Abstract
An indirect flat panel imager (FPI) with programmable avalanche gain and field emitter array (FEA) readout is being investigated for low-dose and high resolution x-ray imaging. It is made by optically coupling a structured x-ray scintillator, e.g., thallium (Tl) doped cesium iodide (CsI), to an amorphous selenium (a-Se) avalanche photoconductor called high-gain avalanche rushing amorphous photoconductor (HARP). The charge image created by the scintillator∕HARP (SHARP) combination is read out by the electron beams emitted from the FEA. The proposed detector is called scintillator avalanche photoconductor with high resolution emitter readout (SAPHIRE). The programmable avalanche gain of HARP can improve the low dose performance of indirect FPI while the FEA can be made with pixel sizes down to 50 μm. Because of the avalanche gain, a high resolution type of CsI (Tl), which has not been widely used in indirect FPI due to its lower light output, can be used to improve the high spatial frequency performance. The purpose of the present article is to investigate the factors affecting the spatial resolution of SAPHIRE. Since the resolution performance of the SHARP combination has been well studied, the focus of the present work is on the inherent resolution of the FEA readout method. The lateral spread of the electron beam emitted from a 50 μm×50 μm pixel FEA was investigated with two different electron-optical designs: mesh-electrode-only and electrostatic focusing. Our results showed that electrostatic focusing can limit the lateral spread of electron beams to within the pixel size of down to 50 μm. Since electrostatic focusing is essentially independent of signal intensity, it will provide excellent spatial uniformity.
Keywords: flat-panel detectors, indirect detection, cesium iodide, amorphous selenium, avalanche gain, field emitter array, high resolution, x-ray imaging, mammography, fluoroscopy
INTRODUCTION
Existing active matrix flat-panel imagers (AMFPI) use a two-dimensional array of thin film transistors (TFT) to read out a charge image generated by an x-ray image sensor, either an x-ray photoconductor or a scintillator coupled with discrete photodiodes for the direct and indirect conversion methods, respectively. While AMFPI provides better image quality than traditional screen films and computed radiography,1, 2 its imaging performance has not yet reached the optimum. Further improvements, especially in spatial resolution and low dose performance, are desirable. Existing AMFPIs have two main limitations: (1) the electronic noise degrades the imaging performance behind dense tissues in low dose application. It has been shown that the detective quantum efficiency (DQE) of mammography AMFPIs at ∼1 mR can be less than 50% of that at 10 mR,3 i.e., the detector is not x-ray quantum noise limited. This problem will be worse with a decrease in pixel size or x-ray dose, such as that used in breast tomosynthesis. (2) The pixel size currently used in mammography AMFPI ranges from 70 to 100 μm. With further decrease in pixel size, the fill factor decreases, resulting in DQE degradation due to noise aliasing (direct FPI) and deleterious effect of electronic noise (both direct and indirect FPI).4, 5 On the other hand, a smaller pixel size may be desirable from a clinical point of view. There is still an ongoing debate on what pixel size is optimal for the detection of calcifications. Studies have shown that del smaller than 100 μm provides better detection and characterization of calcifications.6, 7 Therefore, detector technologies capable of smaller pixel size without comprise in DQE is highly valuable for exploring the optimal pixel size for mammography.
To improve the detector low dose performance with high resolution, we propose a new concept of indirect conversion flat-panel imager with avalanche gain and field emitter array (FEA) readout, which is referred to as scintillator avalanche photoconductor with high resolution emitter readout (SAPHIRE). The concept of SAPHIRE is shown with the schematic drawing in Fig. 1. It consists of a thallium (Tl) doped needle-structured cesium iodide (CsI) scintillator optically coupled (e.g., through fiber optics) to a uniform thin layer (4−25 μm) of avalanche amorphous selenium (a-Se) photoconductor called high-gain avalanche rushing amorphous photoconductor (HARP). The optical photons emitted from the scintillator transmit through a transparent indium tin oxide (ITO) bias electrode of HARP and generate electron-hole pairs near the top interface of the HARP layer. By applying a positive voltage to the ITO electrode through a resistor, holes move toward the bottom (free) surface of the HARP and experience avalanche multiplication under an electric field strength of ESe>70 V∕μm,8, 9 which is an order of magnitude higher than that typically used in direct a-Se x-ray detector. The holes form an amplified charge image at the bottom surface of HARP and are readout with the electron beams generated by a two-dimensional FEA, which is placed at a short distance, e.g., 1 mm, below the scintillator-HARP (SHARP) structure. A mesh electrode, biased with positive potential, is inserted halfway between the FEA and HARP target to minimize the lateral spread of electron beams before they land on the free surface of the HARP target. The amount of charge required to return the bottom surface of the HARP target to ground (cathode) potential is measured by an amplifier connected (through alternating current coupling) to the ITO electrode to form the output signal.
Figure 1.
Schematics showing the concept of the proposed detector SAPHIRE: cross-sectional view showing the operating principles.
The concept of electron beam readout is similar to that used in optical and x-ray vidicons,10 except that the FEA is a compact, two-dimensional source of electron beams, allowing the construction of the detector in the form of a FPI. As an emerging technology for large area flat-panel field emission displays (FEDs),11 FEA has the potential to provide a smaller pixel size than that achieved with the TFT readout. The field emitter (FE) tips (with spacing of ∼1 μm) are connected to base electrodes arranged in columns, and rows of gate electrode are used to control the field emission. The overlapping area between the base and gate electrodes defines the pixel size, and sufficient FE tips can be included in del=50 μm to provide sufficient emission current to produce the wide dynamic range required in medical images. Each pixel is addressed by applying a forward bias between the gate and the base. This driving scheme is very similar to that used in passive matrix liquid crystal displays. The main disadvantage of the passive driving scheme is the pulse delay due to load resistance and capacitance of each line. To alleviate this problem, active matrix FEA has recently been developed, where each pixel is addressed through a transistor.12, 13
Compared to existing AMFPI, SAPHIRE has the following advantages: (1) Programmable avalanche gain gav ensures a wide dynamic range. By increasing the electric field ESe, high gav can be applied at low dose applications (e.g., fluoroscopy or tomosynthesis) to achieve x-ray quantum noise limited performance, while gav is turned off at high dose (e.g., radiography).14 At each gav setting, the signal first increases linearly as a function of radiation exposure. As the exposure exceeds the range that the gav is selected for, the image charge accumulated on the bottom free surface lowers the potential drop across the HARP layer, and causes gav to gradually decrease.15 This ensures that HARP continues to respond to radiation without complete saturation, resulting in a wide dynamic range. (2) FEA may provide smaller pixel size than is possible with TFTs. (3) A high resolution (HR) type of CsI can be used because of the additional gain provided by HARP. HR type CsI has less Lubberts effect, i.e., depth dependent blur, which is the main source of DQE degradation at high spatial frequencies.16, 17 Despite its resolution advantages, HR CsI has not been used widely in commercial indirect FPI due to its lower conversion gain, which makes the detector more susceptible to electronic noise at low exposure levels.5, 18
In this present article, we focus our investigation on the factors affecting the spatial resolution of SAPHIRE. Since breast imaging is the application that requires the highest spatial resolution, we will use a hypothetical mammography detector design as an example for our discussion, although SAPHIRE can be used in a general purpose R∕F detector as well. In Sec. 2, we provide a brief review of the principles of field emission and the electron beam readout method, especially the parallel electron beam readout proposed for SAPHIRE. In Sec. 3 we present our theoretical and simulation methods for investigating the spatial resolution of SAPHIRE, focusing on the factors related to the FEA readout. Two different electron-optical designs for focusing the electron beams are included in our investigation: (1) a basic design incorporating only a mesh electrode, such as that shown in Fig. 1, and (2) electrostatic focusing in addition to the mesh electrode. In Sec. 4, the results for different electron-optical designs are presented and compared from which we draw conclusions about the desired methods for electron beam focusing in SAPHIRE. The properties of lag and noise are also under investigation, they will be submitted in a separate article.
THEORY AND BACKGROUND
The FEA is a practical vacuum microelectronic device built to nanometer tolerance. Several technologies have been invented for manufacturing FEA for flat-panel display applications. They include surface-conduction electron emitter,19 carbon nanotube,20 metal–insulator–metal emitter as ballistic electron surface emitting device,21, 22 metal–insulator–semiconductor in high-efficiency electron emission devices,23 and Spindt-type field emitters.24 The characteristics of different FEA technologies vary in the angular distribution and intensity of the emitted electron beams. In FED, the anode (phosphor) is usually biased with a positive potential at tens of thousands of volts, which accelerate the electrons to high speed and minimize the time the electrons take to reach the anode. As a result, the lateral spread of the electron beam, which is due to the lateral velocity component of the electrons emitted with an oblique angle, is negligible compared to the pixel size. However, in an image sensor, the potential on the bottom surface of the photoconductive target is proportional to the image signal and is typically less than a few tens of volts. Therefore, additional electron-optical focusing methods need to be developed to ensure that the electrons can reach the target without considerable lateral spread. Furthermore, the intensity of the electron beam has to be sufficient for reading out the highest signal current generated in the HARP target.
Compared to the other types of FEA, Spindt-type field emitters have higher emission intensity and narrower angular distribution. They have also proven to be robust and stable over time.25 Prototype Spindt-type FEDs with diagonal dimensions of 20.3 and 28.7 cm have been presented,26, 27 which are approaching the imager size required for medical FPI. For the remainder of this article, we will use the typical characteristics of Spindt-type FEA as an example for our investigation. The most commonly used material for the Spindt-type FE tips is molybdenum (Mo). The FEA is enclosed in high vacuum (10−9 Torr) during operation. To turn on field emission, the cone-shaped Mo cathode is biased at ground potential and a positive bias, Vg (40−100 V), is applied to the gate electrode. This bias condition results in a very high EFE around the emitter tip due to its small hemispherical radius of 300 Å (Refs. 24, 25) and causes field emission. Each pixel consists of a matrix of FE tips and different pixels of a FEA are addressed by orthogonal base and gate lines, i.e., a passive driving scheme. A prototype FEA with del=50 μm with 17×17 tips in each image pixel has been used in a 1 in. diameter optical HARP FEA image sensor.28
The size of the FPI required for x-ray imaging is much larger than that for optical imaging, e.g., 20 cm×25 cm for mammography. A pixel size of 50 μm would result in 4000×5000 pixels. This necessitates division of the ITO signal∕bias electrode into multiple strips for two reasons: (1) One large ITO electrode would result in a large input load capacitance for the amplifier, which leads to increased electronic noise; (2) The large passive load (capacitance and resistance) of each gate and base line of a large area FEA results in driving pulse delay, which requires that each pixel be turned on for at least 0.16 μs. A FEA matrix of 4000×5000 pixels would require 3.2 s to read out pixel-by-pixel. If the ITO electrode is divided into Ns stripes and each connected to a charge amplifier, as shown in Fig. 2, Ns pixels can be turned on simultaneously for parallel readout and increase the readout speed by Ns times. The major benefits of parallel beam readout are the decrease in readout lag and noise, which will be discussed in separate publications. For the discussion in the present article, the parallel beam readout only has an impact on the pixel turn on time (tp), which affects the pixel aperture function (to be discussed in Secs. 3C, 4C).
Figure 2.
3D schematic view of the parallel beam readout method to show the simultaneous emission of electron beams from several pixels of the FEA, one for each ITO strip. The mesh electrode and CsI are removed from the SAPHIRE structure for clarity of illustration.
METHODS
There are several factors affecting the spatial resolution of SAPHIRE: (1) the inherent resolution of the optically coupled SHARP combination; (2) the FEA pixel size; and (3) the lateral spread of the electron beams. The first factor is dominated by the blur in the structured CsI scintillator, since the photon detection and avalanche process in HARP has been shown to have no blur in high definition optical HARP camera that operates with an effective pixel size of 10−20 μm.29, 30 The spatial resolution properties of CsI has been investigated previously14, 16, 31 and the results for a 150 μm thick HR type CsI will be used here in comparison with the other factors. This measured presampling MTF included the inherent image blur of the columnar CsI and the optical properties of the fiber optic faceplate, on which the CsI was deposited. Since the diameters of the CsI columns and the optical fiber are 5−10 μm, their effects (aperture function and sampling) on the MTF are negligible, and the measured MTF is regarded shift invariant. In this section, we will describe the methods for investigating the lateral spread of electrons emitted from the FEA and its effect on the aperture function of the readout method.
Lateral spread of electron beam in FEA readout
The geometric arrangement of FE tips on each pixel of the FEA determines the area of emission. However, the spatial resolution of the FEA readout is determined by the lateral extent and intensity of the electron beam when it reaches the bottom surface of the HARP target. As shown in Fig. 3, the electrons emitted from each FE tip have a finite angular range of up to 40°−50°.32 The lateral travel distance of an electron is equal to the integration of the lateral component of its velocity vx over the time tgt it takes to travel from the gate electrode to the target. Since the value of Vt on the bottom surface of HARP due to x-ray exposure is on the order of several volts, it is necessary to insert a mesh electrode with potential Vm of several hundred volts between the HARP target and the FEA. As shown in Fig. 3, the mesh electrode will accelerate the axial velocity component (vz) of the electrons thus shorten tgt. After the electrons transmit through the mesh electrode, they decelerate due to the reversed electric field, and only those electrons with sufficient initial value of vz can reach the target. The other electrons decelerate to vz=0 before reaching the target, and return to the mesh electrode.
Figure 3.
Schematic diagram showing the lateral spread of electron beams emitted from the FEA.
To reduce the lateral spread of electrons, different electron-optical focusing designs have been investigated to reduce either vx or tgt. In this article two different electron-optical designs were investigated: (1) mesh-electrode-only and (2) electrostatic focusing in addition to the mesh electrode. Beside these two designs, a magnetic focusing design has been implemented for a 1 in. HARP-FEA sensor.28 However, due to the difficulty in maintaining a uniform magnetic field over a large area, it may not be practical for SAPHIRE and will be not discussed in this article. For each electron-optical design, we used a two-step approach to determine the lateral extent and intensity of electron beams: (1) determine the lateral travel distance of a single electron emitted from a single FE tip; and (2) compute the spread and intensity of the electron beam emitted from one pixel of the FEA, which includes a two-dimensional array of tips. These two steps will be described in this subsection (Sec. 3A) and the next subsection (Sec. 3B), respectively.
Mesh-electrode-only
As described above, the simplest method to reduce tgt, and hence, lateral spread is to insert a mesh electrode halfway between the HARP target and the FEA, as shown in Fig. 3. If we assume that all electrons have the same initial energy E=qVg after emission from the gate electrode, the lateral spread can be derived analytically as shown in the Appendix0 and is given by33
| (1) |
where Lgm and Lmt are the gate-mesh and mesh-target distances, respectively. Only the electrons with smaller θ, and hence, higher initial vz can reach the target. So there exists a critical emission angle θc, within which the emitted electrons can reach the target. θc is determined by Vt and Vg of the FEA,33
| (2) |
The maximum value of lateral spread LSmax occurs with critical angle θC and can be obtained by substituting θ in Eq. 1 with θC in Eq. 2,
| (3) |
Equation 3 shows that LSmax depends on the geometry and the operating conditions of the SAPHIRE, e.g., Lgm, Lmt, Vm, and Vg, as well as on the x-ray exposure (i.e., Vt).
Electrostatic focusing
Electrostatic focusing is feasible for large area SAPHIRE. The concept of electrostatic focusing proposed for SAPHIRE is shown in Fig. 4. An additional (focusing) electrode is added on top of the gate electrode for each FE tip, and this structure was referred to as the double-gated Spindt-type emitter.25 Both electrodes are made of Mo through photolithography and are separated by an insulating layer (e.g., SiO2). To deflect the electrons emitted with large θ back to the axial direction, a focusing electrode potential VL that is much lower than gate potential Vg is applied. The choice of VL depends on the geometry of the double-gated tip.
Figure 4.
Cross-sectional view showing the structure of a double-gated Spindt-type emitter with focusing electrodes, which defect the electrons with large emission angle to axial direction.
The lateral spread of electrons with electrostatic focusing cannot be determined analytically. Instead, the electron trajectory was simulated using the finite element method (FEM). A commercial FEM simulation software (COMSOL Multiphsysics®) was used to solve for the electric field distribution between the HARP target and the FEA. Since the size of each FE tip (∼1 μm) is three orders of magnitude smaller than the distance between the target and the gate Lgt (∼1 mm), variable resolution was used to set up the finite element mesh in the simulation. The element size chosen for the vicinity of the FE tip was ∼1 nm to ensure accurate simulation of electron trajectory with the focusing electrode. Since the structure of each FE tip was identical and independent, the simulation was performed on a single FE tip with circular symmetry around the tip. With axial distance of >2 μm above the focusing electrode, the electric field becomes essentially parallel; therefore, the element size for simulation was increased gradually to ∼1 μm to minimize the simulation time. The trajectories of a single electron emitted at different angles were determined with separate runs of the simulation.
Spatial distribution of electron beam intensity
The spatial distribution of the electron beam intensity as it reaches the HARP target, I(x,y), was calculated for each pixel of the FEA with del=50 μm×50 μm. This pixel design has been developed for a prototype 1 in. optical HARP-FEA sensor,28 in which an array of 17×17 FE tips was evenly distributed in an emission area of 20 μm×20 μm in the center of each pixel.
The electron beam intensity from a single tip, I0(x,y), was first obtained by substituting the inverse of the relationship between lateral spread and θ in Sec. 3A, i.e., θ=LS−1(x,y), into the angular intensity distribution of field emission, Iθ(θ), from a single tip, and then converting the result to Cartesian coordinates
| (4) |
The distribution of Iθ(θ) used in our calculation is shown in Fig. 5 and was adapted from the simulation and experimental measurements by Itoh et al. for Spindt-type FEA.32 The beam intensity for one pixel was then calculated by integrating I0(x,y) over all the FE tips in the x and y directions with Nx=Ny=17,
| (5) |
The method described above assumes that the interaction between electrons emitted from the same tip or neighboring tips is negligible. The main mechanism for interaction between electrons is the space charge effect due to the electric field generated by other electrons. This effect is proportional to the electron beam intensity and the square of the electron travel time tgt and inversely proportional to the radius of the electron beam. The space charge effect for the electron beam in a Vidicon has been investigated previously.34 It was found that for a distance of 2 cm between the last focusing (mesh) electrode and the target, the additional spread of electron beam due to the space charge effect was 11 μm with a beam current of 1.6 μA and radius of 166 μm at the mesh electrode. For FEA readout with a similar beam current (2 μA∕pixel), the additional spread due to the space charge effect is expected to be <1 μm because the travel distance Lgt is only 1 mm, which results in a travel time (∼0.18 nS) that is more than an order of magnitude smaller than that in the Vidicon (∼2.13 nS).34 Therefore, this effect can be ignored in our calculation of lateral spread of electron beams.
Figure 5.
Angular distribution of electrons in Spindt-type field emitters, adapted from Ref. 32.
Pixel aperture function
The pixel aperture function of the FEA readout method, MTFFEA(fx,fy), was determined from the Fourier transform of the spatial distribution of the image charge on the target, Qa(x,y), that was read out by each FEA pixel. Qa(x,y) is given by the integral of I(x,y) within the pixel readout time tp,
| (6) |
During electron beam readout, the target potential Vt decreases with time as the electrons reach the target. Since I(x,y) is Vt dependent, Qa(x,y) was calculated numerically by dividing tp into small steps and updating I(x,y) in real time. After Qa(x,y) was calculated for each focusing method, MTFFEA(fx,fy) was obtained through two-dimensional Fourier transform
| (7) |
he presampling MTF of SAPHIRE was then calculated by multiplying the MTFFEA(fx,fy) by the MTF of the SHARP combination
| (8) |
RESULTS AND DISCUSSION
Lateral spread of electron beam in FEA readout
Mesh-electrode-only
The trajectory of electrons emitted with angle θ was determined using Eqs. A5, A6 in the Appendix0, and the corresponding lateral spread was calculated using Eq. 1. The detector geometry and bias conditions used in the calculation, i.e., Vg, Vm, Lgm, and Lmt are listed in Table 1. Shown in Fig. 6 is the trajectory of electrons emitted with the critical angle θc=5.74°. The lateral distance traveled by the electron when it reaches the target corresponds to the maximum lateral spread, which is 59 μm at Vt=0.4 V. Since electrons emitted with θ>5.74° return to the mesh electrons, only a small fraction of the electron beam is utilized for readout. The efficiency increases with higher Vt because electrons with larger emission angle can reach the target. However, the drawback is a larger lateral spread, which increases to 444 μm with Vt=20 V. The major limitation of the mesh-electrode-only design is that lateral spread varies as a function of signal, which will lead to image artifact due to nonuniformity in spatial resolution that cannot be easily corrected. Furthermore, while this magnitude of lateral spread may be acceptable for general radiographic applications with pixel size ranging from del=150−200 μm, it is not adequate for mammography with del=50−100 μm. As shown in Eq. 1, lateral spread can be reduced by increasing Vm or decreasing Lgt, however, this makes the detector much more difficult to manufacture. Therefore, for mammography, additional focusing without reduction in Lgt is desirable.
Table 1.
Detector geometry and bias conditions used for all two electron-optical designs.
| Vg (V) | Vm (V) | Lgm (mm) | Lmt (mm) |
|---|---|---|---|
| 40 | 350 | 0.5 | 0.5 |
Figure 6.
Trajectories of electrons after emission from the FEA and before reaching the HARP target for both electron-optical designs. The mesh electrode is placed half-way between the FEA and the target, which are separated by 1 mm. The detector geometry and bias conditions used for the calculation are shown in Table 1.
Electrostatic focusing
The electron trajectory for electrostatic focusing was obtained using finite element method. The result for the same emission angle of θ=5.74° is shown in Fig. 6 in comparison with the mesh-electrode-only design. It shows that electrostatic focusing can reduce substantially the lateral spread (from 59 to 6 μm). Next we will further examine the dependence of electrostatic focusing on different FEA operating conditions.
Electrostatic focusing electrode potentialVL
With electrostatic focusing, the major factor affecting the trajectory of emitted electrons is the bias potential VL on the electrostatic focusing electrode. In general, VL<Vg is required to produce the focusing effect. Since the focusing electrode is located <1 μm away from the gate, VL has to be chosen carefully to avoid inadequate focusing. The diagrams in Fig. 7 show qualitatively the effect of VL on the trajectory of electrons emitted from different angles. Our strategy in the choice of VL is to minimize underfocusing, which leads to larger lateral spread and loss of electrons. Our simulation of electron trajectories with different VL showed that VL=−2 V provides the best compromise, and this value was used for calculating the results in Fig. 8.
Figure 7.
Conceptual electron trajectories under different VL bias conditions. (a) For electrons emitted with the same angle, different VL results in different lateral spread. (b) For the same VL, electrons emitted with different angle results in different lateral spread.
Figure 8.
Lateral spread of electrons with electrostatic focusing design as a function of electron emission angle θ under the operating conditions in Table 1 and FE tip geometry in Fig. 4. The target potential Vt=0.6 V.
Shown in Fig. 8 is the lateral spread as a function of θ for the operating conditions in Table 1 and FE tip geometry in Fig. 4. The LSmax occurs at θ=20°. This is because electrons emitted with larger θ are deflected by the focusing electrode and result in lower lateral spread. Electrons emitted with θ>28° are absorbed by the focusing electrodes in this simulation setting.
To facilitate the calculation of the pixel aperture function in the next section, the relationship between lateral spread and θ shown in Fig. 8 was fitted with a sixth order polynomial function, and the fitting result is shown for comparison in the same graph.
HARP target free surface potential Vt
In the mesh-electrode-only design, the lateral spread increases substantially with Vt (larger than the desired pixel size) and would result in nonuniformity in spatial resolution. Figure 9 shows the results of lateral spread with electrostatic focusing as a function of Vt for two emission angles: θ=5° and 20°. The figure shows that lateral spread decreases as Vt increases due to the reduction in electron travel time tgt. However, the variation in lateral spread is <2 μm over a Vt range of 0−30 V. Since this variation is much smaller than the desired pixel size, lateral spread can be regarded as essentially independent of Vt, which leads to uniform spatial resolution across the image.
Figure 9.
Lateral spread of electrons with electrostatic focusing as function of target potential Vt for electrons emitted at θ=5° and 20°.
Another difference for electrostatic focusing is that the fraction of emitted electrons utilized for readout is essentially independent of Vt because all electrons emitted with θ<28° can reach the target and contribute to the readout. Whereas with the mesh-electrode-only design, the critical angle θc decreases with decreasing Vt, as indicated by Eq. 2, which means that the readout becomes less efficient as Vt approaches ground (cathode) potential.
Detector geometry and bias conditions: Vm and Lgt
The dependence of lateral spread on Vm is shown in Fig. 10 for θ=5° and 20°. Plotted in the same graph for comparison is the lateral spread versus Vm for mesh-electrode-only design calculated using Eq. 3. It shows that lateral spread decreases with increase in Vm. This is because the lateral spread is the product of vx and tgt in both designs. While vx is reduced substantially by the focusing electrode, tgt can be reduced by increasing Vm or decreasing Lgt. However, since the lateral spread is much lower than that in mesh-electrode-only design, electrostatic focusing is much more tolerant to changes in detector geometry.
Figure 10.
Lateral spread of electron beams with the electrostatic focusing as a function of mesh electrode potential Vm with Vt=1.5 V. For comparison, the result for mesh-electrode-only at Vt=0.4 V is plotted in the same graph.
Spatial distribution of electron beam intensity
The intensity of the electron beam reaching the target from a single tip, I0(x,y) was obtained using Eq. 4. The angular dependence of lateral spread required for Eq. 4 was established using Eq. 1 and Fig. 8 for the two electron-optical designs, respectively. The results of the I0(x,y) calculation are shown in Fig. 11. Since I0(x,y) has circular symmetry, the results are shown as a function of x only (with y=0).
Figure 11.
Comparison of the electron beam intensity on target for a single FE tip, I0(x,y), for the two electron-optical designs.
As shown in Fig. 11, I0(x,y) for the mesh-electrode-only design has a lower intensity and wider spread; and it is not suitable for mammography detectors. For electrostatic focusing, I0(x,y) is the lowest near the center and peaks at the edge. The low intensity at the center is because electrons emitted with small θ are not affected substantially by the focusing electrode and their trajectory is similar to that in the mesh-electrode-only design. However, near the edge, where lateral spread is in the range of 26−27 μm, electrons emitted with a much wider angular range accumulate due to the effect of electrostatic focusing. This can be seen from the lateral spread versus θ graph in Fig. 8, where the curve is essentially flat near the maximum lateral spread of 27 μm for θ between 16° and 23°.
The results of I0(x,y) shown in Fig. 11 were used to calculate I(x,y) for the entire pixel with 17×17 FE tips. The intensity distribution of I(x,y) is shown in gray scale in Fig. 12 for electrostatic focusing. The result for mesh-electron-only design is not shown because the electron beam extends far beyond the pixel size of 50 μm and is not suitable for high resolution imaging applications such as mammography. In Fig. 12, the black square in the center outlines the emitting area of 20 μm×20 μm, the outer square represents the desired del=50 μm, and the boundary of 100 μm×100 μm shows the extent of electron beam spreading. Figure 12 shows that the intensity in the center of the pixel is the lowest, which is resulted from the shape of I0(x,y) shown in Fig. 11.
Figure 12.
Electron beam intensity distribution I(x,y) for each pixel of the FEA with electrostatic focusing design. The boundary of each graph measures 100 μm×100 μm, the outer square represents the pixel size of 50 μm×50 μm; and the small square in the center outlines the emitting area of 20 μm×20 μm.
Pixel aperture function
The spatial distribution of the image charge on the HARP target that is read out by each FEA pixel, Qa(x,y), was obtained using Eq. 6. Besides the parameters list in Table 1, the pixel readout time tp and the initial target potential Vt also affect Qa(x,y). In breast tomosynthesis, which is an emerging three-dimensional (3D) imaging technique using digital mammography detectors, a readout rate of 2−6 frames∕s is required. This translates to tp=1−3 μS if we divide the ITO electrode of SAPHIRE into Ns=128 strips. For simplicity of analysis, Vt=20 V was chosen for the calculation of Qa(x,y). This target potential corresponds to a detector exposure of 7 mR for tomosynthesis, which is 3.5 times the mean exposure.17 For screening mammography, Vt=20 V corresponds to the mean detector exposure of 20 mR. The result of Qa(x,y) calculation for electrostatic focusing is shown in Fig. 13. Since Qa(x,y) has circular symmetry, the result is plotted as a function of x only (with y=0). It shows that the shape of Qa for electrostatic focusing is essentially flat across the pixel, despite the lower I(x,y) intensity at the center. This is because the absolute values of I(x,y) are sufficient to read out 100% of the image charge within tp of 3 μs.
Figure 13.
The spatial distribution of image charge on target, Qa(x,y=0), that is read out by each FEA pixel. The initial Vt=20 V.
The pixel aperture function for electrostatic focusing, MTFFEA_E, was then derived from the Fourier transform of Qa(x,y) using Eq. 7 and the result is shown in Fig. 14. At 5 cycles∕mm, the value of MTFFEA_E is 84%. At 10 cycles∕mm, which is the Nyquist frequency for del=50 μm, the value decreases to 46%. For comparison, the inherent MTF for SHARP (MTFSHARP) and the total detector MTF (MTFsystem) are also shown in Fig. 14. The values used for MTFSHARP were adapted from our previous investigation of 150 μm thick HR type CsI layers.16 Figure 14 shows that the inherent resolution of SHARP is the dominant source of blur in SAPHIRE.
Figure 14.
The presampling MTF calculated from Qa(x,y) in Fig. 13 for the FEA readout method with electrostatic focusing (denoted as _E) methods. The presampling MTF for SHARP combination and the resulting system MTF for SAPHIRE are shown in the same graph.
CONCLUSIONS
A new detector concept SAPHIRE is being investigated to improve the low dose x-ray imaging performance of indirect FPI. In this article, we investigated the spatial resolution aspect of the imaging performance of SAPHIRE. The lateral spread of the electron beam emitted from the FEA, and the resulting pixel aperture function of the FEA readout method was investigated for two different electron-optical designs: mesh-electrode-only and electrostatic focusing. It was found that electrostatic focusing provides a pixel aperture function that is independent of the x-ray signal. It also provides a higher beam current by allowing more emitted electrons to reach the target. Therefore, it is the most promising electron-optical focusing method to be incorporated in the design of SAPHIRE for high resolution, low-dose x-ray imaging applications.
ACKNOWLEDGMENTS
This study is financially supported by the National Institute of Health (1 R01 EB002655) and Army Breast Cancer Research Program (W81XWH0410554). The authors gratefully acknowledge helpful discussion with Dr. Randy Luhta, Dr. Geordi Pang, and Dr. John. A. Rowlands. They would also like to acknowledge stimulating scientific discussion with Dr. Norifumi Egami from NHK Science and Technical Research Laboratory.
APPENDIX: DERIVATION OF ELECTRON TRAJECTORY AND LATERAL SPREAD FOR MESH-ELECTRODE-ONLY DESIGN
In this appendix, the trajectory of the electrons and the resulting lateral spread is derived for the mesh-electrode-only design. The lateral spread is the product of the lateral component of the electron velocity, vx, and the time it takes the electron to travel from the gate to the HARP target, tgt. The electrons were assumed to have an initial kinetic energy of KE0=qVg after emission from the gate electrode with an angle of θ (respect to z axis), hence, vx is given by
| (A1) |
where m is the mass of an electron. Since the electric field is in the z direction, vx remains unchanged as the electrons travel to the target. The potential V(z) and the axial velocity component vz(z) change as functions of position z (z=0 at gate electrode). The value of tgt is given by the sum of the time it takes to travel from the gate to the mesh electrode, tgm, and from the mesh electrode to the target, tmt,
| (A2) |
where agm=[(Vm−Vg)∕Lgm]×q∕m is the electron acceleration between the gate and the mesh and amt=[(Vm−Vt)∕Lmt]×q∕m is the deceleration between the mesh and the target. Using energy conservation, i.e., , we can obtain
| (A3) |
By substituting vz(Lgm) and vz(Lgt) into Eq. A2, tgt can be obtained using
| (A4) |
Thus, the x and z location of an electron at any time t is given by
| (A5) |
and
| (A6) |
Thus, the final lateral spread is given by x(t=tgt) as
| (A7) |
The lateral spread increases as θ increases while the other factors are kept constant. The condition for electrons to reach the target is vz≥0 at z=Lgt. The critical θc occurs when vz=0. Thus, Eq. A3 can be simplified to
| (A8) |
The maximum lateral spread (LSmax) can be obtained by substituting Eq. A8 into Eq. A7,
| (A9) |
References
- Liu X. and Shaw C. C., “a-Si:H∕CsI(Tl) flat-panel versus computed radiography for chest imaging applications: Image quality metrics measurement,” Med. Phys. 10.1118/1.1625102 31, 98–110 (2004). [DOI] [PubMed] [Google Scholar]
- Monnin P., Gutierrez D., Bulling S., Lepori D., Valley J. F., and Verdum F. R., “Performance comparison of an active matrix flat panel imager, computed radiography system, and a screen-film system at four standard radiation qualities,” Med. Phys. 10.1118/1.1843451 32, 343–350 (2005). [DOI] [PubMed] [Google Scholar]
- Zhao W., Ji W. G., Debrie A., and Rowlands J. A., “Imaging performance of amorphous selenium based flat-panel detectors for digital mammography: Characterization of a small area prototype detector,” Med. Phys. 10.1118/1.1538233 30, 254–263 (2003). [DOI] [PubMed] [Google Scholar]
- Zhao W. and Rowlands J. A., “Digital radiology using active matrix readout of amorphous selenium: Theoretical analysis of detective quantum efficiency,” Med. Phys. 10.1118/1.598097 24, 1819–1833 (1997). [DOI] [PubMed] [Google Scholar]
- Jee K. W. et al. , “System performance of a prototype flat-panel imager operated under mammographic conditions,” Med. Phys. 10.1118/1.1585051 30, 1874–1890 (2003). [DOI] [PubMed] [Google Scholar]
- Suryanarayanan S., Karellas A., Vedantham S., Sechopoulos I., and D’Orsi C. J., “Detection of simulated microcalcifications in a phantom with digital mammography: Effect of pixel size,” Radiology 244, 130–137 (2007). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yaffe M. J., in 1999 RSNA Syllabus Categorical Courses in Diagnostic Radiology Physics: Physics Aspect in Breast Imaging-Current and Future Considerations, edited by Haus A. G. and Yaffe M. J. (RSNA, Chicago, 1999), pp. 229–238. [Google Scholar]
- Juska G. and Arlauskas K., “Impact ionization and mobilities of charge-carriers at high electric-fields in amorphous selenium,” Phys. Status Solidi A 10.1002/pssa.2210590151 59, 389–393 (1980). [DOI] [Google Scholar]
- Tsuji K., Takasaki Y., Hirai T., Yamzaki J., and Tanioka K., “Avalanche phenomenon in amorphous selenium,” Optoelectron, Devices Technol. 9, 367–378 (1994). [Google Scholar]
- Luhta R. and Rowlands J. A., “Feasibility of a large area x-ray sensitive vidicon for medical fluoroscopy: Resolution and lag factors,” Med. Phys. 10.1118/1.598075 24, 621–631 (1997). [DOI] [PubMed] [Google Scholar]
- Itoh S. and Tanaka M., “Current status of field-emission displays,” Proc. IEEE 10.1109/JPROC.2002.1002523 90, 514–520 (2002). [DOI] [Google Scholar]
- Nanba M., Hirano Y., Honda Y., Miyakawa K., Ookawa Y., Watabe T., Okazaki S., Egami N., Miya K., Nakamura K., Taniguchi M., Itoh S., and Kobayashi A., “640×480 pixel HARP image sensor with active-matrix Spindt-type FEA,” Proceedings of 13th International Display Workshop, 2006, pp. 1817–1820.
- Negishi N., Matsuba Y., Tanaka R., Nakada T., Sakemura K., Okuda Y., Watanabe A., Yoshikawa T., Ogasawara K., Nanba M., Okazaki S., Tanioka K., Egami N., and Koshida N., “Development of a high resolution active-matrix electron emitter array for application to high-sensitivity image sensing,” J. Vac. Sci. Technol. B 10.1116/1.2709896 25, 661–665 (2007). [DOI] [Google Scholar]
- Zhao W., Li D., Reznik A., Lui B., Hunt D. C., Tanioka K., and Rowlands J. A., “Indirect flat-panel detector with avalanche gain: Fundamental feasibility investigation for SHARP-AMFPI (scintillator HARP active matrix flat panel imager),” Med. Phys. 10.1118/1.2008428 32, 2954–2966 (2005). [DOI] [PubMed] [Google Scholar]
- Zhao W., Li D., Rowlands J. A., Egami N., Takiguchi Y., Nanba M., Honda Y., Ohkawa Y., Kubota M., Tanioka K., Suzuki K., and Kawai T., “An indirect flat-panel detector with avalanche gain for low dose x-ray imaging: SAPHIRE (scintillator avalanche photoconductor with high resolution emitter readout),” Proc. SPIE 10.1117/12.772774 6913, 69130M.1–69130M.11 (2008). [DOI] [Google Scholar]
- Zhao W., Ristic G., and Rowlands J. A., “X-ray imaging performance of structured cesium iodide scintillators,” Med. Phys. 10.1118/1.1782676 31, 2594–2605 (2004). [DOI] [PubMed] [Google Scholar]
- Zhao W., Li D., Reznik A., Lui B., Hunt D. C., Tanioka K., and Rowlands J. A., “Indirect flat-panel detector with avalanche gain: Design and operation of the avalanche photoconductor,” Proc. SPIE 10.1117/12.597295 5745, 352–360 (2005). [DOI] [Google Scholar]
- Antonuk L. E., El-Mohri Y., Siewerdsen J. H., Yorkston J., Huang W., Scarpine V. E., and Street R. A., “Empirical investigation of the signal performance of a high resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation,” Med. Phys. 10.1118/1.597918 24, 51–70 (1997). [DOI] [PubMed] [Google Scholar]
- Nakamoto M., “Advances in field emission displays,” XXIInd International Symposium on Discharges and Electrical Insulation in Vacuum, 2006, Vol. 2, pp. 871–874.
- Sato H., Takegawa H., Yamaji H., Miyake H., Hiramatsu K., and Saito Y., “Fabrication of carbon nanotube array and its field emission property,” J. Vac. Sci. Technol. B 10.1116/1.1667515 22, 1335–1337 (2004). [DOI] [Google Scholar]
- Koshida N., Ozaki T., Sheng X., and Koyama H., “Cold electron-emission from electroluminescent porous silicon diodes,” Jpn. J. Appl. Phys., Part 2 10.1143/JJAP.34.L705 34, L705–L707 (1995). [DOI] [Google Scholar]
- Komoda T., Ichihara T., Honda Y., Aizawa K., and Koshida N., “Ballistic electron surface-emitting cold cathode by porous polycrystalline silicon film formed on glass substrate,” Mater. Res. Soc. Symp. Proc. 638, F4.1 (2000). [Google Scholar]
- Sakemura K., Negishi N., Yamada T., Satoh H., Watanabe A., Yoshikawa T., Ogasawara K., and Koshida N., “Development of an advanced high efficiency electro-emission device,” J. Vac. Sci. Technol. B 10.1116/1.1740759 22, 1367–1371 (2004). [DOI] [Google Scholar]
- Spindt C. A., Brodie I., Humphrey L., and Westerberg E. R., “Physical-properties of thin-film field-emission cathodes with molybdenum cones,” J. Appl. Phys. 10.1063/1.322600 47, 5248–5263 (1976). [DOI] [Google Scholar]
- Brodie I. and Spindt C. A., “Vacuum microelectronics,” Adv. Electron. Electron Phys. 10.1002/9780470141410.ch1 83, 1–106 (1992). [DOI] [Google Scholar]
- Itoh S., Tanaka M., and Tonegawa T., “Development of field emission displays,” J. Vac. Sci. Technol. B 10.1116/1.1691409 22, 1362–1366 (2004). [DOI] [Google Scholar]
- Itoh S., Tanaka M., Tonegawa T., Obara Y., Naito Y., Kobayashi H., and Sato Y., “Development of field emission display,” Proceedings of the 11th International Display Workshops, 2004, pp. 1189–1192.
- Egami N., Nanba M., Takiguchi Y., Miyakawa K., Watabe T., Okazaki S., Osada K., Obara Y., Tanaka M., and Itoh S., “50×50 μm pixel magnetic focus field emitter array image sensor with high-gain avalanche rushing amorphous photoconductor target,” J. Vac. Sci. Technol. B 10.1116/1.2050658 23, 2056–2062 (2005). [DOI] [Google Scholar]
- Tanioka K., Yamazaki J., Shidara K., Taketoshi K., Kawamura T., Hirai T., and Takasaki Y., “Avalanche-mode amorphous selenium photoconductive target for camera tube,” Adv. Electron. Electron Phys. 74, 379–387 (1988). [Google Scholar]
- Kubota M., Kato T., Suzuki S., Maruyama H., Shidara K., Tanioka K., Sameshima K., Makishima T., Tsuji K., Hirai T., and Yoshida T., “Ultrahigh-sensitivity new super-HARP camera,” IEEE Trans. Broadcast 42, 251–258 (1996). [Google Scholar]
- Lubinsky A. R., Zhao W., Ristic G., and Rowlands J. A., “Screen optics effects on detective quantum efficiency in digital radiography: Zero-frequency effects,” Med. Phys. 10.1118/1.2188082 33, 1499–1509 (2006). [DOI] [PubMed] [Google Scholar]
- Itoh S., Watanabe T., Ohtsu K., Tanigushi M., Uzawa S., and Nishimura N., “Experimental study of field-emission properties of the Spindt-type field emitter,” J. Vac. Sci. Technol. B 10.1116/1.588339 13, 487–490 (1995). [DOI] [Google Scholar]
- Takiguchi Y., Nanba M., Osada K., Watabe T., Okazabi S., Egami N., Tanioka K., Tanaka M., and Itoh S., “256×192 pixel field emitter array image sensor with high-gain avalanche rushing amorphous photoconductor target,” J. Vac. Sci. Technol. B 10.1116/1.1667519 22, 1390–1395 (2004). [DOI] [Google Scholar]
- Luhta R. P., “A large-area x-ray-sensitive vidicon,” Ph.D. thesis, University of Toronto, 1997. [Google Scholar]














