Rapid identification of Enterobacteriaceae with the micro-ID system versus API 20E and conventional media.

A L Barry, R E Badal
PMCID: PMC273155  PMID: 385617

Abstract

The Micro-ID system for rapid (4 h) identification of Enterobacteriaceae was evaluated by testing 433 enteric bacilli and 9 other gram-negative bacilli. Each isolate was identified with conventional tubed media and was also tested in the Micro-ID and API 20E systems. The overall accuracy of both systems was 97%. Micro-ID tests for the Voges-Proskauer reaction, indole and H2S production, and ornithine and lysine decarboxylase all demonstrated a 97 to 99% correlation with conventional methods. Only 86% of the Micro-ID urease tests agreed with Christenson urea agar. Two inoculum densities were tested in Micro-ID panels, with 157 stock cultures. Over 90% of the tests were unaffected by changes in inoculum density. Tests with four control strains suggested that the Micro-ID system was more reproducible when a light inoculum was used. The Micro-ID system was found to be a very convenient method for rapid, accurate, and precise identification of the Enterobacteriaceae.

Full text

PDF

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Aldridge K. E., Gardner B. B., Clark S. J., Matsen J. M. Comparison of micro-ID, API 20E, and conventional media systems in identification of Enterobacteriaceae. J Clin Microbiol. 1978 Jun;7(6):507–513. doi: 10.1128/jcm.7.6.507-513.1978. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Butler D. A., Lobregat C. M., Gavan T. L. Reproducibility of the analytab (API 20E) system. J Clin Microbiol. 1975 Oct;2(4):322–326. doi: 10.1128/jcm.2.4.322-326.1975. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Matsen J. M., Sherris J. C. Comparative study of the efficacy of seven paper-reagent strips and conventional biochemical tests in identifying gram-negative organisms. Appl Microbiol. 1969 Sep;18(3):452–457. doi: 10.1128/am.18.3.452-457.1969. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Murray P. R. Standardization of the Analytab Enteric (API 20E) system to increase accuracy and reproducibility of the test for biotype characterization of bacteria. J Clin Microbiol. 1978 Jul;8(1):46–49. doi: 10.1128/jcm.8.1.46-49.1978. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Smith P. B., Tomfohrde K. M., Rhoden D. L., Balows A. API system: a multitube micromethod for identification of Enterobacteriaceae. Appl Microbiol. 1972 Sep;24(3):449–452. doi: 10.1128/am.24.3.449-452.1972. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES