Abstract
We show a new ultra-sensitive laser based analytical technique, intracavity optogalvanic spectroscopy, allowing extremely high sensitivity for detection of 14C labeled carbon dioxide. Capable of replacing large accelerator mass spectrometers, the technique quantifies attomoles of 14C in submicrogram samples. Based on the specificity of narrow laser resonances coupled with the sensitivity provided by standing waves in an optical cavity and detection via impedance variations, limits of detection near 10-15 14C:12C ratios are obtained. Using a 15W 14CO2 laser, a linear calibration with samples from 10-15 to >2 × 10-12 in 14C:12C ratios, as determined by accelerator mass spectrometry, is demonstrated. Possible applications include microdosing studies in drug development, individualized sub-therapeutic tests of drug metabolism, carbon dating and real time monitoring of atmospheric radiocarbon. The method can also be applied to detection of other trace entities.
Carbon 14 (radiocarbon) is an ideal organic tracer having an extremely low natural abundance in living systems, near 1 ppt, and a long half life, 5730 years, ideal for clinical and laboratory tracer experiments. Until recently almost all quantitation of 14C content has been by scintillation detection of the low energy beta particle emitted in its decay. At present there is great interest in the drug development community for pharmacokinetic information on new drug entities using non therapeutic microdoses of labeled drugs, which require much higher analysis sensitivity than is possible with scintillation counting1. Accelerator Mass Spectroscopy (AMS), first developed to extend carbon dating to smaller and older samples, has become the standard method for 14C atom counting in such sensitive bioanalytical tracer studies2, 3. Limitations to wide use of AMS for bioanalytical studies include size, cost and complexity of the analysis system as well as the fact that samples must contain of the order of 0.5 mg total carbon. The laser assisted ratio analyzer (LARA) technique4 is based on the existence of large isotope shifts in molecular spectra, the use of fixed frequency isotopic lasers and sensitive detection via the laser optogalvanic effect (OGE). Research and development during the past decade has demonstrated that LARA’s sensitivity and versatility make new classes of isotope ratio measurement systems possible5,6.
The basic concept of OGE detection has long been used in atomic and molecular spectroscopy7. It is based on the electrical response of a gas discharge to an optical perturbation. If a laser of intensity I, and frequency ν is incident on a cylindrical (length L and radius R) weak electrical discharge, the electrical response, S, of the discharge can be expressed by an integral over the laser-discharge interaction volume:
(1) |
The density of interacting species is n and σ(ν) refers to the laser-species interaction cross section. S can be any discharge parameter related to conductivity and K is a corresponding optogalvanic proportionality constant that depends on the details of the electrical discharge. To a good approximation, on resonance, for cases where light absorption is small and K is independent of n, the average electrical response simplifies to a product:
(2) |
Here, n represents the average molecular density of interacting particles, L, the length of the interaction region, I the average laser intensity and A the average area of the laser beam. Physically, what happens is that the light changes the equilibrium distribution of species, including excited species in the discharge. This affects collision rates, including those between molecules and electrons, leading to a measurable conductivity change, S, of the ionized gas. Optogalvanic spectroscopy for isotopic analysis makes use of the fact that laser resonances in CO2 are isotope dependent8.
In practice, a periodically switched cw laser in resonance with a specific molecular transition changes the population of excited molecular states, which, in turn, changes the electron density and/or temperature in the discharge giving rise to an impedance change varying with the laser intensity. There exist several significant advantages of OGE measurements compared to ion beam and other spectrometric stable isotope measurement technologies:
The measured parameter is electrical and periodic so that filtering and averaging techniques can be used to achieve extremely low noise and background.
No optical measurement is required, thus eliminating the need for collection and dispersion optics and light transducers.
The measured signal is proportional to the product of analyte density, path length and laser power. In contrast with laser absorption techniques or ion counting, the laser power provides a gain, or sensitivity enhancement per molecule, so that signals for dilute isotopes can be amplified relative to the majority species.
The measurement system can be varied, within limits, in order to enhance signal to background ratios by controlling gas mixtures (ranging from 100% to 1% or less CO2, balance N2, CO2 free air or other inert buffer) pressure (typically 1 to 5 mbar) and electrical discharge power (typically 2 to 5 Watts).
Previous success using LARA for breath testing9,10,11,12,13,14,15 and environmental monitoring16 has led to the development of the technique for the detection of 14C reported here. When the laser operates at a particular transition in 14CO2, any resonant change in the impedance of the discharge system is due to the presence of 14CO2 in the discharge. The laser transition chosen for 14CO2 must be well separated in wavelength from any transition in 12CO2 or 13CO2 (1% natural abundance) and from any transition in a CO2 molecule with 18O (.02% natural abundance) or 17O (.004% natural abundance). The discreteness of energy levels and the large isotope shifts in vibrational-rotational transitions allow a line specific 14CO2 laser to be built. The strongest lasing transitions observed in 14CO2 are at 11.8 microns and 11.3 microns, significantly longer in wavelength than lasing transitions of the other “stable” isotope CO2 lasers. Any 14CO2 molecule in the lower or upper laser level is automatically in resonance (absorption or stimulated emission) with the narrow band laser, providing the sharp specificity required for isotope ratio analysis. The nearest 13CO2 and 12CO2 lines are separated by more than 500 linewidths, leading to non resonant cross sections reduced by approximately 10 orders of magnitude. The nonresonant interaction is further reduced for 13CO2 and 12CO2 by rotational state Boltzmann population ratios for the two isotopes. In any case, this small background effect is pressure dependent and measurable and does not seriously limit the ultimate 14CO2 detection limit.
Extension of the LARA technique to 14CO2 was previously established with highly enriched (nano-Curie) 14C samples17. In those studies, an isotope ratio (14C/12C) sensitivity of order 10-9 was demonstrated with a very simple external cavity configuration. A 14CO2 laser at 11.8 microns with a maximum output power of 2 Watts was used. The measurement yielded signal-to-noise ratios of several hundred with short averaging times and a sensitivity limit of about 1 picomole 14CO2. Measurements with non-enriched carbon dioxide confirmed that no memory effect was present and that the system sensitivity was far too low for detection at Modern (1.180 × 10-12 14C/C) levels of 14CO2.
Enhancement of sensitivity in laser absorption experiments often involves increased absorption length in optical cavities-most notably cavity ringdown spectroscopy (CRDS)18 and laser intracavity absorption spectroscopy (ICAS)19. The new standing wave technique reported here is analogous to, but with key differences from, ICAS. The most significant difference from ICAS is that detection is via the OGE rather than optical. The analyte in this case is the same as the lasing species and an OGE exists for stimulated emission and/or absorption.
EXPERIMENTAL SECTION
The layout of the measurement system is shown in Figure 1. The sample cell is placed inside a laser cavity (LaserTech Group Inc. Model LTG250, 15 W output with sealed 14CO2 fill) thus subjecting it to the full saturated laser power. Also inside the laser cavity is a computer controlled shutter used for modulating the 14CO2 laser beam. The laser output beam can be used to monitor power and wavelength as well as for parallel external cell measurements if required. A small electronically modulated 12CO2 laser (Synrad Model 48-1, 5 W output) beam passes through the 14CO2 laser output coupler and the sample cell for single pass 12CO2 measurement. This 12CO2 signal is used for normalization. The quartz sample cell, 10 cm in length and 2.5 cm OD, with ZnSe Brewster angle windows is pumped with an oil free vacuum pump to low pressure. In these studies, a radio frequency (rf) glow discharge is utilized because of its stability and low inherent noise20. With sample pressure held constant-typically between 1 and 5 mbar, a low power (2 to 5 Watt) rf discharge is ignited and maintained via external copper electrodes. The oscillator power supply circuit also monitors the average rf voltage amplitude across the discharge for the OGE signal. The 12CO2 and 14CO2 laser beams simultaneously passing through the sample cell are chopped at different frequencies, and the resulting voltage variation across the cell is measured. The Fourier Transform values of the voltage signal at the two different frequencies yield the 12CO2 and 14CO2 OGE signals.
The sample consists of CO2 with a buffer (95 to 99%) of nitrogen or CO2 free air, or 100% CO2. Calibration samples consist of tanks of research grade CO2 (Air Products) with 14CO2 less than 10-12 (1 Modern) level as determined by AMS measurements at Lawrence Livermore National Laboratory (LLNL). Aliquots of CO2 with varying 14CO2 concentrations were also supplied by LLNL. Measurements could be run in a sealed cell (batch mode) or continuous flow mode with flow rates of 0.1 to 2 sccm.
Time averaged OGE signals were obtained using a National Instruments Data Acquisition System. The virtual instrument designed controlled the shutter, and monitored pressure, 12CO2 concentration and laser wavelength data. An OGE output from the 14CO2 laser tube was used to stabilize laser wavelength and for laser power normalization.
RESULTS and DISCUSSION
Figure 2 shows the OGE response to an enriched sample (9 M) of 22 micromoles CO2 containing 223 attomoles 14CO2. The figure is a 5 second coherent averaged waveform with the 14CO2 laser chopped at 63 Hz. Figure 3 shows a resonance curve obtained by tuning the laser through its gain profile. The optogalvanic effect line shape shows the Voigt profile expected for the discharge conditions. The non resonant background (and the peak intensity) varies with different laser lines, also as expected. Under all conditions, the observed signal is indicative of a greatly enhanced interaction length due to the intracavity standing wave. The maximum enhancement observed over the single pass system is of order 10 million which was initially surprising as in ICAS studies with a single mode laser gain enhancements only of order 100 are expected.
We believe that the observed mega-enhancement is due to the nature of the optogalvanic effect in the optical cavity. In equations 1 and 2 above I is the circulating internal laser power, around 50 W in our case, and L the interaction length is now greatly enhanced due to the standing wave. A simple estimate would have L enhanced only by the number of reflections of a single photon, where the major loss is due to the output coupler. However, for the cw laser system, loss in the OGE cell is balanced by the gain of the lasing medium keeping the internal circulating power constant. Now, the effective interaction length Leff is given by Leff=cΔt, where Δt is given by the time that the laser is on. For a modulation frequency of 63 Hz, Δt is 15.8 msec yielding an effective interaction length of almost 5000 km. This value is close to our observed maximum enhancement of ∼107 for the intracavity configuration compared to single pass. In a multimode ICAS experiment, Δt is given by the spectral saturation time and can be as long as several seconds21. In the ICAS experiment what is observed is a change in laser output. The effect on laser output of the few thousand 14CO2 molecules in the OGE analysis cell is negligible compared to the 1019 active 14CO2 molecules in the laser cavity. However, all photon interactions in the intracavity cell contribute to the OGE.
The sensitivity is so great that the natural abundance of 14C in ambient air can easily be detected. An estimate of the present limit of detection is of order 10-15 14C/12C ratio, similar to AMS. It is known that nitrogen is an ideal buffer gas for the OGE in CO2 for much the same reason that nitrogen is used in the CO2 laser. However, unlike the situation in a laser, where helium is also present to depopulate lower levels, helium is found to reduce the OGE22. Further, we have found that air is a better buffer gas than pure nitrogen, as it inhibits CO2 dissociation. The optogalvanic proportionality factor for 14CO2 in equations 1 and 2 is a complicated function of partial pressure for all gas constituents as well as discharge conditions,23 even changing sign as the system goes from absorption to gain conditions. Also, equations 1 and 2 assume thermal equilibrium, a situation only easily achieved in tens of ms for low 14CO2 concentration. These effects make calibration of the system over large changes in 14CO2 concentration difficult. For radiocarbon dating of small samples and drug development studies near modern enrichment a linear calibration region over more than an order of magnitude has been found using pure CO2 as the carrier gas, at approximately 1 mbar pressure with a Δt near 25msec. Figure 4 shows the calibration from 4.9 ×10-3 Modern to 1.5 Modern samples with the OGE signal plotted against AMS results. A 12 Modern sample showed saturation effects under the same measurement conditions. The slope of the calibration curve, 0.4 is in agreement with calculations due to the non-resonant signal from 13C and 12C discussed above. A detailed model of the intracavity OGE must include laser saturation effects as well as equilibration time constant effects on the OGE parameter K.
The system can run with continuous flow as shown in figure 5 for 1ml CO2 injected into a gas stream of CO2 free air as carrier. The buildup and loss of 12C and 14C is obvious. The saturation effect present when the concentration reaches around 1% is also obvious. Modeling of this effect using ICAS theory (References 18 and 20) and OGE theory 24 is in progress.
CONCLUSION
It is believed that the sensitivity shown in these and certain AMS experiments will allow more rapid drug development by permitting early safe metabolic testing with microdoses of test compounds25. Further, such microdosing studies can be used to customize doses for individual metabolic differences. Though both 14C LARA and AMS aim to quantify 14C in biological samples, there are several significant differences in the technologies. For both techniques, a carbon containing compound is first oxidized to CO2. The analyte in the LARA technique remains in the form of CO2, while AMS machines, at present, require an additional step to reduce the CO2 sample to elemental carbon. There are significant efforts, however, to enable Bio-AMS machines to run effectively with a gas ion source utilizing CO 262. Reasons for these efforts include improving efficiency via fewer processing steps, and higher throughput, as well as enabling smaller sample size. Ultimately it is desired to couple an input separation device such as a gas or liquid chromatograph with fractions oxidized to CO2 and continuously flowed through the system. .
The 12C and/or 13C normalization is straightforward with our system for samples of any enrichment, using 12CO2 and 13CO2 lasers as is now done with existing breath and environmental monitoring instruments. Furthermore, the small sample required and the possibility of continuous flow or batch processing should permit development of combined instrumentation with gas chromatographs or liquid chromatographs coupled to an oxidizer as a front end27, 28. The measurement sensitivity is so great that carbon dating using microgram samples can easily be envisioned.
Limits of precision and sensitivity are yet to be determined but should equal or exceed those of AMS. Use of an external calibration cell and double ratio measurements as is done to determine 13C variations at the part per 10 thousand6 level should also be possible with 14C. The engineering challenge will be in the handling of such small samples and eliminating the ubiquitous background 14CO2. We further note that the technique may be extended to other trace atom or compound analysis with tunable or fixed frequency lasers.
ACKNOWLEDGEMENT
This work has been supported by the US NIH (Grant 5R33RR018280), NSF (Grant DBI0456241) and Merck Research Laboratories as well as by the NIH Resource for Biomedical AMS at Lawrence Livermore National Laboratory who have supplied reference samples. Helpful discussions with John Vogel, Ted Ognibene and Martin Schaden are appreciated.
REFERENCES
- [1].Lappin Graham, Garner R. Colin. Nature Reviews, Drug Discovery. 2003;2:233–240. doi: 10.1038/nrd1037. [DOI] [PubMed] [Google Scholar]
- [2].Turtaltaub KW, Vogel JS. Curr Pharm Des. 2000;6:991–1007. doi: 10.2174/1381612003400047. [DOI] [PubMed] [Google Scholar]
- [3].Ognibene TJ, Bench G, Vogel JS, Peaslee GF, Murov S. Analytical Chemistry. 2003;75:2192–2196. doi: 10.1021/ac026334j. [DOI] [PubMed] [Google Scholar]
- [4].Murnick DE, Peer BJ. Science. 1994;263:945–94. doi: 10.1126/science.8310291. [DOI] [PubMed] [Google Scholar]
- [5].Murnick DE, Colgan MJ, Stonebach DJ. Laser Optogalvanic Isotope Ratio Analysis in Carbon Dioxide. In: Heys GJ, Melillo DG, editors. Synthesis and Applications of Isotopicaly Labeled Compounds 1997. John Wiley and Sons Ltd.; 1998. p. 111. [Google Scholar]
- [6].Murnick DE, Okil JO. Isotopes in Environmental and Health Studies. 2005;41(4):363–371. doi: 10.1080/10256010500384440. [DOI] [PubMed] [Google Scholar]
- [7].Barbieri B, Beverini N, Sasso A. Rev Mod Phys. 1990;62:603–644. [Google Scholar]
- [8].Freed C. Tunable Lasers Handbook. Academic Press; New York: 1995. pp. 63–165. [Google Scholar]
- [9].Braden B, Gelbmann C, Dietrich CF, Caspary WF, Scholmerich J, Lock G. European Journal of Gastroenterology & Hepatology. 2001;13(7):807–10. doi: 10.1097/00042737-200107000-00008. [DOI] [PubMed] [Google Scholar]
- [10].Parente F, Bianchi-Porro G. European Journal of Gastroenterology & Hepatology. 2001;13(7):803–6. doi: 10.1097/00042737-200107000-00007. [DOI] [PubMed] [Google Scholar]
- [11].Savarino V, Landi F, Dulbecco P, Ricci C, Tessieri L, Biagini R, Gatta L, Miglioli M, Celle G, Vaira D. Digestive Diseases & Sciences. 2000;45(11):2168–74. doi: 10.1023/a:1026605021484. [DOI] [PubMed] [Google Scholar]
- [12].Cave DR, Van Zanten V, Laine L. Alimentary Pharmacology & Therapeutics. 1999;13:747. doi: 10.1046/j.1365-2036.1999.00534.x. [DOI] [PubMed] [Google Scholar]
- [13].vanderHulst RW, Hensen EF, vanderEnde A, Kruizinga SP, Homan A, Tytgat GN. Nederlands Tijdschrift voor Geneeskunde. 1999;143:400–404. [PubMed] [Google Scholar]
- [14].Minoli G, Prada A, Schuman R, Murnick D, Rigas B. J. Clin. Gastroenterol; 1998;26:264–266. doi: 10.1097/00004836-199806000-00009. [DOI] [PubMed] [Google Scholar]
- [15].van der Hulst RWM. Gut. 1997;41(suppl 1):A72–73. [Google Scholar]
- [16].Okil JO. PhD thesis. Rutgers University; Newark NJ: 2004. [Google Scholar]
- [17].Murnick DE, Dogru, Ozgur, Ilkmen Nuclear Instruments and Methods in Physics Research. 2007;B 259:786–789. doi: 10.1016/j.nimb.2009.10.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [18].Busch Kenneth W., Busch Marianna A., editors. Cavity-ringdown spectroscopy: an ultratrace-absorption measurement technique; ACS symposium series; American Chemical Society; [New York]. 1999; p. 720. ISSN0097-6156. [Google Scholar]
- [19].Baev VM, Latz T, Toschek PE. App. Phys. B. 1999;69(3):171–202. [Google Scholar]
- [20].May RD, May PH. Rev. Sci. Instruments. 1986;57:2242–5. [Google Scholar]
- [21].Kimble HJ. IEEE J of Quantum Electronics. 1980;QE-16:455–461. [Google Scholar]
- [22].Chaohua Wang, Mingjiang Yang. Chinese Physics Letters. 1986;3:501–504. [Google Scholar]
- [23].Moffet S, Smith ALS. J.Phys D: Appl. Phys. 1984;17:59–70. [Google Scholar]
- [24].Smith ALS, Moffet S. J.Phys D: Appl. Phys. 1984;17:71–78. [Google Scholar]
- [25].Lappin, Graham, et al. Clinical Pharmacology & Therapeutics. 2006;80:203–215. doi: 10.1016/j.clpt.2006.05.008. [DOI] [PubMed] [Google Scholar]
- [26].Ramsey CB, et al. Radiocarbon. 2004;46(1):35–32. [Google Scholar]
- [27].Krummen, et al. Rapid Comm in Mass Spectrometry. 2004;18:2260–2266. doi: 10.1002/rcm.1620. [DOI] [PubMed] [Google Scholar]
- [28].Liberman Rosa G., Tannenbaum Steven R., Hughey Barbara J., Shefer Ruth E., Klinkowstein Robert E., Prakash Chandra, Harriman Shawn P., Skipper Paul L. Anal. Chem. 2004;76:328–334. doi: 10.1021/ac030181y. [DOI] [PubMed] [Google Scholar]