Abstract
At high magnetic field, MR images exhibit large, undesirable signal intensity variations commonly referred to as “intensity field bias”. Such inhomogeneities mostly originate from heterogeneous RF coil B1 profiles and, with no appropriate correction, are further pronounced when utilizing rooted sum of square reconstruction with receive coil arrays. These artifacts can significantly alter whole brain high resolution T1-weighted (T1w) images that are extensively utilized for clinical diagnosis, for gray/white matter segmentation as well as for coregistration with functional time series. In T1 weighted 3D-MPRAGE sequences, it is possible to preserve a bulk amount of T1 contrast through space by using adiabatic inversion RF pulses that are insensitive to transmit B1 variations above a minimum threshold. However, large intensity variations persist in the images, which are significantly more difficult to address at very high field where RF coil B1 profiles become more heterogeneous. Another characteristic of T1w MPRAGE sequences is their intrinsic sensitivity to Proton Density and T2* contrast, which cannot be removed with post-processing algorithms utilized to correct for receive coil sensitivity.
In this paper, we demonstrate a simple technique capable of producing normalized, high resolution T1w 3D-MPRAGE images that are devoid of receive coil sensitivity, Proton Density and T2* contrast. These images, which are suitable for routinely obtaining whole brain tissue segmentation at 7 Tesla, provide higher T1 contrast specificity than standard MPRAGE acquisitions. Our results show that removing the Proton Density component can help identifying small brain structures and that T2* induced artifacts can be removed from the images. The resulting unbiased T1w images can also be used to generate Maximum Intensity Projection angiograms, without additional data acquisition, that are inherently registered with T1w structural images. In addition, we introduce a simple technique to reduce residual signal intensity variations induced by Transmit B1 heterogeneity.
Because this approach requires two 3D images, one divided with the other, head motion could create serious problems, especially at high spatial resolution. To alleviate such inter-scan motion problems, we developed a new sequence where the two contrast acquisitions are interleaved within a single scan. This interleaved approach however comes with greater risk of intra-scan motion issues because of a longer single scan time. Users can choose between these two trade offs depending on specific protocols and patient populations.
We believe that the simplicity and the robustness of this double contrast based approach to address intensity field bias at high field and improve T1 contrast specificity, together with the capability of simultaneously obtaining angiography maps, advantageously counter balance the potential drawbacks of the technique, mainly a longer acquisition time and a moderate reduction in signal to noise ratio.
Keywords: Bias correction, MP-RAGE, T1 contrast, Intensity correction, Ultra high field, Adiabatic Inversion, Brain imaging, B1
Introduction
At high magnetic fields, when the RF wavelength at Larmor frequency is on the order of, or smaller than, the size of an imaged biological target, MR images exhibit undesired intensity variations (often called “intensity field bias”) originating from spatial variations of transmit (B1+) and receive (B1-) RF coil B1 profiles. This is the result of complex interactions between radiofrequency waves and electromagnetic properties of lossy dielectric biological tissues (Adriany et al., 2005b; Bottomley and Andrew, 1978; Collins et al., 2002; Glover et al., 1985; Keltner et al., 1991; Van de Moortele et al., 2005; Vaughan et al., 2004; Vaughan et al., 2001; Wiesinger et al., 2004). Most imaging modalities are adversely impacted by these artifacts, including whole brain high resolution T1-weighted (T1w) images which are extensively utilized for clinical diagnosis, for gray matter/white matter (GW/WM) segmentation, as well as for coregistration with time series images in functional MRI. 3D-T1w images are commonly obtained with Magnetization Prepared (MP) sequences, such as 3D-MP Rapid Acquisition Gradient Echo (3D-MPRAGE) (Mugler and Brookman, 1990), where a bulk amount of T1 contrast comes from an adiabatic non selective inversion pulse, insensitive to B1+ variations within a large range of B1+ values. However, large signal intensity variations are still present in the images, mostly the result of inhomogeneity in receive RF coil profiles and, to a variable degree, to small flip angle spatial variations.
At lower magnetic fields, such B1 induced intensity variations (essentially receive RF coil profile, with further amplification due to root sum of square (RSOS) reconstruction) can be efficiently corrected with a variety of post processing approaches (see (Belaroussi et al., 2006; Wang et al., 2005) and references therein), often including low-pass filtering, signal extrapolation and locally adaptive methods. Such algorithms however are not satisfactory at high field where B1 profiles have much steeper local variations (Van de Moortele et al., 2005; Vaughan et al., 2001; Wiesinger et al., 2004) and cannot be easily approximated or fitted with analytical expressions (e.g. polynomial); consequently, brain segmentation at magnetic fields of 3 Tesla and higher is still a challenging procedure (Belaroussi et al., 2006; Bernstein et al., 2006).
At the acquisition level, 1D shaped and 2D composite RF pulses have been proposed to address RF related inhomogeneity in T1w 3D-MPRAGE (Deichmann et al., 2000; Deichmann et al., 2002; Deichmann et al., 2004). However, transmit B1 distortions in the head at 7 Tesla are quite dramatic (Adriany et al., 2005b; Van de Moortele et al., 2005; Vaughan et al., 2001) and could not necessarily be addressed with RF pulse tools that perform well at 3T. Similar correction at 7T would require either B1 shim or Transmit Sense techniques which, currently, are available only in a few research centers (Adriany et al., 2005b; Katscher et al., 2003; Mao et al., 2006; Setsompop et al., 2006; Vaughan et al., 2001). Other approaches are based on fast B1 mapping (Yarnykh, 2007) or multi parametric mapping (Warntjes et al., 2007), but they require specific attention and non straightforward post-processing tools to deal with parametric fitting issues especially when signal to noise ratio (SNR) is low and/or when local tissue conditions depart from the underlying fitting models.
Beside these RF coil related issues, there is a fast growing interest in taking advantage of higher tissue contrasts at higher field. A partial list of these contrasts includes T1, T2, T2* and SWI (Augustinack et al., 2005; Li et al., 2006; Rooney et al., 2007; Thomas et al., 2008), and expectations are that it will be possible with high spatial resolution images at high field to identify some structures or lesions that, so far, cannot be visualized at lower field (Ge et al., 2008). T1w MPRAGE sequences, however, are intrinsically sensitive to Proton Density and T2* contrast, and T2* shortening in brain tissues at very high field naturally tends to make the latter bias greater whereas it would be beneficial to investigate separately the contributions of each tissue contrast.
Here1, we propose a straightforward correction to produce unbiased high resolution T1w 3D images, relying on two 3D images that are otherwise frequently utilized on clinical scanners: T1w 3D-MPRAGE and 3D-Gradient Echo (3D-GE). This paper is organized as follows:
We show that the normalized ratio between two standard 3D images with different contrast provides substantial reduction of intensity field bias by eliminating signal variations induced by B1-, and yields more accurate T1 contrast, including at high spatial resolution, by removing T2* and Proton Density (PD) components, to the cost of a limited loss in Signal to Noise Ratio (SNR).
We introduce a new sequence where the two contrast acquisitions (MPRAGE and GE) are interleaved, providing a choice between two trade offs: the interleaved sequence which avoids inter-scan registration issues, while two separate acquisitions, with a shorter acquisition for each single scan, reduces the risk of intra-scan head motion issues.
We demonstrate that maximum intensity projection (MIP) angiography maps can be derived from the normalized ratio without additional data acquisition.
We introduce an optional flip angle adjustment scheme allowing for reducing residual bias induced by Transmit B1 heterogeneity.
Theory
Image Ratio: general principle
The technique presented here involves dividing two images with different contrasts in order to eliminate or reduce most non-T1 sources of signal intensity variations, while preserving a large fraction of T1 contrast, as summarized in Table 1. Note that, although we have developed a new sequence interleaving the two image acquisitions, we will describe the principle considering two images acquired separately, which will greatly simplify all subsequent formalism in the manuscript without altering the fundamental principle. The diagram of the sequence is shown in Fig. 1.
Table 1. Contrast components in MPRAGE, PD-GE and P/NP ratio.
