Abstract
Background
Visual assessment of left ventricular ejection fraction (LVEF) is often used in clinical routine despite general recommendations to use quantitative biplane Simpsons (BPS) measurements. Even thou quantitative methods are well validated and from many reasons preferable, the feasibility of visual assessment (eyeballing) is superior. There is to date only sparse data comparing visual EF assessment in comparison to quantitative methods available. The aim of this study was to compare visual EF assessment by two-dimensional echocardiography (2DE) and triplane echocardiography (TPE) using quantitative real-time three-dimensional echocardiography (RT3DE) as the reference method.
Methods
Thirty patients were enrolled in the study. Eyeballing EF was assessed using apical 4-and 2 chamber views and TP mode by two experienced readers blinded to all clinical data. The measurements were compared to quantitative RT3DE.
Results
There were an excellent correlation between eyeballing EF by 2D and TP vs 3DE (r = 0.91 and 0.95 respectively) without any significant bias (-0.5 ± 3.7% and -0.2 ± 2.9% respectively). Intraobserver variability was 3.8% for eyeballing 2DE, 3.2% for eyeballing TP and 2.3% for quantitative 3D-EF. Interobserver variability was 7.5% for eyeballing 2D and 8.4% for eyeballing TP.
Conclusion
Visual estimation of LVEF both using 2D and TP by an experienced reader correlates well with quantitative EF determined by RT3DE. There is an apparent trend towards a smaller variability using TP in comparison to 2D, this was however not statistically significant.
Background
Echocardiography is a noninvasive imaging technique that provides immediate assessment of global and regional LV function. LVEF is the most commonly used parameter of LV systolic function. The assessment of LV function is an important clinical tool and is of great clinical importance in the evaluation and prognosis of patients with coronary artery disease, hypertension, diabetes, valvular disease, and congestive heart failure [1]. The reliable assessment of LV systolic function is important for therapeutic decision making in heart failure [2].
EF can be assessed using a variety of modalities including contrast ventriculography [3], magnetic resonance imaging (MRI) [4] and echocardiography [5]. By far echocardiography is the least expensive, most widely available technique and it is noninvasive, can be performed bedside, is rapid and very safe. LVEF is a measure of LV systolic function that has been shown to predict mortality in a number of conditions [6-8].
However, 2DE has a number of well-known limitations, the two most important perhaps being LV foreshortening and that different projection are not possible to acquire during the same cardiac cycle. The latter is particularly important in patients with variable heart rhythm (i.e. atrial fibrillation). Both these limitations can be overcome using a 3D-array transducer for acquiring the apical four-two-and three chamber views simultaneously in the TP mode.
RT3DE has shown excellent agreement to MRI in measuring both LV volumes and EF and has also proven to be superior to quantitative 2DE regarding EF [9-11]. These data are suggesting that RT3DE is interchangeable with MRI in measuring LVEF.
Quantitative 2D-EF using BPS rule can be somewhat time consuming, and the endocardial border tracing is sometimes difficult to perform, especially in patients with poor image quality, since it is performed on still frames. In clinical practice eyeballing for the evaluation of LVEF can be done faster and is often easier to perform, especially in patients with poor image quality.
Previous studies have demonstrated the value of eyeballing EF [12-16] and one study revealed good correlation with quantitative 2DE measurements [12]. However, the studies comparing to radionuclide methods were performed before the introduction of second harmonics and although eyeballing EF measurement have proven to be robust when performed careful, it is still considered highly subjective and questioned in both clinical and research settings.
The aim of this study was to investigate the correlation and variability of visually assess EF by 2DE and TPE in comparison to quantitative 3DE-EF measurements.
Methods
Study population
We included thirty patients who were referred to the echocardiography laboratory at the Department of Cardiology, Karolinska University Hospital, Huddinge. The mean age of the study population was 39 years (range 26 – 66). Table 1 shows baseline characteristics of the included patients.
Table 1.
Men | Women | |
Number (n) | 23 | 7 |
Age (mean ± SD, years) | 40 ± 11 | 36 ± 16 |
BSA (m2) | 2 ± 0.1 | 1.7 ± 0.1 |
LV-EF* | 55 ± 7 | 60 ± 4 |
Sinus rhythm | 23 | 7 |
HR | 66 ± 9 | 70 ± 13 |
Significant valvular disease | 1 | 1 |
DCM | 1 | 0 |
Diabetes | 2 | 0 |
Previous AMI | 1 | 0 |
Hypertonic | 1 | 1 |
CABG | 1 | 0 |
PM | 1 | 0 |
ICD | 1 | 0 |
Heart failure | 2 | 0 |
HCM | 1 | 0 |
* By 3DE, HR: heart rate, DCM: dilated cardiomyopaty, AMI: acute myocardial infarction, CABG: coronary artery bypass grafting, PM: Pacemaker, ICD: implantable cardiovertor defibrillator, HCM: hypertrophic cardiomyopaty.
Echocardiography
Echocardiographic studies were performed using Vivid 7 Dimension ultrasound equipment, (GE Vingmed Ultrasound, Horten, Norway) by a single experienced operator with patients in a left lateral recumbent position.
Conventional 2DE recordings of standard apical LV 4-and 2 chamber views were acquired using a M4S transducer. Simultaneous TP and RT3DE images were collected using a 3D-array transducer (3V).
When collecting the TP images of the LV, the apical 4-chamber view served as the reference image, and the two other planes were by default displayed with inter-plane angles set to 60 degrees. A figure showing the interposition of imaged planes was simultaneously displayed in a quad screen (figure 1A).
The RT3DE full volume (FV) images (figure 1B) were obtained from an apical window gathered over 4 cardiac cycles during end expiration breath-hold.
The saved 2D, TP and FV cine loops were transferred for a later offline analyze using commercially available software (EchoPAC, GE Vingmed Ultrasound, version 108.0.1).
For the visual 2D-EF evaluation, the 4-and 2 chamber view were displayed simultaneously, and accordingly TP images at different time points in the apical four-two-and three chamber views respectively. Two experienced echocardiographer estimated the 2D and TP LVEF blinded to all clinical data and previous reading. All visual LVEF evaluations were made twice one week apart. The quantitative LVEF evaluations of the RT3D-FV images were made by one reader twice, with one week interval, blinded to the first measurements.
Analyses of the RT3DE data sets were performed using a commercially available semi-automated analysis tool, 4D auto LV volume quantification (4DLVQ, EchoPAC version 108.0.1, GE Vingmed Ultrasound, Horten, Norway). In end-diastole (ED) as well as end-systole (ES) a total of eighteen LV identification landmarks were made: two basal marks at the mitral annulus and one apical mark in the 4-chamber-, 2-chamber- and 3-chamber views respectively (figure 2). Respecting these landmarks, the software automatically delineated the endocardial border in a 3D-model from ED and ES phases. The LVEF is calculated by the software (ED-volume – ES-volume)/ED-volume × 100%. In cases where the automatic delineation of the endocardial border was considered suboptimal the borders could be adjusted manually, although no adjustments were made in this study.
Statistics
All data are expressed as mean ± SD. Linear regression was performed for correlation analysis. Bland and Altman analysis was used for assessment of agreement between the different methods [17]. In order to determine the intraobserver variability all measurements were analyzed twice with one week apart. The inter- and intraobserver variability was measured according to the following formula: (SDdiff × 100%)/total mean × √2, where SDdiff is the SD of difference between measurements. The significance level was set as P < 0.05. Group comparisons of continuous variables were made using analysis of variance (ANOVA). Statistical analysis was performed using standard statistical software (SPSS version 16.0, Inc, Chicago, IL).
Results
All EF measurements, both 2D and 3D were feasible in all patients. EF was 54.7 ± 8.9% by 3DE, 55 ± 8% by 2D eyeballing and 55 ± 9% by TP eyeballing. Statistical analysis showed no significant differences in EF between the 2DE methods and 3DE.
The mean values and mean differences of EF by the different methods are presented in table 2.
Table 2.
Method | mean ± SD (%) | mean difference (%) | ||
FV3D-EF | 54.7 ± 8.9 | -0.1 ± 1.8 | ||
Eyeballing 2DE | 55 ± 8 | 1.3 ± 2.9 | ||
Eyeballing TP | 55 ± 9 | 1.2 ± 2.5 |
Regression analysis
There were excellent correlations between EF measured by 3D and eyeballing 2DE and TP (r = 0.91 and r = 0.95 respectively). Linear regression between 3D vs eyeballing 2DE and between 3D and eyeballing TP is shown in figure 3.
Analysis of agreement
The limits of agreement analysis of EF measured by the different methods are shown in table 3. There was a mean difference of -0.5 ± 3.7% for EF determined by 3D and eyeballing 2D and -0.2 ± 2.9% between 3D and eyeballing TP.
Table 3.
Mean difference ± SD | Limit of agreement | Correlation (r) | SEE | |
3D-EF minus Eyeballing BP | -0.5 ± 3.7 | -7.9 to 6.9 | 0.91 | 3.71 |
3D-EF minus Eyeballing TP | -0.2 ± 2.9 | -6 to 5.6 | 0.95 | 2.90 |
Eyeballing BP minus Eyeballing TP | 0.3 ± 3.5 | -6.7 to 7.3 | 0.92 | 3.31 |
SEE: standard error of estimate.
Reproducibility of EF
Intraobserver variability was 3.8% for eyeballing 2D, 3.2% for eyeballing TP and 2.3% for quantitative 3D-EF. Interobserver variability was 7.5% for eyeballing 2D and 8.4% for eyeballing TP.
Discussion
An optimal method for determining LVEF by echocardiography should be rapid, reliable, and widely applicable in order to utilized routinely in a clinical laboratory. Off-line echocardiographic analysis techniques using software's and operator-performed border tracing meet these criteria poorly, explaining their lack of widespread use despite validation in different laboratories [14,15]. The major advantage of visual estimation of LVEF is that the reader can integrate all information regarding wall motion, atrioventricular plane displacement (AV-plane), etc. Its principal limitations are dependency on the skill of the reader and the inter – and intraobserver variability.
In the present study, eyeballing EF both using 2D and TP mode showed excellent correlation with quantitatively measured EF using RT3DE (r = 0.91 and r = 0.95 respectively) and without any significant bias (-0.5 ± 3.7% and -0.2 ± 2.9% respectively). Intraobserver variability was 3.8% for eyeballing 2D, 3.2% for eyeballing TP and 2.3% for quantitative 3D-EF. Interobserver variability was 7.5% for eyeballing 2D and 8.4% for eyeballing TP.
Lavine et al. showed in concordance to our study excellent correlation (r = 0.90) between 2D eyeballing in comparison to radioventriculography (RVG) [18]. In their study they compared eyeballing EF and wall motion scoring (WMS), showing that WMS had better correlation (r = 0.97) in comparison to RVG. This r-value is almost identical to our TP eyeballing determination, where eyeballing has the advantages of being less time-consuming than WMS analysis.
Two advantages of using TPE are that when EF is determined it is based from same cardiac cycle, and that TP can overcome the problem of LV foreshortening. Another advantage is that the acquisition time is reduced because the transducer is not moved to obtain data from multiple views. Although, this advantages didn't transfer into significant differences between 2D and TP-EF determination in our study. There is an apparent trend towards a smaller variability using TP in comparison to 2D. The variability in this study is very small and is likely to be more pronounced in clinical routine. Thus there might be significant advantage in using TP over 2D for EF assessment in everyday clinical routine.
It might be suggested that the accuracy of eyeballing EF is dependent of the skill of the reader and that visual assessment of EF is accurate only in very experienced hands. However, our results are in concordance with those shown by the other groups, indicating that eyeballing EF commonly can be used with a high level of accuracy [12-16]. Indeed Jensen-Urstad et al. showed in their study that Simpson's rule correlated worst (r = 0.45–0.51) when comparing EF by different methods (eyeballing, WMS and AV-plane) in comparison to radionuclide imaging during acute myocardial infarction [13]. Furthermore, it has been shown in a study by Akinboboye et al [19] that eyeballing EF is a method easy to learn. If a person with no previous experience of echocardiography, after each evaluation is given instant feedback from a experienced reader, about 60 cases in joint reading is required to achieve the same accuracy as an experienced echocardiographer.
Even though quantitative measurement of LVEF is generally recommended [20], visual assessments of EF is commonly used in everyday clinical practice due to both suboptimal image quality and lack of time.
The results of the present and prior studies suggest that eyeballing EF could be accepted for clinical use provided that the variability of the eyeballing EF in the echocardiography laboratory is low. Therefore, the variability of using visual estimation of LVEF should be regularly tested in any echocardiography laboratory using this as the recommended method.
Limitations
The variability in the methods in our study is very low which could limit the generazibility of our data, since a larger variability can be expected in the clinical routine. However we believe that in most laboratories, after adequate training this method could provide reliable estimates of LVEF.
The mean age of the study population was relatively low, and most having normal LVEF. This could partly explain the excellent correlation due to the excepted good image quality. Our findings will therefore have to be confirmed in older populations and also in patients with systolic heart failure.
Conclusion
Visual estimation of LVEF both using 2D and TP by an experienced reader correlates well with quantitative EF determined by RT3DE. There is an apparent trend towards a smaller variability using TP in comparison to 2D, this was however not statistically significant.
Competing interests
The authors declare that they have no competing interests.
Authors' contributions
KS and PG initiated the study. RW and LÅB supervised the study and participated in the interpretation of the results and manuscript preparation. KS and PG performed measurements, made all data conversions, plots and calculations from ultrasound data, and participated in the preparation of the manuscript. KS and AM performed statistical analysis and participated in the interpretation of the results. All authors read and approved the final manuscript.
Contributor Information
Kambiz Shahgaldi, Email: kambiz.shahgaldi@karolinska.se.
Petri Gudmundsson, Email: petri.gudmundsson@mah.se.
Aristomenis Manouras, Email: aristomenis.manouras@karolinska.se.
Lars-Åke Brodin, Email: lars-ake.brodin@syd.kth.se.
Reidar Winter, Email: reidar.winter@karolinska.se.
References
- Hammermeister KE, DeRouen TA, Dodge HT. Variables predictive of survival in patients coronary disease. Cisculation. 1979;59:421–30. doi: 10.1161/01.cir.59.3.421. [DOI] [PubMed] [Google Scholar]
- Cowburn PJ, Cleland JGF, Coats AJS. Risk stratification in chronic heart failure. Eur Heart J. 1998;19:696–710. doi: 10.1053/euhj.1997.0820. [DOI] [PubMed] [Google Scholar]
- Rumberger JA, Behrenbeck T, Bell MR, Breen JF, Johnston DL, Holmes DR, Jr, Enriquez-Sarano M. Determination of ventricular ejection fraction: a comparison of available imaging methods. The Cardiovascular Imaging Working Group. Mayo Clin Proc. 1997;72:860–70. doi: 10.4065/72.9.860. [DOI] [PubMed] [Google Scholar]
- Stratemeier EJ, Thompson R, Brady TJ, Miller SW, Saini S, Wismer GL, Okada RD, Dinsmore RE. Ejection fraction determination by MR imaging: comparison with left ventricular angiography. Radiology. 1986;158:775–777. doi: 10.1148/radiology.158.3.3945753. [DOI] [PubMed] [Google Scholar]
- Erbel R, Schweizer P, Lambertz H, Henn G, Meyer J, Krebs W, Efferts S. Echoventriculography – a simultaneous analysis of two-dimensional echocardiography and cineventriculography. Circulation. 1983;67:205–215. doi: 10.1161/01.cir.67.1.205. [DOI] [PubMed] [Google Scholar]
- Ahnve S, Gilpin E, Henning H. Limitations and advantages of the ejection fraction for defining high risk after acute myocardial infarction. Am J Cardiol. 1986;58:872–8. doi: 10.1016/S0002-9149(86)80002-9. [DOI] [PubMed] [Google Scholar]
- Madsen BK, Hansen JF, Stokholm KH. Chronic cengestive heart failure: description and survival of 190 censecutive patients with a diagnosis of chronic congestive heart failure based on clinical signs and symptoms. Eur Heart J. 1994;15:303–10. doi: 10.1093/oxfordjournals.eurheartj.a060495. [DOI] [PubMed] [Google Scholar]
- Parameshwar J, Keegan J, Sparrow J. Predictors of prognosis in severe chronic heart failure. Am Heart J. 1992;70:359–63. doi: 10.1016/0002-8703(92)90656-g. [DOI] [PubMed] [Google Scholar]
- Mannaerts HF, Heides JA Van Der, Kamp O, Papavassiliu T, Marcus JT, Beek A, Van Rossum AC, Twisk J, Visser CA. Quantification of left ventricular volumes and ejection fraction using freehand transthoracic three-dimensional echocardiography: comparison with magnetic resonance imaging. J Am Soc Echocardiogr. 2003;16:101–9. doi: 10.1067/mje.2003.7. [DOI] [PubMed] [Google Scholar]
- Kuhl HP, Schreckenberg M, Rulands D, Katoh M, Schäfer W, Schummers G, Bucker A, Hanrath P, Franke A. High-resolution transthoracic real-time three-dimensional echocardiography: quantitation of cardiac volumes and function using semi-automatic border detection and comparison with cardiac manetic resonance imaging. J Am Coll Cardiol. 2004;43:2083–90. doi: 10.1016/j.jacc.2004.01.037. [DOI] [PubMed] [Google Scholar]
- Jenkins C, Bricknell K, Hanekom L, Marwick TH. Reproducibility and accuracy of echocardiographic measurements of left ventricular parameters using real-time three-dimensional echocardiography. J Am Coll Cardiol. 2004;44:878–86. doi: 10.1016/j.jacc.2004.05.050. [DOI] [PubMed] [Google Scholar]
- Gudmundsson P, Rydberg E, Winter R, Willenheimer R. Visually estimated left ventricular ejection fraction by echocardiography is closely correlated with formal quantitative methods. Int J Cardiol. 2005;101:209–212. doi: 10.1016/j.ijcard.2004.03.027. [DOI] [PubMed] [Google Scholar]
- Jensenn-Urstad K, Bouvier F, Höjer J, Ruiz H, Hulting J, Samad B, Thorstrand C, Jensen-Urstad M. Comparison of different echocardiographic methods with radionuclide imaging for measuring left ventricular ejection fraction during acute myocardial infarction-treated by thrombolytic theraphy. Am J Cardiol. 1998;81:538–44. doi: 10.1016/S0002-9149(97)00964-8. [DOI] [PubMed] [Google Scholar]
- Mueller X, Stauffer JC, Jaussi a, Goy JJ, Kappenberger L. Subjective visual echocardiographic estimate of left ventricular ejection fraction as an alternative to conventional echocardiographic methods: comparison with contrast angiography. Clin Cardiol. 1991;14:898–902. doi: 10.1002/clc.4960141108. [DOI] [PubMed] [Google Scholar]
- Amico AF, Lichtenberg GS, Reisner SA, Stone CK, Schwartz RG, Meltzer RS. Superiority of visual versus computerized echocardiographic estimation of radionuclide left ventricular ejection fraction. Am Heart J. 1989;1181:259–65. doi: 10.1016/0002-8703(89)90018-5. [DOI] [PubMed] [Google Scholar]
- van't Hof AW, Schipper CW, Gerritsen JG, Reiffers S, Hoomtje JC. Comparison of radionuclide angiography with three echocardiographic parameters of left ventricular function in patients after myocardial infarction. Int J Card Imaging. 1998;14:413–8. doi: 10.1023/A:1006082214191. [DOI] [PubMed] [Google Scholar]
- Bland JM, Altman DG. Statistical methods for assessing agreement between two methods for clinical measurements. 1986;1:307–10. [PubMed] [Google Scholar]
- Lavine SJ, Salacata A. Visual quantitative estimation: Semiquantitative wall motion scoring and determination of ejection fraction. Echocardiogr. 2003;20:401–410. doi: 10.1046/j.1540-8175.2003.03079.x. [DOI] [PubMed] [Google Scholar]
- Akinboboye O, Sumner J, Gopal A, King D, Shen Z, Bardfeld P, Blanz L, Brown EJ., Jr Visual estimation of ejection fraction by two-dimensional echocardiography: the learning curve. Clin Cardiol. 1995;18:726–9. doi: 10.1002/clc.4960181208. [DOI] [PubMed] [Google Scholar]
- Lang RM, Michelle Bierig, Devereux RB, Flachskampf FA, Elyse Patricia, Pellikka PA, Picard MH, Roman MJ, Seward J, Shanewise J, Solomon S, Spencer KT, John Sutton MSt, Stewart W. Recommendations for chamber quantification. Eur J Echocardiogr. 2006;7:79–108. doi: 10.1016/j.euje.2005.12.014. [DOI] [PubMed] [Google Scholar]