Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2009 Nov 10.
Published in final edited form as: J Biomed Opt. 2009 Jul–Aug;14(4):044005. doi: 10.1117/1.3160548

Effect of the chest wall on breast lesion reconstruction

Yasaman Ardeshirpour 1, Minming Huang 1, Quing Zhu 1
PMCID: PMC2775096  NIHMSID: NIHMS147011  PMID: 19725717

Abstract

The chest wall underneath the breast tissue affects near-infrared (NIR) diffusive waves measured with reflection geometry. With the assistance of a co-registered ultrasound, the depth and the tilting angle of the chest wall can be determined and are used to model the breast as a two-layer medium. Finite element method (FEM) is suitable for modeling complex boundary conditions and is adapted to model the breast tissue and chest wall. Four parameters of bulk absorption and reduced scattering coefficients of these two layers are estimated and used for imaging reconstruction. Using a two-layer model, we have systematically investigated the effect of the chest wall on breast lesion reconstruction. Results have shown that chest-wall depth, titling angle, and difference between optical properties of two layers of lesion and reference sites affect the lesion reconstruction differently. Our analysis will be valuable and informative to researchers who are using reflectance geometry for breast imaging. The analysis can also provide guidelines for imaging operators to minimize image artifacts and to produce the best reconstruction results.

Keywords: near-infrared (NIR) light, breast cancer detection, ultrasound, optical tomography

1 Introduction

Optical tomography using near-infrared (NIR) diffused light has shown great promise in distinguishing benign from malignant breast tumors and in assessing the chemotherapy response of breast cancers.1-17 Three typical imaging probe configurations or geometries have been used by researchers to acquire NIR data from multiple sources and detectors for tomographic imaging or spectroscopy of breast lesions. The three imaging geometries are transmission,4,5,7,11,12 ring,3,6,8 and reflection,1,2,13 which measure light transmission, transmission and reflection, and reflection, respectively. In transmission geometry or ring geometry, breasts are either sandwiched between a pair of source and detector planes or surrounded by sources and detectors deployed in a ring or multiple rings. Using these two geometries, lesions close to the chest wall are very difficult to access. In reflection geometry, hand-held probes are typically used 1,2,13 to acquire light reflectance from the surface of the breast. Reflection geometry has a significant advantage of probing reduced breast tissue thicknesses when compared with transmission and ring geometries. In reflection geometry, patients are scanned in a supine position, and the breasts are generally flat and can be further compressed as conventionally done with pulse-echo ultrasound imaging. These factors allow lesions closer to the chest wall to be imaged.

We have adopted the light reflection geometry and also integrated the NIR source and detector fibers with a commercial ultrasound transducer on a hand-held probe for dual-modality breast imaging and for ultrasound-guided optical tomography.13,16 A semi-infinite geometry of an absorbing boundary condition was used for optical reflection measurements, and the standard pulse echo mode was used for ultrasound imaging. Patients were scanned in a supine position, and multiple sets of optical measurements were simultaneously made with ultrasound images of a lesion location and of a contralateral normal breast in the same quadrant. The perturbation was calculated as the difference between lesion and reference measurements. Background (bulk) tissue absorption and reduced scattering coefficients were estimated from the reference measurement, and a modified Born approximation was used to relate the scattered field or perturbation measured at all source–detector pairs to total absorption variations within the sample. Two different imaging grids were used for optical imaging reconstruction. The finer mesh was mapped to the lesion region visible with ultrasound, and the coarser mesh was mapped to background tissue. As a result, the total number of imaging voxels with unknown optical properties was significantly reduced, and the tomographic inversion was well defined and converged in few iterations.13

In general, the breast tissue thickness has been reduced to less than 3 to 4 cm when patients are scanned in supine positions. Therefore, lesions close to the chest wall can be imaged with good sensitivity. However, when the chest wall underneath the breast tissue is present within 1 to 2 cm from the skin surface, it affects the measurements and the reconstructed images. In addition, the chest wall, with respect to the image probe, can be tilted at different angles, which can complicate the measurements. Figure 1 shows simulation and experimental measurements of light reflectance (in a logarithmic scale) versus source-detector distance (ρ). Figures 1(a) and 1(b) are simulated and measured data from a homogenous medium (black), a two-layer phantom with the second layer located at 1.0 cm (blue), and 1.4 cm depth (red), respectively. The interface between top and bottom layers was flat with a zero-degree tilting angle. As seen from (a) and (b), the presence of the second layer changes the slope of light reflectance from that acquired in a homogeneous medium. Figures 1(c) and 1(d) are simulated and experimental data from the homogeneous medium (black) and the two-layer phantom with the interface located at 1.0 cm and 1.4 cm depth and tilted at −14.7 deg with respect to the surface. As seen in Figs. 1(c) and 1(d), the tilting of the interface scatters the light reflection data to a larger angular region compared to those acquired with a flat interface.18

Fig. 1.

Fig. 1

Co-registered ultrasound images of a two-layer phantom with the second layer located at 1.0 cm and 1.4 cm depth (flat interface) (b, top) and −14.7 deg (tilted interface) (d, top). Corresponding light reflectance data (logarithmic scale) versus source–detector distance of the two-layer phantom (bottom). Parts (a) and (c) show simulation results, and (b) and (d) show experimental data. Each figure shows three sets of data obtained from a homogenous medium (black), a two-layer phantom with the second layer located at 1 cm (blue) and 1.4 cm (red) depth, respectively. The slopes of the first order fit to simulated data shown in (c) are −0.94 (homogeneous medium), −1.49 (1.0 cm depth), and −1.21 (1.4 cm depth), respectively. The slopes of experimental data shown in (d) are −0.9 (homogeneous), −1.38 (1.0 cm), and −1.23 (1.4 cm depth), respectively. (Color online only.)

In optical tomography, a typical perturbation approach requires two sets of measurements obtained at lesion breast (lesion or target site) and contralateral normal breast (reference site) to compute the perturbation. Therefore, the chest wall underneath the breast tissue at both sites affects the imaging results. In this paper, we provide systematic analysis of how lesion and reference mismatch in terms of chest-wall depth, tilting angle, and background tissue absorption and reduced scattering coefficient can affect the target quantification and distort the imaging quality. Our analysis will be valuable and informative to researchers who are using reflectance geometry for breast imaging. The analysis can also provide guidelines for imaging operators to minimize the mismatch between the two sites using real-time ultrasound information and to produce the best imaging results. To the best of our knowledge, such analysis is not available in the literature.

2 Methods and Computational Procedures

The problem under study can be mathematically modeled as a cylinder with two distinct layers of different optical properties, as shown in Fig. 2. The layer thicknesses, interface geometry (tilting angle), and lesion size and location are inferred from a co-registered ultrasound (US) measurement. Two sets of measurements are obtained at contralateral normal breast (reference site) and lesion breast (lesion or target site). Measurement data from contralateral normal breast are used to estimate the optical properties of each layer in breast-tissue and chest-wall media based on the optimization method introduced in Sec. 2.2. Both sets of measurements together with US information and two-layer optical properties are used in image reconstruction algorithms addressed in Secs. 2.3 and 2.4.

Fig. 2.

Fig. 2

A B-scan ultrasound image (a) and its corresponding two–layer model (b).

2.1 Forward Model

In our simulations, the frequency domain diffusion approximation and the Robin-type (type III) boundary condition have been adopted for forward computation.19 A commercial software package using finite element method (FEM), COMSOL, has been employed to solve the forward diffusion equation. A 3-D cylindrical mesh is generated for forward calculation. The optical source is modeled as an isotropic point source placed approximately one reduced scattering distance, (1μs1+μa1), underneath the boundary, where μs1 and μa1 are the reduced scattering and absorption coefficients of the first layer. The radius and the height of the cylinder are large enough to approximate the semi-infinite geometry. A smooth surface is used to model the tissue and the chest-wall interface. The chest-wall tilting angle with respect to the probe and the tissue-chest interface location in depth h are determined by the co-registered ultrasound images. Figure 2 gives an example of a two-layer model configuration. Figure 2(a) shows a B-scan ultrasound image of a normal breast with the tilted chest wall marked, and Fig. 2(b) is the corresponding two-layer model generated.

2.2 Estimation of Two-Layer Bulk Absorption and Reduced Scattering Coefficients

FEM has been used to relate the bulk absorption and reduced scattering coefficients of the first and second tissue layers to the photon density wave calculated at the surface as

ϕ(r,ω)=f(μa1,μs1,μa2,μs2), (1)

where μa1, μs1, μa2, and μs2 are absorption and reduced scattering coefficients of the first and second layers, respectively. FEM is desirable for modeling an arbitrary interface of breast tissue and chest wall. A nonlinear optimization algorithm (Nelder-Mead algorithm)20 is used with FEM forward calculation to estimate the two-layer background optical properties.

2.3 Weight Matrix Calculation Using the Two-Layer Model

The forward Jacobian weight matrix Wij = [∂ϕij/∂μaj], which relates the photon density wave perturbation at detector i and imaging voxel j with absorption coefficient change Δμaj, is calculated as

Wij=[ϕ11μa1ϕ1LμaLϕ21μa1ϕ2LμaLϕM1μa1ϕMLμaL], (2)

where M is the total number of measurements, and L is the total number of imaging voxels.19 To speed up the Jacobian matrix calculation and also to improve the matrix inversion, we have used a region of interest (ROI) for the weight matrix calculation and imaging display.13 For simulation and phantom experiments and the reported clinical case, the imaging voxel size within ROI is chosen as 0.25 cm × 0.25 cm × 0.5 cm, and the pixel outside of ROI is chosen as 1.5 cm × 1.5 cm × 1 cm. In all simulations and phantom images, we have chosen an 8 × 8 cm ROI in x and y dimensions for a target size of 1 cm diameter to demonstrate the effect of different mismatches.

Using a larger ROI helps visualize the effects of mismatches; however, it reduces the reconstructed target absorption coefficients. With the ultrasound spatial location guidance, we typically use a tighter ROI of two times the target size to improve the accuracy of reconstructed absorption coefficient. For a 1-cm-diam high-contrast absorber, we can achieve about 80% accuracy when the maximum value of reconstructed absorption distribution is used.21 For a low-contrast absorber of the same size, about 100 to 120% of true values can be achieved. The reconstructed absorption values presented in this manuscript correspond to results with only target depth guidance, and they are lower than those with additional spatial guidance. Therefore, we have used the percentage of the reconstructed target absorption normalized to the no-background mismatch case to highlight the relative changes in simulation and phantom studies.

2.4 Inversion

Last, a dual mesh model and a conjugate gradient method are used for reconstruction of optical absorption properties of lesion with the location information provided by the co-registered ultrasound measurement.

2.5 Experimental System

Our experimental system consists of three laser diodes of 690, 780, and 830 nm and 10 parallel detectors. Each laser diode is sequentially switched to nine source positions on the probe, and 10 parallel detection channels acquire backscattered light simultaneously for each source position. More details on our hand-held imaging probe and the NIR system can be found in Ref. 22.

For phantom experiments, intralipid solution and solid plastisol phantoms have been used to emulate the first and second tissue layers, respectively. A commercial ultrasound probe is located in the middle of the hand-held probe to provide the tilting angle and the depth of the second layer and also the region of the target. Each two-layer phantom is imaged with and without a target. The data without a target is used for estimation of background optical properties, and the difference between the target and background data is used to compute perturbation for imaging reconstruction.

The optical properties of the two layers of the phantom were calibrated separately by using a least-squares fitting procedure detailed in Ref. 23. Hereafter, we will refer to these calibrated values as true background optical properties. Briefly, the gains of different source positions and different detector positions as well as the slopes of amplitude (a logarithmic scale) and phase measurements versus source–detector separation were estimated by fitting the measured data using a semi-infinite model of an absorbing boundary condition. The background absorption and reduced scattering coefficients can be readily obtained from the estimated slopes of amplitude and phase measurements.

We also estimated the two-layer optical properties by fitting the reflection measurements using the nonlinear Nelder-Mead algorithm described in Sec. 2.2. We refer to these background values as fitted optical properties in the following text.

Our data acquired from patients suggest that typical optical absorption coefficients of cancers were in the range of 0.2 cm−1 to 0.3 cm−1.(Ref. 14) For typical benign lesions, the absorption coefficients were in the range of 0.03 cm−1 to 0.16 cm−1. Scattering coefficient of lesions was not consistent. For larger tumors, scattering could be quite low compared with the background; while for smaller tumors, scattering coefficient could be higher or lower than that of the background. Therefore, we have chosen an absorption coefficient of 0.20 cm−1 to represent high-contrast tumors, and 0.07 cm−1 to represent benign lesions. We have used similar reduced scattering coefficients for background and targets.