PD-GE | MP-RAGE | P/NP | ||
---|---|---|---|---|
A | IR-T1w | - | + | + |
B | PD | + | + | - |
T2* | + | + | - | |
|B1-| | + | + | - | |
Sin(α) | + | + | - | |
C | Fast GE T1w | +§ | +† | +‡ |
This approach relies on two fundamental properties: a) the B1+ insensitivity of adiabatic inversion over a large range of transmit B1 magnitude, and b) the reduced amount of T1 weight in fast GE images obtained with a very small flip angle. Because Proton Density (PD) is their main source of contrast, we will indifferently refer to fast GE images as GE-PD or GE images. The numerator of the proposed image division consists of the signal SMP (MP subscript is for MPRAGE) obtained with a 3D-MPRAGE sequence:
[1] |
while the denominator consists of the signal SGE (GE subscript for Gradient Echo) obtained with a similar 3D-GE sequence with all matching parameters, but devoid of the inversion RF pulse and associated delays (see Fig. 1):
[2] |
with ρ for proton density, B1- for receive RF coil sensitivity profile, and MMP,GE for the longitudinal magnetization in either MPRAGE or GE-PD. Note that throughout this paper, a tilde sign upon a variable Mindex indicates magnetization per unit of proton density with . The flip angle is proportional to transmit B1 magnitude (|B1+|), and γ is the gyromagnetic ratio, τ the equivalent duration of a square pulse, and V a scalar factor (Collins et al., 2002). In the dimensionless ratio image, which we will refer to as Prepared over Non Prepared (P/NP) ratio ([P/NP]=[MPRAGE]/[GE]), the pixel intensity SP/NP can be expressed as:
[3] |
It can be seen that multiple non-T1 sources of signal intensity variations can be eliminated in SP/NP: proton density ρ, receive RF coil sensitivity B1-, T2* relaxation and sin(α) (with a small flip angle the latter is approximately proportional to B1+). Note that the elimination of T2* and PD is only possible when MPRAGE and GE images are acquired at the same spatial resolution (and bandwidth), which is always the case in the present study. A similar approach, mostly dedicated to remove RF coil receive B1 profile from the data, has also been independently reported (van Gelderen et al., 2006), but GE images were acquired at a lower spatial resolution than MPRAGE and then heavily smoothed (2D kernel 27.5 × 27.5mm2 FWHM), so that most of the simplifications in Eq. [3] would not apply in that case, as will be discussed and demonstrated later. Here, SP/NP is entirely determined by the longitudinal magnetization per unit of proton density available in each of the two images, and the image ratio technique exploits fundamental differences existing between MMP and MGE with regards to T1 weight. Indeed, MMP is dominated by a bulk amount of T1 contrast approximately constant through space resulting from the adiabatic inversion, and this T1 component, being absent from MGE, is not affected by forming P/NP ratio maps. On the other hand, MMP as well as MGE are non uniformly modulated through space during the fast gradient echo part of the sequences, a direct consequence of spatial heterogeneity of α due to transmit B1 profile distortions, and they do not have the same functional dependence on α. It will be seen, however, that such residual spatial variations of signal induced by α are limited in P/NP maps.
Longitudinal Magnetization in 3D-MPRAGE
In order to describe the magnetization evolution with 3D-MPRAGE sequences (Mugler and Brookman, 1990) we utilize a formalism similar to the one adopted by Deichmann et al. (Deichmann et al., 2000). The generic sequence (first part of the diagram shown in Fig. 1) includes an adiabatic inversion pulse, an inversion delay TI, a host sequence of duration TA consisting of a fast train of N low flip angle (α) gradient echo (GE) acquisitions spanning through the 1st and 2nd dimension of the 3D k-space with a short repetition time TR, followed by an additional delay TD (Foo et al., 1994), before the next adiabatic inversion, looping through the 3rd dimension of k-space (Mugler and Brookman, 1990). The total duration Tcycle between two adiabatic inversion RF pulses is given by Tcycle = TI+TA+TD, with TA = (N-1)TR, where N is the number of small flip angle RF pulses within each fast readout echo train and TA the time between the first and the last of those pulses. During the intervals TI and TD, the longitudinal magnetization M relaxes with constant time T1 towards the value M0 (magnetization at thermal equilibrium), proportional to proton density ρ. However, during the train of fast GE, the overall longitudinal magnetization evolution can be described as relaxing with a shorter time constant T1* (Deichmann et al., 2000; Kaptein et al., 1976), with
[4] |
approaching a steady state value M0* smaller than M0 (in Eq. [4], ln is for natural logarithm). Note that if [α →0], then [T1* →T1]. Here, M0* is given by the standard GE steady state equation:
[5] |
with (in this paper any expression in the form of stands for ).
Assuming that T1 contrast is mostly determined at the k-space origin (Deichmann et al., 2000), and considering a linear phase encoding (LPE) scheme2 along the fast readout echo train, we express the MPRAGE longitudinal magnetization at the sampling time TSLPE = TI + TA/2:
[6] |
where ξ, efficiency of the inversion pulse (Gelman et al., 2001), determines the fraction of longitudinal magnetization actually inverted [-1 ≤ ξ ≤ 1, ideally ξ = 1]. This is equivalent to writing ξ = - cosθ, with θ the effective flip angle achieved along the Z axis at the end of the inversion pulse [ideally θ = 180 °].
One can note that in Eq. [6], if the small flip angle α approaches zero, then T1* approaches T1, the problem reducing then to a simple mono exponential case. For illustration, if we also assume TD≫T1 then after Eq. [6]:
[7] |
Within this extreme approximation, ξ and TSLPE would be the only sequence parameters impacting T1-weight and, assuming that |B1+| is above adiabatic threshold, full inversion would be obtained irrespective of B1+ inhomogeneities [ξ = 1], reducing further the expression into:
[8] |
In the interleaved version, a second train of fast GE (followed by an optional delay TD2 which typically is null) is inserted between TD and the next inversion pulse, with the kspace center of the GE image sampled at TI + TA + TD + TA/2 in the case of linear phase encoding without partial Fourier acquisition. Although the same fundamental principles apply for both separate and interleaved acquisition, we did not include in this manuscript the formalism of the interleaved version which results in a fairly long expression. The main difference compared with separate acquisitions is that the delay TD now also determines how close to steady state the longitudinal magnetization is at the beginning of the second train of fast gradient echo (GE image). As a result, if the delay (TI + TA + TD) is not long enough with regards to T1, the GE image may carry a significant amount of T1 contrast.
Noise propagation
In this section we only address the Signal to Noise Ratio (SNR) available in the ratio images; residual signal intensity variations and impact on tissue contrast are discussed elsewhere. The noise propagation in the ratio x of two variables u and v, with x=ulv, can be expressed as (Bevington, 1969):
[9] |
(σ is the standard deviation). With same noise level in MPRAGE and GE-PD images, i.e. σIM=σMPRAGE=σGE, and assuming that noise in MPRAGE signal (SMP) and GE-PD signal (SGE) are uncorrelated, i.e. σuσv = 0, we can write:
[10] |
As we are mostly interested in comparing SNR in P/NP (SNRP/NP) and in MPRAGE (SNRMP) this can be expressed as:
[11] |
With the technique presented here, SP/NP was typically around 0.65 or less in the white matter and around 0.35 or less in the gray matter with our acquisition parameters (with different sequence parameters different SP/NP values could be found; however, SP/NP was in this range for a variety of parameter sets at 7T). Compared with the source MPRAGE images, this corresponds to a loss in SNR of only ∼6% in gray matter and ∼16% in white matter (see Fig. 2). Fortunately, it can be seen that the largest percentage of SNR loss, although still moderate (16%), occurs for the largest initial SNR (white matter), indicating that the present ratio technique will have only a limited impact on the overall SNR.
Impact of image division on WM/GM contrast
The question addressed in this section is the impact of dividing an MPRAGE image with a GE-PD image on white matter/gray matter contrast, with pre-existing contrast in MPRAGE taken as a reference. For this purpose, we assume all acquisition parameters identical for white and gray matter, including α and B1-, and we also ignore differences in WM/GM T2*'s whose relative weight depend on which particular TE is chosen. Note that we do not attempt to address here T1 contrast issues and Point Spread Function (PSF) optimization in MPRAGE sequences, which have been extensively investigated elsewhere (Deichmann and Haase, 1992; Foo et al., 1994; Haase, 1990; Mugler and Brookman, 1990).
The WM/GM contrast in MPRAGE images and in P/NP maps can be expressed as (see Appendix 1):
[12] |
and
[13] |
with . The value of λ is entirely determined by the GE-PD image (as long as the same value for α is utilized in MPRAGE and GE-PE) and can be expressed as (see Appendix 1 and Eq. [2]):
[14] |
Eq. [13] shows that the WM/GM contrast in Ratio Images decreases with λ>1, increases with λ<1 and is unaltered with λ=l. It is also immediately apparent with Eq. [14] that two GE-PD image components act in opposite directions: the ratio of proton density tends to increase T1 contrast [ρW÷ρG≈0.8], whereas steady state longitudinal magnetization per unit of PD tends to decrease T1 contrast . Very importantly, a simple visual inspection of the GE-PD images is sufficient to determine if λ is smaller than one (GM brighter than WM, PD contrast greater than T1 contrast) or bigger than one (WM brighter than GM, T1 contrast greater than PD contrast). With a small flip angle and a short TR, λ is typically smaller than one, thus overall WM/GM contrast is actually greater in the ratio images than in MPRAGE images because the gain in WM/GM contrast obtained by removing PD contrast is larger than the limited loss in T1 contrast due to the residual T1 weight existing in GE-PD images. This increase in WM/GM contrast also reduces the loss in Contrast to Noise Ratio (CNR) in the ratio images despite a lower SNR. (CNR can even increase in ratio images depending on the acquisition parameters). The ratio between the WM/GM contrast in Ratio Images and the WM/GM contrast in MPRAGE can be written as:
[15] |
Material and methods
Healthy volunteers who had signed a written consent form approved by the Institutional Review Board of the University of Minnesota were imaged at 7 Tesla. The images obtained in separate acquisitions were obtained with standard sequences available on the scanners (MPRAGE and GE-PD). The interleaved sequence (Van de Moortele et al., 2008b) was developed in house for the 7T Siemens console. Because this study is primarily concerned with the impact of forming the ratio between MPRAGE and GE-PD images, rather than with optimizing each individual acquisition, we utilized a variety of existing protocols without attempting to modify their T1 contrast characteristics. For all 3D-MPRAGE images described below, corresponding 3D-GE images acquired separately were obtained with all parameters identical but without inversion pulse and with TI and TD set to zero. One exception concerns the small flip angle and will be discussed later.
Note that the inversion time TI as defined on the Siemens console (TISiemens) refers to the time when the center of the k-space is sampled along the 2nd dimension. In the present manuscript TI actually refers, as suggested elsewhere (Deichmann et al., 2000), to the time when the train of fast GE readouts begins. Thus, with linear phase encoding and in the absence of partial Fourier reconstruction: TISiemens = TI + TA/2.
7 Tesla experiments
Imaging experiments were performed with a 7 Tesla magnet (Magnex Scientific, UK) driven with a Siemens console (Erlangen, Germany). A 16-channel transmit/receive head array coil, as described in (Adriany et al., 2005a; Adriany et al., 2005b) was used for both RF-transmit and signal-receive operations in whole brain imaging experiments. Additional data were obtained in the occipital lobe with another RF coil setup which has been described in details elsewhere (Pfeuffer et al., 2002): briefly, signal reception was obtained with a quadrature RF coil made out of two partially overlapping loops of 6 cm diameters, while RF transmission was produced with a large half volume quadrature surface coil (18 × 12cm2 pair). Multiple sets of parameters were used to acquire 3D-MPRAGE and 3D-GE images, either as separate acquisitions or with the interleaved sequence, and the corresponding details will be found in the figure captions. Note that in all experiments involving acceleration with GRAPPA the number of auto calibration lines was set to 24.
Ratio Images and thresholding
MPRAGE and GE-PD magnitude images were exported from the acquisition console in DICOM format, explicitly excluding any intensity correction filter during image reconstruction at the console. P/NP ratio maps were then obtained by directly dividing MPRAGE images with corresponding GE-PD images. In order to limit meaningless high pixel intensity in pixels divided with noise, a binary mask was defined based on SNR thresholding in GE images. The threshold value was determined manually in order to exclude background noise from the volume of interest. Pixels excluded from this mask were set to zero in the Ratio Images. In order to investigate the impact of using a smoothed reference image we also generated a smoothed reference image by convolving the high resolution GE with a 27.5 × 27.5 mm2 FWHM Gaussian kernel, as described in (van Gelderen et al., 2006).
Motion correction
With data sets obtained in two separate acquisitions inter-scan head motion can occur, translating in edge artifacts in ratio images. In such cases we utilized the FLIRT toolbox (Jenkinson et al., 2002) in the FSL package (Smith et al., 2004) in order to realign the 3D-MPRAGE image onto the 3D-GE image.
Brain segmentation
Based on the normalized ratio images, whole brain segmentation as well as white and gray matter surface rendering can now be obtained as a routine procedure in our center at 7 Tesla using the SurfRelax package (Larsson, 2001). Those specific developments are beyond the scope of this paper and will be detailed elsewhere (Olman et al., 2007). One significant result will, however, be shown in the present manuscript demonstrating successful brain segmentation at 7 Tesla based on high spatial resolution ratio images.
Maximum Intensity Projection Angiography
Gray matter, white matter and CSF inherently exhibit higher signal intensity in low flip angle 3D-GE than in 3D-MPRAGE images, so that their ratio value is always smaller than one (except in locations where B1+ is so weak that inversion pulses do not reach adiabatic conditions). This is not the case for flowing blood. Indeed, we observed that the net signal of flowing blood is greater in 3D-MPRAGE than in 3D-GE images, resulting in values greater than unity in P/PN ratio images for voxels located in vessels. As a result, with an appropriate threshold one can isolate signals originating from vessels in the ratio images, and Maximum Intensity Projection (MIP) maps (Laub, 1990) can then be utilized to visualize those vessels. The short T1 of fat in skin tissues, however, also produce confounding bright signals, a well known issue in MIP images which can be overcome by excluding the corresponding pixel from the analysis. Angiograms were produced with the following steps. First, MPRAGE and GE images were interpolated at twice the acquisition spatial resolution before forming NP/P ratio images. Then skin and skull were removed from the ratio images using morphomathematical tools (erosion/dilation) from a 3D toolbox (SDC Information Systems, Naperville, IL) written for matlab. A minimum intensity threshold was then manually adjusted in order to retain vessels (brighter pixels) while excluding tissues (darker pixels), and the resulting angiography was visualized as MIP maps along axial, sagittal and coronal projections. All computations were performed in Matlab (The Mathworks, Inc.).
Results
Intensity field bias correction
Typical results obtained at 7 Tesla are shown in Fig. 3: intensity variations in MPRAGE images are such that the gray scale cannot be adjusted to accommodate all gray values through brain tissues, or tissue contrast would be lost. One can see that similar variations of signal intensity, on a large spatial scale and with large amplitude, affect both MPRAGE and GE-PD images, whereas P/NP ratio images are much more homogeneous, still preserving excellent gray/white matter contrast. White arrows in the figure are pointing towards areas where signal intensity is either too high or too low when trying to manage a unique gray scale windowing for the whole brain in MPRAGE or GE images. The corresponding areas are within a homogeneous range of gray scale in P/NP images. Note also that the technique performs well from the vertex to the lowest part of the brain. While those images were obtained with a small flip angle of 3°, they do not suffer from a poor SNR, indicating that spatial resolution can be increased further with this approach at high field, as will be shown later with the interleaved sequence.
Initially designed for whole brain T1w imaging at 7T, the ratio technique also proved very efficient with surface RF coils. This is shown in Fig. 4 where 3D MPRAGE images were obtained at 7 Tesla with a half volume RF transmit coil and a 2-loop quadrature receive coil. Such surface coil data typically provide images with steep signal decay as a function of distance to the coil and naturally suffer a loss in SNR when compared with an array of receive coil elements sampled on different receive channels (Roemer et al., 1990). The ratio technique, instead of traditional images with constant thermal noise and large intensity variations, inherently provides images with constant average signal intensity whereas local noise levels are rescaled accordingly. In other words, parts of the brain that are extremely dark in MPRAGE images, such as the cerebellum and central structures in this particular case, have intrinsically very low SNR and a single gray scale display would not accommodate more than a small fraction of the imaged tissues. In P/NP ratio images (Fig. 4), all areas are now equally visible with high T1 contrast. As expected, the noise in P/NP images is much more visible in low SNR areas. It is important to distinguish such increases in noise visibility from actual losses in SNR which, as was discussed in the Theory section, are actually very moderate when forming P/NP ratio images. A source of SNR reduction compared with Fig. 3 is due to using a single receive channel coil versus a 16 receive channels coil, together with lower amount of collected data (smaller matrix size).
Head motion and interleaved sequence
Head motion between two separately acquired scans can naturally be a serious issue as illustrated in Fig. 5 for images obtained with a healthy volunteer. Even limited motion is sufficient to produce evident edge artifacts in the P/NP ratio images which could jeopardize tissue classification procedures. As can be seen in the same figure, it was possible to successfully realign 3D-MPRAGE onto 3D-GE utilizing the standard FLIRT package and eliminate any noticeable artifacts in the ratio obtained with realigned data. This registration, however, typically comes at the price of a certain degree of blurring often induced by the interpolation filter. Naturally, using a higher spatial resolution makes realignment issues even more challenging. The interleaved sequence that we have developed in order to address this issue is virtually insensitive to inter-scan head motion (except for brief jumps in head position shorter than Tcycle), as equivalent k-space lines in MPRAGE and GE are acquired with almost same head position even when head position slowly moves throughout the course of the acquisition. It should be noted that, due to a longer single scan time, the interleaved sequence increases the risk of intra-scan head motion artifacts. Nevertheless, standard segmentation algorithm performed very well even on ratio images obtained with the interleaved sequence, including at a spatial resolution of 0.27 mm3 and without applying any spatial filtering, as illustrated in Fig. 6. Because the current technique provides both unbiased P/NP and uncorrected T1w 3D MPRAGE images, it is straightforward to overlay parametric maps calculated with P/NP ratio onto MPRAGE images, as shown for gray/white matter segmentation (see yellow contours). One can note that, besides cerebral tissues, very large signal intensity variations in skin and skull are remarkably corrected, making skull stripping procedures easier.
Angiogram
Another advantage of removing intensity non-uniformities is to improve vessel visualization with high resolution isotropic MIP angiograms (Fig. 7), helping to identify blood vessels, which can constitute a valuable information for functional MRI analysis as well as for anatomical structural imaging. The angiograms shown in Fig. 7 were obtained by forming Maximum Intensity Projection maps along the X, Y and Z axes, after removing outer non-cerebral structures from the data set shown in Fig. 6. Note that, whereas segmentation was obtained without spatial smoothing (Fig. 6), 3D-GE images (but not 3D-MPRAGE) were spatially smoothed before forming ratio images utilized to produce angiograms. We found that filtering GE images with a 3D-gaussian convolution kernel with a FWHM of ∼1.5mm was a good tradeoff allowing for improving local SNR while preserving vessel structure identification. We hypothesize that the benefit of this filter is in part due to the fact that the blood signal is decreased in GE images compared with MPRAGE, a less favorable situation for noise propagation because GE images are the denominator of the ratio.
Smoothed vs. Unfiltered GE reference: Impact on Brain Structures at High Spatial Resolution
Because of the inherent sensitivity of T1w MPRAGE images to Proton Density and T2* contrast we hypothesized that the appearance of some brain structures may be contaminated with non-T1 components. Such non-T1 contributing patterns, however, are expected to reflect small brain anatomical structures, thus requiring a high spatial resolution reference image to be corrected for. Opposite to this case, receive B1 profiles typically have much smoother spatial patterns and should be successfully removed with a low spatial resolution reference image. To verify these hypotheses high resolution data were obtained with the interleaved sequence, and the MPRAGE images were divided either with the unfiltered GE image (same spatial resolution) or with the GE image smoothed with a 27 × 27.5mm2 Gaussian kernel, as described in (van Gelderen et al., 2006). A few observations can be made about the corresponding results shown in Fig. 8. First of all, unwanted slow spatial variations of signal (mostly corresponding to receive B1 profiles) clearly present in original MPRAGE images (fourth column), were very efficiently removed with both the smoothed (first column) and the unfiltered (second column) GE image, resulting in corrected images shown in the third and fifth column, respectively. On the other hand, when comparing MPRAGE image with MPRAGE/Filtered GE we could not identify high spatial frequency alteration between the two images, which also was expected.
However, the division of the MPRAGE image with the unfiltered high resolution GE reference was capable of unveiling some cerebral structures which were difficult to identify in MPRAGE images (compare the white box contents in the fourth and fifth columns), and these anatomical structures clearly include very high spatial frequency components. This is supported by a comparison between the first and the second rows, which are showing closely matching axial views extracted from two data sets acquired with isotropic 0.51mm3 voxel size (first row) and isotropic 0.3mm3 voxel size (second row). These two data sets (each consisting in MPRAGE + GE 3D images) were acquired with the interleaved sequence in the same volunteer within the same session. In the second row, the higher spatial resolution naturally results in smaller and more numerous structural details being identified in MPRAGE as well as in GE images. Interestingly, this higher spatial resolution also results in smaller and more numerous structural details being unveiled in the MPRAGE/GE image whereas the low spatial signal variations (receive B1) were removed with apparently equal success at both spatial resolutions and with either GE or Filtered GE. The third and fourth rows in Fig. 8 are showing coronal views of a data set obtained in another volunteer with isotropic 0.3mm3 voxel size. Again, inside the white box in the third row some small structures are better identified in MPRAGE/GE compared with MPRAGE.
We assume that the aforementioned unveiled anatomical structures may be due in a large part to the removal of Proton Density which typically cancels a fraction of T1 contrast between gray and white matter in MPRAGE images. However, shorter T2* at higher field can yield losses of signals, especially close to vascular structures, and it may be difficult to determine with certainty the origin of such small signal defects in MPRAGE images. In the fourth row of Fig. 8, the two white arrows point towards small dark spots in GE images, each surrounded with two small bright spots along the Z direction, suggestive of typical susceptibility effects induced by small vessels. Because T2* contribution are the same in MPRAGE and GE images, one would expect most of this aspect to be removed when forming the ratio between these two images. Indeed, these vessel induced patterns essentially disappeared in MPRAGE/GE (fifth column), while they were left intact in MPRAGE/Filtered GE.
When a longer segment of a vessel, rather than just a normal section, is captured within a slice, the pattern of resulting signal losses can become more complex and more difficult to interpret. Such a situation is illustrated in Fig. 9 showing, on three rows, three contiguous parasagittal views extracted from an isotropic 0.3mm3 data set (interleaved acquisition). The pattern of approximately horizontal dark and bright lines enclosed between the brackets, present in both MPRAGE and GE, is considerably attenuated in the ratio MPRAGE/Filtered GE, with smoother and better identified tissue borders.
Besides blood vessels, some brain tissues also have shorter T2*, which in MPRAGE images translates in lower signal intensity that cannot be differentiated from actual T1 contrast, which again affects T1 contrast specificity. Note that here, removing T2* components from MPRAGE images should result in an apparent decrease in tissue contrast. This mechanism is likely to explain, in Fig. 9, the lower visibility of the structure labeled ‘iii’ in MPRAGE/GE compared with MPRAGE. By contrast, both structures labeled ‘i’ and ‘ii’ appear with increased contrast and sharper edges in MPRAGE/GE than in MPRAGE, suggesting the absence of significant T2* component in the two latter structures.
Residual Transmit B1 induced bias
It is known that transmit B1 profiles becomes significantly heterogeneous at 7T (Collins et al., 2002; Van de Moortele et al., 2005; Vaughan et al., 2001). As a consequence, larger residual signal variations are expected in P/NP maps at 7T due to the contribution of a more heterogeneous α. In order to apprehend better this phenomenon we have simulated, based on Eqs. [5] and [6], the impact of α variations on P/NP ratio values for different values of T1, utilizing same acquisition parameters as in Fig. 3. As can be seen in Fig. 10A, our simulations predict that variations in α can have a significant impact on signal intensity in P/NP, (other reports had suggested that the impact of transmit B1 profile would rather disappear in ratio images (van Gelderen et al., 2006)). This plot also reminds that above a certain threshold for α, T1 contrast reduces when α increases, a feature which is known for MPRAGE sequences utilizing linear phase encoding (Deichmann et al., 2000). Strikingly, as shown in Fig. 10B, simulations suggest that simply utilizing a flip angle in MPRAGE larger than in 3D-GE may considerably attenuate the sensitivity of P/NP to α variations, potentially providing an additional tool to obtain more regular P/NP maps at very high field. A preliminary experimental verification can be seen Fig. 11, showing three different ratio images obtained with a volunteer for whom, within a same session at 7 Tesla, both MPRAGE and GE images had been acquired two times, first with a nominal flip angle of 3° and then with a nominal flip angle of 6°. In Fig. 11, the ratio MPRAGE[α=6°]/GE[α=6°] clearly exhibits significant residual intensity bias. To interpret the patterns in those images, it is important to remember that receive B1 profile obtained when summing multiple receive RF coil sources (16 channels here) results in brighter intensity close to the coils (i.e. in the periphery of the brain), and that this receive component is expected to be entirely absent in ratio images, independently of α (see [3]). In Fig. 11, however, spatial patterns of residual brighter intensity fundamentally differ from the sum of receive B1 profiles. Here, in an axial view the brighter signal area includes the center of the brain and is diagonally stretched towards the periphery, and in a coronal view the brighter component concerns the central part of the brain. Those patterns are highly consistent with B1+ profiles previously measured at 7T with transceiver array coils similar to the one utilized here (Adriany et al., 2005b) or with TEM head RF coils (Vaughan et al., 2001). Those brighter areas are considerably attenuated when utilizing a smaller value for α in both MPRAGE and GE, as seen in the ratio image MPRAGE[α=3°]/GE[α=3°]. Interestingly, dividing the larger flip angle MPRAGE[α=6°] with the smaller flip angle GE[α=3°] reduces even further the residual transmit B1 induced intensity variations, consistent with the prediction plot shown in Fig. 10.
Adiabatic inversion
The residual transmit B1 profiles described above are related with the small excitation flip angle α, which is approximately proportional to |B1+|. Adiabatic inversion, however, is expected to be uniform through space, except when |B1+| falls below a minimum threshold. The subsequent local lack of inversion naturally yields a loss in T1 contrast in MPRAGE images, but a weak |B1+| also yields a very small flip angle so that local SNR can be very low, making it difficult to determine if a local dark area rather results from RF coil issues or from anatomical features. As can be seen in Fig. 12, such failures are be easier to identify in P/NP images under the form of a local spot brighter then gray and white matter areas (see details in Fig. 12 caption). Because with the present technique MPRAGE and GE images are available as well, one can easily verify that such bright area in P/NP corresponds to a dark area in GE images and to a loss of local contrast in MPRAGE images. Besides B1+ profiles, other well known sources of inversion failure, especially with non selective inversion pulses, include areas with local susceptibility induced resonance offsets larger than half the bandwidth of the inversion pulse. In such locations, and in the absence of local B1+ defect, the lack of T1 contrast yields high signal intensity in MPRAGE images, which translates in the ratio images in areas brighter than gray and white matter (data not shown). Because such artifacts occur in the vicinity of air/bone/tissue interfaces they are usually easily identified. However, here again a simple visual inspection of the 3 available images with different contrasts allows for an immediate verification.
Discussion
Utilizing the ratio of two standard MR images proves to be a very efficient technique to eliminate the receive coil profile from T1w MPRAGE images at 7Tesla, with a volume or a surface coil. This is expected to greatly facilitate whole brain tissue segmentation as well as co-registration between imaging sessions obtained at different times, using different RF coils and on different scanners. This might be especially advantageous in follow-up studies. The current technique directly relies on unmodified, gold standard T1w images, which are available to the clinician in addition to the unbiased T1w images. In addition to removing signal intensity variations due to coil sensitivities, which are common targets in post processing field bias correction techniques, the current approach also removes T2* and proton-density contributions, providing a more accurate T1 contrast; furthermore, Maximum Intensity Projection angiograms can be generated without additional acquisition.
T2* and Proton Density contrast removal
At low fields, transverse relaxation in brain tissues is very limited during fast GE acquisition because tissue T2*'s are much longer than TE. At high field, however, this assumption might not always be valid due to short T2*'s and due to even shorter T2*'s in the vicinity of small structures surrounded with steep B0 gradients (e.g. small veins). At 7T, for example, T2* values have been reported as ∼25-35ms in white and gray (cortical) matter, ∼20ms in Caudate, ∼16ms in Putamen, and ∼13ms in vessels (Li et al., 2006; Peters et al., 2007; Yacoub et al., 2001). Keeping a short TE at very high spatial resolution in order to limit T2* weight is doable to a certain extent but is constrained by hardware and safety limitations (maximum gradient strength, slew rate) and, especially at high field, tends to significantly increase gradient coil acoustic noise. Furthermore, it has been shown that T1 distribution through the gray matter is significantly more distributed at 7T than at lower field, potentially carrying more information (Rooney et al., 2007). For those reasons it is desirable to remove T2* components from T1w images obtained at high field.
Although Proton Density and T1 values have been reported as highly correlated in the brain (Gelman et al., 2001; Just and Thelen, 1988), they yield opposite contrasts in T1w images, which have inherently some contrast loss due to PD. This loss in T1 contrast is typically recovered in P/NP images (with the interleaved sequence this requires TI + TA + TD to be long enough in order to limit T1 contrast in GE images).
Our experimental results confirm the benefit of removing T2* and PD components from T1 MPRAGE images, resulting in better brain structure visualization at a small spatial scale as well in some restitution of signal in short T2* location. Our results also point towards additional potentiality in utilizing the original T1w MPRAGE and GE images in conjunction with the ratio images.
Residual transmit B1 induced bias
The simplified mathematical model in Eq. [5] and [6] indicates that utilizing very small values for α at 7 Tesla (such as 3°) should result in lower amounts of residual transmit B1 induced signal variations in P/NP ratio (see Fig. 10). Given the higher SNR available at higher field, and considering that T1 contrast is better preserved with smaller values of α, this suggests that very small flip angles (such as 3°) can advantageously be used at very high field (see Fig. 3).
In general, within the range of parameters used in the present study, we found P/NP maps acceptably uniform when utilizing the same α value for MPRAGE and GE, as long as the nominal value for α did not exceed about 5°. Our simulations, however, predicted further improvements by utilizing a larger flip angle in MPRAGE than in GE images (Fig. 10), which is supported by additional experimental data acquired with two different values for α (Fig. 11). Naturally, using a smaller value for α in both MPRAGE and GE means a lower SNR in both images. The two different flip angle approach reduces this SNR penalty by increasing signal intensity in MPRAGE, as long as α is within limits of preserving a T1 contrast predominantly induced by the inversion pulse (Deichmann et al., 2000). In the particular data shown in Fig. 11, the residual transmit bias observed with α=6° was somewhat more pronounced than in other similar data sets. However, our nominal flip angles at 7T were determined with standard calibration routines delivered with the console, that are not reliable at ultra high field as demonstrated by Collins et al. (Collins and Smith, 2001), and depend among other things on which part of the brain (high or low transmit B1) is at the magnet isocenter during the calibration. It is thus possible that the actual average flip angles were somewhat higher than 3° and 6° in this case, but a similar trend (reduced transmit residuals with smaller α value, and further improvement with α [MPRAGE] larger than α [GE]) was consistently found in multiple data sets (not shown). The range of transmit B1 variations through the brain greatly depends on RF coil design which, together with new technologies such as B1 Shimming, could mitigate this issue in the future. However, being able to improve further P/NP quality by choosing two values for α offers the advantage of a straightforward implementation, and optimum values could be calculated on the fly based on acquisition parameters.
Adiabatic inversion
Whereas at lower field RF excitation is most of the time obtained with a large body RF coil, most neuroimaging experiments at 7T utilize head RF coils which are sometimes closely fitting to the head (Adriany et al., 2005a). Relative phase maladjustments between transmit channels can then result in weak transmit B1 amplitude in some areas (Adriany et al., 2005b), a direct consequence of large B1 distortion at high field (Van de Moortele et al., 2005). Although specific solutions (RF coil design, B1 shim, Transmit SENSE, etc.) are under active development, they are not yet available in most imaging centers. Thus, it is important to be able to identify potential local failure in adiabatic inversion at 7 Tesla. We show here (see Fig. 12) that a local spot of weak |B1+|, below adiabatic threshold, typically translates into an abnormally bright areas in P/NP images that are easier to identify than in MPRAGE images. This is because, with an incomplete inversion, the longitudinal magnetization is much closer to thermal equilibrium at the sampling time, whereas the low signal intensity term due to the small flip angle affects both the T1w and reference images, and hence for the most part cancels when forming the ratio. It will be important to evaluate whether this artifact identification criterion would be reliable in the case of tissues alterations under pathological conditions.
