During the first few minutes of life, oxygen saturation (saturation by pulse oximetry, SpO2) increases from intrapartum levels of 30–40%.1 In algorithms for neonatal resuscitation published by the International Liaison Committee for Resuscitation,2 European Resuscitation Council3 and Australian Resuscitation Council,4 clinical assessment of an infant's colour (a measure of oxygenation) and heart rate are used as major action points. However, studies have shown that clinical assessment of colour during neonatal transition is unreliable.5,6 O'Donnell et al6 showed that the SpO2 at which observers perceived infants to be pink varied widely, ranging from 10% to 100%. Assessing colour is difficult and therefore is a poor proxy for tissue oxygenation during the first few minutes of life.
Kattwinkel7 suggested pulse oximetry may help achieve normoxia in the delivery room. The American Heart Association8 suggests that “administration of a variable concentration of oxygen guided by pulse oximetry may improve the ability to achieve normoxia more quickly”. Although “normoxia” and an acceptable time to achieve this during neonatal transition have not been defined, Leone and Finer9 advocate a target “SpO2 of 85 to 90% by three minutes after birth for all infants except in special circumstances”—for example, diaphragmatic hernia or cyanotic congenital heart disease. International surveys show that oximetry is increasingly used during neonatal resuscitation.10,11
To date, there are no evidence‐based guidelines for using oximetry to measure an infant's SpO2 and to guide interventions during neonatal transition after birth. We reviewed the literature to evaluate the evidence on the use of SpO2 measurements immediately after birth.
How does pulse oximetry work?
Pulse oximetry measures SpO2 continuously and non‐invasively, without the need for calibration, and correlates closely with arterial oxygen saturation.12 Pulse oximetry is based on the red and infrared light‐absorption characteristics of oxygenated and deoxygenated haemoglobin. A sensor is placed around a hand or foot and two light‐emitting diodes send red and infrared light through to a photodetector on the other side. The changes in absorption during the arterial pulsatile flow and non‐pulsatile component of the signal are analysed. SpO2 is estimated from the transmission of light through the pulsatile tissue bed. With each heartbeat, there is a surge of arterial blood that momentarily increases arterial blood volume. This results in more light absorption during surges. As peaks occur with each heartbeat, heart rate can also be measured.
Can SpO2 be successfully measured in the minutes after birth?
Seven studies reported between 20% and 100% success in obtaining SpO2 measurements by 1 min after birth.5,13,14,15,16,17,18 By 5 min, the success rate improved to between 63% and 100%.13,15,16,17,19,20,21,22,23 The most common reason for failing to obtain a measurement was motion artefact5,16,18,19,20,24; others were the presence of vernix,25 low perfusion,25 oedema,15 high ambient light,14,24 large infants,24 cracked and wrinkled skin,15 or acrocyanosis.15 Artefacts occurred less often in more recent studies where Masimo signal extraction technology (SET) was used.18,25
Where should the oximeter sensor be applied?
In early studies, investigators placed the sensor over the right Achilles tendon,20,25,26,27 the forefoot19 or midfoot.22,28 Later studies found that measurements were obtained fastest from the right hand,15 probably owing to better perfusion, higher blood pressure and oxygenation in preductal vessels.14,29 Preductal readings were significantly higher than postductal readings soon after birth (p<0.05).5,15,22 By 17 min after birth, there was no longer a significant difference between preductal and postductal measurements (p<0.05).5,15,22
How quickly can an SpO2 reading be obtained?
A sensor can be applied to a baby within 15–20 s of birth,18,21 with the first data obtained at about 50 s after birth.15,18,21 No studies obtained SpO2 data on most infants before 1 min after birth.14,25,26,29 When the Masimo sensor and monitor were used, readings were obtained faster than when the sensor was applied to the infant before connecting it to the oximeter.21
How do SpO2 readings change in the first few minutes after birth?
Some studies report the range of SpO2 at 1, 5 or 10 min (tables 1 and 2); others report the time taken to reach a predetermined SpO2 (table 3). These studies show increases in SpO2 from about 60% at 1 min, but the levels vary widely, with some infants taking >10 min to exceed 90%. Therefore, it may not be appropriate to identify specific SpO2 levels at certain times after birth, which can be used as a trigger to alter an infant's treatment.
Table 1 Observational studies measuring SpO2 in the first few minutes of life in the delivery room where no infant received supplemental oxygen.
Study | Gestation (weeks) | Type of oximeter | Sensor location | n | SpO2 (%) | Comments | |||
---|---|---|---|---|---|---|---|---|---|
1 min | 5 min | 10 min | |||||||
Harris et al20 | >37 | Nellcor N‐100 | Postductal | 32 | 61 (5)* | NA | NA | Vaginal delivery | |
44 | 46 (3)* | C/S | |||||||
Toth et al22 | ⩾35 | Nellcor N‐300 | Preductal | 50 | NA | 84 (48–99)† | 92 (65–99)† | 48 spontaneous deliveries, 2 vacuum extraction | |
Postductal | 78 (42–97)† | 89 (62–99)† | |||||||
Rabi et al29 | ⩾35 | Masimo Radical | Preductal | 45 | NA | 87 (80–95)‡ | NA | Vaginal delivery | Calgary (1049 m) |
81 (75–83)‡ | C/S | Calgary (1049 m) | |||||||
Kamlin et al18 | ⩾31 | Masimo Radical | Preductal | 175 | 63 (53–68)‡ | 90 (79–91)‡ | NA | 51 preterm | |
124 term infants | |||||||||
Gonzales and Salirrosas28 | >37 | Nellcor N‐20 | Postductal | 37 | 42 (2)* | NA | NA | Cerro de Pasco (4340 m) | |
131 | 61 (1)* | Lima (150 m) sea level | |||||||
Gungor et al30 | >37 | Air‐Shields Vickers 19040 | Preductal | 70 | 69 (0.7)§ | 90 (2)§ | NA | No suction | |
70 | 70 (0.7)§ | 80 (2)§ | 92 (0.4)§ | Suction |
C/S, caesarean section; IQR, interquartile range; NA, not available; SpO2, saturation by pulse oximetry.
*Mean (SEM); †mean (range); ‡median (IQR); §mean(SD).
Table 2 Observational studies measuring SpO2 in the first few minutes of life in the delivery room where some infants were treated with 100% oxygen.
Study | Gestation | Type of oximeter | Sensor location | Resuscitation | n | SpO2 % | Comments | ||
---|---|---|---|---|---|---|---|---|---|
1 min | 5 min | 10 min | |||||||
Sendak et al26 | Term and preterm | Nellcor N‐100 | Postductal | No infant received oxygen | 25 | 63* | 89* | NA | Vaginal delivery |
34 | 27 (4)† | 72* | C/S | ||||||
100% oxygen given | 27 | 48 (5)† | 69* | C/S and oxygen | |||||
House et al14 | Term and preterm | Nellcor N‐100 or Ohmeda Biox 3700 | Preductal | 19/38 vaginal deliveries and 53/62 C/S received 100% oxygen | 100 | 58 (22)‡ | 82 (14)‡ | 89 (6)‡ | |
No infant received oxygen | 28 | 78 (9)‡ | 84 (14)‡ | 90 (5)‡ | C/S and vaginal deliveries | ||||
Deckardt et al19 | >37 weeks | Nellcor N‐100 | Postductal | 12 infants received CPAP with 100% oxygen | 35 | 40–75 range | NA | NA | All vaginal deliveries |
Porter et al24 | Term | Ohmeda Biox 3700 | Preductal or postductal | 100% oxygen if poor respiratory effort, central cyanosis, or heart rate <100. Number receiving oxygen not indicated | 96 | 77 (11)‡ | 84 (7)‡ | 89 (6)‡ | C/S and vaginal deliveries |
Rao and Ramji16 | Term and preterm | Novametrix 515A | Preductal | Infants with asphyxia randomised to receive air or 100% oxygen during resuscitation | 95 | 45 (20)‡ | 84 (14)‡ | 91 (10)‡ | Infants with asphyxia enrolled in the Resair 2 study |
Not reported | 30 | 70 (16)‡ | 89 (9)‡ | 94 (2)‡ | Non‐asphyxiated controls | ||||
Dimich et al5 | Not reported | Ohmeda Biox 3700 | Preductal | 7 infants received 100% supplemental oxygen | 100 | 72 (6)‡ | 83 (4)‡ | 91 (5)‡ | 63 vaginal deliveries 37 C/S |
Postductal | 63 (4)‡ | 77 (4)‡ | 87 (6)‡ |
CPAP, continuous positive airway pressure; C/S, caesarean section; NA, not available; SpO2, saturation by pulse oximetry.
