Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 1975 Mar;1(3):302–308. doi: 10.1128/jcm.1.3.302-308.1975

Laboratory experience with a radiometric method for detecting bacteremia.

W A Thiemke, K Wicher
PMCID: PMC275067  PMID: 1100660

Abstract

Two bacteriologic systems for detecting bacteria in blood were compared; the automated radiometric BACTEC and the conventional method used in our laboratory for many years. BACTEC consisted of two bottles with 30 ml and the conventional method with 50 ml of media for aerobes and anaerobes. The BACTEC bottles were inoculated with 2 to 3 ml and the conventional with 4 to 5 ml of blood at the patient's bedside. Out of the 3,045 blood specimens cultured (804 patients), 262 (117 patients) were positive by one or both methods. The conventional system detected 5more cultures. The explanation of the differences is discussed. Positive blood cultures were detected by the BACTEC procedure as early as 6 h after the blood collection. In the first 24 h, on the average, 77% of aerobic organisms were detected by the BACTEC as compared to 48% by the conventional system. All anaerobic BACTEC cultures were positive within 4 days, whereas the conventional system detected at that time 74%. At day 4, 67% of fungi were detected by the BACTEC and only 27% by the conventional system. Of the 3,045 blood cultures examined by the BACTEC, 208 were recorded as false positive with growth index readings ranging from 30 to 59.

Full text

PDF
302

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Bannatyne R. M., Harnett N. Radiometric detection of bacteremia in neonates. Appl Microbiol. 1974 Jun;27(6):1067–1069. doi: 10.1128/am.27.6.1067-1069.1974. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Blazevic D. J., Stemper J. E., Matsen J. M. Comparison of macroscopic examination, routine gram stains, and routine subcultures in the initial detection of positive blood cultures. Appl Microbiol. 1974 Mar;27(3):537–539. doi: 10.1128/am.27.3.537-539.1974. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Brooks K., Sodeman T. Rapid detection of bacteremia by a radiometric system. A clinical evaluation. Am J Clin Pathol. 1974 Jun;61(6):859–866. doi: 10.1093/ajcp/61.6.859. [DOI] [PubMed] [Google Scholar]
  4. COCHRAN W. G. The comparison of percentages in matched samples. Biometrika. 1950 Dec;37(3-4):256–266. [PubMed] [Google Scholar]
  5. Caslow M., Ellner P. D., Kiehn T. E. Comparison of the BACTEC system with blind subculture for the detection of bacteremia. Appl Microbiol. 1974 Sep;28(3):435–438. doi: 10.1128/am.28.3.435-438.1974. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. DeBlanc H. J., Jr, DeLand F., Wagner H. N., Jr Automated radiometric detection of bacteria in 2,967 blood cultures. Appl Microbiol. 1971 Nov;22(5):846–849. doi: 10.1128/am.22.5.846-849.1971. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Finegold S. M., White M. L., Ziment I., Winn W. R. Rapid diagnosis of bacteremia. Appl Microbiol. 1969 Sep;18(3):458–463. doi: 10.1128/am.18.3.458-463.1969. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Hall M., Warren E., Washington J. A., 2nd Comparison of two liquid blood culture media containing sodium polyanethole sulfonate: tryptic soy and Columbia. Appl Microbiol. 1974 Apr;27(4):699–702. doi: 10.1128/am.27.4.699-702.1974. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Larson S. M., Charache P., Chen M., Wagner H. N., Jr Automated detection of Haemophilus influenzae. Appl Microbiol. 1973 Jun;25(6):1011–1012. doi: 10.1128/am.25.6.1011-1012.1973. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Renner E. D., Gatheridge L. A., Washington J. A., 2nd Evaluation of radiometric system for detecting bacteremia. Appl Microbiol. 1973 Sep;26(3):368–372. doi: 10.1128/am.26.3.368-372.1973. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Rosner R. Comparison of macroscopic, microscopic, and radiometric examinations of clinical blood cultures in hypertonic media. Appl Microbiol. 1974 Oct;28(4):644–646. doi: 10.1128/am.28.4.644-646.1974. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Washington J. A., 2nd, Yu P. K. Radiometric method for detection of bacteremia. Appl Microbiol. 1971 Jul;22(1):100–101. doi: 10.1128/am.22.1.100-101.1971. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Zwarun A. A. Effect of osmotic stabilizers on 14 CO 2 production by bacteria and blood. Appl Microbiol. 1973 Apr;25(4):589–591. doi: 10.1128/am.25.4.589-591.1973. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)