Acquisition strategy: scan time, SNR, spatial resolution, head motion
One noticeable limitation of our technique is the additional time required for acquiring a 3D-GE reference image (with an acquisition time typically between half and three quarter of that for 3D-MPRAGE), with resulting SNR in unbiased T1w images rather moderately lower than in 3D-MPRAGE. This would evidently not be an efficient strategy if SNR was a prominent target. In the current study, however, our main goal was not to improve SNR. Rather, we aimed to demonstrate a technique which takes advantage of inherently high SNR and CNR at high field in order to produce high resolution images with high T1 contrast specificity, devoid of T2* and Proton Density contrast and of the large signal variations due to receive coil profiles. For this reason, the results presented here were not evaluated in term of SNR efficiency per unit of time, which actually rather decreases in P/NP compared with MPRAGE alone. Our main criterions rather were to succeed in getting routine tissue classification through the whole brain despite severe B1 distortion, in demonstrating the benefits of higher T1 specificity to identify brain structures at high resolution, in handling inter-scan head motion issues, and in addressing residual transmit B1 bias. We verified that our technique could easily be implemented for a variety of existing T1-weighted MPRAGE protocols (a concern expressed by multiple users). We also verified that, even in much noisier conditions than standard isotropic cubic millimeter resolution, tissue brain segmentation was successfully obtained, even without any spatial filtering (see Fig. 6).
Note that at low spatial resolution, Tcycle can be fairly short in standard (non interleaved) T1w MPRAGE acquisitions. As a result, a significant delay TD would be required to use the interleaved version in order to be close to steady state for the GE acquisition, yielding a low temporal efficiency. This is not the case at higher spatial resolution where TA tends to be significantly longer so that only a relatively short delay TD is sufficient. In the later case, the interleaved acquisition is not significantly longer than the sum of two separate acquisitions. In the present study the additional scan time needed to acquire isotropic cubic millimeter resolution whole brain 3D-GE images at 7 Tesla varied from 2min36s (out of a total of 7min30s for two interleaved images, see Fig. 12) up to 5min45s (separate acquisitions, Fig. 2). For the higher resolution data sets (0.27mm3), the GE image took 5 minutes (out of a total of 14min12s for two interleave images). We found that those additional scan times were not a problem with healthy volunteers; this would naturally be seen differently with patients. Additional investigation will, however, be needed to optimize tradeoffs between spatial resolution, SNR, CNR and acquisition time which depend on multiple acquisition parameters (Tcycle, TR, TE, bandwidth, TI, TD, TA, partial Fourier and parallel imaging acceleration)
Regarding spatial resolution for GE images, another similar approach successfully utilizing a GE reference in order to correct T1w MPRAGE images for receive B1 profile has been independently demonstrated by others at 7T (van Gelderen et al., 2006). However, GE images were acquired at a lower resolution than MPRAGE and then heavily spatially smoothed. This is perfectly suitable for solely removing receive B1 induced intensity field bias, in which case low resolution GE images can indeed be acquired in a short time to efficiently correct for receive B1 bias (scan time = 1min in the referenced work), as shown in our results as well. (it should be reminded that utilizing large smoothing convolution filters still requires the use of specific algorithms in order avoid significant artifacts at the periphery of the brain).
We find, however, considerable interests in acquiring both images at matching spatial resolution, worth in our view the cost of up to 5 minutes of scan time. The resulting high resolution images with high T1 contrast specificity (without T2*, PD and B1 contribution) can be used for applications such as clinical diagnosis, cortical thickness mapping or brain segmentation, as well as to produce angiograms.
Angiograms
We see the MIP angiograms as a very significant addition. Because they are directly derived from the ratio, they should allow for unambiguous identification of vessels location with regards to brain structures without resampling or co-registration error. A remarkable characteristic was the clear and homogeneous separation between vessels and cerebral tissues through the whole brain, which we attributed to the normalizing ratio. Indeed, the vessel information (bright pixels) is present in the MPRAGE images and as such could be in principle directly extracted from these data without forming the ratio images. With this approach, however, it would be impossible to obtain whole brain angiograms in a single step MIP process, because in large areas vessels would be obscured by brain tissues exhibiting larger signal intensity. In the ratio images, a simple intensity threshold (after removal of skin and skull) was sufficient to produce through the whole brain MIP angiograms with almost no signal visible in the background. Optimizing those angiograms while preserving high quality, low bias T1w tissue images will be an important target of our future investigations with this technique. It should be noted that removing T2* and PD contrasts is not expected to have an impact as significant for the angiograms as for the T1-weighted images.
Low SNR mask
When dividing an image with another one, pixels that have only noise or very low SNR in the denominator image generate very high, meaningless values in the ratio image. We simply excluded such pixels with a binary mask derived from the GE images with an intensity threshold which was manually determined. Although we found the resulting ratio images well defined with this procedure, some artifacts are expected to occur on the edges of this mask, due to partial volume between pixels with high PD and pixels with either noise only or very low PD (e.g. inner bone). Imperfect head motion correction for images acquired separately may also yield additional edge artifacts. The potential impact of such edge effects on tissue classification algorithms may have to be investigated, although we rather found that standard segmentation algorithms were working successfully on unbiased P/NP ratio images obtained with the simple mask described above (Olman et al., 2007).
Impact of Receive Coil Arrays
For simplicity we have assumed in the formalism utilized in this paper that signal intensity for both MPRAGE and GE-PD images is proportional to a receive coil sensitivity profile |B1-|. When utilizing receive coil arrays, output magnitude images are usually obtained as the root sum of squares (RSOS) of the individual receive coil images for SNR optimization (Roemer et al., 1990). As a result, |B1-| should be replaced with the root sum of squares of the receive coil sensitivities in Eq. [1] and [2]. However, coil sensitivity profiles concern equally the numerator and the denominator, so that P/NP ratio can be directly formed with RSOS images without modification.
Other limitations
3D-MPRAGE sequences inherently induce a distorted point spread function (PSF) in T1w images due to the evolution of MMPRAGE during the fast train of readout (Deichmann et al., 2000). In the ratio technique proposed here, the 3D-GE image used for the division is acquired in the steady state, thus with very little T1 related PSF distortion. Forming ratio images does not address MPRAGE related PSF. However, to a large extent, any optimized MPRAGE sequence (Deichmann et al., 2000; Foo et al., 1994) could benefit from the present ratio technique. It should be noted that at higher field, due to longer T1's, longer TA's can be used with unchanged TA/T1 ratio. This suggests that higher spatial resolution may be obtained at higher field with MPRAGE sequences without further degrading the T1 PSF. In this study we did not address the issue of Rician noise distribution in magnitude images. However, in most data presented here, SNR was high enough to make such impact negligible.
Conclusion
We have demonstrated at 7 Tesla that forming the ratio of T1w images obtained with adiabatic inversion to GE images obtained without preparation is a simple and very efficient way to correct T1w images for signal variations induced by B1 profiles, Proton Density and T2* contrast. Although T1w MPRAGE images can be corrected for receive B1 profiles with low spatial resolution GE images and appropriate post-processing algorithms, both data need being acquired at high spatial resolution in order to address T2* and Proton Density induced image alterations, rich in high spatial frequencies. The implementation of the introduced technique is straightforward, it does not require sophisticated post processing algorithms, and corrected T1w images are produced in addition to, but not instead of, standard magnetization prepared T1w images which represent a gold standard reference in brain imaging. Furthermore, we demonstrate that ratio images also have the potential of producing Maximum Intensity Projection angiograms, derived from the same data, thus inherently registered with T1w structural images.
In order to address inter-scan head motion induced artifacts when dividing the two images, a new sequence was developed with the two 3D acquisitions interleaved. Users can choose between this interleaved approach, which is however more sensitive to intra-scan head motion because of a longer single scan acquisition, and the two shorter separate acquisitions, more sensitive to inter-scan motion.
It is expected that this simple technique could greatly contribute to the fast growing interest in brain tissue contrast at high field, providing increased T1 specificity and allowing for comparing data obtained in different sessions, with different RF coils or with different scanners.
Acknowledgments
We are grateful to the anonymous reviewers for their very useful comments and for bringing to our knowledge the work referenced in (van Gelderen et al., 2006). We want to express our acknowledgements to Gregor Adriany for RF coil support. This study was supported by BTRR - P41 RR008079, MIND Institute, NIH-P30 NS05709. R01 MH070800, R01 EB000331
Appendix 1.
