Abstract
Antimicrobial elution disks containing amoxicillin-clavulanic acid (Augmentin), cefotetan, ciprofloxacin, or norfloxacin were tested in the Avantage automated susceptibility test system. Performance was compared against an agar diffusion procedure in a three-site collaborative study. Results of 1,500 comparison with amoxicillin-clavulanic acid showed a full accord (agreement of both systems) of 93.6% and an essential accord (agreement excluding minor discrepancies) of 97.6%. Results for cefotetan showed a full accord of 95.1% and an essential accord of 98.3% by the two methods. Results for both ciprofloxacin and norfloxacin were in full accord for more than 98% of tests with gram-negative bacilli and staphylococci, but tests with enterococci gave 38 and 26.1% minor discrepancies (the result of one method was resistant or susceptible and the result of the other method was intermediate), respectively. The results indicated that the Avantage test system is accurate and reliable and provides appropriate determination of bacterial susceptibility with the four antibiotics tested.
Full text
PDF


Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Barnes W. G., Green L. R., Talley R. L. Clinical evaluation of automated antibiotic susceptibility testing with the MS-2 system. J Clin Microbiol. 1980 Oct;12(4):527–532. doi: 10.1128/jcm.12.4.527-532.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Barry A. L., Fass R. J., Anhalt J. P., Neu H. C., Thornsberry C., Tilton R. C., Painter B. G., Washington J. A., 2nd Ciprofloxacin disk susceptibility tests: interpretive zone size standards for 5-microgram disks. J Clin Microbiol. 1985 Jun;21(6):880–883. doi: 10.1128/jcm.21.6.880-883.1985. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Henry D., Kunzer L., Ngui-Yen J., Smith J. Comparative evaluation of four systems for determining susceptibility of gram-positive organisms. J Clin Microbiol. 1986 Apr;23(4):718–724. doi: 10.1128/jcm.23.4.718-724.1986. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kelly M. T., Latimer J. M., Balfour L. C. Comparison of three automated systems for antimicrobial susceptibility testing of gram-negative bacilli. J Clin Microbiol. 1982 May;15(5):902–905. doi: 10.1128/jcm.15.5.902-905.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nolte F. S., Contestable P. B., Lincalis D., Punsalang A., Jr Rapid, direct antibiotic susceptibility testing of blood culture isolates using the Abbott Avantage system. Am J Clin Pathol. 1986 Nov;86(5):665–669. doi: 10.1093/ajcp/86.5.665. [DOI] [PubMed] [Google Scholar]
- Thornsberry C., Anhalt J. P., Washington J. A., 2nd, McCarthy L. R., Schoenknecht F. D., Sherris J. C., Spencer H. J. Clinical laboratory evaluation of the Abbott MS-2 automated antimicrobial susceptibility testing system: report of a collaborative study. J Clin Microbiol. 1980 Sep;12(3):375–390. doi: 10.1128/jcm.12.3.375-390.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]