Abstract
Four-dimensional (4D) radiotherapy is considered as a feasible and ideal solution to accommodate intra-fractional respiratory motion during conformal radiation therapy. With explicit inclusion of the temporal changes in anatomy during the imaging, planning, and delivery of radiotherapy, 4D treatment planning in principle provides better dose conformity. However, the clinical benefits of developing 4D treatment plans in terms of tumor control rate and normal tissue complication probability as compared to other treatment plans based on CT images of a fixed respiratory phase remains mostly unproven. The aim of our study is to comprehensively evaluate 4D treatment planning for nine lung tumor cases with both physical and biological measures using biologically effective uniform dose together with complication-free tumor control probability, P+. Based on the examined lung cancer patients and PTV margin applied, we found similar but not identical curves of DVH, and slightly different mean doses in tumor (up to 1.5%) and normal tissue in all cases when comparing 4D, P0%, and P50% plans. When it comes to biological evaluations, we did not observe definitively PTV size dependence in P+ among these nine lung cancer patients with various sizes of PTV. Moreover, it is not necessary that 4D plans would have better target coverage or higher P+ as compared to a fixed phase IMRT plan. However, on the contrary to significant deviations in P+ (up to 14.7%) observed if delivering the IMRT plan made at end-inhalation incorrectly at end-exhalation phase, we estimated the overall P+, PB, and PI for 4D composite plans that have accounted for intra-fractional respiratory motion.
Keywords: Four-dimensional treatment planning, Biologically effective uniform dose, Treatment planning, Radiobiological objectives, Organ motion
Introduction
With great advances in highly conformal treatment methodologies such as intensity modulated radiation therapy (IMRT) and stereotactic body radiation therapy (SBRT), the most significant problem for today's radiotherapy is management of internal organ motion, which occurs both inter-fractionally and intra-fractionally (1-6). The inter-fractional organ motion can be corrected promptly by aligning the patient position using the bony anatomy and/or internal fiducial markers, which is performed right before the daily treatment. Intra-fractional organ motion, on the other hand, still remains mostly unsolved. Respiratory motion, for example, is one of the most substantial intra-fraction organ motions that introduce undesired error during radiation therapy for lung cancer patients. In current radiation therapy, the practical solution accounting for respiratory motion is to add uniform, population based margins to the clinical target volume (CTV). Expended margins around the CTV, however, usually result in higher doses to normal tissues close to the target. Consequently, to accommodate respiratory motion with suitable protection to normal tissues around the target, four-dimensional (4D) radiotherapy (7-9) is considered as a feasible and ideal resolution.
The workflow of 4D radiotherapy includes 4D computed tomography (4DCT) acquisition and 4D treatment planning followed by 4D delivery. 4DCT are often obtained while the patient breaths freely and therefore the extent of respiratory motion and deformation is captured. To completely describe tumor movement and deformation from phase to phase, deformable image registration needs to be applied after 4DCT acquisition (10). Deformable image registration enables the mapping of dose from multiple anatomic maps at different instants of time to a common geometry. Then, similar treatment plans for different respiratory phases are created based on contours determined specifically for these phases. After the dose distributions of various respiratory phases are established by the treatment planning system, the final 4D composite dose distribution can be calculated by summing up all dose distributions and applying the proper weighting factors to discrepant phases. Although 4D treatment planning in principle provides more accurate dose calculation when organs move during treatment (10), the biological benefits (such as higher tumor control probability) of developing 4D treatment plans as compared to other treatment plans based on CT images of a fixed respiratory phase remains mostly unproven.
Biologically effective uniform dose () was defined as the uniform dose that causes the same tumor control probability or normal tissue complication rate as the actual dose given to the patient (11, 12). In contrast to other biological dose indices (e.g., equivalent uniform dose, EUD), calculation is not model dependent and can be applied to evaluate plans that have more than one target or organs at risk. can be used to estimate the effectiveness of radiation therapy in terms of benefit from tumor control and injury to normal tissues. Furthermore, the concept directly associates a dose with treatment outcome in terms of the complication-free tumor control probability (P+) (12).
The goal of this study is to comprehensively evaluate 4D treatment planning for thoracic tumors with both physical and biological measures using together with P+. Furthermore, comparisons between the 4D treatment plan and MLC-based IMRT plans for fixed respiratory phases (i.e., at end-inhalation and end-exhalation) were performed to estimate the benefit of 4D planning using a more clinical outcome related approach.
Materials and Methods
4D CT Image Acquisition
CT images for each patient were acquired by using a GE Lightspeed Qx/i 4-slice scanner (GE Medical systems, Milwaukee, WI) in axial cine mode. Patients were coached to breath freely and quietly. To generate spatial and temporal coincidence image data, patient breathing traces were recorded with the Varian Real-time Position Management (RPM) system (Varian Medical Systems, Palo Alto, CA) during CT scanning. In brief, the RPM system acquired respiratory motion information by monitoring the patient's abdominal surface motion (8, 9). Then, Advantage 4D software (GE Medical systems, Milwaukee, WI) was used to sort reconstructed axial images into ten respiratory phases according to RPM respiratory data (9). The sorted image sets were labeled by the percentage of the respiratory cycle, in which P0% corresponded to end-inhalation and P50% represented end-exhalation. In this study, we chose end-inhalation, P0%, as the reference phase for our 4D treatment plans to avoid motion artifacts on 4D CT images (8, 10).
Patients
Eight lung cancer patients were randomly selected as candidates for this study. All patients received a fundamental radiotherapy for lung cancer, and had 4DCT images acquired prior to treatment. For each patient, a radiation oncologist delineated the target and the organs at risk manually on each respiratory phase of the 4DCT data sets (i.e., from P0% to P90%, in total, 10 phases). The gross target volume (GTV) was expanded by 1.5 cm uniformly to form the planning target volume (PTV) (13). Table I shows for all the patients the gender, age, the exact treatment site, and the PTV volume. Among the nine cases we listed in Table I, Patient 9 represents an anonymised respiration-correlated 4DCT image set created for 4D related image researches (14). This set of 4DCT images is freely available for download (15).
Table I.
General information of nine lung cancer patients.
| Gender | Age | Treatment Site | PTV (cm3) | |
|---|---|---|---|---|
| Patient 1 | Male | 92 | Rt upper lobe | 50.8 |
| Patient 2 | Male | 74 | Rt lower lobe | 184.5 |
| Patient 3 | Male | 67 | Rt upper lobe | 657.6 |
| Patient 4 | Female | 77 | Lt upper & lower lobes | 125.1 |
| Patient 5 | Female | 71 | Rt hilum | 129.4 |
| Patient 6 | Male | 77 | Rt upper lobe | 391.7 |
| Patient 7 | Male | 67 | Lt upper & right upper lobes | 699.2 |
| Patient 8 | Female | 85 | Rt upper lobe | 235.1 |
| Patient 9 | N/A | N/A | Rt upper lobe | 50.9 |
Treatment Planning and Image Registration
For each of the nine patients, ten MLC-based IMRT plans were developed for different respiratory phases using the Philips (ADAC) Pinnacle3 treatment planning system (ADAC Laboratories, Milpitas, CA, v 8.0d). In other words, each lung cancer patient in this study had ten IMRT plans of fixed-phases (from P0% to P90%) plus a 4D composite plan made with P0% phase as the reference phase. The general prescription of all plans is to provide 60 Gy to at least 95% of the PTV. Dose constraints to organs at risk are listed in Table II. All plans were created independently, and were designed and optimized according to the same radiation oncologist's recommendations and dose constraints. In cases with larger PTVs, the dose to healthy lungs was of primary consideration. A 1.96 × 1.96 × 2.5 mm3 dose-grid resolution was used to compute dose within the whole volume of 3D image sets for each plan (for each respiratory phase).
Table II.
Dose constraints for organs at risk in IMRT plan optimization.
| Organs at risk | Dose constraint |
|---|---|
| Spinal cord | Maximum dose < 4500 cGy |
| Esophagus | D20% < 1500 cGy |
| Heart | D15% < 3000 cGy |
| Total Lung | D25% < 2000 cGy |
For the purpose of image registration, sorted 4DCT image sets of ten respiratory phases were post-processed using an in-house developed software, APT4D (16). In order to save memory space, these CT images were down-sampled by a factor of 2 to a size of 256 × 256 with a pixel resolution of 1.96 × 1.96 mm2, while keeping the cranial-dorsal resolution of 2.5 mm. The deformable image registration method adopted in APT4D was proposed by Thirion (17). The local deformable registrations from CT images of the moving phases to images of the reference phase (i.e., P0% in our study) were performed iteratively using the demons algorithm (17). As suggested from previous research (16), a total of 150 iterations were sufficient to get optimal convergence of the displacement between the fixed and moving images. From results of deformable registration, displacement maps of moving phases to the reference phase were recorded as deformation fields accordingly.
Deformation fields of image registrations for nine moving respiratory phases were then applied to transform 3D dose matrices of corresponding IMRT plans. To compose a final 4D dose distribution, registered 3D dose matrices of ten respiratory phases were averaged by applying equal weighting to each phase. For physical dosimetry evaluation, the 4D composite treatment plan of each patient was compared with the IMRT plans of P0% (reference phase in this study) and P50% in terms of dose statistics (i.e., maximum, mean, and minimum doses) and dose volume histogram (DVH) for the nine lung cases.
Biologically Effective Uniform Dose Evaluation
The uniform dose that causes the same tumor control probability or normal tissue complication rate as the actual dose given to the patient was evaluated using (11, 12). This general definition of can be expressed as the equation below:
| [1] |
The radiobiological model that was used to describe the dose-response relation of tumors and organs at risk was linear-quadratic-Poisson model (18, 19):
| [2] |
where the P(D) is the probability to control the tumor or induce a certain injury to an organ that is irradiated uniformly with a dose D. Since this model takes into account the fractionation effects that are introduced by the irradiation schedule, d, (equals to D/n) is the dose per fraction, and n is the number of fractions. D50 is the dose which gives a response probability of 50% and γ is the maximum normalized value of the dose-response gradient. Variables α and β are the fractionation parameters of the model and account for the early and late effects expected, respectively. The dose-response parameters of the target and organs at risk used in this study were listed in Table III (20-27).
Table III.
Dose-response parameters used in biological dosimetry evaluation.
| D50 (Gy) | γ | s | Endpoint | |
|---|---|---|---|---|
| Spinal cord | 57.00 | 6.70 | 1.00 | Myelitis necrosis |
| Esophagus | 68.00 | 2.8 | 3.40 | Clinical stricture / perforation |
| Heart | 70.70 | 0.96 | 1.00 | Late excess cardiac mortality |
| Lung | 15.90 | 1.00 | 0.17 | Sever radiation pneumonitis, fibrosis |
| PTV | 40.00 | 3.00 | N/A | Control |
To evaluate the biological effectiveness of the 4D composite plan, the concept of P+ and were used (12). P+ is the probability of achieving tumor control without causing complication to normal tissues. In addition, the probability of obtaining benefit from 4D treatments (i.e., tumor control) is denoted as PB, whereas the probability of causing severe injury to normal tissue is denoted as PI. In this study, the main objective of using physical dose indices (e.g., DVHs) for plan evaluation was to estimate the normal tissue tolerance against the optimal target dose needed. The actual probabilities of benefit and injury as well as P+ were carried out using an in-house developed software. Moreover, comprehensive comparisons of the composite 4D IMRT plan against the IMRT plans of fixed phase, P0% and of P50%, were conducted physically and biologically.
Results
It should be noted that 4D composite plans, P0% and P50% plans compared physically and biologically in this study were all MLC-based IMRT plans.
Physical Dose Analysis: DVH
Figure 1 demonstrates the cumulative DVHs of three representative patients with small (patient 1), medium (patient 8), and large (patient 3) PTV sizes, respectively. Each set of DVH curves of the 4D composite plans and P50% plans were superimposed with the results of the reference-phase, i.e. P0%, plans to compare target coverage and normal tissue sparing among these three treatment plans. As shown in Figure 1a, PTV coverage of the P50% plan was inferior compared to both 4D composite and P0% plans while keeping the same protection to normal tissues as the other two. As a result, a part of the PTV volume might be missed if delivering reference-phase plan at other moving respiratory phases. On the other hand, the DVH of the 4D composite plan showed the exact PTV coverage, which is slightly substandard to the reference-phase plan, if we can deliver this 4D composite plan perfectly. The suboptimal DVH curve in PTV of the 4D plan reflected dose perturbations due to respiration as a function of time. In contrast with Figure 1a, Figure 1b illustrated different sparing of the organ at risk, heart, in these three IMRT plans when they had the same optimal PTV coverage. The results in Figure 1b show motion induced effects on the discrepant doses of critical organ(s) proximal to the target during radiation treatments. As shown in Figure 1c, nevertheless, there was no significant difference found in the DVHs of 4D plan, P0% and P50% plans from the patient with large PTV size.
Figure 1.
Cumulative DVHs in PTV (green), healthy lung (red), esophagus (orange), heart (pink), and spinal cord (cyan) of the 4D composite plan (square), and IMRT plans of P0% (solid line) and P50% (dashed line). Results of three patients with (a) small, (b) medium, and (c) large size of PTVs were presented and compared.
Physical Dose Statistics
Tables IV-VIII summarizes mean, maximum, minimum doses, and standard deviations (SD) in PTV, spinal cord, esophagus, heart, and healthy lung, respectively. Mean doses in PTV of these nine lung patients as shown in Table IV fulfilled prescription dose with some differences among the 4D composite plan, P0% and P50% IMRT plans. The dose variety in the target volume as indicated by the SD in Table IV was quite comparable between the 4D plan and reference-phase, P0% plan, but was slightly larger in some cases of comparisons between the 4D plan and P50% plan (e.g., in patient 6).
Table IV.
Physical dose statistics in PTV among the 4D composite plan, P0% and P50% IMRT plans.
| Patient 1 | Patient 2 | Patient 3 | Patient 4 | Patient 5 | Patient 6 | Patient 7 | Patient 8 | Patient 9 | |
|---|---|---|---|---|---|---|---|---|---|
| Mean Dose (Gy) | |||||||||
| 4D (Gy) | 59.7 | 59.8 | 59.9 | 59.6 | 59.8 | 59.7 | 59.8 | 59.7 | 59.7 |
| P0% (Gy) | 60.2 | 59.6 | 59.8 | 59.8 | 60.0 | 59.9 | 59.7 | 60.0 | 59.7 |
| P50% (Gy) | 59.5 | 58.7 | 59.7 | 59.2 | 59.7 | 58.0 | 58.3 | 59.6 | 59.5 |
| Standard Deviation (Gy) | |||||||||
| 4D (Gy) | 0.8 | 1.0 | 0.6 | 1.2 | 1.0 | 1.5 | 1.3 | 0.9 | 0.5 |
| P0% (Gy) | 0.7 | 1.6 | 0.6 | 0.8 | 0.7 | 1.2 | 1.4 | 1.3 | 0.4 |
| P50% (Gy) | 1.6 | 1.3 | 0.8 | 2.7 | 1.5 | 4.7 | 1.3 | 1.4 | 0.5 |
| Maximum Dose (Gy) | |||||||||
| 4D (Gy) | 61.5 | 63.1 | 62.9 | 62.9 | 64.9 | 65.4 | 65.6 | 68.6 | 62.1 |
| P0% (Gy) | 64.8 | 65.1 | 64.2 | 64.5 | 65.2 | 67.1 | 64.5 | 71.2 | 61.8 |
| P50% (Gy) | 65.2 | 63.4 | 63.3 | 63.6 | 65.8 | 63.4 | 65.2 | 69.0 | 62.1 |
| Minimum Dose (Gy) | |||||||||
| 4D (Gy) | 51.6 | 49.0 | 53.8 | 45.0 | 50.1 | 36.3 | 43.4 | 47.1 | 53.0 |
| P0% (Gy) | 56.2 | 44.7 | 55.8 | 52.8 | 54.1 | 40.7 | 45.7 | 46.4 | 57.6 |
| P50% (Gy) | 42.3 | 42.1 | 46.4 | 26.1 | 38.7 | 25.8 | 42.5 | 39.2 | 51.7 |
Table VIII.
Physical dose statistics in healthy lung among the 4D composite plan, P0% and P50% IMRT plans.
| Patient 1 | Patient 2 | Patient 3 | Patient 4 | Patient 5 | Patient 6 | Patient 7 | Patient 8 | Patient 9 | |
|---|---|---|---|---|---|---|---|---|---|
| Mean Dose (Gy) | |||||||||
| 4D (Gy) | 10.2 | 16.7 | 9.9 | 9.1 | 12.8 | 12.3 | 18.8 | 13.4 | 8.4 |
| P0% (Gy) | 10.1 | 15.2 | 9.2 | 9.4 | 12.8 | 11.9 | 18.4 | 12.9 | 8.6 |
| P50% (Gy) | 10.0 | 17.6 | 10.0 | 8.8 | 12.8 | 11.7 | 18.7 | 13.5 | 9.3 |
| Standard Deviation (Gy) | |||||||||
| 4D (Gy) | 15.5 | 13.9 | 19.0 | 12.9 | 19.3 | 19.4 | 18.9 | 18.6 | 14.9 |
| P0% (Gy) | 15.4 | 12.6 | 18.3 | 13.2 | 19.5 | 19.3 | 18.5 | 18.9 | 14.7 |
| P50% (Gy) | 15.5 | 15.2 | 19.1 | 12.8 | 19.4 | 18.1 | 18.8 | 18.8 | 15.1 |
| Maximum Dose (Gy) | |||||||||
| 4D (Gy) | 61.5 | 60.8 | 62.9 | 62.9 | 64.9 | 65.4 | 62.6 | 68.6 | 63.1 |
| P0% (Gy) | 65.5 | 60.5 | 64.2 | 64.5 | 65.5 | 67.1 | 65.2 | 71.2 | 64.0 |
| P50% (Gy) | 65.2 | 62.8 | 63.6 | 62.9 | 65.8 | 63.4 | 62.5 | 69.0 | 64.1 |
| Minimum Dose (Gy) | |||||||||
| 4D (Gy) | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.4 | 0.3 |
| P0% (Gy) | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.4 | 0.3 |
| P50% (Gy) | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.4 | 0.3 |
In general, the physical dose statistics of the 4D composite plan was similar to that of reference-phase IMRT plan in most of these nine cases. The physical dose results of the 4D plans lied between the P0% IMRT plan and the P50% IMRT plan in most patients. However, comparing the doses of the P50% plans with those of P0%, some differences were found in mean (up to 11 Gy in heart, patient 8), maximum (up to 35 Gy in esophagus, patient 8), and minimum (up to 26.7 Gy in PTV, patient 4) doses as well as in SD (up to 10.3 Gy in esophagus, patient 8) of the target and normal tissues.
Biological dose evaluation:
Figure 2 shows biological evaluation in PTV using tumor control probability (TCP) as well as compared with results of physical dose index, mean dose, for the 4D composite plan, P0% and P50% IMRT plans. Based on mean dose and dose distribution in the PTV along with biological parameters acquired from previous researches, the corresponding uniform dose in terms of as well as the TCP could be determined in these lung cancer patients for the different IMRT plans (12). From the results indicated in Figure 2a, the TCPs of these nine patients were relatively high for the dose prescription, which was used in these treatment plans. Comparing the TCPs of the 4D plan with those of the P0% and P50% plans, there was no major difference found for most of the patients. For patient 4 and patient 6, the TCPs of P50% plans were about 4% and 6%, respectively, lower than the TCPs of the 4D and P0% plans. These lower TCPs of P50% IMRT plans might be attributed to larger dose variations in the PTVs.
Figure 2.
Biological dose evaluations in PTV for the 4D composite plans, P0% IMRT plans, and P50% IMRT plans of the nine lung cancer patients. Comparisons of a) tumor control probability and (b) biologically effective uniform dose with (c) physical dose index, mean dose, were illustrated.
Figures 3-6 illustrates the biological evaluation in spinal cord, esophagus, heart, and healthy lung, respectively, using normal tissue complication probability (NTCP) as well as pared with results of physical dose index, mean dose, for the 4D composite plan, P0% and P50% IMRT plans. From the results indicated in Figure 3, although mean dose in spinal cord varied from patient to patient, NTCP of all patients were approximately zero calculated with the endpoint of myelitis necrosis in our study. Deviations in mean (and max) doses of the 4D, P0% and P50% IMRT plans resulted in the same and thus the same complication rate (i.e., ~0% in all cases) in spinal cord.
Figure 3.
Biological dose analyses in spinal cord for the 4D composite plans, P0% IMRT plans, and P50% IMRT plans of the nine lung cancer patients. Comparisons of (a) biologically effective uniform dose with (b) mean dose in spinal cord were illustrated. Normal tissue probabilities (not shown in this figure) of all nine patients were approximately 0.
Figure 6.
Biological dose analyses in healthy lung for the 4D composite plans, P0% IMRT plans, and P50% IMRT plans of the nine lung cancer patients. Comparisons of (a) normal tissue complication probability and (b) biologically effective uniform dose with (c) mean dose in healthy lung were illustrated.
In Figure 4, it is shown that all plans of patient 2 had much higher mean doses in esophagus (up to 35 Gy difference) as compared to other patients. The higher mean dose in esophagus was because the location of tumor in patient 2 was right next to the esophagus, resulting in up to 10 Gy higher Therefore, patient 2 had much higher NTCP (up to 15.7%) than the other patients. When compared NTCP among the 4D, P0% and P50% plans, there was no significant difference discovered for all patients.
Figure 4.
Biological dose analyses in esophagus for the 4D composite plans, P0% IMRT plans, and P50% IMRT plans of the nine lung cancer patients. Comparisons of (a) normal tissue complication probability and (b) biologically effective uniform dose with (c) mean dose in esophagus were illustrated.
Figure 5 demonstrates that patient 2, 6, and 8 had higher mean doses in heart compared to other patients. Patient 8, however, had less deviation in the volume of heart (also from results of Table VI), contributing to less NTCP and thus lower As compared 4D and P50% plans to P0% plan of patient 8, up to 11 Gy difference in mean doses resulted in up to 0.7% deviation, which is negligible, in NTCP.
Figure 5.
Biological dose analyses in heart for the 4D composite plans, P0% IMRT plans, and P50% IMRT plans of the nine lung cancer patients. Comparisons of (a) normal tissue complication probability and (b) biologically effective uniform dose with (c) mean dose in heart were illustrated.
Table VI.
Physical dose statistics in esophagus among the 4D composite plan, P0% and P50% IMRT plans.
| Patient 1 | Patient 2 | Patient 3 | Patient 4 | Patient 5 | Patient 6 | Patient 7 | Patient 8 | Patient 9 | |
|---|---|---|---|---|---|---|---|---|---|
| Mean Dose (Gy) | |||||||||
| 4D (Gy) | 4.6 | 41.2 | 5.7 | 12.3 | 8.5 | 11.2 | 9.8 | 9.9 | 4.1 |
| P0% (Gy) | 4.3 | 41.0 | 6.0 | 12.3 | 8.3 | 11.0 | 9.0 | 4.6 | 4.3 |
| P50% (Gy) | 4.8 | 43.2 | 5.5 | 13.0 | 7.2 | 13.2 | 9.5 | 11.9 | 4.3 |
| Standard Deviation (Gy) | |||||||||
| 4D (Gy) | 4.9 | 21.7 | 8.6 | 13.7 | 8.6 | 8.9 | 9.2 | 11.6 | 5.6 |
| P0% (Gy) | 4.7 | 22.1 | 9.1 | 14.6 | 8.2 | 10.1 | 9.7 | 4.6 | 5.9 |
| P50% (Gy) | 5.4 | 21.5 | 8.2 | 14.8 | 8.4 | 9.8 | 8.8 | 14.9 | 5.7 |
| Maximum Dose (Gy) | |||||||||
| 4D (Gy) | 17.9 | 61.8 | 44.8 | 58.4 | 49.4 | 57.6 | 40.4 | 40.6 | 22.6 |
| P0% (Gy) | 18.9 | 62.8 | 43.1 | 59.7 | 51.4 | 61.3 | 45.4 | 16.9 | 24.3 |
| P50% (Gy) | 18.9 | 62.4 | 42.8 | 58.7 | 49.3 | 58.6 | 40.8 | 52.8 | 22.0 |
| Minimum Dose (Gy) | |||||||||
| 4D (Gy) | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 1.0 | 0.3 | 0.4 | 0.3 |
| P0% (Gy) | 0.3 | 1.0 | 0.3 | 0.3 | 0.3 | 1.0 | 0.3 | 0.4 | 0.3 |
| P50% (Gy) | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 1.0 | 0.3 | 0.4 | 0.3 |
From the results shown in Figure 6, patient 7 had much higher mean doses (up to 9 Gy higher) in 4D, P0% and P50% IMRT plans as compared to other patients, resulting in a high and, furthermore, up to 50% higher NTCP. This much higher NTCP observed in lung of patient 7 was related to the biological parameters of the early radiological pulmonary complications we applied in this study (26, 27).
Biological analyses of the 4D, P0%, and P50% IMRT plans in small (patient 1), medium (patient 8), and large (patient 3) PTV using complication-free tumor control probability, P+, are illustrated in Figure 7. Based on the physical measures using DVH shown in Figure 1 and the results of Tables IV-VIII plus the dose-response probabilities for the target and normal tissues demonstrated in Figures 2-6, Figure 7 illustrates comparisons of overall clinical effectiveness among the 4D plan, P0% and P50% IMRT plans. In this figure, in addition to P+, we calculated the probability of obtaining benefit from radiation treatments (i.e., tumor control), PB, as well as the probability of causing injury to normal tissues, PI, for the 4D, P0%, and P50% IMRT plans. It should be noted that the on the X axis of Figure 7 represents the biologically effective uniform dose calculated based on the radiological characteristics of the target (usually denoted as When using on the dose axis, the dose-response curves are normalized to forcing the response curves of PB of these three plans coincide (12). As a result, the plan, which has the highest P+ and lowest PI as shown in Figure 7 would be considered superior to the rest of plans. Figure 7a shows results of patient 1 with small PTV. From the DVH diagram shown in Figure 1a, it seemed that P0% plan had better target coverage than the other two. The DVHs of normal tissues in these three plans, on the other hand, were all very close. When evaluating the P+ of the 4D, P0% and P50% plans for patient 1, it turned out that P50% IMRT plan was slightly superior compared to the other plans, because of the lowest PI value among these three plans. For patient 8 with medium PTV, all three plans had very close mean doses (4D, 59.7Gy; P0%, 60.0 Gy; P50%, 59.6 Gy) from the results in Table IV. However, as illustrated in Figure 1b and Table VII, the doses in heart from three plans varied considerably (mean dose for 4D, 14.6 Gy; P0%, 6.7 Gy; and P50%, 17.7 Gy). Therefore, the results in Figure 7b showed a lowest P+ with highest PI in P50% IMRT plan. Figure 7c demonstrates the results of patient 3 with large PTV. From Figure 1c, the DVH curves of the target and normal tissues for the three plans were very close to each other. In Figure 7c, however, P0% was the best plan with the highest P+ and lowest PI. While the physical measures of the three dose distributions were similar, the substantial changes in P+ and BEUD indicated significant differences in the clinical effectiveness of these three plans.