With the following contrast definition between signals A and B:
[16] |
where , the WM/GM contrast in either MPRAGE (MP) or in P/NP images becomes:
[17] |
where
[18] |
From Eq. [3] we can write:
[19] |
where SGE is the signal in GE-PD images
and
[20] |
After Eq. [4] and [5] we see that:
[21] |
For all 3D acquisitions considered here (short TR, small flip angle) we obtain TR≪T1*<T1, which allows the approximation (Deichmann and Haase, 1992; Deichmann et al., 1999):
[22] |
[23] |
From Eqs. [23] and [22] we define the parameter λ summarizing the sources of contrast changes between WM and GM induced by forming the image ratio. λ is a function of tissue parameters (ρ, T1) as well as acquisition parameters (TR, α):
[24] |
Finally, T1 contrast in P/NP image can be expressed as:
[25] |
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
Preliminary reports of this work in abstract form include: i) Van de Moortele, P.F., Moeller, S., Ugurbil, K., 2007. Efficient Bias Reduction for High Resolution T1w Imaging in the Human Brain at 7 Tesla. ISMRM Workshop on Advances in High Field MR, Asilomar. ii) Van de Moortele, P.F., Auerbach, E.J., Olman, C., Yacoubd. E., Ugurbil, K., Moeller, S., 2008b. Unbiased High Resolution T1 Weighted Brain Images at High Field with a New Interleaved 3D-MPRAGE/Proton Density GE sequence. Human Brain Mapping 14th Annual Meeting, Sidney. iii) Van de Moortele, P.-F., Auerbach, E.J., Olman, C., Yacoub, E., Ugurbil, K., Moeller, S., 2008a. Which clinical application(s) would best benefit from submillimeter High T1 contrast whole brain images at 7Tesla with simultaneous angiography perfectly coregistered with anatomical images?, Unsolved Problems and Unmet Needs Symposium, ISMRM 14th Annual meeting, Toronto, p. 172.
The exact time when the k-space origin is sampled (TS) is given by TI+(N/2±mod(N,2))*TR where mod(N,2) is the residual after dividing N by 2. However, for simplicity we consider here TS ≈ TI + TA/2 because TR ≪TS. Also, in case of Partial Fourier acquisition, the actual sampling time of the k-space origin is shifted away from the middle of TA.
References
- Adriany G, Ritter J, Van de Moortele PF, Moeller S, Snyder C, Voje B, Vaughan JT, Ugurbil K. A geometrically adjustable 16 Channel Transceive Transmission Line Array for 7 Tesla. ISMRM 13th Scientific Meeting; Miami. 2005a. p. 673. [Google Scholar]
- Adriany G, Van de Moortele PF, Wiesinger F, Moeller S, Strupp JP, Andersen P, Snyder C, Zhang X, Chen W, Pruessmann KP, Boesiger P, Vaughan T, Ugurbil K. Transmit and receive transmission line arrays for 7 Tesla parallel imaging. Magnetic Resonance in Medicine. 2005b;53:434–445. doi: 10.1002/mrm.20321. [DOI] [PubMed] [Google Scholar]
- Augustinack JC, van der Kouwe AJ, Blackwell ML, Salat DH, Wiggins CJ, Frosch MP, Wiggins GC, Potthast A, Wald LL, Fischl BR. Detection of entorhinal layer II using 7Tesla [correction] magnetic resonance imaging. Ann Neurol. 2005;57:489–494. doi: 10.1002/ana.20426. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Belaroussi B, Milles J, Carme S, Zhu YM, Benoit-Cattin H. Intensity non-uniformity correction in MRI: existing methods and their validation. Med Image Anal. 2006;10:234–246. doi: 10.1016/j.media.2005.09.004. [DOI] [PubMed] [Google Scholar]
- Bernstein MA, Huston J, 3rd, Ward HA. Imaging artifacts at 3.0T. J Magn Reson Imaging. 2006;24:735–746. doi: 10.1002/jmri.20698. [DOI] [PubMed] [Google Scholar]
- Bevington PR. Data Reduction and Analysis for the Physical Sciences. McGraw-Hill; 1969. [Google Scholar]
- Bottomley PA, Andrew ER. RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imaging. Physics in Medicine and Biology. 1978;23:630–643. doi: 10.1088/0031-9155/23/4/006. [DOI] [PubMed] [Google Scholar]
- Collins CM, Smith MB. Signal-to-noise ratio and absorbed power as functions of main magnetic field strength, and definition of “90 degrees” RF pulse for the head in the birdcage coil. Magnetic Resonance in Medicine. 2001;45:684–691. doi: 10.1002/mrm.1091. [DOI] [PubMed] [Google Scholar]
- Collins CM, Yang QX, Wang JH, Zhang X, Liu H, Michaeli S, Zhu XH, Adriany G, Vaughan JT, Anderson P, Merkle H, Ugurbil K, Smith MB, Chen W. Different excitation and reception distributions with a single-loop transmit-receive surface coil near a head-sized spherical phantom at 300 MHz. Magnetic Resonance in Medicine. 2002;47:1026–1028. doi: 10.1002/mrm.10153. [DOI] [PubMed] [Google Scholar]
- Deichmann R, Good CD, Josephs O, Ashburner J, Turner R. Optimization of 3-D MP-RAGE sequences for structural brain imaging. NeuroImage. 2000;12:112–127. doi: 10.1006/nimg.2000.0601. [DOI] [PubMed] [Google Scholar]
- Deichmann R, Good CD, Turner R. RF inhomogeneity compensation in structural brain imaging. Magnetic Resonance in Medicine. 2002;47:398–402. doi: 10.1002/mrm.10050. [DOI] [PubMed] [Google Scholar]
- Deichmann R, Haase A. Quantification of T1 values by SNAPSHOT-FLASH NMR Imaging. Journal of Magnetic Resonance. 1992;96:608–612. [Google Scholar]
- Deichmann R, Hahn D, Haase A. Fast T1 mapping on a whole-body scanner. Magnetic Resonance in Medicine. 1999;42:206–209. doi: 10.1002/(sici)1522-2594(199907)42:1<206::aid-mrm28>3.0.co;2-q. [DOI] [PubMed] [Google Scholar]
- Deichmann R, Schwarzbauer C, Turner R. Optimisation of the 3D MDEFT sequence for anatomical brain imaging: technical implications at 1.5 and 3 T. NeuroImage. 2004;21:757–767. doi: 10.1016/j.neuroimage.2003.09.062. [DOI] [PubMed] [Google Scholar]
- Foo TK, Sawyer AM, Faulkner WH, Mills DG. Inversion in the steady state: contrast optimization and reduced imaging time with fast three-dimensional inversion-recovery-prepared GRE pulse sequences. Radiology. 1994;191:85–90. doi: 10.1148/radiology.191.1.8134602. [DOI] [PubMed] [Google Scholar]
- Ge Y, Zohrabian VM, Grossman RI. Seven-Tesla magnetic resonance imaging: new vision of microvascular abnormalities in multiple sclerosis. Arch Neurol. 2008;65:812–816. doi: 10.1001/archneur.65.6.812. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gelman N, Ewing JR, Gorell JM, Spickler EM, Solomon EG. Interregional variation of longitudinal relaxation rates in human brain at 3.0 T: relation to estimated iron and water contents. Magnetic Resonance in Medicine. 2001;45:71–79. doi: 10.1002/1522-2594(200101)45:1<71::aid-mrm1011>3.0.co;2-2. [DOI] [PubMed] [Google Scholar]
- Glover GH, Hayes CE, Pelc NJ, Edelstein WA, Mueller OM, Hart HR, Hardy CJ, O'Donnel M, Barber WD. Comparison of linear and circular polarization for magnetic resonance imaging. J Magn Reson. 1985;64:255–270. [Google Scholar]
- Haase A. Snapshot FLASH MRI: Application to T1, T2, and Chemical Shift Imaging. Magnetic Resonance in Medicine. 1990;13:77–89. doi: 10.1002/mrm.1910130109. [DOI] [PubMed] [Google Scholar]
- Jenkinson M, Bannister P, Brady M, Smith S. Improved optimization for the robust and accurate linear registration and motion correction of brain images. NeuroImage. 2002;17:825–841. doi: 10.1016/s1053-8119(02)91132-8. [DOI] [PubMed] [Google Scholar]
- Just M, Thelen M. Tissue characterization with T1, T2, and proton density values: results in 160 patients with brain tumors. Radiology. 1988;169:779–785. doi: 10.1148/radiology.169.3.3187000. [DOI] [PubMed] [Google Scholar]
- Kaptein R, Dukstra K, Tarr CE. A Single-Scan Fourier Transform Method for Measuring Spin-Lattice Relaxation Times. Journal of Magnetic Resonance. 1976;24:295–300. [Google Scholar]
- Katscher U, Bornert P, Leussler C, van den Brink JS. Transmit SENSE. Magnetic Resonance in Medicine. 2003;49:144–150. doi: 10.1002/mrm.10353. [DOI] [PubMed] [Google Scholar]
- Keltner JR, Carlson JW, Roos MS, Wong STS, Wong TL, Budinger TF. Electromagnetic Fields of Surface Coil in Vivo NMR at High Frequencies. Magnetic Resonance in Medicine. 1991;22:46–480. doi: 10.1002/mrm.1910220254. [DOI] [PubMed] [Google Scholar]
- Larsson J. Imaging vision: functional mapping of intermediate visual processes in man. Karolinska Institutet; Stockholm, Sweden: 2001. [Google Scholar]
- Laub G. Displays for MR angiography. Magnetic Resonance in Medicine. 1990;14:222–229. doi: 10.1002/mrm.1910140208. [DOI] [PubMed] [Google Scholar]
- Li TQ, van Gelderen P, Merkle H, Talagala L, Koretsky AP, Duyn J. Extensive heterogeneity in white matter intensity in high-resolution T2*-weighted MRI of the human brain at 7.0 T. NeuroImage. 2006;32:1032–1040. doi: 10.1016/j.neuroimage.2006.05.053. [DOI] [PubMed] [Google Scholar]
- Mao W, Smith MB, Collins CM. Exploring the limits of RF shimming for high-field MRI of the human head. Magnetic Resonance in Medicine. 2006;56:918–922. doi: 10.1002/mrm.21013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mugler JP, III, Brookman JR. Three dimensional magnetization-prepared rapid gradient-echo imaging (3D MPRAGE) Magnetic Resonance in Medicine. 1990;15:152–157. doi: 10.1002/mrm.1910150117. [DOI] [PubMed] [Google Scholar]
- Olman C, Van de Moortele PF, Ugurbil K, Yacoub E. Neuroscience. San Diego: 2007. Retinotopic mapping with 7 Tesla fMRI: improved resolution and decreased experiment duration; p. 768. [Google Scholar]
- Peters AM, Brookes MJ, Hoogenraad FG, Gowland PA, Francis ST, Morris PG, Bowtell R. T2* measurements in human brain at 1.5, 3 and 7 T. Magn Reson Imaging. 2007;25:748–753. doi: 10.1016/j.mri.2007.02.014. [DOI] [PubMed] [Google Scholar]
- Pfeuffer J, Adriany G, Shmuel A, Yacoub E, Van De Moortele PF, Hu X, Ugurbil K. Perfusion-based high-resolution functional imaging in the human brain at 7 Tesla. Magnetic Resonance in Medicine. 2002;47:903–911. doi: 10.1002/mrm.10154. [DOI] [PubMed] [Google Scholar]
- Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR Phased Array. Magnetic Resonance in Medicine. 1990;16:192–225. doi: 10.1002/mrm.1910160203. [DOI] [PubMed] [Google Scholar]
- Rooney WD, Johnson G, Li X, Cohen ER, Kim SG, Ugurbil K, Springer CS., Jr Magnetic field and tissue dependencies of human brain longitudinal (1)H(2)O relaxation in vivo. Magnetic Resonance in Medicine. 2007;57:308–318. doi: 10.1002/mrm.21122. [DOI] [PubMed] [Google Scholar]
- Setsompop K, Wald LL, Alagappan V, Gagoski B, Hebrank F, Fontius U, Schmitt F, Adalsteinsson E. Parallel RF transmission with eight channels at 3 Tesla. Magnetic Resonance in Medicine. 2006;56:1163–1171. doi: 10.1002/mrm.21042. [DOI] [PubMed] [Google Scholar]
- Smith SM, Jenkinson M, Woolrich MW, Beckmann CF, Behrens TE, Johansen-Berg H, Bannister PR, De Luca M, Drobnjak I, Flitney DE, Niazy RK, Saunders J, Vickers J, Zhang Y, De Stefano N, Brady JM, Matthews PM. Advances in functional and structural MR image analysis and implementation as FSL. NeuroImage. 2004;23(1):S208–219. doi: 10.1016/j.neuroimage.2004.07.051. [DOI] [PubMed] [Google Scholar]
- Thomas BP, Welch EB, Niederhauser BD, Whetsell WO, Jr, Anderson AW, Gore JC, Avison MJ, Creasy JL. High-resolution 7T MRI of the human hippocampus in vivo. J Magn Reson Imaging. 2008;28:1266–1272. doi: 10.1002/jmri.21576. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Van de Moortele PF, Auerbach EJ, Olman C, Yacoub E, Ugurbil K, Moeller S. Which clinical application(s) would best benefit from submillimeter High T1 contrast whole brain images at 7Tesla with simultaneous angiography perfectly coregistered with anatomical images?. Unsolved Problems and Unmet Needs Symposium; ISMRM 14th Annual meeting; Toronto. 2008a. p. 172. [Google Scholar]
- Van de Moortele PF, Akgun C, Adriany G, Moeller S, Ritter J, Collins CM, Smith MB, Vaughan JT, Ugurbil K. B(1) destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil. Magnetic Resonance in Medicine. 2005;54:1503–1518. doi: 10.1002/mrm.20708. [DOI] [PubMed] [Google Scholar]
- Van de Moortele PF, Auerbach EJ, Olman C, Yacoubd E, Ugurbil K, Moeller S. Unbiased High Resolution T1 Weighted Brain Images at High Field with a New Interleaved 3D-MPRAGE/Proton Density GE sequence. Human Brain Mapping 14th Annual Meeting; Sidney. 2008b. [Google Scholar]
- Van de Moortele PF, Moeller S, Ugurbil K. Efficient Bias Reduction for High Resolution T1w Imaging in the Human Brain at 7 Tesla. ISMRM Workshop on Advances in High Field MR; Asilomar. 2007. [Google Scholar]
- van Gelderen P, Koretsy AP, de Zwart JA, Duyn JH. A Simple B1 Correction Method for High Resolution Neuroimaging. 14th Scientific Meeting ISMRM; Seattle, WA. 2006. p. 2355. [Google Scholar]
- Vaughan JT, Adriany G, Snyder CJ, Tian J, Thiel T, Bolinger L, Liu H, DelaBarre L, Ugurbil K. Efficient high-frequency body coil for high-field MRI. Magnetic Resonance in Medicine. 2004;52:851–859. doi: 10.1002/mrm.20177. [DOI] [PubMed] [Google Scholar]
- Vaughan JT, Garwood M, Collins CM, Liu W, DelaBarre L, Adriany G, Andersen P, Merkle H, Goebel R, Smith MB, Ugurbil K. 7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images. Magnetic Resonance in Medicine. 2001;46:24–30. doi: 10.1002/mrm.1156. [DOI] [PubMed] [Google Scholar]
- Wang J, Qiu M, Constable RT. In vivo method for correcting transmit/receive nonuniformities with phased array coils. Magnetic Resonance in Medicine. 2005;53:666–674. doi: 10.1002/mrm.20377. [DOI] [PubMed] [Google Scholar]
- Warntjes JBM, Dahlqvist O, Lundberg P. Novel Method for Rapid, Simultaneous T1, T*2, and Proton Density Quantification. Magnetic Resonance in Medicine. 2007;57:528–537. doi: 10.1002/mrm.21165. [DOI] [PubMed] [Google Scholar]
- Wiesinger F, Van de Moortele PF, Adriany G, De Zanche N, Ugurbil K, Pruessmann KP. Parallel imaging performance as a function of field strength--an experimental investigation using electrodynamic scaling. Magnetic Resonance in Medicine. 2004;52:953–964. doi: 10.1002/mrm.20281. [DOI] [PubMed] [Google Scholar]
- Yacoub E, Shmuel A, Pfeuffer J, Van De Moortele PF, Adriany G, Andersen P, Vaughan JT, Merkle H, Ugurbil K, Hu X. Imaging brain function in humans at 7 Tesla. Magnetic Resonance in Medicine. 2001;45:588–594. doi: 10.1002/mrm.1080. [DOI] [PubMed] [Google Scholar]
- Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magnetic Resonance in Medicine. 2007;57:192–200. doi: 10.1002/mrm.21120. [DOI] [PubMed] [Google Scholar]