The same setup has been used for clinical studies. The study protocol has been approved by a local Institutional Review Board (IRB) committee. Patients have been scanned in a supine position, and multiple sets of optical measurements were made simultaneously with co-registered ultra-sound images of the lesion breast and the normal contralateral breast in the same quadrant as the lesion. Data at the contralateral breast have been used for the estimation of background optical properties as well as the calculation of perturbation for image reconstruction.

3 Results

3.1 Effect of Layer Depth Mismatch between Lesion and Reference

Figure 3 shows simulation results of two-layer interface depth mismatch between the reference and target sites. In the simulation, the second layer of the target site was kept at 1.5 cm depth, and the second layer of the reference site was varied from 1.5 cm, to 1.75 cm, and to 2 cm in Figs. 3(a)–3(c), and to 1.4 cm and 1.3 cm in Figs. 3(d) and 3(e), respectively. The optical properties of two layers were chosen as μa1=0.02, μs1=7.0, μa2=0.1, and μs2=7.0cm1. The target optical properties were chosen as μa=0.20 and μs=7.0cm1. When the second-layer depth of the reference site matches the second-layer depth of the target site, the reconstructed target appears at the correct location with no artifacts [Figs. 3(a-1)]. The reconstructed maximum absorption coefficient was only 39% of the true value because of the use of a larger ROI (8 times the target size) in spatial dimension. With a tighter ROI (two times the target size) in spatial dimension, the reconstructed maximum absorption coefficient reached 78% of the true value [Figs. 3(a-2)]. The scale of figures (a-2) was adjusted to show a better quantification of the target. To highlight the relative changes when the second-layer depth and titling angle as well as the bulk optical properties of the two layers were varied, we have used 8 times target size as the ROI in the rest of the simulation and phantom studies. Therefore, the results are pertinent to optical tomography with target depth guidance only.

Fig. 3.

Fig. 3

Effect of second-layer depth mismatch between the reference site and the target site. The second layer at the reference site was located at (a) 1.5 cm, (b) 1.75 cm, (c) 2 cm, (d) 1.4 cm, and (e) 1.3 cm in depth. The second layer of the target site was located at 1.5 cm depth. Background optical properties at both reference and target sites are μa1=0.02, μs1=7.0, μa2=0.1, and μs2=7.0cm1. The top picture of each part shows the position of the second layer and the target, and the bottom pictures [(a-1) and (b) to (e)] show the reconstructed absorption maps using an ROI of eight times the target size, while the bottom picture (a-2) shows the absorption maps reconstructed with an ROI of two times the target size. The color bar is the absorption coefficient in units of cm−1, and color bars in (a-2) was adjusted to 0.2 cm−1 for better visualization of the target. In the absorption map, each slice presents a spatial image of 8 cm×8 cm obtained from 0.4 cm underneath the probe surface to 2.9 cm in depth, with 0.5-cm spacing between slices. The spatial image dimensions for each slice in the absorption map and the spacing between the slices were kept the same as in Fig. 3 in the following figures except in Fig. 14. (Color online only.)

When the second layer of the reference site is deeper than that of the target site, image artifacts appear. The artifact is more pronounced when the second layer is deeper [Figs. 3(b) and 3(c)]. When the second layer of the reference site was placed at a lower depth than that of the target site, for example, at 1.4 or 1.3 cm, the reconstructed absorption coefficient and the size of the target became lower and smaller [Figs. 3(d) and 3(e)]. The decrease in reconstructed absorption coefficient of the target depends on the optical properties of the two-layer media at the target and reference sites and the depth mismatch.

Table 1 shows the fitted optical properties of the two layer media used in obtaining Fig. 3. The nonlinear Nelder-Mead algorithm discussed in Sec. 2.2 was used for fitting and 1% noise was added to the simulated data. The fitting error of μs2 is larger when the second layer is deeper.24,25 Using the fitted background optical properties, we reconstructed the target again, and the percentage of maximum reconstructed absorption coefficient versus the reference interface depth is shown in Fig. 4. Compared to those obtained using true background values, the results of using fitted background values are essentially the same.

Table 1.

Fitted optical properties of two-layer media used in Fig. 3.

Depth (cm) μa1 (cm−1) μs1(cm1) μa2 (cm−1) μs2(cm1)
True value 0.02 7 0.1 7
1.3 0.023 6.84 0.093 7.63
1.4 0.023 6.87 0.092 7.54
1.5 0.022 6.95 0.094 7.18
1.75 0.016 7.07 0.11 6.6
2.0 0.023 6.85 0.09 10.79

Fig. 4.

Fig. 4

Percentage of maximum reconstructed absorption coefficient of the target normalized to the no-background mismatch case as a function of the two-layer interface depth at the reference site using both calibrated background and fitted background optical properties (simulation data). The two-layer interface at the target side is located at 1.5 cm from the probe.

To validate the simulation results, we performed phantom experiments. An example of phantom experiments is given in Fig. 5. Figure 5(a) shows a B-scan ultrasound image of the two-layer medium with a low-contrast absorber located at (x, y, z)=(0,0,0.9 cm) inside the intralipid solution (target site). The top layer of the phantom was made of a homogeneous 0.8% intralipid solution of calibrated values μa =0.025 cm−1 and μs=7.53cm1, and the bottom layer was a solid phantom made of plastisol of calibrated values μa =0.08 cm−1 and μs=6.5cm1 at 780 nm. The target was a 1.0-cm-diam spherical absorber of calibrated optical properties μa=0.07 cm−1 and μs=5.5cm1 at 780 nm. The plasitsol phantom was used as a chest-wall layer and placed at 1.4 cm underneath the intralipid with zero-degree tilt. Figure 5(b) is a B-scan ultrasound image of the phantom without target (reference site). Figure 5(c) shows the reconstructed absorption map at 780 nm, which is an ideal image obtained without any mismatch. Figure 5(d) shows a B-scan ultrasound image of the two-layer medium with the target. The image is the same as Fig. 5(a). Figure 5(e) is a B-scan ultrasound image of the reference site with the plastisol phantom placed at 2 cm depth underneath the intralipid with zero-degree tilt. Figure 5(f) shows the reconstructed absorption map at 780 nm. The image artifacts pattern is similar to the one obtained from simulations under the same conditions.

Fig. 5.

Fig. 5

Reconstructed absorption map of a 1-cm-diam spherical target with calibrated optical properties μa=0.07 and μs=5.5cm1, located at (x,y,z)=(0,0,0.9 cm). (a) B-scan ultrasound image of a two-layer medium with the target. A plastisol phantom was located at 1.4 cm underneath the probe surface with a zero-degree tilting angle. (b) B-scan ultrasound image of the phantom without the target. (c) Reconstructed target absorption map at 780 nm. (d) B-scan ultrasound image of the two-layer medium with the target [same as (a)]. (e) The two-layer medium with the plastisol phantom located at 2 cm with a zero-degree tilting angle. (f) Reconstructed target absorption map at 780 nm.

A sequence of phantom experiments was performed to further validate the simulation results. The top layer of the phantom was made of a homogeneous 0.8% intralipid solution of calibrated optical properties μa=0.024 cm−1 and μs=8.25cm1, and the bottom layer was a solid phantom made of plastisol of calibrated values μa=0.08 cm−1 and μs=6.5cm1 at 780 nm. The target was a 1.0-cm-diam spherical absorber of calibrated optical properties μa=0.23 cm−1 and μs=5.45cm1 at 780 nm. The two-layer interface was located at 1.5 cm from the probe at the target site, and it varied from 1.2 to 2 cm at the reference site. The target and reference measurements were made from the two-layer medium with and without the target, respectively. The percentage of maximum reconstructed absorption coefficient of the target normalized to the no-background mismatch case using both calibrated and fitted background optical properties is shown in Fig. 6. The fitted optical properties of the two-layer media for the phantom experiments are given in Table 2. The error of the fitted second layer μs2 is larger when the second layer is deeper.24,25 However, the reconstructed target μa is reasonably robust to the fitting error.

Fig. 6.

Fig. 6

Percentage of the maximum reconstructed μa normalized to the no-background mismatch case as a function of the two-layer interface depth at the reference site (phantom data) using both calibrated background and fitted background optical properties values. The two-layer interface at the target site is located at 1.5 cm depth from the probe.

Table 2.

Fitted optical properties of two-layer phantoms.

Depth (cm) μa1 (cm−1) μs1(cm1) μa2 (cm−1) μs2(cm1)
True value 0.023 8.25 0.08 6.5
1.0 0.019 7.28 0.074 9.4
1.3 0.022 9.79 0.10 6.6
1.5 0.023 7.5 0.098 19.45
2.0 0.024 8.25 0.088 10.25

The preceding reported simulation and experimental results can be explained intuitively. As the second layer at the reference site becomes shallower, more photons are absorbed by the second layer, and therefore fever photons are detected at the surface. As a result, the perturbation, which is the normalized difference between lesion data and reference data, is reduced. This reduction in perturbation degrades the target reconstruction accuracy and reduces the target contrast. Certainly, the reduction in the reconstructed target absorption with depth mismatch depends on the optical properties of the two layers and target contrast. On the other hand, as the second layer at the reference site becomes deeper, fever photons are absorbed by the second layer and more are detected at the surface of the reference site. As a result, the perturbation consists of both portions caused by lesion and lesion-reference background mismatch. The perturbation caused by lesion-reference background mismatch produces image artifacts in background regions and increases the reconstructed absorption coefficient of the target.

3.2 Effect of Mismatch Tilting Angle between Lesion and Reference Sites on Image Reconstruction

As shown in Fig. 1, when the second layer is tilted with respect to the imaging probe, the amplitude and phase (not shown) measurements spread over a larger angular region. Figure 7 demonstrates the effect of mismatch between the interface tilting angle in reference and lesion measurements. In this set of simulations, the second-layer depth in both the target and reference measurements was kept at 1.5 cm, and the interface angle at the target site was zero deg. The mismatch at the second layer of the reference site was introduced as −10, −5, 0, 5, or 10 deg in the y or x direction. The optical properties of the two layers were chosen as μa1=0.02, μs1=7.0, μa2=0.1, and μs2=7.0cm1. The target optical properties were chosen as μa=0.20 and μs=7.0cm1. In our system, the ultrasound probe is located at the center of the optical probe. Therefore, with the assistance of the real-time ultra-sound image, the lesion is always positioned in the center of the optical probe. The results show that the mismatch angle introduces a target position shift and reduces the accuracy of the reconstructed absorption coefficient. Figure 8 shows the percentage of maximum reconstructed μa relative to the values obtained with the flat interface (no-background mismatch) as a function of the tilting angle difference between the target and the reference sites. As a result of this lesion-reference tilting angle mismatch, the reconstructed target degrades in accuracy and shifts in position.

Fig. 7.

Fig. 7

Reconstructed absorption maps of a target located at (x,y,z) =(0,0,0.9 cm). The first-and second-layer interfaces of both the target and the reference site were positioned at 1.5 cm depth. The interface at the target site has a zero-degree tilting angle. (a) to (i) sequentially show the absorption maps of the target with a target–reference layer tilting angle mismatch of (a) 0 deg, (b) 10 deg, (c) 5 deg, (d) −10 deg, and (e) −5 deg in the x direction, and of (f) 10 deg, (g) 5 deg, (h) −10 deg, and (i) −5 deg mismatch in the y direction. Optical properties of both the reference and target sites are μa1=0.02, μs1=7.0, μa2=0.1, and μs2=7.0cm1 The top picture of each part shows the position of the second layer and the target, and bottom picture shows the reconstructed absorption map.

Fig. 8.

Fig. 8

Percentage of maximum reconstructed μa relative to the values obtained with the flat interface (no-background mismatch) as a function of the difference between the tilting angle of the two-layer interface at the target and reference sites.

We validated the simulation results with phantom experiments using the same setup and the target as described in the previous section. Figure 9(a) shows a B-scan ultrasound image of a two-layer medium with target located at (x, y, z) =(0,0,0.9 cm). The solid phantom was used as a chest-wall layer and placed underneath the intralipid at 1.4 cm with a zero-degree tilting angle. Figure 9(b) is the B-scan ultrasound image of the same phantom without target and −8 deg titling angle. Figure 9(c) shows the reconstructed absorption map at 780 nm. Similar distortion effects are observed as seen in simulations.

Fig. 9.

Fig. 9

Reconstructed absorption map of 1-cm-diam spherical target located at (0, 0, 0.9 cm) with calibrated optical properties μa =0.07 cm−1 and μs=5.5cm1. (a) B-scan ultrasound image of a two-layer medium with target. The plastisol phantom was located at 1.4 cm with a zero-degree tilting angle. (b) B-scan ultrasound image of the same phantom without the target. The layer was located at 1.4 cm with −8-deg tilting angle. (c) Reconstructed target absorption map at 780 nm.

3.3 Effect of Optical Property Mismatch between Lesion and Reference Media

In this section, we investigate the effect of background optical properties on image reconstruction. In our clinical experiments, background optical properties were estimated from the normal contralateral breast of the same quadrant with the assumption that both breasts at the symmetric locations have similar bulk optical properties. This is not a problem in general (see Sec. 4). However, for patients who have reconstructed contralateral breasts due to prior surgery, the estimated bulk optical properties at the reference site can be different from the lesion site regardless of the absorption properties of the lesion. For patients with contralateral disease, the optical properties of the reference site may also defer from that at the lesion site.

Figure 10 demonstrates the effect of mismatch between the first-layer absorption coefficient of reference site and the lesion site. The first-layer absorption coefficient has been chosen as 0.023 cm−1 (intralipid) and 0.04 cm−1 (intralipid with ink). Other optical properties have calibrated values of μs1=7.5, μa2=0.08, and μs2=6.5cm1. The target is the 1-cm solid phantom with calibrated optical properties of μa=0.23 and μs=5.45cm1. Figure 10 shows the reconstructed image, and Fig. 11 shows the changes in the reconstructed absorption coefficient for cases when the target site has higher, equal, and lower first-layer absorption coefficient compared to the reference site. The comparison has been made using both the true and fitted background optical properties, and the results are essentially the same. The fitted optical properties of the two-layer media were μa1=0.043, μs1=8.07, μa2=0.072, and μs2=13.22cm1 for the case of higher first-layer μa1 at the target site and μa1=0.023, μs1=7.5, μa2=0.098, and μs2=19.45cm1 for the case of lower μa1 at the target site.

Fig. 10.

Fig. 10

Effect of first-layer μa mismatch between the reference site and the target site on the reconstructed image. First-layer optical properties at the target and reference sites are (a) μa1_tar=0.023, μa1_ref=0.023; (b) μa1_tar=0.04, μa1_ref=0.023; (c) μa1_tar=0.04, μa1_ref=0.04; and (d) μa1_tar =0.023, μa1_ref=0.04. The other background optical properties of both the reference and target sites are μs1=7.5, μa2=0.08, and μs2=6.5cm1. The optical properties of the target are μa=0.23 and μs=5.45cm1.

Fig. 11.

Fig. 11

Percentage of maximum reconstructed μa normalized to the no-background mismatch case using both calibrated and fitted background values. Lower μa1_tar shows the results when μa1_tar=0.023 and μa1_ref=0.04 cm−1, and higher μa1_tar shows the results when μa1_tar=0.04, and μa1_ref=0.023 cm−1. The other optical properties of both the target and reference sites are μs1=7.5, μa1=0.08, and μs2=6.5cm1.

The preceding experiments can be explained intuitively. As the first-layer background absorption of the reference site increases, more photons are absorbed and fewer photons are reflected. As a result, the perturbation is reduced as compared with no mismatch, and the reconstructed target contrast is lower and the target mass becomes smaller. On the other hand, lower μa1 of the reference site or higher μa1 of the target site increases the perturbation caused by non-target-related background mismatch, which contributes to image artifacts in the background regions.

The effect of the first layer μs1 mismatch on the reconstructed absorption coefficient of the target is shown in Fig. 12. The top layer of the phantom was made of a homogeneous intralipid solution of different concentrations. The calibrated μs1 varied from 5.8 to 8.2 cm−1 with μa1=0.024 cm−1, and the bottom layer was a solid plastisol phantom with calibrated values of μa=0.08 cm−1 and μs=6.5cm1 at 780 nm. The target was the 1.0-cm-diam spherical absorber of calibrated optical properties μa=0.23 cm−1 and μs=5.45cm1 at 780 nm. While μs1 was changed from 5.8 to 8.2 cm−1 at the target site, it was fixed to 7.0 cm−1 at the reference site. Other optical properties of both reference and target sites were μa1=0.024, μa2=0.08, and μs2=6.5cm1. Figure 13 shows the percentage of reconstructed absorption coefficient of the target normalized to the no-background mismatch case using both true and fitted values of two-layer optical properties. The fitted optical properties of the two-layer medium were μa1 =0.031, μs1=10.49, μa2=0.097, and μs2=7.25cm1.

Fig. 12.

Fig. 12

Effect of the first-layer μs1 mismatch on the reconstructed absorption map of the target. The first-layer reduced scattering coefficient at the target site is (a) μs1=5.8cm1, (b) μs1=6.5cm1, (c) μs1=7cm1, (d) μs1=7.3cm1, (e) μs1=7.7cm1, and (f) μs1=8.15cm1, respectively. μs1 at the reference site is 7.0 cm−1. Other optical properties of both the reference and lesion sites are μa1=0.024, μa2=0.08, and μs2=6.5cm1.

Fig. 13.

Fig. 13

Percentage of maximum reconstructed μa normalized to the no-background mismatch case as a function of first-layer reduced scattering coefficient at the target site using both calibrated and fitted background optical properties. The reduced scattering coefficient is 7.0 cm−1 at the reference site.

As shown in Figs. 12 and 13, a higher background scattering at the target site produces image artifacts, and a lower background scattering causes reduction in target contrast. This is because the higher background scattering at the target site increases the perturbation due to decreased signal strength measured at the target site. The extra perturbation due to background scattering mismatch contributes to the image artifacts in the background region, while the lower background scattering at the lesion site causes the reduction in perturbation due to the increased signal strength measured at the target site. As a result, the reduction in perturbation due to scattering mismatch causes the decrease in reconstructed target absorption.

3.4 Clinical Examples

To evaluate the mismatch in breast imaging, we reconstructed the absorption map of a benign fibroadenoma using three different reference sites acquired at the contralateral breast. Figure 14(a) shows a co-registered ultrasound B-scan of the lesion. The center of the lesion was at 1.2 cm depth, and the chest-wall layer was located approximately at 1.5 cm with an 8-deg tilting angle. The fitted optical properties of the background were μa1=0.049, μs1=9.87, μa2=0.11, μs2=2.9cm1 at 830 nm. Figure 14(b) shows the US B-scan of the first reference and the corresponding reconstructed absorption map at 830 nm. The chest-wall interface at the reference is almost flat and located at 1.5 cm depth. Figure 14(c) shows the second reference used and the reconstructed absorption map. This reference is deeper and tilted compared to the lesion site; therefore, the lesion mass was more spread out compared to part (a). Figure 14(d) shows the third reference and the reconstructed absorption map. This reference is almost flat and deeper than the reference in part (a); therefore, the lesion mass was more spread out compared to part (a). Due to the depth mismatch, some artifacts appeared in the figure; however, since the chest-wall and breast tissue of clinical cases are not homogenous and their interference is not well-defined as in the phantom and simulation studies, the artifacts are not as pronounced as the phantom and simulation. The reconstructed maximum and average μa of lesion for different reference sites are listed in Table 3.

Fig. 14.

Fig. 14

Co-registered ultrasound and reconstructed absorption maps of a benign fibroadenoma. (a) B-scan ultrasound image of the lesion site. The chest-wall layer was located at 1.5 cm with 8 deg tilt. (b) B-scan ultrasound of the first reference site and the reconstructed absorption maps at 830 nm. (c) B-scan ultrasound of the tilted reference site and the reconstructed absorption maps at 830 nm. (d) B-scan ultrasound of the third reference site and the reconstructed absorption maps at 830 nm. Each slice presents a spatial image of 8 cm×8 cm obtained from 0.18 cm underneath the probe surface to 2.68 cm in depth, with 0.5-cm spacing between slices.

Table 3.

Comparison of four parameters for the clinical case shown in Fig. 14: interface depth and tilting angle of the two-layer interface at the reference site and maximum and mean values of the reconstructed target absorption coefficient.

Depth
(cm)
Tilting angle
(deg)
Maximum
target μa
(cm−1)
Mean
target μa
(cm−1)
Figure 14(b) 1.5 0 0.19 0.12
Figure 14(c) 1.7 11 0.19 0.12
Figure 14(d) 1.8 0 0.18 0.13

4 Discussion

In this paper, the effects of depth, tilting angle, and optical property mismatch between the reference and lesion sites of two-layer media on image reconstruction have been systematically investigated. Depth mismatch of the second layer between reference and lesion sites can produce image artifacts or reduce target contrast in reconstructed images. If the second-layer depth at the reference site is shallower than that at the target site, the reconstructed target absorption and target size are reduced. On the other hand, when the second-layer depth at the reference site is deeper than that at the target site, the reconstructed image has multiple artifacts in background regions cased by target-reference background mismatch. Angle mismatch between the target and reference layers reduces the reconstructed target contrast and shifts the target position. Mismatch between optical properties of reference and lesion layers can also introduce artifacts or reduce target contrast.

This study focused on characterizing and understanding the effect of chest-wall mismatch between lesion and reference sites. We have used the percentage of the reconstructed target absorption normalized to the no-background mismatch case to highlight the relative changes in simulation and phantom studies. The ROI used for reconstruction with a finer mesh was eight times larger than the target size. A large ROI helps visualize the effects of mismatches, especially in the background regions. However, a large ROI reduces the reconstructed target absorption coefficients, as demonstrated in Fig. 3(a). With ultrasound spatial location guidance, we typically use a tighter ROI of two times target size to improve light qualification. As shown in Fig. 3(a) and our previous studies, we can achieve about 80% accuracy for high-contrast targets when the maximum value of reconstructed absorption map is used, and about 100 to 120% the true value for low-contrast targets.21 With spatial guidance in addition to target depth from the co-registered ultrasound, the target artifacts at the background region can be recognized and further reduced. In clinical cases, where the breast tissue and chest-wall interfaces are not well-defined, the artifacts are less pronounced than the phantom and simulation studies.

Breasts of each patient are symmetric with respect to density, tissue components, and shape as well as chest-wall position. In our clinical studies, we use real-time co-registered ultrasound to position the hand-held probe at the lesion site and the contralateral reference site of the same quadrant by matching the probe position and the chest-wall layer depth. This effort minimizes the mismatch and produces the best imaging reconstruction results. Because our probe covers about 10-cm diam area of tissue, the fitted optical properties at the carefully chosen reference site are the best representations of the average properties of the contralateral lesion location. The estimated bulk optical properties are quite similar at the symmetric sites. Table 4 lists bulk optical properties obtained from 10 consecutive patients with different chest-wall conditions. The measurements were taken from the same quadrant at left and right breasts without any known disease. The fitting was based on a one-layer semi-infinite model. On average, the bulk absorption and reduced scattering coefficients between left and right differ in absolute value by 0.004 cm−1 (±0.003) and 0.62 cm−1 (±0.52) at 780 nm, respectively. At 830 nm, the bulk absorption and reduced scattering coefficients differ in absolute value by 0.007 cm−1 (±0.005) and 0.85 cm−1 (±0.59), respectively. Note that the background optical absorption is higher for patients 2 to 4 because the fitting is affected by the presence of a shallower chest wall.

Table 4.

Bulk optical properties of 10 consecutive patients obtained from left and right healthy breasts at the same quadrant.

Case# Chest-wall depth
(cm)
Wavelength(nm) Right breast
(μaμs)
(cm−1/cm−1)
Left breast
(μaμs)
(cm−1/cm−1)
1 2.0 780 0.020/3.31 0.027/4.06
830 0.036/4.03 0.044/4.85
2 1.8 780 0.053/5.98 0.055/7.57
830 0.081/6.30 0.090/7.24
3 1.5 780 0.065/6.48 0.066/6.27
(tilted) 830 0.072/6.60 0.090/5.56
4 1.4 780 0.060/4.96 0.059/5.03
(tilted 5 deg) 830 0.083/4.84 0.084/4.62
5 >4 780 0.004/2.87 0.006/2.67
830 0.020/3.18 0.024/2.83
6 2.5 780 0.007/3.55 0.010/3.20
(tilted 11 deg) 830 0.024/3.82 0.027/3.38
7 2.5 780 0.009/2.59 0.003/1.92
(tilted 11 deg) 830 0.026/3.03 0.022/2.18
8 2 780 0.041/3.83 0.044/4.97
(flat) 830 0.051/4.36 0.053/5.68
9 Implants under 1.8 cm 780 0.026/4.08 0.017/2.96
breast tissue 830 0.039/4.50 0.028/3.59
10 2 780 0.020/4.34 0.022/4.23
(tilted 8 deg) 830 0.035/5.16 0.040/4.76

However, this data acquisition process depends on the operators' knowledge and skills. Therefore, the reported studies provide important guidelines to imagers and also help with interpretation of imaging results. In many cases, the perfect matching between chest-wall layer depths at both sites is not obtainable; for example, patients have contralateral diseases. For these cases, the reference should be chosen as close as possible to the lesion site if the chest wall is within 2 cm depth. Currently, we are investigating procedures and algorithms to model the mismatch and to compensate the mismatch-induced perturbation.

Optical properties of breast tissue vary from patient to patient. From literature data and our own experience, the bulk absorption coefficient μa is in the range of 0.01 cm−1 to 0.08 cm−1, and the reduced scattering coefficient μs is in the range of 2.0 cm−1 to 12.0 cm−1. As shown in Table 4, the fitted optical properties using a one-layer semi-infinite model were affected by the chest-wall locations for patients with shallower chest walls. As a result, the reconstructed lesion absorption coefficient is typically higher for this subgroup of patients because the high background value contributes to the total target μa. Our initial experience obtained by fitting the two-layer optical properties of this subgroup of patients suggests21 that the fitted first-layer μa1 is in general lower than the fitted background μa obtained from the one-layer model and that the fitted second-layer μa2 is in general higher than the fitted one-layer μa. The fitted first-layer μs1 is similar or higher than the fitted μs obtained from the one-layer model, and the fitted second-layer μs2 is in general similar or lower than the fitted one-layer μs. Two-layer imaging reconstruction in general improves the accuracy of the reconstructed target μa because of the use of more accurate first-layer background μa1. The optical properties of the chest wall are not available in the literature. The only available data are optical properties of muscles and bones. In Refs. 26-28, muscle was reported to have a large absorption and medium scattering coefficients, with μa in the range of 0.08 cm−1 to 0.27 cm−1 and μs in the range of 5 cm−1 to 10 cm−1. Bone is a highly absorbing and scattering medium. From the literature,29,30 bone was reported to have large absorption and scattering coefficients, with μa in the range of 0.05 cm−1 to 0.15 cm−1 and μs in the range of 20 cm−1 to 30 cm−1. Our data on μa2 agrees with the limited literature data that muscles and bones have higher average μa. More clinical cases are being evaluated for this subgroup of patients, and the statistics will be valuable to breast imagers using diffused wave optical tomography.

The reported studies pertain to the use of optical tomography using a perturbation approach. Other model-based approaches may have advantages over the perturbation approach in this regard. However, breast tissue has a wide range of background absorption and reduced scattering coefficients, and these bulk tissue values change with age, menopause status, hormone regulation, etc.31 The model-based approach may have great difficulty adapting to this wide range of bulk optical properties, while the perturbation approach of using a patient normal site as a control greatly minimizes this problem.

Acknowledgments

This work has been supported by the National Institute of Health (R01EB002136) and the Patrick & Catherine Weldon Donaghue Medical Research Foundation.

References

  • 1.Tromberg BJ, Cerussi A, Shah N, Compton M, Durkin A, Hsiang D, Butler J, Mehta R. Diffuse optics in breast cancer: detecting tumors in pre-menopausal women and monitoring neoadjuvant chemotherapy. Breast Cancer Res. 2005;7:279–285. doi: 10.1186/bcr1358. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Chance B, Nioka S, Zhang J, Conant EF, Hwang E, Briest S, Orel SG, Schnall MD, Czerniecki BJ. Breast cancer detection based on incremental biochemical and physiological properties of breast cancers: a six-year, two-site study. Acad. Radiol. 2005;12:925–933. doi: 10.1016/j.acra.2005.04.016. [DOI] [PubMed] [Google Scholar]
  • 3.Poplack SP, Tosteson TD, Wells WA, Pogue BW, Meaney PM, Hartov A, Kogel CA, Soho SK, Gibson JJ, Paulsen KD. Electromagnetic breast imaging: results of a pilot study in women with abnormal mammograms. Radiology. 2007;243:350–359. doi: 10.1148/radiol.2432060286. [DOI] [PubMed] [Google Scholar]
  • 4.Choe R, Corlu A, Lee K, Durduran T, Konecky SD, Grosicka-Koptyra M, Arridge SR, Czerniecki BJ, Fraker DL, Demichele A, Chance B, Rosen MA, Yodh AG. Diffuse optical tomography of breast cancer during neoadjuvant chemotherapy: a case study with comparison to MRI. Med. Phys. 2005;32:1128–1139. doi: 10.1118/1.1869612. [DOI] [PubMed] [Google Scholar]
  • 5.Heffer E, Pera V, Schütz O, Siebold H, Fantini S. Near-infrared imaging of the human breast: complementing hemoglobin concentration maps with oxygenation images. J. Biomed. Opt. 2004;9:1152–1160. doi: 10.1117/1.1805552. [DOI] [PubMed] [Google Scholar]
  • 6.Gu X, Zhang Q, Bartlett M, Schutz L, Fajardo LL, Jiang H. Differentiation of cysts from solid tumors in the breast with diffuse optical tomography. Acad. Radiol. 2004;11:53–60. doi: 10.1016/s1076-6332(03)00562-2. [DOI] [PubMed] [Google Scholar]
  • 7.Spinelli L, Torricelli A, Pifferi A, Taroni P, Danesini G, Cubeddu R. Characterization of female breast lesions from multi-wavelength time-resolved optical mammography. Phys. Med. Biol. 2005;50:2489–2502. doi: 10.1088/0031-9155/50/11/004. [DOI] [PubMed] [Google Scholar]
  • 8.Schmitz CH, Klemer DP, Hardin R, Katz MS, Pei Y, Graber HL, Levin MB, Levina RD, Franco NA, Solomon WB, Barbour RL. Design and implementation of dynamic near-infrared optical tomographic imaging instrumentation for simultaneous dual-breast measurements. Appl. Opt. 2005;44:2140–2153. doi: 10.1364/ao.44.002140. [DOI] [PubMed] [Google Scholar]
  • 9.Brooksby B, Pogue BW, Jiang S, Dehghani H, Srinivasan S, Kogel C, Tosteson TD, Weaver J, Poplack SP, Paulsen KD. Imaging breast adipose and fibroglandular tissue molecular signatures by using hybrid MRI-guided near-infrared spectral tomography. Proc. Natl. Acad. Sci. U.S.A. 2006;103:8828–8833. doi: 10.1073/pnas.0509636103. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Shah N, Gibbs J, Wolverton D, Cerussi A, Hylton N, Tromberg BJ. Combined diffuse optical spectroscopy and contrast-enhanced magnetic resonance imaging for monitoring breast cancer neoadjuvant chemotherapy: a case study. J. Biomed. Opt. 2005;10:051503. doi: 10.1117/1.2070147. [DOI] [PubMed] [Google Scholar]
  • 11.Zhang Q, Brukilacchio TJ, Li A, Stott JJ, Chaves T, Hillman E, Wu T, Chorlton MA, Rafferty E, Moore RH, Kopans DB, Boas DA. Coregistered tomographic x-ray and optical breast imaging: initial results. J. Biomed. Opt. 2005;10:1–9. doi: 10.1117/1.1899183. [DOI] [PubMed] [Google Scholar]
  • 12.Ntziachristos V, Yodh AG, Schnall MD, Chance B. MRI-guided diffuse optical spectroscopy of malignant and benign breast lesions. Neoplasia. 2002;4:347–354. doi: 10.1038/sj.neo.7900244. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Zhu Q, Chen NG, Kurtzman SH. Imaging tumor angiogenesis by use of combined near-infrared diffusive light and ultrasound. Opt. Lett. 2003;28:337–339. doi: 10.1364/ol.28.000337. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Zhu Q, Huang M, Chen N, Zarfost K, Jagjivant B, Kane M, Hedget P, Kurtzmant SH. Ultrasound-guided optical tomographic imaging of malignant and benign breast lesions: initial clinical results of 19 cases. Neoplasia. 2003;5:379–388. doi: 10.1016/s1476-5586(03)80040-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Zhu Q, Cronin EB, Currier AA, Vine HS, Huang M, Chen N, Xu C. Benign versus malignant breast masses: optical differentiation with US-guided optical imaging reconstruction. Radiology. 2005;237:57–66. doi: 10.1148/radiol.2371041236. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Zhu Q, Kurtzman SH, Hegde P, Tannenbaum S, Kane M, Huang M, Chen NG, Jagjivan B, Zarfos K. Utilizing optical tomography with ultrasound localization to image heterogeneous hemoglobin distribution in large breast cancers. Neoplasia. 2005;7:263–270. doi: 10.1593/neo.04526. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Cerussi A, Hsiang D, Shah N, Mehta R, Durkin A, Butler J, Tromberg BJ. Predicting response to breast cancer neoadjuvant chemotherapy using diffuse optical spectroscopy. Proc. Natl. Acad. Sci. U.S.A. 2007;104:4014–4019. doi: 10.1073/pnas.0611058104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Ardeshirpour Y, Huang M, Zhu Q. Effect of tilted chest-wall on breast lesion reconstruction. Proc. Biomedical Optics (BIOMED) 2008:16–19. doi: 10.1117/1.3160548. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Huang M. 3-D Optical imaging for breast cancer detection and diagnosis. Electrical and Computer Engineering Department, Univ. of Connecticut; 2005. PhD Dissertation. [Google Scholar]
  • 20.Das M, Xu C, Zhu Q. Analytical solution for light propagation in a two-layer tissue structure with a tilted interface for breast imaging. Appl. Opt. 2006;45:5027–5036. doi: 10.1364/ao.45.005027. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Xu C, Das M, Ardeshirpour Y, Zhu Q. An image reconstruction method for a two-layer tissue structure accounts for chest-wall effects in breast imaging. J. Biomed. Opt. 2008;13:064029. doi: 10.1117/1.3041497. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Zhu Q, Xu C, Guo P, Aquirre A, Yuan B, Huang F, Castilo D, Gamelin J, Tannenbaum S, Kane M, Hedge P, Kurtzman S. Optimal probing of optical contrast of breast lesions of different size located at different depths by US localization. Technol. Cancer Res. Treat. 2006;5:365–380. doi: 10.1177/153303460600500408. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Chen NG, Guo PY, Yan SK, Piao DQ, Zhu Q. Simultaneous near-infrared diffusive light and ultrasound imaging. Appl. Opt. 2001;40:6367–6380. doi: 10.1364/ao.40.006367. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Kienle A, Patterson MS, Dögnitz N, Bays R, Wagnières G, Van Den Bergh H. Noninvasive determination of the optical properties of two-layered turbid media. Appl. Opt. 1998;37:779–791. doi: 10.1364/ao.37.000779. [DOI] [PubMed] [Google Scholar]
  • 25.Ripoll J, Ntziachristos V, Culver JP, Pattanayak DN, Yodh AG, Nieto-Vesperinas M. Recovery of optical parameters in multiple-layered diffusive media: theory and experiments. J. Opt. Soc. Am. A Opt. Image Sci. Vis. 2001;18:821–830. doi: 10.1364/josaa.18.000821. [DOI] [PubMed] [Google Scholar]
  • 26.Cheong W, Prahl SA, Welch AJ. A review of the optical property of biological tissues. IEEE J. Quantum Electron. 1990;26:2166–2185. [Google Scholar]
  • 27.Matcher SJ, Cope M, Delpy DT. In vivo measurements of the wavelength dependence of tissue-scattering coefficients between 760 and 900 nm measured with time-resolved spectroscopy. Appl. Opt. 1997;36(1):386–396. doi: 10.1364/ao.36.000386. [DOI] [PubMed] [Google Scholar]
  • 28.Yang Y, Soyemi O, Landry M, Soller B. Influence of a fat layer on the near infrared spectra of human muscle: quantitative analysis based on two-layered Monte Carlo simulations and phantom experiments. Opt. Express. 2005;13(5):1570–1579. doi: 10.1364/opex.13.001570. [DOI] [PubMed] [Google Scholar]
  • 29.Taroni P, Comelli D, Farina A, Pifferi A, Kienle A. Time-resolved diffuse optical spectroscopy of small tissue samples. Opt. Express. 2007;15(6):3301–3311. doi: 10.1364/oe.15.003301. [DOI] [PubMed] [Google Scholar]
  • 30.Xu Y, Iftimia N, Jiang H, Key LL, Bolster MB. Imaging of in vitro and in vivo bones and joints with continuous-wave diffuse optical tomography. Opt. Express. 2001;8(7):447–451. doi: 10.1364/oe.8.000447. [DOI] [PubMed] [Google Scholar]
  • 31.Shah N, Cerussi A, Eker C, Espinoza J, Butler J, Fishkin J, Hornung R, Tromberg B. Noninvasive functional optical spectroscopy of human breast tissue. Proc. Natl. Acad. Sci. U.S.A. 2001;98:4420–4425. doi: 10.1073/pnas.071511098. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES