Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 1986 Jan;23(1):126–128. doi: 10.1128/jcm.23.1.126-128.1986

Comparison of the MicroScan system with the API Staph-Ident system for species identification of coagulase-negative staphylococci.

Z Hussain, L Stoakes, D L Stevens, B C Schieven, R Lannigan, C Jones
PMCID: PMC268585  PMID: 3517037

Abstract

To evaluate the accuracy of the MicroScan System (American Hospital Supply Corp., Sacramento, Calif.) for identification of coagulase-negative staphylococci, we tested 175 clinical isolates of coagulase-negative staphylococci. The results obtained by the MicroScan system were compared with those of the API Staph-Ident system (Analytab Products, Plainview, N.Y.). Forty-three discrepancies between the two systems were resolved by the conventional method of Kloos and Schleifer (W.E. Kloos and K.H. Schleifer, J. Clin. Microbiol. 1:82-88, 1975). The MicroScan and the Staph-Ident systems correctly identified 146 (86.4%) and 154 (88%) of 175 strains, respectively. The API system failed to identify phosphatase-negative Staphylococcus epidermidis. The MicroScan system demonstrated the greatest accuracy in the identification of S. epidermidis and S. saprophyticus, whereas lesser accuracy was achieved with S. hominis, S. warneri, and S. sciuri.

Full text

PDF
126

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Almeida R. J., Jorgensen J. H. Identification of coagulase-negative staphylococci with the API STAPH-IDENT system. J Clin Microbiol. 1983 Aug;18(2):254–257. doi: 10.1128/jcm.18.2.254-257.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Christensen G. D., Parisi J. T., Bisno A. L., Simpson W. A., Beachey E. H. Characterization of clinically significant strains of coagulase-negative staphylococci. J Clin Microbiol. 1983 Aug;18(2):258–269. doi: 10.1128/jcm.18.2.258-269.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Eng R. H., Wang C., Person A., Kiehn T. E., Armstrong D. Species identification of coagulase-negative staphylococcal isolates from blood cultures. J Clin Microbiol. 1982 Mar;15(3):439–442. doi: 10.1128/jcm.15.3.439-442.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Kloos W. E., Schleifer K. H. Simplified scheme for routine identification of human Staphylococcus species. J Clin Microbiol. 1975 Jan;1(1):82–88. doi: 10.1128/jcm.1.1.82-88.1975. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Kloos W. E., Wolfshohl J. F. Identification of Staphylococcus species with the API STAPH-IDENT system. J Clin Microbiol. 1982 Sep;16(3):509–516. doi: 10.1128/jcm.16.3.509-516.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Schleifer K. H., Kloos W. E. A simple test system for the separation of staphylococci from micrococci. J Clin Microbiol. 1975 Mar;1(3):337–338. doi: 10.1128/jcm.1.3.337-338.1975. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)