Table 2.

Inter-day and intra-day assay precision, accuracy of the LC-MS method for the measurement of AEA in rat brain DVC tissue

Inter-day variation
Intra-day variation
Nominal
concentration
(ng mL-1)
Mean
concentration
found (ng mL-1)
Precision
CV%
Mean
accuracy
(%)
%bias Nominal
concentration
(ng mL-1)
Mean
concentration
found (ng mL-1)
Precision
CV%
Mean
accuracy
(%)
%bias
AEA
1 1.1 12.2 106.3 6.3 1 1.0 14.8 102.8 2.8
25 24.5 15.2 97.9 -2.1 25 25.4 1.7 101.6 1.6
50 50.3 14.7 100.5 0.5 50 52.3 5.6 104.6 4.6
DEA
2 1.7 10.2 84.7 -15.3 2 1.8 9.4 91.8 -8.4
50 49.7 20.0 99.4 -0.6 50 54.1 2.5 108.1 8.1
100 100.1 16.6 100.1 0.1 100 106.3 6.0 106.3 6.3
NADA
1 1.0 3.0 103.9 3.9 1 1.0 13.6 100.4 0.4
25 24.5 12.5 98.1 -1.9 25 24.9 1.4 99.4 -0.6
50 50.2 10.6 100.4 0.4 50 50.6 2.2 101.1 1.1

Mean accuracy(%) = 100*(mean concentration/nominal concentration). % bias = 100*(mean concentration - nominal concentration)/nominal concentration. %CV = 100*(standard deviation/mean concentration)