*Mean; †mean (SEM); ‡mean (SD).
Table 3 Time (in min) to reach specified SpO2 levels after birth.
Study | Gestation (weeks) | Type of oximeter | Sensor location | n | Resuscitation | SpO2 | Comments | |||
---|---|---|---|---|---|---|---|---|---|---|
>75% | >80% | >90% | 95% | |||||||
Toth et al22 | ⩾35 | Nellcor N‐3000 | Preductal | 50 | No infant received oxygen | NA | NA | NA | 12 (2–55)* | 48 spontaneous deliveries, 2 vacuum extraction |
Postductal | 50 | 14 (3–55)* | ||||||||
Kamlin et al18 | ⩾31 | Masimo Radical | Preductal | 175 | No infant received oxygen | NA | NA | 5.8 (3.2)† Range 1.3–20.2 | NA | 51 preterm |
124 term infants | ||||||||||
Kopotic and Lindner25 | < 30 | Masimo Radical | Preductal | 15 | Infants initially received 100% oxygen then oxygen adjusted according to SpO2 measurements | NA | 4.4 (1.9–40)‡ | NA | NA | |
Vento et al31 | > 37 | Not reported | Not reported | 22 | Control group | NA | NA | 0.9 (0.4)† | NA | Non‐asphyxiated |
55 | Air | 2.0 (0.7)† | Infants with asphyxia | |||||||
52 | 100% oxygen | 1.8 (0.9)† | ||||||||
Rao and Ramji16 | ⩾31 | Novametrix 515A | Preductal | 95 | Infants with asphyxia randomised to receive air or 100% oxygen during resuscitation | 5 (3–7.25)§ | NA | 7.3 (4.5–11)§ | NA | Infants with asphyxia enrolled in the Resair 2 study |
30 | Not reported | 2.8 (1.6–4.4)§ | NA | 4.3 (2.9–5.8)§ | NA | Non‐asphyxiated control group | ||||
Saugstad et al32 | ⩾31 | Not reported | Not reported | 103 | Air | 1.5 (1.4–1.6)§ | NA | NA | NA | Infants with asphyxia enrolled in the Resair 2 study |
109 | 100% oxygen | 2.5 (1.9–3.1)§ |
NA, not applicable, SpO2, saturation by pulse oximetry.
*Mean (range); †mean (SD); ‡median (range); §median (95% CI).
Does the type of oximeter alter the SpO2 results?
Early oximeters had motion artefact.5,16,18,19,20,24 This has been improved in newer oximeters.25,33 To determine whether the newer oximeters were more reliable than earlier models in the delivery room, Kopotic compared the Masimo SET to the Nellcor Oxismart, with sensors placed on each foot, in 15 newborns of <30 weeks' gestation.25 The Masimo SET provided data for 350 of 362 (96%) min, and the Oxismart provided data for 212 of 362 min (59%; p = 0.0014).25 Leone and Finer9 recommended that oximeters used during neonatal resuscitation should have “minimal averaging time for the SpO2 values coupled with maximum sensitivity”. The combination of these features allows rapid detection of changes in SpO2 and improved SpO2 measurement during periods of low perfusion.34
Does the type of delivery alter the SpO2 after birth?
Harris et al20 found, using an early generation oximeter, that SpO2 was much lower in 44 term elective caesarean‐section deliveries, when compared with 32 term infants delivered vaginally. The mean (standard error, SEM) SpO2 at 1 min was 46% (3%) in the caesarean group and 61% (5%) in the vaginal delivery group (p<0.05), but by 5 min there was no significant difference. They postulated that the difference was due to the increased amount of lung fluid after caesarean section. Kamlin et al18 found that 107 term infants born by elective caesarean section took on average 2 min longer to reach an SpO2 >90% than 68 infants born by spontaneous vaginal delivery. Rabi et al29 found a similar difference in a cohort of 115 infants. In infants of >34 weeks' gestation, the median (interquartile range, IQR) for vaginal births at 5 min was 87% (80–95%) and that for caesarean delivery was 81% (75–83%).29 In contrast, others found no significant difference in SpO2 measurements in infants delivered vaginally or by caesarean section, regardless of the presence or type of anaesthesia.5,14,24
Does resuscitation with air or oxygen affect SpO2 after birth?
Table 4 summarises trials comparing SpO2 measurements at 1, 3 and 5 min in infants with asphyxia randomised to receive air or 100% oxygen during resuscitation. In the Resair 2 study, which enrolled infants weighing >999 g with apnoea and bradycardia at birth, there were no significant differences in time to reach an SpO2 of 75%. The median (95% confidence interval) time to reach an SpO2 of 75% was 1.5 (1.4 to 1.6) min in the group receiving air versus 2.5 (1.9 to 3.1) min in the group receiving oxygen (p = 0.27).32 In this study, the resuscitators were aware of the gas used, whereas Vento et al31 blinded resuscitators to the type of gas used to resuscitate infants with asphyxia. He found no significant difference in time to reach an SpO2 >90% between the two groups with asphyxia. The striking result of these studies is that resuscitating with air or 100% oxygen had little effect on the change in SpO2 in the first 10 min after birth.
Table 4 SpO2 measurements in infants with asphyxia randomised to resuscitation with air or 100% oxygen.
1 min | 5 min | 10 min | ||||
---|---|---|---|---|---|---|
Air | Oxygen | Air | Oxygen | Air | Oxygen | |
Saugstad et al32 | 65 (11)* | 61 (14)* | 86 (10)* | 88 (10)* | 90 (6)* | 91 (7)* |
Saugstad et al35 | 68 (40–82)† | 63 (40–82)† | 90 (66–95)† | 90 (73–96)† | 90 (83–96)† | 92 (79–97)† |
*Mean (SD); †median (5th–95th centile).
Does altitude affect SpO2 after birth?
Gonzales and Salirrosas28 showed that SpO2 was significantly higher in infants born at sea level (Lima 150 m) than in infants born at a higher altitude (Cerro de Pasco 4340 m) from 1 min to 24 h after birth (p<0.01).
Does gestation affect SpO2?
There are two reports of SpO2 measurements in premature infants after birth. In Kopotic and Lindner's25 study of 15 infants born at 24–29 weeks' gestation, the SpO2 was ⩾80% by 4.4 (1.9–40) min (median (range)). In the study by Kamlin et al18 on infants not receiving resuscitation, the time to reach an SpO2 >90% was significantly longer in 54 preterm infants at 6.5 (4.9–9.8) min (median (IQR)) than in 121 term infants at 4.7 (3.3–6.4) min (median (IQR)) (p<0.001). Other studies including premature infants did not report SpO2 for different gestational ages.14,32
Can oximetry be used to measure the effect of resuscitation practices?
Oximetry has been used to measure the effect of clinical interventions, such as oropharyngeal suction and endotracheal intubation during neonatal transition. Three controlled studies show that suctioning does seem to have a negative effect on oxygenation.13,23,30 O'Donnell et al36 measured the effects of attempted endotracheal intubation on SpO2 in the delivery room, and SpO2 often fell during intubation attempts.
Could oximetry be used in the delivery room to improve outcomes?
There are two studies that evaluate the use of oximetry to guide interventions during neonatal transition. Deckardt et al19 used SpO2 readings at 5 min after birth to determine whether infants should receive continuous positive airway pressure (CPAP) with a mask and 100% oxygen. CPAP was used only if the SpO2 was <80% at 5 min and stopped once the SpO2 reached 90%. Kopotic and Lindner25 studied 50 infants at risk for respiratory failure; 25 infants were managed without oximetry and compared with 25 managed with oximetry. Infants managed with oximetry were less likely to be admitted to the special care nursery (32% v 52%; p = 0.04). They also observed the effect of oximetry during resuscitation in 15 infants of <30 weeks' gestation. Initial respiratory care was based on the infant's clinical state and oximetry measurements. Oxygen was started at 100% and adjusted to achieve an SpO2 between 80% and 92%.25 The authors claim that by using pulse oximetry they were able to reduce the fraction of inspired oxygen (FiO2) from 1.0 to, on average, 0.40. The studies by Kopotic and Deckardt, although non‐blinded and non–randomised, suggest that oximetry can improve short‐term outcomes—for example, admission to nursery, the use of oxygen or CPAP. We could find no reports on whether the use of SpO2 measurements immediately after birth alters long‐term outcomes.
