Abstract
BacT/Alert (Organon Teknika Corp., Durham, N.C.) is an automated microbial detection system based on the colorimetric detection of CO2 produced by growing microorganisms. Results of an evaluation of the media, sensor, detection system, and detection algorithm indicate that the system reliably grows and detects a wide variety of bacteria and fungi. Results of a limited pilot clinical trial with a prototype research instrument indicate that the system is comparable to the radiometric BACTEC 460 system in its ability to grow and detect microorganisms in blood. On the basis of these initial findings, large-scale clinical trials comparing BacT/Alert with other commercial microbial detection systems appear warranted.
Full text
PDF




Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Kim M. J., Gottschall R. L., Schwabe L. D., Randall E. L. Effect of agitation and frequent subculturing on recovery of aerobic and facultative pathogens by Roche Septi-Chek and BACTEC blood culture systems. J Clin Microbiol. 1987 Feb;25(2):312–315. doi: 10.1128/jcm.25.2.312-315.1987. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Plorde J. J., Carlson L. G., Dau M. E. Lack of clinical relevance in routine final subcultures of radiometrically negative BACTEC blood culture vials. Am J Clin Pathol. 1982 Nov;78(5):753–755. doi: 10.1093/ajcp/78.5.753. [DOI] [PubMed] [Google Scholar]
- Washington J. A., 2nd, Ilstrup D. M. Blood cultures: issues and controversies. Rev Infect Dis. 1986 Sep-Oct;8(5):792–802. doi: 10.1093/clinids/8.5.792. [DOI] [PubMed] [Google Scholar]
- Weinstein M. P., Mirrett S., Reimer L. G., Reller L. B. Effect of agitation and terminal subcultures on yield and speed of detection of the Oxoid Signal blood culture system versus the BACTEC radiometric system. J Clin Microbiol. 1989 Mar;27(3):427–430. doi: 10.1128/jcm.27.3.427-430.1989. [DOI] [PMC free article] [PubMed] [Google Scholar]