Figure 7.
Biological evaluations on the 4D composite plan, P0% and P50% IMRT plans using PB, PI as well as P+. Results of the same three patients as in Figure 1 with (a) small, (b) medium, and (c) large size of PTVs were presented and compared.
Table VII.
Physical dose statistics in heart among the 4D composite plan, P0% and P50% IMRT plans.
| Patient 1 | Patient 2 | Patient 3 | Patient 4 | Patient 5 | Patient 6 | Patient 7 | Patient 8 | Patient 9 | |
|---|---|---|---|---|---|---|---|---|---|
| Mean Dose (Gy) | |||||||||
| 4D (Gy) | 5.5 | 17.4 | 0.2 | 6.3 | 6.4 | 14.8 | 5.6 | 14.6 | 1.3 |
| P0% (Gy) | 5.1 | 16.7 | 0.2 | 7.4 | 5.8 | 14.9 | 4.9 | 6.7 | 1.6 |
| P50% (Gy) | 5.8 | 17.2 | 0.2 | 5.1 | 6.1 | 12.6 | 5.4 | 17.7 | 1.0 |
| Standard Deviation (Gy) | |||||||||
| 4D (Gy) | 8.2 | 14.6 | 0.2 | 6.5 | 10.0 | 15.2 | 7.7 | 8.0 | 2.0 |
| P0% (Gy) | 7.7 | 14.2 | 0.2 | 7.7 | 9.0 | 15.5 | 6.9 | 3.8 | 2.6 |
| P50% (Gy) | 8.5 | 14.5 | 0.2 | 5.8 | 9.6 | 13.3 | 7.6 | 9.5 | 1.7 |
| Maximum Dose (Gy) | |||||||||
| 4D (Gy) | 51.6 | 61.8 | 0.6 | 44.3 | 60.5 | 63.0 | 48.7 | 46.7 | 18.8 |
| P0% (Gy) | 53.3 | 63.1 | 0.6 | 52.8 | 61.5 | 64.0 | 46.7 | 24.4 | 19.7 |
| P50% (Gy) | 52.3 | 63.1 | 1 | 37.5 | 62.4 | 62 | 49.7 | 55.7 | 20.4 |
| Minimum Dose (Gy) | |||||||||
| 4D (Gy) | 0.3 | 0.7 | 0.3 | 0.3 | 0.3 | 0.7 | 0.3 | 1.4 | 0.3 |
| P0% (Gy) | 0.3 | 0.7 | 0.3 | 0.3 | 0.3 | 0.7 | 0.3 | 0.7 | 0.3 |
| P50% (Gy) | 0.3 | 1.0 | 0.3 | 0.3 | 0.3 | 0.7 | 0.3 | 1.8 | 0.3 |
Discussion
The present study aims to clinically interpret the value of 4D radiation therapy using biological measures in addition to physical dose statistics. The biologically effective uniform dose that we used in this study converted the information of isodose distribution and DVH into more clinically relevant outcomes, expected benefit of tumor control and possible injury to normal tissues. Comparisons among 4D treatment plan and fixed-phase IMRT plans at end-inhalation (P0%) and end-exhalation (P50%) using allowed us to predict the clinical effectiveness owing to respiratory motion induced dose uncertainty.
Several studies have accurately estimated the effect of motion on lung treatments delivered using conventional imaging and planning (2,3, 28). Additionally, investigations have also been done to show improvements achieved when 4DCT and thus 4D treatment planning was incorporated into treatment procedures (5, 8-10, 29). Flampouri et al. (29) has suggested that 4D treatment planning should be performed for patients with tumor motion larger than 12 mm and/or with severe image artifacts in free breathing helical CT. Figure 8 demonstrates the observed volumetric variations of manually contoured PTV for the lung tumor at different respiratory phases for the nine lung patients. The results in this figure implied the rigid and non-rigid tumor motion from phase to phase for these lung cancer patients. Patient 6, for example, had large volume deviation from P0% to P50% (25% difference in volume size). From our physical results of Table IV, the mean dose differences among the 4D, P0% and P50% plans were not significant (59.7 ± 1.5 Gy, 59.9 ± 1.2 Gy, and 58.0 ± 4.7 Gy, respectively). From the results of Figure 2, however, the of the 4D, P0%, P50% plans were significantly different (59.1 Gy, 59.7 Gy, and 51.8 Gy, respectively), resulting in 98.8%, 98.9%, and 93.0% tumor control rates, respectively.
Figure 8.
PTV volume changes in lung tumors of nine patients obtained by manual contouring on 4DCT images.
Variations in P+, PB, and PI among the 4D composite plan, P0% and P50% IMRT plans were summarized in Table IX for the nine lung cancer patients. In this table, the deviation amongst the 4D plan, P0% and P50% IMRT plans is shown quantitatively. Patient 2 among these nine cases had the most significant differences in terms of complication-free tumor control rate when comparing these three plans. From the results of physical dose statistics, the mean doses of the 4D plan, P0% plan and P50% plan in the target (59.8 ± 1.0 Gy, 59.6 ± 1.6 Gy, and 58.7 ± 1.3 Gy, respectively) and normal tissues (e.g., in esophagus: 41.2 ± 21.7 Gy, 41.0 ± 22.1 Gy, and 43.2 ± 21.5 Gy, respectively) were very close. Although physically the three dose distributions were similar, the discrepancies of P+ (8.2%, -6.5%, -14.7% between 4D and P0%, 4D and P50%, and P0% and P50%, respectively ) indicated in Table IX demonstrated that their effectiveness differs significantly.
Table IX.
Deviations of the quantities in P+, PB, PI, and both calculated in terms of benefit and injury among the 4D plan, P0% and P50% IMRT plans.
| Patient 1 | Patient 2 | Patient 3 | Patient 4 | Patient 5 | Patient 6 | Patient 7 | Patient 8 | Patient 9 | |
|---|---|---|---|---|---|---|---|---|---|
| ΔP+ (%) | |||||||||
| P0-4D | 0.70 | 8.20 | 4.20 | -1.90 | -0.30 | 1.20 | 2.60 | 1.50 | 0.70 |
| P50-4D | 0.60 | -6.50 | -1.00 | -3.50 | 0.90 | 0.60 | 0.10 | -1.30 | -0.90 |
| P50-P0 | -0.10 | -14.70 | -5.20 | -1.60 | 1.20 | -0.60 | -2.50 | -2.80 | -1.60 |
| ΔPB (%) | |||||||||
| P0-4D | 0.20 | -0.10 | 0.00 | 0.20 | 0.10 | 0.10 | 0.00 | 0.10 | 0.00 |
| P50-4D | -0.20 | 0.00 | -0.10 | -3.90 | -0.10 | -5.80 | -0.50 | -0.10 | 0.00 |
| P50-P0 | -0.40 | 0.10 | -0.10 | -4.10 | -0.20 | -5.90 | -0.50 | -0.20 | 0.00 |
| ΔPI (%) | |||||||||
| P0-4D | -0.60 | -8.20 | -4.20 | 1.90 | 0.40 | -1.00 | -2.70 | -1.50 | -0.70 |
| P50-4D | -0.80 | 6.50 | 0.90 | -0.50 | 0.10 | -6.50 | -0.60 | 1.20 | 0.90 |
| P50-P0 | -0.20 | 14.70 | 5.10 | -2.40 | -0.30 | -5.50 | 2.10 | 2.70 | 1.60 |
| ΔDB (Gy) | |||||||||
| P0-4D | 0.50 | -0.40 | -0.10 | 0.45 | 0.30 | 0.55 | -0.20 | 0.20 | 0.00 |
| P50-4D | -0.70 | -0.20 | -0.20 | -6.10 | -0.45 | -7.35 | -1.60 | -0.50 | -0.20 |
| P50-P0 | -1.20 | 0.20 | -0.10 | -6.55 | -0.75 | -7.90 | -1.40 | -0.70 | -0.20 |
| ΔD1 (Gy) | |||||||||
| P0-4D | -0.40 | -1.35 | -0.80 | 0.60 | 0.10 | -0.15 | -0.50 | -0.25 | -0.20 |
| P50-4D | -0.45 | 1.10 | 0.15 | -0.15 | 0.05 | -1.05 | -0.10 | 0.20 | 0.25 |
| P50-P0 | -0.05 | 2.45 | 0.95 | -0.75 | -0.05 | -0.90 | 0.40 | 0.45 | 0.45 |
P+, complication-free tumor control probability; PB, tumor control probability; PI, normal tissue complication probability; , biologically effective uniform dose calculated based on the radiological characteristics of the target; , biologically effective uniform dose calculated based on the radiological characteristics of normal tissues.
Major uncertainties involved in the workflow of the 4D treatment planning in this study include variations in contouring of the target (and its margin determination) and organs at risk on 4DCT images, which suffer less from artifact but not artifact-free (29), as well as errors in deformation image registration perform on moving phase images registered to images at reference phase (9). For each of the nine patients in our study, who had ten respiratory phase 4DCT image sets, the GTV and normal tissues were manually delineated and determined by the same oncologist one phase following the other. Contours of normal tissue especially at the boundaries that are generally plagued with motion artifacts, such as the boundary between lung tissue and diaphragm were evaluated by the same physician to minimize errors in delineation. A uniform margin of 1.5 cm around GTV we used for all plans in all nine patients was suggested by Flampouri et al. (29). They concluded that no significant improvement on tumor coverage with 0.5 cm increase on the PTV margin (29). As for errors in deformable image registration, to our best knowledge, there is no perfect image registration algorithm for all anatomical situations. Every deformable registration algorithm has tradeoffs between the accuracy and the computation efficiency. From our previous studies, the deformable image registration algorithm we adapted in the software, APT4D, performed non-rigid registration adequately (16). For lung cancer patients, correlation coefficient calculated after image registration of end-inhalation to end-exhalation 4DCT volumes ranged from 0.992 to 0.999 (16). Concerning variations in the evaluation, the calculation of dose-response probabilities of the tumors and normal tissues was based on recently derived data (26, 27). The fact is that the radiobiological parameters used by different models are now more reliable as compared with similar data in the past (30).
Based on the lung cancer patients that were chosen and the PTV margin that was applied, we found similar but not identical curves of DVH, and slightly different mean doses in tumor and normal tissue in all cases when comparing 4D, P0% and P50% plans. When it comes to biological evaluations, Table IX summarizes substantial deviations in P+ (in patient 2, 3, and 4; up to 14.7%), PB (in patient 4 and 6; up to 5.9%), as well as in PI (in patient 2, 3 and 6; up to 14.7%) among 4D, P0% and P50% plans. From nine lung patients' results in our study, based on both physical and biological analyses for a 4D composite plan, and IMRT plans for fixed respiratory phases (at P0% and P50%), it is not certain that 4D plans have better target coverage or higher expected tumor control rate without severe complication to normal tissues. Nevertheless, from the data of Table IX, significant variations (up to 14.7% difference) in P+ from P0% plan to P50% were found in some patients. These results explained the value of 4D radiation therapy for a consistent P+ through the whole course of the treatment if we are able to deliver 4D treatment plans precisely. In other word, ΔP+, ΔPB, and ΔPI during 4D treatment plan delivery would be 0 in contrast to the deviations we estimated from P0% to P50% for these nine patients in Table IX. The accuracy of 4D treatment plan delivery, which depends mainly on the reproducibility of the breathing cycle acquired at the simulation in latter actual 4D treatment was not examined in this study.
The possible reason that no absolutely benefit in complication-free tumor control probability in our assessments for 4D treatment planning might be because these 4D treatment plans were optimized based on the physical dose constraints rather than biological limitations. When converting the optimized physical dose distribution into biological dose for 4D composite plans, resulting complication-free tumor control probability might not necessarily be in the optimal level as observed in our results. Consequently, it might be essential to implement biological optimization in terms of optimal complication-free tumor control rate for 4D treatment planning of lung cancer patients in future studies.
Conclusions
In this study, by applying both physical and biological evaluations, the 4D radiotherapy plan has been evaluated thoroughly with the consideration of tumor control probability and normal tissue injury rate, which appear to be more related to clinical treatment outcomes. When comparing the 4D composite plans with fixed-phase IMRT plans using physical measures alone, it was hard to decide and conclude which plan would be superior. After taking into account the biological characteristics of the target and normal tissues as demonstrated, the complication-free tumor control probability, P+, determined which plan was better.
Because this study was based on nine patients, it was difficult to make generalized conclusions. From the physical measures of DVH and dose statistics, 4D composite plans provided adequate target coverage and mean dose in PTV and similar sparing to organs at risk as compared to the IMRT plan at reference phase (P0% in our study). From the biological results, P+ of the 4D composite plans for nine patients differed from case to case due to differences in PB and/or in PI. Technical challenges in 4D treatment planning include requirements for 10-20 times more image space and the associated increase in image processing, registration, and segmentation. Consequently, the use of radiobiological parameters is necessary to justify if 4D treatment planning would increase tumor control rates and/or reduce normal tissue complication.
In our study, we did not observe definitively PTV size dependence in P+ among these nine lung cancer patients with PTVs of various absolute sizes. When evaluating the clinical benefit by comparing and P+ with fixed-phase IMRT plans of P0% and P50%, it is not certain that 4D plans had highest expected tumor control probability or lowest normal tissue complication rate. However, if 4D treatment plans can be delivered perfectly, the variations in P+, PB, and PI because of intra-fractional respiratory motion should be close to 0 in contrast to significant deviations observed if delivering the P0% IMRT plan at P50% respiratory phase.
Table V.
Physical dose statistics in spinal cord among the 4D composite plan, P0% and P50% IMRT plans.
| Patient 1 | Patient 2 | Patient 3 | Patient 4 | Patient 5 | Patient 6 | Patient 7 | Patient 8 | Patient 9 | |
|---|---|---|---|---|---|---|---|---|---|
| Mean Dose (Gy) | |||||||||
| 4D (Gy) | 1.7 | 15.3 | 6.0 | 5.5 | 5.5 | 10.8 | 5.8 | 2.2 | 3.1 |
| P0% (Gy) | 1.7 | 15.1 | 6.1 | 5.5 | 5.0 | 9.9 | 5.8 | 2.3 | 3.4 |
| P50% (Gy) | 1.7 | 15.2 | 5.9 | 5.7 | 5.5 | 11.2 | 5.5 | 2.4 | 3.1 |
| Standard Deviation (Gy) | |||||||||
| 4D (Gy) | 4.1 | 13.1 | 8.7 | 7.7 | 9.0 | 12.5 | 8.2 | 4.2 | 4.6 |
| P0% (Gy) | 4.3 | 13.2 | 8.8 | 8.0 | 8.1 | 11.8 | 8.4 | 4.4 | 5.1 |
| P50% (Gy) | 4.3 | 13.0 | 8.6 | 8.0 | 9.5 | 13.0 | 8.2 | 4.6 | 4.6 |
| Maximum Dose (Gy) | |||||||||
| 4D (Gy) | 18.5 | 37.8 | 37.3 | 39.1 | 32.6 | 39.6 | 39.1 | 15.5 | 15.2 |
| P0% (Gy) | 20.8 | 36.8 | 35.4 | 39.1 | 29.9 | 41.7 | 41.8 | 17.6 | 17.5 |
| P50% (Gy) | 19.8 | 37.8 | 36.0 | 44.3 | 34.3 | 39.3 | 39.8 | 18.0 | 15.5 |
| Minimum Dose (Gy) | |||||||||
| 4D (Gy) | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.4 | 0.3 |
| P0% (Gy) | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.4 | 0.3 |
| P50% (Gy) | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.4 | 0.3 |
Acknowledgments
Acknowledgements
This project is sponsored by National Institutes of Health/National Library of Medicine grant (1R01LM009362-01).
Abbreviations
- IMRT
Intensity modulated radiation therapy
- SBRT
Stereotactic body radiation therapy
- 4D
Four-dimensional
- 4DCT
4D computed tomography
- BEUD
Biologically effective uniform dose
- RPM
Real-time Position Management
- P0%
Respiratory phase at end-inhalation
- P50%
respiratory phase at end-exhalation
- GTV
Gross target volume
- PTV
Planning target volume
- DVH
Dose volume histogram
- SD
Standard deviation
- TCP
Tumor control probability
- NTCP
Normal tissue complication probability
- P+
Complication-free tumor control probability
- PB
The probability of obtaining benefit
- PI
The probability of causing severe injury to normal tissue
Biologically effective uniform dose calculated based on the radiological characteristics of the target
Biologically effective uniform dose calculated based on the radiological characteristics of normal tissues
References
- 1.Ohara K, Okumura T, Akisada M, Inada T, Mori T, Yokota H, Calaguas MJ. Irradiation synchronized with respiration gate. Int J Radiat Oncol Biol Phys. 1989;17:853–857. doi: 10.1016/0360-3016(89)90078-3. [DOI] [PubMed] [Google Scholar]
- 2.Lujan AE, Larsen EW, Balter JM, Ten Haken RK. A method for incorporating organ motion due to breathing into 3D dose calculations. Med Phys. 1999;26:715–720. doi: 10.1118/1.598577. [DOI] [PubMed] [Google Scholar]
- 3.Chui CS, Yorke E, Hong L. The effects of intra-fraction organ motion on the delivery of intensity-modulated field with a multileaf collimator. Med Phys. 2003;30:1736–1746. doi: 10.1118/1.1578771. [DOI] [PubMed] [Google Scholar]
- 4.George R, Keall PJ, Kini VR, Vedam SS, Siebers JV, Wu Q, Lauterbach MH, Arthur DW, Mohan R. Quantifying the effect of intrafraction motion during breast IMRT planning and dose delivery. Med Phys. 2003;30:552–562. doi: 10.1118/1.1543151. [DOI] [PubMed] [Google Scholar]
- 5.Vedam S, Docef A, Fix M, Murphy M, Keall P. Dosimetric impact of geometric errors due to respiratory motion prediction on dynamic multileaf collimator-based four-dimensional radiation delivery. Med Phys. 2005;32:1607–1620. doi: 10.1118/1.1915017. [DOI] [PubMed] [Google Scholar]
- 6.Shirato H, Shimizu S, Kitamura K, Onimaru R. Organ motion in image-guided radiotherapy: Lessons from real-time tumor-tracking radiotherapy. Int J Clin Oncol. 2007;12:8–16. doi: 10.1007/s10147-006-0633-y. [DOI] [PubMed] [Google Scholar]
- 7.Keall P. 4-dimensional computed tomography imaging and treatment planning. Semin Radiat Oncol. 2004;14:81–90. doi: 10.1053/j.semradonc.2003.10.006. [DOI] [PubMed] [Google Scholar]
- 8.Keall PJ, Joshi S, Vedam SS, Siebers JV, Kini VR, Mohan R. Four-dimensional radiotherapy planning for DMLC-based respiratory motion tracking. Med Phys. 2005;32:942–951. doi: 10.1118/1.1879152. [DOI] [PubMed] [Google Scholar]
- 9.Rietzel E, Chen GT, Choi NC, Willet CG. Four-dimensional image-based treatment planning: Target volume segmentation and dose calculation in the presence of respiratory motion. Int J Radiat Oncol Biol Phys. 2005;61:1535–1550. doi: 10.1016/j.ijrobp.2004.11.037. [DOI] [PubMed] [Google Scholar]
- 10.Keall P. 4-dimensional computed tomography imaging and treatment planning. Semin Radiat Oncol. 2004;14:81–90. doi: 10.1053/j.semradonc.2003.10.006. [DOI] [PubMed] [Google Scholar]
- 11.Mavroidis P, Lind BK, Van Dijk J, Koedooder K, De Neve W, De Wagter C, Planskoy B, Rosenwald JC, Proimos B, Kappas C, Claudia D, Benassi M, Chierego G, Brahme A. Comparison of conformal radiation therapy techniques within the dynamic radio-therapy project `dynarad'. Phys Med Biol. 2000;45:2459–2481. doi: 10.1088/0031-9155/45/9/302. [DOI] [PubMed] [Google Scholar]
- 12.Mavroidis P, Lind BK, Brahme A. Biologically effective uniform dose for specification, report and comparison of dose response relations and treatment plans. Phys Med Biol. 2001;46:2607–2630. doi: 10.1088/0031-9155/46/10/307. [DOI] [PubMed] [Google Scholar]
- 13.Ekberg L, Holmberg O, Wittgren L, Bjelkengren G, Landberg T. What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer? Radiother Oncol. 1998;48:71–77. doi: 10.1016/s0167-8140(98)00046-2. [DOI] [PubMed] [Google Scholar]
- 14.Vandemeulebroucke J, Sarrut D, Clarysse P. The POPI-model, a point-validated pixel-based breathing thorax model. XVth International Conference on the Use of Computers in Radiation Therapy (ICCR); Toronto, Canada. 2007. [Google Scholar]
- 15. http://www.creatis.insa-lyon.fr/rio/popi-model.
- 16.Lin L, Shi C, Liu Y, Swanson G, Papanikolaou N. Development of a novel post-processing treatment planning platform for 4D radio-therapy. Technol Cancer Res Treat. 2008;7:125–132. doi: 10.1177/153303460800700205. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Thirion JP. Image matching as a diffusion process: An analogy with maxwell's demons. Med Image Anal. 1998;2:243–260. doi: 10.1016/s1361-8415(98)80022-4. [DOI] [PubMed] [Google Scholar]
- 18.Agren Cronqvist AK, Kallman P, Brahme A. Determination of the relative seriality of a tissue from its response to non-uniform dose delivery modeling. In: Baltas D, editor. Clinical Radiobiology. Albert-Ludwigs-University; Freiburg, Germany: 1994. p. 127. [Google Scholar]
- 19.Kallman P, Lind BK, Brahme A. An algorithm for maximizing the probability of complication-free tumour control in radiation therapy. Phys Med Biol. 1992;37:871–890. doi: 10.1088/0031-9155/37/4/004. [DOI] [PubMed] [Google Scholar]
- 20.Emami B, Lyman J, Brown A, Coia L, Goitein M, Munzen-rider JE, Shank B, Solin LJ, Wesson M. Tolerance of normal tissue to therapeutic irradiation. Int J Radiat Oncol Biol Phys. 1991;21:109–122. doi: 10.1016/0360-3016(91)90171-y. [DOI] [PubMed] [Google Scholar]
- 21.Agren Cronqvist AK. PhD thesis. University of Stockholm; Sweden: 1995. Quantification of the response of heterogeneous tumours and organized normal tissues to fractionated radio-therapy. [Google Scholar]
- 22.Okunieff P, Morgan D, Niemierko A, Suit HD. Radiation dose-response of human tumors. Int J Radiat Oncol Biol Phys. 1995;32:1227–1237. doi: 10.1016/0360-3016(94)00475-z. [DOI] [PubMed] [Google Scholar]
- 23.Gagliardi G, Lax I, Ottolenghi A, Rutqvist LE. Long-term cardiac mortality after radiotherapy of breast cancer -- application of the relative seriality model. Br J Radiol. 1996;69:839–846. doi: 10.1259/0007-1285-69-825-839. [DOI] [PubMed] [Google Scholar]
- 24.Gagliardi G, Bjohle J, Lax I, Ottolenghi A, Eriksson F, Liedberg A, Lind P, Rutqvist LE. Radiation pneumonitis after breast cancer irradiation: Analysis of the complication probability using the relative seriality model. Int J Radiat Oncol Biol Phys. 2000;46:373–381. doi: 10.1016/s0360-3016(99)00420-4. [DOI] [PubMed] [Google Scholar]
- 25.Johansson S, Svensson H, Denekamp J. Dose response and latency for radiation-induced fibrosis, edema, and neuropathy in breast cancer patients. Int J Radiat Oncol Biol Phys. 2002;52:1207–1219. doi: 10.1016/s0360-3016(01)02743-2. [DOI] [PubMed] [Google Scholar]
- 26.Rancati T, Wennberg B, Lind P, Svane G, Gagliardi G. Early clinical and radiological pulmonary complications following breast cancer radiation therapy: NTCP fit with four different models. Radio-ther Oncol. 2007;82:308–316. doi: 10.1016/j.radonc.2006.12.001. [DOI] [PubMed] [Google Scholar]
- 27.Ferreira BC, Mavroidis P, Adamus-Gorka M, Svensson R, Lind BK. The impact of different dose-response parameters on biologically optimized IMRT in breast cancer. Phys Med Biol. 2008;53:2733–2752. doi: 10.1088/0031-9155/53/10/019. [DOI] [PubMed] [Google Scholar]
- 28.Mageras GS, Yorke E, Rosenzweig K, Braban L, Keatley E, Ford E, Leibel SA, Ling CC. Fluoroscopic evaluation of diaphragmatic motion reduction with a respiratory gated radiotherapy system. J Appl Clin Med Phys. 2001;2:191–200. doi: 10.1120/jacmp.v2i4.2596. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Flampouri S, Jiang SB, Sharp GC, Wolfgang J, Patel AA, Choi NC. Estimation of the delivered patient dose in lung IMRT treatment based on deformable registration of 4D-CT data and monte carlo simulations. Phys Med Biol. 2006;51:2763–2779. doi: 10.1088/0031-9155/51/11/006. [DOI] [PubMed] [Google Scholar]
- 30.Mavroidis P, Ferreira BC, Papanikotaou N, Svensson R, Kappas C, Lind BK, Brahme A. Assessing the difference between planned and delivered intensity-modulated radiotherapy dose distributions based on radiobiological measures. Clin Oncol (R Coll Radiol) 2006;18:529–538. doi: 10.1016/j.clon.2006.04.007. [DOI] [PubMed] [Google Scholar]








