Abstract
Super-resolution optical microscopy has attracted great interest among researchers in many fields, especially in biology where the scale of physical structures and molecular processes fall below the diffraction limit of resolution for light. As one of the emerging techniques, structured illumination microscopy can double the resolution by shifting unresolvable spatial frequencies into the pass-band of the microscope through spatial frequency mixing with a wide-field structured illumination pattern. However, such a wide-field scheme typically can only image optically thin samples and is incompatible with multiphoton processes such as two-photon fluorescence, which require point scanning with a focused laser beam. Here, we propose two new super-resolution schemes for laser scanning microscopy by generalizing the concept of a spatially nonuniform imaging system. One scheme, scanning patterned illumination (SPIN) microscopy, employs modulation of the excitation combined with temporally cumulative imaging by a nondescanned array detector. The other scheme, scanning patterned detection (SPADE) microscopy, utilizes detection modulation together with spatially cumulative imaging, in this case by a nondescanned single-element detector. When combined with multiphoton excitation, both schemes can image thick samples with three-dimensional optical sectioning and much improved resolution.
It has been well known since the late nineteenth century1 that the resolution of optical microscopes is limited by diffraction. This phenomenon is clearly explained by the theory of Fourier optics2,3 in which the resolution limit of a microscope is given by the extent of its optical transfer function (OTF), the normalized Fourier transform of the point spread function (PSF). The microscope essentially functions as a low-pass filter, rejecting spatial frequencies in the specimen that lie beyond the bandwidth of the OTF.
The length scale of many biological structures and molecular processes is, however, below the diffraction limit, so they cannot be resolved with conventional optical microscopes. This situation has motivated many attempts to break the diffraction limit and achieve super-resolution. Prior knowledge about the object (e.g., its finite size and nonnegativity) has been utilized to recover or extrapolate the object spectrum beyond the pass-band of the microscope.4–13 Nonlinear effects in fluorescence excitation have been exploited to confine fluorescence emission to a volume smaller than defined by the diffraction limit, as demonstrated in two-photon laser scanning microscopy (2PLSM).14–17 Utilizing saturated stimulated emission to spatially restrict the fluorescence PSF, stimulated emission depletion (STED) microscopy achieved a true far-field diffraction-unlimited spatial resolution for the first time.18–23 In particular, a record of 5.8 nm resolution has been recently reported with STED.22 The introduction of photoswitchable molecules enables activation and imaging of a random, sparse set of individual molecules, each of which can be localized to the centroid of its PSF with a precision much smaller than the diffraction limit. This feature has been utilized in recently developed photoactivated localization microscopy (PALM),24–28 fluorescence photoactivation localization microscopy (FPALM),29–31 and stochastic optical reconstruction microscopy (STORM).32–35 Finally, particularly relevant to the present study is the use of nonuniform illumination fields to “shift” out-of-band specimen frequencies into the pass-band of the microscope through spatial frequency mixing with the illumination field, as in structured illumination microscopy (SIM) for wide-field fluorescent imaging.36–42 One particular advantage of SIM over other super-resolution techniques is that SIM does not require special fluorescent probes or high-intensity radiation.
We will briefly review the principles of SIM in the background section. Here it suffices to point out that in current implementations of SIM, the illumination field is generated through interference of coherent light over the entire field of view, using either diffraction gratings41 or spatial light modulators.43 Such wide-field illumination has its limitations. It is incompatible with optical processes that require high intensity of excitation light (e.g., two-photon fluorescence), which is only practical when focused laser beams are scanned through the specimen. Furthermore, wide-field illumination limits imaging depth so typically only thin samples, such as cultured cells, can be effectively imaged.
In this paper, we propose and compare two new super-resolution microscopy schemes. These schemes generalize the concept of a spatially nonuniform imaging system to laser scanning microscopy with either spatial or temporal modulation. Both schemes can image thick samples with optical sectioning when two-photon excitation is used. In addition, they are compatible with nonfluorescent processes such as spontaneous Raman scattering. They achieve maximal resolution improvement by a factor of 2 over conventional wide-field microscopy.
Background
As incoherent imaging is exemplified by fluorescent microscopy, we will use the term “fluorophore” to refer to any incoherent point light source for brevity. Furthermore, we assume that fluorescence excitation and detection are not saturated, so the signal is linear with respect to the distribution of fluorophore concentration. For notational simplicity we write equations mostly for the one-dimensional (1D) system, and only explicitly write equations for the 2D system when significant differences arise between 1D and 2D cases. Without loss of generality, we assume the magnification of the microscope to be unity, and omit unessential constant factors in the equations.
Let hex(•) denote the excitation PSF, and hem(•) the emission PSF. Both PSFs are normalized so that the peak value is 1. Let Iex(r),r ∈ ℝ be the illumination field, the intensity distribution of excitation light, where r indicates position on the specimen plane (Figure 1). Let Iem(r) be the fluorescence emission field, the intensity of fluorescence emitted at point r on the specimen plane. When fluorophore excitation is not saturated, Iem is proportional to the illumination field and to the fluorophore concentration, Iem(r) = Iex(r)s(r).
Let Iim(x),x ∈ ℝ be the light intensity distribution on the image plane (Figure 1). We shall refer to Iim as the image field, which is given by
(1) |
An absorbing mask with transmittance M(x) may be inserted at the image plane, and the signal that is transmitted is given by M(x)Iim(x) as linearity is assumed. To avoid confusion, we reserve the term “image” for the light distribution on the image plane, and refer to final images produced by the detector as “pictures” denoted by p(•).
We will now briefly review the principle of SIM with wide-field illumination (henceforth referred to as WF-SIM) in 1D. Detailed formulas for the 2D case have been given previously38,44 and will not be repeated here. In WF-SIM, the structured illumination field Iex is created through interference of coherent waves, and assumes for example the form
(2) |
where ωr is the spatial frequency of the illumination field in radians per unit length.
In Fourier space, this illumination field is represented as
(3) |
where k is the spatial frequency, δ(•) is the Dirac delta function, and tilde over a function indicates its Fourier transform. In particular, h̃ex and h̃em will denote the excitation and emission OTF, respectively.
Wide-field fluorescence is detected with a charge-coupled device (CCD) camera and gives
(4) |
where ⊗ denotes convolution.
With the structured illumination field given in eq 2, Fourier transform of eq 4 yields
(5) |
where we have used the fact δ(x − x0) ⊗ f(x) = f(x − x0).
Equation 5 shows that frequency components k ± ωr of s̃ are shifted by the structured illumination field into the pass-band given by h̃em. Now each p̃ (k) in the recorded picture is the linear combination of three terms, as shown in eq 5. In order to solve for each term, two additional equations are needed. These extra relations are given by shifting the illumination pattern with respect to the specimen as
(6) |
where ϕ ≡ ωr•r0 indicates displacement in the specimen plane. In Fourier space, such shifts contribute phase factors to the pictures
(7) |
where the sifting property of δ-function has been used, as well as the fact that the Fourier transform of f(x − a) is f̃(k)e−iak. Notice that the phase factor e±iϕ does not depend on the frequency k.
For each frequency k, s̃(k), s̃(k − ωr), and s̃(k + ωr) can be solved by combining eq 5 and 7, if the emission OTF h̃em and the shift φ are known. Out-of-band frequencies in s̃ can be then “shifted back.” Because Iex is diffraction-limited, the maximal spatial frequency it introduces is , the cutoff frequency of h̃ex. The final picture p is band-limited by , the cutoff frequency of h̃em. Therefore, the effective bandwidth of WF-SIM is , which gives approximately twice the resolution of conventional microscopy if the difference between excitation and emission wavelengths is ignored.
In the 2D system, applying the modulation pattern along a particular direction (not necessarily a coordinate axis) extends the bandwidth only along the corresponding direction in the (kx,ky) space. Thus in order to extend the bandwidth isotropically in the (kx,ky) space, it is in theory necessary to apply modulation patterns along an infinite number of directions. In practice, however, modulating along the x- and y-axis gives substantial, although nonisotropic, improvement in resolution,38 whereas using a set of three orientations of the modulation pattern separated by 2π/3 radians gives an almost isotropic improvement in resolution.37
Results. 1. Laser Scanning Microscopy with Modulated Scanning and Temporally Cumulative Imaging
The first scheme we propose is scanning patterned illumination (SPIN) microscopy. In this scheme (Figure 2), excitation intensity is modulated as scanning proceeds. Pictures are cumulatively built up with a nondescanned, spatially resolved detector such as a CCD camera located at the image plane. Similar detector arrangement has been utilized in multifocal multiphoton imaging.45,46
Assume that scanning proceeds at constant speed v ≡ 1, and we shall henceforth use t to refer to both time and the position of the scanning spot center at that particular time. The emission field Iem(r) is given by:
(8) |
where is the normalized peak intensity of the scanning Airy disk. is modulated over time as
(9) |
where ωt is the temporal modulation frequency, and the DC component serves to ensure that .
As scanning proceeds, the detector cumulatively builds up Iim(x,t), which is given in a form similar to eq 1. This temporal integration yields the picture
(10) |
where the order of integration has been exchanged, and the fact that PSFs are even functions is used.
In eq 10 the term plays the role of the structured illumination field Iex in WF-SIM as shown in eq 4. In other words, the spatially structured illumination field has been synthesized through modulation of the peak intensity of the scanning spot. This fact can be better appreciated by examining eq 10 in the Fourier space, which yields
(11) |
has the same form as Ĩex(k) in eq 3. Because of the sifting property of the δ-function, picks up the DC as well as the ± ωt components of h̃ex, which yields the following effective illumination field Ẽeff(k)
(12) |
Because h̃ex is bandwidth-limited with cutoff frequency , the temporal modulation frequency ωt of must be chosen to lie within the pass-band [ ]. The effect of subsequent convolution with s̃(k) and multiplication with h̃em(k) is completely analogous to the process described for WF-SIM and gives the relationship
(13) |
In order to solve for each s̃(k), the intensity modulation pattern should be shifted as described in eq 6 to produce additional pictures p̃±ϕ(k). The same algorithm as used in WF-SIM can then be applied to solve and back-shift s̃(k). Again, the maximal effective bandwidth is given by , just as in WF-SIM. The effect of SPIN is simulated on an image of a (very small) zebra with custom-written Matlab scripts assuming the following parameters: objective lens numerical aperture NA = 1.4, excitation wavelength λex= 488 nm, emission wavelength λem= 535 nm, and temporal modulation period Tt = 1/ωt = 300 nm, which is slightly larger than the resolution limit (~200 nm) to ensure sufficient signal-to-noise ratio. For simplicity we followed Frohn et al.38 and applied the modulation pattern only along the x- and y-axis. As mentioned before, more isotropic resolution improvement can be achieved by applying modulation patterns along a set of 3 orientations. The SPIN reconstructed picture is shown in Figure 4C. The “true” object (Figure 4A) and simulated wide-field fluorescence image (Figure 4B) are given for comparison.
The intensity modulation does not have to be sinusoidal. In two dimensions, a practically simpler implementation is to “jump scan”, that is, scan a few lines, then skip the next few lines, while the intensity on the scanned lines is kept constant. This makes a square wave along the direction perpendicular to the fast axis (say, the x-axis) of the scanning pattern. This square wave contains an infinite number of harmonics of the fundamental spatial frequency ωy along the y-axis. However, ωy can be chosen such that all higher harmonics are cutoff by the excitation OTF, and eq 12 remains valid with only changes in constant factors.
When choosing in two dimensions, one has to consider the available scanning mechanism. Modulation along the coordinate axes is compatible with both point-wise intensity modulation (e.g., through an acousto-optical modulator, AOM) and the “jump scan” discussed above. Modulation oblique to the coordinate axes can be implemented with AOM. Jump scan oblique to the coordinate axes, however, cannot be performed straightforwardly if the two scanning mirrors are not equivalent (e.g., one strictly fast and one slow).
When SPIN is combined with 2PLSM, the nonlinearity in 2-photon excitation yields the following relationship
(14) |
where is the peak intensity of the 2-photon scanning spot, and h2P is the 1-photon PSF at the infrared wavelength used in 2PLSM, which has cutoff frequency . Thus is the effective 2-photon PSF, which has cutoff frequency . If assumes the form as in eq 9, it gives , where the prefactor 1/2 in eq 9 is omitted. Thus the harmonic 2ωt is introduced. If ωt is larger than , this harmonic falls outside the pass-band of the effective 2-photon OTF. Otherwise, this harmonic contributes to p(x), which complicates subsequent manipulations and thus should be avoided. Alternatively, as the modulation pattern is created by controlling laser intensity and can take arbitrary functional forms, a nonsinusoidal pattern can be employed. For instance, as suggested by one reviewer, yields as desired.
2. Laser Scanning Microscopy with Modulated Detection and Spatially Cumulative Imaging
In the above scheme of SPIN, when is a constant, Iim(x,t) has the functional form
(15) |
As hex and hem are both even functions, there is clearly a symmetry between x and t in eq 15 if we ignore the difference between hex and hem. It suggests an alternative scheme for super-resolution. In this scheme (Figure 3), the peak intensity of the scanning spot is kept constant, and modulation is imposed by inserting a mask with transmittance M(x) =(1/2)[1 + cos(ωs•x)] on the image plane, where ωs is the spatial modulation frequency. We thus term this scheme scanning patterned detection (SPADE) microscopy. In SPADE, a single-element detector (e.g., a photomultiplier tube, PMT) with a large detection area is placed behind the mask. The detector sums up all transmitted signals through the mask and assigns the sum to the pixel corresponding to the current scanning position. This scheme produces the following picture
(16) |
where the order of integration has been exchanged, and the evenness of hex and hem is exploited.
Fourier transform of eq 16 yields:
(17) |
where the maximal frequency shift of s̃(k) is limited by the cutoff frequency of the emission OTF h̃em. The overall effective bandwidth, however, is still .
The similarity between eqs 16 and 10 is obvious. Formally, M(x) corresponds to , and the roles of hex and hem are swapped. In order to recover out-of-band frequencies in s̃(k), M(x) also needs to be shifted according to eq 6 to generate additional images. The same algorithmic manipulations are then applied to obtain super-resolved s̃(k). The effect of SPADE is simulated with the same parameters as for SPIN, with spatial modulation period Ts = 1/ωs = 300 nm. The SPADE reconstructed picture is shown in Figure 4D.
Because SPADE applies the modulation to the image plane, its mathematical formalism remains unchanged when it is combined with 2PLSM, except that the effective two-photon excitation PSF should be used.
3. Comparison between SPIN and SPADE
Both SPIN and SPADE depend critically on the fact that light intensity from different incoherent point sources is linearly summed. This property of incoherent imaging enables spatial structures to be first decomposed and encoded into a temporal sequence and subsequently synthesized from the sequence through summation.
The mathematical formalisms of SPIN and SPADE exhibit clear symmetry between eqs 10 and 16, which stems from the symmetry between x and t in eq 15. Mathematically, eqs 10 and 16 are weighed sums over t and x with the temporal and spatial modulation terms being the weights, respectively. Therefore, both SPIN and SPADE can be viewed as linear filtering approaches to the ill-posed problem of solving s(r) from the integral equation eq 15.
The symmetry between SPIN and SPADE also leads to some interesting alternative descriptions of image formation. As Heintzmann et al. pointed out,47 the emission field Iem is linear in Iex only when excitation is not saturated, but is always linear in fluorophore concentration s because fluorophores do not interact. Thus one may introduce the “emittability field” E(x) to represent the dependence of fluorescence emission on excitation intensity and describe image formation with a spatially resolved detector as
(18) |
For image formation in scanning microscopy with single-element detection, we may analogously define D(x) = hem(x) ⊗ M(x), where M(x) is the transmittance on the image plane. Equation 16 then assumes the form
(19) |
We call D(x) “detectability” to emphasize its analogous role to the emittability E(x). When spatial integration is used with scanning, the system behaves as if fluorescence from each point light source is detected with a different detectability, and the limited resolution in p(t) comes from the fact that signals from the entire illuminated spot are assigned to a single pixel.
4. Digital Implementation of SPIN and SPADE
It is in principle possible to implement the spatial or temporal integration utilized in SPIN and SPADE digitally rather than physically. In order to do so, individual pictures corresponding to Iim(x,t) must be taken for each scanning position t. Modulations and integrations are then carried out as weighed sums over these pictures. The sum may be performed either over different pictures, following the integration over t in eq 10, or over different pixels in each picture, following the integration over x in eq 16.
This digital implementation is essentially a linear filter approach to inversion of the pictures. It shares the spirit of some inversion formulas previously proposed.48,49 The difference is that previous approaches involve manipulating an infinite number of frequencies, whereas our schemes only use finite number of “carrier” frequencies.
