Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 1983 Sep;18(3):697–701. doi: 10.1128/jcm.18.3.697-701.1983

Evaluation of a two-minute test for urine screening.

M T Pezzlo, M A Wetkowski, E M Peterson, L M de la Maza
PMCID: PMC270877  PMID: 6355154

Abstract

A study was conducted to evaluate the ability of a urine filtration system (Bac-T-Screen, Marion Laboratories, Inc., Kansas City, Mo.) to detect negative urine cultures within 2 min. A total of 1,000 urine specimens were tested with the Bac-T-Screen and compared with a standard semiquantitative culture plate method and the Autobac system (General Diagnostics, Warner-Lambert Co., Morris Plains, N.J.). Of the 1,000 clean voided urine specimens tested, 246 specimens had colony counts greater than or equal to 10(5) CFU/ml by the culture plate method. Of these, the Bac-T-Screen detected 65.4% (161 of 246), and the Autobac detected 63.0% (155 of 246). When pure cultures of diphtheroids, lactobacilli, and viridans streptococci other than group D and cultures containing multiple organisms were considered to be contaminants and, therefore, were excluded, there were 106 pure cultures of probable pathogens of which the Bac-T-Screen detected 76.4% (81 of 106) and the Autobac detected 90.6% (96 of 106). Some 133 specimens were uninterpretable with the Bac-T-Screen because 36 clogged the filter and 97 left a residual pigment on the filter. A majority of those clogging the filter (69.4%) had positive plate counts, whereas the majority of the pigmented urines had negative plate counts. Of those urine specimens tested. 754 were negative by the culture plate method. The false-positive rates for Bac-T-Screen and Autobac were 16.2 and 5.8%, respectively. As a urine screen, the Bac-T-Screen has a negative predictive value comparable to the Autobac system and has the advantage of being a 2-min test.

Full text

PDF
697

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Hale D. C., Thrupp L. D., Matsen J. M. Evaluation of urine culture screening by light-scatter photometry. Am J Clin Pathol. 1981 Aug;76(2):208–211. doi: 10.1093/ajcp/76.2.208. [DOI] [PubMed] [Google Scholar]
  2. Hale D. C., Wright D. N., McKie J. E., Isenberg H. D., Jenkins R. D., Matsen J. M. Rapid screening for bacteriuria by light scatter photometry (Autobac): a collaborative study. J Clin Microbiol. 1981 Jan;13(1):147–150. doi: 10.1128/jcm.13.1.147-150.1981. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Heinz T., Pezzlo M., Thrupp L. Clinical and laboratory efficiency of rapid screening urine cultures by nephelometry. Am J Clin Pathol. 1982 Mar;77(3):305–311. doi: 10.1093/ajcp/77.3.305. [DOI] [PubMed] [Google Scholar]
  4. Jenkins R. D., Hale D. C., Matsen J. M. Rapid semiautomated screening and processing of urine specimens. J Clin Microbiol. 1980 Mar;11(3):220–225. doi: 10.1128/jcm.11.3.220-225.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. KASS E. H. Asymptomatic infections of the urinary tract. Trans Assoc Am Physicians. 1956;69:56–64. [PubMed] [Google Scholar]
  6. Kelly M. T., Balfour L. C. Evaluation and optimization of urine screening by Autobac. J Clin Microbiol. 1981 Apr;13(4):677–680. doi: 10.1128/jcm.13.4.677-680.1981. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Lamb V. A., Dalton H. P., Wilkins J. R. Electrochemical method for the early detection of urinary-tract infections. Am J Clin Pathol. 1976 Jul;66(1):91–95. doi: 10.1093/ajcp/66.1.91. [DOI] [PubMed] [Google Scholar]
  8. Nicholson D. P., Koepke J. A. The Automicrobic System for urines. J Clin Microbiol. 1979 Dec;10(6):823–833. doi: 10.1128/jcm.10.6.823-833.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Pezzlo M. T., Tan G. L., Peterson E. M., de la Maza L. M. Screening of urine cultures by three automated systems. J Clin Microbiol. 1982 Mar;15(3):468–474. doi: 10.1128/jcm.15.3.468-474.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Ransohoff D. F., Feinstein A. R. Problems of spectrum and bias in evaluating the efficacy of diagnostic tests. N Engl J Med. 1978 Oct 26;299(17):926–930. doi: 10.1056/NEJM197810262991705. [DOI] [PubMed] [Google Scholar]
  11. Wallis C., Melnick J. L., Longoria C. J. Colorimetric method for rapid determination of bacteriuria. J Clin Microbiol. 1981 Sep;14(3):342–346. doi: 10.1128/jcm.14.3.342-346.1981. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES