuable discussions. We would like to thank Mr. S. Seliounine and Dr. E. Tsyganov for experimental assistance with small-animal PET measurements.
This work was supported in part by the NIH/National Cancer Institute U24 CA126608-SAIRP
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- Antich PP, Constantinescu A, et al. FDG Imaging of Lung Metastases in Copenhagen Rats on a Small-Animal PET System with Positron-Encoding Fiber Optic Detectors. J Nuclear Medicine. 2002;43(5):216. [Google Scholar]
- Antich PP, Parkey R, et al. Application of Expectation Maximization Algorithms for Image Resolution Improvement in a Small-Animal PET System. IEEE Trans on Nuclear Science. 2005;53(3):684–690. [Google Scholar]
- Argenti D, Jensen B, et al. A Mass Balance Study to evaluate the Biotransformation and Excretion of [14C]-Triamcinolone Acetonide following oral administration. J Clinical Pharmacology. 2000;40:770–780. doi: 10.1177/00912700022009413. [DOI] [PubMed] [Google Scholar]
- Barrett HH, Myers KJ. Foundations of Image Science. Noboken, New Jersey: Wiley&Sons Inc; 2004. [Google Scholar]
- Beumer JH, Rademaker-Lakhai JM, et al. Trabectedin (Yondelis™, formely ET-743), a Mass Balance Study in patients with Advanced Cancer. Inv New Drags. 2005;23(5):429–436. doi: 10.1007/s10637-005-2902-4. [DOI] [PubMed] [Google Scholar]
- Boellaard R, Nanda C, et al. Effects of Noise, Image Resolution, and ROI Definition on the Accuracy of Standard Uptake Values: a Simulation Study. J Nuclear Med. 2004;45(9):1519–1527. [PubMed] [Google Scholar]
- de Boer JR, Pruim J, et al. l-1-11C -Tyrosine PET in Patients with Laryngeal Carcinomas: Comparison of Standardized Uptake Value and Protein Synthesis Rate. J. Nuclear Medicine. 2002;44(3):341–346. [PubMed] [Google Scholar]
- Clack R, Townsend D, Jevons A. Increased sensitivity and filed of view for a rotating Positron Camera. Phys. Med. Biol. 1984;29(11):1421–1431. doi: 10.1088/0031-9155/29/11/010. [DOI] [PubMed] [Google Scholar]
- Di Chiro G, Brooks R. RPET quantitation: blessing and curse. J. Nuclear Med. 1988;29:1603–1604. [PubMed] [Google Scholar]
- Di Domenico G, Zavattini G, et al. YAP-(S) Small Animal Scanner: Quantitative Results. IEEE Trans on Nuclear Science. 2003;50(5):1351–1356. [Google Scholar]
- Dikmen ZG, Gellert G, Dogan P, et al. A new diagnostic system in cancer research: Bioluminescent Imaging. Turk. J Medical Science. 2005;35:65–70. [Google Scholar]
- Evilevich V, Eilber F, et al. Influence of ROI definition on the results of FDG-PET studies in patients with sarcomas. J Nuclear Medicine. 2007;48:145P. [Google Scholar]
- Foss CA, Mease RC, et al. Radio labeled Small-Molecule Ligands for Prostate Specific Membrane Antigen: In vivo Imaging in Experimental Models of Prostate Cancer. Clinical Cancer Research. 2005;11(11):4022–4028. doi: 10.1158/1078-0432.CCR-04-2690. [DOI] [PubMed] [Google Scholar]
- Gao X, Porter AT, Grignon DJ, et al. Diagnostic and Prognostic Markers for Human Prostate. Cancer Prostate. 1997;31:264–681. doi: 10.1002/(sici)1097-0045(19970601)31:4<264::aid-pros8>3.0.co;2-k. [DOI] [PubMed] [Google Scholar]
- Ghosh A, Heston WD. Tumor target Prostate Specific Membrane Antigen (PSMA) and its regulation in Prostate Cancer. J Cell Biochem. 2004;91:528–539. doi: 10.1002/jcb.10661. [DOI] [PubMed] [Google Scholar]
- Guise TA, Mohammad KS, et al. Basic mechanisms responsible for Osteolytic and Osteoblastic Bone Metastases. Clinical Cancer Research. 2006;12 20 Suppl:6213s–6216s. doi: 10.1158/1078-0432.CCR-06-1007. [DOI] [PubMed] [Google Scholar]
- Hall CL, Bafico A, et al. Prostate Cancer Cells Promote Osteoblastic Bone Metastases through Wnts. Cancer Research. 2005;65(17):7554–7560. doi: 10.1158/0008-5472.CAN-05-1317. [DOI] [PubMed] [Google Scholar]
- Sung-Cheng Huang. Anatomy of SUV. J Nuclear Medicine & Biology. 2000;27:643–646. doi: 10.1016/s0969-8051(00)00155-4. [DOI] [PubMed] [Google Scholar]
- HBOOK. Statistical Analysis and Hostogramming, CERN Program Library Y250. Geneva Switzerland: CERN; 1998. ( www.cern.ch/cernlib). [Google Scholar]
- Jennewein M, Constantinescu A, et al. Molecular imaging of the vascular targeting antibody Vatuximab@ in rat Postate Cancer. Europ J Nucl Med & Molec Imaging. 2004;31:264. [Google Scholar]
- Jennewein M, Constantinescu A, Bergner O, Slavine NV, et al. Radioactive Arsenic Isotopes: New Tools for the Imaging of Tumor Targeting Antibodies. J Molec Imaging & Biology. 2004;16(150):108–117. [Google Scholar]
- Jennewein M, Qaim SM, et al. A New Method for Radiochemical separation of Arsenic from Irradiated Germanium Oxide. Applied Radiation & Isotopes. 2005;63:343–351. doi: 10.1016/j.apradiso.2005.04.005. [DOI] [PubMed] [Google Scholar]
- Jennewein M, Lewis MA, Dhao Dawen, et al. Vascular Imaging of Solid tumors in rats with a Radioactive Arsenic-Labeled Antibody that bins exposed Phosphatidylserine. Clinical Cancer Research. 2008;14:1377–1385. doi: 10.1158/1078-0432.CCR-07-1516. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kesner AL, Dahlbom M, et al. Semiautomated analysis of Small-Animal PET data. J. Nuclear Medicine. 2006;47(7):1181–1186. [PubMed] [Google Scholar]
- Keyes JW., Jr SUV: standard uptake or silly useless value? J. Nuclear Medicine. 1996;36(10):1836–1839. [PubMed] [Google Scholar]
- Kok PJ, van Eerd JE, et al. Biodistribution and Imaging of FDG in rats with LS174T Carcinoma Xenografts and focal Escherichia coli Infection. J. Cancer Biotherapy & Radiopharmaceuticals. 2005;20(3):310–315. doi: 10.1089/cbr.2005.20.310. [DOI] [PubMed] [Google Scholar]
- Leahy RM, Qi J. Statistical approaches in quantitative positron emission Tomography. Statistic and Computing. 2000;10:147–165. [Google Scholar]
- Lewis M, Arbique G, Richer E, Slavine NV, et al. Small-Animal SPECT Imaging. Springer Science Media Inc.; 2005. Projection and Pinhole based data Acquisition for Small-Animal SPECT using Storage Phosphor Technology. Chapter 24; pp. 279–285. [Google Scholar]
- Liang Z, Yoon Y, Votaw J, et al. Silencing of CXCR4 blocks breast cancer Metastasis. Cancer Research. 2005;65(3):967–971. [PMC free article] [PubMed] [Google Scholar]
- Lin M, Wang Z, Kodibagkar VD, Slavine NV, et al. Multimodality Imaging Studies of Bone Matastasisis in an Intraosseous Prostate Cancer Xenograft Mouse Model. Soc. of Nuclear Medicine; 52nd Annual Meeting; Puerto-Rico. 2005. N651850. [Google Scholar]
- Lin X, Combat C, et al. Comparison of 3D Reconstruction with 3D OSEM and with FORE+OSEM for PET. IEEE Trans on Medical Imaging. 2001;20(8):804–814. doi: 10.1109/42.938248. [DOI] [PubMed] [Google Scholar]
- Mason R. Radiology. UT Southwestern at Dallas; 2007. UT Southwestern In Vivo Cancer Cellular and Imaging Program. http://cip.swmed.edu/ICMIC. [Google Scholar]
- McColl R, Faber T, Opperman R, et al. TR3: A Simple Multimodality Image Handling Environment; 10th conf. on Computer Applications to assist Radiology; 1990. pp. 457–462. [Google Scholar]
- National Nuclear Data Center. Brookhaven National Laboratory; 2004 www.nndc.bnl.gov/nudat2.
- PAW - Physics Analysis Workstation. Switzerland: CERN Program Library Q121 CERN Geneva; 1995. ( www.cern.ch/cernlib). [Google Scholar]
- Popperl G, Kreth FW, Herms J, et al. Analysis of 18 F-FET PET for grading of recurrent gliomas: is evaluation of uptake kinetics superior to standard methods. J Nuclear Medicine. 2006;47(3):393–403. [PubMed] [Google Scholar]
- Qi J, Leahy RM. Iterative reconstruction techniques in emission computed Tomography. Physics in Medicine & Biology. 2006;51:R541–R574. doi: 10.1088/0031-9155/51/15/R01. [DOI] [PubMed] [Google Scholar]
- Quarantelli M, Berkouk K, et al. Integrated Software for the Analysis of Brain PET/SPECT Studies with Partial Volume Effect Correction. J Nuclear Medicine. 2004;45:192–201. [PubMed] [Google Scholar]
- Reader AJ, Erlandsson K, et al. Attenuation and scatter correction of List-Mode Data Iterative and Analytic Image Reconstruction Algorithm for 3D PET systems. IEEE Trans on Nuclear Science. 1999;46(6):2218–2226. [Google Scholar]
- Reader AJ, Ally S, Bakatselos F, et al. One-Pass List-Mode EM Algorithm for High Resolution 3D PET Image Reconstruction into Large Arrays. IEEE Trans on Nuclear Science. 2002;49(3):693–699. [Google Scholar]
- Richer E, Slavine NV, Lewis M, et al. Three Dimensional Light Emission Tomography using multiple rotating CCD cameras. J. Molec. Imaging. 2004;3(3):229–230. [Google Scholar]
- Shepp LA, Vardi Y. Maximum Likelihood reconstruction for emission tomography. IEEE Trans on Medical Imaging. 1982;1:113–122. doi: 10.1109/TMI.1982.4307558. [DOI] [PubMed] [Google Scholar]
- Roodman GD. Mechanisms of Bone Metastasis. N Engl J Medicine. 2004;350:1655–1664. doi: 10.1056/NEJMra030831. [DOI] [PubMed] [Google Scholar]
- Roudier MP, Vesselle H, et al. Bone Histology at Autopsy and matched Bone Scintigraphy findings in Patients with Hormone Refractory Prostate Cancer: the effect of Bisphosphonate Therapy on Bone Sintigraphy Results. Clinic Exp Metastasis. 2003;20:171–180. doi: 10.1023/a:1022627421000. [DOI] [PubMed] [Google Scholar]
- Soesbe T, Lewis M, Richer E, Slavine NV, Antich PP. Development and Evaluation of an EMCCD based Gamma Camera. IEEE Trans on Nuclear Science. 2007;54(5):1516–1524. [Google Scholar]
- Slavine NV, Lewis MA, Richer PP. Medical Physics. 1. Vol. 33. 2006. Iterative reconstruction method for light emitting sources based on diffusion equation; pp. 61–68. [DOI] [PubMed] [Google Scholar]
- Slavine NV, Antich PP. Patent disclosure filed with University of Texas Southwestern Medical Center at Dallas in July 2007. 2007. Practical Method for Radioactivity Distribution Analysis in Small-animal PET Cancer Studies©. #UTSD-1999. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smith T. Comparison of two methods of quantitation in Human studies of Biodistribution and Radiation Dosimetry. Phys in Medicine and Biology. 1992;37:1065–1076. doi: 10.1088/0031-9155/37/5/004. [DOI] [PubMed] [Google Scholar]
- Takahashi Y, Otake H, et al. Variety of SUVs among different PET systems in 18F-FDG study. J. Nuclear Medicine. 2007;48:462P. [Google Scholar]
- Thie JA, Hubner KF, Smith GT. The diagnostic utility of the lognormal behavior of PET Standardized Uptake Values in Tumors. J Nuclear Medicine. 2000;41(10):1664–1672. [PubMed] [Google Scholar]
- Truong MT, Erasmus JJ, Munden RF. Focal FDG Uptake in mediastinal brown fat mimicking malignancy: a potential pitfall resolved on PET/CT. Am. J Roentgenol. 2004;183:1127–1132. doi: 10.2214/ajr.183.4.1831127. [DOI] [PubMed] [Google Scholar]
- Tsyganov E, Anderson J, Arbique G, Constantinescu A, et al. UT SWMC Small-Animal Positron Emission Imager. IEEE Trans on Nuclear Science. 2006;53(5):2591–2600. [Google Scholar]
- Vandenberghe S, Asseler YD, et al. Iterative Reconstruction Algorithms in Nuclear Medicine. Computerized Medical Imaging and Graphics. 2001;25:105–111. doi: 10.1016/s0895-6111(00)00060-4. [DOI] [PubMed] [Google Scholar]
- Weerden van WM, Romijn JC. Use of Nude Mouse Xenograft Models in Prostate Cancer Research. The Prostate. 2000;43:263–271. doi: 10.1002/1097-0045(20000601)43:4<263::aid-pros5>3.0.co;2-i. [DOI] [PubMed] [Google Scholar]
- Wells K, Chiverton J, et al. Quantifying the Partial Volume Effect in PET using Benford’s Low. IEEE Trans on Nuclear Science. 2007;54(5):1616–1625. [Google Scholar]
- Virta JR, Tovanen T, et al. 1-11C -Methyl-4-Piperidinyl–N–Butyrate Radiation Dosimetry in Humans by dynamic Organ Specific Evaluation. J Nuclear Medicine. 2008;49(3):347–352. doi: 10.2967/jnumed.107.047233. [DOI] [PubMed] [Google Scholar]
- Wipke BT, Zheng Wang, et al. Staging the Initiation of Autoantibody-Induced Arthritis: a Critical Role for Immune Complexes. J Immunology. 2004:7694–7702. doi: 10.4049/jimmunol.172.12.7694. [DOI] [PubMed] [Google Scholar]
- Wu TT, Sikes RA, et al. Establishing Human Prostate Cancer Cell Xenografts in Bone: Induction of Osteoblastic Reaction by Prostate-Specific Antigen producing Tumors in Athymic and SCID/bg Mice using LNCaP and Lineage-derived Metastatic Sublines. Int. J. Cancer. 1998;77:887–894. doi: 10.1002/(sici)1097-0215(19980911)77:6<887::aid-ijc15>3.0.co;2-z. [DOI] [PubMed] [Google Scholar]
- Wu Y, Zhang X, et al. Micro PET Imaging of Glioma Integrin αvβ3 Expression using 64Cu - labeled Tetrametric RGD Peptide. J Nuclear Medicine. 2005;46(10):1707–1718. [PubMed] [Google Scholar]
- Zhang X, Xiong Z, Wu Y, Cai W, et al. Quantitative PET imaging of tumor Integrin αvβ3 Expression with 18 F-FRGD2. J Nuclear Medicine. 2006;47(1):113–121. [PMC free article] [PubMed] [Google Scholar]
- Zhong H, Weiss E, Siebers JV. Assessment of Dose Reconstruction Errors in Image-guided Radiation therapy. Phys in Medicine & Biology. 2008;53:719–736. doi: 10.1088/0031-9155/53/3/013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zinchenko A, Tsyganov E, Slavine NV, et al. Expectation Maximization Algorithm with Resolution Deconvolution for 3D Image Reconstruction in Small-Animal PET Imager. J Nuclear Medicine. 2003;44(5):165–166. [Google Scholar]













