Abstract
The capability of the Abbott Avantage system to identify 10 species of commonly isolated glucose nonfermentative or oxidase-positive gram-negative bacilli in a 5-h test period was evaluated in a collaborative study. The Avantage nonenteric data base uses 20 biochemical test reactions performed in an expanded Abbott bacterial identification cartridge plus the results of a manual oxidase test. The species included in the Avantage data base are Acinetobacter anitratus, Acinetobacter Iwoffi, Aeromonas hydrophila, Flavobacterium meningosepticum-IIb group, Pseudomonas aeruginosa, Pseudomonas cepacia, Pseudomonas fluorescens-putida group, Pseudomonas maltophilia, Pasteurella multocida, and Plesiomonas shigelloides. The collaborative study included the testing of 200 coded challenge strains in all three laboratories and the subsequent testing of an additional group of 100 to 200 clinical isolates recovered independently by each laboratory. Reference identifications for all isolates were determined by conventional biochemical test reactions. The overall accuracy of identification of the coded challenge strains for the three laboratories was 97%, whereas 95% of 437 clinical isolates and selected stock cultures of clinical derivation were identified correctly.
Full text
PDF





Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Appelbaum P. C., Stavitz J., Bentz M. S., von Kuster L. C. Four methods for identification of gram-negative nonfermenting rods: organisms more commonly encountered in clinical specimens. J Clin Microbiol. 1980 Aug;12(2):271–278. doi: 10.1128/jcm.12.2.271-278.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Barry A. L., Badal R. E. Reliability of early identifications obtained with Enterobacteriaceae-plus biochemical cards in the automicrobic system. J Clin Microbiol. 1982 Aug;16(2):257–265. doi: 10.1128/jcm.16.2.257-265.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Barry A. L., Gavan T. L., Badal R. E., Telenson M. J. Sensitivity, specificity, and reproducibility of the automicrobic system (with the Enterobacteriaceae-plus biochemical card) for identifying clinical isolates of Gram- negative bacilli. J Clin Microbiol. 1982 Apr;15(4):582–588. doi: 10.1128/jcm.15.4.582-588.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Barry A. L., Gavan T. L., Smith P. B., Matsen J. M., Morello J. A., Sielaff B. H. Accuracy and precision of the autobac system for rapid identification of Gram-negative bacilli: a collaborative evaluation. J Clin Microbiol. 1982 Jun;15(6):1111–1119. doi: 10.1128/jcm.15.6.1111-1119.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chester V., Cleary T. J. Evaluation of the Minitek system for identification of nonfermentative and nonenteric fermentative Gram-negative bacteria. J Clin Microbiol. 1980 Oct;12(4):509–516. doi: 10.1128/jcm.12.4.509-516.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Costigan W. J., Hollick G. E. Use of the Autobac IDX system for rapid identification of Enterobacteriaceae and nonfermentative gram-negative bacilli. J Clin Microbiol. 1984 Feb;19(2):301–302. doi: 10.1128/jcm.19.2.301-302.1984. [DOI] [PMC free article] [PubMed] [Google Scholar]
- DiPersio J. R., Dyke J. W., Vannest R. D. Evaluation of the updated MS-2 Bacterial Identification system in comparison with the API 20E system. J Clin Microbiol. 1983 Jul;18(1):128–135. doi: 10.1128/jcm.18.1.128-135.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Isenberg H. D., Sampson-Scherer J. Clinical laboratory evaluation of a system approach to the recognition of nonfermentative or oxidase-producing gram-negative, rod-shaped bacteria. J Clin Microbiol. 1977 Mar;5(3):336–340. doi: 10.1128/jcm.5.3.336-340.1977. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Johnson J. E., Brinkley A. W. Comparison of the AutoMicrobic system and a conventional tube system for identification of nonfermentative and oxidase-positive gram-negative bacilli. J Clin Microbiol. 1982 Jan;15(1):25–27. doi: 10.1128/jcm.15.1.25-27.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kelly M. T., Matsen J. M., Morello J. A., Smith P. B., Tilton R. C. Collaborative clinical evaluation of the Autobac IDX system for identification of gram-negative bacilli. J Clin Microbiol. 1984 Apr;19(4):529–533. doi: 10.1128/jcm.19.4.529-533.1984. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McCracken A. W., Martin W. J., McCarthy L. R., Schwab D. A., Cooper B. H., Helgeson N. G., Prowant S., Robson J. Evaluation of the MS-2 system for rapid identification of Enterobacteriaceae. J Clin Microbiol. 1980 Nov;12(5):684–689. doi: 10.1128/jcm.12.5.684-689.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Oberhofer T. R. Comparison of the API 20E and Oxi/Ferm systems in identification of nonfermentative and oxidase-positive fermentative bacteria. J Clin Microbiol. 1979 Feb;9(2):220–226. doi: 10.1128/jcm.9.2.220-226.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Oberhofer T. R. Use of the API 20E, Oxi/Ferm, and Minitek systems to identify nonfermentative and oxidase-positive fermentative bacteria: seven years of experience. Diagn Microbiol Infect Dis. 1983 Sep;1(3):241–256. doi: 10.1016/0732-8893(83)90024-x. [DOI] [PubMed] [Google Scholar]
- Shayegani M., Lee A. M., McLynn D. M. Evaluation of the Oxi/Ferm tube system for identification of nonfermentative Gram-negative bacilli. J Clin Microbiol. 1978 Jun;7(6):533–538. doi: 10.1128/jcm.7.6.533-538.1978. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shayegani M., Maupin P. S., McGlynn D. M. Evaluation of the API 20E system for identification of nonfermentative Gram-negative bacteria. J Clin Microbiol. 1978 Jun;7(6):539–545. doi: 10.1128/jcm.7.6.539-545.1978. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sielaff B. H., Matsen J. M., McKie J. E. Novel approach to bacterial identification that uses the autobac system. J Clin Microbiol. 1982 Jun;15(6):1103–1110. doi: 10.1128/jcm.15.6.1103-1110.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wellstood-Nuesse S. Comparison of the Minitek system with conventional methods for identification of nonfermentative and oxidase-positive fermentative gram-negative bacilli. J Clin Microbiol. 1979 Apr;9(4):511–516. doi: 10.1128/jcm.9.4.511-516.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]