Conclusion
Before oximetry is advocated for routine use in the delivery room, more research is needed to define normoxia, and more importantly, how to interpret and apply SpO2 readings to clinical practice to improve short‐term and long‐term outcomes.
Abbreviations
CPAP - continuous positive airway pressure
IQR - interquartile range
SET - signal extraction technology
SpO2 - saturation by pulse oximetry
Footnotes
Supported in part by a RWH Postgraduate degree scholarship (CPFOD and COFK) and a NHMRC Practitioner Fellowship (PGD).
Competing interests: None.
References
- 1.East C E, Colditz P B, Begg L M.et al Update on intrapartum fetal pulse oximetry. Aust N Z J Obstet Gynaecol 200242119–124. [DOI] [PubMed] [Google Scholar]
- 2.International Liaison Committee on Resuscitation Part 7: neonatal resuscitation. Resuscitation 200567293–303. [DOI] [PubMed] [Google Scholar]
- 3.Biarent D, Bingham R, Richmond S.et al European Resuscitation Council Guidelines for Resuscitation. Resuscitation 200567S1S97–133. [DOI] [PubMed] [Google Scholar]
- 4.Australian Resuscitation Council Guideline 13.1, Introduction to resuscitation of the newborn infant. 2006. http://www.resus.org.au(accessed 10 Aug 2006)
- 5.Dimich I, Singh P P, Adell A.et al Evaluation of oxygen saturation monitoring by pulse oximetry in neonates in the delivery system. Can J Anaesth 199138985–988. [DOI] [PubMed] [Google Scholar]
- 6.O'Donnell C P F, Kamlin C O F, Davis P G.et alClinical assessment of colour at neonatal resuscitation: The Mullet Study. Washington, DC: PAS, 2005
- 7.Kattwinkel J. Evaluating resuscitation practices on the basis of evidence: the findings at first glance may seem illogical. J Pediatr 2003142221–222. [DOI] [PubMed] [Google Scholar]
- 8.American Heart Association American Heart Association (AHA) Guidelines for cardiopulmonary resuscitation (CPR) and emergency cardiovascular care (ECC) of pediatric and neonatal patients: neonatal resuscitation guidelines. Pediatrics 20061171029–1038. [DOI] [PubMed] [Google Scholar]
- 9.Leone T A, Finer N N. Neonatal resuscitation: beyond the basics. NeoReviews 20056e177–e183. [Google Scholar]
- 10.Leone T A, Rich W, Finer N N. A survey of delivery room resuscitation practices in the United States. Pediatrics 2006117164–175. [DOI] [PubMed] [Google Scholar]
- 11.O'Donnell C P F, Davis P G, Morley C J. Use of supplementary equipment for resuscitation of newborn infants at tertiary perinatal centres in Australia and New Zealand. Acta Paediatr 2005941261–1265. [DOI] [PubMed] [Google Scholar]
- 12.Hay W W, Jr, Brockway J M, Eyzaguirre M. Neonatal pulse oximetry: accuracy and reliability. Pediatrics 198983717–722. [PubMed] [Google Scholar]
- 13.Carrasco M, Martell M, Estol P C. Oronasopharyngeal suction at birth: effects on arterial oxygen saturation. J Pediatr 1997130832–834. [DOI] [PubMed] [Google Scholar]
- 14.House J T, Schultetus R R, Gravenstein N. Continuous neonatal evaluation in the delivery room by pulse oximetry. J Clin Monit 1987396–100. [DOI] [PubMed] [Google Scholar]
- 15.Meier‐Stauss P, Bucher H U, Hürlimann R.et al Pulse oximetry used for documenting oxygen saturation and right‐to‐left shunting immediately after birth. Eur J Pediatr 1990149851–855. [DOI] [PubMed] [Google Scholar]
- 16.Rao R, Ramji S. Pulse oximetry in asphyxiated newborns in the delivery room. Indian Pediatr 200138762–766. [PubMed] [Google Scholar]
- 17.Rao R, Yax S, Rao S.The role of oximetry in the first 10 minutes of age after birth. Washington, DC: PAS, 2005
- 18.Kamlin C O F, O'Donnell C P F, Davis P G.et al Oxygen saturation in healthy infants immediately after birth. J Pediatr 2006148585–589. [DOI] [PubMed] [Google Scholar]
- 19.Deckardt R, Schneider K T, Graeff H. Monitoring arterial oxygen saturation in the neonate. J Perinat Med 198715357–360. [DOI] [PubMed] [Google Scholar]
- 20.Harris A P, Sendak M J, Donham R T. Changes in arterial oxygen saturation immediately after birth in the human neonate. J Pediatr 1986109117–119. [DOI] [PubMed] [Google Scholar]
- 21.O'Donnell C P F, Kamlin C O F, Davis P G.et al Feasibility of and delay in obtaining pulse oximetry during neonatal resuscitation. J Pediatr 2005147698–699. [DOI] [PubMed] [Google Scholar]
- 22.Toth B, Becker A, Seelbach‐Göbel B. Oxygen saturation in healthy newborn infants immediately after birth measured by pulse oximetry. Arch Gynecol Obstet 2002266105–107. [DOI] [PubMed] [Google Scholar]
- 23.Waltman P A, Brewer J M, Rogers B P.et al Building evidence for practice: a pilot study of newborn bulb suctioning at birth. J Midwifery Womens Health 20044932–38. [DOI] [PubMed] [Google Scholar]
- 24.Porter K B, Golhamer R, Mankad A.et al Evaluation of arterial oxygen saturation in pregnant patients and their newborns. Obstet Gynecol 198871354–357. [PubMed] [Google Scholar]
- 25.Kopotic R J, Lindner W. Assessing high‐risk infants in the delivery room with pulse oximetry. Anaesth Analg 200294S31–S36. [PubMed] [Google Scholar]
- 26.Sendak M J, Harris A P, Donham R T. Pulse oximetry in newborn infants in the delivery room. Anaesthesiology 198563433 [Google Scholar]
- 27.Maxwell L G, Harris A P, Sendak M J.et al Monitoring the resuscitation of preterm infants in the delivery room using pulse oximetry. Clin Pediatr (Philadelphia) 19872618–20. [DOI] [PubMed] [Google Scholar]
- 28.Gonzales G F, Salirrosas A. Arterial oxygen saturation in healthy newborns delivered at term in Cerro de Pasco (4340 m) and Lima (150 m). Reprod Biol Endocrinol 2005346. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Rabi Y, Yee W, Chen S Y.et al Oxygen saturation trends immediately after birth. J Pediatr 2006148590–594. [DOI] [PubMed] [Google Scholar]
- 30.Gungor S, Teksoz E, Ceyhan T.et al Oronasopharyngeal suction versus no suction in normal, term and vaginally born infants: a prospective randomised controlled trial. Aust NZ J Obstet Gynaecol 200545453–456. [DOI] [PubMed] [Google Scholar]
- 31.Vento M, Asensi M, Sastre J.et al Oxidative stress in asphyxiated term infants resuscitated with 100% oxygen. J Pediatr 2003142240–246. [DOI] [PubMed] [Google Scholar]
- 32.Saugstad O D, Rootwelt T, Aalen O. Resuscitation of asphyxiated newborn infants with room air or oxygen: an international controlled trial: the Resair 2 study. Pediatrics 1998102e1. [DOI] [PubMed] [Google Scholar]
- 33.Bohnhorst B, Corinna P S, Poets C F. Pulse oximeters' reliability in detecting hypoxemia and bradycardia: comparison between a conventional and two new generation oximeters. Crit Care Med 2000281565–1568. [DOI] [PubMed] [Google Scholar]
- 34.Masimo Corporation Radical signal extraction pulse oximeter operator's manual. Irvine, CA: Masimo, 2004
- 35.Saugstad O D, Siddarth R, Rootwelt T.et al Response to resuscitation of the newborn: early prognostic variables. Acta Paediatr 200594890–895. [DOI] [PubMed] [Google Scholar]
- 36.O'Donnell C P F, Kamlin C O, Davis P G.et al Endotracheal intubation attempts during neonatal resuscitation: success rates, duration, and adverse effects. Pediatrics 2006117e16–e21. [DOI] [PubMed] [Google Scholar]