Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 1981 Oct;14(4):389–392. doi: 10.1128/jcm.14.4.389-392.1981

Comparison of a slide blood culture system with a supplemented peptone broth culture method.

L E Bryan
PMCID: PMC271989  PMID: 7026607

Abstract

A slide blood culture system (Roche Diagnostics, Div. Hoffman-La Roche, Inc., Montreal, Canada; Roche BCB) was compared with a supplemented peptone broth Vacutainer method (Becton, Dickinson & Co., Rutherford, N.J.) on blood samples taken from the same 1,209 patients. Significantly more clinically important isolates were isolates with the Roche BCB system, and technical processing time was reduced. However, significantly more contaminants were isolated with the Roche BCB system, and it could not readily be adapted for anaerobic culture. Contamination was reduced by careful tightening of the slide to the bottle top to prevent any leakage. Overall, the BCB system is a satisfactory method for aerobic blood culture, markedly reducing technical processing time relative to most other blood culture methods.

Full text

PDF

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Roberts F. J. A review of positive blood cultures: identification and source of microorganisms and patterns of sensitivity to antibiotics. Rev Infect Dis. 1980 May-Jun;2(3):329–339. doi: 10.1093/clinids/2.3.329. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES