. 2008 Jul-Sep;12(3):96–101. doi: 10.4103/0972-5229.43677

Prediction of arterial blood gas values from arterialized earlobe blood gas values in patients treated with mechanical ventilation

Azim Honarmand 1,, Mohammadreza Safavi 1
PMCID: PMC2738313  PMID: 19742254

Abstract

Background/Objective:

Arterial blood gas (ABG) analysis is useful in evaluation of the clinical condition of critically ill patients; however, arterial puncture or insertion of an arterial catheter may sometimes be difficult and cause many complications. Arterialized ear lobe blood samples have been described as adequate to gauge gas exchange in acute and chronically ill pediatric patients.

Purpose:

This study evaluates whether pH, partial pressure of oxygen (PO2), partial pressure of carbon dioxide (PCO2), base excess (BE), and bicarbonate (HCO3) values of arterialized earlobe blood samples could accurately predict their arterial blood gas analogs for adult patients treated by mechanical ventilation in an intensive care unit (ICU).

Setting:

A prospective descriptive study

Methods:

Sixty-seven patients who were admitted to ICU and treated with mechanical ventilation were included in this study. Blood samples were drawn simultaneously from the radial artery and arterialized earlobe of each patient.

Results:

Regression equations and mean percentage-difference equations were derived to predict arterial pH, PCO2, PO2, BE, and HCO3-values from their earlobe analogs. pH, PCO2, BE, and HCO3 all significantly correlated in ABG and earlobe values. In spite of a highly significant correlation, the limits of agreement between the two methods were wide for PO2. Regression equations for prediction of pH, PCO2, BE, and HCO3- values were: arterial pH (pHa) = 1.81+ 0.76 × earlobe pH (pHe) [r = 0.791, P < 0.001]; PaCO2 = 1.224+ 1.058 × earlobePCO2 (PeCO2) [r = 0.956, P < 0.001]; arterial BE (BEa) = 1.14+ 0.95 × earlobe BE (BEe) [r= 0.894, P < 0.001], and arterial HCO3- (HCO3-a) = 1.41+ earlobe HCO3(HCO3-e) [r = 0.874, P < 0.001]. The predicted ABG values from the mean percentage-difference equations were derived as follows: pHa = pHe × 1.001; PaCO2 = PeCO2 × 0.33; BEa = BEe × 0.57; and HCO3-a = HCO3-e × 1.06.

Conclusions:

Arterialized earlobe blood gas can accurately predict the ABG values of pH, PCO2, BE, and HCO3- for patients who do not require regular continuous blood pressure measurements and close monitoring of arterial PO2 measurements.

Keywords: Arterialized earlobe blood gas, critically illness, mechanical ventilation

Introduction

Measurement of arterial blood gas tensions is routinely used to assess gas exchange in patients with acute and chronic respiratory disorders. Arterial blood gas (ABG) sampling represents the gold standard method for acquiring patients' acid-base status. The most common complications associated with arterial puncture are pain, arterial injury, aneurysm formation, hemorrhage, and thrombosis with distal ischaemia. The risks increase with repeated arterial punctures, especially with insertion of a catheter when performed by inexperienced individuals.[1] Additionally, this procedure carries a small but appreciable risk of needle stick injury to health care workers, with the consequent risk of transmission of blood borne viruses such as hepatitis C and human immunodeficiency virus (HIV).[2] In the 1960s, it was proposed that blood gas values could be measured using arterialized earlobe blood samples.[3] Arterialized ear lobe blood gas samplings are easier to obtain and a less invasive way of evaluating acid-base status in intensive care unit. It avoids the risks of arterial punctures. It is based on the assumption that provided sufficient vasodilatation can be achieved locally by means of topical application of a vaso-active cream on the earlobe, the arterialized earlobe oxygen tension resembles the arterial oxygen tension due to convergence of arterial and venous oxygen tension.[4] Techniques for sampling arterialized capillary blood from the finger pulp and the earlobe were first described over two decades ago but, although close agreement between arterial values and earlobe samples has been demonstrated in normal subjects, this technique is not in common usage. The main reasons for this appear to be lack of knowledge of its existence and uncertainty over its accuracy.[5] Initial[3,6,7] or more recent[5] studies have concluded that the earlobe method might be accurate enough to replace arterial blood samples for clinical purposes. This opinion is based mainly on positive and strong correlations that have been found between the two methods. However, most investigators have sought correlation using simple regression analysis instead of the method of BLAND and ALTMAN.[8] This approach has probably rendered the conclusion of most studies flawed by statistical bias. In fact, two recent studies using BLAND and ALTMAN[8] analysis have stated that arterialized earlobe PO2 often underestimates arterial PO2, therefore making this method unsuitable for clinical assessment.[9,10] Several studies have shown good correlation between capillary blood, venous blood, and arterial blood gas values in pediatric intensive care units.[1113] However, arterialized earlobe oxygen tension often underestimates arterial oxygen tension[4] and is not fully validated in adult patients with acute respiratory failure receiving mechanical ventilation. The purpose of this study was to investigate the correlation between simultaneous arterial blood gas and arterialized earlobe blood samples and to establish whether pH, partial pressure of oxygen (PO2), partial pressure of carbon dioxide (PCO2), base excess (BE), and bicarbonate (HCO3-) values of arterialized earlobe blood samples could accurately predict their arterial blood gas analogs for patients with acute respiratory failure treated by mechanical ventilation in an intensive care unit (ICU).

Materials and Methods

Sixty seven patients with acute respiratory failure who were admitted to a multidisciplinary adult intensive care unit between May 2005 and August 2005 receiving mechanical ventilation, were enrolled in this prospective descriptive study. The study protocol was approved by the local institutional ethics committee, and written informed consent was obtained from each patient or his or her family. Patients with hypotension (systolic blood pressure less than or equal to 90 mm Hg), hypertension (systolic pressure above 140 with a diastolic pressure above 90), hypothermia (axillary temperature <36°C), and hyperthermia (axillary temperature >38°C), severe sepsis, multiorgan failure, or chronic lung disease were excluded from the study. No patient was in cardiovascular shock.

Blood samples were drawn simultaneously from the radial artery and the arterialized earlobe of each patient in a sitting position. Blood gases were obtained if the patient needed blood gases for clinical decisions. All patients were receiving mechanical ventilation for 48h or more with a fraction of inspired oxygen (FiO2) of 0.5 or less and positive end-expiratory pressure of 5 cm H2O.

Arterial samples

If the patient had an arterial line, they were used for blood sampling. After withdrawing 5 ml of blood from the lines with a non-heparinised syringe, 1 ml of arterial blood was obtained with another similar syringe and transferred, as soon as possible, to a heparinized 0.75 ml capillary tube. If the patient did not have an arterial line, arterial punctures with aseptic precautions were carried out, and blood was transferred directly into an identical capillary tube.

Earlobe samples

After aseptic cleaning, the lateral distal portion of the earlobe was punctured with a scalpel blade (surgical blade No. 11, Troge®Blades, Hamburge, Germany) and blood gases samples were obtained by “contact” with the capillary tube's tip. A short manual massage was necessary in some instances. Two sides of the tube were closed by fingers to avoid air bubbles. All samples were obtained by the investigators. Arterialized samples were collected in heparinized glass capillaries (Modulohm A/S, Vasekaer, Herlev, Denmark) and immediately introduced into the blood gas analyzer (AVL Compact 3, Roche Diagnostics GmbH, Mannheim, Germany) followed within 2 min by arterial samples. pH, PO2, PCO2, BE, and HCO3 values were recorded.

Statistical methods

The statistical analysis for assessing agreement between arterial and arterialized blood gases was performed according to BLAND and ALTMAN.[8] SPSS.11.5 for Windows was used for statistical analysis. Pearson correlation coefficients were determined for correlation and regression equations and mean percentage-difference equations were derived to predict arterial pH, PCO2, BE, and HCO3 - values from their arterialized ear lobe blood gas analogs. For all analyses, P<0.05 was considered statistically significant.

Results

Sixty-seven consecutive adult patients were studied. Blood samples were drawn simultaneously from arterialized earlobe and radial artery. Admission diagnoses included head trauma, abdominal mass, gastrointestinal bleeding, lung contusion, bowel obstruction, myasthenia gravis, epilepsy, pneumonia, meningitis, diabetic ketoacidosis, stroke, brain tumor, lung cancer, and intoxication [Table 1]. The mean (SD) age of the patients was 47.57 (19.51). Fifty (74.6%) were males and 17 (25.4%) were females. No complication in the drawing of blood samples was observed with either method. pH, PCO2, BE, and HCO3 were all significantly correlated in ABG and earlobe samples [Tables 2 and 3]. The range of arterial PO2 values was 6.8–19.5 kPa (51.1–146.5mmHg), mean 11.3 kPa (84.77mmHg). The range and mean of arterial PCO2 values was 3.01–6.5 kPa (22.6–48.6 mmHg) and 4.7 kPa (35.08 mmHg) respectively. The relationships between arterial and earlobe samples for PO2 and PCO2 are shown in [Figure 1a and b]. The correlation coefficients were 0.734 (P<0.0001) and 0.956 (P<0.0001) respectively. Despite this highly significant correlation, regression lines were slightly different from lines of identity, particularly for PO2. In Figure 2a and b, differences between the two methods (arterial - arterialized values) were plotted against means of arterial and arterialized PO2PO2 or PCO2. The mean (bias) ±SD and the range of the differences, as well as the 95% confidence intervals for the lower and upper limit of agreement were reported in Table 4. Arterialized earlobe PO2 was lower than arterial PO2 in most cases, and the difference increased as the arterial PO2 increased. These results show that the limits of agreement for PO2 were wide, reveal a lack of agreement between the two methods. For PCO2, on the other hand, the mean difference between the two methods was close to zero, and the limits of agreement were narrower. Regression equations for prediction of pH, PCO2,, BE, and HCO3 - values were: arterial pH (pHa) = 1.81+ 0.76 × earlobe pH (pHe) [r = 0.791, P < 0.001]; PCO2, = 1.224+ 1.06 × earlobe PCO2, (PeCO2) [r = 0.956, P < 0.001]; arterial BE (BEa) = 1.14+ 0.95 × earlobe BE (BEe) [r= 0.894, P < 0.001], and arterial HCO3- (HCO3-a) = 1.41+ earlobe HCO3 (HCO3-e) [r = 0.874, P < 0.001]. The predicted ABG values from the mean percentage-difference equations were derived as follows: pHa = pHe × 1.001; PCO2, = PeCO2 × 0.33; BEa = BEe × 0.57; and HCO3-a = HCO3-e × 1.06.

Table 1.

Clinical diagnoses of the patients

Diagnosis n (%)
Head trauma 27 (40)
Stroke 11 (16.5)
Pneumonia 8 (12)
Epilepsy 4 (6)
Gastrointestinal bleeding 3 (4.5)
Bowel obstruction 3 (4.5)
Abdominal mass 2 (3)
Lung contusion 2 (3)
Meningitis 2 (3)
Myasthenia gravis 1 (1.5)
Diabetic ketoacidosis 1 (1.5)
Brain tumor 1 (1.5)
Lung cancer 1 (1.5)
Intoxication 1 (1.5)
Total 67 (100)

Table 2.

Correlation of arterial and arterialized earlobe blood samples for pH, PO2, PCO2, BE, and HCO3 in patients

pH PO2 PCO2 BE HCO3





Arterial Earlobe Arterial Earlobe Arterial Earlobe Arterial Earlobe Arterial Earlobe
Arterial 1.000 0.791 1.000 0.734 1.000 0.774 1.000 0.894 1.000 0.874
P<0.001 P <0.001 P <0.001 P <0.001 P <0.001
Earlobe 0.791 1.000 0.734 1.000 0.774 1.000 0.894 1.000 0.874 1.000
P<0.001 P <0.001 P <0.001 P <0.001 P <0.001

Table 3.

Regression of arterial blood gas values on arterialized earlobe blood samples values

Arterial Earlobe Constant (SE) β (SE) R2 SE of estimate
pH 1.81 (0.537) 0.757 (0.073) 0.626 0.029
PCO2 11.438 (2.483) 0.703 (0.071) 0.599 5.271
PO2 10.522 (9.877) 1.093 (0.126) 0.538 16.199
BE 1.138 (0.317) 0.952 (0.059) 0.800 2.183
HCO3 1.406 (1.438) 0.996 (0.069) 0.764 2.644

Figure 1.

Figure 1

Correlation between radial artery and and arterialized earlobe blood gas, a) Partial pressure of oxygen (PO2) values, b) Partial pressure of carbon dioxide (PCO2) values …………., line of identity: ———— regression lines

Figure 2.

Figure 2

a) Differences in arterial –arterialized partial pressure of oxygen values (ΔPO2) plotted against mean of arterial and arterialized PO2 values. b) Differences in arterial-arterialized partial pressure of carbon dioxide values (ΔPCO2) plotted against mean of arterial and arterialized PCO2 values

Table 4.

Limits of agreement of the differences in PO2 and PCO2 values between arterial and arterialized earlobe blood samples

ΔPO2 ΔPCO2


kPa mmHg kPa mmHg
Mean±SD 1.02±2.15 7.7±16.14 0.19±0.78 1.45±5.88
Range 0.04-6.32 0.3-47.4 -3.52-0.32 -26.5-2.4
95% CI of
 mean+ 2SD 3.78-6.89 28.35-51.61 1.37-2.15 10.33-16.09
 mean - 2SD -4.83− -1.68 -36.21− -12.59 -1.77− -0.99 -13.19− -7.43

ΔPO2: Difference in partial pressure of oxygen (arterial-arterialized); ΔPCO2: Difference in partial pressure of carbon dioxide (arterial-arterialized). 95% CI: 95% confidence interval. (1 mmHg=133.32 Pa)

Discussion

Acid-base analysis is essential for management of patients in ICU, yielding valuable information about a variety of disease processes.[1,12,14] Non-invasive methods, such as pulse oximetry, transcutaneous monitoring of oxygen and carbon dioxide, and end tidal carbon dioxide have been proven to be useful,[1] but they do not give information about pH, PO2, BE and bicarbonate.

Arterial blood gases are frequently determined in the ICU. This is, however, an invasive way of monitoring blood gas and there are complications, mostly local hematoma related to arterial puncture. The procedure itself is technically difficult and painful.

Arterial lines are usually placed in unstable patients in the ICU who need close monitoring of PO2. When an arterial line is not in place, arterial or venous blood gas (VBG) values continue to be obtained and used for clinical monitoring and management decisions.

Earlobe blood gas sampling is a less invasive way of evaluating acid-base status in a well perfused patient. If a blood gas value determined by earlobe samples could be used to show patients' acid-base status and guide their management with the same accuracy as arterial sampling, this would be preferable because of ease of blood sample collection.

For many years, clinicians have been looking for alternatives to ABG sampling in both children and adults, and studies have investigated ABG, VBG, and capillary (CBG) blood gas samples and the correlation between the values.[1517]

Numerous studies published many years ago concluded that the method using arterialized earlobe blood for PO2 and PCO2 analysis was accurate enough to replace arterial blood samples for clinical purposes.[3,6,7,1820] However, the validity of this method has been discussed in two studies,[9,10] showing with the analysis of BLAND and ALTMAN,[8] PO2 was usually lower in earlobe than in arterial blood, and that the limits of agreement were wide between the two methods.

Studies comparing ABG and CBG, ABG and VBG samples in diabetic ketoacidosis, and ABG–CBG blood gases values in stable pediatric intensive care unit patients have shown good correlation among ABG, VBG, and CBG samples.[13] However, there have been no studies comparing simultaneously obtained ABG and earlobe samplings in stable adult ICU patients treated by mechanical ventilation.

We studied sixty-seven simultaneously obtained arterial blood gas and arterialized earlobe blood samples of patients and showed pH, PCO2, BE, HCO3 were all correlated in arterial and earlobe blood gases in normotensive and normothermic patients. Although there was a significant correlation for PO2 in these patients, it was lower.

Sauty et al, compared arterial and arterialized earlobe blood samples in 115 consecutive adult patients and concluded that, in adult patients, arterialized earlobe blood PO2 is not a reliable mirror of arterial PO2.[9] The main cause of underestimation of arterial PO2 in earlobe samples is insufficient arterialization of blood, corresponding to a certain venous admixture. The effect of a given venous admixture in earlobe blood depends on the arterio-venous PO2 difference: the larger the arterio-venous PO2 difference, the wider the discrepancy between earlobe and arterial PO2. This is one likely reason for the unreliable PO2 values measured in arterialized earlobe blood in patients breathing 100% oxygen.[19,21,22]

Because the arterio-venous PO2 difference is large in subjects with normal arterial PO2, a small venous admixture in earlobe blood will result in a greater discrepancy between earlobe and arterial PO2. Our data supports this observation.

It must be emphasized that in our study, there was no patient with severe sepsis, multiorgan failure, cardiovascular shock, or chronic lung disease because inclusion of these patients might have caused poorer correlation between ear and arterial PO2values due to venous admixture.

Indeed, [Figure 2a] shows that arterial PO2, when in the normal range, was often markedly underestimated by the earlobe sample. Interestingly and despite fewer normal PO2 values, the study of PITKIN et al,[5] also showed a trend towards increased difference between arterial and arterialized PO2 with increasing mean PO2 values. Accordingly, and as reported by the same authors, we observed a better agreement between the two methods for arterial PO2 values lower than 8.0 kPa (60 mmHg), where the effect of venous admixture is smaller.

On the other hand, there was a good agreement between earlobe and arterial values of PCO2, as previously reported.[5] This reflects the insignificant effect of venous admixture, due to the comparatively smaller arterio-venous PCO2 difference.

The usefulness of each method should be weighed according to its advantages and inconveniences. The advocated advantages of the earlobe method are that it is safe and can be performed by non-medical staff.[23] However, as the collection of an earlobe blood sample must be fully aseptic,[6] the method requires trained personnel.

Although complications of arterial punctures have been described,[24] complications of radial arterial punctures are extremely low and in our experience none were observed in the present experiment.

In conclusion, we showed a good correlation in pH, PCO2, BE, and HCO3 between simultaneous samples of arterialized earlobe and arterial blood in normotensive and normothermic patients receiving mechanical ventilation. So, earlobe blood gas measurements may be useful alternatives to arterial blood gas samples for critically ill patients who do not require regular continuous blood pressure measurements and close monitoring of arterial PO2 measurements. We do not recommend arterialized earlobe blood samples for determining PO2 of arterial blood gas samples.

Footnotes

Source of Support: Nil

Conflict of Interest: None declared.

References

  • 1.Barker WJ. Arterial puncture and cannulation. In: Roberts JR, Hedges JR, editors. Clinical procedures in emergency medicine. 3rd ed. Philadelphia: WB Saunders; 1998. pp. 308–22. [Google Scholar]
  • 2.Kelly AM, McAlpine R, Kyle E. Venous pH can safely replace arterial pH in the initial evaluation of patients in the emergency department. Emerg Med J. 2001;18:340–2. doi: 10.1136/emj.18.5.340. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Langlands JH, Wallace WF. Small blood samples from earlobe puncture: A substitute for arterial puncture. Lancet. 1965;2:315–7. doi: 10.1016/s0140-6736(65)90286-2. [DOI] [PubMed] [Google Scholar]
  • 4.Fajac I, Texereau J, Rivoal V, Dessanges JF, Dinh-Xuan AT, Dall'Ava-Santucci J. Blood gas measurement during exercise: A comparative study between arterialized earlobe sampling and direct arterial puncture in adults. Eur Respir J. 1998;11:712–5. [PubMed] [Google Scholar]
  • 5.Pitkin AD, Roberts CM, Wedzicha JA. Arterialised earlobe blood gas analysis: An underused technique. Thorax. 1994;49:364–6. doi: 10.1136/thx.49.4.364. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Christoforides C, Miller JM. Clinical use and limitations of arterialized capillary blood for PO2 determination. Am Rev Respir Dis. 1968;98:653–7. doi: 10.1164/arrd.1968.98.4.653. [DOI] [PubMed] [Google Scholar]
  • 7.Spiro SG, Dowdeswell IR. Arterialized earlobe blood samples for blood gas tensions. Br J Dis Chest. 1976;70:263–8. doi: 10.1016/0007-0971(76)90042-5. [DOI] [PubMed] [Google Scholar]
  • 8.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10. [PubMed] [Google Scholar]
  • 9.Sauty A, Uldry C, Debétaz LF, Leuenberger P, Fitting JW. Differences in PO2 and PCO2 between arterial and arterialized earlobe samples. Eur Respir J. 1996;9:186–9. doi: 10.1183/09031936.96.09020186. [DOI] [PubMed] [Google Scholar]
  • 10.Dall'Ava-Santucci J, Dessanges JF, Dinh-Xuan AT, Lockhart A. Is arterialized earlobe blood PO2 an acceptable substitute for arterial blood PO2? Eur Respir J. 1996;9:1329–30. doi: 10.1183/09031936.96.09061329. [DOI] [PubMed] [Google Scholar]
  • 11.Escalante-Kanashiro R, Tantalean-Da-Fieno J. Capillary blood gases in a pediatric intensive care unit. Crit Care Med. 2000;28:224–6. doi: 10.1097/00003246-200001000-00037. [DOI] [PubMed] [Google Scholar]
  • 12.Gennis PR, Skovron ML, Aronson ST, Gallagher EJ. The usefulness of peripheral venous blood in estimating acid-base status in acutely ill patients. Ann Emerg Med. 1985;14:845–9. doi: 10.1016/s0196-0644(85)80631-4. [DOI] [PubMed] [Google Scholar]
  • 13.Harrison AM, Lynch JM, Dean JM, Witte MK. Comparison of simultaneously obtained arterial and capillary blood gases in pediatric intensive care unit patients. Crit Care Med. 1997;25:1904–8. doi: 10.1097/00003246-199711000-00032. [DOI] [PubMed] [Google Scholar]
  • 14.Saavedra JM, Harris JG, Li S, Finberg L. Capillary refilling (skin turgor) in the assessment of dehydration. Am J Dis Child. 1991;145:296–8. doi: 10.1001/archpedi.1991.02160030064022. [DOI] [PubMed] [Google Scholar]
  • 15.McGillivray D, Ducharme FM, Charron Y, Mattimoe C, Treherne S. Clinical decisionmaking based on venous versus capillary blood gas values in the well-perfused child. Ann Emerg Med. 1999;34:58–63. doi: 10.1016/s0196-0644(99)70272-6. [DOI] [PubMed] [Google Scholar]
  • 16.Adrogue HJ, Rashad MN, Gorin AB, Yacoub J, Madias NE. Assessing acid-base status in circulatory failure: Differences between arterial and venous blood. N Engl J Med. 1989;320:1312–6. doi: 10.1056/NEJM198905183202004. [DOI] [PubMed] [Google Scholar]
  • 17.Koch G, Wendel H. Comparison of pH, carbon dioxide tension, standard bicarbonate and oxygen tension in capillary blood and in arterial blood during the neonatal period. Acta Pediatr Scand. 1967;56:10–6. [Google Scholar]
  • 18.MacIntyre J, Norman JN, Smith G. Use of capillary blood in measurement of arterial PO2. Br Med J. 1968;3:640–3. doi: 10.1136/bmj.3.5619.640. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Godfrey S, Wozniak ER, Courtenay Evans RJ, Samuels CS. Earlobe blood samples for blood gas analysis at rest and during exercise. Br J Dis Chest. 1971;65:58–64. [PubMed] [Google Scholar]
  • 20.Sadove MS, Thomason RD, Jobgen E. Capillary versus arterial blood gases. Anesth Analg. 1973;52:724–7. [PubMed] [Google Scholar]
  • 21.Gaultier C, Koch G, Helms P. Measurement of blood gases. Eur Respir J. 1989;2:160s–3s. [PubMed] [Google Scholar]
  • 22.Hofford JM, Dowling AS. Value of capillary blood gas analysis in the management of acute respiratory distress. Am Rev Respir Dis. 1976;113:707–8. doi: 10.1164/arrd.1976.113.5.707a. [DOI] [PubMed] [Google Scholar]
  • 23.Dar K, Williams T, Aitken R, Woods KL, Fletcher S. Arterial versus capillary sampling for analysing blood gas pressures. BMJ. 1995;309:24–5. doi: 10.1136/bmj.310.6971.24. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Mortensen JD. Clinical sequelae from arterial needle puncture, cannulation and incision. Circulation. 1967;35:1118–23. doi: 10.1161/01.cir.35.6.1118. [DOI] [PubMed] [Google Scholar]