Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 1989 May;27(5):1027–1030. doi: 10.1128/jcm.27.5.1027-1030.1989

Comparative evaluation of the Roche Cobas IDA and Enterotube II systems for identifying members of the family Enterobacteriaceae.

B Holmes 1
PMCID: PMC267476  PMID: 2663910

Abstract

Two Hoffmann-La Roche products were evaluated in parallel for their ability to identify 321 strains of the family Enterobacteriaceae. The first product was the well-established Enterotube II, which correctly identified 90% of the isolates after 24 h of incubation. The second was the ID-E rotor, originally designed for use in the Cobas Bact but here used in the new Cobas IDA system. The Cobas IDA, used with the full data base held in a microcomputer supplied with the product, identified 87% of strains after 4 h of incubation.

Full text

PDF
1027

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Davis J. R., Stager C. E., Wende R. D., Qadri S. M. Clinical laboratory evaluation of the automicrobic system Enterobacteriaceae biochemical card. J Clin Microbiol. 1981 Oct;14(4):370–375. doi: 10.1128/jcm.14.4.370-375.1981. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Holmes B., Dawson C. A., Pinning C. A. A revised probability matrix for the identification of gram-negative, aerobic, rod-shaped, fermentative bacteria. J Gen Microbiol. 1986 Nov;132(11):3113–3135. doi: 10.1099/00221287-132-11-3113. [DOI] [PubMed] [Google Scholar]
  3. Holmes B., Willcox W. R., Lapage S. P. Identification of Enterobacteriaceae by the API 20E system. J Clin Pathol. 1978 Jan;31(1):22–30. doi: 10.1136/jcp.31.1.22. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Holmes B., Willcox W. R., Lapage S. P., Malnick H. Test reproducibility of the API (20E), Enterotube, and Pathotec systems. J Clin Pathol. 1977 Apr;30(4):381–387. doi: 10.1136/jcp.30.4.381. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Isenberg H. D., Scherber J. S., Cosgrove J. O. Clinical laboratory evaluation of the further improved Enterotube and Encise II. J Clin Microbiol. 1976 Aug;2(2):139–141. [PMC free article] [PubMed] [Google Scholar]
  6. Kelly M. T., Latimer J. M. Comparison of the automicrobic system with API, enterotube, micro-ID, micro-media systems, and conventional methods for identification of Enterobacteriaceae. J Clin Microbiol. 1980 Nov;12(5):659–662. doi: 10.1128/jcm.12.5.659-662.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)