Although physical integration is much faster in the image acquisition process, digital processing may carry one advantage. When the illumination pattern is physical, it cannot take negative values. The mask’s transmittance cannot be negative either, unless special detection configurations are utilized.50 Thus there is always a DC component in the “carrier wave”, corresponding to the h̃ex(0)δ(k) term in eq 12. Typically its amplitude is much larger than the harmonics at ±ωt, which shift the specimen frequencies, because the harmonics are attenuated by the OTF. This has the tendency to bury the harmonics in the noise. With digital implementation of these schemes, the weights can take any value, so the “modulation patterns” can be a single sinusoid wave without the DC offset.
Discussion
In this paper, we propose two new schemes of super-resolution scanning microscopy that improve the resolution by a factor of 2. The first scheme, scanning patterned illumination (SPIN) microscopy, synthesizes an effective structured illumination field through modulation of the peak intensity of the scanning spot and temporally cumulative imaging on a spatially resolved wide-field detector. The second scheme, scanning patterned detection (SPADE) microscopy, exploits the formal symmetry between x and t in the image field Iim as given in eq 15 and uses a mask at the image plane to accomplish the spatial frequency shift. Although we focused on one-photon fluorescence, both schemes can be combined with multiphoton excitation to image thick samples with optical sectioning, as well as nonfluorescent incoherent optical processes such as spontaneous Raman scattering. Neither SPIN nor SPADE per se requires special fluorophores or special light sources to generate extreme light intensities.
The classical resolution limit of optical microscopy holds only when the optical system is linear and space-invariant.51 Scanning microscopy, however, is not space-invariant, because illumination is no longer spatially uniform. Thus out-of-band frequencies in the specimen are actually shifted into the pass-band of the microscope through beating with frequencies in the scanning spot. If a spatially resolved detector is used to take a separate picture for each scanning spot, it is in principle possible to invert the pictures to achieve super-resolution.11,48,52,53 However, recovery of out-of-band specimen frequencies from such imaging schemes is complicated, because the scanning spot contains a continuous distribution of frequencies, rather than a few discrete ones as in WF-SIM. Taking one full image for each location of the scanning spot also severely limits the speed of imaging.
SPIN and SPADE avoid this problem in a way equivalent to heterodyne detection in the spatial frequency domain. They lock-in to a particular “carrier” frequency from the excitation or emission OTF. This “carrier” frequency shifts out-of-band specimen frequencies into the pass-band, which can then be recovered with exactly the same algorithm as in WF-SIM.
Acknowledgments
We thank Sijia Lu, Gene Wei Li, Brian Saar, and Christian Freudiger for helpful discussions. This work was supported by U.S. Department of Energy’s Basic Energy Sciences Program (DE-FG02-07ER15875) to X.S.X., and The Gatsby Charitable Foundation, the HHMI Collaborative Innovation Award, and the Microsoft Research to J.W.L.
References
- 1.Abbe E. Archiv Mikroskop Anat. 1873;9:413–468. [Google Scholar]
- 2.Goodman JW. Introduction to Fourier Optics. 3. Roberts & Company Publishers; Greenwood Village, CO: 2004. [Google Scholar]
- 3.Wilson T, Sheppard C. Theory and Practice of Scanning Optical Microscopy. Academic Press; London: 1984. [Google Scholar]
- 4.Harris JL. J Opt Soc Am. 1964;54(7):931–936. [Google Scholar]
- 5.Rushforth CK, Harris RW. J Opt Soc Am. 1968;58(4):539–545. [Google Scholar]
- 6.Frieden BR. J Opt Soc Am. 1972;62(4):511–518. doi: 10.1364/josa.62.000511. [DOI] [PubMed] [Google Scholar]
- 7.Gerchberg RW. Opt Acta. 1974;21(9):709–720. [Google Scholar]
- 8.Bertero M, Pike ER. Opt Acta. 1982;29(6):727–746. [Google Scholar]
- 9.Bertero M, Boccacci P, Pike ER. Opt Acta. 1982;29(12):1599–1611. [Google Scholar]
- 10.Bertero M, Brianzi P, Parker P, Pike ER. Opt Acta. 1984;31(2):181–201. [Google Scholar]
- 11.Bertero M, De Mol C, Pike ER, Walker JG. Opt Acta. 1984;31(8):923–946. [Google Scholar]
- 12.Conchello JA. J Opt Soc Am A. 1998;15(10):2609–2619. doi: 10.1364/josaa.15.002609. [DOI] [PubMed] [Google Scholar]
- 13.Hom EFY, Marchis F, Lee TK, Haase S, Agard DA, Sedat JW. J Opt Soc Am A. 2007;24(6):1580–1600. doi: 10.1364/josaa.24.001580. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Denk W, Strickler JH, Webb WW. Science. 1990;248(4951):73–6. doi: 10.1126/science.2321027. [DOI] [PubMed] [Google Scholar]
- 15.Zipfel WR, Williams RM, Webb WW. Nat Biotechnol. 2003;21(11):1369–77. doi: 10.1038/nbt899. [DOI] [PubMed] [Google Scholar]
- 16.Helmchen F, Denk W. Nat Methods. 2005;2(12):932–40. doi: 10.1038/nmeth818. [DOI] [PubMed] [Google Scholar]
- 17.Svoboda K, Yasuda R. Neuron. 2006;50(6):823–39. doi: 10.1016/j.neuron.2006.05.019. [DOI] [PubMed] [Google Scholar]
- 18.Klar TA, Engel E, Hell SW. Phys Rev E. 2001;64(6 Pt 2):066613. doi: 10.1103/PhysRevE.64.066613. [DOI] [PubMed] [Google Scholar]
- 19.Willig KI, Rizzoli SO, Westphal V, Jahn R, Hell SW. Nature. 2006;440(7086):935–9. doi: 10.1038/nature04592. [DOI] [PubMed] [Google Scholar]
- 20.Hell SW. Science. 2007;316(5828):1153–1158. doi: 10.1126/science.1137395. [DOI] [PubMed] [Google Scholar]
- 21.Westphal V, Rizzoli SO, Lauterbach MA, Kamin D, Jahn R, Hell SW. Science. 2008;320(5873):246–9. doi: 10.1126/science.1154228. [DOI] [PubMed] [Google Scholar]
- 22.Rittweger E, Han KY, Irvine SE, Eggeling C, Hell SW. Nat Photonics. 2009;3(3):144–147. [Google Scholar]
- 23.Eggeling C, Ringemann C, Medda R, Schwarzmann G, Sandhoff K, Polyakova S, Belov VN, Hein B, von Middendorff C, Schonle A, Hell SW. Nature. 2009;457(7233):1159–62. doi: 10.1038/nature07596. [DOI] [PubMed] [Google Scholar]
- 24.Betzig E, Patterson GH, Sougrat R, Lindwasser OW, Olenych S, Bonifacino JS, Davidson MW, Lippincott-Schwartz J, Hess HF. Science. 2006;313(5793):1642–1645. doi: 10.1126/science.1127344. [DOI] [PubMed] [Google Scholar]
- 25.Shroff H, Galbraith CG, Galbraith JA, White H, Gillette J, Olenych S, Davidson MW, Betzig E. Proc Natl Acad Sci USA. 2007;104(51):20308–13. doi: 10.1073/pnas.0710517105. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Shroff H, Galbraith CG, Galbraith JA, Betzig E. Nat Methods. 2008;5(5):417–23. doi: 10.1038/nmeth.1202. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Manley S, Gillette JM, Patterson GH, Shroff H, Hess HF, Betzig E, Lippincott-Schwartz J. Nat Methods. 2008;5(2):155–7. doi: 10.1038/nmeth.1176. [DOI] [PubMed] [Google Scholar]
- 28.Shtengel G, Galbraith JA, Galbraith CG, Lippincott-Schwartz J, Gillette JM, Manley S, Sougrat R, Waterman CM, Kanchanawong P, Davidson MW, Fetter RD, Hess HF. Proc Natl Acad Sci USA. 2009;106(9):3125–30. doi: 10.1073/pnas.0813131106. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Hess ST, Girirajan TP, Mason MD. Biophys J. 2006;91(11):4258–72. doi: 10.1529/biophysj.106.091116. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Juette MF, Gould TJ, Lessard MD, Mlodzianoski MJ, Nagpure BS, Bennett BT, Hess ST, Bewersdorf J. Nat Methods. 2008;5(6):527–9. doi: 10.1038/nmeth.1211. [DOI] [PubMed] [Google Scholar]
- 31.Gould TJ, Gunewardene MS, Gudheti MV, Verkhusha VV, Yin SR, Gosse JA, Hess ST. Nat Methods. 2008;5(12):1027–30. doi: 10.1038/nmeth.1271. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Rust MJ, Bates M, Zhuang XW. Nat Methods. 2006;3(10):793–795. doi: 10.1038/nmeth929. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Bates M, Huang B, Dempsey GT, Zhuang X. Science. 2007;317(5845):1749–53. doi: 10.1126/science.1146598. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Huang B, Jones SA, Brandenburg B, Zhuang X. Nat Methods. 2008;5(12):1047–52. doi: 10.1038/nmeth.1274. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Huang B, Wang W, Bates M, Zhuang X. Science. 2008;319(5864):810–3. doi: 10.1126/science.1153529. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Lukosz W. J Opt Soc Am. 1966;56(11):1463–1472. [Google Scholar]
- 37.Gustafsson MG. J Microsc. 2000;198(Pt 2):82–7. doi: 10.1046/j.1365-2818.2000.00710.x. [DOI] [PubMed] [Google Scholar]
- 38.Frohn JT, Knapp HF, Stemmer A. Proc Natl Acad Sci USA. 2000;97(13):7232–6. doi: 10.1073/pnas.130181797. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Schermelleh L, Carlton PM, Haase S, Shao L, Winoto L, Kner P, Burke B, Cardoso MC, Agard DA, Gustafsson MG, Leonhardt H, Sedat JW. Science. 2008;320(5881):1332–6. doi: 10.1126/science.1156947. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Kner P, Chhun BB, Griffis ER, Winoto L, Gustafsson MG. Nat Methods. 2009;6(5):339–42. doi: 10.1038/nmeth.1324. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Gustafsson MG, Shao L, Carlton PM, Wang CJ, Golubovskaya IN, Cande WZ, Agard DA, Sedat JW. Biophys J. 2008;94(12):4957–70. doi: 10.1529/biophysj.107.120345. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Shao L, Isaac B, Uzawa S, Agard DA, Sedat JW, Gustafsson MG. Biophys J. 2008;94(12):4971–83. doi: 10.1529/biophysj.107.120352. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Hirvonen LM, Wicker K, Mandula O, Heintzmann R. Eur Biophys J. 2009 doi: 10.1007/s00249-009-0501-6. [DOI] [PubMed] [Google Scholar]
- 44.Gustafsson MGL, Agard DA, Sedat JW. Doubling the lateral resolution of wide-field fluorescence microscopy using structured illumination. In: Conchello JA, Cogswell CJ, Wilson T, editors. Three-Dimensional and Multidimensional Microscopy: Image Acquisition Processing VII. 2000. pp. 141–150. [Google Scholar]
- 45.Bewersdorf J, Pick R, Hell SW. Opt Lett. 1998;23(9):655–7. doi: 10.1364/ol.23.000655. [DOI] [PubMed] [Google Scholar]
- 46.Niesner R, Andresen V, Neumann J, Spiecker H, Gunzer M. Biophys J. 2007;93(7):2519–29. doi: 10.1529/biophysj.106.102459. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Heintzmann R, Jovin TM, Cremer C. J Opt Soc Am A. 2002;19(8):1599–609. doi: 10.1364/josaa.19.001599. [DOI] [PubMed] [Google Scholar]
- 48.Defrise M, De Mol C. Inverse Prob. 1992;8:175–185. [Google Scholar]
- 49.So PT, Kwon HS, Dong CY. J Opt Soc Am A. 2001;18(11):2833–45. doi: 10.1364/josaa.18.002833. [DOI] [PubMed] [Google Scholar]
- 50.Walker JG, Pike ER, Davies RE, Young MR, Brakenhoff GJ, Bertero M. J Opt Soc Am A. 1993;10(1):59–64. [Google Scholar]
- 51.Lohmann AW, Paris DP. J Opt Soc Am. 1965;55(8):1007–1013. [Google Scholar]
- 52.Bertero M, Brianzi P, Pike ER. Inverse Prob. 1987;3:195–212. [Google Scholar]
- 53.Young MR, Davies RE, Pike ER, Walker JG, Bertero M. Europhys Lett. 1989;9(8):773–778. [Google Scholar]