Abstract
The performance of the Septi-Chek AFB system was compared with that of the BACTEC radiometric system and that of Lowenstein-Jensen agar slants (LJ) for detection of mycobacteria in clinical specimens. A total of 642 specimens were cultured; 61 (9.5%) yielded mycobacteria. Mycobacterium tuberculosis (34 isolates) and Mycobacterium avium complex (25 isolates) were the predominant species isolated. Of the 61 culture-positive specimens, 30 were smear positive and 31 were smear negative. Overall, 95% of the positive specimens were detected by Septi-Chek and BACTEC (100% of M. tuberculosis isolates) and 75% by LJ (82% of M. tuberculosis isolates). The mean times to detection were 15 days for BACTEC, 23 days for Septi-Chek, and 27 days for LJ. Of the 30 smear-positive specimens, 100% were recovered by Septi-Chek and BACTEC and 90% were recovered by LJ. Of the 31 smear-negative specimens, 90% were detected by Septi-Chek and BACTEC and 61% were detected by LJ. The Septi-Chek and BACTEC systems are superior to the conventional (LJ) mycobacterial culture method. Although Septi-Chek requires more time for the detection of mycobacteria than BACTEC, it is comparable in terms of overall recovery.
Full text
PDFSelected References
These references are in PubMed. This may not be the complete list of references from this article.
- Abe C., Hosojima S., Fukasawa Y., Kazumi Y., Takahashi M., Hirano K., Mori T. Comparison of MB-Check, BACTEC, and egg-based media for recovery of mycobacteria. J Clin Microbiol. 1992 Apr;30(4):878–881. doi: 10.1128/jcm.30.4.878-881.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Anargyros P., Astill D. S., Lim I. S. Comparison of improved BACTEC and Lowenstein-Jensen media for culture of mycobacteria from clinical specimens. J Clin Microbiol. 1990 Jun;28(6):1288–1291. doi: 10.1128/jcm.28.6.1288-1291.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]
- D'Amato R. F., Isenberg H. D., Hochstein L., Mastellone A. J., Alperstein P. Evaluation of the Roche Septi-Chek AFB system for recovery of mycobacteria. J Clin Microbiol. 1991 Dec;29(12):2906–2908. doi: 10.1128/jcm.29.12.2906-2908.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Giger T., Burkardt H. J. Evaluation of a new biphasic culture system for the recovery of mycobacteria. Eur J Clin Microbiol Infect Dis. 1990 Jun;9(6):428–431. doi: 10.1007/BF01979476. [DOI] [PubMed] [Google Scholar]
- Isenberg H. D., D'Amato R. F., Heifets L., Murray P. R., Scardamaglia M., Jacobs M. C., Alperstein P., Niles A. Collaborative feasibility study of a biphasic system (Roche Septi-Chek AFB) for rapid detection and isolation of mycobacteria. J Clin Microbiol. 1991 Aug;29(8):1719–1722. doi: 10.1128/jcm.29.8.1719-1722.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Piersimoni C., Morbiducci V., De Sio G., Scalise G. Comparative evaluation of the MB-Check system for recovery of mycobacteria from clinical specimens. Eur J Clin Microbiol Infect Dis. 1992 Dec;11(12):1174–1177. doi: 10.1007/BF01961140. [DOI] [PubMed] [Google Scholar]
- Whittier P. S., Westfall K., Setterquist S., Hopfer R. L. Evaluation of the Septi-Chek AFB system in the recovery of mycobacteria. Eur J Clin Microbiol Infect Dis. 1992 Oct;11(10):915–918. doi: 10.1007/BF01962373. [DOI] [PubMed] [Google Scholar]