Abstract
A total of 398 consecutive clinical staphylococcus isolates, of which 205 were coagulase negative and 193 were coagulase positive, were tested in parallel by using AutoMicrobic system Gram-Positive Susceptibility-MIC cards and modified Mueller-Hinton agar containing 4% NaCl and oxacillin (6 micrograms/ml). The AutoMicrobic system cards correctly detected 103 of 104 (99%) oxacillin-resistant coagulase-negative isolates with no reports of false resistance.
Full text
PDF

Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Hansen S. L., Freedy P. K. Variation in the abilities of automated, commercial, and reference methods to detect methicillin-resistant (heteroresistant) Staphylococcus aureus. J Clin Microbiol. 1984 Sep;20(3):494–499. doi: 10.1128/jcm.20.3.494-499.1984. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jorgensen J. H., Redding J., Johnson J. E., Holloway V., Almeida R. J. Rapid recognition of methicillin-resistant Staphylococcus aureus by use of automated test systems. J Clin Microbiol. 1984 Sep;20(3):430–433. doi: 10.1128/jcm.20.3.430-433.1984. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lally R. T., Ederer M. N., Woolfrey B. F. Evaluation of the newly modified AutoMicrobic system gram-positive susceptibility-MIC card for detection of methicillin-resistant Staphylococcus aureus. J Clin Microbiol. 1986 Feb;23(2):387–387. doi: 10.1128/jcm.23.2.387-.1986. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stotler R. W., Meyer M. C. Detection of oxacillin-resistant staphylococci by the AutoMicrobic system. J Clin Microbiol. 1983 Nov;18(5):1205–1211. doi: 10.1128/jcm.18.5.1205-1211.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Thornsberry C., McDougal L. K. Successful use of broth microdilution in susceptibility tests for methicillin-resistant (heteroresistant) staphylococci. J Clin Microbiol. 1983 Nov;18(5):1084–1091. doi: 10.1128/jcm.18.5.1084-1091.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Woolfrey B. F., Lally R. T., Ederer M. N. Evaluation of the automicrobic system for detection of resistance of Staphylococcus aureus to methicillin. J Clin Microbiol. 1984 Apr;19(4):464–467. doi: 10.1128/jcm.19.4.464-467.1984. [DOI] [PMC free article] [PubMed] [Google Scholar]